VENTIlogic LS

VENTIlogic plus

Ventilation device

Description of device and instructions for use

for devices from serial number 20,000

loading

Related Manuals for Weinmann VENTIlogic LS

Summary of Contents for Weinmann VENTIlogic LS

  • Page 1
    VENTIlogic LS VENTIlogic plus Ventilation device Description of device and instructions for use for devices from serial number 20,000…
  • Page 2: Table Of Contents

    Contents Overview ….. 4 Clean the housing….70 Clean coarse dust filter/change fine Special markings on the device .

  • Page 3
    11.11Electromagnetic interference 11.12Electromagnetic interference immunity for life-supporting medical immunity for NON-LIFE-SUPPORTING electrical devices and medical medical electrical devices or electrical systems….123 medical electrical systems ..124 Warranty .
  • Page 4: Overview

    1. Overview English Therapy device, general 5 Power cord coupling 4 Handle 3 Control panel and displays 6 Compartment for replaceable battery available as an option 1 Device outlet port 7 Replaceable battery (optional) 16 Housing fan 8 Card reader 9 Serial ports 14 Device ID plate 13 Power connection…

  • Page 5
    8 Card reader 16 Housing fan Slot for a Weinmann memory card. Therapy data Protects the device from overheating. are stored on the memory card which the doctor can call up.
  • Page 6
    Therapy device with leakage ventilation 21 Sealing plug (2x) 17 Patient circuit 17 Patient circuit 22 Drying adapter 19 Pressure 18 Adapter 20 Creased tube measuring tube 24 Connection for O sensor 23 Exhalation system 17 Patient circuit 22 Drying adapter The air flows to the patient/ventilator interface Required to dry the patient circuit with the aid of through the patient circuit.
  • Page 7
    Therapy device, single patient circuit with patient valve 26 Connection for pressure 27 Test adapter 28 Connection for measuring tube (marked valve control tube blue) 25 Connection for O sensor 29 Single patient circuit 34 Connection for 30 Pressure- device outlet port measurement tube (marked blue)
  • Page 8
    30 Pressure-measurement tube (marked 33 Connection for patient blue) This is where the patient/ventilator interface is For measuring therapy pressure. connected. 31 Valve control tube 34 Connection for device outlet port For controlling (opening and closing) the patient This is where the patient circuit is connected to valve.
  • Page 9
    Therapy device, double patient circuit with patient valve (VENTIlogic LS only) 36 Connection for pressure- measurement tube 38 Connection for 37 Test adapter (marked blue) valve control tube 35 Connection for O sensor 40 Opening for 39 Device connection for…
  • Page 10
    40 Opening for exhaled air 44 Patient valve This is where the patient’s exhaled air is routed For routing the patient’s exhaled air out of the out of the device. patient circuit. 41 Double patient circuit 45 Connection for device inlet port Delivers respiratory air to the patient and from the This is where the patient circuit is connected to patient back to the device.
  • Page 11
    Accessories 49 Replaceable battery WM 27880 50 Carrying bag WM 27976 51 Bacteria filter WM 24148 52 Bacteria filter (only leakage ventilation) WM 27591 (only valve ventilation) 54 Protective bag 53 Set, O sensor WM 15732 WM 27106 49 Replaceable battery WM 27880 53 Set, O sensor WM 15732 Available as an accessory, for mobile power…
  • Page 12: Special Markings On The Device

    1.1 Special markings on the device Left-hand side Oxygen connection: maximum supply rate: 15 l/min at < 1000 hPa VENTIlogic LS: Opening for exhaled air when operated with double patient circuit with patient valve; do not seal opening or block in any other way.

  • Page 13
    Front VENTIlogic LS: Connection for patient’s exhaled air with double patient circuit with patient valve. VENTIlogic plus: Opening is not used with VENTIlogic plus. Jack: electrical connection for oxygen sensor; max. 100 mV DC Connection: pressure measuring tube (marked blue). Therapy pressure 0 — 50 hPa…
  • Page 14
    Servicing label: indicates when the next service is due Safety check label: (in Germany only) marks when the next safety check as per §6 of the German law relating to users of medical devices is required Device inlet port: inlet port for ambient air at room temperature Device ID plate (rear) Type BF application part Protection class II, protective insulation…
  • Page 15: Control Panel

    Control panel 2 Alarm acknowledgement 3 Operating keys 1 On/Off key 9 Program key IPAP PEEP I : E 8 LIAM (insufflation) 4 Dial 7 LED for power supply 6 Menu key 1 On/Off key 6 Menu key For switching the therapy device on and off. For switching from the default display to the menu and vice versa.

  • Page 16
    Default display during therapy 2 Status line 3 Alarm display 5 Battery charge status of internal 4 Active program battery 6 Battery charge status of 1 Active ventilation replaceable battery (optional) mode 7 Memory card symbol 13 Patient circuit 12 Ventilation parameters 8 Respiratory phase change display…
  • Page 17
    9 Access to patient menu 12 Ventilation parameters Use the key adjacent to this menu item to switch The relevant current ventilation parameters are to the patient menu and back to the default displayed depending on the active mode. display. 13 Patient circuit 10 Bar chart for pressure display The relevant text to suit the set patient circuit…
  • Page 18
    Battery not ready for use: – battery defective or – battery too cold or – battery too hot Battery not recognized as a Weinmann battery. Replace battery. Device in internal battery mode. Measured values are written to the SD card Overview…
  • Page 19
    Symbol Significance SD card is write-protected or defective. No data can be recorded. Alarm window: Low-priority alarm triggered Medium-priority alarm triggered High-priority alarm triggered Main window: Plateau signal switched on Plateau signal switched off Overview…
  • Page 20
    Abbreviations used in the display Symbol Significance Status line: S mode active ST mode active T mode active CPAP CPAP mode active PCV mode active PSV mode active aPCV aPCV mode active VENTIlogic VCV mode active ( LS only) VENTIlogic aVCV aVCV mode active ( LS only)
  • Page 21
    Symbol Significance Respiratory phase switch triggered — mandatory Trigger for inspiration blocked during expiration Ti/T Proportion of inspiration time in a respiratory cycle Tidal volume Tidal volume on inspiration Tidal volume on exhalation Inspiration time Exhalation time Mean oxygen concentration (21%) Shown in brackets: measuring cell not calibrated, perform oxygen calibration Oxygen saturation…
  • Page 22: Safety Information In The Instructions

    1.2 Safety information in the instructions for Safety information in these instructions for use is marked as follows: Warning! Warns of risk of injury and potential material damage. Caution! Warns of material damage and potentially false therapy results. Note: Contains useful tips. Overview…

  • Page 23: Description Of Device

    The device can be used in both static or mobile operation, both at home and in appropriate hospital departments. Note VENTIlogic LS is not a ventilator for intensive care purposes in accordance with ISO 80601-2-12. The device can be used for weaning off invasive ventilation and converting to mask ventilation.

  • Page 24: Owner/Operator And User Qualification

    Observe the legal requirements for operation and use (in Germany, the regulations governing owner/operators of medical devices apply in particular). Basic recommendation: get a person authorized by Weinmann to provide you with proper instruction about the handling, use and operation of this medical device.

  • Page 25
    In controlled modes T, PCV and VCV (VENTIlogic LS only) and in assisted-controlled modes ST, PSV, aVCV (VENTIlogic LS only) and aPCV, the doctor can set respiratory frequency in the range from 5 to 45 breaths per minute and inspiration time in the range from 15 % to 67 % of the respiratory period.
  • Page 26
    In S, ST, PSV, aPCV, aVCV (VENTIlogic LS only), SIMV, MPVp and MPVv modes, the doctor can select one of 8 trigger stages for inspiration and one of 14 trigger stages for exhalation (not with aPCV, aVCV, MPVp and MPVv).
  • Page 27
    The device has a remote alarm connection to support the monitoring of patient and device, especially when VENTIlogic LS is used for life-support ventilation. All high and medium- priority alarms, together with the No power supply alarm are passed to this connection.
  • Page 28
    The remote alarm connection can be used to connect the device to the VENTIremote alarm remote alarm case. In hospital, the device can be connected directly to the hospital’s own internal alarm system. 2.3.14 Recording therapy data Therapy data are stored in the device on a removable SD card. The VENTIviews PC software can be used to enable a doctor to evaluate the therapy data.
  • Page 29: Safety Instructions

    3. Safety instructions 3.1 Safety information Read these instructions for use carefully. They are a component of the device and must be available at all times. Use the device exclusively for the intended purpose described (see “2.1 Intended use” on page 23). For your own safety and the safety of your patients and in accordance with the requirements of Directive 93/42/EEC, please note the following.

  • Page 30
    – It is essential to ensure that alarm VT is activated for life-support ventilation (VENTIlogic LS only). Only if these conditions are met can a blockage (stenosis) be detected. – Set the VT alarm appropriately.
  • Page 31
    • Masks of third-party manufacture may only be used following authorization by the manufacturer, Weinmann. The success of therapy is put in jeopardy by the use of unauthorized masks or other types of patient/ventilator interface.
  • Page 32
    • An exhalation system must always be used with leakage ventilation, otherwise the CO concentration in the breathing mask and tube would rise to critical values and thus obstruct breathing. • Ventilation modes MPVv and MPVp for mouthpiece ventilation may only be used on patients with a stable independent respiratory drive.
  • Page 33
    Keep the replaceable battery away from fire or heat. – Only use and charge the replaceable battery with the system provided for it. – Only replace the replaceable battery with a genuine Weinmann replaceable battery. – Children may only use the replaceable battery under supervision.
  • Page 34
    Note: • When planning your time, be aware that at low or very high outdoor temperatures, battery running time is considerably reduced. 3.1.4 Oxygen supply Warning! • If oxygen is being supplied to the respiratory flow, smoking and naked flames are forbidden.
  • Page 35
    • Have servicing and maintenance work carried out only by the manufacturer, Weinmann, or by specialist staff expressly authorized by the manufacturer. • Have modifications to the device carried out only by the manufacturer, Weinmann, or by specialist staff expressly authorized by the manufacturer.
  • Page 36: Contraindications

    3.2 Contraindications The therapy device should not be used or should be used only with particular caution in the case of the following diseases. In the individual case, the decision about therapy is the responsibility of the doctor supervising treatment. •…

  • Page 37: Side Effects

    3.3 Side effects When using the therapy device, the following undesired side effects may occur in short- term or long-term use: • pressure points on the face from the breathing mask and the forehead cushion • reddening of facial skin •…

  • Page 38: Set Up Device

    The volume displayed may deviate from that actually delivered by up to 105 ml. – Use patient circuit WM 27181 from Weinmann to avoid this deviation in the display.

  • Page 39: Patient/Ventilator Interfaces

    Caution! Do not cover the device with blankets etc. The air inlet would be blocked and the device could overheat. This may lead to inadequate therapy and to damage to the device. 1. Connect the power cord to the power connector of the device.

  • Page 40
    It is essential to follow the instructions for use of the patient/ventilator interface and of the patient circuit. 4.3.2 Double patient circuit (VENTIlogic LS only) In addition to the ventilation tube which delivers air to the patient, the pressure- measurement tube and the valve control tube, the double patient circuit also has an exhalation tube which routes exhaled air back to the device and into the ambient air.
  • Page 41: Connect Leakage Ventilation

    The pressure-measurement tube is the same length as the ventilation tube and leads to the Y-connecting piece where the ventilation tube and the exhalation tube are brought together. 4. Connect the valve control tube (5) to the connection of the device marked The valve control tube leads straight from the patient valve to the connection on the device and is therefore shorter than the pressure measuring tube.

  • Page 42
    5. Note that maximum flow rate and the accuracy of dynamic pressure may deviate if you are not using Weinmann tubes. Information relating to a separate exhalation system Caution! Always use an exhalation system. Used air containing carbon dioxide escapes from the patient/ventilator interface (e.g.
  • Page 43: Connect Humidifier

    Follow the instructions for use for the exhalation system and for the patient/ventilator interface. Connect circuit for mouthpiece ventilation Proceed as outlined below to connect the leakage circuit for mouthpiece ventilation (WM 27651). Tube adapter 1. Plug the tube adapter supplied onto the ventilation outlet on the device.

  • Page 44
    Note: The bacteria filter may not be operated on the device for more than 24 hours. Follow the instructions relating to period of use in “6. Hygiene treatment” on page 67. If the therapy device is intended for use by several patients (e.g. in a hospital), a bacteria filter must be used to prevent infections.
  • Page 45: Therapy With Oxygen Supply

    Combination with an oxygen sensor and a humidifier 1. Connect the oxygen sensor directly to the therapy device. 2. Connect the bacteria filter to the outlet of the oxygen sensor. 3. Connect the humidifier to the outlet of the bacteria filter. 4.

  • Page 46
    4. Switch off the device. The safety valve for the oxygen supply automatically shuts off the oxygen supply after 1 minute. Oxygen can be supplied via an oxygen concentrator (e.g. Weinmann Oxymat 3), via the central gas supply system (only with corresponding pressure reducer) of a hospital, in the form of liquid oxygen with a continuous flow or of an oxygen cylinder with a corresponding pressure reducer.
  • Page 47
    2. Plug the oxygen sensor (1) and air management adapter into the T-adapter (3). 3. Plug the T-adapter (3) onto the device outlet port (5). 4. Connect the sensor (1) to the oxygen measuring jack (4) with the aid of the cable. 5.
  • Page 48: Operation In The Event Of A Power Failure

    Note Use only SpO sensors from Weinmann to measure oxygen saturation. 4.8 Operation in the event of a power failure If the power supply should ever fail, the internal battery of the therapy device automatically assumes supply of the device.

  • Page 49: Operation

    5. Operation 5.1 Controls 5.1.1 Function keys The following functions can be called up directly in ventilation mode by pressing the relevant key on the device. IPAP PEEP • LIAM (insufflation) (4) I : E • Acknowledge alarms (3) • Select a program (5) •…

  • Page 50
    5.1.2 Navigating with the dial The dial (1) is the central control of the therapy device. You can use the dial to select menu items, navigate within the menu windows and set values for individual menu items. To familiarize yourself with navigation using the dial, we recommend switching to Menu first.
  • Page 51: Start Up The Device

    Select night mode If you press the dial during therapy, you will activate night mode. The display then goes dark so that only the bar chart with the pressure display is visible. Therapy continues as nor- mal. The display switches back on if you press the dial again or any other key. The display switches back on automatically if an alarm situation arises.

  • Page 52
    4. To switch on the device, press the On/Off key On/Off key briefly. The device is now in ventilation mode. The patient menu can be accessed via the menu key. When Auto switch-on (only with leakage ventilation) is IPAP PEEP activated, you can also put on the patient/ventilator interface and switch on the therapy device by taking a I : E…
  • Page 53: Handling Batteries

    5.3 Handling batteries The device is equipped with an internal battery which supplies the therapy device with power in an emergency. The therapy device can also be equipped with a replaceable battery available as an accessory. 5.3.1 Charging batteries The batteries are charged automatically as soon as the therapy device is connected to the electricity supply.

  • Page 54
    • Only remove the replaceable battery. The internal battery may only be replaced by the manufacturer, Weinmann, or an authorized specialist dealer. • Use only genuine Weinmann replaceable batteries. 1. Press down the latch of the replaceable battery and keep it depressed.
  • Page 55: Activate/Deactivate Auto Switch-On (Only Leakage Ventilation)

    3. Push the replaceable battery into the device until you hear the latch engage. When the device is switched on, the symbol for the replaceable battery appears in the status line and a beep sounds. 4. Use the status line and the Battery menu to see the charge status of the replaceable battery.

  • Page 56: Alarm List

    5.5 Alarm list 5.5.1 Storage of alarms All alarm types listed in the tables “Physiological alarms”and “Technical alarms” are recorded in an alarm list with date, time and duration once the alarm threshold is reached. Up to 200 alarms can be stored. After that, the oldest alarm in each case is overwritten. To call up the alarm list, select the menu item Alarm list in the patient menu using the dial and confirm your selection by pressing the dial.

  • Page 57: Liam Info

    5.7 LIAM info Note A detailed explanation of the LIAM function can be found in the section entitled “5.10 LIAM (insufflation)” on page 58. 1. In the patient menu, use the dial to select the LIAM info menu item. 2. Confirm your selection by pressing the dial. You will find the following values and their residual running times under LIAM info: •…

  • Page 58: Overview

    5.8 Overview Under the menu item Displays > Summary, you can view the current settings and alarms of the respectively configured programs as well as the actual values. 1. In the Patient menu, use the dial to select the Summary menu item. 2.

  • Page 59
    LIAM can only be enabled by the doctor and only triggered during ventilation. The maneuver includes at least one LIAM stroke consisting of insufflation and subsequent exhalation. Pressure Duration in min/h (e.g. 20 min) Interval in sec/min/h (e.g. every 5 min) Cycles (e.g.
  • Page 60
    Pressure Signal which can be 1 sec. plateau phase switched off IPAP LIAM Time LIAM LIAM Initially, the pressure curve within the individual LIAM stroke is comparable with a normal ventilation stroke. When the IPAP pressure level is reached, however, pressure continues to rise in linear form to maximum pressure IPAP (IPAP + P ) and is maintained for…
  • Page 61: Select A Program

    5.10.3 Canceling LIAM LIAM can be interrupted at any time. Press the key to do this. Then LIAM is canceled and the device reverts to the preset ventilation mode. If LIAM is then to be carried out again, begin the process by pressing the key again.

  • Page 62: After Use

    5.12 After use 1. Switch the device to standby by keeping the On/Off key depressed for approx. 2 seconds until the blower switches off. The duration of the previous therapy appears in the display. The device then switches to standby. 2.

  • Page 63
    – ratio of spontaneous inspiration to total number of inspirations in % – ratio of spontaneous exhalation to total number of exhalations in % – tidal volume of last inspiration in ml – mean respiratory minute volume in ml/min – current physiological alarms –…
  • Page 64: Travel With The Therapy Device

    5.13 Travel with the therapy device Traveling by air with the VENTIlogic LS/plus: The terms of commercial transport do not currently apply to individuals who would like to travel with the therapy device (see “3.1.5 Transport/accessories/spare parts/maintenance”…

  • Page 65
    5.13.1 Bags for the therapy device The therapy device has two bags, a protective bag (WM 27106) and a carrying bag for mobile use (WM 27976). The protective bag WM 27106 is supplied and is for protecting the device but not for mobile operation. The carrying bag WM 27976 is available as an accessory and allows the device to be operated on a mobile basis.
  • Page 66
    6. Attach the tube of fabric and the ventilation tubes to the side of the therapy device using the hook-and- loop attachment provided. Tips for use with a replaceable battery • If you are using a replaceable battery, you can change it without having to remove the therapy device from the bag.
  • Page 67: Hygiene Treatment

    6. Hygiene treatment This product may contain disposable items. Disposable items are intended to be used only once. So use these items only once and do not reprocess them. Reprocessing disposable items may impair the functionality and safety of the product and lead to unforeseeable reactions as a result of ageing, embrittlement, wear, thermal load, the effects of chemical processes, etc.

  • Page 68: Clean Leakage Ventilation

    The patient circuits for single and double patient circuits with patient valve (double patient circuit only on VENTIlogic LS) are disposables and cannot be subjected to a hygiene treatment. Follow the instructions for use for the patient circuit in question.

  • Page 69
    6.2.2 Dry the patient circuit using the therapy device If water ever gets into the pressure measuring tube by accident, the leakage circuit must be dried with the aid of the therapy device. This function can only be activated in standby mode. Likewise press the On/Off key to switch the device to standby.
  • Page 70: Clean The Housing

    6.3 Clean the housing Warning! • Risk of electric shock. Switch the device off completely before cleaning (see “Switch device off completely” on page 62). • Ensure that no liquids get into the device. Never immerse the device in disinfectants or other liquids, otherwise damage to the device and thus a hazard to users and patients may result.

  • Page 71: Clean The Fan Filter

    6.5 Clean the fan filter The fan filter protects the housing fan from dirt. To clean the fan filter, proceed as follows: 1. Remove the fan filter according to the instructions in section “Change fan filter” on page 99. 2. Clean the fan filter with fresh running water until it is free of residue. 3.

  • Page 72
    6.8.2 Patient circuit (leakage ventilation) We recommend GIGASEPT FF as disinfectant. When using GIGASEPT FF, take the same steps as described under “6.2 Clean leakage ventilation”. Rinse all parts thoroughly in distilled water following disinfecting. Allow the parts to dry completely.
  • Page 73: Change In Patients

    If the device is to be used for another patient without a bacteria filter being used, it must be subjected to a hygiene treatment beforehand. This must be performed by the manufacturer, Weinmann, or by an authorized specialist dealer. The procedure for hygiene treatment is described in the service sheet and in the servicing and repair instructions for the therapy devices.

  • Page 74: Function Check

    • • • • • • supply If the values/functions quoted below are not met, send the device to your specialist dealer or to the manufacturer, Weinmann, for repair. * These modes are only available with VENTIlogic LS. Function check…

  • Page 75
    7.2.1 Check flow measurement and flow sensors/pressure sensors (leakage ventilation) Note: A function check of the flow sensors/pressure sensors can only be carried out in standby mode. 1. Plug the red drying adapter supplied into the device outlet port. 2. For hospital staff only: Ensure that the patient circuit set on the device is the same as the patient circuit actually in use.
  • Page 76
    3. The subsequent steps are described in “7.2.1 Check flow measurement and flow sensors/pressure sensors (leakage ventilation)” starting from step 2. Function check using a double patient circuit with patient valve (VENTIlogic LS only) 1. Connect the patient circuit to the device.
  • Page 77
    Ensure each time you switch on that two different acoustic signals sound one after the other and that the yellow and red LEDs come on at the same time. 2. Check the No power supply alarm (power supply failure alarm): Start up the therapy device.
  • Page 78: Calibrate Oxygen Sensor (Only Valve Ventilation)

    7.3 Calibrate oxygen sensor (only valve ventilation) 7.3.1 General If oxygen is supplied during therapy, oxygen concentration is measured at the device outlet port so as to ensure that the patient is always adequately supplied with oxygen. To ensure the accuracy of the measurement, calibration should be performed daily. Calibration is necessary in the case of –…

  • Page 79
    7.3.3 Perform calibration Proceed as follows for calibration. 1. Ensure that the oxygen sensor is connected. 2. Operate the device. 3. Press the O key or the menu key . The selection bar is on Calibrate O sensor. 4. Confirm your selection by pressing the dial. Remaining calibration time is displayed.
  • Page 80: Energy Supply

    If supply is not assumed by the battery without interruption, either this or the therapy device is defective. In this case have the device including its internal battery checked by an authorized specialist dealer or by Weinmann. 3. Check battery capacity (shown in the status line of the display).

  • Page 81: Troubleshooting

    8.1 Faults Caution! If faults occur which cannot be eliminated at once, contact the manufacturer, Weinmann, or your specialist dealer immediately to have the device repaired. Do not continue operating the device in order to prevent even greater damage. Fault/fault message…

  • Page 82
    Battery too cold permitted temperature range overtemperature Battery defective Have device repaired Non-approved battery in Use genuine Weinmann battery Replaceable battery not detected Using the WM 27998 battery with a firmware Update the firmware to version 2.9.0 or above version < 2.9.0…
  • Page 83: Alarms

    Fault/fault message Cause of fault Remedy Check SpO sensor and fingertip and clean if Nail varnish, dirty fingers need be signal weak Patient shock Check patient condition 8.2 Alarms A distinction is made between two kinds of alarm. • Physiological alarms are those alarms which affect the patient’s ventilation directly.

  • Page 84
    The default display appears again after the acoustic alarm has been acknowledged. The fault which has not yet been rectified continues to be displayed in the status line and the alarm LED flashes (or stays on) until the fault is rectified. If the fault is not rectified within 120 seconds of the acknowledgement, the acoustic alarm (buzzer) sounds again.
  • Page 85
    Display Alarm Cause of fault Remedy Filter dirty Clean/change filter Adjust headgear/headband so Patient/ventilator that the patient/ventilator interface leaking interface seals, possibly replace it Patient/ventilator Replace patient/ventilator Minimum respiratory interface defective interface volume undershot. High priority Have the settings checked by Settings implausible the doctor supervising treatment…
  • Page 86
    Display Alarm Cause of fault Remedy Check whether the oxygen flow rate prescribed by the doctor is set correctly at the Oxygen flow rate set too oxygen source. Have settings checked by the doctor Minimum oxygen supervising treatment if concentration 2 low appropriate undershot at device…
  • Page 87
    Volume is not exceeded. Low Various possible causes, reached. Have settings checked by the priority, after 10 e.g. reduction in lung (VENTIlogic LS only) doctor supervising treatment breaths, rises to impedance medium priority Filter dirty Clean/change filter Adjust headgear/headband so…
  • Page 88
    8.2.5 Technical alarms Caution! If faults occur which cannot be eliminated at once, contact the manufacturer, Weinmann, or your specialist dealer immediately to have the device repaired. Do not continue operating the device in order to prevent even greater damage. Display…
  • Page 89
    Display Alarm Cause of fault Remedy Internal battery Device defective defective High priority Have device repaired Battery defective Battery defective Have device repaired. Internal battery not An unapproved battery detected is being used Low priority Using the WM 27998 Update the firmware to battery with a firmware version 2.9.0 or above version <…
  • Page 90
    Display Alarm Cause of fault Remedy Fault in power supply Have device repaired. Use Change device High priority Device defective alternative ventilation option measurement Disconnected, defective Check oxygen sensor and Medium priority exhausted or defective replace if necessary sensor Check SpO sensor and, if sensor defective or need be, have it changed or…
  • Page 91
    Display Alarm Cause of fault Remedy Check power cord is firmly connected. If necessary, check the function of the socket by No power supply and Acoustic signal for at connecting a different device internal battery is least 120 seconds, no (e.g.
  • Page 92
    Display Alarm Cause of fault Remedy Check valve control tube for The valve control tube damage, replace patient between the device and circuit if necessary the patient valve is incorrectly connected Connect valve control tube Pressure Device defective Have device repaired permanently High priority Implausible ventilation…
  • Page 93
    Display Alarm Cause of fault Remedy IPC failure Change device High priority Device defective Have device repaired Device switches off. Sensor system failure Change device High priority Sensor system defective Have device repaired Device switches off. System monitoring failure Voltage monitoring Low priority Have device repaired failure…
  • Page 94
    8.2.6 Storage of alarms Once the alarm threshold is reached, all the alarm types listed in tables “8.2.4 Physiological alarms” and “8.2.5 Technical alarms” are recorded in an alarm list with date, time and duration. Up to 200 alarms can be stored. After that, the oldest alarm in each case is overwritten.
  • Page 95: Maintenance And Safety Checks

    9. Maintenance and safety checks 9.1 Intervals We recommend having servicing, safety checks and repair work carried out only by the manufacturer, Weinmann, or by a specialized dealer expressly authorized by the manufacturer. Check both filters regularly for dirt. • The coarse dust filter and the fan filer should be cleaned once a week and changed no later than every 6 months.

  • Page 96: Batteries

    9.1.3 Internal battery • Have the internal battery replaced by Weinmann or an authorized specialist dealer at least every 2 years. 9.1.4 Replaceable battery (if present) • Have the replaceable battery checked by Weinmann or an authorized specialist dealer after 2 years.

  • Page 97: Change Filter

    9.3 Change filter 9.3.1 Change coarse dust filter Use original filters from Weinmann only. Using third-party filters invalidates any claim under warranty and may result in restricted function and bioincompatibility. 1. Press on the latch of the filter compartment lid and take it off.

  • Page 98
    compartment lid into the housing until the latch engages. 9.3.2 Change fine filter The fine filter needs changing when it has gone dark, but in any event after no more than 1000 operating hours. In the latter case, the filter change indicator appears in the display.
  • Page 99
    2. The question Reset filter change? appears. Select YES with the dial and confirm the selection by pressing the dial. If you want to cancel the process, select NO with the dial and press the dial. The process is cancelled. If you select and confirm YES with the dial, the message Filter change reset!appears for approx.
  • Page 100: Change Pressure-Measurement Tube (Only Leakage Ventilation)

    [Medizinprodukte-Betreiberverordnung — Germany only] is 2 years. In addition, servicing must be carried out as a preventive measure by the manufacturer, Weinmann, or by a specialist dealer expressly authorized by the manufacturer, at the following intervals: • after every 8000 operating hours (servicing symbol appears in the display) •…

  • Page 101: Disposal

    – visual inspection for mechanical damage – filter change – cleaning the device – replacement of any defective parts – complete check of device functions and pressure displays – battery change – change of internal battery – final check in accordance with test instruction WM 27104 –…

  • Page 102
    9.6.4 SpO module Do not dispose of the SpO module with domestic waste. To dispose of the SpO module properly, contact a licensed, certified electronic scrap disposal merchant. This address is available from your Environment Officer or from your local authority. The device packaging (cardboard and inserts) can be disposed of in paper recycling facilities.
  • Page 103: Scope Of Supply

    – Hex socket wrench, size 3 WM 24708 Tube adapter (hospital adapter) WM 15880 Instructions for use, EN WM 67781 VENTIlogic LS, single patient circuit with patient valve WM 27950 Parts Order number VENTIlogic LS basic device with panel WM 27951…

  • Page 104
    – Hex socket wrench, size 3 WM 24708 Adapter, leakage ventilation, packed WM 27199 Instructions for use, EN WM 67781 VENTIlogic LS, double patient circuit with patient valve WM 27960 Parts Order number VENTIlogic LS basic basic device with panel WM 27951 Power cord…
  • Page 105
    Bacteria filter (valve ventilation) WM 27591 Tube adapter (hospital adapter) WM 15880 Instructions for use of VENTIlogic LS, EN for patients and caregivers WM 67781 Instructions for use of VENTIlogic LS, EN for medical personnel WM 67801 Quick reference, EN…
  • Page 106
    10.1.2 VENTIlogic plus VENTIlogic plus, leakage ventilation WM 27980 Parts Order number VENTIlogic plus basic device with panel WM 27991 Power cord WM 24177 Protective bag for VENTIlogic plus WM 27106 Set, conversion to leakage ventilation, packed, consisting of: WM 15545 –…
  • Page 107: Accessories And Spare Parts

    10.2 Accessories and spare parts You can order accessories and spare parts separately if required. A current list of accessories and spare parts can be ordered on the Internet at http://www.weinmann-medical.com/en/medical_professionals/products/ventilation/ or via your dealer. Scope of supply…

  • Page 108: Technical Data

    11. Technical data 11.1 Therapy device Therapy device Product class to Directive 93/42/EEC Dimensions W x H x D in cm 24 x 15.3 x 34 Weight approx. 5.9 kg excluding replaceable battery approx. 6.5 kg including replaceable battery Temperature range –…

  • Page 109
    Therapy device Duration of battery charging approx. 6 h process Classification to EN 60601-1 – type of protection against Protection class II electric shock – degree of protection against Type BF electric shock Degrees of protection – Against the ingress of solid particles –…
  • Page 110
    95 % rh (no condensation) and storage Flow at max. speed at Leakage ventilation: 0 hPa 350 l/min Single patient circuit with patient valve: 345 l/min Double patient circuit with patient valve (VENTIlogic LS only): 345 l/min Tolerance ±15 l/min Technical data…
  • Page 111
    Therapy device Flow at max. speed with bacteria Leakage ventilation: 320 l/min filter at 0 hPa: Single patient circuit with patient valve: 330 l/min Double patient circuit with patient valve (VENTIlogic LS only): 330 l/min Tolerance ±15 l/min Filtering and smoothing –…
  • Page 112: System Resistances

    11.2 System resistances System resistance at an air flow rate of 60 l/min at the patient connection opening Therapy device with tube system Therapy device VENTIlogic LS with WM 24130 (leakage with single patient double patient Accessories system) and circuit with patient…

  • Page 113: Oxygen Sensor

    Bacteria filter WM 24148 for leakage ventilation Increase in sound pressure level at a distance of 1 m from the device Max. 0.5 dB(A) in patient position as per EN ISO 17510 1.5 l max. permitted flow (flowing off freely) 300 l/min Internal volume of bacteria filter 85 ml…

  • Page 114: Spo Module

    11.5 SpO module module Dimensions W x H x D 67 x 66 x 28 mm Weight approx. 150 g Cable length up to finger clip sensor 2.5 m Display 0 to 100 % Pulse 0 to 300 beats per minute Temperature range –…

  • Page 115: Analog Box With Therapy Device

    11.6 Analog box with therapy device Channel Measured value Scaling VENTIlogic plus: Channel 1 Mask pressure 0 hPa 55 hPa, VENTIlogic LS: 55 hPa Channel 2 Flow -100 l/min +320 l/min Channel 3 Leakage flow 0 l/min +320 l/min Channel 4…

  • Page 116: Pneumatic Diagrams

    11.7 Pneumatic diagrams 11.7.1 Leakage ventilation Pressure sensor for atmosphere Pressure sensor for patient Patient Exhaled air flushed out Silentflow Pressure sensor via leakage system for device outlet port Flow Bacteria Filter Blower measuring Humidifier filter section insp. valve Flow sensor Non-return valve supply…

  • Page 117
    11.7.2 Single patient circuit with patient valve Pressure Patient sensor for atmosphere Pressure sensor for Patient Exhaled air patient valve Switching valves for exhalation valve Pressure sensor for Bacteria valve control filter Pressure sensor for Humidifier device outlet port Non-return Flow valve with Filter…
  • Page 118
    11.7.3 Double patient circuit with patient valve (VENTIlogic LS only) Pressure sensor for atmosphere Pressure sensor for patient Patient Switching valves for exhalation valve Pressure sensor for Bacteria valve control filter Pressure sensor for Humidifier device outlet port Non-return Flow…
  • Page 119: Safety Distances

    3.80 3.80 7.27 3.50 12.00 12.00 23.00 Further technical data are available from the manufacturer, Weinmann, on request or from the hospital manual and in the Servicing and Repair instructions. The right to make design modifications is reserved. Technical data…

  • Page 120: Emission Of Electromagnetic Interference

    The VENTIlogic LS/VENTIlogic plus is intended for operation in an ELECTROMAGNETIC ENVIRONMENT as outlined below. The customer or the user of the VENTIlogic LS/ VENTIlogic plus should ensure that it is operated in an environment of this type. Measurements of…

  • Page 121: Electromagnetic Interference Immunity

    IMMUNITY The VENTIlogic LS/VENTIlogic plus is intended for operation in the ELECTROMAGNETIC ENVIRONMENT as outlined below. The customer or the user of the VENTIlogic LS/ VENTIlogic plus should ensure that it is used in an environment of this type. INTERFERENCE…

  • Page 122
    IMMUNITY The VENTIlogic LS/VENTIlogic plus is intended for operation in the ELECTROMAGNETIC ENVIRONMENT as outlined below. The customer or the user of the VENTIlogic LS/ VENTIlogic plus should ensure that it is used in an environment of this type. Magnetic fields at power…
  • Page 123: Electromagnetic Interference Immunity For Life-Supporting Medical Electrical Devices And Medical Electrical Systems

    IMMUNITY The VENTIlogic LS/VENTIlogic plus is intended for operation in the ELECTROMAGNETIC ENVIRONMENT as outlined below. The customer or the user of the VENTIlogic LS/ VENTIlogic plus should ensure that it is used in an environment of this type. INTERFERENCE…

  • Page 124: Electromagnetic Interference Immunity For Non-Life-Supporting Medical Electrical Devices Or Medical Electrical Systems

    IMMUNITY The VENTIlogic LS/VENTIlogic plus is intended for operation in the ELECTROMAGNETIC ENVIRONMENT as outlined below. The customer or the user of the VENTIlogic LS/ VENTIlogic plus should ensure that it is used in an environment of this type. INTERFERENCE…

  • Page 125: Warranty

    12. Warranty Weinmann gives the customer a limited manufacturer warranty on new original Weinmann products and any replacement part fitted by Weinmann in accordance with the warranty conditions applicable to the product in question and in accordance with the warranty periods from date of purchase as listed below.

  • Page 128
    Weinmann Geräte für Medizin GmbH + Co. KG Kronsaalsweg 40 D-22525 Hamburg Germany T: +49-(0)40-5 47 02-0 F: +49-(0)40-5 47 02-461 E: info@weinmann-medical.com www.weinmann-medical.com…

Видео инструкция по работе с аппаратом ИВЛ VENTIlogic LS

Видео мануал по использованию аппарата ИВЛ VENTIlogic LS. Удобно, наглядно, понятно.


Серия видео-инструкций поможет разобраться в особенностях работы и настроек аппарата ИВЛ VENTIlogic LS, освоить функционал, разобраться в нюансах. Информация будет полезна и врачам, и пациентам — как новым, так и опытным пользователям.

1. Как начать вентиляцию с аппаратом VENTIlogic LS

2. Демонстрация работы вентилятора VENTIlogic LS

3. Дисплей и различные меню вентилятора VENTIlogic LS 

4. Как использовать вентилятор VENTIlogic LS с контуром с двумя линиями 

5. Как использовать вентилятор VENTIlogic LS с контуром с утечкой

6. Различные программы вентилятора и их активация

7. Как провести калибровку O2 на вентиляторе VENTIlogic LS

8. Как провести контроль работы вентилятора VENTIlogic LS

9. Как провести измерение SpO2 с помощью вентилятора VENTIlogic LS

С заботой о самом ценном, «Центр здорового сна»!


  • Contents

  • Table of Contents

  • Troubleshooting

  • Bookmarks

Available languages

  • EN

  • NL

Quick Links

VENTIlogic LS/VENTIlogic plus

VENTIlogic LS with leakage ventilation / met lekkagesysteem

VENTIlogic LS — single patient circuit with patient valve / systeem

met een slang met patiëntenventiel

VENTIlogic LS — double patient circuit with patient valve /

systeem met dubbele slang met patiëntenventiel

VENTIlogic plus with leakage ventilation / met lekkagesysteem

VENTIlogic plus — single patient circuit with patient valve /

systeem met een slang met patiëntenventiel

Instructions for use for devices up to serial number 9.999 and from

firmware version 2.07 upwards

Gebruiksaanwijzing voor apparaten tot serienummer 9.999 en vanaf

firmwareversie 2.07

loading

Summary of Contents for Weinmann VENTIlogic LS

Specifications:

868/868379-ventilogic_ls.pdf file (22 Apr 2023)

Accompanying Data:

Weinmann VENTIlogic LS Medical Equipment PDF User Manual Manual (Updated: Saturday 22nd of April 2023 09:50:22 AM)

Rating: 4.4 (rated by 27 users)

Compatible devices: MEDUMAT Easy CPR, WM 24900, SOMNOcomfort 2e, JOYCE SilkGel vented, COMBIBAG WM 11000, prismaPSG WM 100 MP, JOYCE Full Face vented 40 hPa, VENTImotion Series.

Recommended Documentation:

Text Version of User Manual Manual

(Ocr-Read Summary of Contents of some pages of the Weinmann VENTIlogic LS Document (Main Content), UPD: 22 April 2023)

  • 176, Weinmann VENTIlogic LS 176 Bediening NL Deze gegevens kunnen met behulp van WEINMANNsupport van de geheugenkaart uitgelezen en weergegeven worden. Of er een geheugenkaart in het apparaat aanwezig is en therapiegegevens geregistreerd worden, herkent u aan het symbool in de statusregel. Als het symbool niet verschijnt is de geheugenkaart…

  • 73, Weinmann VENTIlogic LS Function check 73 EN 6. If the function check is completed successfully, the message Function check ok! appears. If the function check is not successful, the message Device system implausible appears. In this case refer to section “8.1 Faults” on page 79. 7. Press the menu key to return to the standard displ…

  • 14, 14 Overview EN Front 3 VENTIlogic LS: Connection for patient’s exhaled air with double patient circuit with patient valve. VENTIlogic plus: Opening is not used with VENTIlogic plus. 4 Jack: electrical connection for VENTIclick respiratory air humidifier (only leakage ventilation), max. current consumptio…

  • 172, Weinmann VENTIlogic LS 172 Bediening NL Het betreffende drukverloop binnen de afzonderlijke LIAM-slag is aan het begin vergelijkbaar met een normale beademingsslag. Bij het bereiken van het IPAP-drukniveau stijgt de druk echter lineair verder tot op de maximale druk IPAP max (IPAP + ΔP LIAM ) en wordt gedurende een se…

  • 126, 126 Overzicht NL 16 Onderhoudssticker: Geeft aan wanneer het volgende onderhoud vereist is 17 STK-sticker: (alleen in de Bondsrepubliek Duitsland) Geeft aan wanneer de volgende veiligheidstechnische controle overeenkomstig §6 van de „Medizinprodukte-Betreiberverordnung“ (verordening voor de exploitant va…

  • 152, 152 Opstelling van het apparaat NL 4.4.1 Lekkageadapter met standaardconus aansluiten Voor de lekkagewerking is optioneel een adapter verkrijgbaar om slangen met moffen met een binnendiameter van Ø 22mm en een drukmeetslang met een binnendiameter ≤ Ø 5 mm te gebruiken. Wanneer het apparaat met deze…

  • 155, Opstelling van het apparaat 155 NL Wanneer bovendien een bevochtiger en/of een zuurstofsensor aangesloten moet worden, moet een andere volgorde worden aangehouden. Combinatie met een bevochtiger: 1. Sluit het bacteriënfilter direct aan op het therapieapparaat. 2. Sluit de bevochtiger aan op de uitgang …

  • 226, 226 Technische gegevens NL 11.5.2 Systeem met een slang met patiëntenventiel Filter Gebläse Insp. Flow- messstrecke O 2 -Ventil Rückschlag- ventil O 2 -Sensor Patienten- ventil Patient Drucksensor Patient Drucksensor Geräteausgang Flowsensor Drucksensor Atmosphäre O 2 -Einspeisung 0–15 l/min bei < …

  • Weinmann VENTIlogic LS User Manual

  • Weinmann VENTIlogic LS User Guide

  • Weinmann VENTIlogic LS PDF Manual

  • Weinmann VENTIlogic LS Owner’s Manuals

Recommended: AD29112, PCI100MMSC, CVA 4070 EN-CA

  • Alaris Medical Systems 8100 Series

    01202341567899ÿ66    4!»#$%!&'(ÿ)&»ÿ1(#*+,-ÿ/01+234ÿ5677ÿ839:3;<=<>?8/31:@A2ÿ8B;C3,;4 …

    8100 Series 94

  • Stryker BERTEC GOBED PLUS FL20E

    TECHNICAL ASSISTANCE AND PARTS 1 800 428-5025 (Canada) E-mail (Canada only): [email protected] 1 800 327-0770 (United States) Manufactured by Stryker Bertec Medical Inc 72-0439 R1.3 September 2003 F15-43-E-A Printed in Canada OPERATIONS MANUAL ELECTRIC ACUTE CARE BED Model FL20E (US …

    BERTEC GOBED PLUS FL20E 22

  • CAIRE Companion 5

    Companion 5TMUser Manual (English)Manuel d’utilisation (Français)Bedienungsanleitung (Deutsch)Manual del usuario (Español)Manuale dell’utente (Italiano)Manual do utilizador (Português)Gebruikershandleiding (Nederlands)Brugervejledning (Dansk)Bruksanvisning (Norsk)Bruksanvisning (Svenska)Käyttöop …

    Companion 5 276

  • Microlife BP A3 PLUS

    Microlife BP A3 PlusIB BP A3 Plus S-V10-1 4520Revision Date: 2020-10-12 Microlife AGEspenstrasse 1399443 Widnau / Switzerlandwww.microlife.comEN  1ES  8FR  16IT  24DE  32TR  40PT  48NL  56GR  64AR  72 …

    BP A3 PLUS 81

Additional Information:
Popular Right Now:

Operating Impressions, Questions and Answers:

[Page 1] Weinmann VENTIlogic LS

Ventilation device Description of device and instructions for use for devices from serial number 20,000 VENTIlogic LS VENTIlogic plus

[Page 2] Weinmann VENTIlogic LS

Contents 2 Contents EN 1. Overview . . . . . . . . . . . . . . . . . . . . . 4 1.1 Special markings on the device. . . . .12 1.2 Safety information in the instructions for use . . . . . . . . . . . . . . . . . . . . . . .22 2. Description of de…

[Page 3] Weinmann VENTIlogic LS

3 EN 11.11Electromagnetic interference immunity for life-supporting medical electrical devices and medical electrical systems. . . . . . . . . . . . . . 123 11.12Electromagnetic interference immunity for NON-LIFE-SUPPORTING medical electric…

[Page 4] Weinmann VENTIlogic LS

4 Overview EN English 1. Overview Therapy device, general 5 Power cord 6 Compartment for replaceable battery available as an option 3 Control panel and displays 4 Handle 1 Device outlet port 16 Housing fan 8 Card reader 9 Serial ports 13 Power…

[Page 5] Weinmann VENTIlogic LS

Overview 5 EN Key 1 Device outlet port The respiratory air flows to the patient from here via the patient circuit and the patient/ventilator interface. 2 O 2 coupling Serves as an adapter for connecting the oxygen source to the therapy device…

[Page 6] Weinmann VENTIlogic LS

6 Overview EN Therapy device with leakage ventilation Key 17 Patient circuit The air flows to the patient/ventilator interface through the patient circuit. The patient circuit consists of creased hose, pressure measuring tube and adapter. 18 Ada…

[Page 7] Weinmann VENTIlogic LS

Overview 7 EN Therapy device, single patient circuit with patient valve Key 25 Connection for O 2 sensor For connecting an oxygen sensor which can be used to measure oxygen concentration in respiratory air (only with patient circuits with a p…

[Page 8] Weinmann VENTIlogic LS

8 Overview EN 30 Pressure-measurement tube (marked blue) For measuring therapy pressure. 31 Valve control tube For controlling (opening and closing) the patient valve. 32 Patient valve For routing the patient’s exhaled air out of the patien…

[Page 9] Weinmann VENTIlogic LS

Overview 9 EN Therapy device, double patient circuit with patient valve (VENTIlogic LS only) 35 Connection for O 2 sensor For connecting an oxygen sensor which can be used to measure oxygen concentration in respiratory air (only with patient c…

[Page 10] Weinmann VENTIlogic LS

10 Overview EN 40 Opening for exhaled air This is where the patient’s exhaled air is routed out of the device. 41 Double patient circuit Delivers respiratory air to the patient and from the patient back to the device. 42 Connection for patie…

[Page 11] Weinmann VENTIlogic LS

Overview 11 EN Accessories Key 49 Replaceable battery WM 27880 Available as an accessory, for mobile power supply to the therapy device. 50 Carrying bag WM 27976 For mobile use of the therapy device. 51 Bacteria filter WM 24148 (only leakage vent…

[Page 12] Weinmann VENTIlogic LS

12 Overview EN 1.1 Special markings on the device Left-hand side 1 Oxygen connection: maximum supply rate: 15 l/min at < 1000 hPa 2 VENTIlogic LS: Opening for exhaled air when operated with double patient circuit with patient valve; do not sea…

[Page 13] Weinmann VENTIlogic LS

Overview 13 EN Front 3 VENTIlogic LS: Connection for patient’s exhaled air with double patient circuit with patient valve. VENTIlogic plus: Opening is not used with VENTIlogic plus. 4 Jack: electrical connection for oxygen sensor; max. 100 mV …

[Page 14] Weinmann VENTIlogic LS

14 Overview EN 16 Servicing label: indicates when the next service is due 17 Safety check label: (in Germany only) marks when the next safety check as per §6 of the German law relating to users of medical devices is required 18 Device inlet port:…

[Page 15] Weinmann VENTIlogic LS

Overview 15 EN Control panel Key 1 On/Off key For switching the therapy device on and off. 2 Alarm acknowledgement key The alarm acknowledgement key is for the temporary muting of alarms. The LED displays alarms visually. 3 Operating keys For …

[Page 16] Weinmann VENTIlogic LS

16 Overview EN Default display during therapy Key 1 Active ventilation mode The active ventilation mode is displayed at this point in the status line. 2 Status line This is where device status information (such as alarm state display, filter ch…

[Page 17] Weinmann VENTIlogic LS

Overview 17 EN 9 Access to patient menu Use the key adjacent to this menu item to switch to the patient menu and back to the default display. 10 Bar chart for pressure display For the graphical display of therapy pressure. 11 Oxygen concentratio…

[Page 18] Weinmann VENTIlogic LS

18 Overview EN Symbols used in the display Symbol Significance Status line: Filter change required Servicing required Acoustic signal of alarms muted for 120 seconds All physiological alarms deactivated (exception for VENTIlogic LS: In VCV and aV…

[Page 19] Weinmann VENTIlogic LS

Overview 19 EN SD card is write-protected or defective. No data can be recorded. Alarm window: Low-priority alarm triggered Medium-priority alarm triggered High-priority alarm triggered Main window: Plateau signal switched on Plateau signal switc…

[Page 20] Weinmann VENTIlogic LS

20 Overview EN Abbreviations used in the display Symbol Significance Status line: S S mode active ST ST mode active T T mode active CPAP CPAP mode active PCV PCV mode active PSV PSV mode active aPCV aPCV mode active VCV VCV mode active (VENTIlogic …

[Page 21] Weinmann VENTIlogic LS

Overview 21 EN Markings on the packaging T Respiratory phase switch triggered — mandatory B Trigger for inspiration blocked during expiration Ti/T Proportion of inspiration time in a respiratory cycle VT Tidal volume VTi Tidal volume on inspiratio…

[Page 22] Weinmann VENTIlogic LS

22 Overview EN 1.2 Safety information in the instructions for use Safety information in these instructions for use is marked as follows: Warning! Warns of risk of injury and potential material damage. Caution! Warns of material damage and potenti…

[Page 23] Weinmann VENTIlogic LS

Description of device 23 EN 2. Description of device 2.1 Intended use 2.1.1 VENTIlogic LS VENTIlogic LS is used for invasive and non-invasive life-support ventilation in accordance with ISO 10651-2 and for non-life-support mouthpiece ventila…

[Page 24] Weinmann VENTIlogic LS

24 Description of device EN • neurological, muscular and neuromuscular disorders, such as muscular dystrophies, pareses of the diaphragm etc. • central respiratory regulation disorders • hypoventilation syndrome associated with obesity 2.2 …

[Page 25] Weinmann VENTIlogic LS

Description of device 25 EN 2.3.3 Operating status Three operating states are possible on the therapy device: on, off and standby. If the device is switched on, therapy is in progress. In standby, the fan is switched off but the device is ready fo…

[Page 26] Weinmann VENTIlogic LS

26 Description of device EN In S, ST, PSV, aPCV, aVCV (VENTIlogic LS only), SIMV, MPVp and MPVv modes, the doctor can select one of 8 trigger stages for inspiration and one of 14 trigger stages for exhalation (not with aPCV, aVCV, MPVp and MPVv)….

[Page 27] Weinmann VENTIlogic LS

Description of device 27 EN The MPV modes allow breathing as required and are available for leakage ventilation, single patient circuit ventilation and double patient circuit ventilation systems. The MPV modes have no background frequency. A vent…

[Page 28] Weinmann VENTIlogic LS

28 Description of device EN The remote alarm connection can be used to connect the device to the VENTIremote alarm remote alarm case. In hospital, the device can be connected directly to the hospital’s own internal alarm system. 2.3.14 Recor…

[Page 29] Weinmann VENTIlogic LS

Safety instructions 29 EN 3. Safety instructions 3.1 Safety information Read these instructions for use carefully. They are a component of the device and must be available at all times. Use the device exclusively for the intended purpose describ…

[Page 30] Weinmann VENTIlogic LS

30 Safety instructions EN Danger! The alarm will not work if the wrong settings are used! If the VT low alarm has been deactivated, or incorrect settings have been used, then the alarm will not be triggered. If the patient is dependent on the v…

[Page 31] Weinmann VENTIlogic LS

Safety instructions 31 EN patient circuit or, if a single patient circuit is used, that exhaled volume is monitored separately. 3.1.2 Operating the device Warning! • Do not cover the device with blankets etc. The air inlet would be blocked and …

[Page 32] Weinmann VENTIlogic LS

32 Safety instructions EN • An exhalation system must always be used with leakage ventilation, otherwise the CO 2 concentration in the breathing mask and tube would rise to critical values and thus obstruct breathing. • Ventilation modes MPV…

[Page 33] Weinmann VENTIlogic LS

Safety instructions 33 EN resistance of the ventilation system must not exceed 6 hPa at 60 l/min for adults and 6 hPa at 30 l/min for children. 3.1.3 Mobile operation Warning! Risk of injury as a result of handling the replaceable battery incorre…

[Page 34] Weinmann VENTIlogic LS

34 Safety instructions EN Note: • When planning your time, be aware that at low or very high outdoor temperatures, battery running time is considerably reduced. 3.1.4 Oxygen supply Warning! • If oxygen is being supplied to the respiratory flow…

[Page 35] Weinmann VENTIlogic LS

Safety instructions 35 EN interruption to the power supply when the device is switched on and the internal battery is present. • If the therapy device and the batteries have been stored or transported at operating temperatures outside those qu…

[Page 36] Weinmann VENTIlogic LS

36 Safety instructions EN 3.2 Contraindications The therapy device should not be used or should be used only with particular caution in the case of the following diseases. In the individual case, the decision about therapy is the responsibility …

[Page 37] Weinmann VENTIlogic LS

Safety instructions 37 EN 3.3 Side effects When using the therapy device, the following undesired side effects may occur in short- term or long-term use: • pressure points on the face from the breathing mask and the forehead cushion • reddenin…

[Page 38] Weinmann VENTIlogic LS

38 Set up device EN 4. Set up device Warning! Device failure if incorrect patient circuits used! If patient circuits with a diameter smaller than Ø 15 mm are used, the device may overheat. – Use only patient circuits with a diameter of Ø 15 m…

[Page 39] Weinmann VENTIlogic LS

Set up device 39 EN Caution! Do not cover the device with blankets etc. The air inlet would be blocked and the device could overheat. This may lead to inadequate therapy and to damage to the device. 1. Connect the power cord to the power connect…

[Page 40] Weinmann VENTIlogic LS

40 Set up device EN Caution! The patient valve may not be covered when the device is in operation, otherwise the exhaled air may not be routed away, obstructing breathing. Proceed as follows to connect the single patient circuit to the therapy dev…

[Page 41] Weinmann VENTIlogic LS

Set up device 41 EN The pressure-measurement tube is the same length as the ventilation tube and leads to the Y-connecting piece where the ventilation tube and the exhalation tube are brought together. 4. Connect the valve control tube (5) to t…

[Page 42] Weinmann VENTIlogic LS

42 Set up device EN 1. Push the ventilation tube onto the device outlet port of the therapy device. 2. Now connect the pressure-measurement tube to the connection on the device marked . 3. Connect the patient/ventilator interface, e.g. a mask. …

[Page 43] Weinmann VENTIlogic LS

Set up device 43 EN Follow the instructions for use for the exhalation system and for the patient/ventilator interface. Connect circuit for mouthpiece ventilation Proceed as outlined below to connect the leakage circuit for mouthpiece ventilation…

[Page 44] Weinmann VENTIlogic LS

44 Set up device EN Note: The bacteria filter may not be operated on the device for more than 24 hours. Follow the instructions relating to period of use in “6. Hygiene treatment” on page 67. If the therapy device is intended for use by sever…

[Page 45] Weinmann VENTIlogic LS

Set up device 45 EN Combination with an oxygen sensor and a humidifier 1. Connect the oxygen sensor directly to the therapy device. 2. Connect the bacteria filter to the outlet of the oxygen sensor. 3. Connect the humidifier to the outlet of the b…

[Page 46] Weinmann VENTIlogic LS

46 Set up device EN Proceed as follows to supply oxygen: 1. Switch on the therapy device. 2. Connect the O 2 coupling supplied to the connector stub provided on the therapy device. 3. Connect the oxygen source to the O 2 coupling. 4. Start t…

[Page 47] Weinmann VENTIlogic LS

Set up device 47 EN 2. Plug the oxygen sensor (1) and air management adapter into the T-adapter (3). 3. Plug the T-adapter (3) onto the device outlet port (5). 4. Connect the sensor (1) to the oxygen measuring jack (4) with the aid…

[Page 48] Weinmann VENTIlogic LS

48 Set up device EN Note The SpO 2 module supports diagnosis and patient monitoring. The SpO 2 module may only be used for diagnosis in conjunction with other indications of disease and symptoms. No clinical assessment may be made solely on the b…

[Page 49] Weinmann VENTIlogic LS

Operation 49 EN 5. Operation 5.1 Controls 5.1.1 Function keys The following functions can be called up directly in ventilation mode by pressing the relevant key on the device. • LIAM (insufflation) (4) • Acknowledge alarms (3) • Select…

[Page 50] Weinmann VENTIlogic LS

50 Operation EN 5.1.2 Navigating with the dial The dial (1) is the central control of the therapy device. You can use the dial to select menu items, navigate within the menu windows and set values for individual menu items. To familiarize yours…

[Page 51] Weinmann VENTIlogic LS

Operation 51 EN Select night mode If you press the dial during therapy, you will activate night mode. The display then goes dark so that only the bar chart with the pressure display is visible. Therapy continues as nor- mal. The display switches b…

[Page 52] Weinmann VENTIlogic LS

52 Operation EN 4. To switch on the device, press the On/Off key briefly. The device is now in ventilation mode. The patient menu can be accessed via the menu key. When Auto switch-on (only with leakage ventilation) is activated, you can also…

[Page 53] Weinmann VENTIlogic LS

Operation 53 EN 5.3 Handling batteries The device is equipped with an internal battery which supplies the therapy device with power in an emergency. The therapy device can also be equipped with a replaceable battery available as an accessory. …

[Page 54] Weinmann VENTIlogic LS

54 Operation EN 5.3.3 Battery menu This menu gives you a summary of the state of any batteries present. In the patient menu, select the Battery menu item using the dial: • Internal battery: always present and supplies the device with power i…

[Page 55] Weinmann VENTIlogic LS

Operation 55 EN 3. Push the replaceable battery into the device until you hear the latch engage. When the device is switched on, the symbol for the replaceable battery appears in the status line and a beep sounds. 4. Use the status line and the…

[Page 56] Weinmann VENTIlogic LS

56 Operation EN 5.5 Alarm list 5.5.1 Storage of alarms All alarm types listed in the tables “Physiological alarms”and “Technical alarms” are recorded in an alarm list with date, time and duration once the alarm threshold is reached. Up t…

[Page 57] Weinmann VENTIlogic LS

Operation 57 EN 5.7 LIAM info Note A detailed explanation of the LIAM function can be found in the section entitled “5.10 LIAM (insufflation)” on page 58. 1. In the patient menu, use the dial to select the LIAM info menu item. 2. Confirm you…

[Page 58] Weinmann VENTIlogic LS

58 Operation EN 5.8 Overview Under the menu item Displays > Summary, you can view the current settings and alarms of the respectively configured programs as well as the actual values. 1. In the Patient menu, use the dial to select the Summar…

[Page 59] Weinmann VENTIlogic LS

Operation 59 EN LIAM can only be enabled by the doctor and only triggered during ventilation. The maneuver includes at least one LIAM stroke consisting of insufflation and subsequent exhalation. Your doctor uses the Duration parameter to specify …

[Page 60] Weinmann VENTIlogic LS

60 Operation EN Initially, the pressure curve within the individual LIAM stroke is comparable with a normal ventilation stroke. When the IPAP pressure level is reached, however, pressure continues to rise in linear form to maximum pressure IPAP m…

[Page 61] Weinmann VENTIlogic LS

Operation 61 EN 5.10.3 Canceling LIAM LIAM can be interrupted at any time. Press the key to do this. Then LIAM is canceled and the device reverts to the preset ventilation mode. If LIAM is then to be carried out again, begin the process by pres…

[Page 62] Weinmann VENTIlogic LS

62 Operation EN 5.12 After use 1. Switch the device to standby by keeping the On/Off key depressed for approx. 2 seconds until the blower switches off. The duration of the previous therapy appears in the display. The device then switches to sta…

[Page 63] Weinmann VENTIlogic LS

Operation 63 EN – ratio of spontaneous inspiration to total number of inspirations in % – ratio of spontaneous exhalation to total number of exhalations in % – tidal volume of last inspiration in ml – mean respiratory minute volume in ml/mi…

[Page 64] Weinmann VENTIlogic LS

64 Operation EN Proceed as follows to put the memory card back in. 1. Pull on the rubber cover to get at the slot for the memory card. 2. Push the memory card into the slot with the cut-off corner pointing upwards. 3. Briefly press on the card so …

[Page 65] Weinmann VENTIlogic LS

Operation 65 EN 5.13.1 Bags for the therapy device The therapy device has two bags, a protective bag (WM 27106) and a carrying bag for mobile use (WM 27976). The protective bag WM 27106 is supplied and is for protecting the device but not for mob…

[Page 66] Weinmann VENTIlogic LS

66 Operation EN 6. Attach the tube of fabric and the ventilation tubes to the side of the therapy device using the hook-and- loop attachment provided. Tips for use with a replaceable battery • If you are using a replaceable battery, you can cha…

[Page 67] Weinmann VENTIlogic LS

Hygiene treatment 67 EN 6. Hygiene treatment This product may contain disposable items. Disposable items are intended to be used only once. So use these items only once and do not reprocess them. Reprocessing disposable items may impair the func…

[Page 68] Weinmann VENTIlogic LS

68 Hygiene treatment EN 6.1.2 Valve ventilation The patient circuits for single and double patient circuits with patient valve (double patient circuit only on VENTIlogic LS) are disposables and cannot be subjected to a hygiene treatment. Follow t…

[Page 69] Weinmann VENTIlogic LS

Hygiene treatment 69 EN 6.2.2 Dry the patient circuit using the therapy device If water ever gets into the pressure measuring tube by accident, the leakage circuit must be dried with the aid of the therapy device. This function can only be activa…

[Page 70] Weinmann VENTIlogic LS

70 Hygiene treatment EN 6.3 Clean the housing Warning! • Risk of electric shock. Switch the device off completely before cleaning (see “Switch device off completely” on page 62). • Ensure that no liquids get into the device. Never immerse…

[Page 71] Weinmann VENTIlogic LS

Hygiene treatment 71 EN 6.5 Clean the fan filter The fan filter protects the housing fan from dirt. To clean the fan filter, proceed as follows: 1. Remove the fan filter according to the instructions in section “Change fan filter” on page 99…

[Page 72] Weinmann VENTIlogic LS

72 Hygiene treatment EN 6.8.2 Patient circuit (leakage ventilation) We recommend GIGASEPT FF as disinfectant. When using GIGASEPT FF, take the same steps as described under “6.2 Clean leakage ventilation”. Rinse all parts thoroughly in distill…

[Page 73] Weinmann VENTIlogic LS

Hygiene treatment 73 EN 6.9 Change in patients If the device is operated with a bacteria filter, observe the following. • Change bacteria filter WM 27591 or: • sterilize bacteria filter WM 24148 and change the particulate filter inside it. If …

[Page 74] Weinmann VENTIlogic LS

74 Function check EN 7. Function check 7.1 Intervals Perform a function check of the device monthly. One exception to this is the oxygen sensor. The oxygen sensor needs calibrating daily. We recommend checking the battery capacity before each us…

[Page 75] Weinmann VENTIlogic LS

Function check 75 EN 7.2.1 Check flow measurement and flow sensors/pressure sensors (leakage ventilation) Note: A function check of the flow sensors/pressure sensors can only be carried out in standby mode. 1. Plug the red drying adapter supplied…

[Page 76] Weinmann VENTIlogic LS

76 Function check EN 7.2.2 Check flow measurement and flow sensors/pressure sensors (leakage ventilation with leakage adapter for standard tapered connector) 1. Connect the patient circuit to the device. 2. Plug the test adapter supplied onto th…

[Page 77] Weinmann VENTIlogic LS

Function check 77 EN Ensure each time you switch on that two different acoustic signals sound one after the other and that the yellow and red LEDs come on at the same time. 2. Check the No power supply alarm (power supply failure alarm): Start up …

[Page 78] Weinmann VENTIlogic LS

78 Function check EN 7.3 Calibrate oxygen sensor (only valve ventilation) 7.3.1 General If oxygen is supplied during therapy, oxygen concentration is measured at the device outlet port so as to ensure that the patient is always adequately suppl…

[Page 79] Weinmann VENTIlogic LS

Function check 79 EN 7.3.3 Perform calibration Proceed as follows for calibration. 1. Ensure that the oxygen sensor is connected. 2. Operate the device. 3. Press the O 2 key or the menu key . The selection bar is on Calibrate O 2 sensor. 4. Con…

[Page 80] Weinmann VENTIlogic LS

80 Function check EN 7.4 Energy supply 7.4.1 Power supply Connect the device to the power supply. The power supply is working perfectly if the green power supply LED comes on permanently and the standby screen appears in the display. 7.4.2 Inter…

[Page 81] Weinmann VENTIlogic LS

Troubleshooting 81 EN 8. Troubleshooting 8.1 Faults Caution! If faults occur which cannot be eliminated at once, contact the manufacturer, Weinmann, or your specialist dealer immediately to have the device repaired. Do not continue operating th…

[Page 82] Weinmann VENTIlogic LS

82 Troubleshooting EN Calibrate O 2 sensor Initial contact with oxygen sensor Perform calibration of oxygen sensor Oxygen supply has not been turned off Last calibration performed over 24 hours ago Device was switched off completely beforehan…

[Page 83] Weinmann VENTIlogic LS

Troubleshooting 83 EN 8.2 Alarms A distinction is made between two kinds of alarm. • Physiological alarms are those alarms which affect the patient’s ventilation directly. • Technical alarms are those alarms which affect the configuratio…

[Page 84] Weinmann VENTIlogic LS

84 Troubleshooting EN The default display appears again after the acoustic alarm has been acknowledged. The fault which has not yet been rectified continues to be displayed in the status line and the alarm LED flashes (or stays on) until the fau…

[Page 85] Weinmann VENTIlogic LS

Troubleshooting 85 EN VT low Minimum respiratory volume undershot. High priority Filter dirty Clean/change filter Patient/ventilator interface leaking Adjust headgear/headband so that the patient/ventilator interface seals, possibly replace i…

[Page 86] Weinmann VENTIlogic LS

86 Troubleshooting EN O 2 low Minimum oxygen concentration undershot at device outlet port. Medium priority Oxygen flow rate set too low Check whether the oxygen flow rate prescribed by the doctor is set correctly at the oxygen source. Hav…

[Page 87] Weinmann VENTIlogic LS

Troubleshooting 87 EN Pulse high Patient’s maximum pulse rate exceeded Ventilation parameter settings (pressure, volume, frequency, I:E) not suitable Check ventilation parameter setting (pressure, volume, frequency, I:E) and adjust if n…

[Page 88] Weinmann VENTIlogic LS

88 Troubleshooting EN * These alarms are only activated 2 minutes after ventilation starts. 8.2.5 Technical alarms Caution! If faults occur which cannot be eliminated at once, contact the manufacturer, Weinmann, or your specialist dealer immediat…

[Page 89] Weinmann VENTIlogic LS

Troubleshooting 89 EN Internal battery defective High priority Device defective Have device repaired Battery defective Internal battery not detected Low priority Battery defective Have device repaired. An unapproved battery is being used Usin…

[Page 90] Weinmann VENTIlogic LS

90 Troubleshooting EN Fault in power supply Change device High priority Device defective Have device repaired. Use alternative ventilation option O 2 measurement defective Medium priority Disconnected, exhausted or defective sensor Check ox…

[Page 91] Weinmann VENTIlogic LS

Troubleshooting 91 EN Display gone out Acoustic signal for at least 120 seconds, no display. High priority No power supply and internal battery is discharged Check power cord is firmly connected. If necessary, check the function of the socket…

[Page 92] Weinmann VENTIlogic LS

92 Troubleshooting EN Pressure permanently low High priority The valve control tube between the device and the patient valve is incorrectly connected Check valve control tube for damage, replace patient circuit if necessary Connect valve con…

[Page 93] Weinmann VENTIlogic LS

Troubleshooting 93 EN IPC failure Change device High priority Device switches off. Device defective Have device repaired Sensor system failure Change device High priority Device switches off. Sensor system defective Have device repaired System mo…

[Page 94] Weinmann VENTIlogic LS

94 Troubleshooting EN 8.2.6 Storage of alarms Once the alarm threshold is reached, all the alarm types listed in tables “8.2.4 Physiological alarms” and “8.2.5 Technical alarms” are recorded in an alarm list with date, time and duration. …

[Page 95] Weinmann VENTIlogic LS

Maintenance and safety checks 95 EN 9. Maintenance and safety checks 9.1 Intervals We recommend having servicing, safety checks and repair work carried out only by the manufacturer, Weinmann, or by a specialized dealer expressly authorized by t…

[Page 96] Weinmann VENTIlogic LS

96 Maintenance and safety checks EN 9.1.3 Internal battery • Have the internal battery replaced by Weinmann or an authorized specialist dealer at least every 2 years. 9.1.4 Replaceable battery (if present) • Have the replaceable battery checke…

[Page 97] Weinmann VENTIlogic LS

Maintenance and safety checks 97 EN 9.2.2 Storage Internal battery If the device is to be operational at all times, we recommend leaving the device connected to the power supply even if it is not used for prolonged periods. This ensures that the …

[Page 98] Weinmann VENTIlogic LS

98 Maintenance and safety checks EN compartment lid into the housing until the latch engages. 9.3.2 Change fine filter The fine filter needs changing when it has gone dark, but in any event after no more than 1000 operating hours. In the latter c…

[Page 99] Weinmann VENTIlogic LS

Maintenance and safety checks 99 EN 2. The question Reset filter change? appears. Select YES with the dial and confirm the selection by pressing the dial. If you want to cancel the process, select NO with the dial and press the dial. The process…

[Page 100] Weinmann VENTIlogic LS

100 Maintenance and safety checks EN 9.4 Change pressure-measurement tube (only leakage ventilation) 1. Release the sleeve of the creased tube from the adapter. 2. Pull the pressure-measurement tube out of the creased tube. 3. Pull the pressure…