Page 1: Service Manual Europa B

TECNO-GAZ S.p.a.

EUROPA B XP AND EUROPA B XP 24L

TECHNICAL MANUAL

DXBT004 REV_16 THIS MANUAL MUST BE INTEGRATED WITH THE INFORMATION OF

THE USER MANUAL

Page 2: Service Manual Europa B

Page 3: Service Manual Europa B

0434

Council Directive 93/42/CEE

Page 4: Service Manual Europa B

INDEX 1

INDEX

• MAIN COMMANDS

• TECHNICAL SETUP

• CONTROL MENU

• CHECK-UP OF LEVEL PROBES

• CHECK AND CLEAN THE VACUUM PUMP

• TEST OF THE PERIPHERALS CONNECTED TO CBL606

• DIAGRAMS

o PNEUMATIC DIAGRAM o ELECTRIC DIAGRAM

• FUNCTIONAL COMPONENTS

• ERROR MESSAGES

• CYCLE DIAGRAMS

• EXPLODED DRAWING

Page 5: Service Manual Europa B

QUADRO COMANDI / MAIN COMMANDS 1

QUADRO COMANDI / MAIN COMMANDS

Page 6: Service Manual Europa B

TECHNICAL SETUP 1

TECHNICAL SETUP

In this setup you could do the following:

1. change the serial number of the unit 2. change the kind of printer 3. change temperature and pressure offsets 4. change the customer number 5. set the technical transmission

You could enter in this setup with the following procedure:

1. switch-on the unit with the main green button; 2. during the switch-on keep pushed B-START and push at least 5 times B-PUMP until you see on the

display the serial number of the unit; 3. release B-START and B-PUMP.

To go through the various page of the list above you have to use B-START. 1. change the serial number of the unit

B-DOOR selects the number of the serial number that you want to change; the number selected blinks; B-SELECT increment the number selected; B-PUMP decrement the number selected. To SAVE the new serial number keep pushed for at least 10 seconds the 3 button B-DOOR, B-SELECT and B-PUMP.

Page 7: Service Manual Europa B

TECHNICAL SETUP 2

2. change the kind of printer with B-SELECT or B-PUMP you could select the following:

• PRINTER 4050 Only for the unit with the internal printer. 4050 is the name of the internal printer. This is an old model and on the side of the printer you could find the label HTP-4050. The code of this printer is DNVA061. The code of the paper for this printer is DNVA063. The printer is like the one that you see on figure 1.

Figure 1

Page 8: Service Manual Europa B

TECHNICAL SETUP 3

• PRINTER 9050 Only for the unit with the internal printer. 9050 is the name of the internal printer. This is the new version of the model 4050 and on the printer you could find the label HTPN-9050. The code of this printer is DXGA061. The code of the paper for this printer is DNVA063. The printer is like the one that you see on figure 1. This setting is compatible with our new “Custom” printer: on this printer you could find the label CUSTOM PLUS-M2B. The code of this printer is DXBA361. The code of the paper for this printer is DNVA063. The printer is like the one that you see on figure 2.

Figure 2

Page 9: Service Manual Europa B

TECHNICAL SETUP 4

• EXTERNAL PRINTER (STAMP. ESTERNA) This setting is used to connect an external printer to the unit. The external printer is connected to the serial port of the unit. If the serial port is in the front panel then the connector of this port is round, otherwise is rectangular like the COM port on PC computer. The external printer must have the following parameters: a. BAUD RATE = 600 bps b. STOP BITS = 2 c. PARITY = NONE d. BITS PER CHR = 7 Typically we refer to the following model of external printer: MODEL DP1012 (you could find the printer model on the bottom side of the printer) In this model you see 4 buttons on the top. The first button is an empty square, the second is like a @, the third is an 1 and the fourth is a 0. You could enter in the setup by pressing together the button number 1 and the button number 3. Here you could advance by pressing the button number 2 and you could change the selected option by pressing the button number 1. Then you have to press together the buttons number 1 and the button number 2 to save and exit from setup. MODEL DP24S4N (you could find the printer model on the bottom side of the printer) In this model you have 2 buttons on the top (named PRINT and FEED) and 1 switch on the back. To enter in the setup push together the 2 buttons on the top and then switch on the printer by the switch on the back. Now use the PRINT button to change the selected option and the FEED button to select the next option. Exit from setup by switching off the printer with the switch on the back.

• NO PRINTER (STAMP. ASSENTE) Use this setting if you don’t have any of the above or if you want to record cycle data only with a PC.

This setting is automatically saved.

Page 10: Service Manual Europa B

TECHNICAL SETUP 5

3. change temperature and pressure offset To modify/save the offset of the pressure/temperature sensor (buttons that you need to use) please refer to the chapter “CONTROL MENU”.

4. change customer number

This page is used to change a global setting of the unit and it is only for manufacturer setup. Please don’t change this number if you don’t know his function. With the software release 2.56 or newer: if you have an Europa B xp with standard European cycles then ensure that this number is 10; if you have an Europa B xp 24 with standard European cycles then ensure that this number is 11; otherwise ask to the manufacturer. This setting is automatically saved.

5. set the technical transmission

On the LCD you see Tx TECNICO on the first line; in the second line, with B-SELECT and B-PUMP, you could select the following:

• PRESENTE The electronic board will send on the serial port of the unit status information and values of T1, T2 and P1 every second when the unit is in the standard user mode. You need a special software to acquire this information.

• ASSENTE The electronic board will send on the serial port the information that are normally printed during a cycle in a codified format. You need a special software to acquire this information.

Page 11: Service Manual Europa B

CONTROL MENU 1

CONTROL MENU

Attention: in this device it is not recommended to do the temperature probes calibration if the device is younger than 1 years old. This device has 2 temperature probes into the chamber (TEMP1 and TEMP2) and 1 pressure transducer (PRESSURE1). For the temperature probes calibration in this autoclave it is not necessary to work on the electronic board. NOTE: it is not possible to adjust the probe offset when the display shows “OPEN DOOR”. Instead it is possible to do that in every phase of sterilization cycle.

Page 12: Service Manual Europa B

CONTROL MENU 2

For temperature probes offset regulation it is necessary go into the control menu:

1. Push and hold B-DOOR 2. Push and release at least 5 time B-SELECT until the information showed on display change (you can

see for example the picture above) 3. Release B-DOOR

If in the second line from the bottom the display shows T1 then you are in the control menu (you can see for example the picture above) At this point with B-SELECT and B-PUMP is possible to increase and decrease the T1 offset; the offset value is showed on the last line of display. T1 is the temperature probe called TEMP1 in the list of functional components. It is possible to save the value set by keeping pushed B-DOOR and B-PUMP at the same time until you listen the acoustic signal. It is possible to switch between T2 and P1 with B-DOOR. T2 is the second temperature probe and it is called TEMP2 in the list of the functional components. P1 is the pressure transducer and it is called PRESSURE1 in the list of the functional components. If you want to exit from the menu you have to push and hold pushed B-DOOR.

Page 13: Service Manual Europa B

CHECK-UP OF LEVEL PROBES 1

CHECK-UP OF LEVEL PROBES

The level probes are two, one on the clean water tank (TANK1) that we will call LEVEL1 and the other one in the used water tank (TANK2) that we will call LEVEL2: LEVEL1 – this shows the “minimum” and the “maximum” level of the clean water tank LEVEL2 – this shows the maximum filling of the used water tank To check if the probes work correctly it’s necessary to remove the cover of the tank and, with the autoclave turned on, you have to flaw the craft on the auction: then, looking at the front panel you have to verify, that the signal of the level is showed. Clean the crafts before reassembling the cover of the tank. Acting on the craft of LEVEL1 they will be had to turn on the LED indicative following: LEVEL MIN and LEVEL MAX, respectively when the craft is found down at the auction on the probe of level and the summit. Acting on the craft of the LEVEL2 will be had to turn on the LED indicator WASTE LEVEL when the craft is found to the summit of the auction probe. For color wires connection of LEVEL1 please look at the picture in the next page.

Page 14: Service Manual Europa B

CHECK-UP OF LEVEL PROBES 2

BLACK

VIOLET

GREEN

WHITE

BROWN

ORANGE

Page 15: Service Manual Europa B

CHECK AND CLEAN THE VACUUM PUMP 1

CHECK AND CLEAN THE VACUUM PUMP

CPGM020

The failure vacuum to reach the fixed values compromise the sterilisation cycle. Its important the periodically check of the vacuum pump, to guarantee the correct operation of the autoclave. Don’t make maintenance on vacuum pump under warranty (1 year). To do maintenance proceed as follow: 1. Unscrew the fixing screw and take out the pump 2. Place the vacuum pump on a good and clean work table 3. Proceed to open the pump unscrewing the four screw of both head (eight screw at all)

4. Lift at the same time the two heads of the pump and separate them: the heads are jointed by a collector

inserted by pressure 5. Open the two heads by unscrewing the two internal screw (four screw at all)

It’s important don’t invert the two heads when reassembling them

Page 16: Service Manual Europa B

CHECK AND CLEAN THE VACUUM PUMP 2

6. The cleaning of the diaghpram must be done with non-chemical agent products (eg. Alchool) Verify the diaghpram condition by view It’s better to replace the seal and the two o’rings

7. Clean the metal head and the intermediate plate with the same procedure as the diaghpram

Remove completely the impurities to avoid any malfunction

Page 17: Service Manual Europa B

CHECK AND CLEAN THE VACUUM PUMP 3

8. Place the intermediate plate on the pump (everyone on the original position and direction) 9. Insert the seals into the appropriate seat (the seals are all of the same type)

10. Joint the metal heads with the appropriate collector and slowly place on the pump

11. Tighten the two central screws without forced it 12. Tighten the four external screws by forced it

Page 18: Service Manual Europa B

CHECK AND CLEAN THE VACUUM PUMP 4

13. Closed at the end the heads by tighten the two central screw, blocked it but not forced, to avoid to damage the intermediate plates: a high force can broken them (see detail photo)

14. Verify the operating of the pump, by connect it to a pressure transducer (or a manometer) and feeding for

five seconds (no more). The pump must reach a vacuum degree of -0.96 / -0.98 bar: a correct holding means the right operation of the pump In a first time there is lack of pressure of 0.02 / 0.04 bar, afterwards the pressure is aligned with the holding parameter. In case of a little but constantly lack of pressure, it’s necessary to repeat the cleaning and a new control procedure

15. Final assembling of the pump on the autoclave

Page 19: Service Manual Europa B

CHECK AND CLEAN THE VACUUM PUMP 5

Spare parts for vacuum pump:

CPGM027 — CPGM028

CPGM023

CPGM043

Page 20: Service Manual Europa B

CHECK AND CLEAN THE VACUUM PUMP 6

CPGM025

CPGM024

Page 21: Service Manual Europa B

TEST PERIFERICHE CONNESSE A CBL606 1

TEST OF THE PERIPHERALS CONNECTED TO CBL606

With the CBL606 it’s possible to power up one by one all the peripherals connected to this board without following the sterilization cycle. Attention: during the test, for example, if you power up the heating element then this will be always on until you don’t stop it. We suggest you to execute the test only with the door opened to avoid that the unit can go in pressure. In order to execute the test proceed as follows. Insert on the board CBL606 a JUMPER in J3 position. The J3 position in the figure below is between the 2 red lines.

Page 22: Service Manual Europa B

TEST PERIFERICHE CONNESSE A CBL606 2

Switch on the autoclave with main green button. And use the B-START until you read on the display the string USCITA. Then use B-PUMP in order to increase the number indicated on the display. To each number corresponds the activation of one peripheral connected to the board; You could find the number-peripheral correspondence in the list that follows:

1. HEAT1 2. not connected 3. EV7 4. EV5 5. EV2 6. it closes relay «the direction» of BLOCK1 on CBL606 7. BLOCK1 8. not connected 9. it opens relay «the direction» of BLOCK1 on CBL606 10. BLOCK1 11. EV4 12. PUMP1 13. PUMP2 14. FAN1 15. EV3 16. not connected 17. EV6 18. not connected 19. BUZZER on CBL606 20. not connected

Page 23: Service Manual Europa B

SCHEMI / DIAGRAMS

SCHEMI / DIAGRAMS

Page 24: Service Manual Europa B

SCHEMA PNEUMATICO / PNEUMATIC DIAGRAM 1

Page 25: Service Manual Europa B

SCHEMA ELETTRICO / ELECTRIC DIAGRAMS 1

Page 26: Service Manual Europa B

SCHEMA ELETTRICO / ELECTRIC DIAGRAMS 2

Page 27: Service Manual Europa B

FUNCTIONAL COMPONENTS 1

FUNCTIONAL COMPONENTS

Exploded drawing Name Description

53 CBL606B Power board (main electronic board)

8 CBL607 Display board (electronic board to control the display, the main commands and to read water levels)

83C EV2 Solenoid valve typically close (for connections 1/4 gas) 83A EV3 Solenoid valve typically close (for connections 1/4 gas)

81 EV4 Solenoid valve typically close (for connections 1/8 gas) 82 EV5 Solenoid valve typically open (for connections 1/4 gas) 84 EV6 3 way solenoid valve (for connections 1/4 gas)

83B EV7 Solenoid valve typically closed (for connections 1/4 gas) FILTER1 Brass filter FILTER2 Plastic filter

37 FILTER3 Bacteriological filter 89 PRESSURE1 Pressure sensor 87 VALVE1 Non return valve 76 VALVE2 Safety valve 5 MAINS Mains switch (green)

72 P230 Mains input 230 Vac 71A FUSE1 Fuse 12 A 71B FUSE2 Fuse 12 A

73 PEXT Mains auxiliary output (for deionizer connection) 54 FUSE3 Fuse 500 mA

58A TANK1 Clean water tank 58B TANK2 Waste water tank 62A LEVEL1 Level probe for the clean water tank 62B LEVEL2 Level probe for the waste water tank

66 COOLER1 Cooler 67 FAN1 Fan 64 PUMP1 Load pump 57 PUMP2 Vacuum pump 40 VESSEL1 Vessel

27 BLOCK1 Door electromagnet (used to advance the pin to block the handle)

39 SWITCH1 Door open/close switch (push button that is pushed when the user close the door)

29 SWITCH2 Door electromagnet switch (switch activated when the electromagnet pin advance into the handle)

43 HEAT1 Chamber heat 41A TEMP1 Temperature probe 1 41B TEMP2 Temperature probe 2

42 TEMP3 Resetting manual thermostat – heater element 75D OUT-1 Posterior tap to remove water from TANK2 75C OUT-2 Posterior vent output for TANK2 75A OUT-3 Posterior tap to remove water from TANK1 75B OUT-4 Posterior vent output for TANK1 34A OUT-A Frontal output to remove water from TANK2 34B IN-B Frontal input to charge clean water with PUMP1 on TANK1 34C IN-C Frontal input to manual charge clean water on TANK1

74 RS232 RS232 port (to connect the computer) 4 PANEL1 Front panel (see main commands) 6 PRINTER1 Printer

Page 28: Service Manual Europa B

ERROR MESSAGES 1

ERROR MESSAGES E200 The cycle has been started with the LEVEL MIN LED on. You could have this error at the beginning of the cycle. 1. add water into TANK1 2. check LEVEL1 3. check the electrical connections between LEVEL1 and CBL607 4. check or replace CBL607

E250 The cycle has been started with the LEVEL MAX LED off and the autoclave is set with water charge by demineralizer. You could have this error at the beginning of the cycle. 1. refill TANK1

• check the tap in the back of the autoclave (if the water clean system is connected then the tap must be open)

• check if the water is going into the water clean system

• check the installation of the water clean system 2. check if there is 230V on PEXT when the LED LEVEL_MAX is off (if there is no voltage then the problem

could be the board CBL606, FUSE3 or the connection between the 3 object above) 3. check LEVEL1 4. check the electrical connections between LEVEL1 and CBL607 5. check or replace the board CBL607

E300 The cycle has been started with the WASTE LEVEL LED on. You could have this error at the beginning of the cycle. 1. remove water from TANK2 2. check LEVEL2 3. check the electrical connections between LEVEL2 and CBL607

Page 29: Service Manual Europa B

ERROR MESSAGES 2

E401 The cycle has been started with the door open. You could have this error at the beginning of the cycle. Close the door and restart the cycle. If the door isn’t open then SWITCH1 doesn’t feel the handle of the door. 1. check SWITCH1 2. check the electrical connections between SWITCH1 and CBL606:

the problem could be a false contact on the cable connector for the electronic board 3. check the status of the cables 4. check the position of SWITCH1:

you could have problem if the switch is too much within the metallic support panel

E451 The cycle has been started with the door closed but the door is not blocked. You could have this error at the beginning of the cycle. 1. check BLOCK1:

refer to chapter “test of the peripherals connected to CBL606” 2. check the electrical connections between BLOCK1 and CBL606

3. check SWITCH2:

it’s possible that SWITCH2 hasn’t told to CBL606 that the pin of the electromagnet BLOCK1 has blocked the handle

4. check the electrical connections between SWITCH2 and CBL606:

the problem could be a false contact on the cable connector for the electronic board 5. it’s possible that the pin of the electromagnet BLOCK1 couldn’t go into the handle

due to a non correct alignment between the handle hole and the BLOCK1 pin:

• regulate the door as showed in the user manual: if the door is hard to close then the handle heel keep in distance from the metallic panel

• check the position of the electromagnet BLOCK1 on its support

• check the position of the door arm 6. check the position of SWITCH2:

it’s possible that the crank of SWITCH2 hasn’t heard the movement of BLOCK1 pin due to a non correct alignment between the pin and the crank

E461 The vacuum test has been started with the temperature in the chamber greater than 40°C. You could have this error at the beginning of the cycle. Wait that temperature in the chamber fall under 40°C.

Page 30: Service Manual Europa B

ERROR MESSAGES 3

A001 The user has stopped the cycle (by pushing the button B-START). If this alarm comes out every time you start a cycle (after about 10 seconds from the beginning of the cycle) then it’s possible that you have a problem on the front panel buttons.

A011 CBL606 doesn’t communicate with CBL607. Check the connections between the 2 electronic boards: the problem could be the cable that connects the 2 board; in other case we have observed that the problem is the electronic display board CBL607.

A101 The unit hasn’t reached the prefixed vacuum. You could have this alarm during T0, T2, T4 or T10. 1. clean or replace the gasket of the door 2. clean the external border of the chamber 3. adjust the door as explained in the user manual: the handle may close too soft 4. if you find a lot of water in the chamber after the alarm then the problem could be on EV4 5. check EV6 that must open and close alternatively while PUMP2 is working 6. check EV2 and EV7 7. check if EV5 is closed:

you could put a finger on FILTER3 and look if it sucks air (you could feel the air flow only if the leakage is not very small)

8. check if COOLER1 has a small hole 9. check the pipes, the connections and the links near the chamber 10. check if the stoppers/nuts over the chamber are well closed 11. check the silicon gaskets on TEMP1 / TEMP2 12. clean PUMP2 Typically each vacuum phase is about 4 or 5 minutes. If this phase takes more times then you could have one of the problem above.

Page 31: Service Manual Europa B

ERROR MESSAGES 4

A111 / A121 Loose of pressure during the vacuum test. You could have the alarm A111 during T11 and the alarm A121 during T12. 1. clean or replace the gasket of the door 2. clean the external border of the chamber 3. adjust the door as explained in the user manual: the handle may close too soft 4. if you find a lot of water in the chamber after the alarm then the problem could be on EV4 5. check EV2 and EV3 6. check if EV5 is closed:

you could put a finger on FILTER3 and look if it sucks air (you could feel the air flow only if the leakage is not very small)

7. check the pipes, the connections and the links near the chamber 8. check if the stoppers/nuts over the chamber are well closed 9. check the silicon gaskets on TEMP1 / TEMP2

It could be useful to find small hole or broken connection that make fail the vacuum phase with some special products like the one that you see in the picture below:

During the cycle you can put a lot of this product on a suspect link or part: if you have a small hole you will see that the foam will be sucked (during the vacuum phase) or sprayed (if the chamber is in pressure).

Page 32: Service Manual Europa B

ERROR MESSAGES 5

A401 SWITCH1 has indicated to the electronic board that the door has been opened during the cycle or SWITCH2 has indicated that the pin of the electromagnet has been removed from the handle. You could have this alarm during the whole cycle. 1. check SWITCH1 and SWITCH2 2. check the electrical connections between SWITCH1, SWITCH2 and CBL606:

the problem could be a false contact on the cable connector for the electronic board 3. check the status of the cables 4. check the alignment of SWITCH2 on the electromagnet BLOCK1:

the switch could be in a critical position 5. check the position of SWITCH1:

you could have problem if the switch is too much within the metallic support panel

A501 Loose of pressure during the sterilization phase. You could have this alarm during T6. If you have leakage during the sterilization then the pressure decrease and if the pressure goes under about 1.8 Bar during a 134 cycle or under 0.8 Bar during a 121 cycle then you have this alarm. Note that all the solenoid valve are closed during the sterilization so if a solenoid valve doesn’t close well then you could have a loose of pressure. Also check the gasket of the door, the links and the connections on the chamber and the gasket of the temperature probes. Check also the quantity of water that goes into the chamber (as explained in the alarm A651). This alarm message could be related to A651, A661 and A751.

A551 / A552 There is a problem with TEMP1 / TEMP2. The problem could be one of the temperature probe or the electronic board. TEMP1 is connected in the position marked with “T1” on the board CBL606 and TEMP2 is connected in the position marked with “T2”. Change the position of the 2 probe connectors on the board CBL606, reset the alarm and look if it change: from A551 to A552 or from A552 to A551. If the alarm doesn’t change then the problem is the board CBL606. Otherwise the problem is the temperature probe. Check also the status of the probe connectors / cables.

Page 33: Service Manual Europa B

ERROR MESSAGES 6

A641 High pressure in the chamber. You could have this alarm during T5 or T6. If the pressure in the chamber, during the preheating phase or the sterilization phase, is over about 2.4Bar then the unit will show this alarm. This problem could be related to the status of EV3 and EV6 or to a problem on the electronic board CBL606. 1. check the electrical connections between EV3, EV6 and CBL606 2. clean or replace EV3 and EV6 3. replace the electronic board CBL606 4. check the status of PRESSURE1

A651 The temperature in the chamber is over the maximum value for the selected cycle. You could have this alarm during T1, T3, T5 or T6. The unit shows this alarm code during the sterilization phase when the temperature read by TEMP1 or TEMP2 goes over 125°C for the cycle 121 or goes over 138°C for the cycle 134 or during the preheating phases when the temperature goes over 110°C and the pressure is less than 0.3 Bar. Step 1 If the pressure rises quickly after the vacuum phase then check the status of the solenoid valves connected to the chamber: EV2 and EV3. During the preheating phase EV7 is open to discharge the water in the COOLER1, so if you have a problem on EV2 or EV3 that remains open for some reasons then water or steam will escape from the chamber and you’ll have the overheating alarm. In this case the problem could be a solenoid valve, the electronic board or a cable. Step 2 If you have this alarm during the sterilization phase then run a cycle and look the values of TEMP1 and TEMP2 in the control menu: they must be quite closed near the sterilization phase and during this phase; if they aren’t then try to adjust the offset of TEMP1 or the offset of TEMP2 as explained in the example below:

Example 1 If you run a cycle 134 and you find in the control menu the following value: T1 = 134.2°C, T2 = 135.0°C and P1 = 2.1 Bar then increase the offset of TEMP1 of about 0.8°C Example 2 If you run a cycle 134 and you find in the control menu the following value: T1 = 135.0°C, T2 = 134.1°C and P1 = 2.1 Bar then increase the offset of TEMP2 of about 0.9°C

Page 34: Service Manual Europa B

ERROR MESSAGES 7

Example 3 If you run a cycle 134 and you find in the control menu the following value: T1 = 135.0°C, T2 = 134.1°C and P1 = 2.0 Bar then decrease the offset of TEMP1 of about 0.9°C

Note: typically the value of PRESSURE1 is correct and doesn’t need any adjustment. To understand the example above you have to refer to the following temperature-pressure diagrams:

120

121

122

123

124

125

0,9 1 1,1 1,2 1,3 1,4

Bar

°C

133

134

135

136

137

138

1,9 2 2,1 2,2 2,3 2,4

Bar

°C

Page 35: Service Manual Europa B

ERROR MESSAGES 8

Step 3 The cause could be also that the right quantity of water after the vacuum phase isn’t came into the chamber; for example by a closed pipe from TANK1 to the chamber trough EV4. 1. replace TANK1 with a graduated tank and

measure the quantity of water that goes into the chamber; if the chamber is 24 liters then you have to measure: about 0.35 liters after the first 2 vacuum phases and about 0.60 liters after the last vacuum phase; otherwise you have to measure: about 0.20 liters after the first 2 vacuum phases and about 0.45 liters after the last vacuum phase

2. if the quantities above are correct then

• check the hole in TANK1 that could be closed by some impurities

• check the pipe from TANK1 to FILTER2 that could be closed

• remove the water in TANK1 by OUT-3 until turn-on the LED of the minimum clean water level; remove all the remaining water in TANK1 by OUT-3 and put it in a graduated can; if the water is less than 0.8 liters then you have to replace TANK1

3. if they aren’t correct then

• check and clean EV4

• check FILTER2

• check the filter in the chamber

• check the connection and the link from EV4 to the chamber

• ask to the manufacturer if you need an upgraded microcontroller Step 4 This alarm could be also caused by a small loose of steam during the cycle:

• try to adjust the door as indicated in the user manual (if the door closes too softly then steam could came out)

• check and clean the solenoid valve connected to the chamber: EV2, EV3 and EV4

• check the connections and the links near the chamber (they could have a small hole, check it with a small mirror)

• check the gasket of TEMP1 and TEMP2 (during the cycle put a small mirror in the back of the chamber near TEMP1 and TEMP2)

Page 36: Service Manual Europa B

ERROR MESSAGES 9

Step 5 This alarm could be also caused by a problem on temperature reading:

• replace the electronic board CBL606

• replace TEMP1 and TEMP2 If you replace the temperature probes make attention to the silicon gasket – Figure 1. Every time you remove a temperature probe, remove also the old silicon gasket in the chamber and replace it. When you mount the new probe don’t close too much the screw in Figure 1 because you could damage the silicon gasket: put 2 fingers on the probe (from the internal side of the chamber) and try to push and pull the probe, if it’s fixed then you have closed enough the screw.

Figure 1 Figure 2 Figure 3

On the Figure 4 you see the silicon gasket that comes out from the hole: this is ok and it’s another signal that the probe has been well fixed.

Figure 4

Page 37: Service Manual Europa B

ERROR MESSAGES 10

Step 6 This alarm could be also caused by an arrangement phenomenon of the heating element over the chamber that will generate a non correct conduction of the heat:

• cut the top of the insulation band in 4 points (near the screws that fix the heating band) and screw them more

• check the position of the heating band: if it is over the weddings of the chamber top link then unscrew the fixing screws and move the heating band outside

• replace the heating band

A661 The temperature probes show different values during the sterilization phase: the difference between TEMP1 and TEMP2 is more than 2°C. You could have this alarm during T6.

This case is like A651 (follow the step 2, 3, 4, 5 and 6)

A701 The pressure in the chamber doesn’t rise correctly. You could have this alarm during T1, T3, T5. 1. check the thermo switch of the heating band:

this element could have opened by an over temperature problem 2. check if you have 230V on the contacts for the heating band in the CBL606

during the preheating phase: if not then replace the CBL606 3. the heating band could be broken: replace it

Page 38: Service Manual Europa B

ERROR MESSAGES 11

ATTENTION The A701 is shown when the unit take too much time to make the preheating phase, for example because the heating band isn’t working. If the heating band isn’t working then could be damaged the electronic board so the switch on the board doesn’t close (in this case you have to replace the CBL606). Also if the heating band isn’t working then it’s possible that in a previous cycle the unit is gone in over temperature and the thermoswitch has opened. The unit could be gone in over temperature for some different case: — no water in the chamber after the vacuum phase; — water is escaped during the preheating (for example an EV doesn’t close) — the switch on the electronic board that control the heating band is broken and the heating band is always

on So, when you have A701, it’s necessary also to keep in note the information above.

A751 The temperature in the chamber is under the minimum value for the selected cycle. You could have this alarm during T6. The unit shows this alarm code during the sterilization phase when the temperature read by TEMP1 or TEMP2 goes under 121°C for the cycle 121 or goes over 134°C for the cycle 134. This case is like A651 (follow the step 2, 3, 4, 5 and 6)

A761 The pressure sensor doesn’t work or there are problems with the connections 1. check the connections between PRESSURE1 and CBL606 2. replace PRESSURE1 3. replace CBL606

A801 The pressure in the chamber doesn’t fall correctly after the preheating phases or after the sterilization phase. You could have this alarm when the unit discharges the pressure in the chamber. 1. check and clean EV2 and EV7:

it’s possible that some impurity has gone into the hydraulic circuit so one of the 2 solenoid valve above could be blocked

2. check the cable and the connection between EV2, EV7 and CBL606 3. clean FILTER1 4. check and clean the filter in the chamber:

it’s possible that something blocks the flow of water / steam in the bottom hole / filter of the chamber

5. replace CBL606: it’s possible that the electronic doesn’t send the command to open the 2 valve above

Page 39: Service Manual Europa B

ERROR MESSAGES 12

ATTENTION

If the unit isn’t running a cycle and the pressure is increasing (for example because there is a lot of steam in the chamber and the unit is hot) then, over a certain level of pressure, the electronic board open the valve EV2 and EV7 to discharge the pressure. If the pressure doesn’t fall then the unit will show A801.

A901 Power has been removed from the unit during a cycle. It’s possible that you have a problem on the power line or someone has disconnected the power cable or switched-off the unit during a cycle. 1. reset the alarm and restart the cycle 2. check the status of the cables and the connections 3. check the status of the main green button

Page 40: Service Manual Europa B

CYCLE DIAGRAMS 1

CYCLE DIAGRAMS

PLOT 1

Pressure diagram when we push the key B-START for a cycle of 121°C or 134°C

PLOT 2

Pressure diagram when we push the key B-START for a vacuum test

Page 41: Service Manual Europa B

DISEGNO ESPLOSO / EXPLODED DRAWING

DISEGNO ESPLOSO / EXPLODED DRAWING

EUROPA B XP

Page 42: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_0

Page 43: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_1

Page 44: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_3

Page 45: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_1 15/06/2005 Sostituiti raccordi per carico e scarico acqua Cambia anche il pannello frontale in quanto cambiano i fori di fissaggio Replaced link for charge and discharge water Replaced also the front panel because it’s changed the position of the hole for the link above Rev_3 20/03/2007 Sostituito raccordo per carico acqua Eliminato: n°1 DANA111 n°1 CM20011 n°1 CM05028 Replaced link for water charging. Deleted: n°1 DANA111 n°1 CM20011 n°1 CM05028

Page 46: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 1 / EXPLODED DRAWING – PART 1

Rev_0

Page 47: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 1 / EXPLODED DRAWING – PART 1

Rev_1

Page 48: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 1 / EXPLODED DRAWING – PART 1

Rev_1 06/06/06 Sostituiti DHYA035, DXBA332 Rev_10 con DXBA335+CEST012 (TEMP3) e DXBA332 Rev_11 Replaced DHYA035 , DXBA332 Rev_10 with DXBA335+CEST012 and DXBA332 Rev_11

Page 49: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 2 / EXPLODED DRAWING – PART 2

Rev_0

Page 50: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 2 / EXPLODED DRAWING – PART 2

Rev_1

Page 51: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 2 / EXPLODED DRAWING – PART 2

Rev_1 25/07/2005 Sostituite sonde CESL006 con CESL009 Replaced level probe CESL006 with CESL009

Page 52: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 3 / EXPLODED DRAWING – PART 3

Rev_0

Page 53: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_0

Page 54: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_1

Page 55: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_2

Page 56: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_1 15/06/2005 Sostituiti raccordi in plastica CPRP003 con i raccordi in metallo DANA111 Sostituito pannello frontale DXBA307 con il DXBA304 Replaced plastic link CPRP003 with metallic connection DANA111 Replaced front panel DXBA307 with DXBA304 Rev_2 20/03/2007 Sostituito raccordo per carico acqua Eliminato: n°1 DANA111 n°1 CM20011 n°1 CM05028 Replaced link for water charging. Deleted: n°1 DANA111 n°1 CM20011 n°1 CM05028

Page 57: Service Manual Europa B

DISEGNO ESPLOSO / EXPLODED DRAWING

DISEGNO ESPLOSO / EXPLODED DRAWING

EUROPA B XP 24L

Page 58: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_0

Page 59: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_1

Page 60: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_2

Page 61: Service Manual Europa B

DISEGNO ESPLOSO – COMPLESSIVO / EXPLODED DRAWING – GLOBAL VIEW

Rev_1 08/06/06 Cambiate viti di fissaggio DXBA326 e DXBA323 Changed screw to fix the DXBA326 and the DXBA323 Rev_2 20/03/2007 Sostituito raccordo per carico acqua Eliminato: n°1 DANA111 n°1 CM20011 n°1 CM05028 Replaced link for water charging. Deleted: n°1 DANA111 n°1 CM20011 n°1 CM05028

Page 62: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 1 / EXPLODED DRAWING – PART 1

Rev_0

Page 63: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 2 / EXPLODED DRAWING – PART 2

Rev_0

Page 64: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 3 / EXPLODED DRAWING – PART 3

Rev_0

Page 65: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_0

Page 66: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_1

Page 67: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_2

Page 68: Service Manual Europa B

DISEGNO ESPLOSO – PARTE 4 / EXPLODED DRAWING – PART 4

Rev_1 08/06/06 Cambiate viti di fissaggio DXBA326 e DXBA323 Changed screw to fix the DXBA326 and the DXBA323 Rev_2 20/03/2007 Sostituito raccordo per carico acqua Eliminato: n°1 DANA111 n°1 CM20011 n°1 CM05028 Replaced link for water charging. Deleted: n°1 DANA111 n°1 CM20011 n°1 CM05028

Tecno-gaz Europa B EVO Instructions For Use Manual

  1. Manuals
  2. Brands
  3. Tecno-gaz Manuals
  4. Laboratory Equipment
  5. Europa B EVO
  6. Instructions for use manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Istruzioni per l’uso

ITALIANO

Instructions for use

ENGLISH

Mode d’emploi

FRANCAIS

Instrucciones de uso

ESPANOL

Bedienungsanleitung

DEUTCH

DXBT200 Rev_3

loading

Related Manuals for Tecno-gaz Europa B EVO

Summary of Contents for Tecno-gaz Europa B EVO

  • Page 1
    Istruzioni per l’uso ITALIANO Instructions for use ENGLISH Mode d’emploi FRANCAIS Instrucciones de uso ESPANOL Bedienungsanleitung DEUTCH DXBT200 Rev_3…
  • Page 2
    Costruito da / The Manufacturer / Fabriqué par / Fabricado por / Hersteller: MEDILINE ITALIA s.r.l. 0434 Via 8 marzo, 4 43025 Corte Tegge – Cavriago (RE) ITALY…
  • Page 3
    ITALIANO Gentile Cliente, La ringraziamo per aver scelto la nostra autoclave, sapremo ricambiare la Sua fiducia con la massima attenzione e un servizio sicuramente adeguato alle Sue aspettative. L’autoclave da Lei scelta è la massima espressione in fatto di tecnologia applicata. Può…
  • Page 4
    ITALIANO INDICE CONSULTAZIONE MANUALE GLOSSARIO DESCRIZIONE VISTA ANTERIORE / POSTERIORE SICUREZZA MARCATURA DI SICUREZZA DISPOSITIVI DI SICUREZZA NOTE DI SICUREZZA SMALTIMENTO IMPIEGO E DESTINAZIONE D’USO DELL’AUTOCLAVE DATI TECNICI MECCANICI ELETTRICI CAMERA SERBATOIO ACQUA PULITA SERBATOIO ACQUA USATA FILTRO BATTERIOLOGICO ACCESSORI DISIMBALLAGGIO INSTALLAZIONE ISTRUZIONI DI UTILIZZO…
  • Page 5
    ITALIANO 10.4 HELIX TEST 10.5 VACUUM TEST 11. TABELLE E DIAGRAMMI 12. LETTURA DEL REPORT DI CICLO 13. MESSAGGI DI ERRORE E ALLARMI 14. PROGRAMMAZIONE DISPLAY 14.1 SELEZIONE LINGUA 14.2 SELEZIONE ORA 14.3 SELEZIONE MINUTI 14.4 SELEZIONE SECONDI 14.5 SELEZIONE DATA 14.6 SELEZIONE MESE 14.7…
  • Page 6
    ITALIANO CONSULTAZIONE MANUALE Glossario VUOTO Pressione inferiore alla pressione atmosferica CAMERA Corpo cilindrico in acciaio che alloggia i dispositivi da sterilizzare FILTRO Dispositivo atto a trattenere particelle TRAY Cassetto di supporto del carico da sterilizzare PORTATRAY Struttura di supporto dei tray in camera di sterilizzazione Dispositivo atto ad eliminare i sali minerali dell’acqua proveniente dalla rete DEMINERALIZZATORE idrica…
  • Page 7
    ITALIANO FIG.2…
  • Page 8
    ITALIANO Filtro Batteriologico Carico Manuale Acqua demineralizzata con Tappo Carico Acqua demineralizzata Scarico Acqua Usata utilizzata Display Stampante Connessione RS232 Connettore incubatore biologico B-TEST Interruttore Generale Ciclo 121° C (valido per materiale imbustato e non imbustato) Ciclo 134° C (valido per materiale imbustato e non imbustato) Ciclo 134°…
  • Page 9
    ITALIANO SICUREZZA Marcatura di sicurezza TENSIONE PERICOLOSA TOGLIERE TENSIONE PRIMA DI RIMUOVERE IL COPERCHIO ALTA TEMPERATURA CONNESSIONE A TERRA Dispositivi di sicurezza L’autoclave è fornita dei seguenti dispositivi di sicurezza: -) Valvola di sicurezza tarata 2.4 bar 0/+10% -) Blocco elettromagnetico per evitare l’apertura del portello durante l’esecuzione del ciclo -) Termostato di sicurezza Note di Sicurezza •…
  • Page 10
    TIPO A: turbine, manipoli e dispositivi con fori ciechi o di piccole dimensioni. Questo dispositivo è certificato per la sterilizzazione dei seguenti carichi di materiali* : EUROPA B EVO EUROPA B EVO 24 Materiali solidi liberi ed imbustati max kg. 5 max kg. 7 Materiali cavi di tipo A e B max kg.
  • Page 11
    ITALIANO DATI TECNICI Meccanici EUROPA B EVO EUROPA B EVO 24 +5° C ÷ +40° C Temperatura di lavoro Altitudine MAX 2.000 m Umidità relativa MAX a 30° C Umidità relativa MAX a 40° C Dimensioni ingombro 510 X 390 X 590 mm…
  • Page 12
    ITALIANO Serbatoio Acqua Pulita EUROPA B EVO EUROPA B EVO 24 Senza Con demineralizz. demineralizz. Volume 4,5 l Vedere istruzioni Cicli eseguibili demineralizzatore Materiale polietilene Serbatoio Acqua Usata EUROPA B EVO EUROPA B EVO 24 Volume 4,5 l Cicli eseguibili…
  • Page 13
    ITALIANO ACCESSORI Portatray EUROPA B EVO EUROPA B EVO 24 Materiale Alluminio anodizzato Dimensioni (L x H x P) 192 x 165 x 280 mm 192 x 200 x 470 mm Immagine FIG.3 FIG.4 Dotazione prevista Portatray (opzionali) EUROPA B EVO…
  • Page 14
    ITALIANO Trays (opzionali) EUROPA B EVO EUROPA B EVO 24 Alluminio anodizzato o Materiale acciaio inox Dimensioni (L x H x P) 184 x 17 x 460 mm Immagine FIG.9 Dotazione prevista Chiave estrazione tray e regolazione portello Utilizzare per estrarre e manovrare i trays (FIG.11) e per poter regolare il portello (paragrafo 14.4)
  • Page 15
    ITALIANO Raccordo carico acqua Utilizzarlo con l’imbuto in plastica con tubo per caricamento manuale acqua e con tubo carico acqua completo di filtro Immagine FIG.14 Dotazione prevista Tubo carico acqua completo di filtro Utilizzare per caricamento acqua manuale (paragrafo 8.2) Immagine FIG.15 Dotazione prevista…
  • Page 16
    ITALIANO Piedini distanziale posteriore in plastica Applicare il piedino distanziale posteriore in plastica nella parte posteriore dell’autoclave (FIG.1–pos.0) per garantire un’adeguata ventilazione nel caso in cui l’autoclave venga posizionata vicino ad una parete Immagine FIG.18 Dotazione prevista Tubi per scarico posteriore utenze 1- Tubo per troppo pieno acqua pulita: collegare una estremità…
  • Page 17
    ITALIANO Cavo alimentazione Collegare l’estremità del connettore (FIG.21–pos.B) al pannello posteriore (FIG.1–pos.6) successivamente la spina (FIG.21–pos.A) direttamente alla presa di alimentazione dell’impianto elettrico. Immagine FIG.21 Dotazione prevista DISIMBALLAGGIO L’autoclave viene spedita in un imballo idoneo al trasporto, alla movimentazione e alla protezione dell’autoclave stessa.
  • Page 18
    ITALIANO INSTALLAZIONE FIG.22 EUROPA B EVO EUROPA B EVO 24 610 mm 590 mm 730 mm 50 mm 410 mm 426 mm (anteriore) 410 mm (anteriore) 455 mm (posteriore) 444 mm (posteriore) 400 mm 515 mm 1 — Installare l’autoclave in ambienti idonei alla sterilizzazione.
  • Page 19
    ITALIANO 8 — Lasciare all’interno della camera di sterilizzazione solo il portatray con i trays, posizionare tutti gli altri accessori in un vano esterno a disposizione degli operatori. 9 — Non appoggiare mai sull’autoclave giornali, vassoi, contenitori con liquidi ecc… 10 — Non appoggiarsi mai al portello quando è…
  • Page 20
    ITALIANO ISTRUZIONI DI UTILIZZO Dopo aver installato l’autoclave, procedere alla preparazione e all’utilizzo. Accensione dell’autoclave Premere l’interruttore generale (FIG.2–pos.19) Carico serbatoio acqua pulita Collegare al “tubo carico acqua completo di filtro” (in dotazione) il “raccordo carico acqua “ FIG.14, e inserirlo nel raccordo frontale dell’autoclave (FIG.2–pos.13). Inserire l’altro capo del tubo col filtro all’interno del contenitore dell’acqua demineralizzata o distillata.
  • Page 21
    ITALIANO Caratteristiche acqua da utilizzare TABELLA LIVELLI QUALITATIVI STABILITI DALLA NORMATIVA UNI EN 13060: 2009 CEN STANDARD UNI EN 13060 : 2009 ≤ Residuo evaporazione mg/l ≤ Ossido di silicio mg/l ≤ Ferro mg/l ≤ Cadmio 0.005 mg/l ≤ Piombo 0.05 mg/l ≤…
  • Page 22
    ITALIANO Inizio ciclo di sterilizzazione Scegliere il programma di sterilizzazione più idoneo al carico predisposto, premendo il pulsante B-SELECT. Ogni singola pressione, cambia la selezione del programma a quello successivo. Dopo aver scelto il programma avviare il ciclo premendo il pulsante B-START. La porta verrà…
  • Page 23
    ITALIANO Interrompere un ciclo di sterilizzazione Un ciclo di sterilizzazione può essere volontariamente interrotto, premendo il pulsante B-START per almeno 2 secondi. L’autoclave emetterà un suono, andrà in decompressione e sul display comparirà il messaggio di allarme A001 (CICLO INTERROTTO). Per resettare l’allarme, mantenere premuti contemporaneamente i pulsanti sotto evidenziati, sino all’intervento del segnale acustico: B-SELECT…
  • Page 24
    ITALIANO CICLI DI STERILIZZAZIONE Descrizione cicli L’autoclave è corredata di tre serie di cicli: A — cicli operativi B — cicli notte C — cicli test Cicli operativi Tutti i cicli operativi, hanno il sistema di vuoto frazionato, pertanto possono sterilizzare materiali cavi, porosi, solidi, sia liberi che imbustati.
  • Page 25
    ITALIANO TEST DI CONTROLLO AUTOCLAVE 10.1 Integratore chimici Sono test che sfruttano le proprietà di sostanza coloranti capaci di modificare il proprio aspetto se vengono adeguatamente esposte al calore e alla pressione in tempi adeguati. In quanto interagiscono con tutti i parametri del ciclo (pressione,temperatura, tempo). Il superamento del test ne certifica la corretta stabilità.
  • Page 26
    ITALIANO Le fiale devono essere inserite nella camera di sterilizzazione, anche con presenza di carico. UTILIZZO Eseguire il ciclo, al termine estrarre la fiala con attenzione in quanto calda e in pressione. Lasciarla raffreddare per circa 10 minuti poi attivarla, utilizzando l’apposita pinza in dotazione, mantenendola sempre in verticale.
  • Page 27
    ITALIANO La valutazione è semplice e rapida. RESPONSO Se il viraggio risulta uniforme (come in figura) il test è perfettamente riuscito, diversamente il test non è superato, pertanto l’autoclave non è in grado di sterilizzare correttamente i corpi porosi. Non esiste una periodicità predefinita, se non nelle zone ove vi sono leggi regionali specifiche.
  • Page 28
    Non esiste una periodicità predefinita, se non nelle zone ove vi sono leggi regionali specifiche. In tutti i casi onde potere avere un costante PERIODICITA’ controllo della validità dell’autoclave si consiglia di eseguire il test almeno una volta ogni settimana. WIR EMPFEHLEN, NUR VON TECNO-GAZ S.p.A. EMPFOHLENEN TEST ZU BENUTZEN…
  • Page 29
    ITALIANO TABELLE E DIAGRAMMI TABELLA STERILIZZAZIONE TEMPO TEMPO BANDA DI BANDA DI ASCIUG. PRESSIONE TEMPERATURA ESPOSIZ. CICLI T5 (Minuti) DI LAVORO DI LAVORO ( bar relativi ) ( °C ) (Minuti) 121° C 1.04 ÷ 1.30 121 ÷ 125 (imbustati e non imbustati) 134°…
  • Page 30
    ITALIANO LETTURA DEL REPORT DI CICLO Versione del software Numero seriale Data e Descrizione del ciclo selezionato (con i valori principali di setup) Ore – Minuti – Secondi Pressione della camera Temperatura principale della camera Fase del ciclo Fine del ciclo (OK significa che il ciclo si è…
  • Page 31
    ITALIANO MESSAGGI DI ERRORE E ALLARMI I messaggi di errore e di allarmi sono evidenziati tramite un codice alfanumerico, composto da una lettera e da 3 cifre. Il prefisso “E” è relativo ad errori dell’operatore, ripristinabili dall’operatore stesso. Il prefisso “A” è relativo ad allarmi, anomalie dell’autoclave, dopo aver attuato il rimedio, se il problema persiste bisogna richiedere assistenza telefonica.
  • Page 32
    ITALIANO Portello aperto durante ciclo o problema A 401 Controllare il sistema di chiusura chiusura A 501 Pressione non mantenuta Resettare e ripetere il ciclo A 551 Anomalia 1° sensore temperatura Resettare e ripetere i l ciclo A 552 Anomalia 2° sensore temperatura Resettare e ripetere i l ciclo A 641 Sovrapressione…
  • Page 33
    ITALIANO PROGRAMMAZIONE DISPLAY Fare riferimento alla FIG.2. 14.1 Selezione Lingua Premere B-START e contemporaneamente l’interruttore generale (MAINS). Tenere premuto B-START fino a quando sul display compare il nome della lingua impostata. Premendo B-SELECT si modifica la lingua [avanti]. Premendo B-PUMP si modifica la lingua [indietro].
  • Page 34
    ITALIANO 14.7 Selezione Anno Premendo B-DOOR compare anno. Premendo B-SELECT si modifica anno [avanti]. Premendo B-PUMP si modifica anno [indietro]. 14.8 Visualizzazione Pressione Atmosferica Premendo B-START compare la pressione atmosferica memorizzata. NON MODIFICARE ! 14.9 Selezione Carico Acqua Premendo B-START compare l’impostazione carico acqua : •…
  • Page 35
    ITALIANO SOLUZIONE A PROBLEMI OPERATIVI In molti casi, alcuni allarmi o errori sono determinati da non attenzione o non conoscenza di alcuni aspetti tecnici ed operativi. Qui di seguito vengono elencati alcuni casi di anomalie con relative soluzioni. 15.1 L’autoclave non asciuga correttamente •…
  • Page 36
    ITALIANO 15.6 La stampante dell’autoclave non stampa • il portellino portarotolo non e’ chiuso correttamente, sul display compare la lettera “D”. Aprire e richiudere il portellino della stampante accertandosi che sia chiuso correttamente. • manca la carta, sul display compare la lettera “P”. Aprire il portello e inserire il rotolo di carta termografica.
  • Page 37
    ITALIANO DEMINERALIZZATORE 16.1 Impostazione sistemi di demineralizzazione L’autoclave è predisposta anche per il caricamento dell’acqua demineralizzata attraverso il SISTEMA IONICO o SISTEMA OSMOSI (accessorio opzionale). Tali prodotti sono utilizzati per eliminare i sali minerali dell’acqua proveniente dalla rete idrica per alimentare l’autoclave. L’operatore prima di installare il sistema IONICO o OSMOSI deve programmare l’autoclave seguendo le istruzioni sotto indicate: Accendere l’autoclave premendo l’interruttore generale (FIG.2–POS.19) e tenere premuto…
  • Page 38
    ITALIANO 16.2 Collegamento dei sistemi di demineralizzazione Qui riportiamo il particolare del collegamento specifico all’autoclave del tubo di alimentazione dell’acqua e della connessione della spina elettrica: Spegnere l’autoclave se è accesa (FIG.2–pos.19) • • • • Chiudere il rubinetto posto a monte dell’impianto di demineralizzazione •…
  • Page 39
    ITALIANO FIG.31 16.3 Cambio cartuccia resina — filtro L’autoclave segnala sul display lo stato di esaurimento delle resine o del filtro con il messaggio “SOSTITUIRE RESINE” (cod.ricambio 242-S) o “SOSTITUIRE FILTRO” (cod.ricambio 248-S-2). Sostituire la cartuccia come indicato nel manuale del demineralizzatore; dopo sostituzione, premere il pulsante B-PUMP fino all’intervento del segnale acustico, per resettare…
  • Page 40
    Chi non segue tutte le varie fasi del processo di sterilizzazione, può inficiare il risultato finale. TECNO-GAZ S.p.A produce tutti i dispositivi per ogni fase del ciclo di sterilizzazione. Bisogna pertanto : 1 — Predisporre un preciso protocollo operativo scritto,da distribuire a tutti gli operatori ed addetti alla sterilizzazione.
  • Page 41
    ITALIANO ISTRUZIONE PER LA SPEDIZIONE DELL’AUTOCLAVE Nel caso si debba spedire l’autoclave per riparazione, controllo, ripristino, revisione, validazione, seguire le seguenti indicazioni obbligatorie: • Utilizzare l’imballo originale, se questo non è più in Vostro possesso, utilizzare un imballo adeguato. La merce viaggia con rischio a carico del mittente. •…
  • Page 42
    ITALIANO PROCEDURE PER SERVIZIO ED ASSISTENZA Il reparto riparazioni è una entità autonoma di MEDILINE ITALIA SRL (TECNO-GAZ GROUP), costituito da una struttura articolata e professionale, in conformità con quelle che sono le nostre procedure espresse nel nostro “Manuale Qualità”.
  • Page 43
    È comunque escluso il diritto alla sostituzione dell’intero apparecchio. Per quanto riguarda i componenti applicati o integrati prodotti con marchio TECNO-GAZ S.P.A. o da terzi e dotati di proprio certificato di garanzia, valgono le condizioni, i limiti e le esclusioni riportati sui certificati stessi.
  • Page 44
    é conforme alle istruzioni di TECNO-GAZ S.p.A. ed é stata eseguita da personale non autorizzato; c) l’apparecchiatura venga riparata, modificata o comunque manomessa dall’Acquirente o da terzi non autorizzati;…
  • Page 45
    ITALIANO MANUTENZIONE Una corretta manutenzione dell’autoclave assicura un buon funzionamento della stessa e un sicuro risparmio in termini di tempo e costi dovuti ad assistenza. Le seguenti operazioni sono da intendersi obbligatorie ed eseguibili dagli operatori. Pulizia della camera 20 cicli oppure 1 volta la settimana Pulire periodicamente la camera, asportando eventuali depositi o detriti, eviterete così…
  • Page 46
    ITALIANO Regolazione portello ogni 2 mesi Per preservare l’integrità funzionale della macchina, bisogna regolare la pressione di chiusura del portello, agendo sul regolatore del portello stesso, utilizzando la chiave a doppia funzione (FIG.10) in dotazione (cod.ricambio DANA008). Ruotare in senso antiorario (pos.A) per aumentare la pressione di chiusura.
  • Page 47
    ITALIANO Le seguenti operazioni sono consigliate dal produttore. Tali operazioni devono essere eseguite esclusivamente da tecnici formati e autorizzati dal produttore. ogni 1 anno / ogni 1000 cicli * Sostituzione del filtro carico acqua ogni 1 anno / ogni 1000 cicli * Controllo del filtro acqua ingresso flussimetro ogni 1 anno / ogni 1000 cicli * Pulizia filtro Y in ottone a valle del radiatore…
  • Page 48
    ITALIANO NOTE…
  • Page 49
    ENGLISH Dear Client, Thank you for having chosen our autoclave, we know how to exchange your fidelity, with maximum attention and service definitely corresponding to your expectations. The autoclave you have chosen is the maximum expression of applied technology. It can be used for dental, medical, and aesthetic applications and in general in all the sectors in which there is the need of performing a safe sterilization of tools and materials of any kind: solid, hollow, and porous loads, both wrapped and non-wrapped.
  • Page 50: Table Of Contents

    ENGLISH TABLE OF CONTENTS CONSULTING THE MANUAL GLOSSARY FRONT VIEW / REAR VIEW DESCRIPTION SAFETY SAFETY MARKING SAFETY DEVICES SAFETY NOTICES DISPOSAL AUTOCLAVE USE AND APPLICATIONS TECHNICAL DATA MECHANICAL DATA ELECTRICAL DATA CHAMBER CLEAN WATER TANK USED WATER TANK BACTERIOLOGICAL FILTER ACCESSORIES UNPACKING INSTALLATION…

  • Page 51
    ENGLISH 10.4 HELIX TEST 10.5 VACUUM TEST 11. TABLE AND DIAGRAMS 12. CYCLE PRINTOUT READING 13. ERROR MESSAGES AND ALARMS 14. PROGRAMMING AND PARAMETERS DISPLAY 14.1 LANGUAGE SELECTION 14.2 HOURS SELECTION 14.3 MINUTES SELECTION 14.4 SECONDS SELECTION 14.5 DATE SELECTION 14.6 MONTH SELECTION 14.7…
  • Page 52: Consulting The Manual

    ENGLISH CONSULTING THE MANUAL Glossary VACUUM Pressure below atmospheric pressure CHAMBER Cylindrical steel body that, during the cycle, is filled with saturated steam FILTER Device for retaining particles TRAY Drawer for holding the load to be sterilized TRAY HOLDER Structure for holding the trays in the sterilization chamber DEMINERALIZER Device for eliminating mineral salts in mains water HELIX TEST…

  • Page 53
    ENGLISH FIG.2…
  • Page 54
    ENGLISH Bacteriological filter Manual water filling with funnel Water filling with pump Used water draining Display Printer RS232 Connector B-TEST Connector General switch 121° C Cycle (valid for wrapped and unwrapped materials) 134° C Cycle (valid for wrapped and unwrapped materials) Cycle 134°…
  • Page 55: Safety

    ENGLISH SAFETY Safety Marking HAZARDOUS VOLTAGE DISCONNECT THE POWER BEFORE REMOVING THE LID HIGH TEMPERATURE EARTH CONNECTION Safety Devices The following safety devices are installed: -) Safety valve set at 2.4 bar 0/+10% -) Electromagnetic lock to prevent the door from opening while the cycle is running -) Resistance over temperature thermostats Safety Notes •…

  • Page 56: Disposal

    TYPE A: turbines, hand pieces and devices with blind or small holes This device has been certified for the sterilization of the following materials *: EUROPA B EVO EUROPA B EVO 24 Wrapped and unwrapped solid max kg. 5 max kg. 7…

  • Page 57: Technical Data

    ENGLISH TECHNICAL DATA Mechanical Data EUROPA B EVO EUROPA B EVO 24 +5° C ÷ +40° C Working temperature Maximum altitude 2.000 m MAX relative humidity at 30° C MAX relative humidity at 40° C Dimensions of space occupied 510 X 390 X 730 mm…

  • Page 58: Clean Water Tank

    ENGLISH Clean Water tank EUROPA B EVO EUROPA B EVO 24 Without With Demineralizer Demineralizer Volume 4,5 l Read demineralizer Usable cycles instruction polyethylene Material Used Water tank EUROPA B EVO EUROPA B EVO 24 Volume 4,5 l Usable cycles…

  • Page 59: Accessories

    ENGLISH ACCESSORIES Tray holder EUROPA B EVO EUROPA B EVO 24 Material Aluminium anodized Size (L x H x P) 192 x 165 x 280 mm 192 x 200 x 470 mm Picture FIG.3 FIG.4 Envelope standard Tray holder (optional)

  • Page 60
    ENGLISH Trays (optional) EUROPA B EVO EUROPA B EVO 24 Aluminium anodized or Material stainless steel Size (L x H x P) 184 x 17 x 460 mm Picture FIG.9 Envelope standard Tray extraction and door adjustment wrench Use for extract the trays (FIG.11) and for door adjustement (par.14.4) Picture FIG.11…
  • Page 61
    ENGLISH Connection for water filling pipe and plastic funnel Use with plastic funnel and water filling pipe with filter Picture FIG.14 Envelope standard Water filling pipe with filter Use to charge water with pump (par.8.2) Picture FIG.15 Envelope standard Water discharghe pipe Use to empty used water tank (par.8.9) Picture FIG.16…
  • Page 62
    ENGLISH Rear spacer Put the spacer in the autoclave’s back panel (FIG.1–pos.0) It’s necessary for guarantee a good ventilation if you place the autoclave near a wall. Picture FIG.18 Envelope standard Pipes for discharge utilities 1- Pipe for clean water overflow: Connect one pipe end into the water overflow pipe fitting (FIG.1–pos.1), and put the other end in a tank or in the discharge (demineralizer version).
  • Page 63: Unpacking

    ENGLISH Power supply cord Take the power supply cord provided and insert the female plug (FIG.21–pos.B) in the socket of the back panel of the autoclave (FIG.1–pos.6). Then insert the male plug (FIG.21–pos.A) in the electric plug of the system. Picture FIG.21 Envelope standard…

  • Page 64: Installation

    ENGLISH INSTALLATION FIG.22 EUROPA B EVO EUROPA B EVO 24 610 mm 590 mm 590 mm 50 mm 410 mm 426 mm (front) 426 mm (front) 455 mm (rear) 455 mm (rear) 400 mm 400 mm 1 — Install the autoclave in an environment suitable for carrying out sterilization procedures.

  • Page 65
    ENGLISH 8 — Leave inside the sterilization chamber only the tray carrier with the trays. All other accessories should be positioned in a separate space available to operators. 9 — Never place on the autoclave newspapers, trays, liquid containers, etc… 10 — Do not lean on the door when it is open.
  • Page 66: Usage Instruction

    ENGLISH USAGE INSTRUCTION After having installed the autoclave, proceed with preparations and use. Turn on the autoclave Press the general switch (FIG.2–pos.19) Loading pure water tank Connect the water filling pipe (provided) with the connection for water filling pipe and plastic funnel PICT.14.

  • Page 67
    ENGLISH Characteristics of the water to be used TABLE SHOWING THE QUALITY LEVELS LAID BY THE UNI EN 13060 : 2009 CEN STANDARD UNI EN 13060 : 2009 ≤ Evaporation residue mg/l ≤ Silicon oxide mg/l ≤ Iron mg/l ≤ Cadmium 0.005 mg/l…
  • Page 68: End Of Cycle

    ENGLISH Starting the sterilization cycle Choose sterilization program most adaptable of load prepared and press the button B-SELECT. Each individual pressure changes program selection to the next. After having chosen the program start the cycle by pressing the button B-START, the door will lock automatically and the cycle will begin.

  • Page 69: Interruption Of Sterilization Cycle

    ENGLISH Interruption of sterilization cycle A sterilization cycle can be voluntarily interrupted by pressing the button B-START for at least 2 seconds. The autoclave will emit a beep, go into decompression and on the display will appear the message ALLARM 001 (INTERRUPTED CYCLE). To reset error, maintain pressed simultaneously the buttons below indicated, until the intervention of an acoustic signal: B-SELECT…

  • Page 70: Sterilization Cycles

    ENGLISH STERILIZATION CYCLES Descriptions cycles The autoclave has three series of cycles: A — operation cycles B — night cycles C — test cycles Operation cycles All operation cycles have a system of fractioned vacuum, which therefore permits sterilization of materials that are hollow, porous, solid, both free and packaged. Temperatures can be selected from 121°…

  • Page 71: Autoclave Validation Tests

    ENGLISH AUTOCLAVE VALIDATION TESTS 10.1 Chemical integrators These tests (spare code 200/S and 215-S) are based on the properties of indicator dyes that are able to modify their own aspect if they are adequately exposed to heat and pressure according to given times, since they interact with all the cycle parameters (pressure, temperature, time).

  • Page 72: Bowie-Dick Test

    ENGLISH The ampoules shall have to be put into the sterilization chamber even when loaded. Carry out the cycle and then take out the ampoule by taking care since it is hot and pressurized. Leave it cool down for about 10 minutes and then activate it by using the supplied crusher by taking care so as to keep the ampoule always upright.

  • Page 73: Helix Test

    ENGLISH the assessment is simple and rapid. RESPONSE If the colour change turns out to be uniform (as shown in the figure) the test perfectly succeeded. On the contrary the test failed, hence the autoclave cannot properly sterilize porous bodies. There is no preset testing interval, apart from areas to which specific regional regulations apply.

  • Page 74: Vacuum Test

    ENGLISH BEFORE THE TEST RESPONSE TEST FAILED TEST SUCCEEDED There is no preset testing interval, apart from areas to which specific PERIODICITY regional regulations apply. Hence in order to constantly monitor the autoclave performance, it is advisable to perform the test at least once every 30 days.

  • Page 75
    ENGLISH TABLES AND CHARTS STERILIZATION TABLE DRYING PRESS.WORK EXPOSED TEMPERATURE TIME CYCLES MIN AND MAX WORK MIN AND TIME T4 T5 (Minutes) MAX ( ° C ) (Minutes) (relative bar) 121° C 1.04 ÷ 1.30 121 ÷ 125 (wrapped and unwrapped) 134°…
  • Page 76: Cycle Printout Reading

    ENGLISH CYCLE PRINTOUT READING Software release Serial number Date and Description of the selected cycle Hours – Minutes – Seconds Chamber pressure Chamber main temperature Cycle phase End of cycle (OK means that the cycle has been correctly completed) Cycle number Operator (blank for signature)

  • Page 77
    ENGLISH MESSAGES OF ERROR OR ALARMS Messages of error are emphasized through an Alfa-numerical code, consisting in a letter and three numbers. The “E” code is relative to errors of the operator, which can be corrected by same operator. The “A” code is relative to alarms, irregularities of the autoclave, after having carried out the remedy, if the problem persists it is necessary to request telephonic assistance.
  • Page 78
    ENGLISH Hatch opende during the cycle or problem A 401 Check the closing system closing A 501 Pressure not maintened Reset and restart cycle A 551 Irregular 1° temperature sensor Reset and restart cycle A 552 Irregular 2° temperature sensor Reset and restart cycle A 641 Overly high pressure…
  • Page 79: Language Selection

    ENGLISH PROGRAMMING DISPLAY Please, see the FIG.2. 14.1 Language selection Press simultaneously B-START and MAINS switch. Hold down B-START until on the display appears the name of the language programmed. Press B-SELECT to modify the language [forward]. Press to B-PUMP modify the language [backwards].

  • Page 80: Year Selection

    ENGLISH 14.7 Year selection Press B-DOOR year appears. Press B-SELECT to modify year [forward]. Press B-PUMP to modify year [backwards]. 14.8 Visualize atmospheric pressure Press B-START and compare memorized atmospheric pressure appears. DO NOT MODIFY! 14.9 Water loading selection Press and release B-START to select next screen until you see setting the water load: •…

  • Page 81: Solution To Operation Problems

    ENGLISH SOLUTIONS TO OPERATIONAL PROBLEMS In many cases some alarms or errors are determined by lack of attention or lack of familiarity with some technical and operational aspects. Following are listed some cases of irregularity with relative solutions. 15.1 Autoclave does not dry correctly •…

  • Page 82: Autoclave Printer Does Not Print

    ENGLISH 15.6 Autoclave printer does not print • Paper carrier cover is not closed correctly, on the display appears the letter “d”. open and close the printer cover checking that is correctly closed. • Paper missing, on the display appears the letter “p”. open the cover and insert the roll of thermo-graphic paper.

  • Page 83: Demineralizer

    ENGLISH DEMINERALIZER 16.1 Set of demineralizer system Autoclave is already prepared also for plug IONIC SYSTEM or OSMOSI SYSTEM for charging water into the autoclave (optional accessories). These applications are accessories of autoclave. These applications have been designed for remove minerals in the water before to charge the autoclave.

  • Page 84
    If the unit at the beginning of the cycle shows E250 then you have to fill the clean water tankuntil the maximun is reached. NOTE: for the connection of demineralization system to the autoclave tecno-gaz refer to indications in the manual.
  • Page 85: Change Resins Of Filter

    ENGLISH FIG.31 16.3 Change resins — filter Autoclave display indicates the state of finishing of the resin with the message “REPLACE THE RESINS” (spare code 242-S) or “REPLACE THE FILTER” (spare code 248-S-2). Substitute the resin cartridge as indicated in the deminaralizer manual. After substitution, press button B-PUMP until the “beep”, to reset the autoclave.

  • Page 86: Standard Sterilization Procedure

    Autoclave is the final link of this system. Those who do not follow all of the various phases of the sterilization system will be unsatisfied with the final result. TECNO-GAZ S.p.A. produces all the devices for each phase of the sterilization cycle. Therefore it is necessary 1 — Prepare a precise written operational protocol to distribute to all operators assigned to sterilization.

  • Page 87: Instructions For Autoclave Shipping

    ENGLISH INSTRUCTION FOR AUTOCLAVE SHIPPING In case the autoclave must be shipped for repairs, controls, reactivation, revisions, validations follow the obligatory indications below: • Use the original packaging; if this is no longer in your possession, use adequate packaging. The merchandise travels at risk to the sender. •…

  • Page 88: Procedures For Service And Assistance

    ENGLISH PROCEDURE FOR SERVICE AND ASSISTANCE The repair service dept is an autonomous division of MEDILINE ITALIA SRL (TECNO- GAZ GROUP), comprising a complex and professional structure, in compliance with the procedures set in our “Quality Assurance Manual”. • When the autoclave arrives at MEDILINE ITALIA SRL it is consigned directly to the service center, which codes and creates the technical assistance document.

  • Page 89
    The right to replace the entire machine is excluded. As for components either applied or added to TECNO-GAZ products or by third parties, having their own guarantee certificate, the conditions, limits and exclusions indicated in the relevant certificates apply.
  • Page 90
    TECNO-GAZ instructions and by authorized personnel; c) the equipment is repaired, modified or tampered with by the Purchaser or by non- authorized third parties;…
  • Page 91: Maintenance

    ENGLISH MAINTENANCE Correct maintenance of the autoclave assures correct functioning of it and a secure saving in terms of time and costs for assistance and maintenance. The following operations are compulsory feasible by operators. Cleaning of chamber Every 20 cycles or once a week Clean periodically the chamber, remove eventual deposits or debris, thus avoiding the introduction in the discharging circuit of material which can cause obstructions.

  • Page 92
    ENGLISH Door regulation Every 2 mouths To regulate the door lock pressure, it is necessary to adjust the regulator of the same door using the double function wrench (extraction tray + wrench) (FIG.10) provided (spare code DANA008). Turn anti clockwise (pos.A) to increase lock pressure. Turn clockwise (pos.B) to decrease the lock pressure Substitution printer paper Thermographic paper roll 57 mm wide (spare code DNVA363).
  • Page 93
    ENGLISH The following steps are recommended by the manufacturer. These operations must be performed only by trained technicians authorized by the manufacturer Replacement of the water loading filter every 1 year / every 1000 cycles * Control of the water filter of flowmeter inlet every 1 year / every 1000 cycles * Cleaning Y brass filter downstream of the radiator every 1 year / every 1000 cycles *…
  • Page 94
    ENGLISH NOTE…
  • Page 95
    FRANÇAIS Cher client, ous vous remercions d’avoir choisi notre autoclave et saurons mériter votre confiance avec la plus grande attention et un service en mesure de répondre à votre attente. L’autoclave que vous avez choisi est la meilleure expression en fait de technologie appliquée Il peut avoir des applications dans le domaine dentaire, médical, esthétique et en général dans tous les secteurs où…
  • Page 96
    FRANÇAIS INDEX CONSULTATION MANUEL GLOSSAIRE DESCRIPTION VUE AVANT/ARRIERE SÉCURITÉ MARQUAGE DE SÉCURITÉ DISPOSITIF DE SÉCURITÉ NOTES DE SÉCURITÉ MISE AU REBUT UTILISATION ET APPLICATIONS DE L’AUTOCLAVE DONNÉES TECHNIQUES MÉCANIQUES ÉLECTRIQUES CHAMBRE RÉSERVOIR EAU PROPRE RÉSERVOIR EAU USÉE FILTRE BACTÉRIOLOGIQUE ACCESSOIRES DÉBALLAGE INSTALLATION MODE D’EMPLOI…
  • Page 97
    FRANÇAIS 10.4 TEST HELIX 10.5 TEST DE VIDE 11. TABLES ET TABLEAUX 12. LECTURE DU RAPPORT DU CYCLE 13. MESSAGES D’ERREUR ET ALARMES 14. PROGRAMMATION DE L’ECRAN 14.1 SÉLECTION LANGUE 14.2 SÉLECTION HEURE 14.3 SÉLECTION MINUTES 14.4 SÉLECTION SECONDES 14.5 SÉLECTION DATE 14.6 SÉLECTION MOIS…
  • Page 98
    FRANÇAIS CONSULTATION MANUEL Glossarie VIDE Pression inférieure à la pression atmosphérique CHAMBRE Corps cylindrique en acier qui loge les appareils à stériliser FILTRE Dispositif pour retenir les particules PLATEAU Tiroir de support de la charge à stériliser PORTE-PLATEAUX Structure de support des plateaux dans la chambre de stérilisation Dispositif pour éliminer les sels minéraux de l’eau qui vient du réseau DEMINERALISATEUR hydrique…
  • Page 99
    FRANÇAIS FIG.2…
  • Page 100
    FRANÇAIS Filtre bactériologique Remplissage manuel eau Remplissage eau Évacuation eau usée Écran Imprimeur Connexion RS232 Le connecteur de l’incubateur B-TEST Interrupteur général Cycle 121° C (valable pour les matériaux ensaché et non ensaché) Cycle 134° C (valable pour les matériaux ensaché et non ensaché) Cycle 134°…
  • Page 101
    FRANÇAIS SÉCURITÉ Marquage de sécurité TENSION DANGEREUSE COUPER LE COURANT AVANT D’ENLEVER LE COUVERCLE TEMPÉRATURE ÉLEVÉE MISE A LA TERRE Dispositifs de sécurité Les dispositifs de sécurité sont les suivants : ± -) Soupape de sécurité étalonnée à 2,4 bar -) Verrouillage électromagnétique pour prévenir l’ouverture du panneau pendant le cycle -) Thermostat de sécurité…
  • Page 102
    TYPE A : turbines, tubes, manipules et dispositifs avec des trous des petites dimensions Ce dispositif a été certifié pour la stérilisation des suivants instruments*: EUROPA B EVO EUROPA B EVO 24 Matériels ferreux ou solides ensaches max kg. 5 max kg. 7 Matériels cavità…
  • Page 103
    FRANÇAIS DONNÉES TECHNIQUES Mécaniques EUROPA B EVO EUROPA B EVO 24 +5° C ÷ +40° C Température d’exercice Altitude max. 2.000 m Humidité relative MAX. à 30° C Humidité relative MAX. à 40° C Dimensions hors tout (L x H x P)
  • Page 104
    FRANÇAIS Réservoir eau propre EUROPA B EVO EUROPA B EVO 24 Avec Sans demineralizz. demineralizz. 4,5 l Volume Vour les instructions Nombre de cycles réalisables demineralizers Polyéthylène Composition Réservoir eau usée EUROPA B EVO EUROPA B EVO 24 4,5 l Volume Nombre de cycles réalisables…
  • Page 105
    FRANÇAIS ACCESSOIRES Porte-plateaux EUROPA B EVO EUROPA B EVO 24 Composition Aluminium Anodisé Dimensions (L x H x P) 192 x 165 x 280 mm 192 x 200 x 470 mm Image FIG.3 FIG.4 Dotation prévue Porte-Plateaux (facultatif) EUROPA B EVO…
  • Page 106
    FRANÇAIS Plateaux (facultatif) EUROPA B EVO EUROPA B EVO 24 Composition Aluminium Anodisé Acier Inoxydable Dimensions (L x H x P) 184 x 17 x 460 mm Image FIG.9 Dotation prévue Pince d’extraction plateau et réglage porte A employer pour extraire et manutentionner les plateaux (FIG.11) et pour pouvoir régler la porte (paragrafo 14.4)
  • Page 107
    FRANÇAIS Raccord charge eau A employer avec l’entonnoir en plastique avec tuyau pour charge manuelle de l’eau et avec tuyau charge eau avec filtre Image FIG.14 Dotation prévue Entonnoir en plastique avec tuyau A employer pour charge manuelle de l’eau (paragrafo 8.2) Image FIG.15 Dotation prévue…
  • Page 108
    FRANÇAIS Pied d’écartement arrière en plastique Mettre le pied d’écartement dans la partie arrière de l’autoclave (FIG.1–pos.0) pour assurer un aérage adéquat en cas de positionnement de l’autoclave près d’une paroi. Image FIG.18 Dotation prévue Tuyaux pour vidange arrière des eaux 1- Tuyau pour trop-plein eau propre: Connecter une extrémité…
  • Page 109
    FRANÇAIS Cordon d’alimentation Connecter l’extrémité du connecteur (FIG.21–pos.B) au panneau arrière (FIG.1–pos.6) ensuite brancher la fiche (FIG.21–pos.A) directement à la prise de secteur. Image FIG.21 Dotation prévue DÉBALLAGE L’autoclave est expédié dans un emballage protecteur idoine lui permettant d’être manutentionné facilement. L’emballage ne doit pas subir de heurts, il doit être manipulé avec attention en évitant de le faire rouler ou tomber d’une hauteur supérieure à…
  • Page 110
    FRANÇAIS INSTALLATION FIG.22 EUROPA B EVO EUROPA B EVO 24 610 mm 590 mm 730 mm 50 mm 410 mm 426 mm ( 403 mm ( avant avant 455 mm ( 444 mm ( arriere arriere 400 mm 1 — Installez l’autoclave dans un environnement idoine à la stérilisation.
  • Page 111
    FRANÇAIS 8 — Ne laissez dans la chambre de stérilisation que le porte-plateaux et les plateaux. Rangez tous les autres accessoires dans un logement externe à la disposition du personnel. 9 — Ne posez jamais sur l’autoclave des journaux, des plateaux, des conteneurs de liquide, etc 10 — Ne jamais s’appuyer au panneau lorsqu’il est ouvert.
  • Page 112
    FRANÇAIS MODE D’EMPLOI Après avoir installé l’autoclave, procédez à la préparation et à l’utilisation. Allumage autoclave En appuyant sur l’interrupteur général (FIG.2–pos.19) Remplissage réservoir eau pure Connecter le “raccord charge eau” FIG.14 au “tuyau charge eau avec filtre” (standard), et l’insérer dans l’orifice du raccord frontal de l’autoclave (FIG.2–pos.13).
  • Page 113
    FRANÇAIS Caractéristiques de l’eau a utiliser TABLEAU NIVEAUX QUALITATIFS ÉTABLIS PAR LA NORME UNI EN 13060 : 2009 CEN STANDARD UNI EN 13060 : 2009 ≤ Résidu d’évaporation mg/l ≤ Oxyde de silicium mg/l ≤ mg/l ≤ Cadmium 0.005 mg/l ≤…
  • Page 114
    FRANÇAIS Début cycle de stérilisation Sélectionnez le programme de stérilisation plus idoine au matériel à stériliser en appuyant sur le bouton B-SELECT. Chaque pression commute la sélection du programme au suivant. Après avoir sélectionné le programme, démarrez le cycle en appuyant sur le bouton B-START, la porte se verrouillera automatiquement et le cycle aura lieu.
  • Page 115
    FRANÇAIS Interruption du cycle de stérilisation Un cycle de stérilisation peut être volontairement interrompu en appuyant sur le bouton B-START pendant 2 secondes au moins. L’autoclave produira un son, ira en décompression et le message d’erreur A001 s’affichera sur le visuel (CYCLE INTERROMPU). Pour restaurer l’erreur, appuyez simultanément sur les boutons illustrés ci-dessous jusqu’au déclenchement du signal acoustique : B-SELECT…
  • Page 116
    FRANÇAIS CYCLES DE STÉRILISATION Description cycles L’autoclave est doté de trois séries de cycles : A — cycles opérationnels B — cycles opérationnels-cycles nuit C — cycles tests Cycles opérationnels Tous les cycles opérationnels ont le système de vide fractionné, ils peuvent donc stériliser des matériels creux, poreux, solides, aussi bien en vrac qu’ensachés.
  • Page 117
    FRANÇAIS ESSAIS DE CONTROLE AUTOCLAVE 10.1 Intégrateurs chimiques Ces sont des essais qui exploitent les propriétés des colorants étant capables de modifier leur aspect si exposés d’une manière adéquate à la chaleur et à la pression pour un certain délai, puisqu’ils interagissent avec tous les paramètres du cycle (pression, température, temps).
  • Page 118
    FRANÇAIS Il faut positionner les ampoules dans la chambre de stérilisation, même à la présence de charge. UTILISER Effectuer le cycle, à la fin enlever l’ampoule en faisant attention puisqu’elle est chaude et sous pression. La faire refroidir pour environ 10 minutes, ensuite l’activer, à l’aide du concasseur à…
  • Page 119
    FRANÇAIS L’évaluation est simple et rapide. Si RÉPONSE le virage résulte uniforme (comme dans la figure) l’essai a réussi parfaitement, autrement l’essai n’a réussi, conséquent l’autoclave n’est pas à même de stériliser correctement les charges poreuses. Il n’y a pas d’intervalles préétablis pour les contrôles périodiques, sauf dans les zones où…
  • Page 120
    En tous FRÉQUENCE cas, pour avoir un contrôle constant de l’efficacité de l’autoclave, on conseille d’effectuer l’essai au moins une fois par semaine. ON CONSEIL D’UTILISER EXCLUSIVEMENT LES TESTS PROPOSES PAR TECNO-GAZ S.p.A.
  • Page 121
    FRANÇAIS TABLES ET TABLEAUX TABLEAU STÉRILISATION DURÉE PLAGE DE PLAGE DE DURÉE SÉCHAGE PRESSION TEMPERATURE CYCLES EXPOSITION T5 (Minutes) DE TRAVAIL DE TRAVAIL T4 (Minutes) ( bars relatifs ) ( ° C ) 121° C 1.04 ÷ 1.30 121 ÷ 125 (ensaché…
  • Page 122
    FRANÇAIS LECTURE DU RAPPORT DU CYCLE Version du logiciel Numéro série Date et Description du cycle sélectionné (avec les principales valeurs de configuration) Heures – Minutes – Secondes Pression de la chambre Température principale de la chambre Phase de cycle Fin du cycle (OK signifie que le cycle s’est terminé…
  • Page 123
    FRANÇAIS MESSAGES D’ERREUR ET D’ALARME Les messages d’erreur sont signalés par un code alphanumérique composé d’une lettre et de 3 chiffres. Le ”préfixe E” concerne les erreurs de l’opérateur et peut être rétablie par l’opérateur lui- même. La “lettre A” concerne des alarmes relatives à l’autoclave. Si le problème persiste après avoir effectué…
  • Page 124
    FRANÇAIS Porte ouverte pendant le cycle ou problème Contrôler le système de A 401 de fermeture fermeture A 501 Pression non maintenue Effacer et répéter le cycle A 551 Anomalie 1° capteur température Effacer et répéter le cycle A 552 Effacer et répéter le cycle Anomalie 2°…
  • Page 125
    FRANÇAIS PROGRAMMATION DE L’ECRAN Faire référence à la FIG.2. 14.1 Sélection langue Appuyez sur B-START et simultanément sur l’interrupteur (MAINS). Appuyez B-STARTjusqu’à ce que la langue souhaitée s’affiche. La langue est modifiée [en avant] en appuyant sur B-SELECT. La langue est modifiée [en arrière] en appuyant sur B-PUMP 14.2 Sélection heure L’heure s’affiche en appuyant sur B-START.
  • Page 126
    FRANÇAIS 14.7 Sélection année L’année s’affiche en appuyant sur B-DOOR. L’année est modifiée [en avant] en appuyant sur B-SELECT. L’année est modifiée [en arrière] en appuyant sur B-PUMP 14.8 Affichage pression atmosphérique La pression atmosphérique enregistrée s’affiche en appuyant sur B-START. NE PAS MODIFIER ! 14.9 Sélection remplissage eau Appuyer sur le poussoir…
  • Page 127
    FRANÇAIS PETIT GUIDE DE DÉPANNAGE Dans la plupart des cas, certaines alarmes ou erreurs sont provoquées par une inattention ou une ignorance de certains aspects techniques et opérationnels. Ce paragraphe décrit les problèmes plus fréquents pouvant surgir et les solutions pour y remédier 15.1 L’autoclave ne sèche pas correctement •…
  • Page 128
    FRANÇAIS 15.6 L’imprimante de l’autoclave n’imprime pas • le volet du porte-rouleau n’est pas bien ferme, la lettre “D” s’affiche. ouvrez et refermez le volet de l’imprimante en vérifiant sa fermeture. • plus de papier, la lettre “P” s’affiche. Ouvrez le volet et introduisez le rouleau de papier thermosensible.
  • Page 129
    L’autoclave est prédisposé aussi pour le chargement de l’eau déminéralisé par le SYSTEME IONIQUE ou SYSTEME D’OSMOSE (accessoires en option). Ces systèmes sont accessoires des autoclaves TECNO-GAZ S.p.A. (pas en dotation). Ces produits sont utilises pour éliminer les sels minéraux de l’eau qui vient du réseau hydrique pour alimenter l’autoclave.
  • Page 130
    Au debut du cycle de sterilisation, l’autoclave signale E250, il signifie que on a demarre le cycle avec le niveau d’eau pure bas; donc, il est necessaire remplir le reservoir d’eau pure. NOTE: pour le branchement du déminéralisateur au autoclaves TECNO-GAZ, veuillez-vous reporter aux indications du manuel d’instructions correspondant…
  • Page 131
    FRANÇAIS FIG.31 16.3 Changement cartouche resines du demineralisateur L’autoclave signale sur le display l ‘état de déplétion, des résines ou du filtre, avec le message “ECHANGER RESINES“ (code 242-S) “ECHANGER FILTRE“ (code 248-S-2). Changez la cartouche de résine comme illustré dans le manuel du déminéralisateur.
  • Page 132
    L’autoclave est le maillon final de ce système. Le non respect d’une des phases du cycle de stérilisation peut nuire au résultat final. TECNO-GAZ S.p.A. fabrique tous les dispositifs pour chaque phase du cycle de stérilisation. Il faut : 1 — Prévoir un protocole opérationnel écrit précis à diffuser auprès de tout le personnel préposé…
  • Page 133
    FRANÇAIS INSTRUCTIONS POUR LE RETOUR DE L’AUTOCLAVE En cas de retour au fabricant de l’autoclave pour réparation, révision ou validation, suivez les indications suivantes pour son expédition : • Utilisez l’emballage d’origine, s’il n’est plus en votre possession, utilisez un emballage adéquat.
  • Page 134
    FRANÇAIS PROCÉDURES A SUIVRE POUR BÉNÉFICIER DU SAV Le département réparation est une entité autonome de MEDILINE ITALIA SRL (TECNO- GAZ GROUP), qui se compose d’une structure articulée et professionnelle, conformément aux procédures indiquées dans notre “Manuel Qualité”. • Lorsque l’autoclave arrive chez MEDILINE ITALIA s.r.l. elle est réceptionnée directement par l’atelier de réparation qui le codifie et crée une fiche d’assistance technique.
  • Page 135
    4) LIMITES : la garantie donne droit au remplacement ou à la réparation gratuits des composants défectueux. Le remplacement de l’appareil tout entier reste toutefois exclu. Pour les composants installés ou intégrés fabriqués sous la marque TECNO-GAZ ou par des tiers fournissant leur propre certificat de garantie, ce sont les conditions, les limites et les exclusions indiquées sur ces certificats qui sont valables.
  • Page 136
    à des causes non imputables à des défauts de fabrication ; b) l’installation n’est pas conforme aux instructions de TECNO-GAZ et a été effectuée par du personnel non autorisé ;…
  • Page 137
    FRANÇAIS ENTRETIEN Un entretien correct de l’autoclave garantit son bon fonctionnement et un gain assuré en termes de pertes de temps et de coûts dus à l’intervention du SAV et de l’entretien. Les suivantes opérations sont obligatoires et à exécuter par l’opérateur. Nettoyage de la chambre Chaque 20 cycles ou 1 fois par semaine Nettoyer périodiquement la chambre et le filtre, et enlever tout éventuel dépôt ou débris,…
  • Page 138
    FRANÇAIS Régulation de la porte Chaque 2 mois Pour régler la pression de fermeture de la porte, agissez sur le régulateur dudit porte avec la clé à double fonction (FIG.10) fournie en dotation (code DANA008). Tournez dans le sens opposé des aiguilles d’une montre (pos.A) pour augmenter la pression de fermeture.
  • Page 139
    FRANÇAIS Les stivante opérations sont conseillées par le producteur. Ces operations doivent etre executes par des techniciens formés et autorisés par le producteur. Remplacement du filtre chargement eau Chaque 1 an / 1000 cycles* Controle du filtre eau entrée fluximetre Chaque 1 an / 1000 cycles* Nettoyage filtre laiton Y en aval au radiateur Chaque 1 an / 1000 cycles*…
  • Page 140
    FRANÇAIS En plus quelques dispositions régionales/nationales, établient la condition obligatoire d’une validation de l’appareil, qu’en controle les paramètres de stérilisation. La validation périodique doit etre toujours faite par des techniciens spécialisés avec des instruments de mesures aptes. Dans le cas que la validation ne soit pas obligatoire, la validation meme devient une opération volontaire avec le but de certifier que l’autoclave est parfaitement fonctionnante et apte à…
  • Page 141
    ESPAÑOL Estimado Cliente, Les agradecemos por elegir nuestro autoclave y sepan que sabremos corresponder a su confianza, con la máxima atención y un servicio adecuado a sus expectativas. El autoclave que eligieron es la máxima expresión de tecnología aplicada. Puede utilizarse en el sector dental, médico, estético y, por lo general, en todos los sectores que requieren la esterilización segura de todo tipo de instrumentos y materiales: sólidos, huecos, porosos, libres y envasados.
  • Page 142
    ESPAÑOL ÍNDICE CONSULTA DEL MANUAL GLOSARIO DESCRIPCIÓN VISTA DELANTERA / TRASERA SEGURIDAD MARCACIÓN DE SEGURIDAD DISPOSITIVO DE SEGURIDAD NOTAS SOBRE LA SEGURIDAD ELIMINACIÓN FINALIDAD DE EMPLEO DEL AUTOCLAVE DATOS TÉCNICOS MECÁNICOS ELÉCTRICOS CÁMARA DEPÓSITO AGUA LIMPIA DEPÓSITO AGUA USADA FILTRO BACTERIOLÓGICO ACCESORIOS DESEMBALADO INSTALACIÓN…
  • Page 143
    ESPAÑOL 10.4 TEST DE HÉLICE 10.5 TEST DE VACÍO 11. LAS TABLAS Y LAS DIAGRAMAS 12. LECTURA DEL INFORME DE CICLO 13. MENSAJES DE ERROR Y ALARMAS 14. PROGRAMACIÓN DE PARÁMETROS DISPLAY 14.1 SELECCIÓN IDIOMA 14.2 SELECCIÓN HORA 14.3 SELECCIÓN MINUTOS 14.4 SELECCIÓN SEGUNDOS 14.5…
  • Page 144
    ESPAÑOL CONSULTA DEL MANUAL Glosario VACÍO Presión inferior a la presión atmosférica CELDA Cuerpo cilíndrico de acero que aloja los dispositivos a esterilizar FILTRO Dispositivo apto a retener las partículas TRAY Cajón de soporte de la carga a esterilizar PORTATRAY Estructura de soporte de las bandejas en la celda de esterilización Dispositivo apto para eliminar los sales minerales del agua DESMINERALIZADOR…
  • Page 145
    ESPAÑOL FIG.2…
  • Page 146
    ESPAÑOL Filtro Bacteriológico Carga manual de Agua Llenado con agua Drenaje agua Usada Display Impresora Conexión RS232 Incubadora de conector biológico B-TEST Interruptor general Ciclo 121° C (válida para materiales envasado e no envasado) Ciclo 134° C (válida para materiales envasado e no envasado) Ciclo 134°…
  • Page 147
    ESPAÑOL SEGURIDAD Marcación de Seguridad TENSIÓN PELIGROSA QUITAR TENSIÓN ANTES DE RETIRAR LA TAPA ALTA TEMPERATURA CONEXIÓN DE TIERRA Dispositivos de Seguridad Los Dispositivos de Seguridad son los siguientes: ± -) Válvula de Seguridad calibrada a 2.4 bar -) Bloqueo electromagnético para impedir la apertura de la tapa durante la ejecución del ciclo -) Termostato de seguridad Notas sobre la Seguridad •…
  • Page 148
    TIPO A: turbinas, empuñaduras y dispositivos con orificios ciegos o de pequeñas dimensiones El autoclave ha sido certificado para la esterilizaciòn de los siguientes materiales*: EUROPA B EVO EUROPA B EVO 24 Materiales solidos ensobrados y no max kg. 5 max kg. 7…
  • Page 149
    ESPAÑOL DATOS TECNICOS Mecánicos EUROPA B EVO EUROPA B EVO 24 +5° C ÷ +40° C Temperatura de trabajo Altitud MAX 2.000 m Humedad relativa MAX a 30° C Humedad relativa MAX a 40° C Dimensiones totales (L x H x P)
  • Page 150
    ESPAÑOL Depósito Agua Limpia EUROPA B EVO EUROPA B EVO 24 Senza Con demineralizz. demineralizz. 4,5 l Volumen Leer impostaciòn del Ciclos efectuables sistema de demineralizaciòn polietileno Material Depósito Agua Usada EUROPA B EVO EUROPA B EVO 24 Volumen 4,5 l…
  • Page 151
    ESPAÑOL ACCESORIOS Portabandejas EUROPA B EVO EUROPA B EVO 24 Material Aluminio Anodizado Dimensiones (L x H x P) 192 x 165 x 280 mm 192 x 200 x 470 mm Cuadro FIG.3 FIG.4 Dotación prevista Portabandeja (opcional) EUROPA B EVO…
  • Page 152
    ESPAÑOL Bandejas (opcional) EUROPA B EVO EUROPA B EVO 24 Aluminio Anodizado / de Material acero inoxidable Dimensiones (L x H x P) 184 x 17 x 460 mm Cuadro FIG.9 Dotación prevista Llave para extracción del tray y regulación de la portezuela Utilízala para extraer y manejar los trays (FIG.11) y también para regular la portezuela…
  • Page 153
    ESPAÑOL Racor para carga de agua Utilízalo con el embudo de plástico con tubo para cargar el agua en manual y con el tubo para carga de agua completo con filtro Cuadro FIG.14 Dotación prevista Tubo para carga de agua completo con filtro Utilízalo para cargar el agua en manual (apartado 8.2) Cuadro FIG.15…
  • Page 154
    ESPAÑOL Patilla separadora de plástico posterior Aplica la patilla separadora de plástico en la parte posterior del autoclave (FIG.1–pos.0) para asegurar la ventilación adecuada en caso de que se posicione la autoclave demasiado cerca de una pared Cuadro FIG.18 Dotación prevista Tubos de descarga posterior de dependencias 1- Tubo para rebosadero agua limpia: Une un extremo del tubo al rebosadero posterior del agua limpia (FIG.1–pos.1), y el otro extremo a un contenedor para recuperación del agua…
  • Page 155
    ESPAÑOL Cable de alimentación eléctrica Une el extremo del conector (FIG.21–pos.B) al panel trasero (FIG.1–pos.6) y seguidamente enchufa la clavija (FIG.21–pos.A) directamente en la toma de alimentación del sistema eléctrico. Cuadro FIG.21 Dotación prevista DESEMBALADO El autoclave será expedido en embalaje idóneo que protegerá el contenido y permitirá su fácil transporte y manejo.
  • Page 156
    ESPAÑOL INSTALACIÓN FIG.22 EUROPA B EVO EUROPA B EVO 24 610 mm 590 mm 730 mm 50 mm 410 mm 426 mm (trasero) 410 mm (trasero) 455 mm (delatero) 444 mm (delatero) 400 mm 515 mm 1 — Instalar el autoclave en ambientes idóneos para la esterilización.
  • Page 157
    ESPAÑOL 8 — Dejar en el interior de la cámara de esterilización solo el portabandeja con las bandejas, todos los demás accesorios deben ser guardados en el compartimiento externo a disposición de los utilizadores. — Nunca deben dejarse sobre la máquina periódicos, bandejas, contenedores con líquidos, etc ……
  • Page 158
    ESPAÑOL INSTRUCCIONES PARA EL USO Después de instalada, el autoclave deberá ser preparada para la utilización. Accionamento del autoclave Pulsando el interruptor general (FIG.2–pos.19) Llenado depósito agua limpia Conecta el “racor para carga de agua “ FIG.14 al “tubo para carga de agua completo con filtro”…
  • Page 159
    ESPAÑOL Caracteristicas del agua utilizada TABLA NIVELES CUALITATIVOS ESTABLECIDOS POR LA NORMATIVA UNI EN 13060 : 2009 CEN STANDARD UNI EN 13060 : 2009 ≤ Residuo evaporazione mg/l ≤ Ossido di silicio mg/l ≤ Ferro mg/l ≤ Cadmio 0.005 mg/l ≤…
  • Page 160
    ESPAÑOL Inicio ciclo de esterilización Elegir el programa de esterilización más idóneo para la carga introducidapulsando el botón B-SELECT Cada opresión del botón alterará la selección del programa pasando a la opción sucesiva. Después de elegir el programa será posible accionar el ciclo pulsando el botón B-START, la tapa se bloqueará…
  • Page 161
    ESPAÑOL Interrupción de un ciclo de esterilización Un ciclo de esterilización podrá ser interrumpido voluntariamente pulsándose el botón B-START por aproximadamente 2 segundos. El autoclave emitirá un sonido y pasará a la descompresión mostrando en el display el mensaje de error A001 (CICLO INTERRUMPIDO). Para reponer a cero el error deben mantenerse oprimidos simultáneamente los botones que destacamos a continuación hasta que intervenga la señal acústica: B-SELECT…
  • Page 162
    ESPAÑOL CICLOS DE ESTERILIZACIÓN Descripción de los ciclos El autoclave dispone de tres series de ciclos: A — ciclos operativos B — ciclos nocturnos C — ciclos de testes Ciclos operativos Todos los ciclos operativos disponen de un sistema de vacío fraccionado y por lo tanto pueden esterilizar materiales huecos, porosos, sólidos, libres o envasados.
  • Page 163
    ESPAÑOL TEST DE CONTROL DEL AUTOCLAVE 10.1 Integradores químicos Este tipo de test utiliza las propiedades de sustancias colorantes capaces de modificar su aspecto al exponerlas al calor y a la presión durante los tiempos adecuados ya que interactúan con todos los parámetros del ciclo (presión, temperatura, tiempo). Al superar el test se demuestra que la estabilidad es correcta.
  • Page 164
    ESPAÑOL Las ampollas deberán ser introducidas en la cámara de esterilización, incluso cuando esté llena. Realizar el ciclo y, una vez finalizado, extraer la ampolla prestando mucha atencíon ya que estará caliente y bajo presión. Dejarla enfriar durante unos 10 minutos y, a continuación, activarla utilizando la pinza especial proporcionada y manteniéndola siempre en posición vertical.
  • Page 165
    ESPAÑOL La evaluación del resultado es RESPUESTA sencilla y rápida. Si el aspecto es uniforme (tal y como se indica en la imagen) significa que el test es positivo, caso contrario autoclave puede esterilizar correctamente los cuerpos porosos. No existen intervalos de tiempo predefinidos, con excepción de las FRECUENCIA zonas reguladas por leyes regionales específicas.
  • Page 166
    No existen intervalos de tiempo predefinidos, con excepción de las zonas reguladas por leyes regionales específicas. Para poder FRECUENCIA mantener un control constante de la validez del autoclave, se recomienda realizar el test por lo menos una vez a la semana. CONSEJAMOS EL UTILIZO DE LAS PRUEBAS TECNO-GAZ S.P.A.
  • Page 167
    ESPAÑOL LAS TABLAS Y LAS DIAGRAMAS CUADRO DE ESTERILIZACIÓN TIEMPO TIEMPO RANGO DE RANGO DE SECADO PRESIÓN TEMPERATURA CICLOS T5 (Minuti) DE TRABAJO DE TRABAJO (°C ) (Minuti) (bar relativos) 121° C 1.04 ÷ 1.30 121 ÷ 125 (envasado e no envasado) 134°…
  • Page 168
    ESPAÑOL LECTURA DEL INFORME DE CICLO Versión del software Número de serie Descripción del ciclo seleccionado (con los valores principales de configuración) Horas – Minutos – Segundos Presión de la cámara Temperatura principal de la cámara (T1) Fase del ciclo Fin del ciclo (OK significa que el ciclo ha finalizado correctamente)
  • Page 169
    ESPAÑOL MENSAJES DE ERROR Y ALARMAS Los mensajes de error estarán destacados por un código alfanumérico compuesto de una letra y de 3 cifras. El sufijo “E” está realcionado con errores del operador que el mismo operador puede restablecer. El sufijo “A” está relacionado con alarmas, anomalías del autoclave; si el problema persiste tras la reparación debe contactarse con la asistencia telefónica.
  • Page 170
    ESPAÑOL Tapa abierta durante el ciclo o problema de A 401 Controlar sistema de cierre cierre A 501 Presión no mantenida Reiniciar y repetir el ciclo A 551 Anomalía 1° sensor temperatura Reiniciar y repetir el ciclo A 552 Anomalía 2° sensor temperatura Reiniciar y repetir el ciclo A 641 Superpresión…
  • Page 171
    ESPAÑOL PROGRAMACIÓN DE PARÁMETROS DISPLAY Consulta la FIG.2. 14.1 Selección Idioma Pulsar B-START y simultáneamente el interruptor (MAINS). Mantener pulsado B-START hasta que aparezca en el display el nombre del idioma requerido. Pulsando B-SELECT se modificará el idiomai [avance]. Pulsando B-PUMP se modificará…
  • Page 172
    ESPAÑOL 14.7 Selección Año Pulsando B-DOOR aparece el año. Pulsando B-SELECT modificase el año [hacia adelante]. Pulsando B-PUMP modificase el año [hacia atrás]. 14.8 Visualización Presión Atmosférica Pulsando B-START aparece la presión atmosférica almacenada. NO MODIFICABLE ! 14.9 Selección Carga Agua Pulsando B-START aparece la programación del llenado del agua:…
  • Page 173
    ESPAÑOL SOLUCIÓN DE PROBLEMAS OPERATIVOS En muchos casos, algunas de las alarmas o de los errores estarán originados por la desatención o el desconocimiento de algunos apsectos técnicos y operativos. A continuación relataremos algunos de los casos de anomalías con respectivas soluciones. 15.1 El autoclave no efectua secado correcto •…
  • Page 174
    ESPAÑOL 15.6 La impresora del autoclave no imprime • la tapa del portacilindro no está cerrada correctamente, en el display aparece la letra “D”. Abrir y cerrar la tapa de la impresora comprobando su correcto cierre. • falta el papel, aparece en el display la letra “P”. Abrir la tapa e introducir el rollo de papel termográfico.
  • Page 175
    SISTEMA IONICO o SISTEMA OSMOSIS. (accesorios opcionales). Estos sistemas son accesorios de los autoclaves TECNO-GAZ S.p.A. (no en dotación). Ese producto se utiliza para eliminar las sales minerales del agua procedente de la red hídrica para alimentar el autoclave.
  • Page 176
    NOTA: para la conexión los sistemas de desmineralizaciòn a autoclaves TECNO-GAZ, consulte lo indicado en el manual de desmineralizador.
  • Page 177
    ESPAÑOL FIG.32 16.3 Cambio cartucho resina – filtre El autoclave pone en el display el esaurimento de las resinas o del filtre con el mesate “SUBSTITUCIÒN RESINAS” (codico de repuesto 242-S) o “SUBSTITUCIÒN FILTRES” (codico de repuesto 248-S-2). Sustituir el cartucho de resina conforme lo indicado en el manual de desmineralizador; Terminada la sustitución debe pulsarse el botón B-PUMP hasta la intervención de la…
  • Page 178
    El autoclave es el eslabón final de ese sistema. El que no efectúe todas las fases del sistema de esterilización perjudicará el resultado final. TECNO-GAZ S.p.A. produce todos los dispositivos para todas las fases del ciclo de esterilización. Se requiere por lo tanto: 1 — Predisponer un protocolo operativo por escrito preciso y distribuirlo entre todos los operadores y encargados de la esterilización.
  • Page 179
    ESPAÑOL INSTRUCCION PARA EXPEDICIÓN DEL AUTOCLAVE Cuando haya la necesidad de expedirse el autoclave para reparaciones, control, rearme, revisión, validación, sigan obligatoriamente las instrucciones a continuación: • Utilizar el embalaje original, si ese embalaje ya no se encuentra en las manos del cliente, deberá…
  • Page 180
    PROCEDIMIENTOS PARA EL SERVICIO Y ASISTENCIA El departamento encargado de las reparaciones constituye una sección autónoma de MEDILINE ITALIA SRL (TECNO-GAZ GROUP) y su organización se basa en una estructura compleja y profesional, en conformidad con nuestros procedimientos indicados en el “Manual de Calidad”.
  • Page 181
    Queda excluido el derecho a la sustitución de todo el aparato. En lo referente a los componentes aplicados o integrados con productos con marca TECNO-GAZ, o de terceros, y acompañados por su certificado de garantía, son válidas las condiciones, límites y exclusiones indicadas en tales certificados.
  • Page 182
    TECNO-GAZ S.p.A. y si fue llevada a cabo por personal no autorizado; c) el equipo es reparado, modificado o, en algún modo, alterado por el Comprador o por terceros no autorizados;…
  • Page 183
    ESPAÑOL MANTENIMIENTO El correcto mantenimiento del autoclave garantizará su buen funcionamiento y el ahorro seguro de tiempo y costes debidos a asistencia y mantenimiento. Las seguente operaciones son obligatorias por los operadores Limpieza de la cámara 20 ciclos o una vez a la semana Limpia periódicamente la cámara y el filtro eliminando posibles depósitos o detritos para evitar así, introducir en el circuito de desagüe, materiales que puedan crear obstrucciones.
  • Page 184
    ESPAÑOL Regulaciòn puerta Cada 2 meses Para regular la presión de cierre de la puerta debe actuarse con el regulador de la puerta utilizando la llave de doble función (FIG.10) en dotación. Girar a izquierdas (pos.A) para aumentar la presión de cierre. Girar a derechas (pos.B) para reducir la presión de cierre.
  • Page 185
    ESPAÑOL Las siguente operaciones son consejadas de parte del productor y tienen que ser realizada por tecnicos autorizados. Substitucion filtre y carga agua Cada año / cada 100 ciclos * Control filtre agua entrada fluximetro Cada año / cada 100 ciclos * Limpieza filtre Y de latòn a bajo del radiador Cada año / cada 100 ciclos * Limpieza elecrtovalvula carga agua E4…
  • Page 186
    ESPAÑOL NOTAS…
  • Page 187
    Ort aufzubewahren. Der Autoklav darf nur für den vom Hersteller vorgesehenen Einsatz verwendet werden. Für Installationsarbeiten, die Instandhaltung und den Service wenden Sie sich ausschließlich an von der Firma TECNO-GAZ autorisierte Techniker. Bitte verwenden und bestellen Sie nur Original-Ersatzteile der TECNO-GAZ…
  • Page 188
    DEUTCH INHALT DAS HANDBUCH KONSULTIEREN GLOSSAR BESCHREIBUNG DER VORDER-/RÜCKSEITE SICHERHEIT SICHERHEITSKENNZEICHNUNG SCHUTZEINRICHTUNGEN ANMERKUNGEN ZUR SICHERHEIT ENTSORGUNG VERWENDUNG UND GEBRAUCHSZIEL DES AUTOKLAVS TECHNISCHE DATEN MECHANISCHE DATEN ELEKTRISCHE DATEN KAMMER REINWASSERTANK SCHMUTZWASSERTANK BAKTERIOLOGISCHER FILTER ZUBEHÖR DEN AUTOKLAV AUSPACKEN INSTALLATION GEBRAUCHSANWEISUNGEN DEN AUTOKLAV EINSCHALTEN DEN REINWASSERTANK FÜLLEN EIGENSCHAFTEN DES ZU VERWENDENDEN WASSER DAS MATERIAL IN DEN AUTOKLAVEN GEBEN…
  • Page 189
    DEUTCH 10.4 HELIXTEST 10.5 VAKUUMTEST 11. TABELLEN UND DIAGRAMME 12. LESEN DES ZYKLUSBERICHTS 13. FEHLERMELDUNGEN UND ALARME 14. PROGRAMMIERUNG UND DISPLAY-PARAMETER 14.1 DIE SPRACHE EINSTELLEN 14.2 DIE UHRZEIT EINSTELLEN — STUNDE 14.3 DIE UHRZEIT EINSTELLEN — MINUTEN 14.4 DIE UHRZEIT EINSTELLEN — SEKUNDEN 14.5 DAS DATUM EINSTELLEN — TAG 14.6…
  • Page 190: Das Handbuch Konsultieren

    DEUTCH DAS HANDUBUCH KONSULTIEREN Glossar VAKUUM Druck, dessen Wert unter dem Luftdruck liegt Zylindrischer Körper aus Stahl in den die zu sterilisierenden Geräte KAMMER gegeben werden FILTER Vorrichtung, die Partikel zurückhält TABLETT Gestell, auf das das zu sterilisierende Material gelegt wird TABLETTHALTER Halterung für die Tabletts in der Sterilisationskammer ENTSALZER…

  • Page 191
    DEUTCH FIG.2…
  • Page 192
    DEUTCH Bakteriologischer Filter Manueller Wassereinlauf Wassereinlauf Schmutzwasserablauf Display Drucker Anschluss RS232 Stecker B-TEST Inkubator Hauptschalter 121° C Zyklus (geeignet für verpacktes und unverpacktes Material) 134° C Zyklus (geeignet für verpacktes und unverpacktes Material) 134° -Zyklus Flash Prion-Zyklus Vacuum Test Helix – B&D Test Wassermindeststand Wasserhöchststand Höchststand Brauchwasser…
  • Page 193: Sicherheit

    DEUTCH SICHERHEIT Sicherheitskennzeichnung GEFÄHRLICHE SPANNUNG VOR ABNAHME DES DECKELS STROMZUFUHR UNTERBRECHEN HOHE TEMPERATUR ERDUNG Schutzeinrichtungen Folgenden Schutzeinrichtungen sind am Gerät installiert: -) Schutzventil geeicht auf 2.4 bar 0/+10% -) Elektromagnetische Sperre, verhindert das Öffnen der Tür während dem Zyklus -) Sicherheitsthermostat Anmerkungen zur Sicherheit •…

  • Page 194: Entsorgung

    Blindöffnungen oder kleinen Öffnungen Der Autoklav kann im Zahn- Arzt- und Kosmetiksbereich und im Allgemeinem in allen Instruments- und Materialsterilizationsbereichen eingesetzt werden: EUROPA B EVO EUROPA B EVO 24 Eisenmaterial oder festkörper max kg. 5 max kg. 7 Körper mit hohlkörper max kg.

  • Page 195: Technische Daten

    DEUTCH TECHNISCHE DATEN Mechanische Daten EUROPA B EVO EUROPA B EVO 24 +5° C ÷ +40° C Betriebstemperatur Maximale Meereshöhe 2.000 m Maximale relative Feuchte bei 30° C Maximale relative Feuchte bei 40° C Raumbedarf (L x H x P)

  • Page 196: Reinwassertank

    DEUTCH Reinwassertank EUROPA B EVO EUROPA B EVO 24 Ohne Wasseraufbereitung Wasseraufbereitung 4,5 l Volumen Bitte s. die Ausführbare Zyklen Gebrauchsanleitung von Wasseraufbereitung Material Polyäthylen Schmutzwassertank EUROPA B EVO EUROPA B EVO 24 Volumen 4,5 l Ausführbare Zyklen Material Polyäthylen Maximale 50°…

  • Page 197: Zubehör

    DEUTCH ZUBEHÖR Tablettaufnahmegestell EUROPA B EVO EUROPA B EVO 24 Material Eloxiertem aluminium Abmessungen (L x H x P) 192 x 165 x 280 mm 192 x 200 x 470 mm Abbildung FIG.3 FIG.4 Standardausstattung Portatray (opzionali) EUROPA B EVO…

  • Page 198
    DEUTCH Zusätzliches Tablettaufnahmegestell EUROPA B EVO EUROPA B EVO 24 Eloxiertem aluminium oder Material rostfreies stal Abmessungen (L x H x P) 184 x 17 x 460 mm Abbildung FIG.9 Standardausstattung Schlüssel für das Herausnehmen des Tabletts und zur Türeinstellung Für das Herausnehmen und die Handhabung der Tabletts (FIG.11) und um die Tür…
  • Page 199
    DEUTCH Anschlussstück Wasserzulauf Zusammen mit dem Plastikschlauch für die manuelle Wasserbefüllung und mit dem mit dem Filter bestückten Wasserzulaufschlauch zu verwenden. Abbildung FIG.14 Standardausstattung Wasserauffüllschlauch mit Filter Für das Einfüllen des Wassers von Hand (Absatz 8.2) Abbildung FIG.15 Standardausstattung Wasserablassschlauch Für das Ablassen des Brauchwassers (Absatz 8.9) Abbildung FIG.16…
  • Page 200
    DEUTCH Abstandsfuß für die Geräterückseite aus Plastik Den Abstandsfuß aus Plastik auf der Rückseite des Autoklaven anbringen (FIG.1–pos.0) damit eine angemessene Belüftung gewährleistet ist, auch wenn der Autoklav nahe einer Wand aufgestellt wird. Abbildung FIG.18 Standardausstattung Schläuche für den rückseitigen Ablass der Verbraucher 1- Überlaufschutz Reinwasser : Ein Ende des Schlauchs an den hinteren Überlaufschutz Reinwasser (FIG.1–pos.1), anschließen, das andere Ende an einen Auffangbehälter für entsalzenes Wasser oder Abwasser (Ausführung mit Entsalzer) anschließen.
  • Page 201: Den Autoklav Auspacken

    DEUTCH Speisekabel Das Ende der Steckers (FIG.21–pos.B) zu hintere Tafel (FIG.1–pos.6) anschließen und anschließend den Stecker (FIG.21–pos.A) direkt an die Versorgungsbuchse der elektrischen Anlagen anschließen. Abbildung FIG.21 Standardausstattung DEN AUTOKLAV AUSPACKEN Der Autoklav wird angemessen verpackt, sodass er bequem transportiert und verlagert werden kann und der Packungsinhalt geschützt ist.

  • Page 202: Installation

    DEUTCH INSTALLATION FIG.22 EUROPA B EVO EUROPA B EVO 24 610 mm 590 mm 730 mm 50 mm 410 mm 426 mm (rückseitig) 410 mm (rückseitig) 455 mm (vorderseitig) 444 mm (vorderseitig) 400 mm 515 mm 1 — Den Autoklav an einem für die Sterilisierung angemessenen Ort installieren.

  • Page 203
    DEUTCH 8 — In der Sterilisierkammer nur das Tablettaufnahmegestell mit den Tabletts lassen. Das gesamte Zubehör ist in einem dazu bestimmten und den Bediener zugänglichem Fach unterzubringen. 9 — Auf dem Autoklaven dürfen keine Zeitungen, Schalen, Behälter mit Flüssigkeiten, usw. abgestellt werden.
  • Page 204: Gebrauchsanweisungen

    DEUTCH GEBRAUCHSANWEISUNGEN Nachdem der Autoklav installiert wurde, kann er angewendet werden. Den autoclav einschalten Drücken den Hauptschalter (FIG.2–pos.19) Den reinwassertank fullen An den “Wasserzulaufschlauch mit Filter” (mitgeliefert) das “Wasserzulaufanschlussstück“ FIG.14, anschließen und dann in die Öffnung des frontalen Anschlusses am Autoklaven (FIG.2–pos.13) stecken.

  • Page 205
    DEUTCH Eigenschaften des zu verwendeten wassers TABELLE DER VON DER RICHTLINIE UNI EN 13060: 2009 FESTGELEGTEN QUALITÄTSSTANDARDS CEN STANDARD UNI EN 13060 : 2009 ≤ Evaporationsrückstände mg/l ≤ Siliziumoxyd mg/l ≤ Eisen mg/l ≤ Kadmium 0.005 mg/l ≤ Blei 0.05 mg/l Schwermetallrückstände, ausgenommen Eisen, ≤…
  • Page 206: Den Sterilisierzyklus Starten

    DEUTCH Den sterilisierzyklus starten Das der Ladung entsprechende Sterilisierprogramm anwählen, dazu die Taste B-SELECT drücken. Bei jedem Drücken der Taste schaltet die Anwahl auf das folgende Programm. Nach der Programmanwahl mit der Taste B-START den Zyklus starten; die Tür wird automatisch blockiert und der Zyklus startet.

  • Page 207: Einen Sterilisierzyklus Unterbrechen

    DEUTCH Einen sterilisierzyklus unterbrechen Ein Sterilisierzyklus kann unterbrochen werden, dazu die Taste B-START mindestens 2 Sekunden lang drücken. Der Autoklav gibt einen Signalton ab, schaltet aus Unterdruck und es erscheint die Fehlermeldung A001 (ZYKLUS UNTERBROCHEN). Um den Fehler rückzusetzen, die unten angezeigten Tasten zugleich drücken bis der Signalton ertönt: B-SELECT B-PUMP…

  • Page 208: Sterilisierzyklus

    DEUTCH STERILISIERZYKLUS Beschreiburg der zyklen Der Autoklav besitzt drei Zyklusserien: A — Arbeitszyklen B — Arbeitszyklen — Nachtzyklen C — Testzyklen Arbeitszyklen Alle Arbeitszyklen arbeiten mit dem System des fraktionierten Vakuums, daher können mit diesen Zyklen Hohlkörper, poröse Körper, Feststoffe, verpacktes und unverpacktes Material sterilisiert werden.

  • Page 209
    DEUTCH KONTROLLTESTS FÜR AUTOKLAV 10.1 Chemische Zusätze Dies sind Tests, welche die Eigenschaften von Färbemitteln nutzen, die ihr Aussehen verändern, wenn sie der Wärme und dem Druck in entsprechenden Zeiten angemessen ausgesetzt werden. Da sie mit allen Zyklusparametern (Druck, Temperatur, Zeit) interagieren. Das Bestehen des Tests bescheinigt die korrekte Stabilität.
  • Page 210
    DEUTCH Die Ampullen müssen auch bei vorhandener Ladung in die Sterilisierungskammer gelegt werden. VERWENDEN Zyklus durchführen, Ende Ampulle vorsichtig herausnehmen, da sie warm ist und unter Druck steht. Sie circa 10 Minuten abkühlen lassen, danach unter Verwendung der entsprechenden mitgelieferten Schlüssel aktivieren, indem sie stets senkrecht gehalten wird.
  • Page 211
    DEUTCH Die Auswertung ist schnell und ANTWORT einfach. Wenn Tonung einheitlich (wie Abbildung), war der Test erfolgreich, andernfalls wurde der Test nicht bestanden und der Autoklav ist also nicht in der Lage, poröse Körper korrekt zu sterilisieren. Es besteht keine vorbestimmte Regelmäßigkeit, es sei denn in jenen Gebieten, in denen es spezifische regionale Gesetze gibt.
  • Page 212
    DEUTCH TEST AM BEGINN ANTWORT TEST NICHT ERFOLGREICH TEST ERFOLGREICH Es besteht keine vorbestimmte Regelmäßigkeit, es sei denn in jenen PERIODIZITÄT Gebieten, in denen es spezifische regionale Gesetze gibt. Auf allen Fällen wird für eine ständige Kontrolle der Wirksamkeit des Autoklavs die Durchführung des Tests mindestens einmal in 30 Tagen empfohlen.
  • Page 213: Tabellen Und Diagramme

    DEUTCH TABELLEN UND DIAGRAMME STERILISIERTABELLE TROCKNUNG PROZESS- BETRIEBSTEM BETRIEBSDRU S-ZEIT ZEIT PERATURBERE CYCLES CKBEREICH T5 (Minuten) ( relative Bar ) ( °C ) (Minuten) 121° C 1.04 ÷ 1.30 121 ÷ 125 (verpackt und unverpackt) 134° C 2.02 ÷ 2.38 134 ÷…

  • Page 214: Lesen Des Zyklusberichts

    DEUTCH LESEN DES ZYKLUSBERICHTS Softwareversion Seriennummer Beschreibung des gewählten Zyklus (mit den wichtigsten Setup-Werten) Stunden – Minuten – Sekunden Druck der Kammer Haupttemperatur der Kammer (T1) Zyklusphase Zyklusende (OK bedeutet, dass der Zyklus korrekt beendet wurde) Zyklusnummer Benutzer (Platz für die Unterschrift)

  • Page 215
    DEUTCH FEHLER- UND STÖRUNGSMELDUNGEN Die Fehlermeldungen werden mit einem alphanumerischen Code dargestellt, dieser besteht aus einem Buchstaben und 3 Ziffern. Der Buchstabe «E» bezieht sich auf Fehler durch den Bediener, die von diesem rückgesetzt werden können. Der Buchstabe «A» bezieht sich auf Störungen am Autoklaven. Sollte nach der Fehlerbehebung das Problem weiterhin auftreten, ist der Kundendienst anzurufen.
  • Page 216
    DEUTCH Tür offen während dem Zyklus oder A 401 Das Schließsystem überprüfen problem Tür- Schließsystem Rücksetzen und den Zyklus A 501 Druck nicht aufrechterhalten wiederholen Rücksetzen und den Zyklus A 551 Störung 1. Temperaturfühler wiederholen Rücksetzen und den Zyklus A 552 Störung 2.
  • Page 217: Die Sprache Einstellen

    DEUTCH PROGRAMMIERUNG DER DISPLAY-PARAMETER Siehe FIG.2 14.1 Die Sprache einstellen Die Taste B-START und den Schalter MAINS zugleich drücken. Die Taste B-START gedrückt halten bis auf dem Display die eingestellte Sprache erscheint. Mit der Taste B-SELECT wird auf die folgende Sprache geschalten. Mit der Taste B-PUMP wird auf die letzte Sprache zurückgeschaltet.

  • Page 218: Das Datum Einstellen — Jahr

    DEUTCH 14.7 Das Datum einstellen — Jahr Die Taste B-DOOR drücken, es erscheint das Jahr. Mit der Taste B-SELECT wird das Jahr verändert [vorwärts]. Mit der Taste B-PUMP wird das Jahr verändert [rückwärts]. 14.8 Luftdruckanzeige Die Taste B-START. Es erscheint der gespeicherte Luftdruck. NICHT VERÄNDERN! 14.9 Anwahl Wasserzulauf Wenn Sie…

  • Page 219: Behebung Auftretender Probleme

    15.1 Der autoklav trocknet nicht ordnungsgemass • • • • den bakteriologischen filter gegen einen neuen originalen der TECNO-GAZ ersetzen. • • • • es wurden keine originalen tabletts, sondern andersartige ohne oder mit einer andersartigen lochung verwendet Verwenden Sie nur Original-Tabletts, möglichst aus Aluminium.

  • Page 220: Flecken Auf Den Instrumenten

    DEUTCH 15.6 Der drucker des autoklaven drunkt nicht • die klappe der papierrolle ist nicht geschlossen, auf dem Display erscheint der Buchstabe «D». Öffnen Sie die Klappe. Schließen Sie nun die Klappe wieder und achten Sie darauf, dass diese richtig zu ist. •…

  • Page 221: Wasseraufbereitung

    DEUTCH WASSERAUFBEREITUNG 16.1 Einstellung von wasseraufbereitung systeme Der Autoklav ist so entworfen worden, dass das demineralisierte Wasser durch das IONIC SYSTEM oder das OSMOSIS SYSTEM einlaufen kann (Extra-Zubehör). Diese Systeme sind notwendig, um die Mineralsalzen zu entfernen, die im Stadtwasser enthalten sind.

  • Page 222: Verbindung Von Wasseraufbereitungsysteme

    GAZ AUTOKLAVEN, DIE DAS ERLAUBEN. WENN DER AUTOKLAV AM ZYKLUSBEGINN AUF E250 SCHALTET, BEDEUTET DAS, DASS EIN ZYKLUS MIT EINEM ZU GERINGEN REINWASSERSTAND GESTARTET WURDE. DAHER MUSS DER REINWASSERTANK GEFÜLLT WERDEN MERKE ANSCHLIESSEN WASSERAUFBEREITUNGSYSTEME TECNO-GAZ AUTOKLAVEN SIEHE ANGABEN IN DER GEBRAUCHSANLEITUNG DES AUTOKLAVEN.

  • Page 223: Wechseln Von Patrone Und Filter

    DEUTCH FIG.32 16.3 Wechseln von patrone und filter Der Autoklav zeigt im Display mit dem Nachricht PATRONE WECHSELN (Ersatzteil-Nr. 242-S) oder FILTER WECHSELN (Ersatzteil-Nr. 248-S-2), wenn die Patrone oder den Filter erschöpft sind . Nach dem Auswechseln, zum Rücksetzen des Autoklaven, die Taste B-PUMP drücken bis der Signalton ertönt.

  • Page 224: Standardvorgänge Sterilisierung

    Phasen Sterilisierungssystems nicht einhält, könnte das Endergebnis beeinträchtigen. TECNO-GAZ S.p.A. stellt alle Geräte für jede Phase des Sterilisierprozesses her. Dementsprechend ist Folgendes zu beachten: 1 — Ein präzises schriftliches Betriebsprotokoll anfertigen das allen Bedienern und mit der Sterilisierung beauftragten Personen ausgehändigt wird. TECNO-GAZ S.p.A. kann Ihnen Ihren Anforderungen angepasste Protokolle liefern.

  • Page 225
    DEUTCH STANDARDVORGÄNGE STERILISIERUNG Wenn der Autoklav für Reparaturen, Revisionen, Wiederherstellung der Funktionstüchtigkeit oder Gültigkeit verschickt werden muss, sind folgende Anweisungen unbedingt zu beachten: • Die Original-Verpackung verwenden. Sollte Ihnen diese nicht mehr zur Verfügung stehen, eine angemessene Verpackung verwenden. Der Absender trägt die Verantwortung für die Versendung der Ware.
  • Page 226: Anweisungen Für Den Versand Des Autoklaven

    DEUTCH ANWEISUNGEN FÜR DEN VERSAND DES AUTOKLAVEN Die Reparaturabteilung ist eine unabhängige Einheit von MEDILINE ITALIA srl (TECNO- GAZ GROUP), die sich aus einer gegliederten und professionellen Struktur laut unseren in unserem “Qualitätshandbuch” angegebenen Verfahren zusammensetzt. • Wenn der Autoklav bei der MEDILINE ITALIA SRL eintrifft, wird er direkt von der Reparaturabteilung angenommen.

  • Page 227: Garantie

    4) GARANTIEGRENZEN: Die Garantie verleiht Anspruch auf den kostenlosen Austausch oder die kostenlose Reparatur der defekten Bauteile. Es besteht kein Anspruch auf den Austausch des kompletten Geräts. Bezüglich der mit dem Markenzeichen TECNO-GAZ oder durch Dritte gefertigten Bauteile, die mit einem eigenen Garantieschein ausgestattet sind, gelten die in diesen Garantiescheinen angegebenen Bedingungen, Grenzen und Ausschließungen.

  • Page 228
    Flüssigkeiten, Blitzschlag, Naturereignisse oder Witterungseinflüsse oder auf andere nicht auf Fabrikationsmängel zurückführbare Schäden zurückzuführen sind; b) wenn die Installation nicht gemäß den Anleitungen von TECNO-GAZ erfolgt und durch nicht autorisiertes Personal durchgeführt wird; c) wenn das Gerät durch den Käufer oder nicht autorisierte Dritte repariert, abgeändert oder umgebaut wird;…
  • Page 229: Instandhaltung

    DEUTCH INSTANDHALTUNG Eine ordnungsgemäße Instandhaltung des Autoklaven sichert einen einwandfreien Betrieb und eine Zeit- und Kostenersparnis in Folge auf Service- und Wartungseingriffe. Die Folgenden Verfahren sind Pflicht, und sollen von Bedienern durchgeführt werden. Reiningung vom Kammer Alle 20 Zyklen oder alle Woche Die Kammer und der Filter sind regelmäßig zu reinigen.

  • Page 230
    DEUTCH Die regelmäßig die eventuellen Rückstände die sich an der Dichtung absetzen mit Wasser und dem mitgelieferten Schwamm (nicht abrasive Seite) oder mit einem feuchten Tuch entfernen (Ersatzteil-Nr. DANA038). Regelung von der Tür Alle 2 Monate Um den Schließdruck der Tür einzustellen, mit dem mitgelieferten Schlüssel mit Doppelfunktion (Tablettentnahme + Schlüssel) — (Ersatzteil-Nr.
  • Page 231
    DEUTCH Die folgende Verfahren werden vom Hersteller empfohlen. Sie müssen nur von geschulten und vom Hersteller freigegebenen Technikern durchgeführt werden. Umtausch vom Wassereinlauffilter Alle Jahre / Alle 1000 Zyklen* Überprüfung vom Wassereinlauffilter in Alle Jahre / Alle 1000 Zyklen* Durchflußmesser Reinigung vom Filter Y-förmig aus Messing Alle Jahre / Alle 1000 Zyklen* Reinigung von der EV4 — Wassereinlaufventil…
  • Page 232
    DEUTCH Einige Staat- oder Regionorm verschreiben die Validierung vom Gerät, die die Sterilisierungsparameter überprüft. Die o. g. Validierung soll immer von hoch spezialisierten Technikern durchgeführt werden, die geeignete Instrumenten benutzen. Falls die Validierung nicht obligatorisch ist, gilt als Erklärung, dass der Steri richtig funktioniert, und dass er für die Sterilisierung von den mitgeteilten Beladungen geeignet ist.
  • Page 234
    Il presente manuale deve sempre accompagnare il prodotto, in adempimento alle Direttive Comunitarie Europee. TECNO-GAZ, si riserva il diritto di apporre modifiche al presente documento senza dare alcun pre-avviso. La ditta TECNO-GAZ si riserva la proprietà del presente documento e ne vieta l’utilizzo o la divulgazione a terzi senza il proprio benestare.

Tecno-gaz Europa B evo Instructions for use | Manualzz

Istruzioni per l’uso
ITALIANO
Instructions for use
ENGLISH
Mode d’emploi
FRANCAIS
Instrucciones de uso
ESPANOL
Bedienungsanleitung
DEUTCH
DXBT200 Rev_4
Costruito da / The Manufacturer / Fabriqué
par / Fabricado por / Hersteller:
0434
MEDILINE ITALIA s.r.l.
Via 8 marzo, 4
43025 Corte Tegge – Cavriago (RE)
ITALY
ITALIANO
Gentile Cliente,
La ringraziamo per aver scelto la nostra autoclave, sapremo ricambiare la Sua fiducia con
la massima attenzione e un servizio sicuramente adeguato alle Sue aspettative.
L’autoclave da Lei scelta è la massima espressione in fatto di tecnologia applicata.
Può essere utilizzata nel campo dentale, medicale, estetico ed in generale in tutti i settori
in cui si ha la necessità di procedere ad una sicura sterilizzazione di tutti i tipi di strumenti
e materiali: solidi, cavi, porosi, liberi ed imbustati.
Prima di utilizzare questa autoclave, La invitiamo a leggere con massima attenzione il
manuale d’uso e successivamente alloggiarlo in un luogo accessibile a tutti gli operatori
addetti alla sterilizzazione.
L’autoclave deve essere utilizzata solo ed esclusivamente per gli usi previsti dal
costruttore.
Per l’installazione, manutenzione ed assistenza si rivolga esclusivamente a tecnici
autorizzati.
La invitiamo ad usare e richiedere esclusivamente ricambi originali.
1
ITALIANO
INDICE
1. CONSULTAZIONE MANUALE
1.1
GLOSSARIO
1.2
DESCRIZIONE VISTA ANTERIORE / POSTERIORE
2. SICUREZZA
2.1
MARCATURA DI SICUREZZA
2.2
DISPOSITIVI DI SICUREZZA
2.3
NOTE DI SICUREZZA
2.4
SMALTIMENTO
3.
IMPIEGO E DESTINAZIONE D’USO DELL’AUTOCLAVE
4. DATI TECNICI
4.1
MECCANICI
4.2
ELETTRICI
4.3
CAMERA
4.4
SERBATOIO ACQUA PULITA
4.5
SERBATOIO ACQUA USATA
4.6
FILTRO BATTERIOLOGICO
5.
ACCESSORI
6.
DISIMBALLAGGIO
7.
INSTALLAZIONE
8. ISTRUZIONI DI UTILIZZO
8.1
ACCENSIONE DELL' AUTOCLAVE
8.2
CARICO SERBATOIO ACQUA PULITA
8.3
CARATTERISTICHE ACQUA DA UTILIZZARE
8.4
CARICO MATERIALI IN AUTOCLAVE
8.5
INIZIO CICLO DI STERILIZZAZIONE
8.6
FINE CICLO DI STERILIZZAZIONE
8.7
SCARICO MATERIALI STERILIZZATI
8.8
SCARICO ACQUA UTILIZZATA
8.9
INTERROMPERE UN CICLO DI STERILIZZAZIONE
9. CICLI DI STERILIZZAZIONE
9.1
DESCRIZIONE CICLI
9.2
CICLI OPERATIVI
9.3
CICLI NOTTE
9.4
CICLI TEST
10. TEST DI CONTROLLO AUTOCLAVE
10.1
INTEGRATORI CHIMICI
10.2
INDICATORE BIOLOGICO
10.3
BOWIE&DICK TEST
2
ITALIANO
10.4
10.5
HELIX TEST
VACUUM TEST
11. TABELLE E DIAGRAMMI
12. LETTURA DEL REPORT DI CICLO
13. MESSAGGI DI ERRORE E ALLARMI
14. PROGRAMMAZIONE DISPLAY
14.1 SELEZIONE LINGUA
14.2 SELEZIONE ORA
14.3 SELEZIONE MINUTI
14.4 SELEZIONE SECONDI
14.5 SELEZIONE DATA
14.6 SELEZIONE MESE
14.7 SELEZIONE ANNO
14.8 VISUALIZZAZIONE PRESSIONE ATMOSFERICA
14.9 SELEZIONE CARICO ACQUA
15. SOLUZIONE A PROBLEMI OPERATIVI
15.1 L’AUTOCLAVE NON ASCIUGA CORRETTAMENTE
15.2 LA CAMERA DELL’AUTOCLAVE DIVENTA BIANCA
15.3 LA CAMERA DELL’AUTOCLAVE PRESENTA MACCHIE VERDI
15.4 IL CICLO DI STERILIZZAZIONE SI INTERROMPE
15.5 L’AUTOCLAVE NON RICEVE I COMANDI
15.6 LA STAMPANTE DELL’AUTOCLAVE NON STAMPA
15.7 MACCHIE SUGLI STRUMENTI
16. DEMINERALIZZATORE
16.1 IMPOSTAZIONE SISTEMI DI DEMINERALIZZAZIONE
16.2 COLLEGAMENTO SISTEMI DI DEMINERALIZZAZIONE
16.3 CAMBIO CARTUCCIA RESINA E FILTRO
17. PROCEDURE STANDARD STERILIZZAZIONE
18. ISTRUZIONE PER LA SPEDIZIONE DELL’AUTOCLAVE
19. PROCEDURE PER SERVIZIO ED ASSISTENZA
20. GARANZIA
A. MANUTENZIONE
3
ITALIANO
CONSULTAZIONE MANUALE
1
1.1
Glossario
VUOTO
Pressione inferiore alla pressione atmosferica
CAMERA
Corpo cilindrico in acciaio che alloggia i dispositivi da sterilizzare
FILTRO
TRAY
PORTATRAY
Dispositivo atto a trattenere particelle
Cassetto di supporto del carico da sterilizzare
Struttura di supporto dei tray in camera di sterilizzazione
DEMINERALIZZATORE
Dispositivo atto ad eliminare i sali minerali dell'acqua proveniente dalla rete
idrica
HELIX TEST
BOWIE&DICK TEST
VACUUM TEST
Test di verifica della penetrazione del vapore in un corpo poroso
FILTRO
BATTERIOLOGICO
Dispositivo atto a trattenere le particelle in sospensione nell'aria superiori a
0,3 µm.
SOVRAPRESSIONE
Pressione superiore a quella standard per il ciclo considerato
Test di verifica della penetrazione del vapore in un corpo cavo
Test di verifica del mantenimento del grado di vuoto in camera
SOVRATEMPERATURA Temperatura superiore a quella standard per il ciclo considerato
Componente sostituibile (cartuccia) del demineralizzatore che elimina tutti i
sali di tipo inorganico
RESINE
1.2
Descrizione vista anteriore / posteriore
Piedino distanziale
Troppo pieno acqua demineralizzata
Rubinetto di scarico acqua demineralizzata
(collegamento demineralizzatore)
Valvola di sicurezza
Troppo pieno acqua utilizzata – scarico condensa
Rubinetto di scarico acqua utilizzata
Alimentazione elettrica principale
Collegamento presa elettrica esclusivamente
demineralizzatore
Fusibile 12A
FIG.1
Fusibile 12A
Fusibile 500mA
4
ITALIANO
FIG.2
5
ITALIANO
Filtro Batteriologico
Carico Manuale Acqua demineralizzata con Tappo
Carico Acqua demineralizzata
Scarico Acqua Usata utilizzata
Display
Stampante
Connessione RS232
Connettore incubatore biologico B-TEST
Interruttore Generale
Ciclo 121°C
(valido per materiale imbustato e non imbustato)
Ciclo 134°C
(valido per materiale imbustato e non imbustato)
Ciclo 134°C Flash
Ciclo 134°C Prion
Vacuum Test
Helix / Bowie&Dick Test
Minimo livello acqua demineralizzata
Massimo livello acqua demineralizzata
Massimo livello acqua utilizzata
6
B-START
Pulsante START-STOP
B-PUMP
Pulsante Caricamento Acqua demineralizzata
B-SELECT
Pulsante Selezione Ciclo
B-DOOR
Pulsante Apertura porta
ITALIANO
2
2.1
SICUREZZA
Marcatura di sicurezza
TENSIONE PERICOLOSA
TOGLIERE TENSIONE PRIMA
DI RIMUOVERE IL
COPERCHIO
ALTA TEMPERATURA
CONNESSIONE A TERRA
2.2
Dispositivi di sicurezza
L'autoclave è fornita dei seguenti dispositivi di sicurezza:
-) Valvola di sicurezza tarata 2.4 bar 0/+10%
-) Blocco elettromagnetico per evitare l’apertura del portello durante l’esecuzione del ciclo
-) Termostato di sicurezza
2.3
Note di Sicurezza
• Il produttore è responsabile del prodotto immesso sul mercato ai sensi della normativa
vigente. La responsabilità decade nel momento in cui vengono eseguite operazioni sul
dispositivo, o su parte di esso, da personale non qualificato o con l’utilizzo di parti di
ricambio non originali.
• Il locale dove si installa l’autoclave non deve essere a rischio potenziale di esplosione
e/o incendio.
• L’autoclave deve essere installata in un ambiente conforme ai requisiti legislativi vigenti.
• Il connettore dell’ incubatore (FIG.2-pos.18), deve essere utilizzato solo ed
esclusivamente per il collegamento del B-TEST.
Poiché è una connessione a 230 V a.c. indipendente dall’interruttore generale,
non toccare il connettore con mani bagnate o umide e proteggerlo da eventuali
schizzi d’acqua.
7
ITALIANO
2.4
Smaltimento
Questo prodotto è soggetto alla direttiva 2002/96/EC del Parlamento
europeo e del Consiglio dell’Unione europea sui rifiuti di apparecchiature
elettriche (RAEE). Nelle giurisdizioni che adottano tale direttiva, il prodotto
è stato lanciato sul mercato in data successiva al 13 agosto 2005 e non
deve essere smaltito come rifiuto domestico non riciclabile. Utilizzare le
apposite strutture RAEE di raccolta locali per lo smaltimento di questo
prodotto oppure attenersi alle disposizioni vigenti.
3
IMPIEGO E DESTINAZIONE D’USO DELL’AUTOCLAVE
L’autoclave è in grado di sterilizzare le tre tipologie di carico previste dalla norma
EN13060, specificatamente :
MATERIALI FERROSI Strumenti senza cavità e senza ostacoli per la penetrazione del
O SOLIDI
vapore
CORPI POROSI
Materiali semplici o composti che possono assorbire i fluidi
(tessuti, camici, garze, medicazioni ecc…)
CORPI CAVI
Materiali o dispositivi con cavità, ostruzioni ecc…
Questi si suddividono in due tipologie, classificate secondo
lunghezza e diametro. Indicativamente
TIPO B: cannule, tubi o dispositivi con passaggi considerevoli
TIPO A: turbine, manipoli e dispositivi con fori ciechi o di piccole
dimensioni.
Questo dispositivo è certificato per la sterilizzazione dei seguenti carichi di materiali* :
EUROPA B EVO
EUROPA B EVO 24
Materiali solidi liberi ed imbustati
max kg. 5
max kg. 7
Materiali cavi di tipo A e B
max kg. 5
max kg. 7
max kg. 1,5
max kg. 2
Materiali porosi
*Valido solo per i paesi europei
L’autoclave deve essere utilizzata, esclusivamente, per la
sterilizzazione di strumenti e materiali compatibili con il sistema di
sterilizzazione a vapore. Accertarsi sempre che i carichi sottoposti a
sterilizzazione possano sopportare le temperature del ciclo
prescelto.
8
ITALIANO
4
4.1
DATI TECNICI
Meccanici
EUROPA B EVO
Temperatura di lavoro
Altitudine MAX
Umidità relativa MAX a 30°C
Umidità relativa MAX a 40°C
Dimensioni ingombro (L x H x P)
Ingombro portello aperto
Peso (serbatoi vuoti)
Peso (serbatoi pieni)
Peso per area di supporto
Livello potenza sonora
4.2
+5°C ÷ +40°C
2.000 m
80%
50%
510 X 390 X 590 mm
510 X 390 X 730 mm
300 mm
54 kg
58 kg
63 kg
67 kg
2
2058 N/m
2058 N/m2
< 70 db A
Elettrici
Tensione alimentazione
Potenza
Frequenza
Cavo alimentazione
Fusibili
Calore trasmesso
4.3
EUROPA B EVO 24
EUROPA B EVO
EUROPA B EVO 24
230 V a.c. +/-10 % single phase
1.7 kW
2.2 kW
50 / 60 Hz
2 + 1 x 1mm2
6. 3 x 32 – 12A (x2) / 6.3 x 32 – 500mA (x1)
3.6 E6 J / ora
Camera
Pressione lavoro MAX
Vuoto MAX
Temperatura MAX
Materiale
Dimensioni
EUROPA B EVO
EUROPA B EVO 24
2.4 bar (relativi)
- 0.9 bar (relativi)
138 °C
Inox AISI 304
Ø 245 x 318 mm
Ø 245 x 500 mm
9
ITALIANO
4.4
Serbatoio Acqua Pulita
EUROPA B EVO
EUROPA B EVO 24
Con demineralizz.
Volume
Cicli eseguibili
4,5 l
Vedere istruzioni
demineralizzatore
4
Materiale
4.5
Serbatoio Acqua Usata
Volume
Cicli eseguibili
Materiale
Temperatura max. acqua di
scarico
EUROPA B EVO 24
4,5 l
4
2
polietilene
50°C
Filtro Batteriologico
EUROPA B EVO
Diametro
Capacità filtrante
10
2
polietilene
EUROPA B EVO
4.6
Senza
demineralizz.
EUROPA B EVO 24
56 mm
0.3 µm
ITALIANO
05
ACCESSORI
Portatray
EUROPA B EVO
EUROPA B EVO 24
Alluminio anodizzato
192 x 165 x 280 mm
192 x 200 x 470 mm
Materiale
Dimensioni (L x H x P)
Immagine
FIG.3
FIG.4
Dotazione prevista
1
Portatray (opzionali)
EUROPA B EVO
Alluminio anodizzato
193 x 200 x 280 mm
(200 x 193 x 280) mm
Materiale
Dimensioni (L x H x P)
EUROPA B EVO 24
/
/
/
Immagine
FIG.5
Dotazione prevista
1
/
Trays
EUROPA B EVO
Materiale
Dimensioni (L x H x P)
EUROPA B EVO 24
Alluminio anodizzato
184 x 17 x 286 mm 184 x 17 x 286 mm + 184 x 17 x 140 mm
Immagine
+
FIG.6
Dotazione prevista
4
FIG.7
FIG.8
4+4
11
ITALIANO
Trays (opzionali)
EUROPA B EVO
Materiale
/
Dimensioni (L x H x P)
/
Immagine
/
Dotazione prevista
/
EUROPA B EVO 24
Alluminio anodizzato o
acciaio inox
184 x 17 x 460 mm
FIG.9
4
Chiave estrazione tray e regolazione portello
Utilizzare per estrarre e manovrare i trays (FIG.11) e per poter regolare il portello
(paragrafo 14.4)
Immagine
FIG.11
FIG.10
Dotazione prevista
1
Imbuto in plastica con tubo
Utilizzare per caricamento acqua manuale (paragrafo 8.2)
Immagine
FIG.12
Dotazione prevista
1
Spugna pulizia camera e guarnizione portello
Utilizzarla per pulire la camera di sterilizzazione e la guarnizione portello (par.14.2-14.3)
Immagine
FIG.13
Dotazione prevista
12
1
ITALIANO
Raccordo carico acqua
Utilizzarlo con l’imbuto in plastica con tubo per caricamento manuale acqua e con tubo
carico acqua completo di filtro
Immagine
FIG.14
Dotazione prevista
1
Tubo carico acqua completo di filtro
Utilizzare per caricamento acqua manuale (paragrafo 8.2)
Immagine
FIG.15
Dotazione prevista
1
Tubo scarico acqua
Utilizzare per scaricare l’acqua utilizzata (paragrafo 8.9)
Immagine
FIG.16
Dotazione prevista
1
Tubo rilsan
Collegare una estremità del tubo al Troppo pieno acqua utilizzata – scarico condensa
(FIG.1–pos.4), l’altra ad un contenitore per recupero acqua e condensa.
Immagine
FIG.17
Dotazione prevista
1
13
ITALIANO
Piedini distanziale posteriore in plastica
Applicare il piedino distanziale posteriore in plastica nella parte posteriore dell’autoclave
(FIG.1–pos.0) per garantire un’adeguata ventilazione nel caso in cui l’autoclave venga
posizionata vicino ad una parete
Immagine
FIG.18
Dotazione prevista
1
Tubi per scarico posteriore utenze
1- Tubo per troppo pieno acqua pulita: collegare una estremità del tubo al troppo pieno
posteriore acqua pulita (FIG.1–pos.1), l’altra ad un contenitore per recupero acqua
demineralizzata o in scarico (versione col demineralizzatore).
2- Tubo per rubinetto scarico acqua utilizzata: collegare una estremità del tubo al raccordo
FIG.20 ed avvitarlo al rubinetto di scarico acqua utilizzata (FIG.1–pos.5), l’altra ad un
contenitore per recupero acqua utilizzata .
Immagine
FIG.19
Dotazione prevista
2
Raccordo per scarico posteriore acqua utilizzata
Collegare una estremità del tubo per scarico acqua utilizzata al raccordo FIG.20 ed
avvitarlo al rubinetto di scarico acqua utilizzata (FIG.1-pos.5).
Immagine
FIG.20
Dotazione prevista
14
1
ITALIANO
Cavo alimentazione
Collegare l’estremità del connettore (FIG.21–pos.B) al pannello posteriore (FIG.1–pos.6)
successivamente la spina (FIG.21–pos.A) direttamente alla presa di alimentazione
dell’impianto elettrico.
Immagine
FIG.21
Dotazione prevista
06
1
DISIMBALLAGGIO
L’autoclave viene spedita in un imballo idoneo al trasporto, alla movimentazione e alla
protezione dell’autoclave stessa.
L’imballo non deve subire urti, deve essere maneggiato con cura evitando di farlo rotolare
o farlo cadere da altezze superiori a 16 cm.
Nel caso non siano disponibili attrezzature per la movimentazione maneggiare l’autoclave
imballata sempre in due persone.
L’autoclave è supportata con un pallet in legno e racchiusa in un cartone ondulato e
rinforzato internamente da composti di cartone.
Per disimballare l’autoclave, aprire il cartone ondulato, rimuovere le parti di rinforzo ed
estrarla utilizzando le cinghie in dotazione.
La movimentazione deve avvenire solo con l’utilizzo delle cinghie e
mediante l'impiego minimo di due persone.
Non sollevare mai l’autoclave prendendola dalla parte inferiore del portello
o del quadro comandi, in quanto questa operazione errata, potrebbe
danneggiare l’apparecchiatura.
ATTENZIONE: Conservate sempre l’imballo originale.
All’interno dell’imballo troverete:
• MANUALE D’USO: che dovrete leggere attentamente e riporre in un luogo accessibile a
tutti gli operatori, addetti alla sterilizzazione.
• CERTIFICAZIONI E DICHIARAZIONI DI CONFORMITA’ : da conservare.
• TAGLIANDO DI GARANZIA: che dovrete compilare e spedire.
15
ITALIANO
07
INSTALLAZIONE
FIG.22
EUROPA B EVO
A
B
C
D
E
F
EUROPA B EVO 24
610 mm
590 mm
730 mm
50 mm
410 mm
426 mm (anteriore)
410 mm (anteriore)
455 mm (posteriore)
444 mm (posteriore)
400 mm
515 mm
1 - Installare l’autoclave in ambienti idonei alla sterilizzazione.
2 - Il locale deve essere adeguatamente illuminato ed areato, come previsto dalle direttive
in vigore.
3 - Installare l’autoclave lontano da fonti di calore e schizzi d’acqua.
4 - Posizionare l’autoclave su un piano idoneo a supportarne il peso (min.80 kg) e di
adeguate dimensioni.
5 - Il piano di appoggio dell’autoclave deve essere perfettamente orizzontale.
6 - Posizionare l’autoclave ad una altezza tale che l’utilizzatore possa ispezionare la
totalità della camera di sterilizzazione e riesca a pulirla con facilità.
7 - Aprire il portello dell’autoclave e togliere tutte le buste, che imballano i singoli
accessori, contenuti all’interno della camera di sterilizzazione.
16
ITALIANO
8 - Lasciare all’interno della camera di sterilizzazione solo il portatray con i trays,
posizionare tutti gli altri accessori in un vano esterno a disposizione degli operatori.
9 - Non appoggiare mai sull’autoclave giornali, vassoi, contenitori con liquidi ecc…
10 - Non appoggiarsi mai al portello quando è aperto.
11 – Lasciare uno spazio di almeno 5 cm nella parte posteriore utilizzando il piedino
distanziale posteriore in plastica (FIG.1–pos.0 / FIG.18) e nei fianchi dell’ apparecchio in
modo da garantire la ventilazione necessaria.
12 - Effettuare i collegamenti dei tubi nella parte posteriore (capitolo 5)
13 - Accertarsi sempre che l’impianto elettrico a cui si allaccia l’autoclave sia conforme alle
norme vigenti e dimensionato in maniera adeguata alle caratteristiche dell’apparecchio.
14 - Prendere il cavo di alimentazione elettrica in dotazione e innestare la presa femmina
sulla spina del pannello posteriore dell’autoclave (FIG.1 –pos.6)
15 - Collegare la spina elettrica all’impianto assicurandosi che l’alimentazione sia 230 Vac2000Va
IMPORTANTE:
Evitare di connettersi con prolunghe, riduzioni od adattatori; in caso contrario
potrebbero crearsi microinterruzioni con conseguente segnalazione di allarme.
16 - Accendere l’autoclave premendo l’interruttore generale (FIG.2–pos.19) e aprire il
portello della autoclave stessa. Attendere alcuni secondi, vi saranno due segnalazioni
sonore che informano sull’acquisizione dei parametri relativi all’allineamento barometrico
automatico, contestualmente sul display comparirà la scritta “PORTA APERTA”.
IMPORTANTE:
Non selezionare mai un comando prima delle due segnalazioni sonore, l’autoclave
non accetterà la programmazione prescelta.
17 - Caricare il serbatoio di acqua pulita seguendo le indicazioni del capitolo 8.
18 - Se l’autoclave viene installata con il sistema di demineralizzazione, attenersi alle
istruzioni contenute nell’imballo di questo dispositivo e nel capitolo 16.
17
ITALIANO
08
ISTRUZIONI DI UTILIZZO
Dopo aver installato l’autoclave, procedere alla preparazione e all’utilizzo.
8.1
Accensione dell’autoclave
Premere l’interruttore generale (FIG.2–pos.19)
8.2
Carico serbatoio acqua pulita
Collegare al “tubo carico acqua completo di filtro” (in dotazione) il “raccordo carico acqua “
FIG.14, e inserirlo nel raccordo frontale dell’autoclave (FIG.2–pos.13).
Inserire l’altro capo del tubo col filtro all’interno del contenitore dell’acqua demineralizzata
o distillata.
A questo punto premere il pulsante B-PUMP per azionare la pompa di carico acqua e
mantenerlo premuto finché non apparirà il conto alla rovescia.
La pompa carica il serbatoio dell'acqua pulita interno all’autoclave. Se il livello massimo
non viene raggiunto entro 180 secondi, la pompa si ferma automaticamente e sarà quindi
necessario premere nuovamente il pulsante B-PUMP.
La pompa si ferma automaticamente quando il livello massimo è stato raggiunto.
Nel caso in cui la pompa di carico acqua non funzioni, il riempimento del serbatoio può
essere effettuato manualmente dall’operatore nel seguente modo:
•
•
•
•
•
•
Accendere l’autoclave.
Togliere il tappo posto frontalmente sull’autoclave (FIG.2–pos.12)
Inserire il raccordo carico acqua col tubo connesso all’imbuto (FIG.14)
Versare l’acqua distillata nell’imbuto tenendolo più alto del raccordo di carico
Versare fino a quando il led di livello massimo non si sarà acceso (FIG.2–pos.2).
A carico completato, estrarre il raccordo e riposizionare il tappo.
Con connessione a mezzo demineralizzatore, se il livello massimo di acqua non è stato
raggiunto, il funzionamento dell’autoclave sarà inibito.
18
ITALIANO
8.3
Caratteristiche acqua da utilizzare
TABELLA LIVELLI QUALITATIVI STABILITI DALLA NORMATIVA UNI EN 13060: 2009
CEN STANDARD UNI EN 13060 : 2009
Residuo evaporazione
10
mg/l
≤
Ossido di silicio
1
mg/l
≤
Ferro
0.2
mg/l
≤
Cadmio
0.005
mg/l
≤
Piombo
0.05
mg/l
≤
Resti di metalli pesanti tranne ferro, cadmio, piombo
0.1
mg/l
≤
Cloruro (Cl’)
2
mg/l
≤
Fosfato (P20s)
0.5
mg/l
≤
Conduttività (a 20°C)
15
≤
µs/cm
Valore Ph (grado di acidità)
5 ÷ 7,5
Aspetto
Incolore, limpido, senza depositi
Durezza (E Ioni di terra alcalina)
0.02
mmol/l
≤
8.4
Carico materiali in autoclave
Disporre i materiali da sterilizzare sui trays in dotazione, facendo attenzione:
• non sovrapporre mai i materiali
• disporre gli strumenti imbustati, sempre con la parte carta verso l’alto
• non mettere mai i materiali a contatto con la camera di sterilizzazione o con il portello di
chiusura
• disporre pinze e forbici con le lame aperte
FIG.23
Terminato il carico, chiudere il portello dell’autoclave. Sul display comparirà la scritta
“PORTA CHIUSA”.
19
ITALIANO
8.5
Inizio ciclo di sterilizzazione
Scegliere il programma di sterilizzazione più idoneo al carico predisposto, premendo il
pulsante B-SELECT.
Ogni singola pressione, cambia la selezione del programma a quello successivo.
Dopo aver scelto il programma avviare il ciclo premendo il pulsante B-START. La porta
verrà bloccata automaticamente ed il ciclo avrà inizio.
Durante il ciclo il display visualizzerà tutti i parametri ed informazioni relative.
8.6
Fine ciclo di sterilizzazione
Un segnale acustico avviserà gli operatori dell’avvenuto ciclo di sterilizzazione e sul
display comparirà il messaggio “FINE CICLO”.
Sbloccare la porta premendo il pulsante B-DOOR. Nel caso vi sia presenza di pressione,
all’interno della camera, il pulsante non azionerà lo sblocco. Attendere la completa
depressurizzazione della camera e ripetere l’operazione oppure, solo in caso di
emergenza, premere contemporaneamente B-DOOR + B-PUMP.
A portello sbloccato, tirare la maniglia della porta ed aprire.
8.7
Scarico materiali sterilizzati
Indossare idonei dispositivi di protezione individuale in accordo con le normative vigenti in
materia di sicurezza e igiene sul lavoro. Estrarre i trays utilizzando l’apposita chiave in
dotazione (FIG.10), lasciare condizionare gli strumenti e riporli in ambienti dove non
possano subire contaminazioni.
8.8
Scarico acqua utilizzata
Quando il led di livello acqua utilizzata (FIG.2–pos.28) si accende bisogna procedere allo
svuotamento del serbatoio di raccolta dell’acqua esausta.
Se non si provvede, il funzionamento dell’autoclave è inibito.
Prendere il tubo in dotazione (FIG.16), ed inserirlo nella raccordo di scarico acqua usata
posto frontalmente all’autoclave (FIG.2–pos.14), svitare la ghiera agendo in senso
antiorario (compiere 2 giri completi), contestualmente mettere l’altro capo del tubo in un
contenitore; l’acqua per caduta andrà nel contenitore stesso.
IMPORTANTE:
A - Il tubo alloggiato nel contenitore di raccolta, non deve mai, lambire o essere immerso
nell’acqua scaricata, diversamente si avrà una situazione di risucchio.
B - Attendere sempre che l’acqua di scarico sia fuoriuscita totalmente. Il led di livello
massimo acqua utilizzata, si spegne quando ancora vi è acqua nel serbatoio, pertanto non
utilizzarlo come riferimento per questa operazione.
Al termine dello scarico avvitare la ghiera e togliere il tubo.
20
ITALIANO
8.9
Interrompere un ciclo di sterilizzazione
Un ciclo di sterilizzazione può essere volontariamente interrotto, premendo il pulsante
B-START per almeno 2 secondi.
L’autoclave emetterà un suono, andrà in decompressione e sul display comparirà il
messaggio di allarme A001 (CICLO INTERROTTO).
Per resettare l’allarme, mantenere premuti contemporaneamente i pulsanti sotto
evidenziati, sino all’intervento del segnale acustico:
B-SELECT
B-PUMP
B-DOOR
ATTENZIONE: Se la stampante è in funzione il reset dell’allarme è inibito
21
ITALIANO
09
9.1
CICLI DI STERILIZZAZIONE
Descrizione cicli
L’autoclave è corredata di tre serie di cicli:
A - cicli operativi
B - cicli notte
C - cicli test
9.2
Cicli operativi
Tutti i cicli operativi, hanno il sistema di vuoto frazionato, pertanto possono sterilizzare
materiali cavi, porosi, solidi, sia liberi che imbustati.
Le temperature selezionabili sono 121°C – 134°C.
Normalmente i cicli 121°C, si utilizzano per termoplasti ci o materiali sensibili, mentre i cicli
134°C, si utilizzano per tutti gli altri materiali.
In tutti i casi, seguire sempre le indicazioni, date dai produttori degli strumenti o dispositivi,
da sterilizzare.
Esistono anche 2 cicli speciali:
- ciclo Prion, studiato per il morbo di Creutzfeldt-Jakob (sindrome della mucca pazza)
- ciclo Flash, creato per la sterilizzazione rapida di strumenti o dispositivi non imbustati.
9.3
Cicli notte
L’autoclave, è dotata di uno speciale dispositivo economizzatore.
Tale dispositivo permette di eseguire cicli di sterilizzazione in assenza dell’operatore.
Al termine del ciclo, se la porta non viene aperta, l’autoclave si stabilizza, poi si spegne
automaticamente, resterà acceso soltanto l’interruttore generale (FIG.2-pos.19)
All’arrivo degli operatori, sarà sufficiente premere qualsiasi pulsante per riaccendere
l’autoclave e leggere l’esito del ciclo sul display. Oltre a ciò la stampante avrà redatto
regolarmente il rapporto scritto relativo al ciclo stesso.
9.4
Cicli test
I clcli di test disponibili sono:
- Helix test
- Bowie&Dick test
- Vacuum test
22
ITALIANO
10
TEST DI CONTROLLO AUTOCLAVE
10.1
Integratore chimici
Sono test che sfruttano le proprietà di sostanza coloranti capaci di modificare il proprio
aspetto se vengono adeguatamente esposte al calore e alla pressione in tempi adeguati.
In quanto interagiscono con tutti i parametri del ciclo (pressione,temperatura, tempo). Il
superamento del test ne certifica la corretta stabilità.
UTILIZZO
Gli integratori chimici (cod.ricambio 200/S e 215-S) vanno inseriti
all’interno della camera di sterilizzazione prima di dare avvio al ciclo,
anche con presenza di carico.
Gli integratori possono essere liberi e posizionati sui tray, oppure
possono essere inseriti all’interno delle buste di sterilizzazione
insieme ai materiali da sterilizzare.
Inserire gli integratori nelle buste è una procedura che si consiglia
nelle strutture dove vi sono diversi operatori, infatti così facendo si
certifica l’avvenuta sterilizzazione di ogni singolo strumento.
Se il viraggio rimane nella zona
bianca, il test non è superato
RESPONSO
Se il viraggio entra nella finestra
SAFE, il test è superato
PERIODICITA’
10.2
Non esiste una periodicità predefinita, se non nelle zone ove vi sono
leggi regionali specifiche. In tutti i casi gli integratori chimici sono i
test più economici ed immediati, pertanto, onde potere avere un
costante controllo della validità dell’autoclave, si consiglia l’utilizzo
su tutti i cicli, oppure almeno una volta al giorno.
Indicatore biologico
Il test (cod.ricambio 262-S) serve a dimostrare la capacità dell’autoclave in merito alla
distruzione di tutti i microrganismi. E’ rappresentato da una preparazione standardizzata di
spore, che possiedono caratteristiche biologiche, alta resistenza al calore, e che sono da
ritenere un mezzo di controllo di assoluta tranquillità e sicurezza (ATCC 7953).
Il test si presenta sottoforma di fiale, e non è patogeno, tossico e pirogenico.
23
ITALIANO
UTILIZZO
Le fiale devono essere inserite nella camera di sterilizzazione,
anche con presenza di carico.
Eseguire il ciclo, al termine estrarre la fiala con attenzione in quanto
calda e in pressione.
Lasciarla raffreddare per circa 10 minuti poi attivarla, utilizzando
l’apposita pinza in dotazione, mantenendola sempre in verticale.
L’indicatore chimico posto sull’etichetta della fiala sarà virato dal blu
al nero.
Successivamente, inserire la fiala in un incubatore biologico a 57°C
per 48 ore, unitamente ad una fiala test non processata, ma
comunque attivata con l’apposita chiave.
RESPONSO
Dopo 48 ore estrarre la fiala processata dall’incubatore e valutare il
responso. Se la fiala ha virato sul giallo vuole dire che l’autoclave
non ha superato il test e vi è crescita batterica. Se la fiala resta
colore viola indica che non vi sono microrganismi in crescita,
pertanto l’autoclave ha superato il test. La fiala test ovviamente
virerà sempre al giallo in quanto non processata, e servirà solo
come termine di confronto.
Al termine del test smaltire le fiale nei rifiuti urbani solidi, si consiglia
però di sottoporre la fiala ad un ulteriore ciclo di sterilizzazione a
121°C
PERIODICITA’
10.3
Non esiste una periodicità predefinita, se non nelle zone ove vi sono
leggi regionali specifiche. In tutti i casi i test biologici sono i test più
concreti, pertanto onde potere avere un costante controllo della
validità dell’autoclave, si consiglia di eseguire il test almeno una
volta ogni 90 giorni.
Bowie&Dick test
E’ un test fisico (cod.ricambio 268/S) che permette di verificare la capacità di penetrazione
del vapore nei corpi porosi. Per il test viene utilizzato un “pacco prova” standardizzato e
conforme alle normative tecniche attualmente in vigore.
UTILIZZO
24
Il test va eseguito a camera vuota. Il Bowie&Dick deve essere
posizionato sul tray centrale dell’autoclave.
Eseguire l’apposito ciclo, indicato nel display dell’autoclave, al
termine estrarre il pacco, aprirlo e controllare il viraggio del foglio con
indicatore chimico posto all’interno.
ITALIANO
RESPONSO
PERIODICITA’
10.4
La valutazione è semplice e rapida.
Se il viraggio risulta uniforme (come
in figura) il test è perfettamente
riuscito, diversamente il test non è
superato, pertanto l’autoclave non è
in grado di sterilizzare correttamente
i corpi porosi.
Non esiste una periodicità predefinita, se non nelle zone ove vi sono
leggi regionali specifiche. In tutti i casi, onde potere avere un
costante controllo della validità dell’autoclave, si consiglia di
eseguire il test almeno una volta ogni 30 giorni.
Helix test
E’ un test fisico (cod.ricambio 267-S) che permette di verificare la capacità di penetrazione
del vapore nei corpi cavi. Per il test viene utilizzato un sistema standardizzato e conforme
alle norme tecniche in vigore.
UTILIZZO
Il test va inserito a camera vuota.
Nella capsula posta all’estremità del test inserire l’apposito strip, poi
posizionare il test sul tray centrale dell’autoclave.
Eseguire l’apposito ciclo, indicato nel display dell’autoclave, al
termine estrarre il test aprire la capsula e controllare il viraggio dello
strip indicatore.
25
ITALIANO
TEST ALL’ORIGINE
RESPONSO
TEST NON RIUSCITO
TEST RIUSCITO
PERIODICITA’
10.5
Non esiste una periodicità predefinita, se non nelle zone ove vi sono
leggi regionali specifiche. In tutti i casi onde potere avere un costante
controllo della validità dell’autoclave si consiglia di eseguire il test
almeno 1 volta ogni 30 giorni.
Vacuum test
E’ una prova di tenuta della camera o prova della perdita del vuoto.
Il controllo ha lo scopo di verificare che durante il ciclo non si verifiche infiltrazione di aria
attraverso le tenute della camera (guarnizioni, valvole, ecc…)
Il ciclo va eseguito con camera vuota.
UTILIZZO
Si seleziona l’apposito ciclo indicato nel display dell’autoclave e si
attiva.
L’autoclave esegue automaticamente il ciclo seguendo precise
procedure tecniche. Successivamente l’esito finale viene emesso
dalla stampante tramite una eventuale connessione informatica.
RESPONSO
Il responso è immediato e viene rilasciato dalla stampante, la quale
emette tutti i valori dal ciclo eseguito e anche la valutazione finale.
PERIODICITA’
Non esiste una periodicità predefinita, se non nelle zone ove vi sono
leggi regionali specifiche. In tutti i casi onde potere avere un costante
controllo della validità dell’autoclave si consiglia di eseguire il test
almeno una volta ogni settimana.
SI CONSIGLIA DI UTILIZZARE ESCLUSIVAMENTE
TEST PROPOSTI DA TECNO - GAZ S.p.A
26
ITALIANO
11
TABELLE E DIAGRAMMI
121°C
(imbustati e non
imbustati)
134°C
(imbustati e non
imbustati)
134°C
Flash
134°C
Prion
134°C
Helix / Bowie&Dick
18L
18
24L
BANDA DI
PRESSIONE
DI LAVORO
( bar relativi )
BANDA DI
TEMPERATURA
DI LAVORO
( °C )
17
13
1.04 ÷ 1.30
121 ÷ 125
5
17
13
2.02 ÷ 2.38
134 ÷ 138
4
2
4
2.02 ÷ 2.38
134 ÷ 138
20
17
13
2.02 ÷ 2.38
134 ÷ 138
2.02 ÷ 2.38
134 ÷ 138
3.5
4
-0.9
Vacuum
CICLI
TEST
CICLI
TEMPO
ASCIUG.
T5 (Minuti)
TEMPO
ESPOSIZ.
T4
(Minuti)
CICLI OPERATIVI
TABELLA STERILIZZAZIONE
I tempi di preriscaldo e di frazionamento del vuoto possono variare a seconda delle
condizioni dell’autoclave da 25 – 35 minuti che andranno a sommarsi con i tempi dei cicli
riportati in tabella. Tutti i cicli di sterilizzazione hanno 3 fasi di vuoto.
DIAGRAMMA DI CICLO
T1,T2,T3
fasi di preriscaldo e di
frazionamento del vuoto
T4
sterilizzazione
T5
asciugatura
FIG.24
27
ITALIANO
12
LETTURA DEL REPORT DI CICLO
Versione del software
Numero seriale
Data e Descrizione del ciclo selezionato
(con i valori principali di setup)
Ore – Minuti – Secondi
Pressione della camera
Temperatura principale della camera
Fase del ciclo
Fine del ciclo (OK significa che il ciclo si
è concluso correttamente)
Numero di ciclo
Operatore (spazio per la firma)
28
ITALIANO
13
MESSAGGI DI ERRORE E ALLARMI
I messaggi di errore e di allarmi sono evidenziati tramite un codice alfanumerico,
composto da una lettera e da 3 cifre.
Il prefisso “E” è relativo ad errori dell’operatore, ripristinabili dall’operatore stesso.
Il prefisso “A” è relativo ad allarmi, anomalie dell’autoclave, dopo aver attuato il rimedio, se
il problema persiste bisogna richiedere assistenza telefonica.
In caso di visualizzazione di un messaggio di allarme (suffisso "A") il
ciclo è da ritenersi NON ANDATO A BUON FINE: occorrerà ripetere
tutte le operazioni di preparazione e sterilizzazione.
Per resettare allarmi ed errori, mantenere premuti contemporaneamente i pulsanti sotto
evidenziati, sino all’intervento del segnale acustico:
B-DOOR + B-PUMP + B-SELECT
ERRORE
E 200
E 250
CAUSA
RIMEDIO
Basso livello acqua nel serbatoio acqua
Riempire il serbatoio
demineralizzata
Basso livello acqua pulita con caricamento
Riempire il serbatoio
da demineralizz.
E 300
Livello massimo acqua utilizzata
Svuotare serbatoio
E 401
Porta aperta
Chiudere con attenzione la porta
E 451
Porta non bloccata
Aprire la porta e richiudere
E 461
Alta temperatura in camera per VACUUM
TEST (oltre 40°C)
Spegnere ed attendere 10 min.
A 250
Inizio ciclo senza acqua
Controllare i sistemi di
demineralizz.
A 001
Ciclo interrotto
Resettare e riavviare il ciclo
A 011
Anomalia pc display
Spegnere e riaccendere
l’autoclave
A 101
Vuoto non raggiunto in 10 min.
Resettare e ripetere il ciclo
A 111
A 121
Vuoto non mantenuto su prima fase
VACUUM TEST
Vuoto non mantenuto su seconda fase
VACUUM TEST
Resettare e ripetere il ciclo
Resettare e ripetere il ciclo
29
ITALIANO
A 401
Portello aperto durante ciclo o problema
chiusura
Controllare il sistema di chiusura
A 501
Pressione non mantenuta
Resettare e ripetere il ciclo
A 551
Anomalia 1° sensore temperatura
Resettare e ripetere i l ciclo
A 552
Anomalia 2° sensore temperatura
Resettare e ripetere i l ciclo
A 641
Sovrapressione
Resettare e ripetere il ciclo
A 651
Sovratemperatura
Resettare e ripetere il ciclo
A 661
Anomalie raffronto temperature
Resettare e ripetere il ciclo
A 701
Pressione di lavoro non raggiunta
Resettare e ripetere il ciclo
A 751
Bassa temperatura
Resettare e ripetere il ciclo
A 761
Anomalia nel sensore di pressione
Resettare e ripetere il ciclo
A 801
Anomalia scarico pressione
Resettare e ripetere il ciclo
A 901
Tensione principale troppo bassa –
Temporanea mancanza di alimentazione
elettrica
Resettare e ripetere il ciclo.
In caso di ripetizione verificare
erogazione energia elettrica.
SOSTITUIRE LE RESINE Resine IONIC
SYSTEM esaurite
Seguire le indicazione del del
manuale paragrafo 16.3
SOSTITUIRE IL FILTRO Resine OSMOSI SYSTEM
esaurite
Seguire le indicazione del del
manuale paragrafo 16.3
D
Portello stampante non
chiuso
Aprire e richiudere il portello della stampante
P
Manca la carta della
stampante
Inserire la carta
Lo
Bassa tensione sulla
linea di alimentazione
Provare a collegare la macchina alla rete elettrica in un
altro punto; se il problema persiste contattare un
elettricista
30
ITALIANO
14
PROGRAMMAZIONE DISPLAY
Fare riferimento alla FIG.2.
14.1
Selezione Lingua
Premere B-START e contemporaneamente l’interruttore generale (MAINS). Tenere
premuto B-START fino a quando sul display compare il nome della lingua impostata.
Premendo B-SELECT si modifica la lingua [avanti].
Premendo B-PUMP si modifica la lingua [indietro].
14.2
Selezione Ora
Premendo B-START compare l’ora.
Premendo B-SELECT si modifica l’ora [avanti].
Premendo B-PUMP si modifica l’ora [indietro].
14.3
Selezione Minuti
Premendo B-DOOR compare minuti.
Premendo B-SELECT si modifica minuti [avanti].
Premendo B-PUMP si modifica minuti [indietro].
14.4
Selezione Secondi
Premendo B-DOOR compare secondi. Non modificare.
14.5
Selezione Data
Premendo B-START compare giorno (lampeggia).
Premendo B-SELECT si modifica giorno [avanti].
Premendo B-PUMP si modifica giorno [indietro].
14.6
Selezione Mese
Premendo B-DOOR compare mese (lampeggia).
Premendo B-SELECT si modifica mese [avanti].
Premendo B-PUMP si modifica mese [indietro].
31
ITALIANO
14.7
Selezione Anno
Premendo B-DOOR compare anno.
Premendo B-SELECT si modifica anno [avanti].
Premendo B-PUMP si modifica anno [indietro].
14.8
Visualizzazione Pressione Atmosferica
Premendo B-START compare la pressione atmosferica memorizzata.
NON MODIFICARE !
14.9
Selezione Carico Acqua
Premendo B-START compare l’impostazione carico acqua :
• CARICAMENTO CON POMPA
• IONIC SYSTEM
• OSMOSI SYSTEM
Premendo B-SELECT si modifica l’impostazione.
Per salvare l’impostazione premere B-DOOR e tenerlo premuto fino al segnale acustico
“BEEP”
Per uscire premere B-START e tenerlo premuto fino a quando non si sente il segnale
acustico “BEEP”.
I dati impostati sono memorizzati. Spegnere l’autoclave tramite l’interruttore MAINS.
32
ITALIANO
15
SOLUZIONE A PROBLEMI OPERATIVI
In molti casi, alcuni allarmi o errori sono determinati da non attenzione o non conoscenza
di alcuni aspetti tecnici ed operativi.
Qui di seguito vengono elencati alcuni casi di anomalie con relative soluzioni.
15.1
L’autoclave non asciuga correttamente
• sostituire il filtro batteriologico con uno nuovo originale
• sono stati utilizzati tray non originali, di diverso materiale, senza fori o con foratura
diversa. Si consiglia di utilizzare solo tray originali.
• gli strumenti non sono stati disposti correttamente. Attenersi rigorosamente alle
indicazione del par.8.4
15.2
La camera dell’ autoclave è diventata bianca
• cambiare immediatamente il tipo di acqua utilizzata, utilizzare acqua demineralizzata o
distillata, come specificatamente indicato nei capitoli precedenti e procedere poi alla
pulizia della camera.
• il colore biancastro puo’ essere conseguenza dell’evaporazione di materiali organici,
presenti sugli strumenti. Provvedere a sottoporre gli strumenti ad una azione di
detersione più idonea ed approfondita.
• verificare l’eventuale impianto di demineralizzazione installato
15.3
La camera dell’ autoclave presenta macchie verdi-bluastre
• non vi e’ stato un corretto risciacquo degli strumenti dopo la fase di detersione,
sciacquare con maggior attenzione e scrupolo gli strumenti. se le macchie sono evidenti
richiedere assistenza tecnica telefonica.
15.4
Il ciclo di sterilizzazione si interrompe
• controllare se l’autoclave e’ collegata alla rete elettrica con prolunghe, riduzioni,
adattatori, nel caso togliere questi accessori e collegare l’autoclave direttamente alla
presa elettrica.
15.5
L’autoclave non riceve i comandi
• l’autoclave sta effettuando l’allineamento barometrico automatico, attendere il doppio
segnale sonoro dopo l’apertura del portello, poi impostare le funzioni.
• il serbatoio di acqua demineralizzata è vuoto, il led di livello minimo è acceso,
provvedere al riempimento di acqua pura.
• il serbatoio di acqua utilizzata e’ pieno, il led di livello massimo è acceso, provvedere
allo scarico dell’acqua esausta.
33
ITALIANO
15.6
La stampante dell’autoclave non stampa
• il portellino portarotolo non e’ chiuso correttamente, sul display compare la lettera “D”.
Aprire e richiudere il portellino della stampante accertandosi che sia chiuso
correttamente.
• manca la carta, sul display compare la lettera “P”. Aprire il portello e inserire il rotolo di
carta termografica.
• il rotolo della carta e’ montato con la parte termica verso l’interno, aprire il portellino e
girare il rotolo della carta in posizione corretta.
15.7
Macchie sugli strumenti
• gli strumenti diventano gialli, residuo di liquido chimico che con il caldo si è fissato sugli
strumenti. Non si è eseguito un risciacquo adeguato.
• la camera di sterilizzazione presenta macchie gialle. E' stato immesso nella camera
strumentario con presenza di liquido chimico che cadendo si è fissato grazie al calore.
Non si è eseguito un risciacquo adeguato.
• gli strumenti presentano macchie biancastre, lo sciacquo è stato effettuato con acqua
molto calcarea e gli strumenti non sono stati asciugati. Come ultimo sciacquo si
consiglia di utilizzare acqua demineralizzata e asciugare accuratamente gli strumenti.
• gli strumenti si sono anneriti, ciò è dovuto al fatto che gli strumenti hanno all’interno
forte componentistica di carbonio.
34
ITALIANO
16
16.1
DEMINERALIZZATORE
Impostazione sistemi di demineralizzazione
L’autoclave è predisposta anche per il caricamento dell’acqua demineralizzata attraverso il
SISTEMA IONICO o SISTEMA OSMOSI (accessorio opzionale).
Tali prodotti sono utilizzati per eliminare i sali minerali dell’acqua proveniente dalla rete
idrica per alimentare l’autoclave.
L’operatore prima di installare il sistema IONICO o OSMOSI deve programmare
l’autoclave seguendo le istruzioni sotto indicate:
Accendere l’autoclave premendo l’interruttore generale (FIG.2–POS.19) e tenere premuto
B-START finché non appare la scritta che riporta la lingua selezionata.
Premere e rilasciare B-START
appare:
per selezionare la schermata successiva finchè non
• CARICAMENTO CON POMPA
• IONIC SYSTEM
• OSMOSI SYSTEM
Per cambiare l’impostazione usare B-SELECT.
Per salvare l’impostazione premere B-DOOR e tenerlo premuto fino al segnale acustico
“BEEP”
Per uscire premere B-START e tenerlo premuto fino a quando non si sente il segnale
acustico “BEEP”.
I dati impostati sono memorizzati.
ATTENZIONE:
Il numero riportato sotto al tipo di caricamento indica approssimativamente quanti cicli di
sterilizzazione si possono ancora fare col filtro (nel caso del sistema ad osmosi) o con le
resine (nel caso del sistema a scambio ionico) attualmente in uso.
E’ buona norma ricordarsi di resettare questo numero al valore massimo quando si cambia
il filtro o le resine sul sistema attualmente in uso.
Dopo aver salvato il tipo di caricamento in uso, per resettare al valore massimo questa
informazione bisogna tenere premuto il tasto B-PUMP fino al segnale acustico (circa 10
secondi) nella schermata di attesa inizio ciclo.
35
ITALIANO
16.2
Collegamento dei sistemi di demineralizzazione
Qui riportiamo il particolare del collegamento specifico all’autoclave del tubo di
alimentazione dell’acqua e della connessione della spina elettrica:
Spegnere l’autoclave se è accesa (FIG.2–pos.19)
• Chiudere il rubinetto posto a monte dell’impianto di demineralizzazione
• Installare il demineralizzazione come indicato nel manuale del demineralizzatore
stesso;
• Avvolgere il filetto maschio del raccordo-portatubo con teflon o altro componente che
garantisca la tenuta all’acqua;
• Avvitare il raccordo-portatubo sul filetto femmina dello scarico dell’acqua pulita
(FIG.1–pos.2)
• Inserire il tubo in uscita dal demineralizzatore nel raccordo-portatubo ora avvitato
all’autoclave;
• Inserire la spina del demineralizzatore nella presa (FIG.1–pos.7) nel retro
dell’autoclave;
• Aprire il rubinetto posto a monte dell’impianto di demineralizzazione;
• Controllare che non vi siano delle perdite d’acqua;
• Accendere l’autoclave;
• Eseguire uno o più cicli di sterilizzazione per controllare il funzionamento della
connessione eseguita e controllare soprattutto le perdite.
A fine giornata chiudere sempre il rubinetto posto a monte dell’impianto di
demineralizzazione
Collegare i sistemi di demineralizzazione soltanto ad autoclavi predisposte
Se all’ inizio del ciclo l’autoclave segnala E250 significa che è stato avviato il
ciclo con un basso livello d’acqua pulita. E' pertanto necessario riempire il
serbatoio acqua pulita
NOTA:
per il collegamento dei sistemi di demineralizzazione alle autoclavi, fare riferimento
anche a quanto indicato sul manuale dei sistemi di demineralizzazione.
36
ITALIANO
FIG.31
16.3
Cambio cartuccia resina - filtro
L’autoclave segnala sul display lo stato di esaurimento delle resine o del filtro con il
messaggio “SOSTITUIRE RESINE” (cod.ricambio 242-S) o “SOSTITUIRE FILTRO”
(cod.ricambio 248-S-2).
Sostituire la cartuccia come indicato nel manuale del demineralizzatore; dopo sostituzione,
premere il pulsante B-PUMP fino all’intervento del segnale acustico, per resettare
l’autoclave.
37
ITALIANO
17
PROCEDURE STANDARD STERILIZZAZIONE
Sterilizzare vuol dire adottare precisi protocolli operativi ed applicare una metodologia di
sterilizzazione. L’autoclave è il punto chiave di questa metodologia.
Chi non segue tutte le varie fasi del processo di sterilizzazione, può inficiare il risultato
finale.
TECNO-GAZ S.p.A produce tutti i dispositivi per ogni fase del ciclo di sterilizzazione.
Bisogna pertanto :
1 - Predisporre un preciso protocollo operativo scritto,da distribuire a tutti gli operatori
ed addetti alla sterilizzazione.
TECNO-GAZ SpA può fornire questi protocolli personalizzati.
2 - Seguire con scrupolo ed attenzione,tutte le fasi del ciclo di sterilizzazione:
DISINFEZIONE fase obbligatoria, per salvaguardare la sicurezza degli operatori
preposti.,da attuarsi con immersione in liquidi chimici o termo disinfezione;
DETERSIONE la fase più importante che assicura la rimozione di tutti i tipi di
residui, chimici ed organici. Lo strumento più idoneo sono le vasche ad ultrasuoni;
ASCIUGATURA fase indispensabile, che evita corrosioni degli strumenti e
interferenze al ciclo di sterilizzazione;
IMBUSTAMENTO fase indispensabile per il mantenimento della sterilità nel
tempo
STERILIZZAZIONE fase finale sterilizzazione a vapore.
38
ITALIANO
18
ISTRUZIONE PER LA SPEDIZIONE DELL’AUTOCLAVE
Nel caso si debba spedire l’autoclave per riparazione, controllo, ripristino, revisione,
validazione, seguire le seguenti indicazioni obbligatorie:
•
Utilizzare l’imballo originale, se questo non è più in Vostro possesso, utilizzare un
imballo adeguato. La merce viaggia con rischio a carico del mittente.
•
Spedire solo l’autoclave (non inserire alcun componente contenuto nel kit
accessori).
•
Pulire accuratamente la camera di sterilizzazione e l’autoclave nel suo complessivo,
prima di spedirla. Nel caso giunga sporca e con residui, l’autoclave verrà rispedita
non riparata, oppure verrà sottoposta ad azione di pulizia e disinfezione.
•
Scaricare sempre il serbatoio di acqua demineralizzata, tramite il raccordo posto
posteriormente all’autoclave (FIG.2–pos.2)
•
Scaricare sempre il serbatoio di acqua utilizzata, tramite l’apposito raccordo posto
posteriormente all’autoclave (FIG.2–pos.5)
•
Indicare per iscritto, ed inserire nell’imballo un documento ove si indichi con
precisione l’anomalia riscontrata o il servizio di cui si intende beneficiare.
•
Spedire sempre in porto franco, diversamente saranno addebitate le spese di
trasporto sostenute.
ATTENZIONE:
Tutti gli imballi non originali che ci perverranno, verranno smaltiti.
Le autoclavi Vi saranno ritornate con imballi nuovi ed originali, questo per assicurare
massima tutela alla vostra autoclave durante il trasporto. Il costo dell’imballo vi sarà
addebitato.
39
ITALIANO
19
PROCEDURE PER SERVIZIO ED ASSISTENZA
Il reparto riparazioni è una entità autonoma di MEDILINE ITALIA SRL (TECNO-GAZ
GROUP), costituito da una struttura articolata e professionale, in conformità con quelle
che sono le nostre procedure espresse nel nostro “Manuale Qualità”.
•
In caso di guasto l’autoclave, viene presa in consegna direttamente dal reparto
riparazioni che la codifica e crea la scheda di assistenza tecnica.
•
Entro un massimo di 48 ore viene visionata dai tecnici e viene stilata la diagnosi
tecnica con preventivo di spesa per il servizio richiesto. Se l’autoclave non viene
resa come richiesto nel capitolo 18, i tempi espressi si possono dilatare
sensibilmente.
•
Il preventivo di spesa verrà inoltrato al cliente distributore che lo trasmetterà al
cliente finale, per presa d’atto e sottoscrizione.
•
Dopo aver ricevuto il preventivo sottoscritto per accettazione, l’autoclave verrà
messa in lavorazione e verrà spedita nei tempi, indicati sul modulo del preventivo.
•
La merce verrà rispedita con imballo originale mediante spedizioniere del cliente.
IMPORTANTE:
Se dopo aver eseguito i rimedi, il problema si ripresenta, interpellare l’assistenza
telefonica, presso MEDILINE ITALIA srl +39 0522 - 94 29 97.
40
ITALIANO
20
GARANZIA
1) DURATA: il prodotto è coperto da un periodo di garanzia di 12 (dodici) mesi o 1200
cicli.
2) ESCLUSIONE: sono da escludersi dalla garanzia:
a) le avarie causate da mancata manutenzione ordinaria dovute a trascuratezza
dell’Utilizzatore o ad un uso improprio del prodotto;
b) controlli periodici e manutenzione;
c) riparazioni o sostituzioni di parti soggette ad usura, fragili o di durata non
prevedibile, a meno che non sia provata la loro difettosità al momento della
consegna;
d) guasti relativi a mano d'opera, trasferta del personale tecnico e trasporto;
e) guasti o danni derivanti da un uso improprio e da errori di utilizzo;
f) guasti o danni derivanti da impurità nei sistemi di alimentazione di acqua e aria,
eventi chimici o elettrici straordinari;
g) guasti o danni derivanti da uso di detergenti, disinfettanti, sostanze o processi
sterilizzanti, non espressamente indicati nel manuale d'uso e manutenzione;
h) naturale viraggio delle parti in plastica.
3) VERBALE DI INSTALLAZIONE E COLLAUDO: condizione essenziale per accedere alla
garanzia é la restituzione dell’apposito verbale di installazione e collaudo, debitamente
compilato in tutte le sue parti e controfirmato dall’Utilizzatore stesso e dal Tecnico
Installatore. La restituzione del verbale dovrà avvenire entro 15 giorni dall'installazione,
pena il decadimento della garanzia.
4) LIMITI: la garanzia dà diritto alla sostituzione o riparazione gratuita dei componenti
difettosi. È comunque escluso il diritto alla sostituzione dell’intero apparecchio. Per quanto
riguarda i componenti applicati o integrati prodotti con marchio TECNO-GAZ S.P.A. o da
terzi e dotati di proprio certificato di garanzia, valgono le condizioni, i limiti e le esclusioni
riportati sui certificati stessi.
L’effettuazione di una o più riparazioni nel periodo di garanzia non comporta alcuna
modifica della data di scadenza della garanzia stessa.
5) CONTESTAZIONI: nel caso di contestazione da parte dell’Acquirente sull’applicazione
della garanzia, sulla qualità o sulle condizioni della apparecchiatura consegnata,
l’Acquirente non potrà in alcun caso sospendere e/o ritardare i pagamenti.
6) CONTROVERSIE: in caso di controversia sull’applicazione ed interpretazione del
presente Certificato di Garanzia, sarà competente il Tribunale di PARMA (Italia), ovunque
sia stato stipulato il contratto di acquisto dell’apparecchiatura.
7) DEROGHE: particolari deroghe, concesse in merito alle presenti norme di garanzia, non
implicano il riconoscimento di alcun diritto nei confronti dell’Acquirente e saranno da
considerarsi limitatamente concesse in relazione al caso specifico.
8) ALTRO: per quanto non espressamente previsto dal presente Certificato di Garanzia
valgono le norme contenute nel Codice Civile e nelle Leggi in materia della Repubblica
Italiana.
41
ITALIANO
La garanzia decade se:
a) l’apparecchiatura presenta danneggiamenti dovuti a caduta, esposizione a fiamme,
rovesciamenti di liquidi, fulmini, calamità naturali, eventi atmosferici, o comunque da cause
non imputabili a difetti di fabbricazione;
b) l’installazione non é conforme alle istruzioni di TECNO-GAZ S.p.A. ed é stata eseguita
da personale non autorizzato;
c) l’apparecchiatura venga riparata, modificata o comunque manomessa dall’Acquirente o
da terzi non autorizzati;
d) all’atto della richiesta di intervento in garanzia il numero di matricola del prodotto risulti
asportato, cancellato, contraffatto, ecc.;
e) non venga restituito il verbale di installazione e collaudo compilato e firmato entro 15
giorni dall'installazione;
f) l’Acquirente sospenda e/o ritardi per qualunque motivo il pagamento di qualunque
somma dovuta in relazione all’acquisto e/o alla manutenzione dell’attrezzatura;
g) non venga rispettata la manutenzione periodica programmata o altra prescrizione
prevista nel manuale d'uso e manutenzione.
42
ITALIANO
A
MANUTENZIONE
Una corretta manutenzione dell’autoclave assicura un buon funzionamento della stessa e
un sicuro risparmio in termini di tempo e costi dovuti ad assistenza. Le seguenti operazioni
sono da intendersi obbligatorie ed eseguibili dagli operatori.
Pulizia della camera
20 cicli oppure 1 volta la settimana
Pulire periodicamente la camera, asportando eventuali depositi o detriti, eviterete così di
immettere nel circuito di scarico materiali che possono creare ostruzioni. Per una buona
pulizia usare esclusivamente acqua demineralizzata e la spugna abrasiva in dotazione
(parte non abrasiva - FIG.13).
Da eseguire assolutamente a camera fredda per evitare scottature - Non utilizzare mai
solventi, detergenti, soluzioni chimiche, disincrostanti o altri prodotti similari.
Pulizia del filtro camera
20 cicli oppure 1 volta la settimana
Tirare verso l’alto il filtro (cod.ricambio DXBA091), prestando attenzione a non
danneggiarlo, lavarlo con acqua demineralizzata e asciugarlo con panno asciutto e
pulito. Ricollocarlo quindi nella sua sede, facendo attenzione che sporga di circa 15mm.
Pulizia dei tray e dei portatray
20 cicli oppure 1 volta la settimana
Pulire con spugna in dotazione (parte non abrasiva) imbevuta di acqua demineralizzata
Cambio del filtro batteriologico
200 cicli oppure quando assume
una colorazione scura
Sostituire il filtro batteriologico (FIG.2–pos.11) ruotandolo in senso antiorario per svitarlo
e in senso orario per avvitarlo. Utilizzare esclusivamente filtri originali (cod.ricambio
DAVA101)
Pulizia della guarnizione del portello
20 cicli oppure 1 volta la settimana
Periodicamente asportare eventuali residui che si depositano sulla circonferenza della
guarnizione (cod.ricambio DANA038), utilizzando la spugna in dotazione (parte non
abrasiva) inumidita con acqua demineralizzata
A-1
ITALIANO
Regolazione portello
ogni 2 mesi
Per preservare l’integrità funzionale della macchina, bisogna regolare la pressione di
chiusura del portello, agendo sul regolatore del portello stesso, utilizzando la chiave a
doppia funzione (FIG.10) in dotazione (cod.ricambio DANA008). Ruotare in senso
antiorario (pos.A) per aumentare la pressione di chiusura. Ruotare in senso orario
(pos.B) per diminuire la pressione di chiusura.
A
Sostituzione carta stampante
B
-
Rotolo di carta termografica di larghezza 57 mm (cod.ricambio DNVA363). La carta
termografica deve essere usata solo da un lato. Aprire il portello della stampante
premendo il pulsante inserire il rotolo di carta prestando attenzione al verso di rotazione.
Tirare la carta fino a farla uscire e chiudere lo sportello; la stampante è ora pronta per la
stampa.
A-2
ITALIANO
Le seguenti operazioni sono consigliate dal produttore. Tali operazioni devono essere
eseguite esclusivamente da tecnici formati e autorizzati dal produttore.
Sostituzione del filtro carico acqua
ogni 1 anno / ogni 1000 cicli *
Controllo del filtro acqua ingresso flussimetro
ogni 1 anno / ogni 1000 cicli *
Pulizia filtro Y in ottone a valle del radiatore
ogni 1 anno / ogni 1000 cicli *
Pulizia dell’elettrovalvola carico acqua EV4
ogni 1 anno / ogni 1000 cicli *
Sostituzione Valvole Pompa del vuoto
ogni 1 anno / ogni 1000 cicli *
Pulizia del radiatore
ogni 1 anno / ogni 1000 cicli *
Pulizie delle ventole
ogni 1 anno / ogni 1000 cicli *
Controllo integrità circuito pneumatico
ogni 1 anno / ogni 1000 cicli *
Controllo integrità circuito elettrico
ogni 1 anno / ogni 1000 cicli *
Controllo isolante termico camera di sterilizzazione
ogni 1 anno / ogni 1000 cicli *
Controllo serraggio viti macchina e fascia riscaldante
ogni 1 anno / ogni 1000 cicli *
Controllo sonde di temperatura
ogni 1 anno / ogni 1000 cicli *
Controllo valvola di sicurezza
attenersi alle indicazione del
produttore (vedere istruzioni
allegate)
Ingrassare la parte mobile della leva maniglia
ogni 1 anno / ogni 1000 cicli *
Controllo sonde di livello
ogni 1 anno / ogni 1000 cicli *
Controllo stato serbatoio
ogni 1 anno / ogni 1000 cicli *
* vale la condizione che viene raggiunta prima
Inoltre alcune disposizioni regionali/nazionali, stabiliscono l’obbligatorietà di una
validazione della macchina che ne controlla i parametri di sterilizzazione. La validazione
periodica deve essere sempre eseguita da tecnici specializzati tramite opportuni strumenti
di misurazione. Nel caso di non obbligatorietà la validazione è una operazione volontaria
che ha lo scopo di certificare che l’autoclave è perfettamente funzionante ed idonea alla
sterilizzazione dei carichi per cui è stata progettata e costruita.
A-3
ITALIANO
NOTE
A-4
ENGLISH
Dear Client,
Thank you for having chosen our autoclave, we know how to exchange your fidelity, with
maximum attention and service definitely corresponding to your expectations.
The autoclave you have chosen is the maximum expression of applied technology. It can
be used for dental, medical, and aesthetic applications and in general in all the sectors in
which there is the need of performing a safe sterilization of tools and materials of any kind:
solid, hollow, and porous loads, both wrapped and non-wrapped.
Before using this autoclave, we invite you to read with maximum attention the user’s
manual and then keep it in a place accessible to all operators in charge of sterilization.
The autoclave must be used only and exclusively for the uses foreseen by the constructor.
For installation, maintenance and assistance ask exclusively for a technician
authorized. We invite you to use and ask for exclusively original spare parts.
1
ENGLISH
TABLE OF CONTENTS
1. CONSULTING THE MANUAL
1.1
GLOSSARY
1.2
FRONT VIEW / REAR VIEW DESCRIPTION
2. SAFETY
2.1
SAFETY MARKING
2.2
SAFETY DEVICES
2.3
SAFETY NOTICES
2.4
DISPOSAL
3.
AUTOCLAVE USE AND APPLICATIONS
4. TECHNICAL DATA
4.1
MECHANICAL DATA
4.2
ELECTRICAL DATA
4.3
CHAMBER
4.4
CLEAN WATER TANK
4.5
USED WATER TANK
4.6
BACTERIOLOGICAL FILTER
5.
ACCESSORIES
6.
UNPACKING
7.
INSTALLATION
8. USAGE INSTRUCTION
8.1
TURNING ON AUTOCLAVE
8.2
LOADING PURE WATER TANK
8.3
CHARACTERISTICS OF WATER TO USE
8.4
LOADING MATERIALS IN AUTOCLAVE
8.5
BEGINNING STERILIZATION CYCLE
8.6
END OF CYCLE
8.7
UNLOADING STERILIZED MATERIAL
8.8
DISCHARGING USED WATER
8.9
INTERRUPTION OF STERILIZATION CYCLE
9. STERILIZATION CYCLES
9.1
CYCLES DESCRIPTION
9.1
OPERATION CYCLES
9.2
NIGHT CYCLES
9.3
TEST CYCLES
10. AUTOCLAVE VALIDATION TESTS
10.1
CHEMICAL INTEGRATORS
10.2
BIOLOGICAL INDICATORS
10.3
BOWIE-DICK TEST
2
ENGLISH
10.4
10.5
HELIX TEST
VACUUM TEST
11. TABLE AND DIAGRAMS
12. CYCLE PRINTOUT READING
13. ERROR MESSAGES AND ALARMS
14. PROGRAMMING AND PARAMETERS DISPLAY
14.1 LANGUAGE SELECTION
14.2 HOURS SELECTION
14.3 MINUTES SELECTION
14.4 SECONDS SELECTION
14.5 DATE SELECTION
14.6 MONTH SELECTION
14.7 YEAR SELECTION
14.8 VISUALIZE ATMOSPHERIC PRESSURE
14.9 WATER LOADING SELECTION
15. SOLUTION TO OPERATION PROBLEMS
15.1 AUTOCLAVE DOES NOT DRY CORRECTLY
15.2 AUTOCLAVE CHAMBER BECOMES WHITE
15.3 AUTOCLAVE CHAMBER PRESENTS GREEN STAINS
15.4 INTERRUPTION OF THE STERILIZATION CYCLE
15.5 AUTOCLAVE DOES NOT RECEIVE COMMANDS
15.6 AUTOCLAVE PRINTER DOES NOT PRINT
15.7 STAINS ON INSTRUMENTS
16. DEMINERALIZER
16.1 SET OF DEMINERALIZER SYSTEM
16.2 CONNECTION AND PLUG OF DEMINERALIZER SYSTEM
16.3 CHANGE RESINS OF FILTER
17. STANDARD STERILIZATION PROCEDURE
18. INSTRUCTIONS FOR AUTOCLAVE SHIPPING
19. PROCEDURES FOR SERVICE AND ASSISTANCE
20. GUARANTEE
A. MAINTENANCE
3
ENGLISH
CONSULTING THE MANUAL
1
1.1
Glossary
VACUUM
Pressure below atmospheric pressure
CHAMBER
Cylindrical steel body that, during the cycle, is filled with saturated steam
FILTER
TRAY
TRAY HOLDER
Device for retaining particles
Drawer for holding the load to be sterilized
Structure for holding the trays in the sterilization chamber
DEMINERALIZER
Device for eliminating mineral salts in mains water
HELIX TEST
BOWIE - DICK TEST
Test for checking the penetration of steam in a hollow object
Test for checking the penetration of steam in a porous object
VACUUM TEST
Test for checking that the degree of vacuum in the chamber is
maintained
BACTERIOLOGICAL
FILTER
OVER PRESSURE
OVER
TEMPERATURE
RESINES
1.2
Device for retaining particles in suspension in the air larger than 0.3µm.
Pressure higher than the standard one for the cycle considered
Temperature higher than the standard one for the cycle considered
Replaceable component (cartridge) of the demineralizer that eliminates
all inorganic salts
Front view / Rear view description
Spacer
Clean water overflow
Clean water drain tap
(Demineralizer tap)
Safety Valve
Used water overflow - condensate drain
Used water drain tap
Main power supply
electrical socket only demineralizer
Fuse 12A
FIG.1
Fuse 12A
Fuse 500mA
4
ENGLISH
FIG.2
5
ENGLISH
Bacteriological filter
Manual water filling with funnel
Water filling with pump
Used water draining
Display
Printer
RS232 Connector
B-TEST Connector
General switch
121°C Cycle
(valid for wrapped and unwrapped materials)
134°C Cycle
(valid for wrapped and unwrapped materials)
Cycle 134°C Flash
Cycle 134°C Prion
Vacuum Test
Helix / Bowie&Dick Test
Clean Water Minimum Level
Clean Water Maximum Level
Used Water Maximum Level
B-START
START-STOP Button
B-PUMP
Button for water filling with pump
B-SELECT
B-DOOR
6
Cycle selection button
Door opening button
ENGLISH
2
2.1
SAFETY
Safety Marking
HAZARDOUS VOLTAGE
DISCONNECT THE POWER
BEFORE REMOVING THE LID
HIGH TEMPERATURE
EARTH CONNECTION
2.2
Safety Devices
The following safety devices are installed:
-) Safety valve set at 2.4 bar 0/+10%
-) Electromagnetic lock to prevent the door from opening while the cycle is running
-) Resistance over temperature thermostats
2.3
Safety Notes
• The manufacturer is liable for the marketed product in accordance with current
regulations. The manufacturer's liability will expire when operations are carried out on
the device, or a part of it, by unskilled personnel or using non-original spare parts.
• There should be no potential risk of explosion and/or fire in the room where the
autoclave is installed.
• The autoclave should be installed in a special well-ventilated room.
• The incubator’s connector (FIG.2-pos.18), should be used only for the B-TEST.
Since it is a 230 V a.c. connection independent from the main switch, do not
touch the connector with wet or moist hands and protect it against possible
water splashes.
7
ENGLISH
2.4
Disposal
This product is subject to Directive 2002/96/EC of the European
Parliament and the Council of the European Union on Waste of Electrical
and Electronic Equipment (WEEE) and, in jurisdictions adopting that
Directive, is marked as being put on the market after August 13, 2005, and
should not be disposed of as unsorted public waste. Please utilise your
local WEEE collection facilities in the disposition and otherwise observe all
applicable requirements.
3
AUTOCLAVE USE AND APPLICATIONS
The autoclave is able to sterilize the three types of load provided for by the standard
EN13060, especially:
METAL OR SOLID
MATERIALS
Instruments with no cavities and no obstacles to the penetration
of steam
POROUS OBJECTS
Simple or composite materials that can absorb fluids (fabrics,
gowns, surgical gauzes, dressings, etc)
HOLLOW OBJECTS
Materials or devices with cavities, obstructions, etc.
These are subdivided into two types, classified according to the
length and diameter of the cavity. Approximately:
TYPE B: cannulas, tubes or devices with large passages.
TYPE A: turbines, hand pieces and devices with blind or small
holes
This device has been certified for the sterilization of the following materials *:
Wrapped and unwrapped solid
materials for a maximum load of
Type A/B hollow materials for a
maximum load of
Porous materials for a maximum
load of
EUROPA B EVO
EUROPA B EVO 24
max kg. 5
max kg. 7
max kg. 5
max kg. 7
max kg. 1,5
max kg. 2
* Only for European Countries
The autoclave shall only have to be used for the sterilization of tools
and materials being compatible with the steam sterilization system.
Always make sure that the loads that need to undergo sterilization
can stand the temperatures scheduled for the selected cycle.
8
ENGLISH
4
4.1
TECHNICAL DATA
Mechanical Data
Working temperature
Maximum altitude
MAX relative humidity at 30°C
MAX relative humidity at 40°C
Dimensions of space occupied
(L x H x P)
Space occupied with open door
Weight (tank empty)
Weight (tank full)
Weight of area of support
Potential sound level
4.2
510 X 390 X 590 mm
510 X 390 X 730 mm
300 mm
58 kg
67 kg
2058 N/m2
54 kg
63 kg
2058 N/m2
< 70 db A
Electrical Data
Power voltage
Power
Frequency
Power cord
Fuses
Heat transmitted
4.3
EUROPA B EVO
EUROPA B EVO 24
+5°C ÷ +40°C
2.000 m
80%
50%
EUROPA B EVO
EUROPA B EVO 24
230 V a.c. +/-10 % single phase
1.7 kW
2.2 kW
50 / 60 Hz
2 + 1 x 1mm2
6. 3 x 32 – 12A (x2) / 6.3 x 32 – 500mA (x1)
3.6 E6 J / ora
Chamber
MAX working pressure
MAX empty
MAX Temperature
Material
Size
EUROPA B EVO
EUROPA B EVO 24
2.4 bar (related)
- 0.9 bar (related)
138 °C
Inox AISI 304
Ø 245 x 318 mm
Ø 245 x 500 mm
9
ENGLISH
4.4
Clean Water tank
EUROPA B EVO
EUROPA B EVO 24
With Demineralizer
Volume
Usable cycles
4,5 l
Read demineralizer
instruction
4
Used Water tank
EUROPA B EVO
Volume
Usable cycles
Material
Max temperature used water
4.6
EUROPA B EVO 24
4,5 l
4
2
polyethylene
50°C
Bacteriological filter
EUROPA B EVO
Diameter
Filtering capacity
10
2
polyethylene
Material
4.5
Without
Demineralizer
EUROPA B EVO 24
56 mm
0.3 µm
ENGLISH
05
ACCESSORIES
Tray holder
EUROPA B EVO
EUROPA B EVO 24
Aluminium anodized
192 x 165 x 280 mm
192 x 200 x 470 mm
Material
Size (L x H x P)
Picture
FIG.4
FIG.3
Envelope standard
1
Tray holder (optional)
EUROPA B EVO
Aluminium anodized
193 x 200 x 280 mm
(200 x 193 x 280) mm
Material
Size (L x H x P)
EUROPA B EVO 24
/
/
Picture
/
FIG.5
Envelope standard
1
/
Trays
EUROPA B EVO
Material
Size (L x H x P)
EUROPA B EVO 24
Aluminium anodized
184 x 17 x 286 mm 184 x 17 x 286 mm + 184 x 17 x 140 mm
Picture
+
FIG.6
Envelope standard
4
FIG.7
FIG.8
4+4
11
ENGLISH
Trays (optional)
EUROPA B EVO
Material
/
Size (L x H x P)
/
Picture
/
Envelope standard
/
EUROPA B EVO 24
Aluminium anodized or
stainless steel
184 x 17 x 460 mm
FIG.9
4
Tray extraction and door adjustment wrench
Use for extract the trays (FIG.11) and for door adjustement (par.14.4)
Picture
FIG.11
FIG.10
Envelope standard
1
Plastic funnel with pipe
Use to charge water in manual metod (par.8.2)
Picture
FIG.12
Envelope standard
1
Chamber and door gasket clearing sponge
Use to clean sterilization chamber and door gasket (par.14.2-14.3)
Picture
FIG.13
Envelope standard
12
1
ENGLISH
Connection for water filling pipe and plastic funnel
Use with plastic funnel and water filling pipe with filter
Picture
FIG.14
Envelope standard
1
Water filling pipe with filter
Use to charge water with pump (par.8.2)
Picture
FIG.15
Envelope standard
1
Water discharghe pipe
Use to empty used water tank (par.8.9)
Picture
FIG.16
Envelope standard
1
Rilsan pipe
Connect one pipe end into the used water overflow - condensate drain pipe fitting (FIG.1 –
pos.4), and put the other end in a tank
Picture
FIG.17
Envelope standard
1
13
ENGLISH
Rear spacer
Put the spacer in the autoclave’s back panel (FIG.1–pos.0) It’s necessary for guarantee a
good ventilation if you place the autoclave near a wall.
Picture
FIG.18
Envelope standard
1
Pipes for discharge utilities
1- Pipe for clean water overflow: Connect one pipe end into the water overflow pipe fitting
(FIG.1–pos.1), and put the other end in a tank or in the discharge (demineralizer version).
2- Pipe for discharging used water: Connect one pipe end at the pipe fitting PICT.20 and
screw it into the used water tap (FIG.1–pos.5), then put the other end in a tank or in the
discharge.
Picture
FIG.19
Envelope standard
2
Pipe fitting for back discharge used water
Connect the pipe for discharging used water into the pipe fitting for back discharge used
water PICT.20 and screw into the used water tap (FIG.1-pos.5).
Picture
FIG.20
Envelope standard
14
1
ENGLISH
Power supply cord
Take the power supply cord provided and insert the female plug (FIG.21–pos.B) in the
socket of the back panel of the autoclave (FIG.1–pos.6). Then insert the male plug
(FIG.21–pos.A) in the electric plug of the system.
Picture
FIG.21
Envelope standard
06
1
UNPACKING
The autoclave is shipped in a suitable package to be transported and moved easily and to
protect its contents.
The package must not be jostled, it must be moved with care avoiding letting it turn over or
fall from a height superior to 16 cm.
In case autonomous handling means are not available handle the packaged autoclave
always in two persons.
Autoclave is placed on wooden pallet and packed with corrugated cardboard internal and
external application.
For remove the autoclave to its pack please remove first the corrugated cardboard.
For lift the autoclave please use the belts.
The autoclave shall have to be handled by at least two people and by using
belts only.
Do not lift the autoclave from the inferior part of the door or command
panel, this incorrect operation can create problems of a mechanical nature.
ATTENTION: Always conserve original packaging.
Inside the package you will find:
• USER’S MANUAL : to be read with attention and kept in a place available to all
operators assigned to sterilization.
• CERTIFICATION : which must be conserved.
• GUARANTEE LABEL: which must be completed and send.
15
ENGLISH
07
INSTALLATION
FIG.22
EUROPA B EVO
A
B
C
D
E
F
EUROPA B EVO 24
610 mm
590 mm
590 mm
50 mm
410 mm
426 mm (front)
455 mm (rear)
400 mm
426 mm (front)
455 mm (rear)
400 mm
1 - Install the autoclave in an environment suitable for carrying out sterilization procedures.
2 - The room must be adequately lit and ventilated, as provided for in the directives in
force.
3 - Install the autoclave away from heat sources and water splashes.
4 - Position the autoclave on a surface that is able to bear its weight. The minimum
recommended load bearing capacity of the surface is 80 kg.
5 - The supporting surface must be perfectly horizontal and must not have inclinations.
6 - Place the autoclave at a height that allows the user to inspect and easily clean all parts
of the sterilization chamber.
7 - Open the door of the autoclave and remove from inside the sterilization chamber all
packages which contain the single accessories.
16
ENGLISH
8 - Leave inside the sterilization chamber only the tray carrier with the trays. All other
accessories should be positioned in a separate space available to operators.
9 - Never place on the autoclave newspapers, trays, liquid containers, etc…
10 - Do not lean on the door when it is open.
11 – Leave a space of at least 5 cm in the rear of the autoclave, using the spacer (FIG.1–
pos.0 / FIG.18), and sides of the unit to ensure the ventilation required
12 - Connect the overflow connectors to the special drains with the pipes provided
(chapter 5)
13 - Always make sure the electrical system to which the autoclave is to be connected is in
conformity with the legislation in force and sized to suit the specifications of the said
appliance.
14 - Take the power supply cable supplied and plug the plug on the rear panel of the
autoclave into the female socket (FIG.1 –pos.6)
15 - Connect the electric plug to the system and make sure the power supply is 230 Vac 2000 Va minimum
IMPORTANT:
Do not make the connections using extensions, reducers or adapters as this could
create micro outages with consequent generation of alarm signals.
16 - Turn on the autoclave by pushing the general switch (FIG.2–pos.19) and open the
door to the autoclave. Wait a few seconds, there will be two beeps which indicate the
reaching of relative parameters for automatic atmospheric alignment, at the same time the
display will indicate “OPEN DOOR”.
IMPORTANT:
Never select a control before hearing the two acoustic signals: the autoclave will
not accept the selected program
17 - Load the pure water tank following the indications in the section entitled usage
instructions (chapter 8).
Successively run the test cycle, always following the indications expressed in the section
usage instructions (chapter 8).
18 - If the autoclave is installed with the demineralizer system, follow the instructions
contained in the packaging of this device and the demineralizer section (chapter 16).
17
ENGLISH
08
USAGE INSTRUCTION
After having installed the autoclave, proceed with preparations and use.
8.1
Turn on the autoclave
Press the general switch (FIG.2–pos.19)
8.2
Loading pure water tank
Connect the water filling pipe (provided) with the connection for water filling pipe and
plastic funnel PICT.14. Then engage the connection in the charging hole (FIG.2–pos.13).
Insert the other end of the pipe with filter in the demineralized or distilled water container.
Press the button B-PUMP to start the water loading pump and press it until the countdown
start.
The pump loads the tank inside the autoclave. If maximum level is not reached within 180
seconds the pump automatically stops, therefore it is necessary to press the button again.
The pump stops automatically when maximum level is reached.
If you wish to use this method of filling or in the event of malfunctioning of the pump, the
filling of the tank can be carried out manually by the operator in the following way:
•
•
•
•
•
•
Turn on autoclave.
Remove the connection cover (FIG.2–pos.12)
Insert in this opening the hose connected to the funnel (FIG.14)
Pour distilled water in the funnel, keeping it higher than the loading top.
Add until the led of maximum level turns on (FIG.2–pos.2).
Loading completed, twist on the top.
With connection to a deionizer, if the maximum water level has not been reached, the
functioning of the autoclave will be obstructed.
18
ENGLISH
8.3
Characteristics of the water to be used
TABLE SHOWING THE QUALITY LEVELS LAID BY THE UNI EN 13060 : 2009
CEN STANDARD UNI EN 13060 : 2009
Evaporation residue
10
mg/l
≤
Silicon oxide
1
mg/l
≤
Iron
0.2
mg/l
≤
Cadmium
0.005
mg/l
≤
Lead
0.05
mg/l
≤
mg/l Remains of heavy metals apart from iron,
0.1
mg/l
≤
cardium, lead
Chloride (Cl’)
2
mg/l
≤
Phosphate (P20s)
0.5
mg/l
≤
Conductivity (at 20°C)
15
≤
µs/cm
Ph value (acidity level)
5 ÷ 7,5
Transparent, clear, without deposits
Appearance
Hardness (and alkaline earth ions)
0.02
mmol/l
≤
8.4
Loading the material into the autoclave
Place the materials to be sterilized on the trays provided:
• never superpose the materials
• arrange the wrapped instruments, with the paper part facing upwards
• never bring the materials into contact with the sterilization chamber or the closing door
• put the scissors and dental forceps with the blade open
FIG.23
When you have place all the instruments close the door and
CLOSED” on the display.
you can read “DOOR
19
ENGLISH
8.5
Starting the sterilization cycle
Choose sterilization program most adaptable of load prepared and press the button
B-SELECT.
Each individual pressure changes program selection to the next.
After having chosen the program start the cycle by pressing the button B-START, the door
will lock automatically and the cycle will begin.
During the cycle, the display will show all parameters and relative information.
8.6
End of cycle
An acoustic signal will warn the operators that the sterilization cycle has been completed
and the “END OF CYCLE “ message will be displayed.
When the acoustic signal stops, release the door by pressing the B-DOOR button. If there
is pressure inside the chamber, the button will not activate the release device. Wait until
the chamber is completely depressurized, then repeat the operation or, in emergency
situations only, simultaneously press B-DOOR + B-PUMP. When the door is released, pull
the door handle to open it.
8.7
Unloading, preserving and storing the sterilizied materials
Wear personal protection equipment in accordance with the existing laws.
Extract the trays using the special spanner provided (FIG.10), and place the autoclave
load on a clean, dry surface Then store the treated instruments in the special, prepared
compartments.
8.8
Discharging used water
When the led of used water level (FIG.2–pos.28) goes on, proceed with emptying. If not
emptied, the autoclave function is inhibited.
Take the pipe provided (FIG.16), and insert it in the used water discharge plug, on the
front of the autoclave (FIG.2–pos.14), unscrew the metal ring anticlockwise (to execute 2
complete turns), simultaneously place the other end of the pipe in a container, the water
will descend into the same container.
IMPORTANT:
A - The hose located in the receiving container, must not ever, touch or be immersed in
the discharged water, otherwise there will be a situation of sucking up.
B - Always wait until the unloading water is totally emptied. The led of maximum used
water level will turn off when there is still water in the tank, therefore do not use as a
reference for this operation.
To the term of the drainage to screw the metal ring and remove the tube.
20
ENGLISH
8.9
Interruption of sterilization cycle
A sterilization cycle can be voluntarily interrupted by pressing the button B-START for at
least 2 seconds. The autoclave will emit a beep, go into decompression and on the display
will appear the message ALLARM 001 (INTERRUPTED CYCLE).
To reset error, maintain pressed simultaneously the buttons below indicated, until the
intervention of an acoustic signal:
B-SELECT
B-PUMP
B-DOOR
ATTENTION: You can’t reset the alarm if the printer works
21
ENGLISH
09
9.1
STERILIZATION CYCLES
Descriptions cycles
The autoclave has three series of cycles:
A - operation cycles
B - night cycles
C - test cycles
9.2
Operation cycles
All operation cycles have a system of fractioned vacuum, which therefore permits
sterilization of materials that are hollow, porous, solid, both free and packaged.
Temperatures can be selected from 121°C – 134°C.
Normally the cycles of 121°C are used for thermoplasti cs or sensitive materials, while the
134°C cycles are used for all other materials.
In all cases always follow the indications given by the manufacturer of the instruments or
devices to be sterilized.
There are also two special cycles:
- Prion cycle, studied for the Crutzfeldt-Jacobs disease (mad cow disease)
- Flash cycle created for rapid sterilization of unpackaged instruments and devices.
9.3
Night cycles
The autoclave is provided with a special economizer device.
This device allows the execution of sterilization cycles without the presence of an operator.
At cycle end, if the door is not opened, the autoclave stabilizes and then shuts off
automatically, only the general switch remains on (FIG.2-pos.19)
Upon the operator’s arrival it is sufficient to press any button to re-start the autoclave and
read the cycle result on the display. Besides that the printer will have regularly prepared
the written report relative to the same cycle.
9.4
Test Cycles
The available test cycles are the following:
- Helix test
- Bowie-Dick test
- Vacuum test
22
ENGLISH
10
10.1
AUTOCLAVE VALIDATION TESTS
Chemical integrators
These tests (spare code 200/S and 215-S) are based on the properties of indicator dyes
that are able to modify their own aspect if they are adequately exposed to heat and
pressure according to given times, since they interact with all the cycle parameters
(pressure, temperature, time). If the test is successfully completed, the correct stability is
validated.
The chemical integrators are to be put into the sterilization chamber
before starting the cycle, even with the load already in.
USE
The integrators can either be unwrapped and positioned into the trays
or they can be wrapped into the sterilization bags together with the
materials to be sterilised.
In any case it is advisable to wrap the integrators where there are
many operators. As a matter of fact by doing so the sterilization of
every single tool will be certified.
If the colour change remains
within the white area, the test
failed
RESPONSE
if the colour change enters the
SAFE
window,
the
test
succeded
PERIODICITY
10.2
There is no preset testing interval, apart from areas to which specific
regional regulations apply. In any case all the chemical integrators are
the cheapest and quickest available tests, hence in order to
constantly monitor the autoclave performance, it is advisable to use
them on all of the cycles or at least once a day.
Biological indicators
The test (spare code 262-S) is used to prove the autoclave capability as to the complete
destruction of all microorganisms.
It is represented by a standardized spore preparation, having biological characteristics,
high resistance to heat, which is to be deemed an absolutely safe control tool (ATCC
7953). It consists of spore ampoules. It is non pathogenic, non toxic and non pyrogenic.
23
ENGLISH
USE
The ampoules shall have to be put into the sterilization chamber even
when loaded.
Carry out the cycle and then take out the ampoule by taking care
since it is hot and pressurized.
Leave it cool down for about 10 minutes and then activate it by using
the supplied crusher by taking care so as to keep the ampoule always
upright. The chemical indicator placed on the label of the ampoule will
have changed its colour from blue into black.
Subsequently incubate the ampoule into a biological incubator at
57°C for 48 hours, together with an unprocessed test am poule, that
has anyway be activated through the relevant crusher.
RESPONSE
After 48 hours take the processed ampoule out of the incubator and
assess the test result. If the ampoule has changed its colour to yellow
this means that the autoclave has failed the test and there is bacterial
growth. If the ampoule remains purple, this means that there is no
bacterial growth and hence the test succeeded. The test ampoule of
course will always change to yellow since it has not been processed
and will only be used as term of comparison.
At the end of the test dispose of the ampoule in the solid urban waste.
It is anyway advisable to perform another sterilization cycle at 121° C
on the ampoule.
PERIODICITY
10.3
There is no preset testing interval, apart from areas to which specific
regional regulations apply. In any case the biological tests are the
more concrete ones. Hence in order to constantly monitor the
autoclave performance, it is advisable to perform the test at least
once every 90 days.
Bowie-Dick test
It is a physical test (spare code 268/S) enabling to verify steam penetration capability into
porous loads. For the test a “test pack” is used which is standardized and complying with
the technical standards being currently in force.
USE
24
Perform the test with empty chamber. The Bowie-Dick test pack shall
have to be positioned on the autoclave central tray.
Perform the cycle, indicated in the display of the autoclave, and at
the end remove the pack, open it and check the sheet colour change
through the chemical indicator placed inside.
ENGLISH
RESPONSE
PERIODICITY
10.4
the assessment is simple and rapid.
If the colour change turns out to be
uniform (as shown in the figure) the
test perfectly succeeded. On the
contrary the test failed, hence the
autoclave cannot properly sterilize
porous bodies.
There is no preset testing interval, apart from areas to which specific
regional regulations apply. Hence in order to constantly monitor the
autoclave performance, it is advisable to perform the test at least
once every 30 days.
Helix test
It is a physical test (spare code 267-S) enabling to verify steam penetration capability into
hollow loads. For the test a standardized system is used which complies with the technical
standards in force.
Put the test device into the empty chamber.
USE
Fit the suitable indicator strip into the test device end cap, then
position the test device on the central tray of the autoclave.
Perform the suitable cycle, indicated in the autoclave display, then
remove the test device, open the cap and check the indicator strip
colour change.
25
ENGLISH
BEFORE THE TEST
RESPONSE
TEST FAILED
TEST SUCCEEDED
PERIODICITY
10.5
There is no preset testing interval, apart from areas to which specific
regional regulations apply. Hence in order to constantly monitor the
autoclave performance, it is advisable to perform the test at least
once every 30 days.
Vacuum test
It is a chamber leak test or a vacuum leak test. This test aims at verifying that during the
cycle no air infiltration takes place through the chamber seals (gaskets, valves, etc…)
Perform the test with empty chamber.
USE
Select the suitable cycle indicated on the autoclave display. The
cycle is enabled.
The autoclave automatically performs the cycle by complying with
precise technical procedures. Then the final result is printed out
through the printer connection.
RESPONSE
PERIODICITY
The test result is immediate and is printed by the printer which prints
out all the values relating to the performed cycle as well as the final
assessment.
There is no preset testing interval, apart from areas to which specific
regional regulations apply. Hence in order to constantly monitor the
autoclave performance, it is advisable to perform the test at least
once per week.
WE RECOMMEND TO USE ONLY TEST PROPOSED
BY TECNO - GAZ S.p.A
26
ENGLISH
11
TABLES AND CHARTS
STERILIZATION TABLE
18
17
13
1.04 ÷ 1.30
121 ÷ 125
5
17
13
2.02 ÷ 2.38
134 ÷ 138
4
2
4
2.02 ÷ 2.38
134 ÷ 138
20
17
13
2.02 ÷ 2.38
134 ÷ 138
2.02 ÷ 2.38
134 ÷ 138
3.5
4
-0.9
Vacuum
OPERATIVE
CYCLE
121°C
(wrapped and
unwrapped)
134°C
(wrapped and
unwrapped)
134°C
Flash
134°C
Prion
134°C
Helix / Bowie&Dick
TEMPERATURE
WORK MIN AND
MAX ( °C )
TEST
CYCLE
CYCLES
DRYING
PRESS.WORK
EXPOSED
TIME
TIME T4 T5 (Minutes) MIN AND MAX
(Minutes)
(relative bar)
18L 24L
The pre-heating time and fractionation of the vacuum can vary depending on the
conditions of the autoclave from 25 - 35 minutes that will add up with the times of the
cycles shown in table. All sterilization cycles have 3 stages of vacuum.
CYCLE DIAGRAM
T1,T2,T3
pre heater fase and
vacuum fases
T4
sterilization
T5
drying
FIG.24
27
ENGLISH
12
CYCLE PRINTOUT READING
Software release
Serial number
Date and Description of the
selected cycle
Hours – Minutes – Seconds
Chamber pressure
Chamber main temperature
Cycle phase
End of cycle (OK means that the
cycle has been correctly completed)
Cycle number
Operator (blank for signature)
28
ENGLISH
13
MESSAGES OF ERROR OR ALARMS
Messages of error are emphasized through an Alfa-numerical code, consisting in a letter
and three numbers.
The “E” code is relative to errors of the operator, which can be corrected by same
operator.
The “A” code is relative to alarms, irregularities of the autoclave, after having carried out
the remedy, if the problem persists it is necessary to request telephonic assistance.
If an alarm message is displayed (CODE “A”) the cycle is to be
considered aborted: it will be necessary to repeat all the preparation
and sterilization operations.
To reset alarms and errors, maintain pressed the buttons here indicated until the
intervention of a “BEEP”
B-DOOR + B-PUMP + B-SELECT
ERROR
CAUSE
REMEDY
E 200
Low water level
E 250
Low level water cleaned up with loadingfrom
Fill tank
demineralizer
E 300
Maximum level used water
Empty tank
E 401
Door open
Colse door with attention
E 451
Door unlocked
Open door and close again
E 461
High temperature in chamber for vacuum
test (above 40° C)
Shut off and wait 10 minutes
A 250
Cycle begins without water
Control demineralizer device
A 001
Interrupted cycle
Reset and restart cycle
A 011
Irregular pc display
Shut off and restart autoclave
A 101
Vacuum not reached in 10 minutes
Reset and restart cycle
A 111
Vacuum not maintened for first phase VT
Reset and restart cycle
A 121
Vaccum not maintened for second phase VT Reset and restart cycle
Fill tank
29
ENGLISH
A 401
Hatch opende during the cycle or problem
closing
Check the closing system
A 501
Pressure not maintened
Reset and restart cycle
A 551
Irregular 1° temperature sensor
Reset and restart cycle
A 552
Irregular 2° temperature sensor
Reset and restart cycle
A 641
Overly high pressure
Reset and restart cycle
A 651
Overlay high temperature
Reset and restart cycle
A 661
Irregular temperature comparison
Reset and restart cycle
A 701
Working pressure not reached
Reset and restart cycle
A 751
Low temperature
Reset and restart cycle
A 761
Irregular pressor sensor
Reset and restart cycle
A 801
Irregular unloading pressure
Reset and restart cycle
A 901
Main voltage too low - Temporary lack of
electric energy
Reset and restart cycle,
Inform eletric Energy provider
REPLACE RESINS Resins IONIC
SYSTEM exhaust
Replacement of resins; refer to
paragraph 16.3
REPLACE FILTER Resins OSMOSI SYSTEM
exhaust
Replacement of filter; refer to
paragraph 16.3
D
The printer’s door is not
closed
Try to close the printer’s door
P
Paper is missing
Insert paper
Lo
There is low voltage on
the line
Try to connect the machine to the net electrical worker in
another point; if the problem persists to contact a
electrician
30
ENGLISH
14
PROGRAMMING DISPLAY
Please, see the FIG.2.
14.1
Language selection
Press simultaneously B-START and MAINS switch. Hold down B-START until on the
display appears the name of the language programmed.
Press B-SELECT to modify the language [forward].
Press to B-PUMP modify the language [backwards].
14.2
Hours selection
Press B-START the hours appears.
Press B-SELECT to modify hour [forward].
Press B-PUMP to modify hour [backwards].
14.3
Minuts selection
Press B-DOOR minutes appear.
Press B-SELECT to modify minutes [forward].
Press B-PUMP to modify minutes [backwards].
14.4
Seconds selection
Press B-DOOR seconds appear. Do not modify.
14.5
Date selection
Press B-START day appears (flashing).
Press B-SELECT to modify day [forward].
Press B-PUMP to modify day [backwards].
14.6
Month selection
Press B-DOOR month appears (flashing).
Press B-SELECT to modify month [forward].
Press B-PUMP to modify day [backwards].
31
ENGLISH
14.7
Year selection
Press B-DOOR year appears.
Press B-SELECT to modify year [forward].
Press B-PUMP to modify year [backwards].
14.8
Visualize atmospheric pressure
Press B-START and compare memorized atmospheric pressure appears.
DO NOT MODIFY!
14.9
Water loading selection
Press and release B-START to select next screen until you see setting the water load:
• CHARGING BY PUMP
• IONIC SYSTEM
• OSMOSI SYSTEM
Press B-SELECT to modify water supply loading.
Press B-DOOR to save the modify and maintain pressed until the sound of the “BEEP”.
Press B-START to exit menu and maintain pressed until the sound of the “BEEP”.
The data inserted are memorized. Turn off the autoclave using the MAINS switch.
32
ENGLISH
15
SOLUTIONS TO OPERATIONAL PROBLEMS
In many cases some alarms or errors are determined by lack of attention or lack of
familiarity with some technical and operational aspects. Following are listed some cases of
irregularity with relative solutions.
15.1
Autoclave does not dry correctly
• substitute the bacteriological filter, with a new original.
• original trays have not been used, but other trays without perforations or diverse
perforations. use only original trays, possibly in aluminum. request additional original
trays.
• Instruments have not been arranged correctly. Follow the indication of par.8.4
15.2
Autoclave chamber becomes white
• Change immediately the type of water used, use demineralized or distilled water, as is
specifically indicated in the previous chapter and then proceed with cleaning the
chamber.
• The whitish color can be a consequence of the evaporation of organic materials present
on the instruments. instruments should be cleaned with a detergent action more suitable
and thoroughly.
• check the demineralizer (IONIC SYSTEM or OSMOSI SYSTEM)
15.3
Autoclave chamber presents blueish green stains
• Instruments have not been correctly rinsed cleanse and rinse with major attention the
instruments. if the stains are evident request telephonic assistance.
15.4
Sterilization cycle interrupts without apparent motive
• Control if autoclave is connected to the electrical net with extensions, reductions,
adapters, if so remove this accessories and connect the autoclave directly to the electric
socket.
15.5
Autoclave does not receive commands
• Autoclave pre-arranges bar automatic alignment wait for the double beep after opening
the door, then program the functions.
• Pure water tank is empty, minimum led level is on, fill with pure water
• Used water tank is full. maximum level led is on, discharge used water.
33
ENGLISH
15.6
Autoclave printer does not print
• Paper carrier cover is not closed correctly, on the display appears the letter “d”. open
and close the printer cover checking that is correctly closed.
• Paper missing, on the display appears the letter “p”. open the cover and insert the roll
of thermo-graphic paper.
• Roll of paper is mounted with thermo side towards inside, open the cover and turn the
roll of paper to the correct position.
15.7
Stains on instruments
• Instruments become yellow, chemical liquid residual has with heat been fixed on the
instruments. they were not sufficiently rinsed.
• Sterilization chamber presents yellow spots, instruments were introduced in the
chamber with chemical liquid which dripped and is fixed due to heat. rinsing was not
sufficient.
• Instruments present white spots, rinsing was done with water with high level of calcium
and the instruments were not dried. for last rinse it is advisable to used demineralized
water and dry carefully the instruments.
• Instruments are darkened, this is caused by the fact that the instruments have internally
a strong component of carbon.
34
ENGLISH
16
16.1
DEMINERALIZER
Set of demineralizer system
Autoclave is already prepared also for plug IONIC SYSTEM or OSMOSI SYSTEM for
charging water into the autoclave (optional accessories).
These applications are accessories of autoclave.
These applications have been designed for remove minerals in the water before to charge
the autoclave.
First to plug the hydraulic connections it is necessary set the software of autoclave
following this instructions.
Switch-on the autoclave with the MAINS (FIG.2–POS.19) switch and keep pressed
B-START ((during the switch on) until you see a string reporting the selected language.
Press and release B-START to select next screnn until you setting the water load:
• CHARGING BY PUMP
• IONIC SYSTEM
• OSMOSI SYSTEM
To change this setting use B-SELECT.
To save keep pressed B-DOOR until you hear a “BEEP”.
To exit keep pressed B-START until you hear a “BEEP”.
Set data are stored in memory.
WARNING:
The number you can see on display, under the kind of water system used, it’s a counter. It
gives you an indication about the number of cycles possible to make until the end of filter
(in case of osmosi system) or until the end of resins (in case of ionic system).
When you change the filter or the resins, it’s necessary reset the number of the counter
after saving of the kind of system used.
For reset the counter it’s necessary keep push the button B-PUMP until the acoustic signal
of the buzzer (it takes about 10 second) in the main page of the menu where you can start
the sterilization cycle.
35
ENGLISH
16.2
Connection of demineralizer system
Here are indicated the specific connection to the autoclave of the water supply hose and
the connection of the electric plug.
•
•
•
•
•
•
•
•
•
•
•
Turn off the autoclave if it is on (FIG.2–pos.19)
Close the faucet located above the demineralization system
Install the demineralizer as indicated in its manual;
Twist the male screw of the connection-hose carrier with Teflon or other component
which guarantees the sealing of water;
Screw the connection on the Demineralizer tap (FIG.1–pos.2)
Insert the hose exit of demineralizer to the connection-hose carrier already screwed to
the autoclave;
Insert the plug of demineralizer to the socket (FIG.1–pos.7) on the back of the
autoclave
Open the faucet located above the demineralization system;
Control that there is no water loss;
Turn on the autoclave;
Run one or more sterilization cycles to control the functioning of the connection and
check above all for leaking
At the end of each day always close the faucet located above the
demineralization system to prevent flooding
Connect of demineralizer only to preprogrammed autoclave
If the unit at the beginning of the cycle shows E250 then you have to fill the
clean water tankuntil the maximun is reached.
NOTE: for the connection of demineralization system to the autoclave tecno-gaz
refer to indications in the manual.
36
ENGLISH
FIG.31
16.3
Change resins - filter
Autoclave display indicates the state of finishing of the resin with the message “REPLACE
THE RESINS” (spare code 242-S) or “REPLACE THE FILTER” (spare code 248-S-2).
Substitute the resin cartridge as indicated in the deminaralizer manual. After substitution,
press button B-PUMP until the “beep”, to reset the autoclave.
37
ENGLISH
17
STANDARD STERILIZATION PROCEDURE
Sterilization means adopting precise operational protocol and applying a system.
Autoclave is the final link of this system. Those who do not follow all of the various phases
of the sterilization system will be unsatisfied with the final result. TECNO-GAZ S.p.A.
produces all the devices for each phase of the sterilization cycle. Therefore it is necessary
to:
1 - Prepare a precise written operational protocol to distribute to all operators assigned
to sterilization.
TECNO-GAZ S.p.A. can provide these personalized protocols.
2 - Follow carefully and with attention all of the sterilization cycle phases:
DISINFECTION obligatory phase, in base of that is obligatory for the operator
security, to be done with immersion in chemical liquids or thermo-disinfection.
CLEANING the most important phase which insures the removal of all types of
residual, chemical and organic. The most suitable instruments are ultrasound baths.
DRYING essential phase to avoid corrosion of instruments and interference of the
sterilization cycle.
PACKAGING essential phase for maintaining sterile the treated instruments.
STERILIZATION final phase of steam sterilization
38
ENGLISH
18
INSTRUCTION FOR AUTOCLAVE SHIPPING
In case the autoclave must be shipped for repairs, controls, reactivation, revisions,
validations follow the obligatory indications below:
•
Use the original packaging; if this is no longer in your possession, use adequate
packaging. The merchandise travels at risk to the sender.
•
Ship the autoclave only (do not include any component contained in the
accessories kit).
•
Carefully clean the sterilization chamber and autoclave in general before shipping.
In case it arrives dirty or with residual the autoclave with be returned without being
repaired, or it will be put through a cleansing action and disinfection.
•
Always empty the clean water tank through the attachment located on the back of
the autoclave (FIG.2–pos.2)
•
Always empty the used water tank through the attachment located on the back of
the autoclave (FIG.2–pos.5)
•
Indicate by letter and insert in the package a document which indicates precisely
the irregularity or service desired.
•
Ship at your expense, otherwise you will be billed for shipping
ATTENTION:
All non original packaging which arrives will be disposed of.
Autoclave will be returned with new and original packing to insure maximum protection for
your autoclave during shipping. Cost of packaging will be charged to client.
39
ENGLISH
19
PROCEDURE FOR SERVICE AND ASSISTANCE
The repair service dept is an autonomous division of MEDILINE ITALIA SRL (TECNOGAZ GROUP), comprising a complex and professional structure, in compliance with the
procedures set in our “Quality Assurance Manual”.
•
When the autoclave arrives at MEDILINE ITALIA SRL it is consigned directly to the
service center, which codes and creates the technical assistance document.
• Within 48 hours, it is examined by technicians and technical analysis along with cost
estimate is prepared for the service requested. If the autoclave is not delivered as
requested in the chapter instructions for shipping, above times can sensibly differ.
• Cost estimate will be sent to the distributor client, who will transmit it to the final
client for vision and approval.
•
After MEDILINE ITALIA SRL receives written acceptance of the cost estimate, the
autoclave will be serviced and reshipped according to times indicated on the cost
estimate.
•
Ship at your expense, otherwise you will be billed for shipping.
IMPORTANT:
If after following the remedies, the problem persists, call telephone assistance at
MEDILINE ITALIA srl +39 0522 - 94 29 97.
40
ENGLISH
20
WARRANTY
1) DURATION: the product is covered by a 12 (twelve) month or 1200 cycles guarantee.
2) EXCLUSION: the guarantee excludes:
a) faults originating from non-performance of scheduled maintenance and due to
negligence or improper use of the product by the User;
b) periodic checks and maintenance;
c) repair or replacement of parts subject to wear that are fragile or have an
unpredictable life, unless they prove faulty at the time of their delivery;
d) faults caused by personnel or originating from technical personnel transfer and
transportation;
e) faults or damage originating from improper use or operating errors;
f) faults or damage originating from contamination in the water and air infeed systems,
extraordinary chemical agents or electrical events;
g) faults or damage originating from the use of detergents, disinfecting agents,
sterilizing fluids or processes that are not described explicitly in the instruction
manual for operation and maintenance;
h) natural color change of plastic components.
3) INSTALLATION AND TESTING REPORT: an essential prerequisite for guarantee
validity is the return of the installation and testing report after all its sections have been
filled in by the User and the Fitter. The report must be returned within 15 days after
installation, otherwise the guarantee becomes null and void.
4) LIMITS: the guarantee entitles its holder to repairs or replacement of faulty components
free of charge. The right to replace the entire machine is excluded. As for components
either applied or added to TECNO-GAZ products or by third parties, having their own
guarantee certificate, the conditions, limits and exclusions indicated in the relevant
certificates apply.
Under no circumstances shall carrying out of one or more than one repair during the
guarantee term change the guarantee expiry date.
5) CLAIMS: in the event of the Purchaser making a claim relating to the application of the
guarantee and the quality or conditions of the delivered equipment, the said Purchaser
shall in no way suspend and/or delay payments.
6) DISPUTES: if a dispute arises on the application and interpretation of this Certificate of
guarantee, the Court of PARMA (Italy) shall be the court of jurisdiction, regardless of the
location in which the purchase agreement was stipulated.
7) DEROGATIONS: any special derogations from the guarantee terms described in this
section shall not imply the granting to the Purchaser of any right and shall be considered
as granted for a specific case only.
8) MISCELLANEOUS: for any other issue, which is not covered by this Certificate of
Guarantee, the rules described in the Civil Code and existing relevant Laws in the
Republic of Italy apply.
41
ENGLISH
The guarantee becomes null and void in the event that:
a) the equipment presents damage caused by a fall, exposure to flames, liquid spillage,
natural events, bad weather conditions or by any other cause that is not due to
manufacturing defects;
b) installation has not been performed in line with TECNO-GAZ instructions and by
authorized personnel;
c) the equipment is repaired, modified or tampered with by the Purchaser or by nonauthorized third parties;
d) when asking for an intervention under guarantee, the product serial number has been
removed, deleted, counterfeited, etc.;
e) the installation and testing report is not returned suitably filled in and signed within 15
days from installation;
f) the Purchaser suspends and/or delays payment of amounts due, for any reason
whatsoever, relating to the purchase and/or maintenance of the equipment;
g) scheduled periodic maintenance or any other instruction envisaged in the manual for
operation and maintenance is not carried out.
42
ENGLISH
A
MAINTENANCE
Correct maintenance of the autoclave assures correct functioning of it and a secure saving
in terms of time and costs for assistance and maintenance. The following operations are
compulsory feasible by operators.
Cleaning of chamber
Every 20 cycles or once a week
Clean periodically the chamber, remove eventual deposits or debris, thus avoiding the
introduction in the discharging circuit of material which can cause obstructions. For
correct cleaning use only demineralized water and the abrasive sponge provided (non
abrasive side - FIG.13).
To execute absolutely with cold chamber to avoid burns - Never use solvents,
detergents, chemical solutions, descaling agents or other similar products.
Cleaning of chambre filter
Every 20 cycles or once a week
Pull up the filter (spare code DXBA091), paying attention not to damage it, wash with
demineralized water and dry with cloth. Then replace in the seat, making sure that
protrudes about 15mm.
Cleaning of trays and tray holder
Every 20 cycles or once a week
Clean with sponge supplied (non abrasive side) and demineralized water.
Change of the bacteriological filter
Every 200 cycles or when it assumes
a dark color
Substitute the bacteriological filter (FIG.2–pos.10) turning it counterclockwise to unscrew
and clockwise to screw. Use only original filter (spare code DAVA101).
Cleaning of the door gasket
Every 20 cycles or once a week
Periodically remove eventual residuals which are deposited on the circumference of the
seal (spare code DANA038) using water and the sponge provided (non abrasive side),
or else a moistened cloth.
A-1
ENGLISH
Door regulation
Every 2 mouths
To regulate the door lock pressure, it is necessary to adjust the regulator of the same
door using the double function wrench (extraction tray + wrench) (FIG.10) provided
(spare code DANA008). Turn anti clockwise (pos.A) to increase lock pressure. Turn
clockwise (pos.B) to decrease the lock pressure
A
Substitution printer paper
B
-
Thermographic paper roll 57 mm wide (spare code DNVA363). Thermographic paper can
be printed on one side only. Open the printer cover by pushing the button place the paper
roll by paying attention to the rotation direction, pull the paper until it comes out and close
the cover; the printer is ready for printing.
A-2
ENGLISH
The following steps are recommended by the manufacturer. These operations must be
performed only by trained technicians authorized by the manufacturer.
Replacement of the water loading filter
every 1 year / every 1000 cycles *
Control of the water filter of flowmeter inlet
every 1 year / every 1000 cycles *
Cleaning Y brass filter downstream of the radiator
every 1 year / every 1000 cycles *
EV4 (water loading solenoid valve) cleaning
every 1 year / every 1000 cycles *
Vacuum pump valves replacement
every 1 year / every 1000 cycles *
Cleaning of the radiator
every 1 year / every 1000 cycles *
Fans cleaning
every 1 year / every 1000 cycles *
Control integrity of the pneumatic circuit
every 1 year / every 1000 cycles *
Control integrity of the electric circuit
every 1 year / every 1000 cycles *
Control thermal insulation of the chamber
every 1 year / every 1000 cycles *
Control tightening screws
every 1 year / every 1000 cycles *
Control temperature probes
every 1 year / every 1000 cycles *
Safety valve check
Follow the guidelines of the
manufacturer (see instructions)
Grease the moving part of the lever
every 1 year / every 1000 cycles *
Level probes check
every 1 year / every 1000 cycles *
state of tank check
every 1 year / every 1000 cycles *
* valid the condition of the reached first
Regional/national regulations fix the mandatory of autoclave validation that checks
starilization parameters. Periodic validation must always be performed by trained
technicians using appropriate measurement instruments. In the case of non-mandatory,
validation is a voluntary operation which aims to certify that the autoclave is fully functional
and fit for sterilization of loads for which it was designed and built.
A-3
ENGLISH
NOTE
A-4
FRANÇAIS
Cher client,
Nous vous remercions d'avoir choisi notre autoclave et saurons mériter votre confiance
avec la plus grande attention et un service en mesure de répondre à votre attente.
L’autoclave que vous avez choisi est la meilleure expression en fait de technologie
appliquée Il peut avoir des applications dans le domaine dentaire, médical, esthétique et
en général dans tous les secteurs où il y a la nécessité d’effectuer une stérilisation en
toute sécurité de tous les types d’instruments et matériaux: solides, creux, charges
poreuses, emballés ou non.
Avant d'utiliser cet autoclave, nous vous invitons à lire attentivement le contenu de ce
mode d'emploi puis de le ranger dans un lieu accessible à tout le personnel préposé à la
stérilisation.
L’autoclave doit exclusivement être utilisé pour l'usage prévu par le fabricant.
Pour l'installation, l'entretien et le SAV adressez-vous exclusivement à un personnel
agréé. Nous vous invitons aussi à utiliser et à demander exclusivement des pièces
de rechange d'origine
1
FRANÇAIS
INDEX
1. CONSULTATION MANUEL
1.1
GLOSSAIRE
1.2
DESCRIPTION VUE AVANT/ARRIERE
2. SÉCURITÉ
2.1
MARQUAGE DE SÉCURITÉ
2.2
DISPOSITIF DE SÉCURITÉ
2.3
NOTES DE SÉCURITÉ
2.4
MISE AU REBUT
3.
UTILISATION ET APPLICATIONS DE L’AUTOCLAVE
4. DONNÉES TECHNIQUES
4.1
MÉCANIQUES
4.2
ÉLECTRIQUES
4.3
CHAMBRE
4.4
RÉSERVOIR EAU PROPRE
4.5
RÉSERVOIR EAU USÉE
4.6
FILTRE BACTÉRIOLOGIQUE
5.
ACCESSOIRES
6.
DÉBALLAGE
7.
INSTALLATION
8. MODE D'EMPLOI
8.1
ALLUMAGE AUTOCLAVE
8.2
REMPLISSAGE RÉSERVOIR EAU PURE
8.3
CARACTERISTIQUES DE L'EAU A UTILISER
8.4
REMPLISSAGE MATÉRIEL DANS AUTOCLAVE
8.5
DÉBUT CYCLE DE STÉRILISATION
8.6
FIN DU CYCLE DE STÉRILISATION
8.7
ENLÈVEMENT DU MATÉRIEL STÉRILISE
8.8
ÉVACUATION EAU USÉE
8.9
INTERRUPTION DU CYCLE DE STÉRILISATION
9. CYCLES DE STÉRILISATION
9.1
DESCRIPTION CYCLES
9.2
CYCLES OPÉRATIONNELS
9.3
CYCLES NUIT
9.4
CYCLES TEST
10. ESSAIS DE CONTROLE AUTOCLAVE
10.1
INTEGRATEURS CHIMIQUES
10.2
INDICATEUR BIOLOGIQUE
10.3
TEST DE BOWIE&DICK
2
FRANÇAIS
10.4
10.5
TEST HELIX
TEST DE VIDE
11. TABLES ET TABLEAUX
12. LECTURE DU RAPPORT DU CYCLE
13. MESSAGES D’ERREUR ET ALARMES
14. PROGRAMMATION DE L’ECRAN
14.1 SÉLECTION LANGUE
14.2 SÉLECTION HEURE
14.3 SÉLECTION MINUTES
14.4 SÉLECTION SECONDES
14.5 SÉLECTION DATE
14.6 SÉLECTION MOIS
14.7 SÉLECTION ANNÉE
14.8 AFFICHAGE PRESSION ATMOSPHÉRIQUE
14.9 SÉLECTION REMPLISSAGE EAU
15. PETIT GUIDE DE DÉPANNAGE
15.1 L’AUTOCLAVE NE SÈCHE PAS CORRECTEMENT
15.2 LA CHAMBRE DE L'AUTOCLAVE DEVIENT BLANCHE
15.3 LA CHAMBRE DE L'AUTOCLAVE PRÉSENTE DES TACHES VERTES
15.4 LE CYCLE DE STÉRILISATION A ÉTÉ INTERROMPU
15.5 L’AUTOCLAVE NE REÇOIT PAS LES COMMANDES
15.6 L'IMPRIMANTE DE L'AUTOCLAVE N'IMPRIME PAS
15.7 TACHES SUR LES INSTRUMENTS
16. DÉMINÉRALISATEUR
16.1 CONFIGURATION DES SYSTEMES DE DEMINERALISATION
16.2 CONNEXION DES SYSTÈMES DE DÉMINÉRALISATION
16.3 CHANGEMENT CARTOUCHE RESINES DU DEMINERALISATEUR
17. PROCÉDURES STANDARD DE STÉRILISATION
18. INSTRUCTIONS POUR LE RETOUR DE L'AUTOCLAVE
19. PROCÉDURES A SUIVRE POUR BÉNÉFICIER DU SAV
20. GARANTIE
A. ENTRETIEN
3
FRANÇAIS
CONSULTATION MANUEL
1
1.1
Glossarie
VIDE
Pression inférieure à la pression atmosphérique
CHAMBRE
Corps cylindrique en acier qui loge les appareils à stériliser
FILTRE
PLATEAU
PORTE-PLATEAUX
Dispositif pour retenir les particules
Tiroir de support de la charge à stériliser
Structure de support des plateaux dans la chambre de stérilisation
TEST HELIX
TEST DE BOWIE DICK
TEST DE VIDE
Dispositif pour éliminer les sels minéraux de l’eau qui vient du réseau
hydrique
Test de vérification de la pénétration de la vapeur dans un corps
creuxde vérification de la pénétration de la vapeur dans un corps
Test
poreux
Test
de vérification du maintien du degré de vide dans la chambre
FILTRE
BACTERIOLOGIQUE
Dispositif pour retenir les particules supérieures à 0,3 µm en
suspension dans l’air.
DEMINERALISATEUR
SURPRESSION
SURTEMPERATURE
RESINES
1.2
Pression supérieure à la pression standard pour le cycle considéré
Température supérieure à la standard pour le cycle considéré
Composant remplaçable (cartouche) du déminéralisateur qui élimine
tous les sels inorganiques
Description vue avant / arriere
Pied d’écartement
Excedent eau pure
Vidange eau pure – connexion deioniseur
Soupape de sécurité
Excedent eau utiliés – decharge condensation
Vidange eau utilisée
Alimentation électrique principale
prise pour alimentation électrique exclusivement
réservée au deioniseur
Fusible 12A
FIG.1
Fusible 12A
Fusible 500mA
4
FRANÇAIS
FIG.2
5
FRANÇAIS
Filtre bactériologique
Remplissage manuel eau
Remplissage eau
Évacuation eau usée
Écran
Imprimeur
Connexion RS232
Le connecteur de l’incubateur B-TEST
Interrupteur général
Cycle 121°C
(valable pour les matériaux ensaché et non ensaché)
Cycle 134°C
(valable pour les matériaux ensaché et non ensaché)
Cycle 134°C Flash
Cycle 134°C Prion
Test de vide
Helix – B&D Test
Niveau minimum eau
Niveau maximum eau
Niveau maximum eau usée
B-START
Bouton START-STOP
B-PUMP
Bouton Remplissage eau
B-SELECT
B-DOOR
6
Bouton Sélection Cycle
Bouton Déverrouillage panneau
FRANÇAIS
2
2.1
SÉCURITÉ
Marquage de sécurité
TENSION DANGEREUSE
COUPER LE COURANT AVANT
D'ENLEVER LE COUVERCLE
TEMPÉRATURE ÉLEVÉE
MISE A LA TERRE
2.2
Dispositifs de sécurité
Les dispositifs de sécurité sont les suivants :
-) Soupape de sécurité étalonnée à 2,4 bar ± 10%
-) Verrouillage électromagnétique pour prévenir l'ouverture du panneau pendant le cycle
-) Thermostat de sécurité
2.3
Notes de sécurité
• Le constructeur est responsable du produit introduit sur le marché aux sens de la
réglementation en vigeur. La responsabilité déchoit dans l'instant où elles sont
exécutées des opérations sur le dispositif, o sur une partie d'elle, de personnel non
qualifié ou avec l'utilise des parties de rechange pas originaux.
• L'autoclave ne doit pas fonctionner dans un environnement comportant un risque de
déflagration et/ou d'incendie.
• Les autoclaves doivent être installés dans une atmosphère conformément à la condition
législative en vigueur sur l'utilisation de pays.
• Le connecteur de l’incubateur (FIG.2-pos.18), doit être employé uniquement pour la
connexion du B-TEST
Comme il y a une connexion à 230 V c.a. indépendante de l’interrupteur
général, ne pas toucher le connecteur les mains mouillées ou humides et le
protéger contre les éclaboussures d’eau.
7
FRANÇAIS
2.4
Mise au rebut
Ce produit est soumis a la directive 2002/96/CEE du Parlement européen
et du Conseil de l'Union européenne relative aux déchets d'équipements
électriques et électroniques (DEEE). Par ailleurs, dans les juridictions
adoptant cette directive, il est marqué comme mis sur le marché après le
13 août 2005. Il ne doit donc pas être jeté avec vos ordures ménagères
courantes (déchets municipaux non triés). Pour vous dé débarrasser de ce
produit, veuillez faire appel à vos services de collecte des DEEE et
observer toutes les obligations en vigueur.
3
UTILISATION ET APPLICATIONS DE L’AUTOCLAVE
L’ autoclave est en mesure pour stériliser les trois typologies de chargement prévues en la
règle EN 13060, spécifiéeme :
MATÉRIELS FERREUX Instruments sans cavités et sans entrave à la pénétration de la
OU SOLIDES
vapeur
PARTIES POREUX
Matériels simples ou composés qui peuvent absorber les fluides
(tissés, blouses, gazes, pansements etc...)
PARTIES AVEC
CAVITE’
Matériels ou dispositifs avec cavité, obstructions etc...
Ceux-ci se subdivisent en deux typologies, classifiées selon
longueur et diamètre:
TYPE B : canules, tubes ou dispositifs avec des passages larges
TYPE A : turbines, tubes, manipules et dispositifs avec des trous
des petites dimensions
Ce dispositif a été certifié pour la stérilisation des suivants instruments*:
EUROPA B EVO
EUROPA B EVO 24
Matériels ferreux ou solides ensaches
max kg. 5
max kg. 7
Matériels cavità de type A et B
max kg. 5
max kg. 7
max kg. 1,5
max kg. 2
Matériels poreax
* Valable uniquement pour les pays européens
Il faut employer l’autoclave exclusivement pour la stérilisation
d’instruments et matériaux compatibles avec le système de
stérilisation à la vapeur. S’assurer que les charges à stériliser peuvent
supporter les températures du cycle sélectionné.
8
FRANÇAIS
4
4.1
DONNÉES TECHNIQUES
Mécaniques
EUROPA B EVO
Température d'exercice
Altitude max.
Humidité relative MAX. à 30°C
Humidité relative MAX. à 40°C
Dimensions hors tout (L x H x P)
Encombrement panneau ouvert
Poids (réservoirs vides)
Poids (réservoirs pleins)
Poids par aire de support
Emission sonore
4.2
+5°C ÷ +40°C
2.000 m
80%
50%
510 X 390 X 590 mm
510 X 390 X 730 mm
300 mm
54 kg
58 kg
63 kg
67 kg
2
2058 N/m
2058 N/m2
< 70 db A
Électriques
Tension alimentation
Puissance
Fréquence
Cordon alimentation
Fusibles
Chaleur transmise
4.3
EUROPA B EVO 24
EUROPA B EVO
EUROPA B EVO 24
230 V a.c. +/-10 % single phase
1.7 kW
2.2 kW
50 / 60 Hz
2 + 1 x 1mm2
6. 3 x 32 – 12A (x2) / 6.3 x 32 – 500mA (x1)
3.6 E6 J / ora
Chambre
Pression d'exercice MAX.
Vide MAX.
Température MAX.
Composition
Dimensions
EUROPA B EVO
EUROPA B EVO 24
2.4 bar (relatifs)
- 0.9 bar (relatifs)
138 °C
Inox AISI 304
Ø 245 x 318 mm
Ø 245 x 500 mm
9
FRANÇAIS
4.4
Réservoir eau propre
EUROPA B EVO
EUROPA B EVO 24
Avec
Sans
demineralizz.
4,5 l
Volume
Nombre de cycles réalisables
Vour les instructions
demineralizers
4
Réservoir eau usée
EUROPA B EVO
Volume
Nombre de cycles réalisables
Composition
Température max. eau usée
4.6
EUROPA B EVO 24
4,5 l
4
2
Polyéthylène
50°C
Filtre bactériologique
EUROPA B EVO
Diamètre
Capacité filtrante
10
2
Polyéthylène
Composition
4.5
demineralizz.
EUROPA B EVO 24
56 mm
0.3 µm
FRANÇAIS
05
ACCESSOIRES
Porte-plateaux
EUROPA B EVO
EUROPA B EVO 24
Aluminium Anodisé
192 x 165 x 280 mm
192 x 200 x 470 mm
Composition
Dimensions (L x H x P)
Image
FIG.4
FIG.3
Dotation prévue
1
Porte-Plateaux (facultatif)
EUROPA B EVO
Aluminium Anodisé
193 x 200 x 280 mm
(200 x 193 x 280) mm
Composition
Dimensions (L x H x P)
EUROPA B EVO 24
/
/
Image
/
FIG.5
Dotation prévue
1
/
Plateaux
EUROPA B EVO
Composition
Dimensions (L x H x P)
184 x 17 x 286
mm
EUROPA B EVO 24
Aluminium Anodisé
184 x 17 x 286 mm + 184 x 17 x 140 mm
Image
+
FIG.6
Dotation prévue
4
FIG.7
FIG.8
4+4
11
FRANÇAIS
Plateaux (facultatif)
EUROPA B EVO
Composition
EUROPA B EVO 24
Aluminium Anodisé  Acier
Inoxydable
184 x 17 x 460 mm
/
Dimensions (L x H x P)
/
Image
/
Dotation prévue
/
FIG.9
4
Pince d’extraction plateau et réglage porte
A employer pour extraire et manutentionner les plateaux (FIG.11) et pour pouvoir régler la
porte (paragrafo 14.4)
Image
FIG.11
FIG.10
Dotation prévue
1
Entonnoir en plastique avec tuyau
A employer pour charge manuelle de l’eau (paragrafo 8.2)
Image
FIG.12
Dotation prévue
1
Eponge pour nettoyage chambre et garniture porte
A employer pour nettoyer la chambre de stérilisation et la garniture de la porte (par.14.214.3)
Image
FIG.13
Dotation prévue
12
1
FRANÇAIS
Raccord charge eau
A employer avec l’entonnoir en plastique avec tuyau pour charge manuelle de l’eau et
avec tuyau charge eau avec filtre
Image
FIG.14
Dotation prévue
1
Entonnoir en plastique avec tuyau
A employer pour charge manuelle de l’eau (paragrafo 8.2)
Image
FIG.15
Dotation prévue
1
Tuyau vidange eau
A employer pour vider l’eau usée (paragrafo 8.9)
Image
FIG.16
Dotation prévue
1
Tubo rilsan
Connecter une extrémité du tuyau au Trop-plein eau usée – purge vapeur (FIG.1 – pos.4),
l’autre à un bac pour la récupération de l’eau et de la vapeur.
Image
FIG.17
Dotation prévue
1
13
FRANÇAIS
Pied d’écartement arrière en plastique
Mettre le pied d’écartement dans la partie arrière de l’autoclave (FIG.1–pos.0) pour
assurer un aérage adéquat en cas de positionnement de l’autoclave près d’une paroi.
Image
FIG.18
Dotation prévue
1
Tuyaux pour vidange arrière des eaux
1- Tuyau pour trop-plein eau propre: Connecter une extrémité du tuyau au trop-plein
arrière eau propre (FIG.1–pos.1), l’autre à un bac pour la récupération de l’eau
déminéralisée ou en cours de vidange (version avec déminéralisateur).
2- Tuyau pour robinet vidange eau usée: Connecter une extrémité du tuyau au raccord
FIG.20 et le visser au robinet de vidange de l’eau usée (FIG.1–pos.5), l’autre à un bac de
récupération de l’eau usée .
Image
FIG.19
Dotation prévue
2
Raccord pour vidange arrière eau usée
Connecter une extrémité du tuyau pour vidange eau usée au raccord FIG.20 et le visser
au robinet de vidange de l’eau usée (FIG.1-pos.5).
Image
FIG.20
Dotation prévue
14
1
FRANÇAIS
Cordon d'alimentation
Connecter l’extrémité du connecteur (FIG.21–pos.B) au panneau arrière (FIG.1–pos.6)
ensuite brancher la fiche (FIG.21–pos.A) directement à la prise de secteur.
Image
FIG.21
Dotation prévue
06
1
DÉBALLAGE
L’autoclave est expédié dans un emballage protecteur idoine lui permettant d'être
manutentionné facilement. L’emballage ne doit pas subir de heurts, il doit être manipulé
avec attention en évitant de le faire rouler ou tomber d'une hauteur supérieure à 16 cm.
En cas d'absence d'engins de levage, l'autoclave emballé doit toujours être manutentionné
par deux personnes. L’autoclave est soutenue avec une palette en bois et elle est
contenue dans un carton ondulé et renforcé intérieurement par des composés de carton.
Pour déballer l’autoclave, il faut ouvrir le carton ondulé, lever les renforcements et la
déballer en utilisant les courroies en dotation.
La manutention doit se faire uniquement à l’aide des courroies et par deux
personnes au moins.
Ne soulevez jamais l’autoclave en la saisissant par la partie inférieure du
panneau ou du tableau de commandes au risque de provoquer des
cassures mécaniques.
ATTENTION: L'emballage d'origine doit être conservé.
A l’intérieur de l'emballage, vous trouverez:
• MANUEL DE MODE D'EMPLOI: à lire attentivement et à ranger dans un endroit
accessible à tout le personnel préposé à la stérilisation.
• CERTIFICATION ET LA DÉCLARATION DE CONFORMITÉ: être maintenu.
• BON DE GARANTIE: que vous devez remplir et envoyer.
15
FRANÇAIS
07
INSTALLATION
FIG.22
EUROPA B EVO
A
B
C
D
E
F
EUROPA B EVO 24
610 mm
590 mm
730 mm
50 mm
410 mm
426 mm (avant)
403 mm (avant)
455 mm (arriere)
444 mm (arriere)
400 mm
431 mm
1 - Installez l’autoclave dans un environnement idoine à la stérilisation.
2 - La pièce doit être bien éclairée et ventilée, comme prévu par la législation en vigueur.
3 - Installez l’autoclave loin des sources de chaleur et des projections d'eau.
4 - Placez l’autoclave sur un plan idoine à en supporter le poids (min.80 kg) et de
dimensions adéquates.
5 - Le plan d'appui de l’autoclave doit être parfaitement horizontal.
6 - Placez l’autoclave à une hauteur telle que l'utilisateur puisse contrôler toute la chambre
de stérilisation et la nettoyer aisément.
7 - Ouvrez le panneau de l’autoclave et enlevez tous les sachets contenant les
accessoires et qui ont été rangés dans la chambre de stérilisation.
16
FRANÇAIS
8 - Ne laissez dans la chambre de stérilisation que le porte-plateaux et les plateaux.
Rangez tous les autres accessoires dans un logement externe à la disposition du
personnel.
9 - Ne posez jamais sur l’autoclave des journaux, des plateaux, des conteneurs de
liquide, etc
10 - Ne jamais s'appuyer au panneau lorsqu'il est ouvert.
11 – Laisser un espace de 5cm minimum dans la partie arrière par le biais du pied
d’écartement arrière en plastique (FIG.1–pos.0 / FIG.18) et dans les côtés de l’appareil
pour assurer l’aérage nécessaire.
12 - Effectuer les connexions des tuyaux dans la partie arrière (chapitre 5)
13 - Avant de brancher l'autoclave, vérifiez si l'installation électrique est conforme à la
législatin en vigueur en la matière et si elle est adaptée aux caractéristiques de l'appareil.
14 - Branchez la prise femelle du cordon d'alimentation en dotation sur la fiche du tableau
situé au dos de l’autoclave (FIG.1–pos.6)
15 - Branchez la prise électrique à l'installation en vérifiant que l'alimentation corresponde
bien à 230 Vac- 2000Va.
IMPORTANT:
Evitez l'utilisation de rallonges, réductions ou adaptateurs au risque de produire
des micro-interruptions de courant et par conséquent des déclenchements
d'alarme.
16 - Mettez l’autoclave en route en appuyant sur l'interrupteur général (FIG.2–pos.19) et
ouvrez le panneau de l'autoclave. Attendez quelques secondes les deux signalisations
sonores indiquant l'acquisition des paramètres relatifs à l'alignement atmosphérique
automatique. En même temps, le message “PORTE OUVERTE” s'affichera sur le visuel.
IMPORTANT:
Ne sélectionnez jamais une commande avant d'avoir entendu les deux
signalisations sonores car l'autoclave ne l'accepterait pas.
17 - Remplissez le réservoir d'eau pure en suivant les indications du chapitre 8.
18 - Si l’autoclave est installé avec le dispositif de Déminéralisation, suivez les instructions
reportées sur l'emballage de ce dispositif et chapitre 16.
17
FRANÇAIS
08
MODE D'EMPLOI
Après avoir installé l’autoclave, procédez à la préparation et à l'utilisation.
8.1
Allumage autoclave
En appuyant sur l'interrupteur général (FIG.2–pos.19)
8.2
Remplissage réservoir eau pure
Connecter le “raccord charge eau” FIG.14 au “tuyau charge eau avec filtre” (standard), et
l’insérer dans l’orifice du raccord frontal de l’autoclave (FIG.2–pos.13).
Introduire l’autre extrémité du tuyau avec filtre dans le bac de l’eau déminéralisée ou
distillée.
A ce moment-là appuyer sur le bouton B-PUMP pour actionner la pompe de charge eau et
le maintenir abaissé jusqu’à l’affichage du compte à rebours .
La pompe remplit le réservoir interne de l'autoclave. Si le niveau maximum n'est pas
atteint dans les 180 secondes, la pompe s'arrête automatiquement et il faudra à nouveau
appuyer sur le bouton B-PUMP.
La pompe s'arrête automatiquement lorsque le niveau maximum a été atteint.
Si la pompe de remplissage de l'eau ne devait pas fonctionner, le réservoir pourrait être
rempli manuellement par l'opérateur en procédant de la façon suivante:
• Allumer l’autoclave..
• Enlever le bouchon placé frontalement sur l’autoclave (FIG.2–pos.12)
• Introduire le raccord charge eau dans cet orifice avec le tuyau connecté à l’entonnoir
(FIG.14)
• Verser l’eau distillée dans l’entonnoir en positionnant ce dernier plus en haut par rapport
au raccord de charge.
• Verser jusqu’à l’éclairage du led de niveau maximum (FIG.2–pos.2).
• La charge terminée, extraire le raccord et repositionner le bouchon.
Si le niveau maximum d'eau n'est pas atteint avec un déionisateur branché, le
fonctionnement de l'autoclave sera invalidé.
18
FRANÇAIS
8.3
Caractéristiques de l'eau a utiliser
TABLEAU NIVEAUX QUALITATIFS ÉTABLIS PAR LA NORME UNI EN 13060 : 2009
CEN STANDARD UNI EN 13060 : 2009
Résidu d'évaporation
10
mg/l
≤
Oxyde de silicium
1
mg/l
≤
Fer
0.2
mg/l
≤
Cadmium
0.005
mg/l
≤
Plomb
0.05
mg/l
≤
Résidus de métaux lourds sauf fer, cadmium, plomb
0.1
mg/l
≤
Chlorure (Cl')
2
mg/l
≤
Phosphate (P20s)
0.5
mg/l
≤
Conductivité (à 20°C)
15
≤
µs/cm
Valeur pH (degré d'acidité)
5 ÷ 7,5
Aspect
incolore, limpide, sans dépôts
Dureté (et ions de terre alcaline)
0.02
mmol/l
≤
8.4
Remplissage matériel dans l'autoclave
Disposez le matériel à stériliser sur les plateaux en dotation.
• ne superposez jamais le matériel
• les instruments ensachés doivent toujours être disposés avec la partie papier vers le
haut
• le matériel ne doit jamais entrer en contact avec la chambre de stérilisation ou le
panneau de fermeture
• ranger pinces et ciseaux les lames ouvertes
FIG.23
Après le chargement, fermez le panneau de l'autoclave. Le message “PORTE FERMÉE”
s'affichera sur le visuel
19
FRANÇAIS
8.5
Début cycle de stérilisation
Sélectionnez le programme de stérilisation plus idoine au matériel à stériliser en appuyant
sur le bouton B-SELECT. Chaque pression commute la sélection du programme au
suivant. Après avoir sélectionné le programme, démarrez le cycle en appuyant sur le
bouton B-START, la porte se verrouillera automatiquement et le cycle aura lieu.
Pendant le cycle, le visuel affiche tous les paramètres et les informations correspondantes
8.6
Fin du cycle de stérilisation
Un avertissement sonore préviendra les opérateurs de la fin du cycle de stérilisation et le
message “FIN CYCLE” sera affiché.
A la fin du signal acoustique, déverrouillez la porte en appuyant sur le bouton B-DOOR. Si
la chambre n'est pas complètement dépressurisée, l'actionnement du bouton n'aura aucun
effet. Attendez la dépressurisation complète de la chambre et répétez l'opération. En cas
d'urgence appuyez simultanément sur B-DOOR + B-PUMP
Une fois le panneau déverrouillé, tirez la poignée de la porte et ouvrez.
8.7
Enlèvement du materie stérilise
Mettre les équipements de protection individuelle conformément aux normes en vigueur
en matière d’hygiène et sécurité du travail. Extraire les plateaux en employant la pince
livrée (FIG.10), laisser conditionner les instruments et les stocker dans un milieu stérile.
8.8
Évacuation eau usée
Lorsque le témoin de niveau eau usée (FIG.2–pos.28) s'allume, il faut la vider. Si on ne
vide pas l'eau usée, le fonctionnement de l'autoclave est invalidé.
Se non si provvede, il funzionamento dell’autoclave è inibito.
Prenez le tuyau avec le raccord en dotation (FIG.16), et introduisez le dans la prise de
vidage de l'eau usée située sur la façade de l’autoclave (FIG.2–pos.14), Dévisser sur la
bague dans le sense inverse aux aiguilles d'une montre (compiere 2 giri completi), placez
l'autre extrémité du tuyau dans un conteneur. L'eau usée tombera par gravité dans ledit
conteneur.
IMPORTANT:
A - le tuyau logé dans le conteneur de réception ne doit jamais effleurer ou être plongé
dans l'eau usée car cela créerait un vortex.
B - attendez toujours que l'eau usée soit complètement vidangée, étant donné que le
témoin de niveau maximum de l'eau usée s'éteint lorsqu'il y a encore de l'eau dans le
réservoir, ne l'utilisez pas comme indicateur dans cette opération.
A la fin du vidage visser le métal sonnent et enlever le tube.
20
FRANÇAIS
8.9
Interruption du cycle de stérilisation
Un cycle de stérilisation peut être volontairement interrompu en appuyant sur le bouton
B-START pendant 2 secondes au moins.
L’autoclave produira un son, ira en décompression et le message d'erreur A001 s'affichera
sur le visuel (CYCLE INTERROMPU).
Pour restaurer l'erreur, appuyez simultanément sur les boutons illustrés ci-dessous
jusqu'au déclenchement du signal acoustique :
B-SELECT
B-PUMP
B-DOOR
ATTENTION: Si l’imprimante est en fonction la remise à zéro de l’alarme est inhibée
21
FRANÇAIS
09
9.1
CYCLES DE STÉRILISATION
Description cycles
L’autoclave est doté de trois séries de cycles :
A - cycles opérationnels
B - cycles opérationnels-cycles nuit
C - cycles tests
9.2
Cycles opérationnels
Tous les cycles opérationnels ont le système de vide fractionné, ils peuvent donc stériliser
des matériels creux, poreux, solides, aussi bien en vrac qu'ensachés. Les températures
peuvent être sélectionnées de 121°C à 134°C.
Normalement les cycles à 121°C s'utilisent pour thermopla stiques ou matériel sensible
tandis que les cycles à 134°C s'utilisent pour tous les aut res matériels.
Dans tous les cas suivez toujours les indications des fabricants des instruments ou des
dispositifs à stériliser.
Il existe aussi 2 cycles spéciaux :
- cycle Prion, étudié pour le variant de la maladie de Creutzfeld-Jacob (vache folle)
- cycle Flash conçu pour stériliser rapidement des instruments ou des dispositifs en vrac.
9.3
Cycles nuit
L’autoclave est équipé d'un dispositif économiseur d'énergie. Ce dispositif permet
d'effectuer des cycles de stérilisation en l'absence du personnel.
A la fin du cycle, si la porte n'est pas ouverte, l’autoclave se stabilise, puis s'éteint
automatiquement, seul l'interrupteur général restera allumé (FIG.2-pos.19)
Au retour du personnel, il suffira d'appuyer sur n'importe quel bouton pour rallumer
l'autoclave et lire le résultat du cycle sur le visuel. En plus, l'imprimante aura imprimé le
rapport concernant ledit cycle.
9.4
Cycles test
Les cycles test disponibles sont:
- Test Hélix
- Test de Bowie&Dick
- Test de vide
22
FRANÇAIS
10
10.1
ESSAIS DE CONTROLE AUTOCLAVE
Intégrateurs chimiques
Ces sont des essais qui exploitent les propriétés des colorants étant capables de modifier
leur aspect si exposés d’une manière adéquate à la chaleur et à la pression pour un
certain délai, puisqu’ils interagissent avec tous les paramètres du cycle (pression,
température, temps). La réussite de l’essai détermine la correcte stabilité de l’appareil.
UTILISER
Il faut positionner les intégrateurs chimiques (code 200/S e 215-S) à
l’intérieur de la chambre de stérilisation avant de démarrer le cycle,
même à la présence de charge.
Les intégrateurs peuvent être non emballés et positionnés sur les
plateaux, ou ils peuvent être emballés dans les sachets de
stérilisation avec les matériaux à stériliser.
Il est pourtant conseillable d’emballer toujours les intégrateurs dans
les sachets dans les structures où il y a plusieurs opérateurs; de cette
façon la stérilisation de chaque instrument sera certifiée.
Si le virage reste dans la zone
blanche, l’essai n’a pas réussi
RÉPONSE
Si le virage entre dans la fenêtre
SAFE, l’essai a réussi
FRÉQUENCE
10.2
Il n’y a pas d’intervalles préétablis pour les contrôles périodiques,
sauf dans les zones où il y a des lois régionales spécifiques. En tout
cas les intégrateurs chimiques sont les dispositifs d’essais les plus
économiques et immédiats, par conséquent pour un contrôle constant
de l’efficacité de l’autoclave on conseille de les utiliser sur tous les
cycles, ou une fois par jour minimum.
Indicateur biologique
Cet essai (code 262-S) sert à prouver la capacité de l’autoclave quant à la destruction de
tous les microorganismes. Il se compose de spores standard, ayant des caractéristiques
biologiques, une haute résistance à la chaleur et pouvant être considérées comme un
moyen de contrôle absolument sûr et fiable (ATCC 7953). Le dispositif d’essai se présente
sous forme d’ampoules et il n’est ni pathogène, ni toxique, ni pyrogénique.
23
FRANÇAIS
UTILISER
Il faut positionner les ampoules dans la chambre de stérilisation,
même à la présence de charge.
Effectuer le cycle, à la fin enlever l’ampoule en faisant attention
puisqu’elle est chaude et sous pression.
La faire refroidir pour environ 10 minutes, ensuite l’activer, à l’aide du
concasseur à mâchoires en la gardant toujours en position verticale.
L’indicateur chimique placé sur l’étiquette de l’ampoule aura viré de la
couleur bleue à la noire.
Ensuite introduire l’ampoule dans un incubateur biologique à 57°C
pour 48 heures, avec une ampoule d’essais non traitée, mais de
toute façon activée avec le concasseur à mâchoires.
RÉPONSE
Après 48 heures enlever l’ampoule traitée de l’incubateur et évaluer
le résultat. Si l’ampoule a viré au jaune cela signifie que l’autoclave
n’a pas réussi l’essai et qu’il y a croissance bactérienne. Si l’ampoule
reste de couleur violet cela signifie qu’il n’y a pas de croissance de
microorganismes et donc l’autoclave a réussi l’essai. Evidemment
l’ampoule d’essai virera toujours au jaune puisqu’elle n’est pas traitée
et servira uniquement comme terme de comparaison.
A la fin de l’essai découler les ampoules dans les déchets solides
urbains. On conseille de toute façon d’effectuer un ultérieur cycle de
stérilisation à 121°C sur l’ampoule
FRÉQUENCE
10.3
Il n’y a pas d’intervalles préétablis pour les contrôles périodiques,
sauf dans les zones où il y a de lois régionales spécifiques. En tout
cas les dispositifs d’essais biologiques sont les dispositifs d’essais les
plus concrets, par conséquent pour un contrôle constant de l’efficacité
de l’autoclave on conseille d’effectuer l’essai au moins une fois
tous les 90 jours.
Test de Bowie&Dick
Il s’agit d’un essai physique (code 268/S) qui permet de vérifier la capacité de pénétration
de la vapeur dans les charges poreuses. Pour cet essai on utilise un « paquet-test »
standard et conforme aux réglementations techniques actuellement en vigueur.
UTILISER
24
Il faut effectuer l’essai la chambre vide.
Il faut positionner le Bowie&Dick sur le plateau central de l’autoclave.
Effectuer le cycle prévu à cet effet qui est indiqué dans l’afficheur de
l’autoclave, à la fin enlever le paquet-test, l’ouvrir et contrôler le
virage de la feuille avec indicateur chimique placé à l’intérieur.
FRANÇAIS
RÉPONSE
FRÉQUENCE
10.4
L’évaluation est simple et rapide. Si
le virage résulte uniforme (comme
dans la figure)
l’essai a réussi
parfaitement, autrement l’essai n’a
pas
réussi,
par
conséquent
l’autoclave n’est pas à même de
stériliser correctement les charges
poreuses.
Il n’y a pas d’intervalles préétablis pour les contrôles périodiques,
sauf dans les zones où il y des lois régionales spécifiques. En tous
cas, pour avoir un contrôle constant de l’efficacité de l’autoclave, on
conseille d’effectuer l’essai au moins tous les 30 jours.
Test Hélix
Il s’agit d’un essai physique permettant de vérifier la capacité de pénétration de la vapeur
dans les corps creux. Pour l’essai on utilise un système standard et conforme aux
réglementations techniques en vigueur.
Il faut introduire le dispositif d’essai la chambre vide.
UTILISER
Dans la capsule placée à l’extrémité du dispositif d’essai introduire
l’indicateur, ensuite positionner le dispositif d’essai sur le plateau
central de l’autoclave.
Effectuer le cycle prévu à cet effet qui est indiqué dans l’afficheur de
l’autoclave, à la fin enlever le dispositif d’essai, ouvrir la capsule et
contrôler le virage de l’indicateur.
25
FRANÇAIS
AVANT L’ESSAI
RÉPONSE
ESSAI NON REUSSI
ESSAI REUSSI
FRÉQUENCE
10.5
Il n’y a pas d’intervalles préétablis pour les contrôles périodiques,
sauf dans les zones où il y des lois régionales spécifiques. En tous
cas, pour avoir un contrôle constant de l’efficacité de l’autoclave, on
conseille d’effectuer l’essai au moins 1 fois tous 30 jours.
Test de vide
Il s’agit d’un essai d’étanchéité de la chambre ou d’un essai pour établir la perte de vide.
Ce contrôle a le but de vérifier que pendant le cycle il n’y ait pas d’infiltrations d’air à
travers les étanchéités de la chambre (garnitures, vannes, etc…)
Il faut effectuer le cycle la chambre vide.
UTILISER
On sélectionne le cycle prévu à cet effet qui est indiqué dans
l’afficheur de l’autoclave et le cycle s’active.
L’autoclave effectue automatiquement le cycle selon de précises
procédures techniques. Ensuite le résultat final est imprimé par
l’imprimante grâce à une éventuelle connexion informatique.
RÉPONSE
FRÉQUENCE
Le résultat est immédiat et c’est l’imprimante qui l’émet avec toutes
les valeurs du cycle effectué et l’évaluation finale aussi.
Il n’y a pas d’intervalles préétablis pour les contrôles périodiques,
sauf dans les zones où il y des lois régionales spécifiques. En tous
cas, pour avoir un contrôle constant de l’efficacité de l’autoclave, on
conseille d’effectuer l’essai au moins une fois par semaine.
ON CONSEIL D’UTILISER EXCLUSIVEMENT LES
TESTS PROPOSES PAR TECNO-GAZ S.p.A.
26
FRANÇAIS
11
TABLES ET TABLEAUX
18L
24L
18
17
13
1.04 ÷ 1.30
121 ÷ 125
5
17
13
2.02 ÷ 2.38
134 ÷ 138
4
2
4
2.02 ÷ 2.38
134 ÷ 138
20
17
13
2.02 ÷ 2.38
134 ÷ 138
2.02 ÷ 2.38
134 ÷ 138
T4 (Minutes)
121°C
(ensaché e non
ensaché)
134°C
(ensaché e non
ensaché)
134°C
Flash
134°C
Prion
134°C
Helix / Bowie&Dick
3.5
4
PLAGE DE
TEMPERATURE
DE TRAVAIL
( °C )
-0.9
Vacuum
TEST
CYCLES
CYCLES
DURÉE
SÉCHAGE
T5 (Minutes)
PLAGE DE
PRESSION
DE TRAVAIL
( bars relatifs )
DURÉE
EXPOSITION
DUTY CYCLE
TABLEAU STÉRILISATION
Les temps de préchauffage et de fractionnement du vide peurent changer, par rapport aux
conditiones de l’autoclave, des 25 - 30 min. Ces temps vont se sommer aux temps des
cycles indiqués dans la table. Tout les cycles de stérilisation ont 3 phases de vide.
SCHÉMA DE CYCLE
T1,T2,T3
les étapes de
préchauffage et le
fractionnement du vide
T4
Stérilisation
T5
Séchage
FIG.24
27
FRANÇAIS
12
LECTURE DU RAPPORT DU CYCLE
Version du logiciel
Numéro série
Date et Description du cycle
sélectionné (avec les principales
valeurs de configuration)
Heures – Minutes – Secondes
Pression de la chambre
Température principale de la
chambre
Phase de cycle
Fin du cycle (OK signifie que le cycle
s’est terminé correctement)
Numéro de cycle
Opérateur (espace pour la signature)
28
FRANÇAIS
13
MESSAGES D'ERREUR ET D'ALARME
Les messages d'erreur sont signalés par un code alphanumérique composé d'une lettre et
de 3 chiffres.
Le ”préfixe E” concerne les erreurs de l'opérateur et peut être rétablie par l'opérateur luimême.
La “lettre A” concerne des alarmes relatives à l'autoclave. Si le problème persiste après
avoir effectué la solution décrite, contactez le SAV par téléphone du fabricant.
En cas d’affichage d’un message d’alarme (suffixe "A") le cycle doit
être considéré NON réussi: il faudra répéter toutes les opérations de
préparation et de stérilisation.
Pour effacer les alarmes et les erreurs, appuyez simultanément sur les boutons indiqués
ci-dessous jusqu'au déclenchement du signal acoustique :
B-DOOR + B-PUMP + B-SELECT
ERREUR
CAUSE
SOLUTION
E 200
Niveau eau insuffisant
Remplir le reservoir
E 250
Niveau d'eau pure bas, avec remplissage
par déminéralisateur
Remplir le reservoir
E 300
Niveau maximum eau usée
Vider le réservoir
E 401
Porte ouverte
Fermer la porte avec attention
E 451
Mauvais blocage de la porte
Ouvrir et refermer la porte
E 461
Température chambre trop élevée pour le
Vacuum test (plus de 40°C)
Éteindre et attendre 10 minutes
A 250
Début cycle sans eau
Contrôler le déminéralisateur
A 001
Cycle interrompu
Effacer et redémarrer le cycle
A 011
Anomalie pc visuel
Éteindre et rallumer l’autoclave
A 101
Vide pas atteint en 10 minutes
Effacer et répéter le cycle
A 111
Vide pas maintenu sur première phase VT Effacer et répéter le cycle
A 121
Vide pas maintenu sur deuxième phase VT Effacer et répéter le cycle
29
FRANÇAIS
A 401
Porte ouverte pendant le cycle ou problème Contrôler le système de
de fermeture
fermeture
A 501
Pression non maintenue
Effacer et répéter le cycle
A 551
Anomalie 1° capteur température
Effacer et répéter le cycle
A 552
Anomalie 2° capteur température
Effacer et répéter le cycle
A 641
Surpression
Effacer et répéter le cycle
A 651
Surchauffe
Effacer et répéter le cycle
A 661
Anomalies comparaison températures
Effacer et répéter le cycle
A 701
Pression d'exercice pas atteinte
Effacer et répéter le cycle
A 751
Température trop basse
Effacer et répéter le cycle
A 761
Anomalie sur le capteur de pression
Effacer et répéter le cycle
A 801
Anomalie dépressurisation
Effacer et répéter le cycle
A 901
Absence temporaire d’alimentation
électrique. Tension principale trop basse
Effacer et répéter le cycle
Contacter la compagnie de
distribution de l'électricité
REMPLACER LES RESINE Résines
IONIC SYSTEM épuisé
Suivre les indications du paragraphe 16.3
de ce manuel
REMPLACER LE FILTRE Résines
OSMOSI SYSTEM épuisé
Suivre les indications du paragraphe 16.3
de ce manuel
D
La porte de l'imprimeur
n'est pas fermée
Essayez de fermer la porte de l'imprimeur
P
Le papier est absent
Insérez le papier
Lo
Il y a bas voltege sur la
ligne
Pour essayer de relier la machine à l'ouvrier électrique net
dans un autre point; si le problème persiste pour entrer en
contact avec un électricien
30
FRANÇAIS
14
PROGRAMMATION DE L’ECRAN
Faire référence à la FIG.2.
14.1
Sélection langue
Appuyez sur B-START et simultanément sur l'interrupteur (MAINS).
Appuyez B-STARTjusqu'à ce que la langue souhaitée s'affiche.
La langue est modifiée [en avant] en appuyant sur B-SELECT.
La langue est modifiée [en arrière] en appuyant sur B-PUMP
14.2
Sélection heure
L'heure s'affiche en appuyant sur B-START.
L'heure est modifiée [en avant] en appuyant sur B-SELECT.
L'heure est modifiée [en arrière] en appuyant sur B-PUMP
14.3
Sélection minutes
Les minutes s'affichent en appuyant sur B-DOOR.
Les minutes sont modifiées [en avant] en appuyant sur B-SELECT.
Les minutes sont modifiées [en arrière] en appuyant sur B-PUMP.
14.4
Sélection secondes
Les secondes s'affichent en appuyant sur B-DOOR. Ne pas modifier.
14.5
Sélection date
Le jour s'affiche (clignotant) en appuyant sur B-START.
Le jour est modifié [en avant] en appuyant sur B-SELECT
Le jour est modifié [en arrière] en appuyant sur B-PUMP.
14.6
Sélection mois
Le mois s'affiche (clignotant) en appuyant sur B-DOOR .
Le mois est modifié [en avant] en appuyant sur B-SELECT.
Le mois est modifié [en arrière] en appuyant sur B-PUMP.
31
FRANÇAIS
14.7
Sélection année
L'année s'affiche en appuyant sur B-DOOR.
L'année est modifiée [en avant] en appuyant sur B-SELECT.
L'année est modifiée [en arrière] en appuyant sur B-PUMP
14.8
Affichage pression atmosphérique
La pression atmosphérique enregistrée s'affiche en appuyant sur B-START.
NE PAS MODIFIER !
14.9
Sélection remplissage eau
Appuyer sur le poussoir B-START pour l’établissement chargement d’eau :
• CHARGEMENT PAR POMPE
• IONIC SYSTEM
• OSMOSI SYSTEM
Appuyer sur le poussoir B-SELECT pour modifier l’établissement.
Appuyer sur le poussoir B-DOOR jusqu’à l’intervention du “BEEP”.
Appuyez sans lâcher sur B-START jusqu'à ce que vous entendiez le bip.
Les paramètres sont à présent enregistrés. Éteignez l’autoclave avec l'interrupteur
MAINS.
32
FRANÇAIS
15
PETIT GUIDE DE DÉPANNAGE
Dans la plupart des cas, certaines alarmes ou erreurs sont provoquées par une inattention
ou une ignorance de certains aspects techniques et opérationnels.
Ce paragraphe décrit les problèmes plus fréquents pouvant surgir et les solutions pour y
remédier
15.1
L’autoclave ne sèche pas correctement
• remplacez le filtre bactériologique par un neuf d'origine
• les plateaux utilises ne sont pas d'origine, sans trous ou avec une perforation différente.
Utilisez seulement des plateaux d'origine, en aluminium si possible. Demandez des
plateaux d'origine.
• le matériel à stériliser n'est pas disposé correctement. Se conformer strictement aux
instructions au paragraphe 8.4
15.2
La chambre de l'autoclave est devenue blanche
• changez immédiatement le type d'eau utilisée, utilisez de l'eau déminéralisée ou
distillée comme indiqué spécifiquement dans les chapitres précédents puis nettoyez
ensuite la chambre.
• la couleur blanchâtre peut être une conséquence de l'évaporation de matériaux
organiques présents sur le matériel à stériliser. Recommencez le nettoyage du matériel
à stériliser d'une façon plus idoine et approfondie.
• vérifier le système de déminéralisation installé
15.3
La chambre de l'autoclave prèsente des taches vertes
• le matériel à stériliser n'a pas été rince correctement après le lavage. Refaites le
rinçage du matériel à stériliser d'une façon plus idoine et approfondie. Si les tâches sont
évidentes, contactez le sav du fabricant par téléphone.
15.4
Le cycle de stérilisation ètè interrompu
• si l’autoclave est branché avec des rallonges, réductions ou adaptateurs, enlevez ces
accessoires et branchez directement l’autoclave à la prise électrique.
15.5
L’autoclave ne reçoit pas les commandes
• l’autoclave est en train d'effectuer l'alignement barique automatique, attendez le double
signal sonore après l'ouverture du panneau programmez les fonctions.
• le réservoir d'eau pure est vide, le témoin de niveau minimum est allumé, remplissez-le
avec de l'eau pure.
• le réservoir d'eau est plein, le témoin de niveau maximum est allumé, videz l'eau usée
33
FRANÇAIS
15.6
L'imprimante de l'autoclave n'imprime pas
• le volet du porte-rouleau n'est pas bien ferme, la lettre “D” s'affiche. ouvrez et refermez
le volet de l'imprimante en vérifiant sa fermeture.
• plus de papier, la lettre “P” s'affiche. Ouvrez le volet et introduisez le rouleau de papier
thermosensible.
• le rouleau de papier est monté avec la face thermosensible vers l'intérieur. Ouvrez le
volet et tourner le rouleau de papier dans l'autre sens.
15.7
Tâches sur les instruments
• le matériel à stériliser a jauni en raison d'un résidu de liquide chimique qui, avec la
chaleur, s'est fixé sur le matériel. Ceci est dû à un mauvais rinçage.
• la chambre de stérilisation présente des taches jaunes car du matériel contenant du
liquide chimique y a été introduit. en tombant, le liquide s'est fixé par effet de la chaleur.
Ceci est dû à un mauvais rinçage.
• le matériel à stériliser présente des taches blanchâtres car le rinçage a été effectué
avec de l'eau très calcaire et que le matériel n'a pas été bien séché. En dernier rinçage,
il est conseillé d'utiliser de l'eau déminéralisée et de sécher soigneusement le matériel à
stériliser.
• le matériel à stériliser a noirci ceci est dû au fait qu'il contient un grand pourcentage de
carbone
34
FRANÇAIS
16
16.1
DÉMINÉRALISATEUR
Configuration des systèmes de déminéralisation
L’autoclave est prédisposé aussi pour le chargement de l’eau déminéralisé par le
SYSTEME IONIQUE ou SYSTEME D’OSMOSE (accessoires en option).
Ces systèmes sont accessoires des autoclaves TECNO-GAZ S.p.A. (pas en dotation).
Ces produits sont utilises pour éliminer les sels minéraux de l’eau qui vient du réseau
hydrique pour alimenter l’autoclave. Avant d’installer le système IONIQUE ou d’OSMOSE,
l’opérateur doit programmer l’autoclave en suivant les instructions sous indiquées :
Appuyez sur l’interrupteur géneral (vert) et tenir appuyé B-START (FIG.2–POS.19)
jusqu’à quand la langue sélectionnée ne parait sur le visual
Appuyer et relacher B-START pour sélectionner la page-écran suivant, jusqu’à quand ne
parait :
• CHARGEMENT PAR POMPE
• IONIC SYSTEM
• OSMOSI SYSTEM
Pour changer l’affichage, appuyer sur le poussoir B-SELECT
Pour confirmer le l’affichage, appuyer sur le poussoir B-DOOR , sans lacher, jusqu’à
quand on n’écout l’indication acoustique ‘BEEP’
Per uscire premere B-START e tenerlo premuto fino a quando non si sente il segnale
acustico “BEEP”.
Les données configurées sont mémorisées.
ATTENTION:
Le numéro indiqué sous le type de chargement indique approximativement les cycles de
stérilisation que on peut encore démarrer avec le filtre (en cas du système d’osmose) ou
avec les résines (dans le cas du système ionique) actuellement en usage.
Il est de règle de se souvenir de réinitialiser cet numéro au valeur maximum quand on
change le filtre ou les résines sur le système actuellement en usage.
Après avoir sauvé le type de chargement en usage, pour réinitialiser au valeur maximum
cette information on doit appuyer sur le poussoir B-PUMP jusqu’à l’intervention du signal
acoustique (à peut près 10 seconds) dans l’écran d’attente de début cycle.
35
FRANÇAIS
16.2
Connexion des systèmes de déminéralisation
Ici on reporte le détail de l’enclenchement spécifique à l’autoclave du tuyau d’alimentation
de l’eaux et de la connexion de la fiche électrique :
•
•
•
•
•
•
•
•
•
•
•
Éteignez l’autoclave si elle est allumée (FIG.2–pos.19).
Fermez le robinet situé en amont du déminéralisateur.
Installez le déminéralisateur en suivant les indications reportées dans son manuel.
Appliquez du Téflon ou un autre matériau garantissant l'étanchéité à l'eau sur le filet
mâle du raccord porte-tuyau.
Vissez le raccord porte-tuyau sur le filet femelle du tuyau d'évacuation de l'eau propre
(FIG.1–pos.2)
Introduisez le tuyau du déminéralisateur dans le raccord porte-tuyau vissé à présent à
l’autoclave;
Introduisez la fiche du déminéralisateur dans la prise (FIG.1–pos.7) au dos de
l'autoclave;
Ouvrez le robinet situé en amont du déminéralisateur;
Vérifiez s'il n'y a pas de fuites d'eau;
Tourner à l'autoclave;
Effectuez un ou plusieurs cycles de stérilisation pour vérifier le fonctionnement du
branchement à peine effectué en vérifiant surtout l'absence de fuites.
En fin de journée, fermez toujours le robinet situe en amont du
demineralisateur afin de prévenir les inondations
Branchez les systemes de déminéralisation seulement a des autoclaves
TECNO-GAZ prédisposées.
Au debut du cycle de sterilisation, l'autoclave signale E250, il signifie que on
a demarre le cycle avec le niveau d'eau pure bas; donc, il est necessaire
remplir le reservoir d'eau pure.
NOTE:
pour le branchement du déminéralisateur au autoclaves TECNO-GAZ, veuillez-vous
reporter aux indications du manuel d'instructions correspondant
36
FRANÇAIS
FIG.31
16.3
Changement cartouche resines du demineralisateur
L’autoclave signale sur le display l ‘état de déplétion, des résines ou du filtre, avec le
message “ECHANGER RESINES“ (code 242-S) ou
“ECHANGER FILTRE“
(code 248-S-2). Changez la cartouche de résine comme illustré dans le manuel du
déminéralisateur. Après avoir remplacé la cartouche, appuyez sur le bouton B-PUMP
jusqu'au déclenchement du signal acoustique, pour effacer l’autoclave.
37
FRANÇAIS
17
PROCÉDURES STANDARD DE STÉRILISATION
Stériliser veut dire adopter des protocoles opérationnels précis et appliquer un système.
L’autoclave est le maillon final de ce système. Le non respect d'une des phases du cycle
de stérilisation peut nuire au résultat final.
TECNO-GAZ S.p.A. fabrique tous les dispositifs pour chaque phase du cycle de
stérilisation. Il faut :
1 - Prévoir un protocole opérationnel écrit précis à diffuser auprès de tout le
personnel préposé à la stérilisation. TECNO-GAZ S.p.A. peut fournir ces protocoles
personnalisés
2 - Suivre scrupuleusement et avec attention, toutes les phases du cycle de
stérilisation :
DÉSINFECTION phase obligatoire selon la législation en vigueur, qui doit ê
tre effectuée par immersion dans des liquides chimiques ou par désinfection
thermique.
DÉTERSION la phase plus importante qui assure l'élimination de tous les
résidus chimiques et organiques. L'instrument plus idoine est le bac à
ultrasons.
SÉCHAGE phase indispensable pour prévenir la corrosion du matériel à
stériliser et les interférences avec le cycle de stérilisation
ENSACHAGE phase indispensable pour assurer la stérilisation du matériel.
STÉRILISATION phase finale stérilisation à vapeur
38
FRANÇAIS
18
INSTRUCTIONS POUR LE RETOUR DE L'AUTOCLAVE
En cas de retour au fabricant de l’autoclave pour réparation, révision ou validation, suivez
les indications suivantes pour son expédition :
•
Utilisez l'emballage d'origine, s'il n'est plus en votre possession, utilisez un
emballage adéquat. La marchandise voyage aux risques et périls de l'expéditeur.
•
Expédier seulement l’autoclave (n’introduire aucun composant contenu dans le kit
des accessoires).
•
Nettoyez soigneusement la chambre de stérilisation et l’autoclave dans son
ensemble avant de l'expédier. Si elle arrive sale et avec des résidus, elle sera
réexpédiée non réparée ou sera soumise à un nettoyage et désinfection.
•
Videz toujours le réservoir d'eau propre à travers le raccord placé au dos de
l’autoclave (FIG.2–pos.2)
•
Videz toujours le réservoir d'eau usée à travers le raccord placé au dos de
l’autoclave (FIG.2–pos.5)
•
Joignez une lettre d'explication précisant l'anomalie constatée ou le service dont
vous entendez bénéficier.
•
Expédiez toujours en port franc, sinon les frais de transport vous seront facturés.
ATTENTION:
Tous les emballages non d'origine seront éliminés.
L'autoclave vous sera retourné dans un emballage neuf et d'origine afin d'assurer la plus
grande protection pendant le transport. L'emballage vous sera facturé à prix coûtant.
39
FRANÇAIS
19
PROCÉDURES A SUIVRE POUR BÉNÉFICIER DU SAV
Le département réparation est une entité autonome de MEDILINE ITALIA SRL (TECNOGAZ GROUP), qui se compose d’une structure articulée et professionnelle, conformément
aux procédures indiquées dans notre “Manuel Qualité”.
•
Lorsque l’autoclave arrive chez MEDILINE ITALIA s.r.l. elle est réceptionnée
directement par l'atelier de réparation qui le codifie et crée une fiche d'assistance
technique.
•
dans un délai maximum de 48 heures, l'autoclave est vu par les techniciens et un
diagnostic technique est posé ainsi qu'un devis. Si l’autoclave n'a pas été retourné
comme indiqué dans le paragraphe 16 instruction pour retourner l'autoclave, les
délais indiqués ci-dessus peuvent sensiblement se rallonger.
•
Le devis sera fait parvenir au distributeur qui le transmettra au client final pour
acceptation.
•
Après que MEDILINE ITALIA srl aura reçu le devis signé pour acceptation,
l'autoclave sera mis en réparation et sera expédié dans le délai indiqué sur le devis.
•
L'autoclave sera réexpédié dans son emballage d'origine à travers le transporteur
du client.
IMPORTANT:
Si le problème persiste après avoir effectué la solution décrite, contactez le SAV du
fabricant par téléphone au numéro suivant: MEDILINE ITALIA srl +39 0522 - 94 29 97.
40
FRANÇAIS
20
GARANTIE
1) DUREE: le produit est couvert par une garantie de 12 (douze) mois ou 1200 cycles.
2) EXCLUSION: sont exclus de la garantie:
a) les problèmes liés au manque d’entretien ordinaire, à une négligence de l’utilisateur
ou à une utilisation impropre du produit ;
b) les contrôles périodiques et l’entretien ;
c) les réparations ou les remplacements de pièces soumises à usure, fragiles ou dont
la durée de vie est imprévisible, à moins que leur défaut n’ait été prouvé au moment
de la livraison ;
d) les problèmes relatifs à la main d’œuvre, au déplacement du personnel technique
et au transport ;
e) les problèmes ou les dommages liés à une utilisation impropre et à une utilisation
erronée ;
f) les problèmes ou les dommages liés à des impuretés dans les systèmes
d'alimentation en eau et en air, les opérations chimiques ou électriques
extraordinaires ;
g) les problèmes ou les dommages liés à l’utilisation de détergents, désinfectants,
substances ou processus de stérilisation, non expressément indiqués dans le
manuel d’utilisation et d’entretien ;
h) le virage normal des pièces en plastique.
3) PROCÈS-VERBAL D’INSTALLATION ET D’ESSAI: ce procès-verbal d’installation et
d’essai dûment rempli, dans sa totalité, signé par l’utilisateur et le technicien installateur
doit être renvoyé à l’entreprise. Cette condition est essentielle pour pouvoir bénéficier de
la garantie. Ce procès-verbal devra être renvoyé dans un délai de 15 jours à compter de
l’installation, sous peine d’expiration de la garantie.
4) LIMITES : la garantie donne droit au remplacement ou à la réparation gratuits des
composants défectueux. Le remplacement de l’appareil tout entier reste toutefois exclu.
Pour les composants installés ou intégrés fabriqués sous la marque TECNO-GAZ ou par
des tiers fournissant leur propre certificat de garantie, ce sont les conditions, les limites et
les exclusions indiquées sur ces certificats qui sont valables.
La réalisation d’une ou de plusieurs réparations durant la période de garantie ne comporte
aucune modification de la date d’expiration de la garantie.
5) CONTESTATIONS : en cas de contestation de l’Acheteur sur l’application de la
garantie, sur la qualité ou sur les conditions de l’équipement livré, l’Acheteur ne pourra en
aucun cas suspendre et/ou retarder les paiements.
6) LITIGES : tout litige sur l’application et l’interprétation de ce Certificat de garantie sera
du ressort du Tribunal de PARME (Italie) indépendamment du lieu où le contrat d’achat de
l’équipement a été stipulé.
7) DÉROGATIONS : les dérogations spéciales, accordées dans le cadre des présentes
normes de garantie, ne comportent la reconnaissance d'aucun droit pour l’Acheteur et
devront être considérées comme étant accordées dans la limite de chaque cas spécifique.
8) AUTRE : pour tout ce qui n’est pas expressément prévu par ce Certificat de garantie, ce
sont les normes contenues dans le Code Civil et dans les Lois en matière de la
République Italienne qui sont valables.
41
FRANÇAIS
La garantie expire si :
a) l’équipement est endommagé suite à une chute, l’exposition aux flammes, les
renversements de liquides, la foudre, les calamités naturelles, les intempéries ou toutefois
suite à des causes non imputables à des défauts de fabrication ;
b) l’installation n’est pas conforme aux instructions de TECNO-GAZ et a été effectuée par
du personnel non autorisé ;
c) l’équipement est réparé, modifié ou toutefois altéré par l’Acheteur ou par des tiers non
autorisés ;
d) au moment de la demande d’intervention sous garantie, le numéro de série du produit a
été enlevé, effacé, falsifié, etc. ;
e) le procès-verbal d’installation et d’essai rempli et signé n’est pas renvoyé dans les 15
jours suivants l’installation ;
f) l’Acheteur suspend et/ou retarde, pour quelle que raison que ce soit, le paiement de
toute somme due pour l’achat et/ou l’entretien de l’équipement ;
g) l’entretien périodique programmé ou toute autre prescription prévue dans le manuel
d’utilisation et d’entretien n’est pas respecté.
42
FRANÇAIS
A
ENTRETIEN
Un entretien correct de l'autoclave garantit son bon fonctionnement et un gain assuré en
termes de pertes de temps et de coûts dus à l'intervention du SAV et de l'entretien.
Les suivantes opérations sont obligatoires et à exécuter par l’opérateur.
Nettoyage de la chambre
Chaque 20 cycles ou 1 fois par semaine
Nettoyer périodiquement la chambre et le filtre, et enlever tout éventuel dépôt ou débris,
vous éviterez ainsi d’introduire dans le circuit de vidange des matériaux qui peuvent
causer des obstructions. Pour une bon nettoyage employer exclusivement de l’eau
déminéralisée et l’éponge en dotation (partie pas abrasive – FIG.13).
Cette opération doit etre exécutée absolument avec chambre froide, pour éviter des
brulures. - N'utilisez jamais de solvants, détergents, solutions chimiques, détartrants ou
autres produits similaires.
Nettoyage du filtre chambre
Chaque 20 cycles ou 1 fois par semaine
Tirer vers l’haut le filter (code DXBA091), avec l’attention de ne le dommager pas, le
laver avec de l’eau déminéralisée et le sécher avec un drop sec et nettoyé. Mettre le
filtre dans son siège, faisant attention qu’il sort de à peu près 15 mm.
Nettoyage du plateaux et porte-plateaux
Chaque 20 cycles ou 1 fois par semaine
Nettoyer avec l’éponge en dotation (partie pas abrasive) mouillée avec de l’eau
déminéralisée.
Changement de filtre bactériologique
Chaque 200 cycle ou quand il devient
plus foncé
Changez le filtre bactériologique (FIG.2–pos.10) Tourner le filtre dans le sens inverse
des aiguilles d’une montre pour le dévisser et dans le sens des aiguilles d’une montre
pour le visser. Utilisez seulement et exclusivement des filtres d'origine (code DAVA101).
Nettoyage joint de la porte
Chaque 20 cycles ou 1 fois par semaine
Éliminez régulièrement les résidus qui pourraient se déposer sur le pourtour de la
garniture (code DANA038, avec de l'eau déminéralisée et l'éponge en dotation (partie
pas abrasive).
A-1
FRANÇAIS
Régulation de la porte
Chaque 2 mois
Pour régler la pression de fermeture de la porte, agissez sur le régulateur dudit porte
avec la clé à double fonction (FIG.10) fournie en dotation (code DANA008). Tournez
dans le sens opposé des aiguilles d'une montre (pos.A) pour augmenter la pression de
fermeture. Tournez dans le sens des aiguilles d'une montre (pos.B) pour diminuer la
pression de fermeture.
A
Changement du papier de l'imprimante
B
-
Rouleau de papier thermographique de 57 mm de large. Le papier thermographique doit
être imprimé sur un côté seulement (code DNVA363). Ouvrir le capot de l’imprimante en
poussant le bouton insérer le rouleau de papier en faisant attention à la direction de
rotation , tirer le papier jusqu’à le faire ressortir et refermer le capot; l’imprimante est prête
pour l’impression.
A-2
FRANÇAIS
Les stivante opérations sont conseillées par le producteur. Ces operations doivent etre
executes par des techniciens formés et autorisés par le producteur.
Remplacement du filtre chargement eau
Chaque 1 an / 1000 cycles*
Controle du filtre eau entrée fluximetre
Chaque 1 an / 1000 cycles*
Nettoyage filtre laiton Y en aval au radiateur
Chaque 1 an / 1000 cycles*
Nettoyage de l’électrovanne charge eau EV4
Chaque 1 an / 1000 cycles*
Remplacement valves Pompe à vide
Chaque 1 an / 1000 cycles*
Nettoyage du radiateur
Chaque 1 an / 1000 cycles*
Nettoyage des ventilateurs
Chaque 1 an / 1000 cycles*
Controle intégrité du circuit pneumatique
Chaque 1 an / 1000 cycles*
Controle intégrité circuit électrique
Chaque 1 an / 1000 cycles*
Controle isolant thermique chambre de stérilisation
Chaque 1 an / 1000 cycles*
Controle serrage vis
Chaque 1 an / 1000 cycles*
Controle sondes de température
Chaque 1 an / 1000 cycles*
Controle valve de sécurité
Respencter les indication du
producteur (voir instructions
annexées)
Graisser la partie mobile du levier de la poignée
Chaque 1 an / 1000 cycles*
Controler les sondes de niveau
Chaque 1 an / 1000 cycles*
Controler les conditions du reservoir
Chaque 1 an / 1000 cycles*
* il est valable la première condition atteinte.
A-3
FRANÇAIS
En plus quelques dispositions régionales/nationales, établient la condition obligatoire
d’une validation de l’appareil, qu’en controle les paramètres de stérilisation. La validation
périodique doit etre toujours faite par des techniciens spécialisés avec des instruments de
mesures aptes. Dans le cas que la validation ne soit pas obligatoire, la validation meme
devient une opération volontaire avec le but de certifier que l’autoclave est parfaitement
fonctionnante et apte à la stérilisation des charges pour lesquels elle a èté projeté et
fabriquée
NOTE
A-4
ESPAÑOL
Estimado Cliente,
Les agradecemos por elegir nuestro autoclave y sepan que sabremos corresponder a su
confianza, con la máxima atención y un servicio adecuado a sus expectativas.
El autoclave que eligieron es la máxima expresión de tecnología aplicada.
Puede utilizarse en el sector dental, médico, estético y, por lo general, en todos los
sectores que requieren la esterilización segura de todo tipo de instrumentos y materiales:
sólidos, huecos, porosos, libres y envasados.
Antes de utilizar el autoclave les rogamos leer con la máxima atención el manual de uso y
que lo guarden en un lugar accesible a todos los operadores encargados de la
esterilización.
El autoclave deberá ser utilizada exclusivamente para las operaciones previstas por el
fabricante.
La instalación, el mantenimiento y asistencia requeriá exclusivamente la
intervención de técnicos autorizador. Les rogamos que utilicen y pidan
exclusivamente repuestos originales.
1
ESPAÑOL
ÍNDICE
1. CONSULTA DEL MANUAL
1.1
GLOSARIO
1.2
DESCRIPCIÓN VISTA DELANTERA / TRASERA
2. SEGURIDAD
2.1
MARCACIÓN DE SEGURIDAD
2.2
DISPOSITIVO DE SEGURIDAD
2.3
NOTAS SOBRE LA SEGURIDAD
2.4
ELIMINACIÓN
3.
FINALIDAD DE EMPLEO DEL AUTOCLAVE
4. DATOS TÉCNICOS
4.1
MECÁNICOS
4.2
ELÉCTRICOS
4.3
CÁMARA
4.4
DEPÓSITO AGUA LIMPIA
4.5
DEPÓSITO AGUA USADA
4.6
FILTRO BACTERIOLÓGICO
5.
ACCESORIOS
6.
DESEMBALADO
7.
INSTALACIÓN
8. INSTRUCCIONES PARA EL USO
8.1
ACCIONAMIENTO DEL AUTOCLAVE
8.2
LLENADO DEPÓSITO AGUA PURA
8.3
CARACTERISTICAS DEL AGUA UTILIZADA
8.4
CARGA DE MATERIALES EN AUTOCLAVE
8.5
INICIO CICLO DE ESTERILIZACIÓN
8.6
FINAL CICLO DE ESTERILIZACIÓN
8.7
EXTRACCIÓN MATERIALES ESTERILIZADOS
8.8
DRENAJE DEL AGUA UTILIZADA
8.9
INTERRUPCIÓN DE UN CICLO DE ESTERILIZACIÓN
9. CICLOS DE ESTERILIZACIÓN
9.1
DESCRIPCIÓN DE LOS CICLOS
9.2
CICLOS OPERATIVOS
9.3
CICLOS NOCTURNOS
9.4
CICLOS DE TESTE
10. TEST DE CONTROL DEL AUTOCLAVE
10.1
INTEGRADORES QUÍMICOS
10.2
INDICADOR BIOLÓGICO
10.3
TEST DE BOWIE&DICK
2
ESPAÑOL
10.4
10.5
TEST DE HÉLICE
TEST DE VACÍO
11. LAS TABLAS Y LAS DIAGRAMAS
12. LECTURA DEL INFORME DE CICLO
13. MENSAJES DE ERROR Y ALARMAS
14. PROGRAMACIÓN DE PARÁMETROS DISPLAY
14.1 SELECCIÓN IDIOMA
14.2 SELECCIÓN HORA
14.3 SELECCIÓN MINUTOS
14.4 SELECCIÓN SEGUNDOS
14.5 SELECCIÓN FECHA
14.6 SELECCIÓN MES
14.7 SELECCIÓN AÑO
14.8 VISUALIZACIÓN PRESIÓN ATMOSFÉRICA
14.9 SELECCIÓN CARGA AGUA
15. SOLUCIÓN DE PROBLEMAS OPERATIVOS
15.1 AUTOCLAVE NO EFECTUA SECADO CORRECTO
15.2 LA CÁMARA DEL AUTOCLAVE CAMBIA A COLOR BLANCO
15.3 LA CÁMARA DEL AUTOCLAVE PRESENTA ,MANCHAS VERDES
15.4 EL CICLO DE ESTERILIZACIÓN SE INTERRRUMPE
15.5 EL AUTOCLAVE NO RECIBE LOS MANDOS
15.6 LA IMPRESORA DEL AUTOCLAVE NO IMPRIME
15.7 MANCHAS EN LOS INSTRUMENTOS
16. DESMINERALIZADOR
16.1 IMPOSTACIÒN DEL SISTEMA DE DEMINERALIZACIÒN
16.2 CONEXION SISTEMA DE DESMINERALIZACIÒN
16.3 CAMBIO CARTUCHO RESINA FILTRE
17. PROCEDIMIENTOS ESTÁNDAR DE ESTERILIZACIÓN
18. INSTRUCCIÓN PARA EXPEDICIÓN DEL AUTOCLAVE
19. PROCEDIMIENTOS PARA EL SERVICIO Y ASISTENCIA
20. GARANTÍA
A.
MANTENIMIENTO
3
ESPAÑOL
CONSULTA DEL MANUAL
1
1.1
Glosario
VACÍO
CELDA
FILTRO
TRAY
PORTATRAY
DESMINERALIZADOR
HELIX TEST
BOWIE - DICK TEST
VACUUM TEST
FILTRO
BACTERIOLÓGICO
SOBREPRESIÓN
Presión inferior a la presión atmosférica
Cuerpo cilíndrico de acero que aloja los dispositivos a esterilizar
Dispositivo apto a retener las partículas
Cajón de soporte de la carga a esterilizar
Estructura de soporte de las bandejas en la celda de esterilización
Dispositivo apto para eliminar los sales minerales del agua
proveniente de la red hídrica
Teste para comprobación de la penetración del vapor en un
cuerpo hueco
Teste para comprobación de la penetración del vapor en un
cuerpo
poroso
Teste para
comprobación del mantenimiento del grado de vacío
en la celda
Dispositivo apto para retener las partículas en suspensión en el
aire superiores a 0,3 µm.
Presión superior a la estándar para el ciclo considerado
SOBRETEMPERATURA Temperatura superior a la estándar para el ciclo considerado
Componente reemplazable (cartucha) del desmineralizador que
elimina todos los sales de tipo inorgánico
RESINAS
1.2
Descripción vista delantera / trasera
Patilla separadora
Válvula de rebose de agua limpia
Grifo de drenaje de agua limpia – conexión
desmineralizador
Válvula de Seguridad
Válvula de rebose del agua usada - descarga
condensación
Grifo de drenaje de agua usada
Alimentación Eléctrica Principal
Toma para alimentación eléctrica solo y
exclusivamente desmineralizador
Fusible 12A
FIG.1
Fusible 12A
Fusible 500mA
4
ESPAÑOL
FIG.2
5
ESPAÑOL
Filtro Bacteriológico
Carga manual de Agua
Llenado con agua
Drenaje agua Usada
Display
Impresora
Conexión RS232
Incubadora de conector biológico B-TEST
Interruptor general
Ciclo 121°C
(válida para materiales envasado e no envasado)
Ciclo 134°C
(válida para materiales envasado e no envasado)
Ciclo 134°C Flash
Ciclo 134°C Prion
Vacuum Test
Helix / Bowie&Dick Test
Mínimo nivel de agua
Máximo nivel de agua
Máximo nivel de agua usada
B-START
Botòn START-STOP
B-PUMP
Botòn llenado Agua
B-SELECT
B-DOOR
6
Botòn Selección Ciclo
Botòn Tapa abierta
ESPAÑOL
2
2.1
SEGURIDAD
Marcación de Seguridad
TENSIÓN PELIGROSA
QUITAR TENSIÓN ANTES DE
RETIRAR LA TAPA
ALTA TEMPERATURA
CONEXIÓN DE TIERRA
2.2
Dispositivos de Seguridad
Los Dispositivos de Seguridad son los siguientes:
-) Válvula de Seguridad calibrada a 2.4 bar ± 10%
-) Bloqueo electromagnético para impedir la apertura de la tapa durante la ejecución del ciclo
-) Termostato de seguridad
2.3
Notas sobre la Seguridad
• El fabricante es responsable del producto puesto en comercio cumpliendo la normativa
en vigor. La responsabilidad caducará cuando personas no capacitadas realicen
operaciones con el dispositivo o partes del mismo o cuando se utilicen piezas de
repuesto no originales.
• Los locales donde se instale el autoclave no deben estar expuestos a riesgos
potenciales de explosiones y/o incendios.
• El autoclave debe estar instalado en un ambiente conforme a los requisitos legislativos
en vigor.
• El conector de la incubadora (FIG.2-pos.18), debe utilizarse únicamente para la
conexión de B-TEST.
Trátandose de una conexión de 230 V c.a. independiente del interruptor general,
se recomienda no tocar el conector con las manos mojadas o húmedas y
protegerlo de posibles salpicaduras de agua.
7
ESPAÑOL
2.4
Eliminación
Este producto cumple todos los requisitos establecidos por la Directiva
2002/96/CE del Parlamento Euoropeo y del Consejo de la Unión Europea sobre
Residuos de Aparatos Eléctricos y Electrónicos (RAEE). En los estados
miembros que adoptaron la citada Directiva, deberá inicarse en el producto que
fue puesto en el mercado después del 13 de agosto de 2005 y que no debe
eliminarse como residuo urbano no reciclable. Utilice las instalaciones locales de
recogida de RAEE que correspondan para deshacerse de este producto de
acuerdo con todos los requisitos estipulados.
3
FINALIDAD DE EMPLEO DEL AUTOCLAVE
El autoclave es capaz de esterilizar las tres tipologías de carga previstas por la norma
EN13060, y específicamente:
MATERIALES
Instrumentos sin cavidades y sin obstáculos para la penetración del
FERROSOS O SÓLIDOS vapor
CUERPOS POROSOS
Materiales sencillos o compuestos que puedan absorber los fluidos
(tejidos, batas, gasas, vendas, etc.)
CUERPOS HUECOS
Materiales o dispositivos con cavidades, obstrucciones, etc.
Estos están subdivididos en dos tipos, clasificados en función de la
longitud y del diámetro. Indicativamente:
TIPO B: cánulas, tubos o dispositivos con pasadas considerables
TIPO A: turbinas, empuñaduras y dispositivos con orificios ciegos o
de pequeñas dimensiones
El autoclave ha sido certificado para la esterilizaciòn de los siguientes materiales*:
EUROPA B EVO
EUROPA B EVO 24
max kg. 5
max kg. 7
max kg. 5
max kg. 7
max kg. 1,5
max kg. 2
Materiales solidos ensobrados y no
ensobrados para un cargo maximo de
Materiales concavos de tipologia a y b
para un cargo maximo de
Materiales porosos para un cargo
maximo de
* Válido solamente para los países europeos
El autoclave debe utilizarse, sólo y exclusivamente para la esterilización de
instrumentos y materiales compatibles con el sistema de esterilización por
vapor. Asegurese siempre de que los elementos a esterilizar puedan
soportar la temperatura del ciclo seleccionado.
8
ESPAÑOL
4
4.1
DATOS TECNICOS
Mecánicos
EUROPA B EVO
Temperatura de trabajo
Altitud MAX
Humedad relativa MAX a 30°C
Humedad relativa MAX a 40°C
Dimensiones totales (L x H x P)
Dimensiones con tapa abierta
Peso (depósitos vacíos)
Peso (depósitos llenos)
Peso por área de soporte
Nivel de potencia sonora
4.2
+5°C ÷ +40°C
2.000 m
80%
50%
510 X 390 X 590 mm
510 X 390 X 730 mm
300 mm
54 kg
58 kg
63 kg
67 kg
2
2058 N/m
2058 N/m2
< 70 db A
Eléctricos
Tensión alimentación
Potencia
Frecuencia
Cable alimentación
Fusibles
Calor transmitido
4.3
EUROPA B EVO 24
EUROPA B EVO
EUROPA B EVO 24
230 V a.c. +/-10 % single phase
1.7 kW
2.2 kW
50 / 60 Hz
2 + 1 x 1mm2
6. 3 x 32 – 12A (x2) / 6.3 x 32 – 500mA (x1)
3.6 E6 J / ora
Cámara
Presión trabajo MAX
Vacío MAX
Temperatura MAX
Material
Dimensiones
EUROPA B EVO
EUROPA B EVO 24
2.4 bar (relativi)
- 0.9 bar (relativi)
138 °C
Inox AISI 304
Ø 245 x 318 mm
Ø 245 x 500 mm
9
ESPAÑOL
4.4
Depósito Agua Limpia
EUROPA B EVO
EUROPA B EVO 24
Con demineralizz.
4,5 l
Volumen
Ciclos efectuables
Leer impostaciòn del
sistema de
demineralizaciòn
4
Depósito Agua Usada
EUROPA B EVO
Volumen
Ciclos efectuables
Material
Temperatura max. agua de
drenaje
4.6
EUROPA B EVO 24
4,5 l
4
2
polietileno
50°C
Filtro Bacteriológico
EUROPA B EVO
Diàmetro
Capacidad de filtrado
10
2
polietileno
Material
4.5
Senza
demineralizz.
EUROPA B EVO 24
56 mm
0.3 µm
ESPAÑOL
05
ACCESORIOS
Portabandejas
EUROPA B EVO
EUROPA B EVO 24
Aluminio Anodizado
192 x 165 x 280 mm
192 x 200 x 470 mm
Material
Dimensiones (L x H x P)
Cuadro
FIG.4
FIG.3
Dotación prevista
1
Portabandeja (opcional)
EUROPA B EVO
Aluminio Anodizado
193 x 200 x 280 mm
(200 x 193 x 280) mm
Material
Dimensiones (L x H x P)
EUROPA B EVO 24
/
/
/
Cuadro
FIG.5
Dotación prevista
1
/
Bandejas
EUROPA B EVO
Material
Dimensiones (L x H xP)
184 x 17 x 286
mm
EUROPA B EVO 24
Aluminio Anodizado
184 x 17 x 286 mm + 184 x 17 x 140 mm
Cuadro
+
FIG.6
Dotación prevista
4
FIG.7
FIG.8
4+4
11
ESPAÑOL
Bandejas (opcional)
EUROPA B EVO
Material
/
Dimensiones (L x H x P)
/
Cuadro
/
Dotación prevista
/
EUROPA B EVO 24
Aluminio Anodizado / de
acero inoxidable
184 x 17 x 460 mm
FIG.9
4
Llave para extracción del tray y regulación de la portezuela
Utilízala para extraer y manejar los trays (FIG.11) y también para regular la portezuela
(apartado 14.4)
Cuadro
FIG.11
FIG.10
Dotación prevista
1
Embudo de plástico con tubo
Utilízalo para la carga manual del agua (apartado 8.2)
Cuadro
FIG.12
Dotación prevista
1
Paño para la limpieza de la celda y del burlete de la portezuela
Utilízalo para limpiar la celda de esterilización y el burlete de la portezuela (apartado A)
Cuadro
FIG.13
Dotación prevista
12
1
ESPAÑOL
Racor para carga de agua
Utilízalo con el embudo de plástico con tubo para cargar el agua en manual y con el tubo
para carga de agua completo con filtro
Cuadro
FIG.14
Dotación prevista
1
Tubo para carga de agua completo con filtro
Utilízalo para cargar el agua en manual (apartado 8.2)
Cuadro
FIG.15
Dotación prevista
1
Tubo de desagüe
Utilízalo para descargar el agua usada (apartado 8.9)
Cuadro
FIG.16
Dotación prevista
1
Tubo rilsan
Une un extremo del tubo al Rebosadero del agua usada – descarga de condensación
(FIG.1 – pos.4), el otro extremo debe acoplarse a un contenedor para recolección del
agua y del vapor condensado.
Cuadro
FIG.17
Dotación prevista
1
13
ESPAÑOL
Patilla separadora de plástico posterior
Aplica la patilla separadora de plástico en la parte posterior del autoclave (FIG.1–pos.0)
para asegurar la ventilación adecuada en caso de que se posicione la autoclave
demasiado cerca de una pared
Cuadro
FIG.18
Dotación prevista
1
Tubos de descarga posterior de dependencias
1- Tubo para rebosadero agua limpia: Une un extremo del tubo al rebosadero posterior del
agua limpia (FIG.1–pos.1), y el otro extremo a un contenedor para recuperación del agua
desmineralizada o a un punto de desagüe (versión con desmineralizador).
2- Tubo para grifo de descarga del agua usada: Une un extremo del tubo al racor FIG.20 y
ensambla en él el grifo para el desagüe del agua usada (FIG.1–pos.5), y el otro extremo
del tubo a un contenedor para recoger el agua.
Cuadro
FIG.19
Dotación prevista
2
Racor para desagüe del agua usada posterior
Conecta un extremo del tubo de desagüe del agua usada al racor FIG.20 y fija en él el
grifo de desagüe del agua usada (FIG.1-pos.5).
Cuadro
FIG.20
Dotación prevista
14
1
ESPAÑOL
Cable de alimentación eléctrica
Une el extremo del conector (FIG.21–pos.B) al panel trasero (FIG.1–pos.6) y
seguidamente enchufa la clavija (FIG.21–pos.A) directamente en la toma de alimentación
del sistema eléctrico.
Cuadro
FIG.21
Dotación prevista
06
1
DESEMBALADO
El autoclave será expedido en embalaje idóneo que protegerá el contenido y permitirá su
fácil transporte y manejo.
El embalaje no deberá sufrir choques, requeriendo manejo atento que evite vuelcos y
caídas de alturas superiores a los 16 cm.
En caso de imposibilidad de utilizar medios de desplazamiento autónomos, el instrumento
debe ser manejado dentro de su embalaje y siempre con 2 personas para las
operaciones. El autoclave està puesta arriba de un pallet de madera y cerrada en un
cartòn ondulado y reforzado para dentro con compuestos de cartòn.
Para desembalar el autoclave, abrir el cartòn ondulado, remover las partes de refuerzo y
extraerlo utilizzando las correas en dotaciòn.
La manipulación sólo podrá realizarse mediante el uso de correas y por
mínimo dos personas.
Nunca deberá levantarse el autoclave asiéndolo por la parte inferior de la
tapa o del panel de mandos pues esa operación además de errónea puede
originar problemas de naturaleza mecánica.
ATTENZIONE: Consérvese siempre el embalaje original.
Dentro del embalaje se encuentran :
• MANUAL DE USO: que deberán leer atentamente guardandolo en lugar accesible a
todos los operadores responsables de la esterilización.
• CERTIFICADOS DE CONFORMIDAD: que deberán guardar.
• BOLETÍN DE GARANTÍA: que deberán completar enviando una parte.
15
ESPAÑOL
07
INSTALACIÓN
FIG.22
EUROPA B EVO
A
B
C
D
E
F
EUROPA B EVO 24
610 mm
590 mm
730 mm
50 mm
410 mm
426 mm (trasero)
410 mm (trasero)
455 mm (delatero)
444 mm (delatero)
400 mm
515 mm
1 - Instalar el autoclave en ambientes idóneos para la esterilización.
2 - El local debe estar adecuadamente iluminado y aireado, conforme lo previsto por las
directivas en vigor.
3 - Instalar el autoclave lejos de fuentes de calor y salpicaduras de agua.
4 - Colocar el autoclave sobre una superficie idónea que pueda soportar el peso (min.80
kg) y que tenga las dimensiones apropiadas.
5 - La superficie de apoyo de la autoclave debe estar perfectamente horizontal.
6 - Colocar el autoclave a una altura que permita al utilizador la inspección total de la
cámara de esterilización y que posibilite la realización de la limpieza.
7 - Abrir la tapa del autoclave y retirar todos los sobres que envolucran los accesorios
que se encuentran en el interior de la cámara de esterilización.
16
ESPAÑOL
8 - Dejar en el interior de la cámara de esterilización solo el portabandeja con las
bandejas, todos los demás accesorios deben ser guardados en el compartimiento externo
a disposición de los utilizadores.
9 - Nunca deben dejarse sobre la máquina periódicos, bandejas, contenedores con
líquidos, etc …
10 - Nunca apoyarse a la tapa cuando esté abierta.
11 – Crea un espacio de por lo menos 5 cm detrás y por los lados del aparato utilizando la
patilla espaciadora de plástico (FIG.1–pos.0 / FIG.18) garantizando la ventilación
necesaria.
12 - Efectúa la conexión de los tubos en la parte trasera (capítulo 5)
13 - Comprobar siempre si el sistema eléctrico al que se enlaza el autoclave cumple las
normas vigentes y que tenga las dimensiones adecuadas a las características del aparato.
14 - Tome el cable de poder y participar el receptáculo hembra en el enchufe en el panel
posterior de la autoclave (FIG. –pos.6)
15 - Conectar el enchufe eléctrico al sistema comprobando una alimentación de 230 Vac2000Va
IMPORTANTE:
Evitar el enlace de extensiones, reductores o adaptadores pues pueden originar
microinterrupciones con consecuente señalización de alarma.
16- Activar el autoclave pulsando el interruptor general (FIG.2–pos.19) y abrir la tapa del
autoclave. Aguardar algunos segundos, habrá dos señales acústicas que informarán de la
adquisición de los parámetros relativos a
alineación atmosférica automática y
simultáneamente aparecerá en el display el mensaje “PUERTA ABIERTA”.
IMPORTANTE:
Nunca debe seleccionarse un mando antes de las dos señales acústicas pues el
autoclave no aceptará el programa elegido.
17- Llenar el depósito con agua pura siguiendo las instrucciones del capítulo 8.
18 - Si se instala el autoclave con el sistema de desionización, deben respetarse las
instrucciones proporcionadas en el embalaje del dispositivo (capítulo 16).
17
ESPAÑOL
08
INSTRUCCIONES PARA EL USO
Después de instalada, el autoclave deberá ser preparada para la utilización.
8.1
Accionamento del autoclave
Pulsando el interruptor general (FIG.2–pos.19)
8.2
Llenado depósito agua limpia
Conecta el “racor para carga de agua “ FIG.14 al “tubo para carga de agua completo con
filtro” (en dotación) introduciéndolo en la abertura del racor frontal del autoclave (FIG.2–
pos.13).
Coloca el otro extremo del tubo con filtro dentro del contenedor del agua desmineralizada
o destilada.
Presiona ahora el pulsador B-PUMP para accionar la bomba para carga de agua y
manténgalo pulsado hasta visualizarse la cuenta atrás.
La bomba llenará el depósito interno del autoclave. No alcanzándose el nivel máximo en
el plazo de 180 segundos la bomba se parará automáticamente siendo necesario pulsar
nuevamente el botón de accionamiento B-PUMP.
La bomba parará automáticamente al alcanzarse el nivel máximo.
En caso de fallo en el funcionamiento de la bomba de entrega de agua el llenado del
depósito podrá efectuarse en manual y el operador deberá:
• Acciona el autoclave.
• Extrae la tapa frontal del autoclave (FIG.2–pos.12)
• Coloca en esa abertura el racor para carga de agua con tubo acoplado al embudo
(FIG.14)
• Vierte el agua destilada en el embudo manteniéndolo en posición más alta respecto del
racor de carga
• Vierte el agua destilada en el embudo manteniéndolo en posición más alta respecto del
racor de carga (FIG.2–pos.2).
• Completada la introducción del líquido, extrae el racor y repón la tapa.
Con enlace a desionizador el funcionamiento del autoclave estará inhibido en caso de
nivel máximo no alcanzado.
18
ESPAÑOL
8.3
Caracteristicas del agua utilizada
TABLA NIVELES CUALITATIVOS ESTABLECIDOS POR LA NORMATIVA UNI EN 13060 : 2009
CEN STANDARD UNI EN 13060 : 2009
Resíduos de evaporación
10
mg/l
≤
Óxidos de silício
1
mg/l
≤
Hierro
0.2
mg/l
≤
Cadmio
0.005
mg/l
≤
Plomo
0.05
mg/l
≤
Restos de metales pesados excepto hierro, cadmio,
0.1
mg/l
≤
plomo
Cloruro (Cl’)
2
mg/l
≤
Fosfato (P20s)
0.5
mg/l
≤
Conductividad (a 20°C)
15
≤
µs/cm
Valor pH (grado de acidez)
5 ÷ 7,5
Aspecto
Incolor, limpido, sin depósitos
Dureza (E Ioni de tierra alcalina)
0.02
mmol/l
≤
8.4
Carga de materiales en autoclave
Disponer los materiales por esterilizar en las bandejas, atención:
• nunca sobreponer los materiales
• disponer los instrumentos envasados siempre con la cara de papel hacia arriba
• nunca permitir el contacto entre los materiales y la cámara de esterilización o la tapa
cerrada
• colocar las pinzas y las tijeras con las hojas abiertas
FIG.23
Al completar la carga, cerrar la tapa del autoclave. En el display aparecerá el mensaje
“PUERTA CERRADA”.
19
ESPAÑOL
8.5
Inicio ciclo de esterilización
Elegir el programa de esterilización más idóneo para la carga introducidapulsando el
botón B-SELECT
Cada opresión del botón alterará la selección del programa pasando a la opción sucesiva.
Después de elegir el programa será posible accionar el ciclo pulsando el botón B-START,
la tapa se bloqueará automáticamente y el ciclo empezará.
Durante el ciclo, el display visualizará todos los parámetros y las informaciones
correspondientes.
8.6
Final ciclo de esterilización
Una señal acústica avisa a los operadores de que el ciclo de esterilización ha finalizado y
en la pantalla aparece el mensaje “CICLO FINALIZADO”.
Al terminar la señal acústica, desbloquear la tapa oprimiendo el botón B-DOOR.
Habiendo presión en el interior de la cámara, el botón no accionará el desbloqueo.
Aguardar la completa despresurización de la cámara y repetir la operación ó, sólo en
casos de emergencia, pulsar simultáneamente B-DOOR + B-PUMP .
Desbloqueada la tapa, tirar de la empuñadura y abrirla.
8.7
Extracción materiales esterilizados
Utiliza equipos de protección individual idóneos cumpliendo las normativas vigentes en
materia de seguridad e higiene laboral. Extrae los trays con el auxilio de la llave específica
suministrada con el aparato (FIG.10). Deja que los utensilios se acondicionen antes de
guardarlos en ambiente al amparo de contaminaciones.
8.8
Drenaje del agua utilizada
Al encenderse el led de nivel de agua usada (FIG.2–pos.28) deberá accionarse el
vaciado. Sin el vaciado, el funcionamiento del autoclave estará inhibido.
Introducir el tubo (FIG.16), en la toma de desagüe del agua usada situada en la parte
frontal del autoclave (FIG.2–pos.14), desenroscar la anilla girando a izquierdas (dos
vueltas completas), introducir la otra extremidad del tubo en un recipiente; el agua verterá
en el contenedor por gravedad.
IMPORTANTE:
A - El tubo alojado en el contenedor de captación nunca deberá rozar ni sumergirse en el
agua drenada pues son posibles situaciones de reflujo.
B - Aguardar siempre hasta que el agua haya salido totalmente. El led de nivel máximo de
agua usada se apagará antes que el agua termine en el depósito y por ello no debe ser
considerado como referencia en esa operación.
Al terminar el drenaje al tornillo la anilla y quitar el tubo.
20
ESPAÑOL
8.9
Interrupción de un ciclo de esterilización
Un ciclo de esterilización podrá ser interrumpido voluntariamente pulsándose el botón
B-START por aproximadamente 2 segundos.
El autoclave emitirá un sonido y pasará a la descompresión mostrando en el display el
mensaje de error A001 (CICLO INTERRUMPIDO).
Para reponer a cero el error deben mantenerse oprimidos simultáneamente los botones
que destacamos a continuación hasta que intervenga la señal acústica:
B-SELECT
B-PUMP
B-DOOR
ATENCIÓN: Con impresora en funcionamiento, el reset de la alarma estará inhibido.
21
ESPAÑOL
09
9.1
CICLOS DE ESTERILIZACIÓN
Descripción de los ciclos
El autoclave dispone de tres series de ciclos:
A - ciclos operativos
B - ciclos nocturnos
C - ciclos de testes
9.2
Ciclos operativos
Todos los ciclos operativos disponen de un sistema de vacío fraccionado y por lo tanto
pueden esterilizar materiales huecos, porosos, sólidos, libres o envasados.
Las temperaturas seleccionables son 121°C – 134°C.
Normalmente los ciclos de 121°C, son utilizados para ter moplásticos o materiales
sensibles mientras que los ciclos de 134°C, se utilizan par a todos los demás materiales.
Seguir siempre las instrucciones proporcionadas por los productores de los instrumentos
o de los dispositivos esterilizados.
Habrá además 2 ciclos especiales:
- Ciclo Prión, estudiado para la enfermedad de Creutzfeldt-Jacobs (vaca loca)
- Ciclo Flash creado para el esterilizado rápido de instrumentos o dispositivos no
empaquetados.
9.3
Ciclos nocturnos
El autoclave, posee un dispositivo especial para el ahorro. Ese dispositivo permitirá
realizar ciclos de esterilizado durante la ausencia de los operadores. Al completarse el
ciclo, no abriéndose la tapa, el autoclave se estabilizará para entonces apagarse
automáticamente dejando encendido sólo el interruptor general (FIG.2-pos.19). Con la
llegada de los operarios, será suficiente oprimir uno de los botones para accionar
nuevamente el autoclave y leer el resultado del ciclo en el display. Además de lo anterior
la impresora habrá imprimido regularmente el informe correspondiente al ciclo realizado.
9.4
Ciclos de Test
Los ciclos de test disponibles son:
- Test de Hélice
- Test Bowie&Dick
- Test de vacío
22
ESPAÑOL
10
10.1
TEST DE CONTROL DEL AUTOCLAVE
Integradores químicos
Este tipo de test utiliza las propiedades de sustancias colorantes capaces de modificar su
aspecto al exponerlas al calor y a la presión durante los tiempos adecuados ya que
interactúan con todos los parámetros del ciclo (presión, temperatura, tiempo). Al superar
el test se demuestra que la estabilidad es correcta.
USO
Los integradores químicos (codico de repuesto 200/S e 215-S) deben
ser introducidos en el interior de la cámara de esterilización antes de
inciar el ciclo, incluso si está llena.
Los integradores pueden ser libres y posicionados directamente en la
bandejan o pueden ser introducidos dentro de los sobres de
esterilización junto con los materiales a esterilizar.
Se aconseja introducir los integradores en los sobres en el caso de
estructuras donde trabajen varios operadores para asegurarse de
que cada instrumento ha sido esterilizado.
Si la marca se queda en la zona
blanca, el test non ha sido
superado
RESPUESTA
Si la marca alcanza la zona de
SAFE, el test ha sido superado
FRECUENCIA
10.2
No existen intervalos de tiempo predefinidos, con excepción de las
zonas reguladas por leyes regionales específicas. En todo caso los
integradores químicos son el tipo de test más económico e inmediato
para comprobar la eficacia del autoclave. Se recomienda su uso en
todos los ciclos, o por lo menos una vez al día.
Indicador biológico
El objetivo de este test (codico de repuesto 262-S) es demostrar la eficacia del autoclave
con respecto a la destrucción de todos los microorganismos.
Consta de un preparado estandardizado de esporas con características biológicas
específicas, elevada resistencia al calor, y que se consideran un medio de control
extremadamente fiable y seguro (ATCC 7953). Para realizar este test se utilizan ampollas.
El test no es patógeno ni tóxico y es pirogénico
23
ESPAÑOL
USO
Las ampollas deberán ser introducidas en la cámara de esterilización,
incluso cuando esté llena.
Realizar el ciclo y, una vez finalizado, extraer la ampolla prestando
mucha atencíon ya que estará caliente y bajo presión.
Dejarla enfriar durante unos 10 minutos y, a continuación, activarla
utilizando la pinza especial proporcionada y manteniéndola siempre
en posición vertical. El indicador químico posicionado en la etiqueta
de la ampolla cambiará del azul al negro.
A continuación, introducir la ampolla en un encubador biológico a 57º
durante 48 horas junto con una ampolla de test no procesada pero
que ha sido activada con la pinza especial.
RESPUESTA
Después de 48 horas, extraer la ampolla procesada del incubador y
evaluar el resultado. Si la ampolla se ha vuelto amarillenta, significa
que el autoclave no ha superado el test y que existe produccion de
bacterias en su interior. Si la ampolla mantiene el color violeta
significa que no existen microorganismos y, por lo tanto, que el
autoclave ha superado correctamente el test. La ampolla de test
siempre cambiará su color a amarillo si no está procesada y su única
función es servir de elemento de comparación.
Una vez finalizado el test es posible eliminar las ampollas con los
residuos urbanos pero se recomienda someter la ampolla a un ciclo
adicional de esterilización.
FRECUENCIA
10.3
No existen intervalos de tiempo predefinidos, con excepción de las
zonas reguladas por leyes regionales específicas. En todo caso el
test biológico es el tipo de test más efectivo y para mantener un
control constante de la validez del autoclave se recomienda realizar
el test por lo menos una vez cada 90 días.
Test de Bowie&Dick
Es un test físico (codico de repuesto 268/S) que permite comprobar la capacidad de
penetración del vapor en los cuerpos porosos. Para realizar el test se utiliza un “paquete
de prueba” estandardizado y conforme a las normativas técnicas en vigor.
USO
24
El test deberà realizarse con la cámara vacía. Posicionar el test de
Bowie&Dick en la bandeja central del autoclave. Realizar el ciclo
adecuado indicado en la pantalla del autoclave. Al finalizar el ciclo,
extraer el paquete, abrirlo y comprobar el aspecto del papel con
indicador quimíco posicionado en su interior.
ESPAÑOL
La evaluación del resultado es
sencilla y rápida. Si el aspecto es
uniforme (tal y como se indica en la
imagen) significa que el test es
positivo, en caso contrario el
autoclave no puede esterilizar
correctamente los cuerpos porosos.
RESPUESTA
FRECUENCIA
10.4
No existen intervalos de tiempo predefinidos, con excepción de las
zonas reguladas por leyes regionales específicas. Para poder
mantener un control constante de la validez del autoclave, se
recomienda realizar el tes por lo menos una vez cada 30 días.
Test de hélice
Es un test de tipo físico (codico de repuesto 267-S) que permite comprobar la capacidad
de penetración del vapor en los cuerpos huecos. El test se realiza utilizando un sistema
estandardizado y conforme con las normas técnicas en vigor.
El test deberà realizarse con la cámara vacía.
USO
Introducir en la cápsula posicionada en el extremo del test la tira
especial y, a continuación, posicionar el test en la bandeja central del
autoclave.
Realizar el ciclo adecuado indicado en la pantalla del autoclave. Al
finalizar el ciclo, extraer el test y abrir la cápsula: comprobar el
aspecto de la tira indicadora.
25
ESPAÑOL
TEST ORIGINAL
RESPUESTA
TEST NEGATIVO
TEST POSITIVO
FRECUENCIA
10.5
No existen intervalos de tiempo predefinidos, con excepción de las
zonas reguladas por leyes regionales específicas. Para poder
mantener un control constante de la validez del autoclave, se
recomienda realizar el tes por lo menos una vez cada 30 días.
Test de vacìo
Es un prueba de estanqueidad de la cámara o prueba de la pérdida del vacío.
Esta comprobación tiene como objetivo comprobar que no existan infiltraciones de aire en
los elementos de estanqueidad de la cámara durante el ciclo (juntas, válvulas, etc...).
El ciclo deberà realizarse con la cámata vacía.
USO
Seleccionar el ciclo adecuado indicado en la pantalla del autoclave
para activarlo.
El autoclave realiza el ciclo de forma automática siguiendo
procedimientos técnicos bien definidos. A continuación, el resultado
final es imprimido por la impresora a través de una conexión
informática.
RESPUESTA
FRECUENCIA
El resultado es comunicado de forma inmediata a través de la
impresora y contiene todos los valores del ciclor realizado junto con
una evaluación final.
No existen intervalos de tiempo predefinidos, con excepción de las
zonas reguladas por leyes regionales específicas. Para poder
mantener un control constante de la validez del autoclave, se
recomienda realizar el test por lo menos una vez a la semana.
CONSEJAMOS EL UTILIZO DE LAS PRUEBAS
TECNO-GAZ S.P.A.
26
ESPAÑOL
11
LAS TABLAS Y LAS DIAGRAMAS
121°C
(envasado e no
envasado)
134°C
(envasado e no
envasado)
134°C
Flash
134°C
Prion
134°C
Helix / Bowie&Dick
24L
RANGO DE
PRESIÓN
DE TRABAJO
(bar relativos)
RANGO DE
TEMPERATURA
DE TRABAJO
(°C )
18L
18
17
13
1.04 ÷ 1.30
121 ÷ 125
5
6
12
2.02 ÷ 2.38
134 ÷ 138
4
2
4
2.02 ÷ 2.38
134 ÷ 138
20
17
2.02 ÷ 2.38
134 ÷ 138
2.02 ÷ 2.38
134 ÷ 138
3.5
4
-0.9
Vacuum
CICLOS
TEST
CICLOS
TIEMPO
SECADO
T5 (Minuti)
TIEMPO
EXP
T4
(Minuti)
CICLOS DE
FUNCIONAMIENTO
CUADRO DE ESTERILIZACIÓN
Los intervalos de precalentamiento y de fracción de vacío pueden variar en función de las
condiciones del autoclave de 25 – 30 minutos a los que se añdirán los intervalos de los
ciclos proporcionados en la tabla. Los ciclos de esterilización se desarrollan en tres fases
de vacío.
DIAGRAMA DEL CICLO
T1,T2,T3 =
precalentamiento /
fracción de vacío
T4
esterilización
T5
secado
FIG.24
27
ESPAÑOL
12
LECTURA DEL INFORME DE CICLO
Versión del software
Número de serie
Descripción del ciclo seleccionado
(con los valores principales de
configuración)
Horas – Minutos – Segundos
Presión de la cámara
Temperatura principal de la cámara
(T1)
Fase del ciclo
Fin del ciclo
(OK significa que el ciclo ha finalizado
correctamente)
Número del ciclo
Operador (espacio para la firma)
28
ESPAÑOL
13
MENSAJES DE ERROR Y ALARMAS
Los mensajes de error estarán destacados por un código alfanumérico compuesto de una
letra y de 3 cifras.
El sufijo “E” está realcionado con errores del operador que el mismo operador puede
restablecer.
El sufijo “A” está relacionado con alarmas, anomalías del autoclave; si el problema
persiste tras la reparación debe contactarse con la asistencia telefónica.
Al visualizarse un mensaje de alarma (sufijo "A") el ciclo debe
considerarse SIN éxito: deben repetirse todas las operaciones de
preparación y de esterilización.
Para restablecer alarmas y errores deben mantenerse pulsados simultáneamente los
botones que a continuación destacamos hasta la intervención de la señal acústica:
B-DOOR + B-PUMP + B-SELECT
ERRORE
CAUSA
RIMEDIO
E 200
Nivel Bajo de agua
Llenar el depósito
E 250
Nivel bajo de agua limpia con llenado por
desmineralizador
Llenar el depósito
E 300
Nivel máximo de agua usada
Vaciar el depósito
E 401
Tapa abierta
Cerrar con atención la tapa
E 451
Tapa no bloqueada
Abrir la tapa y cerrarla
E 461
Alta temperatura en cámara para Vacuum
test (mas que 40°C)
Apagar y aguardar 10 min.
A 250
Inicio ciclo sin agua
Controlar el dispositivo
desmineraliz.
A 001
Ciclo interrumpido
Reiniciar y accionar el ciclo
A 011
Anomalía pc display
Apagar y volver a accionar el
autoclave
A 101
Vacío no alcanzado en 10 min.
Reiniciar y repetir el ciclo
A 111
Vacío no mantenido en primera fase VT
Reiniciar y repetir el ciclo
A 121
Vacío no mantenido en segunda fase VT
Reiniciar y repetir el ciclo
29
ESPAÑOL
A 401
Tapa abierta durante el ciclo o problema de
Controlar sistema de cierre
cierre
A 501
Presión no mantenida
Reiniciar y repetir el ciclo
A 551
Anomalía 1° sensor temperatura
Reiniciar y repetir el ciclo
A 552
Anomalía 2° sensor temperatura
Reiniciar y repetir el ciclo
A 641
Superpresión
Reiniciar y repetir el ciclo
A 651
Sobretemperatura
Reiniciar y repetir el ciclo
A 661
Anomalías comparación temperaturas
Reiniciar y repetir el ciclo
A 701
Presión de trabajo no alcanzada
Reiniciar y repetir el ciclo
A 751
Baja temperatura
Reiniciar y repetir el ciclo
A 761
Anomalía en sensor de presió
Reiniciar y repetir el ciclo
A 801
Anomalía descarga presión
Reiniciar y repetir el ciclo
A 901
Tensión principal muy baja
Pérdida momentánea de alimentación
eléctrica
Consultar la empresa de
suministro de Energía eléctrica
REEMPLAZAR LAS RESINAS Resinas Seguir las instrucciones en elapartado 17.3
de IONIC SYSTEM agotado
del manual
REEMPLAZAR EL FILTRO Filtro de
OSMOSI SYSTEM agotado
Seguir las instrucciones en elapartado 17.3
del manual
D
Tapa impresora no
cerrada
Abrir y cerrar la tapa de la impresora
P
Falta papel en impresora
Introducir el papel
Lo
Baja tensión en la línea
de alimentación
Intentar conectar la máquina a la red eléctrica en otro
punto; si el problema persiste contactar con electricista
30
ESPAÑOL
14
PROGRAMACIÓN DE PARÁMETROS DISPLAY
Consulta la FIG.2.
14.1
Selección Idioma
Pulsar B-START y simultáneamente el interruptor (MAINS).
Mantener pulsado B-START hasta que aparezca en el display el nombre del idioma
requerido.
Pulsando B-SELECT se modificará el idiomai [avance].
Pulsando B-PUMP se modificará el idoma [atrás].
14.2
Selección Hora
Pulsando B-START aparece la hora.
Pulsando B-SELECT modificase la hora [hacia adelante].
Pulsando B-PUMP modificase la hora [hacia atrás].
14.3
Selección Minutos
Pulsando B-DOOR aparecen los minutos.
Pulsando B-SELECT modificanse los minutos [hacia adelante].
Pulsando B-PUMP modificanse los minutos [hacia atrás].
14.4
Selección Segundos
Pulsando B-DOOR aparecen los segundos. No modificables.
14.5
Selección Fecha
Pulsando B-START aparecen los días (destello).
Pulsando B-SELECT
Pulsando B-PUMP
14.6
modificanse los días [hacia adelante].
modificanse los días [hacia atrás].
Selección Mes
Pulsando B-DOOR aparecen los meses (destello).
Pulsando B-SELECT
Pulsando B-PUMP
modificanse los meses [hacia adelante].
modificanse los meses [hacia atrás].
31
ESPAÑOL
14.7
Selección Año
Pulsando B-DOOR aparece el año.
Pulsando B-SELECT modificase el año [hacia adelante].
Pulsando B-PUMP modificase el año [hacia atrás].
14.8
Visualización Presión Atmosférica
Pulsando B-START aparece la presión atmosférica almacenada.
NO MODIFICABLE !
14.9
Selección Carga Agua
Pulsando B-START aparece la programación del llenado del agua:
• CARGA CON BOMBA
• IONIC SYSTEM
• OSMOSI SYSTEM
Pulsando B-SELECT se modificase la impostaciòn.
Pulsar B-SELECT mantenerlo oprimido hasta oirse un “BEEP”.
Pulsar B-START y mantenerlo oprimido hasta oirse el “BEEP”.
Los datos programados será almacenados. Apagar el autoclave por el interruptor MAINS.
32
ESPAÑOL
15
SOLUCIÓN DE PROBLEMAS OPERATIVOS
En muchos casos, algunas de las alarmas o de los errores estarán originados por la
desatención o el desconocimiento de algunos apsectos técnicos y operativos. A
continuación relataremos algunos de los casos de anomalías con respectivas soluciones.
15.1
El autoclave no efectua secado correcto
• sustituir el filtro bacteriológico con uno original
• no fueron utilizadas bandejas originales sino bandejas distintas sin perforaciones o con
perforación distinta. utilizar exclusivamente bandejas originales posiblemente en
aluminio. solicitar bandejas adicionales originales.
• los instrumentos no fueron acomodados correctamente. Siga la indicación de par.8.4
15.2
La cámara del autoclave cambia a color blanco
• cambiar inmediatamente el tipo de agua utilizada, utilizar agua desmineralizada o
destilada conforme especificado en los capítulos anteriores y limpiar la cámara.
• el color blanco puede ser consecuencia de la evaporación de materiales orgánicos
presentes en los instrumentos. someter los instrumentos a una acción de detersión más
idónea y profunda.
• Comprobar al grupo de demineralización del agua instalado
15.3
La cámara del autoclave presenta manchas verdes-azuladas
• non vi e’ stato un corretto risciacquo degli strumenti dopo la fase di detersione,
sciacquare con maggior attenzione e scrupolo gli strumenti. se le macchie sono evidenti
richiedere assistenza tecnica telefonica.
15.4
El ciclo de esterilización se interrumpe
• controlar si el autoclave está conectada a la red eléctrica con extensiones, reducciones,
adaptadores y de ser así eliminar tales accesorios y conectar el autoclave directamente
a la toma de la red eléctrica.
15.5
El autoclave no recibe los mandos
• el autoclave está respondiendo a la alineación bárica automática, aguardar la doble
señal acústica tras la apertura de la tapa para entonces parametrizar las funciones.
• el depósito de agua pura está vacío, el led de nivel minimo está encendido abastecer
con agua pura.
• el depósito de agua usada está lleno, el led de nivel máximo está encendido, drenar el
agua usada.
33
ESPAÑOL
15.6
La impresora del autoclave no imprime
• la tapa del portacilindro no está cerrada correctamente, en el display aparece la letra
“D”. Abrir y cerrar la tapa de la impresora comprobando su correcto cierre.
• falta el papel, aparece en el display la letra “P”. Abrir la tapa e introducir el rollo de
papel termográfico.
• el cilindro de papel está montado con la cara térmica hacia dentro, abrir la tapa y
colocar el cilindro en la posición correcta
15.7
Manchas en los instrumentos
• los instrumentos pasan al color amarillo, resíduos de líquido químico que con el calor
se fijan en los instrumentos. No fueron adecuadamente aclarados.
• la cámara de esterilización presenta manchas amarillas, ha sido puesto en la cámara
un instrumental con líquido químico que cayendo pudo fijarse gracias al calor. No fue
efectuado el aclarado adecuado.
• los instrumentos muestran manchas blancas, el aclarado fue realizado con agua
demasiadamente calcárea y los instrumentos no se secaron. Para el último aclarado es
aconsejable utilizar agua desmineralizada además de secar perfectamente los
instrumentos.
• los instrumentos se oscurecieron, y ello se debe a la fuerte presencia de carbono en su
interior.
34
ESPAÑOL
16
16.1
DESMINERALIZADOR
Impostaciòn del sistema de demineralizaciòn
El autoclave está apta también para el abastecimiento de agua desmineralizada atraves
de los sistemas de cargo automatico SISTEMA IONICO o SISTEMA OSMOSIS.
(accesorios opcionales).
Estos sistemas son accesorios de los autoclaves TECNO-GAZ S.p.A. (no en dotación).
Ese producto se utiliza para eliminar las sales minerales del agua procedente de la red
hídrica para alimentar el autoclave.
Es necesario programar el autoclave para cargar el agua con los sistemas IONICO o
OSMOSIS, segun las indicaciones puestas a bajo:
Accionar el autoclave pulsando el interruptor general (FIG.2–POS.19) y mantener pulsado
B-START hasta que aparezca el mensaje que indique el idioma elegido.
Pulsar y dejar B-START para seleccionar la pantalla sucesiva hasta que aparezca:
• CARGA CON BOMBA
• IONIC SYSTEM
• OSMOSI SYSTEM
Para cambiar la programación usar B-SELECT.
Para almacenar el programa debe pulsarse B-DOOR y mantenerlo oprimido hasta oirse
la señal acústica “BEEP”
Para salir debe pulsarse B-START y mantenerlo oprimido hasta oirse la señal acústica
“BEEP”
Los datos han sido memorizados.
ATENCIÒN:
Il numero puesto a bajo de la tipologia de cargamiento esplica cuantos ciclos se pueden
mas o menos hacer con el filtre o con las resinas.
Se conseja de hacer el reset de este numero cuando se cambia el filtro o las resinas
despues de haber guardado la tipologia de cargamiento en uso.
Para hacer el reset al maximo valor de esta informaciòn es necesario mantener pulsada
B-PUMP hasta el señal acústica en la panoramica de espera empievo ciclo.
35
ESPAÑOL
16.2
Conexion de los sistema de desmineralization
Aqui ponemos el particolar de la conexion del autoclave con el conduco de alimentaciòn
del agua y de la conexion al enchufe:
•
•
•
•
•
•
•
•
•
•
•
Apagar el autoclave si está encendida (FIG.2–pos.19)
Cerrar el grifo aguas arriba del sistema de desmineralización
Instalar el desmineralizador conforme indicado en el manual del desmineralizador;
Envolver el roscado macho del racorportatubo con teflon u otro componente que
garantice la estanqueidad contra el agua;
Enroscar el racor-portatubo en el filete hembra del desagüe del agua Limpia
(FIG.1–pos.2)
Introducir el tubo que sale de desmineralizador en el racor-portatubo recién acoplado al
autoclave;
Introducir el enchufe del desmineralizador en la toma (FIG.1–pos.7) por detrás del
autoclave;
Abrir el grifo aguas arriba del sistema de desmineralización;
Controlar si hay fugas de agua;
Accionar el autoclave;
Efectuar uno o más ciclos de esterilización para controlar el funcionamiento de la
conexión realizada y controlar principalmente las fugas
Al final del día cerrar siempre el grifo aguas arriba del sistema de
desmineralización para evitar inundaciones
Conectar los sistemas de desmineralizaciòn solo con autoclaves TECNOGAZ predispuestas.
Si al iniciar el ciclo el autoclave entra en E250 signifca que fue accionado el
ciclo con un nivel de agua limpia bajo y por lo tanto deberá efectuarse el
llenado del depósito con agua limpia.
NOTA:
para la conexión los sistemas de desmineralizaciòn a autoclaves TECNO-GAZ,
consulte lo indicado en el manual de desmineralizador.
36
ESPAÑOL
FIG.32
16.3
Cambio cartucho resina – filtre
El autoclave pone en el display el esaurimento de las resinas o del filtre con el mesate
“SUBSTITUCIÒN RESINAS” (codico de repuesto 242-S) o “SUBSTITUCIÒN FILTRES”
(codico de repuesto 248-S-2).
Sustituir el cartucho de resina conforme lo indicado en el manual de desmineralizador;
Terminada la sustitución debe pulsarse el botón B-PUMP hasta la intervención de la
señal acústica para entoces reiniciar el autoclave.
37
ESPAÑOL
17
PROCEDIMIENTOS ESTANDAR DE ESTERILIZACIÓN
Esterilizar significa adoptar protocolos operativos específicos y aplicar un sistema.
El autoclave es el eslabón final de ese sistema.
El que no efectúe todas las fases del sistema de esterilización perjudicará el resultado
final. TECNO-GAZ S.p.A. produce todos los dispositivos para todas las fases del ciclo de
esterilización.
Se requiere por lo tanto:
1 - Predisponer un protocolo operativo por escrito preciso y distribuirlo entre todos los
operadores y encargados de la esterilización.
TECNO GAZ SPA puede suministrar tales protocolos personalizados.
2 - Seguir escrupulosa y atentamente todas las fases del ciclo de esterilización:
DESINFECCIÓN fase compulsoria que es obligatorio para salvaguardar
la seguridad de los operadores que debe actuarse por sumersión en líquidos
químicos o por termodesinfección;
DETERSIÓN la fase más que garantiza la eliminación de todos los tipos de
resíduos químicos y orgánicos. El instrumento más idóneo son las cubetas de
ultrasonidos;
SECADO fase indispensable, que evita corrosión de los instrumentos e
interferencias en el ciclo de esterilización;
ENVASADO fase indispensable para mantener estériles los instrumentos
tratados.
ESTERILIZACIÓN fase final esterilización por vapor.
38
ESPAÑOL
18
INSTRUCCION PARA EXPEDICIÓN DEL AUTOCLAVE
Cuando haya la necesidad de expedirse el autoclave para reparaciones, control, rearme,
revisión, validación, sigan obligatoriamente las instrucciones a continuación:
•
Utilizar el embalaje original, si ese embalaje ya no se encuentra en las manos del
cliente, deberá utilizarse un embalaje adecuado. La mercancía viaja por riesgo y
cargo del remitente.
•
Enviar sólo el autoclave (no incluir ninguno de los componentes presentes en el kit
de accesorios).
•
Limpiar esmeradamente la cámara de esterilización y el autoclave completa antes
de expedirlas La llegada de un autoclave sucio o con resíduos determinará su
rechazo sin reparaciones ó una acción de limpieza y desinfección.
•
Siempre drenar el depósito de agua limpia por el racor que se encuentra por de
trás del autoclave (FIG.2–pos.2)
•
Siempre debe drenarse el depósito de agua usada por el racor que se encuentra
por detrás del autoclave (FIG.2–pos.5)
•
Indicar por escrito e incluir en el embalaje, un documento en que se indique
precisamente la anomalía detectada o el servicio que se requiere.
•
Expedir siempre puerto franco de lo contrario serán adeudados los gastos de
transporte.
ATENCIÓN:
Todos los embalajes no originales que nos envíen serán eliminados.
Los autoclaves les serán devueltos con embalajes nuevos y originales ello para garantizar
la máxima protección del autoclave durante el transporte.
El embalaje les será adeudado.
39
ESPAÑOL
19
PROCEDIMIENTOS PARA EL SERVICIO Y ASISTENCIA
El departamento encargado de las reparaciones constituye una sección autónoma de
MEDILINE ITALIA SRL (TECNO-GAZ GROUP) y su organización se basa en una
estructura compleja y profesional, en conformidad con nuestros procedimientos indicados
en el “Manual de Calidad”.
•
Llegando a la MEDILINE ITALIA SRL, el autoclave pasa directamente a la oficina
de reparaciones que aplicará un código al aparato creando una ficha de asistencia
técnica.
•
Dentro de un plazo máximo de 48 horas, el aparato será sometido a inspección por
los técnicos y se redactará el diagnóstico técnico con presupuesto de gastos por el
servicio requerido. En caso de no devolución del autoclave conforme lo requerido
en el capítulo 19, los plazos indicados pueden aumentar considerablemente.
•
El presupuesto de gastos será enviado al cliente distribuidor que lo transmitirá al
cliente final para información y suscripción.
•
Después que MEDILINE ITALIA SRL habrá recibido el presupuesto suscrito por
aceptación el autoclave será sometido a reparación y devuelto en los plazos
conforme lo indicado en el módulo del presumpuesto.
•
La mercancía será enviada con embalaje original por medio del transportador del
clinete.
IMPORTANTE:
Si el problema persiste aunque se haya aplicado la solución correspondiente diríjanse a la
asistencia telefónica de: MEDILINE ITALIA srl +39 0522 – 94 29 97.
40
ESPAÑOL
20
GARANTÍA
1) DURACIÓN: el producto está amparado por un período de garantía de 12 (doce)
meses o 1200 ciclos.
2) EXCLUSIÓN: no están incluidos en la garantía:
a)
las averías causadas por la carencia de mantenimiento ordinario debido a la
negligencia del usuario o a un uso inadecuado del producto;
b)
controles periódicos y mantenimiento;
c)
reparaciones o sustitución de las piezas sujetas a desgaste, frágiles o de
duración imprevisible, salvo que se haya probado, en el momento de su
entrega, que presentan defectos;
d)
averías relativas a mano de obra; viaje del personal técnico y transporte;
e)
averías o daños producidos por un uso inadecuado o por errores en la
utilización;
f)
averías o daños producidos por impurezas en los sistemas de alimentación del
agua y del aire, accidentes químicos o eléctricos extraordinarios;
g)
averías o daños causados por la utilización de detergentes, substancias o
procesos de esterilización que no estén indicados expresamente en el manual
de uso y mantenimiento;
h)
cambio de color natural de las piezas de plástico.
3) INFORME DE INSTALACIÓN Y ENSAYO: una condición esencial para acceder a la
garantía es la devolución del informe de instalación y ensayo, correctamente
cumplimentado y firmado por el Usuario y por el Técnico Instalador. El informe deberá
devolverse antes de transcurridos 15 días a partir de la instalación, pudiéndose incurrir en
la caducidad de la garantía.
4) LÍMITES: la garantía da derecho a la sustitución o reparación gratis de los
componentes defectuosos. Queda excluido el derecho a la sustitución de todo el aparato.
En lo referente a los componentes aplicados o integrados con productos con marca
TECNO-GAZ, o de terceros, y acompañados por su certificado de garantía, son válidas
las condiciones, límites y exclusiones indicadas en tales certificados.
Las reparaciones llevadas a cabo durante el período de garantía no modifican la fecha de
vencimiento de la garantía.
5) RECLAMACIONES: en el caso de reclamación del Comprador sobre la aplicación de la
Garantía, sobre la calidad o sobre las condiciones del equipo entregado, el Comprador no
tendrá la facultad de suspender ni retrasar los pagos, por ningún motivo.
6) CONTROVERSIAS: en caso de controversia sobre la aplicación e interpretación del
presente Certificado de Garantía, será competente el Tribunal de PARMA (Italia),
dondequiera que haya sido estipulado el contrato de compra del aparato.
7) EXCEPCIONES: cualquiera excepción especial, concedida con relación a las presentes
normas de garantía, no implica el reconocimiento de ningún derecho al Comprador, y se
considerará concedida exclusivamente con relación al caso específico.
8) OTROS: Todo aquello que no esté expresamente previsto en este Certificado de
Garantía estarán regido por las normas contenidas en el Código Civil y en las leyes en
materia de la República Italiana.
41
ESPAÑOL
La garantía caduca si:
a) el equipo presenta averías provocadas por caídas, exposición al fuego, vertido de
líquidos, rayos, calamidades naturales, eventos atmosféricos, u otras causas no
imputables a defectos de fabricación;
b) la instalación no es conforme a las instrucciones de TECNO-GAZ S.p.A. y si fue llevada
a cabo por personal no autorizado;
c) el equipo es reparado, modificado o, en algún modo, alterado por el Comprador o por
terceros no autorizados;
d) al pedir el servicio bajo garantía faltara el número de matrícula del aparato, o si
estuviera cancelado, falsificado, etc.;
e) no se devuelve el informe de instalación y ensayo cumplimentado y firmado antes de
transcurridos 15 días a partir de la instalación;
f) el Comprador suspende o retrasa, por cualquier motivo, el pago de cualquier suma
debida con relación a la compra o al mantenimiento del equipo;
g) no se respeta el mantenimiento periódico programado u otra disposición prevista en el
manual de uso y mantenimiento
42
ESPAÑOL
A
MANTENIMIENTO
El correcto mantenimiento del autoclave garantizará su buen funcionamiento y el ahorro
seguro de tiempo y costes debidos a asistencia y mantenimiento. Las seguente
operaciones son obligatorias por los operadores
Limpieza de la cámara
20 ciclos o una vez a la semana
Limpia periódicamente la cámara y el filtro eliminando posibles depósitos o detritos para
evitar así, introducir en el circuito de desagüe, materiales que puedan crear
obstrucciones. Para realizar correctamente la limpieza utiliza únicamente agua y la
esponja suministrada (parte no abrasivo - FIG.13).
Ejecutar solo a camara fria para evitar quemaduras - Nunca utilizar solventes,
detergentes, soluciones químicas desincrustantes u otros productos similares.
Limpieza filtre camara
20 ciclos o una vez a la semana
Tirarar el filtre (codico de repuesto DXBA091) hacia arriba, sin dañarlo, lavar con agua
desmineralizada y secar con un trapo limpio y seco.Reposicionar en su sitio el filtre
cuidando que sobresale de 15 mm
Limpieza bandejas y portabandejas
20 ciclos o una vez a la semana
Limpiar con esponja no abrasiva en dotaciòn empapada con agua desmineralizada
Sustitución del filtro bacteriológico
200 ciclos o cuando llega a ser
oscura
Sustituir el filtro bacteriológico (FIG.2–pos.10) Gira el filtro a izquierdas para
desenroscarlo y a derechas para enroscarlo. Utilizar exclusivamente filtros originales
(codico de repuesto DAVA101)
Limpieza de la junta de la tapa
20 ciclos o una vez a la semana
Deberán eliminarse periódicamente los resíduos que se acumulan en la circunferencia
de la junta (codico de repuesto DAVA101) usando para ello agua y la esponja en
dotación (parte no abrasiva), ó un paño húmedo.
A-1
ESPAÑOL
Regulaciòn puerta
Cada 2 meses
Para regular la presión de cierre de la puerta debe actuarse con el regulador de la
puerta utilizando la llave de doble función (FIG.10) en dotación. Girar a izquierdas
(pos.A) para aumentar la presión de cierre. Girar a derechas (pos.B) para reducir la
presión de cierre.
A
Sustitución del Papel de la Impresora
B
-
Rodillo de papel termográfico con 57 mm de ancho. El papel termográfico debe ser
utilizado solo por un lado. Levantar la tapa de la impresora empujando el botón introducir
el rodillo de papel atentos al sentido de rotación, tirar del papel hasta que salga y cerrar la
tapa; la impresora estará así lista para imprimir.
A-2
ESPAÑOL
Las siguente operaciones son consejadas
realizada por tecnicos autorizados.
de parte del productor y tienen que ser
Substitucion filtre y carga agua
Cada año / cada 100 ciclos *
Control filtre agua entrada fluximetro
Cada año / cada 100 ciclos *
Limpieza filtre Y de latòn a bajo del radiador
Cada año / cada 100 ciclos *
Limpieza elecrtovalvula carga agua E4
Cada año / cada 100 ciclos *
Substituciòn valvula bomba vacìo
Cada año / cada 100 ciclos *
Limpieza radiador
Cada año / cada 100 ciclos *
Limpieza abanicos
Cada año / cada 100 ciclos *
Control integridad circuito Pneumatico
Cada año / cada 100 ciclos *
Control integridad circuito electrico
Cada año / cada 100 ciclos *
Control isolante termico camara de esterilizaciòn
Cada año / cada 100 ciclos *
Comprobaciòn de los tornillos de adjuste
Cada año / cada 100 ciclos *
Control sondas de temperadura
Cada año / cada 100 ciclos *
Control valvula seguridad
Siga las instrucciones del
fabricante (ver instrucciones)
Engrasar la parte movil del tiradòr
Cada año / cada 100 ciclos *
Control sondas de nivel
Cada año / cada 100 ciclos *
Control tanque
Cada año / cada 100 ciclos *
* vale la codiciòn que se alcanza como primera
Algunas disposiciones nacionales obligan la validacciòn anual de la maquina que controla
los parametros de esterilizaciòn. La validacciòn periodica tiene que ser hecha por tecnicos
autorizados de parete del productor con instrumentos adecuados. En el caso no sea
obbligatoria, la validacciòn es una operaciòn volontaria con el fin de certificar que el
autoclave funciona perfectamente y esteriliza las cargas de instrumentos por los cuales ha
sido proyectada.
A-3
ESPAÑOL
NOTAS
A-4
DEUTCH
Lieber Kunde,
Vielen Dank für die Wahl unseres Autoklaven. Wir werden uns mit großer Aufmerksamkeit
und einem sicherlich Ihren Ansprüchen entsprechenden Kundendienst dieses Vertrauens
würdig erweisen.
Der von Ihnen gewählte Autoklav ist der maximale Ausdruck in Sachen angewandter
Technologie. Er kann im zahnmedizinischen, medizinischen und ästhetischen Bereich
sowie generell in allen Sektoren verwendet werden, in denen die Notwendigkeit einer
sicheren Sterilisierung aller Geräte und feste, hohle, poröse, lose und verpackte
Materialien, besteht.
Wir bitten Sie die Bedienungsanleitung vor Gebrauch des Autoklaven aufmerksam zu
lesen und diese an einem für alle - mit der Sterilisierung beauftragten - Bediener
zugänglichen Ort aufzubewahren.
Der Autoklav darf nur für den vom Hersteller vorgesehenen Einsatz verwendet werden.
Für Installationsarbeiten, die Instandhaltung und den Service wenden Sie sich
ausschließlich an von der Firma TECNO-GAZ autorisierte Techniker. Bitte
verwenden und bestellen Sie nur Original-Ersatzteile der TECNO-GAZ
1
DEUTCH
INHALT
1. DAS HANDBUCH KONSULTIEREN
1.1
GLOSSAR
1.2
BESCHREIBUNG DER VORDER-/RÜCKSEITE
2. SICHERHEIT
2.1
SICHERHEITSKENNZEICHNUNG
2.2
SCHUTZEINRICHTUNGEN
2.3
ANMERKUNGEN ZUR SICHERHEIT
2.4
ENTSORGUNG
3.
VERWENDUNG UND GEBRAUCHSZIEL DES AUTOKLAVS
4. TECHNISCHE DATEN
4.1
MECHANISCHE DATEN
4.2
ELEKTRISCHE DATEN
4.3
KAMMER
4.4
REINWASSERTANK
4.5
SCHMUTZWASSERTANK
4.6
BAKTERIOLOGISCHER FILTER
5.
ZUBEHÖR
6.
DEN AUTOKLAV AUSPACKEN
7.
INSTALLATION
8. GEBRAUCHSANWEISUNGEN
8.1
DEN AUTOKLAV EINSCHALTEN
8.2
DEN REINWASSERTANK FÜLLEN
8.3
EIGENSCHAFTEN DES ZU VERWENDENDEN WASSER
8.4
DAS MATERIAL IN DEN AUTOKLAVEN GEBEN
8.5
DEN STERILISIERZYKLUS STARTEN
8.6
ZYKLUSENDE
8.7
DAS STERILISIERTE MATERIAL HERAUSNEHMEN
8.8
DAS SCHMUTZWASSER ABLASSEN
8.9
EINEN STERILISIERZYKLUS UNTERBRECHEN
9. STERILISIERZYKLUS
9.1
BESCHREIBUNG DER ZYKLEN
9.2
ARBEITSZYKLEN
9.3
NACHTZYKLEN
9.4
TESTZYKLEN
10. KONTROLLTESTS FÜR DEN AUTOKLAV
10.1
CHEMISCHE ZUSÄTZE
10.2
BIOLOGISCHER INDIKATOR
10.3
BOWIE-DICK-TEST
2
DEUTCH
10.4
10.5
HELIXTEST
VAKUUMTEST
11. TABELLEN UND DIAGRAMME
12. LESEN DES ZYKLUSBERICHTS
13. FEHLERMELDUNGEN UND ALARME
14. PROGRAMMIERUNG UND DISPLAY-PARAMETER
14.1 DIE SPRACHE EINSTELLEN
14.2 DIE UHRZEIT EINSTELLEN - STUNDE
14.3 DIE UHRZEIT EINSTELLEN - MINUTEN
14.4 DIE UHRZEIT EINSTELLEN - SEKUNDEN
14.5 DAS DATUM EINSTELLEN - TAG
14.6 DAS DATUM EINSTELEN - MONAT
14.7 DAS DATUM EINSTELLEN - JAHR
14.8 LUFTDRUCKANZEIGE
14.9 ANWAHL WASSERZULAUF
15. BEHEBUNG AUFTRETENDER PROBLEME
15.1 DER AUTOKLAV TROCKNET NICHT ORDNUNGSGEMÄSS
15.2 DIE KAMMER DES AUTOKLAVEN WIRD WEISS
15.3 DIE KAMMER DES AUTOKLAVEN HAT GRÜNE FLECKEN
15.4 DER STERILISIERZYKLUS WIRD ABGEBROCHEN
15.5 DER AUTOKLAV EMPFÄNGT KEINE ARBEITSANWEISUNGEN
15.6 DER DRUCKER DES AUTOKLAVEN DRUCKT NICHT
15.7 FLECKEN AUF DEN INSTRUMENTEN
16. WASSERAUFBEREITUNG
16.1 EINSTELLUNG VON WASSERAUFBEREITUNG SYSTEME
16.2 VERBINDUNG VON WASSERAUFBEREITUNGSYSTEME
16.3 WECHSELN VON PATRONE UND FILTER
17. STANDARDVORGÄNGE STERILISIERUNG
18. ANWEISUNGEN FÜR DEN VERSAND DES AUTOKLAVEN
19. SERVICE UND KUNDENDIENST
20. GARANTIE
A. INSTANDHALTUNG
3
DEUTCH
DAS HANDUBUCH KONSULTIEREN
1
1.1
Glossar
VAKUUM
Druck, dessen Wert unter dem Luftdruck liegt
KAMMER
Zylindrischer Körper aus Stahl in den die zu sterilisierenden Geräte
gegeben werden
FILTER
TABLETT
TABLETTHALTER
Vorrichtung, die Partikel zurückhält
Gestell, auf das das zu sterilisierende Material gelegt wird
Halterung für die Tabletts in der Sterilisationskammer
ENTSALZER
Vorrichtung für die Eliminierung der Mineralsalze im Netzwasser
HELIX TEST
Test zur Überprüfung der Penetrationskapazität von Dampf in einem
Hohlkörper
BOWIE - DICK TEST
Test zur Überprüfung der Penetrationskapazität von Dampf in einem
porösen Körper
VAKUUMTEST
BAKTERIOLOGISCHER
FILTER
ÜBERDRUCK
ÜBERTEMPERATUR
HARZE
1.2
Test zur Überprüfung des Vakuumbeständigkeitsgrades in der Kammer
Vorrichtung, die in der Luft schwebende Partikel mit einer Größe über
0,3 µm zurückhält.
Druck, dessen Wert über dem Standarddruck des gewünschten
Betriebsablaufs liegt
Temperatur, deren Wert über dem Standardwert des gewünschten
Betriebsablaufs liegt
Austauschbarer Bestandteil (Kartusche) des Entsalzers, der alle
anorganischen Salze eliminiert
Beschreibung der vorder-/rückseite
Abstandsfuß
Überlaufschutz Reinwasser
Hahn für Reinwasserablauf – Anschluss Entsalzer
Schutzventil
Überlaufschutz Schmutzwasser– Kondenswasserablauf
Hahn für Schmutzwasserablauf
Hauptstromversorgung
Buchse für Stromversorgung NUR ENTSALZER
12A-Sicherung
FIG.1
12A-Sicherung
Fusibile 500mA
4
DEUTCH
FIG.2
5
DEUTCH
Bakteriologischer Filter
Manueller Wassereinlauf
Wassereinlauf
Schmutzwasserablauf
Display
Drucker
Anschluss RS232
Stecker B-TEST Inkubator
Hauptschalter
121°C Zyklus
(geeignet für verpacktes und unverpacktes Material)
134°C Zyklus
(geeignet für verpacktes und unverpacktes Material)
134°-Zyklus Flash
Prion-Zyklus
Vacuum Test
Helix – B&D Test
Wassermindeststand
Wasserhöchststand
Höchststand Brauchwasser
B-START
START-STOP-Taste
B-PUMP
Uploaded Taste Entionisiertes Wasser
B-SELECT
B-DOOR
6
Zyklusanwahl Taste
Türverriegelung öffnen Taste
DEUTCH
2
2.1
SICHERHEIT
Sicherheitskennzeichnung
GEFÄHRLICHE SPANNUNG
VOR ABNAHME DES DECKELS
STROMZUFUHR
UNTERBRECHEN
HOHE TEMPERATUR
ERDUNG
2.2
Schutzeinrichtungen
Folgenden Schutzeinrichtungen sind am Gerät installiert:
-) Schutzventil geeicht auf 2.4 bar 0/+10%
-) Elektromagnetische Sperre, verhindert das Öffnen der Tür während dem Zyklus
-) Sicherheitsthermostat
2.3
Anmerkungen zur Sicherheit
• Als Hersteller haftet die Firma Mediline Italia Srl gemäß der geltenden Richtlinie für das
auf den Markt gebrachte Produkt. Die Verantwortbarkeit verfällt unverzüglich wenn nicht
qualifiziertes Personal an der Maschine oder ihren Teilen Eingriffe vornimmt oder nicht
originale Ersatzteile verwendet werden.
• Der Raum in dem der Autoklav installiert wird darf keine potentielle Explosionsund/oder Brandgefahr aufweisen.
• Der Autoklav darf nur in einem Raum installiert werden der den gesetzlichen
Vorschriften entspricht.
• Der Stecker des Inkubator (FIG.2-pos.18), darf nur für den Anschluss des B-TEST
verwendet werden.
Da es sich um einen vom Hauptschalter unabhängigen Anschluss zu 230V
Wechselstrom handelt, den Verbinder nicht mit nassen oder feuchten Händen
anfassen und ihn vor eventuellen Wasserspritzern schützen.
7
DEUTCH
2.4
Entsorgung
Dieses Produkt unterliegt der Richtlinie 2002/96/EG über Elektro- und ElektronikAltgeräte (WEEE) des Europäischen Parlaments bzw. des EU-Ministerrats.
Dieses Gerät ist in Ländern, die diese Richtlinie bereits als Gesetz verabschiedet
haben, nicht vor dem 13. August 2005 im Handel erhältlich und sollte nicht im
Rahmen des normalen Hausmülls entsorgt werden. Nutzen Sie bitte zur
Entsorgung dieses Geräts die entsprechenden Rücknamesysteme in Ihrer
Gemeinde und beachten Sie auch alle weiteren Vorschriften in diesem Kontext.
3
ZWECKBESTIMMUNG DES AUTOKLAVEN
Der Autoklav sterilisiert entsprechend der Norm EN13060 drei Materialtypen, im
Einzelnen:
EISENMATERIAL ODER Instrumente ohne Hohlkörper und ohne Behinderungen für das
FESTKÖRPER
Dampfeindringen
PORÖSE KÖRPER
Einfache Materialien oder Komposite, welche die Flüssigkeiten
aufnehmen können (Stoffe, Kittel, Verbandstoffe, usw…)
KÖRPER MIT
HOHLKÖRPER
Materialien oder Geräte mit Hohlkörpern, Behinderungen,
usw…Diese werden in zwei Kategorien unterteilt und indikativ
nach Länge und Durchmesser klassifiziert
KATEGORIE B Kanülen, Leitungen oder Geräte mit großen
Durchgängen
KATEGORIE A Turbinen, Gerätegriffe und Geräte mit
Blindöffnungen oder kleinen Öffnungen
Der Autoklav kann im Zahn- Arzt- und Kosmetiksbereich und im Allgemeinem in allen
Instruments- und Materialsterilizationsbereichen eingesetzt werden:
EUROPA B EVO
EUROPA B EVO 24
Eisenmaterial oder festkörper
max kg. 5
max kg. 7
Körper mit hohlkörper
max kg. 5
max kg. 7
max kg. 1,5
max kg. 2
Poröse Körper
* Gültig nur für Europäische Ländern
Der Autoklav darf ausschließlich für die Sterilisierung von mit dem System
der Dampfsterilisierung kompatiblen Geräten und Materialien verwendet
werden. Vergewissern Sie sich stets, dass die zu sterilisierenden
Ladungen die Temperaturen des gewählten Zyklus vertragen können.
8
DEUTCH
4
4.1
TECHNISCHE DATEN
Mechanische Daten
EUROPA B EVO
Betriebstemperatur
Maximale Meereshöhe
Maximale relative Feuchte bei 30°C
Maximale relative Feuchte bei 40°C
Raumbedarf (L x H x P)
Raumbedarf bei offener Tür
Gewicht (vollen tanks)
Gewicht (vollen tanks)
Gewicht für Auflagefläche
Geräuschpegel
4.2
+5°C ÷ +40°C
2.000 m
80%
50%
510 X 390 X 590 mm
510 X 390 X 730 mm
300 mm
54 kg
58 kg
63 kg
67 kg
2
2058 N/m
2058 N/m2
< 70 db A
Elektrische Daten
Versorgungsspannung
Leistung
Frequenz
Speisekabel
Sicherungen
übertragene Wärme
4.3
EUROPA B EVO 24
EUROPA B EVO
EUROPA B EVO 24
230 V a.c. +/-10 % single phase
1.7 kW
2.2 kW
50 / 60 Hz
2 + 1 x 1mm2
6. 3 x 32 – 12A (x2) / 6.3 x 32 – 500mA (x1)
3.6 E6 J / ora
Kammer
Maximaler Arbeitsdruck
Maximales Vakuum
Höchsttemperatur
Material
Abmessungen
EUROPA B EVO
EUROPA B EVO 24
2.4 bar (relativi)
- 0.9 bar (relativi)
138 °C
Inox AISI 304
Ø 245 x 318 mm
Ø 245 x 500 mm
9
DEUTCH
4.4
Reinwassertank
EUROPA B EVO
EUROPA B EVO 24
Mit
Wasseraufbereitung
4,5 l
Volumen
Ausführbare Zyklen
Bitte s. die
Gebrauchsanleitung von
Wasseraufbereitung
4
Material
4.5
Schmutzwassertank
Volumen
Ausführbare Zyklen
Material
Maximale
Ablasswassertemperatur
EUROPA B EVO 24
4,5 l
4
2
Polyäthylen
50°C
Bakteriologischer Filter
EUROPA B EVO
Durchmesser
Filterkapazität
10
2
Polyäthylen
EUROPA B EVO
4.6
Ohne
Wasseraufbereitung
EUROPA B EVO 24
56 mm
0.3 µm
DEUTCH
05
ZUBEHÖR
Tablettaufnahmegestell
EUROPA B EVO
EUROPA B EVO 24
Eloxiertem aluminium
192 x 165 x 280 mm
192 x 200 x 470 mm
Material
Abmessungen (L x H x P)
Abbildung
FIG.4
FIG.3
Standardausstattung
1
Tablettaufnahmegestell (facultativ)
EUROPA B EVO
Eloxiertem aluminium
193 x 200 x 280 mm
(200 x 193 x 280) mm
Material
Abmessungen (L x H x P)
EUROPA B EVO 24
/
/
Abbildung
/
FIG.5
Standardausstattung
1
/
Tabletts
EUROPA B EVO
Material
Abmessungen
(L x H x P)
184 x 17 x 286
mm
EUROPA B EVO 24
Eloxiertem aluminium
184 x 17 x 286 mm + 184 x 17 x 140 mm
Abbildung
+
FIG.6
Standardausstattung
4
FIG.7
FIG.8
4+4
11
DEUTCH
Zusätzliches Tablettaufnahmegestell
EUROPA B EVO
Material
/
Abmessungen (L x H x P)
/
Abbildung
/
EUROPA B EVO 24
Eloxiertem aluminium oder
rostfreies stal
184 x 17 x 460 mm
FIG.9
Standardausstattung
/
4
Schlüssel für das Herausnehmen des Tabletts und zur Türeinstellung
Für das Herausnehmen und die Handhabung der Tabletts (FIG.11) und um die Tür
einzustellen (Absatz 14.4)
Abbildung
FIG.11
FIG.10
Standardausstattung
1
Trichter aus Plastik mit Schlauch
Für das Einfüllen des Wassers von Hand (Absatz 8.2)
Abbildung
FIG.12
Standardausstattung
1
Schwamm für die Reinigung der Kammer und der Türdichtung
Für das Reinigen der Sterilisationskammer und der Türdichtung (par.14.2-14.3)
Abbildung
FIG.13
Standardausstattung
12
1
DEUTCH
Anschlussstück Wasserzulauf
Zusammen mit dem Plastikschlauch für die manuelle Wasserbefüllung und mit dem mit
dem Filter bestückten Wasserzulaufschlauch zu verwenden.
Abbildung
FIG.14
Standardausstattung
1
Wasserauffüllschlauch mit Filter
Für das Einfüllen des Wassers von Hand (Absatz 8.2)
Abbildung
FIG.15
Standardausstattung
1
Wasserablassschlauch
Für das Ablassen des Brauchwassers (Absatz 8.9)
Abbildung
FIG.16
Standardausstattung
1
Rilsan-schlauch
Ein Ende des Schlauchs an den Überlaufschutz Brauchwasser - Kondensatablass (FIG.1–
pos.4), anschließen, das andere Ende an einen Wasser- und Kondensatauffangbehälter
anschließen.
Abbildung
FIG.17
Standardausstattung
1
13
DEUTCH
Abstandsfuß für die Geräterückseite aus Plastik
Den Abstandsfuß aus Plastik auf der Rückseite des Autoklaven anbringen (FIG.1–pos.0)
damit eine angemessene Belüftung gewährleistet ist, auch wenn der Autoklav nahe einer
Wand aufgestellt wird.
Abbildung
FIG.18
Standardausstattung
1
Schläuche für den rückseitigen Ablass der Verbraucher
1- Überlaufschutz Reinwasser : Ein Ende des Schlauchs an den hinteren Überlaufschutz
Reinwasser (FIG.1–pos.1), anschließen, das andere Ende an einen Auffangbehälter für
entsalzenes Wasser oder Abwasser (Ausführung mit Entsalzer) anschließen.
2- Schlauch für den Hahn des Brauchwasserablaufs : Ein Ende des Schlauchs an das
Anschlussstück FIG.20 anschließen und diese am Brauchwasserablaufhahn (FIG.1–
pos.5), aufdrehen, das andere Ende an einen Auffangbehälter für das Brauchwasser
anschließen.
Abbildung
FIG.19
Standardausstattung
2
Anschlussstück für rückseitigen Brauchwasserablauf
Ein Ende des Schlauchs für Brauchwasserablauf an das Anschlussstück FIG.20
anschließen und dieses am Brauchwasserablaufhahn (FIG.1-pos.5) aufdrehen.
Abbildung
FIG.20
Standardausstattung
14
1
DEUTCH
Speisekabel
Das Ende der Steckers (FIG.21–pos.B) zu hintere Tafel (FIG.1–pos.6) anschließen und
anschließend den Stecker (FIG.21–pos.A) direkt an die Versorgungsbuchse der
elektrischen Anlagen anschließen.
Abbildung
FIG.21
Standardausstattung
06
1
DEN AUTOKLAV AUSPACKEN
Der Autoklav wird angemessen verpackt, sodass er bequem transportiert und verlagert
werden kann und der Packungsinhalt geschützt ist. Der Karton muss gegen Stöße
geschützt werden, ist vorsichtig zu behandeln, sollte nicht auf die Seite gekippt werden
und darf nicht tiefer als 16 cm fallen. Wenn keine selbsttätigen Verlagerungssysteme zur
Verfügung stehen, ist der verpackte Autoklav von 2 Personen zu verlagern. Der Autoklav
liegt auf eine Holzpalette e wurde in einem abgestützten Wellenkarton eingepackt. Um den
Autoklav auszupacken, öffnen Sie den Karton, entfernen Sie die Abstützungsteilen, und
nehmen Sie den Autoklav mit den in der Verpackung enthaltenen Griffen aus.
Die Handhabung darf nur unter Verwendung der Riemen und unter Einsatz
von mindestens zwei Personen erfolgen.
Nie Aufzug, durch Ergreifen des Autoklaven unten an der Tür oder am
Armaturenbrett, da diese falsch, könnte das Gerät beschädigen.
ACHTUNG: Die Original-Verpackung ist über die gesamte Betriebsdauer des Geräts
aufzubewahren.
In der Verpackung befinden sich:
• BEDIENUNGSANLEITUNG : Diese muss aufmerksam gelesen und dann an einem für
alle - mit der Sterilisierung beauftragten - Bediener zugänglichen Ort aufbewahrt
werden.
• ZERTIFIZIERUNGEN : Diese müssen sorgfältig aufbewahrt werden.
• GARANTIESCHEIN : Dieser muss ausgefüllt und Teile davon zurückgesandt werden.
15
DEUTCH
07
INSTALLATION
FIG.22
EUROPA B EVO
A
B
C
D
E
F
EUROPA B EVO 24
610 mm
590 mm
730 mm
50 mm
410 mm
426 mm (rückseitig)
410 mm (rückseitig)
455 mm (vorderseitig)
444 mm (vorderseitig)
400 mm
515 mm
1 - Den Autoklav an einem für die Sterilisierung angemessenen Ort installieren.
2 - Der Raum muss angemessen, und gemäß den geltenden Richtlinien, beleuchtet und
belüftet sein.
3 - Den Autoklav nicht in der Nähe von Wärmequellen und Wasserspritzern installieren.
4 - Den Autoklav auf einer Fläche mit entsprechender Tragfähigkeit (geringste 80kg) und
angemessener Größe aufstellen.
5 - Die Aufstellfläche des Autoklaven muss absolut eben sein.
6 - Den Autoklaven auf einer Höhe aufstellen, die dem Bediener die volle Einsicht in die
Sterilisierkammer gestattet und die Reinigung ermöglicht.
7 - Die Tür des Autoklaven öffnen und alle Beutel, für das Verpacken einzelner
Instrumente, welche in der Kammer aufbewahrt sind, herausnehmen.
16
DEUTCH
8 - In der Sterilisierkammer nur das Tablettaufnahmegestell mit den Tabletts lassen. Das
gesamte Zubehör ist in einem dazu bestimmten und den Bediener zugänglichem Fach
unterzubringen.
9 - Auf dem Autoklaven dürfen keine Zeitungen, Schalen, Behälter mit Flüssigkeiten, usw.
abgestellt werden.
10 - Sich niemals auf der Tür abstützen, wenn diese offen steht.
11 – Um eine ordnungsgemäße Belüftung zu garantieren, muss stets einen
Mindestabstand von 5 cm auf der Geräterückseite gegeben sein. Nutzen Sie dazu den
Abstandsfuß für die Geräterückseite (FIG.1–pos.0 / FIG.18) und die Geräteseiten.
12 - Die Schläuche auf der Rückseite anschließen (Kapitel 5)
13 - Vergewissern Sie sich stets, dass das Stromnetz an das der Autoklav angeschlossen
wird, den geltenden Richtlinien und den für das Gerät erforderlichen Daten entspricht.
14 - Das mitgelieferte Stromkabel an der Steckverbindung auf der Autoklav-Rückseite
anschließen (FIG.1 –pos.6)
15 - Das andere Kabelende an einer Steckdose 230 Vac - 2000 Va des Stromnetzes
anschließen.
WICHTIG
Vermeiden Sie Verlängerungskabel und Adapterstecker, denn es könnten
Mikrounterbrechungen der Stromzufuhr entstehen, die zu einer Störmeldung führen
würden.
16 - Den Autoklaven am Hauptschalter (FIG.2–pos.19) einschalten und seine Tür öffnen.
Einige Sekunden warten. Zwei Signaltöne geben an, dass die Parameter für die
automatische Anpassung an den Luftdruck erfasst werden. Auf dem Display erscheint die
Schrift "TÜR OFFEN".
WICHTIG
Niemals ein Programm anwählen bevor die beiden Signaltöne abgegeben wurden,
da der Autoklav sonst die angewählten Programme nicht annimmt.
17 - Den Reinwassertank füllen, beachten Sie dabei die Anweisungen im kapitel 8.
18 - Wenn der Autoklav mit einem Entsalzungssystem installiert wird, beachten Sie die
Anweisungen in der mit dem Entionisierer mitgelieferten Anleitung kapitel 16.
17
DEUTCH
08
GEBRAUCHSANWEISUNGEN
Nachdem der Autoklav installiert wurde, kann er angewendet werden.
8.1
Den autoclav einschalten
Drücken den Hauptschalter (FIG.2–pos.19)
8.2
Den reinwassertank fullen
An den “Wasserzulaufschlauch mit Filter” (mitgeliefert) das “Wasserzulaufanschlussstück“
FIG.14, anschließen und dann in die Öffnung des frontalen Anschlusses am Autoklaven
(FIG.2–pos.13) stecken.
Das andere Ende des Schlauchs mit Filter in den Behälter für das entsalzene oder
destillierte Wasser stecken.
An dieser Stelle die Taste B-PUMP drücken und so die Wasserzulaufpumpe einschalten;
die Taste so lange drücken bis die Rückwärtszählung beginnt.
Die Tankfüllpumpe befüllt den Innentank des Autoklaven. Sollte der Höchststand binnen
180 Sekunden nicht erreicht werden, schaltet die Pumpe automatisch ab. Sie kann nur
durch erneutes Betätigen der B-PUMP Taste neugestartet werden.
Die Pumpe stoppt automatisch wenn der Höchststand erreicht ist.
Ist ein Entionisierer vorgeschaltet und der Wasserhöchststand wird nicht erreicht, ist der
Betrieb der Autoklaven unterbunden:
• Den Autoklav einschalten.
• Den Verschluss auf der Front des Autoklaven (FIG.2–pos.12) abnehmen.
• Den an den Trichter angeschlossenen Wasserzulaufschlauch in diese Öffnung stecken
(FIG.14)
• Destilliertes Wasser in den Trichter geben, diesen dabei höher als das
Einlaufverbindungsstück halten.
• Einfüllen bis die Led für den Höchststand aufleuchtet (FIG.2–pos.2).
• Nach dem Auffüllvorgang das Anschlussstück herausnehmen und den Verschluss
wieder aufsetzen.
Ist ein Entionisierer vorgeschaltet und der Wasserhöchststand wird nicht erreicht, ist der
Betrieb der Autoklaven unterbunden.
18
DEUTCH
8.3
Eigenschaften des zu verwendeten wassers
TABELLE DER VON DER RICHTLINIE UNI EN 13060: 2009 FESTGELEGTEN
QUALITÄTSSTANDARDS
CEN STANDARD UNI EN 13060 : 2009
Evaporationsrückstände
10
mg/l
≤
Siliziumoxyd
1
mg/l
≤
Eisen
0.2
mg/l
≤
Kadmium
0.005
mg/l
≤
Blei
0.05
mg/l
≤
Schwermetallrückstände, ausgenommen Eisen,
0.1
mg/l
≤
Kadmium, Blei
Chlorid (Cl')
2
mg/l
≤
Phosphat (P20s)
0.5
mg/l
≤
Leitfähigkeit (bei 20°C)
15
≤
µs/cm
PH-Wert (Säurewert)
5 ÷ 7,5
Aussehen
farblos, klar, ohne Ablagerungen
Härte (E-Ionen Alkalierde)
0.02
mmol/l
≤
8.4
Das material in den autoklav geben
Das zu sterilisierende Material auf die mitgelieferten Tabletts legen:
• das Material niemals übereinander legen
• die verpackten Instrumente stets mit der Papierseite nach oben gerichtet anordnen
• das Material darf die Sterilisierkammer oder das Tür niemals berühren
• Greifer und Scheren mit geöffneten Klingen einlegen
FIG.23
Nach dem Laden des Autoklaven, dessen Türe schließen. Auf dem Display erscheint die
Schrift "TÜR GESCHLOSSEN“.
19
DEUTCH
8.5
Den sterilisierzyklus starten
Das der Ladung entsprechende Sterilisierprogramm anwählen, dazu die Taste B-SELECT
drücken. Bei jedem Drücken der Taste schaltet die Anwahl auf das folgende Programm.
Nach der Programmanwahl mit der Taste B-START den Zyklus starten; die Tür wird
automatisch blockiert und der Zyklus startet. Während dem Zyklus zeigt das Display alle
diesbezüglichen Parameter und Informationen.
8.6
Zyklusende
Ein Piepton meldet den Benutzern das Ende des Sterilisierungszyklus und auf dem
Display wird die Meldung “ZYKLUSENDE” angezeigt. Nach dem Signalton, die Sperre de
Tür durch Drücken der Taste Sollte Druck in der Kammer anliegen, lässt sich die Tür
durch diese Taste nicht entsperren. Die Vervollständigung des Druckabbaus abwarten und
die Vorgang wiederholen oder im Notfall die Tasten + zugleich drücken.
Nachdem die Türsperre gelöst wurde, am Türgriff ziehen und diese somit öffnen.
8.7
Das sterilisierte material herausnehmen
In Übereinstimmung mit den geltenden Vorschriften hinsichtlich der Sicherheit und
Hygiene am Arbeitsplatz die passende Persönliche Schutzausrüstung tragen. Die Tabletts
mit dem mitgelieferten Schlüssel (FIG.10), herausnehmen, die Instrumente auskühlen
lassen und an einem Ort aufbewahren an dem sie nicht kontaminiert werden können.
8.8
Das schmutzwasser ablassen
Wenn die Led für den Brauchwasserstand (FIG.2–pos.28) aufleuchtet, diesen leeren.
Wenn der Tank nicht geleert wird, ist der Betrieb des Autoklaven unterbunden.
Den mitgelieferten (FIG.16), Schlauch mit der Steckverbindung an den
Schmutzwasserauslass auf der Front des Autoklaven (FIG.2–pos.14), anschließen, Den
Ringen linksrum aufschrauben (2 vollständige Umdrehungen). Das andere Ende des
Schlauchs in einen Auffangbehälter halten; das Wasser strömt durch die Schwerkraft in
den Auffangbehälter.
WICHTIG :
A - Das in den Auffangbehälter gehaltene Schlauchende muss stets über dem
Schmutzwasserpegel sein, andernfalls würde das Brauchwasser aufgesogen werden.
B - Der Inhalt im Schmutzwassertank des Autoklaven muss immer ganz abgelassen
werden. Die Led für den Höchststand im Schmutzwassertank geht aus, obwohl sich noch
Wasser im Tank befindet, daher gilt diese Led nicht als Hinweis für diesen Vorgang.
Am Ende des Abfluss die Nutmutter schrauben und das Rohr abnehmen.
20
DEUTCH
8.9
Einen sterilisierzyklus unterbrechen
Ein Sterilisierzyklus kann unterbrochen werden, dazu die Taste B-START mindestens 2
Sekunden lang drücken. Der Autoklav gibt einen Signalton ab, schaltet aus Unterdruck
und es erscheint die Fehlermeldung A001 (ZYKLUS UNTERBROCHEN).
Um den Fehler rückzusetzen, die unten angezeigten Tasten zugleich drücken bis der
Signalton ertönt:
B-SELECT
B-PUMP
B-DOOR
ACHTUNG: Wenn der Drucker in Betrieb ist, ist die Quittierung der Störung nicht
möglich.
21
DEUTCH
09
9.1
STERILISIERZYKLUS
Beschreiburg der zyklen
Der Autoklav besitzt drei Zyklusserien:
A - Arbeitszyklen
B - Arbeitszyklen - Nachtzyklen
C - Testzyklen
9.2
Arbeitszyklen
Alle Arbeitszyklen arbeiten mit dem System des fraktionierten Vakuums, daher können mit
diesen Zyklen Hohlkörper, poröse Körper, Feststoffe, verpacktes und unverpacktes
Material sterilisiert werden. Die anwählbaren Temperaturen liegen zwischen 121°C und
134°C. Normalerweise nutzt man die 121°C-Zyklen für t hermoplastisches und
empfindliches Material, während die 134°C-Zyklen bei a llen anderen Materialien
angewendet werden. In allen Fällen sind die Anweisungen des Instrumenten- und
Geräteherstellers zu beachten.
Außerdem stehen 2 Sonderzyklen zur Verfügung:
- Prion-Zyklus, dieser Zyklus wurde für die Kreuzfeld-Jacobs-Krankheit (BSE) entwickelt
wurde.
- Flash-Zyklus, dieser Zyklus wurde für das schnelle Sterilisieren der unverpackten
Instrumente und Geräte entwickelt.
9.3
Arbeitszyclen - nachtzyklen
Der Autoklav ist mit einem speziellem Economy-Gerät ausgestattet.
Dank dieser Einrichtung kann ein Sterilisierzyklus in der Abwesenheit des Bedieners
durchgeführt werden. Wenn nach Abschluss der Sterilisierung die Tür nicht geöffnet wird,
stabilisiert sich der Autoklav und schaltet dann automatisch ab. Dabei bleibt nur der
Hauptschalter eingeschaltet (FIG.2-pos.19).
Der Bediener muss bei seinem Zurückkommen eine beliebige Taste drücken, um den
Autoklaven wieder einzuschalten und kann dann am Display die Ergebnis Zykluses
ablesen. Bei dieser Vorgehensweise druckt der Drucker wie gewohnt den Zyklus-Bericht
aus.
9.4
Testzyklen
Es stehen die folgenden Testzyklen zur Verfügung:
- Helixtest
- Bowie-Dick-Test
- Vakuumtest
22
DEUTCH
10
10.1
KONTROLLTESTS FÜR AUTOKLAV
Chemische Zusätze
Dies sind Tests, welche die Eigenschaften von Färbemitteln nutzen, die ihr Aussehen
verändern, wenn sie der Wärme und dem Druck in entsprechenden Zeiten angemessen
ausgesetzt werden.
Da sie mit allen Zyklusparametern (Druck, Temperatur, Zeit) interagieren. Das Bestehen
des Tests bescheinigt die korrekte Stabilität.
VERWENDEN
Die chemischen Zusätze (Ersatzteil-Nr. 200/S e 215-S) müssen auch
bei vorhandener Ladung vor dem Zyklusstart in die
Sterilisierungskammer gelegt werden.
Die Zusätze können lose sein und auf den Ablagen positioniert
werden oder in die Sterilisierungsbeutel zusammen mit den zu
sterilisierenden Materialien gelegt werden.
Das Einlegen der Zusätze in die Beutel ist ein Verfahren, das sich in
jenen Strukturen empfiehlt, in denen mehrere Benutzer arbeiten.
Dadurch wird nämlich die erfolgte Sterilisierung jedes einzelnen
Geräts bescheinigt.
Wenn die Tonung im weißen
Bereich bleibt, wurde der Test
nicht bestanden
ANTWORT
Wenn die Tonung in das Fenster
SAFE eintritt, wurde der Test
bestanden
PERIODIZITÄT
10.2
Es besteht keine vorbestimmte Regelmäßigkeit, es sei denn in jenen
Gebieten, in denen es spezifische regionale Gesetze gibt. Auf allen
Fällen sind die chemischen Zusätze die wirtschaftlichsten und
unmittelbarsten Tests, um so eine ständige Kontrolle der Wirksamkeit
des Autoklavs zu haben, empfiehlt sich die Verwendung in allen
Zyklen bzw. mindestens einmal täglich.
Biologischer Indikator
Der Test (Ersatzteil-Nr. 262-S) dient zum Nachweis der Leistungsfähigkeit des Autoklavs
in Sachen Vernichtung aller Mikroorganismen. Er besteht aus einer Standardzubereitung
von Sporen, die biologische Eigenschaften und hohe Wärmebeständigkeit besitzen und
die als eine absolut ruhige und sichere Kontrollmethode (ATCC 7953) anzusehen sind.
Den Test gibt es in Ampullenform und er ist nicht pathogen, giftig und Fieber erzeugend.
23
DEUTCH
VERWENDEN
Die Ampullen müssen auch bei vorhandener Ladung in die
Sterilisierungskammer gelegt werden.
Den Zyklus durchführen, am Ende die Ampulle
herausnehmen, da sie warm ist und unter Druck steht.
vorsichtig
Sie circa 10 Minuten abkühlen lassen, danach unter Verwendung der
entsprechenden mitgelieferten Schlüssel aktivieren, indem sie stets
senkrecht gehalten wird. Der chemische auf dem Etikett der Ampulle
angebrachte Indikator hat die Farbe von Blau in Schwarz geändert.
Anschließend die Ampulle zusammen mit einer nicht prozessierten
aber dennoch mit dem entsprechenden Schlüssel aktivierten Ampulle
bei 57°C für 48 Stunden in einen biologischen Inkubato r legen.
ANTWORT
Nach 48 Stunden die prozessierte Ampulle aus dem Inkubator
herausnehmen und die Antwort auswerten. Wenn die Ampulle die
Farbe In Gelb geändert hat, bedeutet dies, dass der Autoklav den
Test nicht bestanden hat und es ein Bakterienwachstum gibt. Wenn
die Ampulle Violett bleibt, zeigt dies an, dass es keine wachsenden
Mikroorganismen gibt und somit der Autoklav den Test bestanden
hat. Die Testampulle ist stets Gelb getont, da sie nicht prozessiert
wurde und dient nur zum Vergleich.
Am Ende des Tests die Ampullen im Hausmüll entsorgen, aber es
empfiehlt sich die Ampulle einem weiteres Sterilisierungszyklus bei
121°C zu unterziehen
PERIODIZITÄT
10.3
Es besteht keine vorbestimmte Regelmäßigkeit, es sei denn in jenen
Gebieten, in denen es spezifische regionale Gesetze gibt. Auf allen
Fällen sind die biologischen Tests die konkretesten, um so eine
ständige Kontrolle der Wirksamkeit des Autoklavs zu haben,
empfiehlt sich die Durchführung des Tests mindestens einmal in
90 Tagen.
Bowie&Dick-Test
Dies ist ein physikalischer Test (Ersatzteil-Nr. 268/S), der die Überprüfung des
Eindringvermögens des Dampfs in poröse Körper zulässt. Für den Test wird ein
standardisiertes und mit den zurzeit geltenden technischen Richtlinien übereinstimmendes
“Testpaket” verwendet.
VERWENDEN
24
Der Test wird mit leerer Kammer durchgeführt. Der Bowie-Dick muss
auf der Mittelablage des Autoklavs positioniert werden. Den auf dem
Display des Autoklavs angezeigten entsprechenden Zyklus
durchführen, am Ende das Paket herausnehmen und die Tonung
des Bogens mit darin befindlichem chemischen Indikator
kontrollieren.
DEUTCH
ANTWORT
PERIODIZITÄT
10.4
Die Auswertung ist schnell und
einfach.
Wenn
die
Tonung
einheitlich
ist
(wie
auf
der
Abbildung), war der Test erfolgreich,
andernfalls wurde der Test nicht
bestanden und der Autoklav ist also
nicht in der Lage, poröse Körper
korrekt zu sterilisieren.
Es besteht keine vorbestimmte Regelmäßigkeit, es sei denn in jenen
Gebieten, in denen es spezifische regionale Gesetze gibt. Auf allen
Fällen wird für eine ständige Kontrolle der Wirksamkeit des
Autoklavs die Durchführung des Tests mindestens einmal in 30
Tagen empfohlen.
Helixtest
Dies ist ein physikalischer Test (Ersatzteil-Nr. 267-S), der die Überprüfung des
Eindringvermögens des Dampfs in Hohlkörper zulässt. Für den Test wird ein mit den
geltenden technischen Richtlinien übereinstimmendes Standardsystem verwendet.
VERWENDEN
Der Test wird in die leere Kammer gelegt.
In die am Ende des Tests angebrachte Kapsel den entsprechenden
Streifen einlegen, anschließend den Test auf der mittleren Ablage
des Autoklavs positionieren.
Den auf dem Display des Autoklavs angezeigten entsprechenden
Zyklus durchführen, am Ende den Test herausnehmen, die Kapsel
öffnen und die Tonung des Indikatorstreifens kontrollieren.
25
DEUTCH
ANTWORT
TEST AM BEGINN
TEST NICHT ERFOLGREICH
TEST ERFOLGREICH
PERIODIZITÄT
10.5
Es besteht keine vorbestimmte Regelmäßigkeit, es sei denn in jenen
Gebieten, in denen es spezifische regionale Gesetze gibt. Auf allen
Fällen wird für eine ständige Kontrolle der Wirksamkeit des
Autoklavs die Durchführung des Tests mindestens einmal in 30
Tagen empfohlen.
Vakuumtest
Dies ist ein Test für Kammerdichtheit oder für Vakuumverlust.
Die Kontrolle zielt auf die Überprüfung ab, dass es zu keinen Luftinfiltrationen über die
Kammerdichtungen (Dichtungen, Ventile, usw…) kommt.
Der Zyklus mit leerer Kammer durchgeführt werden.
VERWENDEN
Es wird der entsprechende auf dem Display des Autoklavs
angezeigte Zyklus gewählt und eingeschaltet.
Der Autoklav führt den Zyklus unter Befolgung genauer technischer
Verfahren durch. Anschließend wird das Endergebnis von einem
Drucker mit einem eventuellen Computeranschluss ausgegeben.
ANTWORT
PERIODIZITÄT
Die Antwort ist umgehend und wird vom Drucker ausgegeben, der
alle vom Zyklus ausgeführten Werte sowie auch die
Abschlussbewertung ausgibt.
Es besteht keine vorbestimmte Regelmäßigkeit, es sei denn in jenen
Gebieten, in denen es spezifische regionale Gesetze gibt. Auf allen
Fällen wird für eine ständige Kontrolle der Wirksamkeit des
Autoklavs die Durchführung des Tests mindestens einmal pro
Woche empfohlen.
WIR EMPFEHLEN, NUR VON TECNO-GAZ S.p.A.
EMPFOHLENEN TEST ZU BENUTZEN
26
DEUTCH
11
TABELLEN UND DIAGRAMME
STERILISIERTABELLE
18
17
13
1.04 ÷ 1.30
121 ÷ 125
5
17
13
2.02 ÷ 2.38
134 ÷ 138
4
2
4
2.02 ÷ 2.38
134 ÷ 138
20
17
13
2.02 ÷ 2.38
134 ÷ 138
2.02 ÷ 2.38
134 ÷ 138
3.5
4
-0.9
Vacuum
SCHALTSPIELE
121°C
(verpackt und
unverpackt)
134°C
(verpackt und
unverpackt)
134°C
Flash
134°C
Prion
134°C
Helix / Bowie&Dick
BETRIEBSTEM
BETRIEBSDRU
PERATURBERE
CKBEREICH
ICH
( relative Bar )
( °C )
TESTZYKLEN
CYCLES
PROZESS- TROCKNUNG
S-ZEIT
ZEIT
T5
(Minuten)
T4
(Minuten)
18L
24L
Die Zeiten für das Vorerwärmen und Fraktionieren des Vakuums können je nach
Vorraussetzungen des Autoklaven zwischen 25-35 Minuten variieren. Alle
Sterilisierungszyklen durchlaufen 3 Vakuumphasen.
ZYKLUSDIAGRAMM
T1,T2,T3
Die Zeiten für das
Vorerwärmen und
Fraktionieren des
Vakuums
T4
Sterilisier-zeit
T5
Trocknungs-zeit
FIG.24
27
DEUTCH
12
LESEN DES ZYKLUSBERICHTS
Softwareversion
Seriennummer
Beschreibung des gewählten Zyklus
(mit den wichtigsten Setup-Werten)
Stunden – Minuten – Sekunden
Druck der Kammer
Haupttemperatur der Kammer (T1)
Zyklusphase
Zyklusende
(OK bedeutet, dass der Zyklus korrekt
beendet wurde)
Zyklusnummer
Benutzer
(Platz für die Unterschrift)
28
DEUTCH
13
FEHLER- UND STÖRUNGSMELDUNGEN
Die Fehlermeldungen werden mit einem alphanumerischen Code dargestellt, dieser
besteht aus einem Buchstaben und 3 Ziffern.
Der Buchstabe "E" bezieht sich auf Fehler durch den Bediener, die von diesem
rückgesetzt werden können.
Der Buchstabe "A" bezieht sich auf Störungen am Autoklaven. Sollte nach der
Fehlerbehebung das Problem weiterhin auftreten, ist der Kundendienst anzurufen.
Wenn eine Störungsmeldung erscheint (Nachsilbe „A“) gilt der Zyklus
als NICHT KORREKT AUSGEFÜHRT: alle Vorbereitungen und die
Sterilisation müssen wiederholt werden.
Um die Störungen und die Fehler rückzusetzen, die unten dargestellten Tasten zugleich
drücken bis der Signalton ertönt:
B-DOOR + B-PUMP + B-SELECT
FEHLER
URSACHE
ABHILFE
E 200
Niedriger Wasserstand
Den Tank füllen
E 250
Niedriger Reinwasserstand mit Zulauf von
Entsalzer
Den Tank füllen
E 300
Schmutzwassertank voll
Den Tank leeren
E 401
Tür offen
Die Tür vorsichtig schließen
E 451
Tür nicht blockiert
E 461
A 250
A 001
A 011
A 101
A 111
A 121
Die Tür öffnen und erneut
schließen
Zu hohe Temperatur in der Kammer für
Ausschalten und 10 Minuten
Vakuumtest (über 40°C)
warten
Das Gerät wasseraufbereitung
Zyklusstart ohne Wasser
systeme überprüfen
Den Zyklus rücksetzen und
Zyklus unterbrochen
neustarten
Den Autoklaven aus-und wieder
Störung PC Display
einschalten
Rücksetzen und den Zyklus
Nach 10 min wurde kein Vakuum aufgebaut
wiederholen
Vakuum nicht aufrechterhalten in der ersten Rücksetzen und den Zyklus
Phase VT
wiederholen
Vakuum nicht aufrechterhalten in der
Rücksetzen und den Zyklus
zweiten Phase VT
wiederholen
29
DEUTCH
A 401
Tür offen während dem Zyklus oder
problem Tür- Schließsystem
A 501
Druck nicht aufrechterhalten
A 551
Störung 1. Temperaturfühler
A 552
Störung 2. Temperaturfühler
A 641
Überdruck
A 651
Übertemperatur
A 661
Störung Temperaturvergleich
A 701
Arbeitsdruck nicht erreicht
A 751
Zu geringe Temperatur
A 761
Störung an Druckfühler
A 801
Störung Druckablass
A 901
Hauptdruck zu gering
Vorübergehend unterbrochene
Stromversorgung
Das Schließsystem überprüfen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Rücksetzen und den Zyklus
wiederholen
Wenden Sie sich an den
Stromversorger
DIE HARZE WECHSELN die Patrone
IONIC SYSTEM sind erschöpft
im Absatz 16.3 der Anleitung
DIE HARZE FILTER die Patrone
OSMOSIS SYSTEM sind erschöpft
im Absatz 16.3 der Anleitung
D
Druckerklappe nicht
geschlossen
Die Klappe des Druckers öffnen und wieder schließen
P
Kein Papier im Drucker
Papier einlegen
Zu geringe Spannung in
Lo
der Versorgungsleitung
30
Schließen Sie die Maschine an einer anderen Stelle des
Stromnetzes an; bleibt das Problem bestehen wenden
Sie sich an einen Elektriker
DEUTCH
14
PROGRAMMIERUNG DER DISPLAY-PARAMETER
Siehe FIG.2
14.1
Die Sprache einstellen
Die Taste B-START und den Schalter MAINS zugleich drücken.
Die Taste B-START gedrückt halten bis auf dem Display die eingestellte Sprache
erscheint.
Mit der Taste B-SELECT wird auf die folgende Sprache geschalten.
Mit der Taste B-PUMP wird auf die letzte Sprache zurückgeschaltet.
14.2
Selezione Ora
Die Taste B-START drücken, es erscheint die Stunde.
Mit der Taste wird di B-SELECT e Stunde verstellt [vorwärts].
Mit der Taste B-PUMP wird die Stunde verstellt [rückwärts].
14.3
Die Uhrzeit einstellen - Minuten
Die Taste B-DOOR drücken, es erscheinen die Minuten.
Mit der Taste B-SELECT werden die Minuten verstellt [vorwärts].
Mit der Taste B-PUMP werden die Minuten verstellt [rückwärts].
14.4
Die Uhrzeit einstellen - Sekunden
Die Taste B-DOOR drücken, es erscheinen die Sekunden. Nicht verändern.
14.5
Das Datum einstellen - Tag
Die Taste B-START drücken, es erscheint der Tag (blinkt).
Mit der Taste B-SELECT wir der Tag verändert [vorwärts].
Mit der Taste B-PUMP wird der Tag verändert [rückwärts].
14.6
Das Datum einstellen – Monat
Die Taste B-DOOR drücken, es erscheint der Monat (blinkt).
Mit der Taste B-SELECT wird der Monat verändert [vorwärts].
Mit der Taste B-PUMP wird der Monat verändert [rückwärts].
31
DEUTCH
14.7
Das Datum einstellen - Jahr
Die Taste B-DOOR drücken, es erscheint das Jahr.
Mit der Taste B-SELECT wird das Jahr verändert [vorwärts].
Mit der Taste B-PUMP wird das Jahr verändert [rückwärts].
14.8
Luftdruckanzeige
Die Taste B-START. Es erscheint der gespeicherte Luftdruck.
NICHT VERÄNDERN!
14.9
Anwahl Wasserzulauf
Wenn Sie B-START drücken, können Sie den Wassereinlauf wählen:
• EINLAUF DURCH PUMPE
• IONIC SYSTEM
• OSMOSI SYSTEM
Wenn Sie B-SELECT drücken, können Sie die Einstellung wechseln.
Zum Speichern, drücken Sie B-DOOR, bis wann den „Pieptone“ hören können
Die Taste B-START gedrückt halten bis der Signalton zu hören ist.
Die eingestellten Daten sind nun gespeichert. Den Autoklaven am Schalter MAINS
ausschalten.
32
DEUTCH
15
BEHEBUNG AUFTRETENDER PROBLEME
In vielen Fällen werden Störungen oder Fehler durch die Nichtbeachtung oder die
Unkenntnis hinsichtlich einiger technischer Aspekte und der Funktionsweise des Geräts
verursacht. Im Folgenden sind einige Störungen und deren Behebungsmaßnahmen
aufgeführt.
15.1
Der autoklav trocknet nicht ordnungsgemass
• den bakteriologischen filter gegen einen neuen originalen der TECNO-GAZ ersetzen.
• es wurden keine originalen tabletts, sondern andersartige ohne oder mit einer
andersartigen lochung verwendet Verwenden Sie nur Original-Tabletts, möglichst aus
Aluminium. Fordern Sie zusätzliche Original-Tabletts an.
• die instrumente wurden nicht ordnungsgemass in den autoklav gelegt. Folgen Sie den
Hinweis auf Absatz 8.4
15.2
Die kammer des autoklaven wird weiss
• sofort das verwendete wasser wechseln, verwenden Sie entsalzenes oder destilliertes
Wasser gemäß den Angaben in den vorigen Kapiteln und reinigen Sie die Kammer.
• die weissliche farbe kann eine Folge der Verdunstung von auf den Instrumenten
abgelagertem organischem Material sein. Reinigen Sie die Instrumente sorgfältiger.
• gegebenenfalls die Entsalzungsanlage überprüfen.
15.3
Die kammer des autoklaven hat grun-blauliche flecken
• die instrumente wurden nach dem Reinigungsvorgang nicht ordnungsgemass
nachgespult, die Instrumente nach dem Reinigen sorgfältiger mit Wasser abspülen.
Sollten die Flecken sehr auffällig sein, wenden Sie sich an den Service-Telefondienst.
15.4
Der sterilisierzyklus wird, ohne ersichtlichen grund, abgebrochen
• überprüfen sie ob der autoklav mit verlängerungskabel oder adaptersteckern an das
stromnetz angeschlossen ist; gegebenenfalls diese zwischenteile entfernen und den
autoklaven direkt an das stromnetz anschließen.
15.5
Der autoklav empfangt keine arbeitsanweisungen
• der autoklav führt gerade die automatische anpassung an den luftdruck durch, warten
sie nach dem öffnen der tür auf den zweifachen signalton und geben sie erst dann die
gewünschten funktionen ein.
• der reinwassertank ist leer, die led für den mindestpegel leuchtet, füllen sie den tank mit
sauberem wasser auf.
• der brauchwassertank ist voll, die led für den höchststand leuchtet lassen sie das
brauchwasser ab.
33
DEUTCH
15.6
Der drucker des autoklaven drunkt nicht
• die klappe der papierrolle ist nicht geschlossen, auf dem Display erscheint der
Buchstabe "D". Öffnen Sie die Klappe. Schließen Sie nun die Klappe wieder und achten
Sie darauf, dass diese richtig zu ist.
• kein papier, auf dem Display erscheint der Buchstabe "P". Öffnen Sie die Klappe und
legen Sie eine Rolle Thermopapier ein.
• die hitzeempfindliche seite des papiers zeigt nach innen, öffnen Sie die Klappe und
legen Sie die Papierrolle richtig ein.
15.7
Flecken auf den instrumenten
• die instrumente werden gelb, rückstände chemischer flüssigkeit die sich durch die hitze
an den instrumenten festgesetzt haben. um dies zu verhindern, die instrumente besser
abwaschen.
• die sterilisierkammer hat gelbe flecken, es wurden instrumente mit rückständen
chemischer flüssigkeit in die kammer gelegt, diese flüssigkeit tropfte durch die
hitzeeinwirkung auf den boden. um dies zu verhindern, die instrumente besser
abgewaschen.
• die instrumente haben weissliche flecken, die instrumente wurden mit sehr kalkhaltigem
wasser abgespült und nicht abgetrocknet. für den letzten abspülvorgang sollte
entsalzenes wasser verwendet und darauf hin die instrumente sorgfältig abgetrocknet
werden.
• die instrumente haben eine schwarze farbe angenommen, dies ist auf einen hohen
kohlegehalt im inneren der instrumente zurückzuführen.
34
DEUTCH
16
16.1
WASSERAUFBEREITUNG
Einstellung von wasseraufbereitung systeme
Der Autoklav ist so entworfen worden, dass das demineralisierte Wasser durch das IONIC
SYSTEM oder das OSMOSIS SYSTEM einlaufen kann (Extra-Zubehör).
Diese Systeme sind notwendig, um die Mineralsalzen zu entfernen, die im Stadtwasser
enthalten sind.
Der Bediener soll die folgenden Normen respektieren, um den Autoklav so einzustellen,
dass entweder IONIC oder OSMOSIS System ergänzt werden kann:
Den Autoklaven am Hauptschalter (FIG.2–POS.19) einschalten und die Taste B-START
gedrückt halten bis der Schriftzug mit der angewählten Sprache erscheint.
Die Taste B-START drücken und loslassen, um die folgende Seite anzuwählen bis
folgendes erscheint:
• EINLAUF DURCH PUMPE
• IONIC SYSTEM
• OSMOSIS SYSTEM
Mit der Taste B-SELECT die Einstellung ändern.
Um die Einstellung zu speichern, die Taste B-SELECT solange gedrückt halten bis der
Signalton ertönt.
Um diese Einstellungen zu beenden, die Taste B-START solange gedrückt halten bis der
Signalton ertönt.
Die eingestellten Daten werden gespeichert.
ACHTUNG:
Das lesbare Nummer, das unter dem Wassereinlauftyp gezeigt wird, bedeutet, wie viele
Sterilisierungszyklen entweder mit dem Filter oder mit der Patrone noch möglich sind.
Es ist wichtig, sich zu erinnern, das o. g. Nummer auf Null zu stellen, wenn Sie den Filter
oder die Patrone wechseln.
Nachdem Sie das Wassereinlauftyp gewählt haben, sollen Sie den Druckknopf B-PUMP
auf der Seite für den Zyklusanfang drücken, bis wann Sie den Piepton hören, um das
Nummer auf Null zu stellen.
35
DEUTCH
16.2
Verbindung von wasseraufbereitungsysteme
Hier wird zusammengefasst, wie den Einlaufrohr für Wasser dem Autoklav verbindet wird,
und wie den Stecker verbindet wird:
•
•
•
•
•
•
•
•
•
•
•
Gegebenenfalls den Autoklav ausschalten (FIG.2–pos.19)
Den Hahn vor der Entsalzungsanlage schließen
Den Einlaufrohr gemäß den Anweisungen im entsprechenden Handbuch montieren;
Das Außengewinde des Schlauch-Anschlussstücks mit Teflon oder ähnlichem
umwickeln, um seine Dichtigkeit zu garantieren;
Das Schlauch-Anschlussstück am Innengewinde des Reinwasserauslauf (FIG.1–pos.2)
Den Gummischlauch an den Auslauf des Einlaufrohr am Schlauch-Anschlussstück, das
nun am Autoklav angeschraubt ist, anschließen;
Den Stecker des Einlaufrohr in die Buchse (FIG.1–pos.7) auf der Rückseite des
Autoklaven stecken;
Den Hahn vor der Entsalzungsanlage öffnen;
Die Anlage auf Lecks überprüfen;
Den Autoklav einschalten;
Einen oder mehrere Sterilisierumläufe durchführen und die Funktionstüchtigkeit der
Anschlüsse überprüfen. Vergewissern Sie sich, dass keine Wasserlecks anliegen.
AM ENDE DES ARBEITSTAGES STETS ÜBERPRÜFEN, DASS DER HAHN VOR
DER ENTSALZUNGSANLAGE ZU IST, SODASS
ÜBERSCHWEMMUNGSGEFAHR AUSGESCHLOSSEN IST
VERBINDEN SIE DIE WASSERAUFBEREITUNGSYSTEME NUR MIT TECNOGAZ AUTOKLAVEN, DIE DAS ERLAUBEN.
WENN DER AUTOKLAV AM ZYKLUSBEGINN AUF E250 SCHALTET,
BEDEUTET DAS, DASS EIN ZYKLUS MIT EINEM ZU GERINGEN
REINWASSERSTAND GESTARTET WURDE. DAHER MUSS DER
REINWASSERTANK GEFÜLLT WERDEN
MERKE:
ZUM ANSCHLIESSEN DIE WASSERAUFBEREITUNGSYSTEME AN TECNO-GAZ
AUTOKLAVEN SIEHE ANGABEN IN DER GEBRAUCHSANLEITUNG DES AUTOKLAVEN.
36
-
DEUTCH
FIG.32
16.3
Wechseln von patrone und filter
Der Autoklav zeigt im Display mit dem Nachricht PATRONE WECHSELN (Ersatzteil-Nr.
242-S) oder FILTER WECHSELN (Ersatzteil-Nr. 248-S-2), wenn die Patrone oder den
Filter erschöpft sind .
Nach dem Auswechseln, zum Rücksetzen des Autoklaven, die Taste B-PUMP drücken bis
der Signalton ertönt.
37
DEUTCH
17
STANDARDVORGÄNGE STERILISIERUNG
Sterilisieren bedeutet, präzise Betriebsprotokolle und ein System anwenden. Der Autoklav
ist das Schlussglied dieses Systems. Wer die verschiedenen Phasen des
Sterilisierungssystems nicht einhält, könnte das Endergebnis beeinträchtigen.
TECNO-GAZ S.p.A. stellt alle Geräte für jede Phase des Sterilisierprozesses her.
Dementsprechend ist Folgendes zu beachten:
1 - Ein präzises schriftliches Betriebsprotokoll anfertigen das allen Bedienern und mit
der Sterilisierung beauftragten Personen ausgehändigt wird. TECNO-GAZ S.p.A. kann
Ihnen Ihren Anforderungen angepasste Protokolle liefern.
2 - Alle Phase des Sterilisierumlauf strengstens befolgen:
DESINFIZIERUNG das ist verbindlich um die Sicherheit der Arbeiter zu schutzen,
anzuwenden bei Eintauchen in chemische Flüssigkeiten oder
Thermodesinfizierung.
REINIGUNG die wichtigste Phase in der alle chemischen und organischen
Rückstände entfernt werden. Das beste Instrument für diesen Vorgang sind die
Ultraschallbecken.
TROCKNEN unbedingt erforderliche Phase, dank der Korrosion der Instrumente
und Interferenzen im Sterilisierzyklus vermieden wird.
VERPACKEN unbedingt erforderliche Phase für die Erhaltung der Keimfreiheit der
behandelten Instrumente.
STERILISIERUNG Endphase Dampfsterilisierung
38
DEUTCH
18
STANDARDVORGÄNGE STERILISIERUNG
Wenn der Autoklav für Reparaturen, Revisionen, Wiederherstellung der
Funktionstüchtigkeit oder Gültigkeit verschickt werden muss, sind folgende Anweisungen
unbedingt zu beachten:
•
Die Original-Verpackung verwenden. Sollte Ihnen diese nicht mehr zur Verfügung
stehen, eine angemessene Verpackung verwenden. Der Absender trägt die
Verantwortung für die Versendung der Ware.
•
Nur den Autoklav versenden (kein im Zubehörbausatz enthaltenes Bestandteil
einlegen).
•
Reinigen Sie, vor dem Versenden, den gesamten Autoklav und insbesondere die
Sterilisierkammer. Sollte der Autoklav verschmutzt oder mit Rückstände zur
Reparatur geschickt werden, wird er unrepariert zurückgeschickt oder gereinigt und
desinfiziert.
•
Der Reinwassertank muss leer sein. Der Tank wird über das Anschlussstück auf
der Rückseite des Autoklaven (FIG.2–pos.2) geleert.
•
Stets den Brauchwassertank leeren. Der Tank wird über das Anschlussstück auf
der Rückseite des Autoklaven (FIG.2–pos.5) geleert.
•
Schriftlich und der Verpackung beigelegt die aufgetretene Störung oder die
gewünschte Dienstleistung genau angeben.
•
Das Gerät frei Haus senden, andernfalls werden die Transport in Rechnung gestellt.
ACHTUNG:
Nicht-Original-Verpackung die bei uns eintrifft, wird entsorgt.
Der Autoklav wird Ihnen in einer neuen Original-Verpackung zugesandt, damit die
Unversehrtheit Ihres Geräts während dem Transport gewahrt wird. Diese neue
Verpackung wird Ihnen in Rechnung gestellt.
39
DEUTCH
19
ANWEISUNGEN FÜR DEN VERSAND DES AUTOKLAVEN
Die Reparaturabteilung ist eine unabhängige Einheit von MEDILINE ITALIA srl (TECNOGAZ GROUP), die sich aus einer gegliederten und professionellen Struktur laut unseren in
unserem “Qualitätshandbuch” angegebenen Verfahren zusammensetzt.
•
Wenn der Autoklav bei der MEDILINE ITALIA SRL eintrifft, wird er direkt von der
Reparaturabteilung angenommen. Dort erhält er eine Bearbeitungsnummer und es
wird eine Service-Karteikarte angelegt.
•
Binnen maximal 48 Stunden wird der Autoklav von Technikern begutachtet und die
technische Diagnose mit Kostenvoranschlag für die erforderliche Leistung erstellt.
Wenn der Autoklav nicht gemäß den Anweisungen im Kapitel ANWEISUNGEN
FÜR DEN VERSAND DES AUTOKLAVEN versendet wird, kann die
Auftragsbearbeitung länger dauern.
•
Der Kostenvoranschlag wird an die zuständige Vertriebsfirma geschickt. Diese
senden ihn dann zur Kenntnisnahme und Annahme durch Unterschrift an den
Endkunden.
•
Nachdem bei MEDILINE ITALIA srl der unterzeichnete Kostenvoranschlag
eingegangen ist, wird der Autoklav repariert und gemäß den im Kostenanschlag
angegebenen Zeiten an den Kunden zurückgeschickt.
•
Die Ware wird
zurückgeschickt.
von
der
Kundenspedition
in
der
Original-Verpackung
WICHTIG
Sollte auch nach der Durchführung der Abhilfemaßnahme das Problem weiterhin
auftreten, rufen Sie den Service-Telefondienst der MEDILINE ITALIA srl unter folgender
Nummer an: +39 0522 94 29 97.
40
DEUTCH
20
GARANTIE
1) DAUER: Auf das Produkt wird eine Garantie von 12 Monaten (zwölf) oder 1200 zyklen
gewährt.
2) AUSSCHLIESSUNGEN: Von der Garantie ausgeschlossen sind:
a) Auf fehlende Wartung oder Nachlässigkeit des Anwenders oder auf einen
unsachgemäßen Gebrauch des Produkts zurückführbare Störungen und Anomalien;
b) Regelmäßige Kontrollen und Wartung;
c) Reparatur und Austausch von Verschleißteilen, empfindlichen Teilen und Teilen
mit einer nicht schätzbaren Lebensdauer, ausgenommen, die Mängel wurden bei
Empfang des Produkts festgestellt und mitgeteilt;
d) Arbeitskräfte, Entsendung des technischen Personals und Transport;
e) Störungen oder Schäden, die auf einen unsachgemäßen Gebrauch und
Anwendungsfehler zurückzuführen sind;
f) Störungen oder Schäden, die auf Verunreinigungen in den Wasser- und
Luftversorgungssystemen, chemische oder elektrische Ereignisse zurückzuführen
sind;
g) Störungen oder Schäden, die auf nicht ausdrücklich in den Bedienungs- und
Wartungsanleitungen
angegebene
Reinigungsmittel,
Desinfektionsmittel,
Sterilisierstoffe oder –verfahren zurückzuführen sind.
h) Natürliche Verfärbung der Kunststoffteile.
3) INSTALLATIONS- UND ABNAHMEPROTOKOLL: Wesentliche Voraussetzung für die
Inanspruchnahme der Garantieleistungen ist die Rücksendung des vollständig
ausgefüllten und vom Anwender unterzeichneten speziellen Installations- und
Abnahmeprotokolls. Dieses Protokoll muss innerhalb von 15 Tagen nach der Installation
zurückgesandt werden, da andernfalls die Garantie verfällt.
4) GARANTIEGRENZEN: Die Garantie verleiht Anspruch auf den kostenlosen Austausch
oder die kostenlose Reparatur der defekten Bauteile. Es besteht kein Anspruch auf den
Austausch des kompletten Geräts. Bezüglich der mit dem Markenzeichen TECNO-GAZ
oder durch Dritte gefertigten Bauteile, die mit einem eigenen Garantieschein ausgestattet
sind, gelten die in diesen Garantiescheinen angegebenen Bedingungen, Grenzen und
Ausschließungen.
Die Durchführung von einer oder mehreren Reparaturen während der Garantiefrist führt zu
keiner Verlängerung des Garantieanspruchs.
5) BEANSTANDUNGEN: Bei Beanstandungen seitens des Käufers bezüglich der
Ausübung der Garantieleistungspflicht und der Qualität oder des Zustands des gelieferten
Geräts ist der Käufer auf keinen Fall berechtigt, die Bezahlung zu verzögern und/oder zu
unterbrechen.
6) STREITIGKEITEN: Für Streitigkeiten bezüglich der Anwendung und Auslegung dieses
Garantiescheins ist unabhängig von dem Ort, an dem der Kaufvertrag für das Gerät
abgeschlossen wurde, das Gericht PARMA (Italien) zuständig.
7) AUSNAHMEN: Bezüglich dieser Garantiebedingungen gewährte Ausnahmen führen zu
keiner Anerkennung von weiteren Rechten des Käufers und beziehen sich einzig und
allein auf den spezifischen Fall.
8) WEITERES: Für alle nicht ausdrücklich in diesem Garantieschein erwähnten Umstände
gelten die anwendbaren Gesetze des Bürgerlichen Gesetzbuchs der Italienischen
Republik.
41
DEUTCH
Der Garantieanspruch verfällt in folgenden Fällen:
a) wenn das Gerät Beschädigungen aufweist, die auf Sturz, Feueraussetzung,
ausgegossene Flüssigkeiten, Blitzschlag, Naturereignisse oder Witterungseinflüsse oder
auf andere nicht auf Fabrikationsmängel zurückführbare Schäden zurückzuführen sind;
b) wenn die Installation nicht gemäß den Anleitungen von TECNO-GAZ erfolgt und durch
nicht autorisiertes Personal durchgeführt wird;
c) wenn das Gerät durch den Käufer oder nicht autorisierte Dritte repariert, abgeändert
oder umgebaut wird;
d) wenn bei Inanspruchnahme der Garantieleistungen festgestellt wird, dass die
Gerätnummer entfernt, getilgt, gefälscht, usw. wurde;
e) wenn das Installations- und Abnahmeprotokoll nicht innerhalb von 15 Tagen nach der
Installation ausgefüllt und unterzeichnet zurückgesandt wird;
f) wenn der Käufer die Bezahlung jedwelcher für den Kauf des Geräts und/oder dessen
Wartung geschuldeter Betrags aus jedwelchen Gründen verzögert oder unterbricht;
g) wenn die in den Bedienungs- und Wartungsanleitungen vorgeschriebene
programmierte Wartung nicht in den angegebenen Zeitabständen durchgeführt wird.
42
DEUTCH
A
INSTANDHALTUNG
Eine ordnungsgemäße Instandhaltung des Autoklaven sichert einen einwandfreien Betrieb
und eine Zeit- und Kostenersparnis in Folge auf Service- und Wartungseingriffe. Die
Folgenden Verfahren sind Pflicht, und sollen von Bedienern durchgeführt werden.
Reiningung vom Kammer
Alle 20 Zyklen oder alle Woche
Die Kammer und der Filter sind regelmäßig zu reinigen. Dabei sind eventuelle
Ablagerungen und Schmutz zu entfernen. Auf diese Weise wird verhindert, dass
Schmutz in den Auslass gelangt und diesen verstopft. Für eine ordnungsgemäße
Reinigung nur Wasser und den mitgelieferten Scheuerschwamm verwenden (Bitte
benutzen Sie die Seite, die nicht abschleifend ist. - FIG.12).
Das o. G. Verfahren muss nur durchgeführt werden, wenn der Kammer kalt ist, um
Brandwunde zu vermeiden - Niemals Lösungsmittel, Reinigungsmittel, chemische
Lösungen, Verkrustungen lösende Säuren oder andere ähnliche Produkte verwenden.
Reiningung vom Kammerfilter
Alle 20 Zyklen oder alle Woche
Ziehen Sie den Filer nach oben (Ersatzteil-Nr. DXBA091), aber passen Sie auf, den
Filter nicht zu beschädigen. Waschen sie den Filter mit aufgereinigtem Wasser, und
trocknen Sie mit einem trocknen und sauberen Tuch. Stellen Sie den Filter wieder ein,
und passen Sie auf, dass der Filter 15 mm vorsteht.
Reinigung von Tray und Trayhalter
Alle 20 Zyklen oder alle Woche
Machen Sie die Tray und den Trayhalter sauber mit dem zusammengelieferten
Schwamm mit Wasser. Bitte benutzen Sie die Seite, die nicht abschleifend ist.
Den bakteriologischen filter ersetzen
Alle 200 Zyklen oder wenn der
Filter dunkel wird
Der bakteriologische Filter (FIG.2–pos.10) linksrum drehen um ihn abzudrehen und
rechtsrum drehen, um ihn aufzudrehen. Bauen Sie nur Original-Filter ein (Ersatzteil-Nr.
DAVA101)
Die turdichtung reinigen
Alle 20 Zyklen oder alle Woche
A-1
DEUTCH
Die regelmäßig die eventuellen Rückstände die sich an der Dichtung absetzen mit
Wasser und dem mitgelieferten Schwamm (nicht abrasive Seite) oder mit einem
feuchten Tuch entfernen (Ersatzteil-Nr. DANA038).
Regelung von der Tür
Alle 2 Monate
Um den Schließdruck der Tür einzustellen, mit dem mitgelieferten Schlüssel mit
Doppelfunktion (Tablettentnahme + Schlüssel) - (Ersatzteil-Nr. DANA008) die
entsprechende Regelschraube der Tür verstellen. Linksrumdrehen (pos.A) um den
Schließdruck zu erhöhen. Rechtsrumdrehen (pos.B) um den Schließdruck zu senken.
A
Das drunckerpapier wechseln
B
-
Rolle thermografisches Papier mit einer Breite von 57 mm (cod.ricambio DNVA363). Das
thermografische Papier darf nur einseitig verwendet werden. Die Klappe am Drucker
öffnen, indem Sie die Taste betätigen, die Papierrolle einschieben, beachten Sie dabei die
Drehrichtung, das Papier herausziehen und die Klappe schließen; der Drucker ist jetzt
betriebsbereit .
A-2
DEUTCH
Die folgende Verfahren werden vom Hersteller empfohlen. Sie müssen nur von geschulten
und vom Hersteller freigegebenen Technikern durchgeführt werden.
Umtausch vom Wassereinlauffilter
Überprüfung vom Wassereinlauffilter in
Durchflußmesser
Reinigung vom Filter Y-förmig aus Messing
Reinigung von der EV4 - Wassereinlaufventil
Umtausch von den Ventilen in der Vakuumpumpe
Reinigung vom Kühler
Reinigung von Lüfter
Überprüfung von Luftkreis
Überprüfung von Elektrokreis
Überprüfung vom thermischen Isolierband um dem
Kammer
Überprüfung von den Schrauben und von der Heizung
vom Kammer
Überprüfung von Temperaturfühler
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Überprüfung von Sicherheitsventil
Beachten Sie die Anleitungen
vom Hersteller*
Schmierung vom bewegenden Teil vom Türgriff
Überprüfung vom Niveaufühler
Überprüfung von den Tankzustände
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
Alle Jahre / Alle 1000 Zyklen*
* Im Anhang mit dem früheren erreichten Termin
A-3
DEUTCH
Einige Staat- oder Regionorm verschreiben die Validierung vom Gerät, die die
Sterilisierungsparameter überprüft. Die o. g. Validierung soll immer von hoch
spezialisierten Technikern durchgeführt werden, die geeignete Instrumenten benutzen.
Falls die Validierung nicht obligatorisch ist, gilt als Erklärung, dass der Steri richtig
funktioniert, und dass er für die Sterilisierung von den mitgeteilten Beladungen geeignet
ist.
NOTES
A-4
TECNO-GAZ S.p.A.
Strada Cavalli N°4 • 43038
Sala Baganza • Parma • ITALIA
Tel. +39 0521 83.80
Fax. +39 0521 83.33.91
I
Il presente manuale deve sempre accompagnare il prodotto, in adempimento alle Direttive Comunitarie Europee. TECNO-GAZ, si riserva
il diritto di apporre modifiche al presente documento senza dare alcun pre-avviso. La ditta TECNO-GAZ si riserva la proprietà del
presente documento e ne vieta l’utilizzo o la divulgazione a terzi senza il proprio benestare.
EN
This manual must always be kept with the product, in complying with the Directives of European Community.
TECNO-GAZ reserves the right to modify the enclosed document without notice. TECNO-GAZ reserves the property of the document and
forbids others to use it or spread it without its approval.
F
Ce manuel doit toujours accompagner l'appareil conformément aux Directives de la Communauté européenne.
TECNO-GAZ se réserve le droit d'y apporter des modifications sans aucun préavis. TECNO-GAZ se réserve la propriété de ce manuel.
Toute utilisation ou divulgation à des tiers est interdite sans son autorisation.
E
El presente manual siempre deberá acompañar el producto al cual pertenece, cumpliendo las Directivas Comunitarias Europeas.
TECNO-GAZ, reserva el derecho de aportar alteraciones al presente documento sin avisos previos. La empresa TECNO-GAZ reserva a si
los derechos de propiedad del presente documento prohiebiendo su utilización o divulgación a terceros sin que haya expresado su
consentimiento.
D
www.tecnogaz.com
Diese Anleitung muss dem Gerät in Übereinstimmung mit den Richtlinien der Europäischen Gemeinschaft beigelegt werden. TECNOGAZ behält sich dar Recht vor ohne Vorankündigung Änderungen in dieser Dokumentation vorzunehmen. Die Firma TECNO-GAZ ist
Eigentümer der vorliegenden Dokumentation und verbietet die Nutzung von Drittenoder die Weitergabe an Dritte ohne entsprechende
Genehmigung .

Автоклав Europa B Evo 24 (TECNO-GAZ)

Europa B Evo — автоклав, разработанный специалистами компании Tecno-Gaz совместно с ведущими научно-исследовательскими структурами данной отрасли.

Europa B Evo незаменим для всех медицинских и стоматологических клиник и во всех областях, где необходима безопасная стерилизация всех видов материалов: полых, пористых, твердых, упакованных и неупакованных.

Серия Europa B Evo является успешным продолжением достижений компании Tecno-Gaz за последние годы, что подтверждается тысячами профессионалов по всему миру, отдающими предпочтение продукции Tecno-Gaz.

  • Встроенное подключение для биологического инкубатора B-Test. Справа находится разъем для подключения биологического инкубатора, управляемого непосредственно автоклавом.
  • Электронный контроль. Europa B Evo управляется новой системой электроники, которая оснащена микропроцессором последнего поколения.
  • Стерилизационные циклы. 4 цикла стерилизации: 1210С и 1340С для упакованных и неупакованных инструментов, Prion для болезни Крейтцфельдта — Якоба, Flash — ускоренный цикл для использования инструментов сразу после стерилизации, а не для хранения; а также наличие трех тест-циклов.
  • Система охлаждения. Новая система охлаждения с двухкорпусным радиатором.
  • Безопасность дверцы. 3 типа защитных систем в дверце автоклава:
    1. Первая система механическая и электромагнитная: включается при старте стерилизационного цикла и полностью блокирует возможность открытия дверцы.
    Эта система может быть отключена в конце цикла нажатием кнопки разблокировки, но только в том случае, когда давление в камере положительное и близко к нулю.
    2. Вторая система электрическая: включается автоматически когда регистрируется попытка силового открытия двери.
    3. Третья система динамическая и действует на структуру уплотнителя двери.
  • Совокупность систем безопасности
    — сертифицированный защитный клапан
    — стерилизационная камера выполнена в соответствии со стандартом PED
    — изоляция камеры выполнена с использованием сертифицированных керамических материалов.
  • Регулировка давления. Автоматическая регулировка давления обеспечивается сразу после включения аппарата без дополнительного вмешательства.
  • Ночной цикл. Автоматический ночной цикл с системой предотвращения скопления конденсата, автоматическим выключением и эффектом памяти.
  • Обеспечение основного водоснабжения. Система деминерализации воды Pura Plus — это чрезвычайно компактная часть оборудования для деминерализации воды, созданная для непосредственной установки на автоклав, что гарантирует постоянное водоснабжение. Является опцией для автоклавов Europa B Evo, входит в основную комплектацию для автоклава Europa B Evo 24.
  • Встроенный принтер.
  • PID-контроллер. Интегральная система для контроля и управления всеми параметрами оборудования. Контролирует значения температуры и давления во время стерилизационного цикла, следит за их колебаниями, при необходимости добавляя воду при приближении к превышению порога температуры.

Дополнительные материалы

  • Автоклав Europa B Evo 24 (TECNO-GAZ)

Возможности автоклава Europa B 24 Evo:

  • встроенное подключение для биологического инкубатора B-Test (справа находится разъем для подключения биологического инкубатора, управляемого непосредственно автоклавом);
  • электронный контроль (EuropaBEvo управляется новой системой электроники, которая оснащена микропроцессором последнего поколения);
  • стерилизационные циклы (4 цикла стерилизации: 121°С и 134°С для упакованных и неупакованных инструментов, Prion для болезни Крейтцфельдта — Якоба, Flash
  • ускоренный цикл для использования инструментов сразу после стерилизации, а не для хранения; а также наличие трех тест-циклов);
  • система охлаждения (новая система охлаждения с двухкорпусным радиатором.
  • безопасность дверцы (3 типа защитных систем в дверце автоклава: первая система механическая и электромагнитная: включается при старте стерилизационного цикла и полностью блокирует возможность открытия дверцы, эта система может быть отключена в конце цикла нажатием кнопки разблокировки, но только в том случае, когда давление в камере положительное и близко к нулю; вторая система электрическая: включается автоматически когда регистрируется попытка силового открытия двери; третья система динамическая и действует на структуру уплотнителя двери);
  • совокупность систем безопасности (сертифицированный защитный клапан, стерилизационная камера выполнена в соответствии со стандартом PED, изоляция камеры выполнена с использованием сертифицированных керамических материалов);
  • регулировка давления (автоматическая регулировка давления обеспечивается сразу после включения аппарата без дополнительного вмешательства);
  • ночной цикл (автоматический ночной цикл с системой предотвращения скопления конденсата, автоматическим выключением и эффектом памяти);
  • обеспечение основного водоснабжения (система деминерализации воды PuraPlus – это чрезвычайно компактная часть оборудования для деминерализации воды, созданная для непосредственной установки на автоклав, что гарантирует постоянное водоснабжение, она является опцией для автоклавов Europa B Evo, входит в основную комплектацию для автоклава Europa B Evo 24);
  • встроенный принтер;
  • PID-контроллер (пропорционально-интегрально-дифференциальный контролер, который является вершиной контролеров в управляющем контуре с обратной связью, его структура технологически продвинутого контроля является простой и гибкой. PID обеспечивает постоянный контроль и мониторинг всей операционной системы автоклава, действуя автоматически в случае отклонений или неправильных значений в технических и электронных параметрах, система ведет себя как «реальный техник» с диагностикой в реальном времени и возможностью вмешательства., что снижает возможность отклонений и тревожных срабатываний, гарантируя максимальную надежность и минимизируя время работы аппарата).

    Типы циклов и их продолжительность при полной загрузке:

  • 38 мин в режиме 121°С
  • 24 мин в режиме 134°С
  • 32 мин в режиме Flash
  • 58 мин в режиме Prion

    Технические характеристики:

  • Объем стерилизационной камеры: 24 л
  • Размеры (ШxВxД): 510 х 390 х 730 мм
  • Вес: 58 кг
  • Питающее напряжение: 230 В, 50/60 Гц
  • Потребляемая мощность: 2,2 кВт

    Размеры стерилизационной камеры

  • Диаметр камеры: 245 мм
  • Глубина камеры: 500 мм

    Опции

  • Встроенный принтер: да
  • Подключение к биологическому инкубатору: да
  • Ручная блокировка дверцы: да
  • Возможность подключения системы деминерализации PURA: да
  • Наличие системы деминерализации PURA в комплекте поставки по умолчанию: да
  • Ночной цикл: да
  • Система P.I.D.: да
  • Автоматическая регулировка давления: да

Понравилась статья? Поделить с друзьями:
  • Автокад инструкция для начинающих скачать бесплатно
  • Автозапуск пандора с брелка как делать инструкция
  • Автокад для чайников инструкция с картинками 2022
  • Автозапуск на автомобиль инструкция starline
  • Автозапуск аллигатор как завести с брелка инструкция