-
Contents
-
Table of Contents
-
Troubleshooting
-
Bookmarks
Quick Links
DOMINA PLUS B
Water Steam Sterilizer
User Manual
Rev. 8
Date: March 2017
OM1004U
Related Manuals for dental X DOMINA PLUS B
Summary of Contents for dental X DOMINA PLUS B
-
Page 1
DOMINA PLUS B Water Steam Sterilizer User Manual Rev. 8 OM1004U Date: March 2017… -
Page 2
DentalX. Information in this manual is subject to change without any warning or prior notice by Dental X and does not represent a commitment for the vendor. -
Page 3: Table Of Contents
8.5 CLASS B ADDITIONAL ALARMS …………………. 27 CONNECTIONS ……………………….28 9.1 CONNECTION TO AN EXTERNAL PRINTER ………………. 28 9.2 INTEGRATED PRINTER (OPTION) ………………..28 9.3 CONNECTION TO A COMPUTER (ONLY FOR SERVICE) …………… 29 OPERATOR’S NOTES ……………………..30 APPENDIX: SERVICE BOOK Domina Plus B…
-
Page 4: General
FORWARDER, DENTAL X AND ITS AREA AGENT. CONFORMITY TO EUROPEAN DIRECTIVES Domina Plus B is produced by Dental X and conforms to electromagnetic compatibility standards in ac- cordance with the Medical Devices Directive 93/42/EEC, with European Standard EN 13060 and with the Pressure Equipment Directive (PED) 97/23/CE.
-
Page 5: Familiarization
DentalX 2. FAMILIARIZATION PACKAGE DIMENSIONS AND WEIGHT Total weight: 55 Kg Store the package for future shipment. UNPACKAGING Domina Plus B…
-
Page 6: Overall Dimensions
57 Kg USABLE SPACE IN THE CHAMBER Diameter: 240 mm Depth: 384 mm Capacity: 17,5 liters Useful dimensions per tray: 315 x 214 mm (x 2), 315 x 168 mm (x 2) Useful tray volume : 10 liters Domina Plus B…
-
Page 7: Safety Features
Elapsed 30 minutes from the end of the cycle without opening the door or without activating a front panel button, the unit will switch off automatically. This function is not operating if no sterilization cycle has been carried out and completed. Domina Plus B…
-
Page 8: Precautions And Intended Use
The unit is not to be used to sterilize liquids and flammable materials. The unit is designed for indoor use only. Do not use in presence of anesthetic or flammable gas. To avoid an excessive level of humidity, properly air the room where the unit is installed. Domina Plus B…
-
Page 9
the device. INTENDED USE Domina Plus B is a small table steam generator designed to decontaminate dental and medical products in accordance with ISO EN 13060 requirements. Its intended use is to perform: • Class B sterilisation The sterilisation of all wrapped or non-wrapped, solid, type A hollow-load products and porous products as represented by the test loads. -
Page 10: Front And Rear View
DentalX FRONT AND REAR VIEW Domina Plus B…
-
Page 11: Standard Accessories
20% to 85%, atmospheric pressure from 750 mBar to 1050 mBar and alti- tude between 0 and +2000 meters. Storage conditions: temperature -10°C 50°C, humidity without condensation 10-95%, atmospheric pressure 500-1060 mBar. NEVER USE IN PRESENCE OF INFLAMMABILE ANESTHETIC GAS Domina Plus B…
-
Page 12: Technical Specifications
Pollution degree: 2 Transient overvoltage protection: II Monitoring of the water conductivity Maximum volume available on the trays: 10 l Maximum chamber temperature: 135°C (-0/+2°C) Safety valve working pressure : 2,55 bar Pressurised container conforming to Directive 97/23/EC (PED) Domina Plus B…
-
Page 13: Installation
(ex. near a wall). The sterilizer can also be installed in a cabinet. In this case, provide adequate space (>10 cm) for aeration. Domina Plus B…
-
Page 14: Getting Started
(see next page). Then push the key to confirm the displayed value and start the automatic initialization procedure with the entry of water in the hydraulic circuit and chamber. Domina Plus B…
-
Page 15: Notes On Altitude Compensation
AL8 or AL5 alarm should be signaled. CONVERSION NOTE: to obtain meters multiply feet by 0.3048. These operations can be carried out by qualified personnel. Wrong settings can affect the quality of the sterilization. Domina Plus B…
-
Page 16: Operating Instructions
Test : allows to carry out the Bowie & Dick Test with sterilizer normally turned on, or the Vacuum Test with sterilizer in stand-by and chamber temperature lower than 35°C. Domina Plus B…
-
Page 17: Running A Sterilizazion Cycle
TEST Not a sterilization COMPLY drying 10’ (6’ with vacuum + 4’ 2,04-2,25 bar or equivalent cycle cod. 1300) ventilation) Vacuum test Temperature under 35°C TEST < 35°C Empty chamber Not a sterilization cycle (*) Liquids excluded Domina Plus B…
-
Page 18: Starting A Sterilization Cycle
The display Time shows the countdown of this phase. Follows the forced ventilation phase through the bacterial filter – the display Time shows also the countdown of this phase. Domina Plus B…
-
Page 19: Stopping The Cycle
Remove the load and check for the presence of water into the chamber. In case of wrapped instruments, we suggest to replace with new bags. Before loading the chamber again, dry it carefully and wait 10 minutes to allow the water to evaporate and be drained completely. Domina Plus B…
-
Page 20: Topping Up And Draining The Tanks
CAUTION! The water contained in the used water tank is to be regarded as biologically con- taminating, therefore when this tank is emptied, suitable precautions should be taken. The dis- posal of recovery water needs to be done in accordance with the national or local regulation. Domina Plus B…
-
Page 21: Programming
Press key measurement unit SET UNIT PSI Push again the key to exit the programming mode The equipment is normally preset in the factory with “°C” and “bar” measurement units; the default lan- guage is English (L2). Domina Plus B…
-
Page 22: Special Cycle S5
At the end of the special cycle S5, only the Led READY will switch on — not STERILIZE — to sig-nal that the efficiency of the cycle selected by the operator has not been validated by the manu-facturer. Domina Plus B…
-
Page 23: Maintenance
In order to extend the sterilizer life, we recommend to carry out an accurate cleaning of instruments; one of the main causes of an early equipment wear is the settlement and accumulation of debris and fragments for inadequately in- strument cleaning, and consequent stains, fouling and progressive clogging of filters, electro-valves and tubing. Domina Plus B…
-
Page 24: Cleaning/Replacing The Water Inlet Filter
The frequency of these checks is regulated at local country level; check the regula- tions in force. For any explanations or information, contact your authorized reseller or directly Dental X. 6.5.1 Bowie & Dick test The test can be performed at any time with the sterilizer turned on and operating.
-
Page 25: Troubleshooting
Push the key Electro-valve 4 energized (open) Power Push the key Electro-valve 5 energized (closed) Test Push the key Condenser’s fan powered Exit the diagnosis and return to the normal mode Push the key Domina Plus B…
-
Page 26: Initial Auto-Test
< 0,5 mg/l < 0,1 mg/l Conductivity (at 20°C) < 15 uS < 3 uS from 5 to 7,5 from 5 to 7 appearance colorless, clean, without sediment colorless, clean, without sediment hardness < 0,02 mmol/l < 0,02 mmol/l Domina Plus B…
-
Page 27: Alarms
Need for the installation procedure. Chapter 3.2) NEED TEST Detected a pre-warning alarm See table B Clean the door gasket and repeat the test. TEST FAIL Negative result of the Vacuum Test Call for a technical service Domina Plus B…
-
Page 28: Pre-Warning Alarms
(see § 6.1), and the sterilization phase the Vacuum Test. AL 10 Pressure too high Call for a technical service. Check the door tightness. Perform, if needed, AL 11 Pressure too low the automatic cleaning cycle (see § 6.1). Domina Plus B…
-
Page 29: Class B Additional Alarms
Solution AL 18 Drying phase Drying interrupted Dry the load AL 31 Drying phase Vacuum not sufficient Excess of load NOTE: The Class B alarms may occur only for the programs 1, 2, S2, S3 and S4. Domina Plus B…
-
Page 30: Connections
The printer cable cannot exceed 3 meters. The serial port of the equipment can be connected to a printer with serial interface RS232. Please contact Dental X for further information. Switch-on the printer, Switch-on the sterilizer. The report is automatically printed on the course of the cycle and with the following data content: date and time of the process –…
-
Page 31: Connection To A Computer (Only For Service)
Do not remove it if the device is not in use. Do not connect devices not supported by the manufacturer. THE RJ45 CONNECTOR IS NOT DESIGNED FOR A LAN CONNECTION, THIS OPER- ATION WILL DESTROY THE MOST OF THE ELECTRONIC PART OF THE AUTO- CLAVE. Domina Plus B…
-
Page 32: 10. Operator’s Notes
DentalX 10. OPERATOR’S NOTES Domina Plus B…
-
Page 33
DentalX Domina Plus B… -
Page 34: Appendix Service Book
“ NEED SERVICE”. This feature is in compliance with the norm EN13060 and answers to spe- cific safety requirements. The maintenance activities must be performed by authorized technicians (provided with card or certificate released by Dental X) and reported on this Service Book. Model …………..
-
Page 35
APPENDIX SERVICE BOOK □ □ □ □ Sterilizer OK Sterilizer OK Date …………….Date …………….No. of cycle ….Cleaning cycles ……No. of cycle ….Cleaning cycles ……Aborted cycles … Alarm code ……..Aborted cycles ..Alarm code ……..Technician name: …………. -
Page 36
APPENDIX SERVICE BOOK □ □ □ □ Sterilizer OK Sterilizer OK Date …………….Date …………….No. of cycle ….Cleaning cycles ……No. of cycle ….Cleaning cycles ……Aborted cycles ..Alarm code ……..Aborted cycles … Alarm code ……..Technician name: …………. -
Page 37
APPENDIX SERVICE BOOK □ □ □ □ Sterilizer OK Sterilizer OK Date …………….Date …………….No. of cycle ….Cleaning cycles ……No. of cycle ….Cleaning cycles ……Aborted cycles … Alarm code ……..Aborted cycles ..Alarm code ……..Technician name: …………. -
Page 40
DENTAL X S.p.A. Via Marzotto 11 36031 Dueville (VI) Italy Tel. +39 0444 367400 Fax +39 0444 367436 e-mail: dentalx@dentalx.it http://www.dentalx.it Dental X 2017 Printed in Italy Sous réserve de modifi- Ci riserviamo il diritto di Reservados los derechos de Änderungen im Zuge techni-…
Domina Plus B, Dental X – это автоклав класса B!Он подходит для ежедневной стерилизации медицинских инструментов абсолютно любой конфигурации и формы без исключений!
Это настоящий шедевр от итальянских производителей!
Аппарат имеет следующие необходимые функции:
- автоматическое управление циклом стерилизации
- надёжную электромеханическую фиксацию дверцы с тройной системой блокировки!
- усиленный предвакуум (до 4-х циклов)
- 4 ёмких лотка увеличенного размера для стерилизации
- 7 стандартных программ стерилизации для разных типов объектов
- возможность индивидуального режима (t: 105˚С-135˚С)
- систему вакуумной сушки
- штатную автоматическую подачу воды
- тесты B&D и VACUUM-тест для подтверждения качества стерилизации
Ускоренная программа стерилизации требует лишь 12 минут с учётом сушки!
Настоящий шедевр!
Здесь собраны все находки, начиная от уникальной медно-никелевой камеры и заканчивая применением двухмоторной вакуумной помпы. Domina Plus B каждую секунду думает за пользователя и выдает лучшее решение! Размещение разноразмерных лотков увеличивает и без того значительный эффективный объём для размещения объектов стерилизации.
Отличная скорость!
Автоклавы Dental X одни из самых быстрых потому что используют медно-никелевые камеры, имеющие очень высокую теплопроводность.
Эффективное использование пространства!
Быстрая и равномерная разогреваемость камеры позволяет использовать бОльшую часть объема, что позволяет увеличить как размеры лотков, так и их количество. При одинаковых объёмах камеры в автоклавах Dental X одновременно стерилизуется до 20-30% инструментов больше!
Хорошая вместимость!
В четырёх больших лотках размещается до 6 кг инструментов, Эффективный объём Domina Plus B не меньше такого же объёма автоклава на 23-24 литра от других брендов!
Высокая экономичность
За счёт высокой скорости разогрева и эффективности использования внутреннего объёма камеры Domina Plus B тратит до 40-50% меньше электрической энергии на поддержания необходимых режимов температуры и давления. Средняя мощность лишь 1 кВт против порядка 1,6-1,8 кВт у автоклавов аналогичной производительности.
Современные технологии стерилизации
Современная технология трехфазного пульсирующего предвакуума гарантирует, что горячий пар проникнет во все полости, тонкие трубочки и пористый материал, а вакуумная сушка обеспечит полную стерильность на завершающей стадии.
Надёжный корпус
Корпус сделан из медицинского пластика, рассчитанного на регулярную обработку дезинфицирующими средствами без появлений изменений цвета и неровностей поверхности.
Бесшумность работы!
Благодаря встроенному высокопроизводительному вакуумному насосу автоклав Domina Plus B работает совершенно бесшумно и создает уровень вакуума, достаточный для проведения качественной стерилизации инструментов.
Простота управления и контроля
Панель управления очень проста, а удобство и наглядность стерилизации обеспечивает LED-дисплей и система звуковой сигнализации.
Надёжная автоматика
Полностью автоматическое управление, включая трёхфакторную автоматическую блокировку дверцы, позволяет быть на 100% уверенным в высоком качестве результатов обработки!
Простая отчетность для архива данных
Для регистрации данных о стерилизации предусмотрена опция размещения встроенного принтера и порта USB-адаптера.
Domina Plus B, Dental X – это высочайшее качество и технологии на службе Вашего бизнеса!
Главная/Автоклавы медицинские/Автоматический автоклав DOMINA PLUS B 18L класса B с вакуумной сушкой и быстрым циклом
- Описание
- Параметры
- Отзывы
- Инструкция
- Аксессуары
- Модификации
- Описание
DOMINA PLUS B 18L — автоматический автоклав с вакуумной сушкой и быстрым циклом, соединивший в себе передовые комплектующие, компактные формы, эргономичный дизайн и замечательные показатели работы. Процесс стерильной обработки инструментов полностью автоматизирован, благодаря современной компьютерной технике. К системе можно подключить как специальный, так и стационарный принтер, который позволит в печатном виде получать информацию о проводимом стерилизационном цикле.
—
DOMINA PLUS B 18L
Технические характеристики:
- Объем камеры 18 л
- 4 лотка
- Фракционный вакуум
- 4 цикла предвакуума
- Вакуумная сушка
- 7 стандартных программ + программа, задаваемая пользователем
- Тестовые программы: Bowie&Dick-тест, Vacuum-тест
- Температура стерилизации 105-135оС
- Подключение к внешнему принтеру
- Встроенный принтер (опция)
- USB-комплект (опция)
- Габариты 443x590x428 мм
- Размеры камеры ø240х384 мм
- Максимальная потребляемая мощность 1900 Вт
- Вес нетто 55 кг
check
Нам доверяют
С нами работают известные мировые производители
Обновление каталога
Каталог товаров регулярно расширяется и пополняется
Гарантия возврата
Не понравился товар? Мы вернем деньги
Быстрая доставка
Быстрая доставка по всей территории России
1
Оставьте заявку
Заполните заявку на сайте или позвоните нам
2
Мы перезваниваем
Перезваниваем вам и обговариваем детали заказа
3
Доставляем товар
Осуществляем доставку по указанному вами адресу
4
Производите оплату
Вы производите оплату любым удобным способом
Курьерская доставка
Курьерская доставка работает с 9.00 до 19.00. Когда товар поступит на склад, курьерская служба свяжется для уточнения деталей. Специалист предложит выбрать удобное время доставки и уточнит адрес. Осмотрите упаковку на целостность и соответствие указанной комплектации.
Почтовая доставка
Почтовая доставка через почту России. Когда заказ придет в отделение, на ваш адрес придет извещение о посылке. Перед оплатой вы можете оценить состояние коробки: вес, целостность. Вскрывать коробку самостоятельно вы можете только после оплаты заказа. Один заказ может содержать не больше 10 позиций и его стоимость не должна превышать 100 000 р.
DOMINA PLUS B
Water Steam Sterilizer
User Manual
Rev. 5
Date: December 2011
0120
© DentalX
2011
DOMINA PLUS B fulfills all the directions in force concerning the safety, and the built-in parameters has been properly set by the manufacturer in order to warranty effective sterilization if proper loading conditions are followed.
Please, read carefully this manual before using the machine; an improper utilization of the sterilizer should carry on defective sterilization with unattended consequences.
In case of doubt or questions, please call the agent.
Thanks for the confidence given.
All rights reserved
No portion of this publication can be printed, transmitted, rewritten, stored in a data recovery system, translated in any foreign or computer language, in any form or through any devices, without written consent by
DentalX.
Information in this manual is subject to change without any warning or prior notice by Dental X and does not represent a commitment for the vendor.
DOMINA PLUS B is trademark of Dental X.
DENTAL X S.p.A.
Via Marzotto 11
36031 Dueville (VI) Italy
Tel. +39 0444 367400
Fax +39 0444 367436 e-mail: [email protected] http://www.dentalx.it
SERVICE
Italy Tel. +39 0444 367415
Foreign Tel. +39 02 76275234
DentalX
TABLE OF CONTENTS
1.
GENERAL ………………………………………………………………………………………………………………………………….. 2
1.1
INTRODUCTION …………………………………………………………………………………………………………………. 2
1.2
CONFORMITY TO EUROPEAN DIRECTIVES …………………………………………………………………………. 2
2.
FAMILIARIZATION ……………………………………………………………………………………………………………………… 3
2.1
PACKAGE DIMENSIONS AND WEIGHT ………………………………………………………………………………… 3
2.2
UNPACKAGING ………………………………………………………………………………………………………………….. 3
2.3
OVERALL DIMENSIONS ………………………………………………………………………………………………………. 4
2.4
USABLE SPACE IN THE CHAMBER ……………………………………………………………………………………… 4
2.5
SAFETY FEATURES ……………………………………………………………………………………………………………. 5
2.6
PRECAUTIONS …………………………………………………………………………………………………………………… 6
2.7
FRONT AND REAR VIEW …………………………………………………………………………………………………….. 7
2.8
STANDARD ACCESSORIES ………………………………………………………………………………………………… 8
2.9
TECHNICAL SPECIFICATIONS …………………………………………………………………………………………….. 8
2.9.1
Environment operating conditions …………………………………………………………………………… 8
3.
INSTALLATION ………………………………………………………………………………………………………………………….. 9
3.1
BASIC REQUIREMENTS ……………………………………………………………………………………………………… 9
3.2
GETTING STARTED ………………………………………………………………………………………………………….. 10
3.3
NOTES ON ALTITUDE COMPENSATION …………………………………………………………………………….. 11
4.
OPERATING INSTRUCTIONS ……………………………………………………………………………………………………. 12
4.1
FRONT PANEL COMMAND/SIGNALING ………………………………………………………………………………. 12
4.2
RUNNING A STERILIZAZION CYCLE ………………………………………………………………………………….. 13
4.2.1
The available sterilization cycle …………………………………………………………………………….. 13
4.2.2
Starting a sterilization cycle ………………………………………………………………………………….. 14
4.3
STOPPING THE CYCLE …………………………………………………………………………………………………….. 15
4.4
TOPPING UP AND DRAINING THE TANKS ………………………………………………………………………….. 16
4.4.1
Topping up the main tank …………………………………………………………………………………….. 16
4.4.2
Draining the used water tank ………………………………………………………………………………… 16
5.
PROGRAMMING ………………………………………………………………………………………………………………………. 17
5.1
DATE AND TIME ……………………………………………………………………………………………………………….. 17
5.2
MEASUREMENT UNIT AND OPTIONS ………………………………………………………………………………… 17
5.3
SPECIAL CYCLE S5 ………………………………………………………………………………………………………….. 18
6.
MAINTENANCE ………………………………………………………………………………………………………………………… 19
6.1
AUTOMATIC PERIODIC CLEANING CYCLE ………………………………………………………………………… 19
6.2
CLEANING THE INSTRUMENTS BEFORE THE STERILIZATION …………………………………………… 19
6.3
CLEANING/REPLACING THE WATER INLET FILTER …………………………………………………………… 20
6.4
REGULAR STERILITY TESTS …………………………………………………………………………………………….. 20
6.4.1
Bowie & Dick test ………………………………………………………………………………………………… 20
6.4.2
Vacuum Test ………………………………………………………………………………………………………. 20
7.
TROUBLESHOOTING ……………………………………………………………………………………………………………….. 21
7.1
MANUAL DIAGNOSIS ………………………………………………………………………………………………………… 21
7.2
INITIAL AUTO-TEST ………………………………………………………………………………………………………….. 22
7.2.1
Water quality check …………………………………………………………………………………………….. 22
8.
ALARMS ………………………………………………………………………………………………………………………………….. 23
8.1
GENERAL …………………………………………………………………………………………………………………………. 23
8.2
WARNING MESSAGES ……………………………………………………………………………………………………… 23
8.3
PRE-WARNING ALARMS …………………………………………………………………………………………………… 24
8.4
ABORTED CYCLE ALARMS ……………………………………………………………………………………………….. 24
8.5
CLASS B ADDITIONAL ALARMS …………………………………………………………………………………………. 25
9.
CONNECTIONS ………………………………………………………………………………………………………………………… 26
9.1
CONNECTION TO AN EXTERNAL PRINTER………………………………………………………………………… 26
9.2
INTEGRATED PRINTER (OPTION) ……………………………………………………………………………………… 26
9.3
CONNECTION TO A COMPUTER (ONLY FOR SERVICE) ……………………………………………………… 27
10.
OPERATOR’S NOTES ………………………………………………………………………………………………………………. 28
APPENDIX: SERVICE BOOK
DOMINA PLUS B 1
2
DentalX
1. GENERAL
1.1 INTRODUCTION
Object of this manual is to supply instructions for the operators in order to allow:
correct installation
right use
proper maintenance of the sterilizer
The machine must be installed and operated according to the procedures described in this manual.
The user is responsible for what concerns the fulfillment in the legal subject concerning installation and operation of the sterilizer.
If the machine is not correctly installed and operated or the appropriate maintenance is not carried out, the manufacturer cannot be considered responsible for any possible breaks and malfunctions.
Please, check for the packing integrity and no evident damages or missing parts (see delivery note).
IN CASE OF DAMAGES OR MISSING PARTS, PLEASE IMMEDIATELY INFORM AND IN DETAIL THE
FORWARDER, DENTAL X AND ITS AREA AGENT.
1.2 CONFORMITY TO EUROPEAN DIRECTIVES
DOMINA PLUS B is produced by Dental X and complies with the electromagnetic compatibility standards in conformity with the Medical Device Directive 93/42/CEE and with the norm EN 13060.
This equipment has been developed and manufactured using high quality material and parts that can be recycled and reused.
This symbol means that electrical and electronic equipment, at the end of their lifetime, must be disposed separately from the household waste. Dispose this unit by carrying it to the local collection/disposal centre. Sanctions are applied in case the regulations concerning waste disposal are not met.
Help us to preserve the environment in which we live!
0120
Mark CE 0120 applied on the rear panel points out the conformity with the Directive
93/42/CEE and warrants the customer that the equipment is safe and according with the international standards .
DOMINA PLUS B
DentalX
DOMINA PLUS B
2. FAMILIARIZATION
2.1 PACKAGE DIMENSIONS AND WEIGHT
Total weight: 67 Kg
Store the package for future shipment.
2.2 UNPACKAGING
3
DentalX
2.3 OVERALL DIMENSIONS
Net weight
Full load weight
55 Kg
67 Kg
2.4 USABLE SPACE IN THE CHAMBER
Diameter:
Depth
Capacity
240 mm
384 mm
17,5 liters
Useful dimensions per tray: 315 x 214 mm (x 2),
Useful tray volume :
315 x 168 mm (x 2)
10 liters
4
DOMINA PLUS B
DentalX
2.5 SAFETY FEATURES
The sterilizer features several devices that assure the full safety for of operators.
Door with dual lock control
An electromechanical device allows the door to be opened only if the following conditions are met:
— unit power supplied and turned on
— no current alarms
— internal pressure not dangerous for the operator
For further safety, to unblock the door at the end of cycle or in case of alarm it is necessary to press the Start/Stop button.
If the equipment is turned off with open door, do not apply force on the handle to try to
close the door. To close it, the equipment must be on.
Protection against overpressure – safety valve and decompression valve
Safety valve — it takes over when the pressure inside the chamber exceeds the value of 2,55 bar. To verify the efficiency of the valve, when the unit is cold and switched off, unscrew the black cap of the valve, when a “click” is heard pull it gently : you have to feel that the cap is free to move. The valve can not be adjusted or serviced for safety reasons. Follow the maintenance program described in the appendix SERVICE BOOK to assure the safety of the unit.
Decompression valve – it takes over when the pressure inside the camera exceeds the value of 2,4 bar; an acoustic signal warns the operator and the display shows the message ALARM
10.
Overheating protection
The temperature inside the chamber is programmed so as not to exceed the limit of 142 °C; in case of fault, a further protection takes over to prevent the temperature from rising over 150 °C.
Blackout protection
In case of power blackout during the sterilization cycle, the pressure inside the chamber is completely released to the environment value. At the restore of the power supply, the display will show the message BLACK OUT.
Automatic switch off
Elapsed 30 minutes from the end of the cycle without opening the door or without activating a front panel button, the unit will switch off automatically.
This function is not operating if no sterilization cycle has been carried out and completed
.
DOMINA PLUS B 5
6
DentalX
2.6 PRECAUTIONS
The European norms concerning safety and sterilization process defines the following figures:
OPERATOR: the person operating the unit to achieve the expected result.
RESPONSIBLE AUTHORITY: person or group responsible for the use and maintenance of the unit, he or she also has to make sure that.
— all personnel who operate or maintain the equipment are trained in its operation and in its safe use.
— there is regular training of all personnel concerned with the operation and maintenance of the equipment, including emergency procedures for any toxic, flammable, explosive or pathogenic material released into the environment.
— records of attendance at training are maintained, and evidence of understanding demonstrated.
The purpose of this manual is to provide suitable use instructions for both figures: however it does not give instructions concerning the STERILIZATION PROCEDURE and the cautions to be followed to prevent contamination of instruments and/or personnel using the unit that is a assignment of the RESPONSIBLE AUTHORITY of the practice
unit.
We wish to point out the following risks:
— The sterilization is a process that works by means of water steam under pressure and high temperatures; when the load is removed from the sterilization chamber always use tools and wear personal protections suitable for handling hot instruments.
— On opening the sterilizer, especially in case the cycle has been aborted, a small quantity of hot water steam or condensate can be released in the environment; be careful when opening the door.
— If the condensation cycle is not brought to an end, the load, the trays and the tray support as well as the chamber inner space are ALWAYS to be considered as potentially contaminating elements, as long as a subsequent sterilization cycle has not been successfully completed.
— The water contained in the recovery tank is to be regarded as biologically contaminating, therefore when this tank is emptied, suitable precautions should be taken.
The disposal of recovery water needs to be done in accordance with the national or local regulation.
its use.
Check the integrity of the draining pipe before
— To prevent cross contaminations during the loading and unloading steps, open the door with clean hands or wear uncontaminated gloves to avoid contaminating the door handle, do not use gloves worn during the instruments decontamination step for this operation; when the sterilized instruments are removed from the chamber, always use uncontaminated gloves.
—
In case of contact with hot water, steam or contaminated materials rinse with fresh water and seek for medical help.
SYMBOLS
On the panels of the unit and in this manual, potential hazards and the parts that can be dangerous at high temperatures, are marked with danger symbols.
WARNING: instruments and chamber are very hot
Contamination risk
CAUTION, risk of danger
Documentation needs to be consulted
Read carefully this user manual because wrong use may expose the user to health risks.
This symbol indicates the presence of additional important notes about the use
The water steam sterilizer is designed to be used for the sterilization of reusable medical instruments that can be steam sterilized in a range of temperatures between 121°C and 135°C; any attempt of sterilizing instruments that are not fit for undergoing this process can result in hazard for the operator: it can also lead to potentially serious faults and damage the sterilizer‟s safety mechanisms.
The unit is not to be used to sterilize liquids and flammable materials.
The unit is designed for indoor use only.
Do not use in presence of anesthetic or flammable gas.
To avoid an excessive level of humidity, properly air the room where the unit is installed.
DOMINA PLUS B
2.7 FRONT AND REAR VIEW
Bacterial filter
Door
Safety valve connection
Distilled water tank top filling point
Door handle
Control panel
Door gasket
Temperature sensor
Air line filter
On-board printer (option)
Water inlet filter
Vent hole for used water tank
Automatic draining of used water tank (option) : max flow 20 ml/minute, water T < 70°C
Chamber water inlet
PC interface
Quick connector for distilled water tank drain
Quick connector for used water tank drain
Water outflow filter
Safety valve
DentalX
Condenser/Cooler
External printer interface
Mains fuses
Power supply switch
Power supply cord
DOMINA PLUS B 7
DentalX
2.8 STANDARD ACCESSORIES
Filter spanner
(P/N 105228)
Water filter
(P/N 105320)
Drain tube
(P/N 101041A)
Tray rack (P/N 105078)
Tray holder
(PN/ 105609)
Cleaning tabs
8
Warranty
Bacterial filter
(P/N 021008)
User manual
2+2 Trays
(P/N 105076 — 105077)
the agent, to the manufacturer; for want of this the warranty will decline.
To make the warranty active it is necessary that a copy of the supplied Unit Passport is sent, through
2.9 TECHNICAL SPECIFICATIONS
Chamber dimensions
Chamber capacity
Ø = 240 mm
Depth = 384 mm
17,5 l
Auto-switching-off
Dual water tank elapsed 30‟ from the end of a cycle and without any action
4 liter each (used and clean water tank)
Maximum load
Warming-up time
Sterilization time
Drying time
External dimensions
Net weight
Power supply voltage
Frequency
Max consumption
Average consumption
Standby consumption
AC fuses
4 kg (solid load)
1,5 kg (porous load)
20‟ from room temperature
10‟ with pre-heated chamber from 3‟ to 90‟ depending on the selected cycle from 3‟ to 14‟ depending on the selected cycle
443 x 590 x 428 mm
(L x D x H)
55 Kg
200 – 250 Vac
50/60 Hz
1920 W
1000 W
1 W
2 x 12,5A T (type 6.3 x 32 CT ) —
IEC 127
2.9.1 Environment operating conditions
Vacuum pump
Bacterial filter
20 l/ minute- 0.97 bar
0.3 µm al 99.97 %
Differential heating system — SDR
Transmitted heat in environment at 23°C: 0,22 kJ/h
Sound emission: 52 dB/A at 1 m
Working cycle: continuous
Pollution grade: 2
Transient overvoltage protection: II
Monitoring of the water conductivity
Maximum volume available on the trays: 10 l
Maximum chamber temperature: 135°C (-0/+2°C)
Safety valve working pressure : 2,55 bar
The sterilizer is designed to operate in environments with temperatures between 3 °C and 40 °C, relative humidity not greater than 95%, atmospheric pressure from 750 mBar to 1050 mBar and altitude between
0 and +2500 meters.
NEVER USE IN PRESENCE OF INFLAMMABILE ANESTHETIC GAS
DOMINA PLUS B
DentalX
3. INSTALLATION
3.1 BASIC REQUIREMENTS
1. Check that the mains voltage of your electrical installation matches the value indicated on the equipment plate, the electrical socket is capable to supply at least 10A and is provided with an earth connection. In case the installation makes inaccessible the power supply switch, provides for a proper electrical breaker
The manufacturer will not be responsible for damages to people or things caused by an unsuitable electrical installation or missing of the earth connection.
2. The equipment must be installed on a flat surface; adjust the front feet to have a slight slope and so make the outflow of water easier during the drain phase.
WARNING: do not place the unit on surfaces which could cause fire or fume if hot items fall from the equipment.
3. For the correct operation, it is mandatory to leave a free space of at least 4 cm on the rear side of the equipment.
4. Do not install the equipment near heat sources, in humid or not well aired environment; in a sterilization room is required a minimum of 10 air changes per hour, a recirculating ventilation system can not be used.
5. On the rear panel it is located the safety valve, if it operates for overpressure it may be released hot steam: locate the unit to eliminate the risk of burning for the operator (ex. near a wall).
The sterilizer can also be installed in a cabinet. In this case, provide adequate space (>10 cm) for aeration.
200/250Vac
I> or = 10A
>4 cm
1
2
F
S
DOMINA PLUS B 9
DentalX
3.2 GETTING STARTED
These operations must be carried out only by qualified personnel, wrong settings could affect the quality of the sterilization
.
Check the electrical requirements and connect the power supply cable to the mains socket.
1. The equipment is delivered with empty tank and is therefore necessary to fill it completely with demineralized water. Fill the tank without exceeding the maximum level reference marked on the tank opening.
The use of low quality demineralized water could create scale deposits on the instruments, inside the chamber and on the trays. Carefully read the label of the demineralized water container. Do not use tap water even if treated with filters or sweet-
eners.
Do not use water for batteries, other liquids or additives
which can cause irreversible damage to the equipment and
be a risk for the operator.
2. Turn on the equipment with the main switch on the back panel.
This switch can be left ON since the electrical consumption in stand-by is almost nil.
3. Remove the basket and the trays from the chamber and close the door.
When the unit is off, the door is blocked; if the block persists, turn
off and then on again the unit.
4. Hold on the button value (100 m).
1
and push the key
Power
; the display shows the message <SET ALT 100 MT> with the set altitude
Use both keys
1
and
2
to adjust the value as necessary according to the actual site altitude (see next page).
Then push the key
Set
to confirm the displayed value and start the automatic initialization procedure with the entry of water in the hydraulic circuit and chamber.
10
DOMINA PLUS B
DentalX
5. At the end of the initialization procedure the Led READY turns on; open the door and dry the chamber with a clean cloth.
In case of procedure steps not correctly followed, the display will show one of the following warning messages:
DOOR OPEN:
ADD H2O:
NEED INST: door not closed lack of water initialization procedure not carried out
In this case, repeat the procedure.
With procedure already been performed, the display will show OFF and the door stays blocked. To
unblock it, push the button
Power
.
The sterilizer is ready for use.
Arrange basket and trays in the chamber and select the sterilization cycle.
See Chapter 4 «OPERATING INSTRUCTIONS».
3.3 NOTES ON ALTITUDE COMPENSATION
For a proper operation of the pressure control devices, an altitude compensation feature has been introduced.
During the installation procedure it is necessary to set the altitude value (referred to sea level) for the site where the unit operates. This procedure must be carried out every time the unit is moved to a site with different altitude from the one previously set.
During the factory test the equipment is set at a default value of 100 meters and can be left unchanged for real altitude values between 0 and 200 meters, since a ±100 meter error does not affect the equipment operation.
To be sure of the sterilization process, it is important that the altitude tolerance from the current value does not exceed 200 meters; otherwise, the vacuum devices should be additional loaded, and false or premature AL8 or AL5 alarm should be signaled.
CONVERSION NOTE: to obtain meters multiply feet by 0.3048.
These operations can be carried out by qualified personnel. Wrong settings can affect the quality of the sterilization
.
DOMINA PLUS B 11
DentalX
4. OPERATING INSTRUCTIONS
4.1 FRONT PANEL COMMAND/SIGNALING
The front panel is equipped with control keys, signaling Led‟s and displays. A slight push on a key will activate the command.
LCD Displays: visualize (from the top) the value of the parameters Time (Led “Time” turned on), Temperature (measure unit: °C or °F depending on the setting, related Led turned on) and Pressure (measure unit: bar or psi depending on the setting, related Led turned on); the upper display visualize, instead of time, possible alarms occurred during the cycle (Led
“ Alarm” turned on).
Phase in progress indication: VACUUM – STERILIZE – DRY
– READY — turned on or flashing during the phases of the cycle.
Clean water tank level indication: turned on for maximum or minimum level.
Used water tank level indication: turned on for maximum level.
Current program indication: 1 – 2 – F – S – BOWIE & DICK
— turned on for the selected program.
Program selection keys :
1
Program 1 — Sterilization at 134°C, 5 min., 3 vacuum pulses: for wrapped and unwrapped solid instruments
2
Program 2 — Sterilization at 121°C, 20min., 3 vacuum pulses: for porous instruments and fabrics
F
Program 3 — Fast sterilization at 134°C, 3min., 1 vacuum pulse
S
Program 4 – Four preset cycles (S1, S2, S3 and S4) + 1 cycle (P5) programmable by the operator
Key
Start/Stop
: controls the start or the stop of the selected cycle, unlocks the door at the end of the cycle or in case of alarm occurred during the cycle.
Key
Power
: enables the command panel, the auto-test at the switching-on and the pre-heating resistance.
Key
Set
: allows to set: current date/time, measuring unit, report printout language, temperature/process time/ vacuum pulse number for the programmable cycle.
Key
Test
: allows to carry out the Bowie & Dick Test with sterilizer normally turned on, or the Vacuum
Test with sterilizer in stand-by and chamber temperature lower than 35°C.
12
DOMINA PLUS B
DentalX
4.2 RUNNING A STERILIZAZION CYCLE
1. Turn on the equipment by means of the rear switch.
Display TIME is showing the current time
Display TEMP. is showing OFF
Display PRESS is showing the current day and month
2. Press the key
Power
and wait for a few seconds the initial auto-test completion; during this time the parameter set-points and the components currently tested will appear in a sequence on the display.
Over the auto-test, the display TIME will show again the current time, the display PRESS the value of the current pressure, the display
TEMP the value of the current chamber temperature (if lower than 35 °C the display will show the message “
low”). The microprocessor enables the pre-heating step in order to rise the chamber temperature up to 100 °C.
During this phase the temperature reading on the display is inaccurate, because no steam
being there.
3. Arrange the material to be sterilized on the trays, load the chamber and close the door.
4. Check that Led H2O MIN (main tank signaling) is off. If not, fill up the tank with demineralized water up to the lighting of Led H2O MAX.
4.2.1 The available sterilization cycle
Key
1
2
F
Parameters
134°C, 5„ three pre-vacuum pulses, drying 10‟ (6‟ with vacuum + 4‟ with ventilation)
121°C, 20‟ three pre-vacuum pulses, drying 11‟ (7‟ with vacuum + 4‟ ventilation)
134°C, 3‟ single pre-vacuum pulse drying 3‟ (2‟ with vacuum + 1‟ ventilation)
Load type (*)
Solid, porous, hollows type A and B, wrapped
(ref. EN868)
Solid instruments, unwrapped
S
Cycle Parameters
B
B
S
134-137°C
2,04-2,25 bar
121-124°C
1,04-1,24bar
134-137°C
2,12-2,30 bar
Solid, porous, hollows type A and B
Solid, porous, hollows type A and B, wrapped
(ref. EN868)
Disinfection
105-108°C
0,21-0,35 bar
B
B
B
134-137°C
2,04-2,25 bar
121-124°C
1,04-1,24bar
135-138°C
2,12-2,30 bar
S5
Test B&D
Parameters to be set and range:
Temp.: 105 – 135°C
Process Time: 3‟ – 90‟
Pre-vacuum pulses: 1, 3 or 4
Drying (vacuum + ventilation):
3‟+2‟, 6‟+4‟, 8‟+6‟
134°C, 3,5‟ three pre-vacuum pulses drying 10‟ (6‟ with vacuum + 4‟ ventilation)
Depends on the parameters set
Test B&D (3M
COMPLY
TM cod. 1300)
TM
TEST
105-138°C
0,21-2,30 bar
134-137°C
2,04-2,25 bar
Max load
4Kg solid or 1,5kg porous or a combination of both in proportion.
4Kg solid
4Kg solid or 1,5kg porous or a combination of both in proportion.
B&D test pack or equivalent
Vacuum test Temperature under 35°C TEST < 35°C Empty chamber
Check test
Helix test
EN13060 Par 10.6
Helix test
EN13060 Par 10.6
Solid load, unwrapped
EN13060 Par 10.5
NA not a sterilization cycle
Helix test
EN13060 Par 10.6
Helix test
EN13060 Par 10.6
Helix test
EN13060 Par 10.6
To be defined by the operator
NA
Not a sterilization cycle
NA
Not a sterilization cycle
(*) Liquids excluded
DOMINA PLUS B 13
DentalX
To select the programs
1 , 2 or F press the relating key
.
To select one of the option
S programs, hold down the key S and press the key 1 or 2 to visualize and select on the lower LCD display the choice available (S1, S2, S3, S4 or S5).
The indication Led of the selected program turns on, and the displays will show for 5 seconds the parameters of the cycle currently selected.
4.2.2 Starting a sterilization cycle
Press the key START/STOP to start the selected cycle.
The programs F, S1 and S5 do not warrant the Class B sterilization; to start these types of cycle,
hold down the key
F
(or
S
) and push the key
Start/Stop
.
The door is locked and stays locked throughout the cycle duration.
The parameters of the selected cycle are shown once again for 10 seconds, then the sterilizer starts and runs the cycle phases automatically. The various steps of the cycle are microprocessor controlled and sequentially shown on the display; in this way the operator can monitor the progress of the sterilization phases and times.
The signaling for the various phases of the cycle are reported below:
Led VACUUM turned on
Display TIME starts to record the cycle duration
Display PRESS shows the chamber pressure value
Display TEMP. shows the chamber temperature value
The program indication Led (1, 2, 3 or 4) starts to flash.
Vacuum phase (water entry in the chamber and pre-vacuum pulses)
During this phase the microprocessor enables the vacuum pump and enters a water dose in the chamber. Led VACUUM is flashing. This phase will be repeated more times and should require 10 to 20 minutes depending on the chamber conditions and load type to be processed. Pump operation may be slightly noisy.
For unwrapped solid instruments we recommend to use the cycle F. In this way the sterilization time will be faster, and power consumption reduced
.
Sterilize phase
Reached the preset parameter values, Led
VACUUM turns off and Led STERILIZE turns on. The display Time starts the countdown marking the time remaining to the end of the sterilization process, the Pressure and Temperature displays show respectively the P and T values of the steam.
The sterilization phase is followed by the decompression phase, with the display Pressure showing the decreasing pressure values down to 0. Again, the display Time will start the countdown of the decompression phase. Based on the our experiences, the decompression time has been slightly extended in order to minimize the thermal shock consequent to the status change of the steam.
Drying phase
Over the decompression phase, Led STERILIZE starts to flash to signal the completion of the sterilization process. At the same time the Led DRY turns on, signaling the start of the drying phase.
Throughout this phase, the chamber heaters keep the chamber warm according to a microprocessor-controlled logic, the vacuum pump comes again into operation to eliminate the residual steam.
The display Time shows the countdown of this phase. Follows the forced ventilation phase through the bacterial filter – the display Time shows also the countdown of this phase.
14
DOMINA PLUS B
DentalX
End of cycle
As soon as the drying is over, Led
DRY turns off, and Led READY and STERILIZE turn on. A 10second alert signal is generated to draw the operator attention. The chamber heaters are set at reduced power (pre-heating) until the door is open. The display Time shows the total time of the cycle, the Pressure and Temperature displays show respectively the current P and T of the chamber.
At the end of cycle F, S1 or S5 only the Led READY will light and not STERILIZE to signal that the cycle selected by the operator does not warrant a Class B sterilization; the display will show the cycle counter.
To unlock the door before opening it, press the key
Start/Stop
.
The cycle is over and the load can be taken out.
ATTENTION: instrument and chambers are very hot
Opening the door, the displays will show again the current time, chamber temperature and pressure, and the sterilizer is ready for a new cycle.
In case of on-board printer or external printer connected, a report will be issued during the cycle phases with the more significant data; the report can be filed as proof of the sterilization process performed.
The operator can arrange other load on the trays and start a new sterilization cycle. with the advantage of shorter heating-up time as the chamber is already warm, or press key Power to put the unit in stand-by status (OFF on the display).
If the door is not opened or a key pressed within 30 minutes, the unit switches to stand-by status
(OFF) automatically
.
Should any failure or error occur during the cycle, Led
ALARM turns on, the display Time will show the type of alarm (see chapter ALARM) and the door remains locked. To unlock the door, press key
Start/Stop
.
ATTENTION: instrument and chambers are very hot
Contamination risk
4.3 STOPPING THE CYCLE
To stop the sterilization cycle, press key
Start/Stop
. The display Time shows the message “
MANU
STOP”.
Before opening the door, make sure that the display Pressure is showing the value 0. A safety device will anyway prevent from opening the door if the chamber is over-pressurized.. To unlock the door, press key
Start/Stop
.
Remove the load and check for the presence of water into the chamber. In case of wrapped instruments, we suggest to replace with new bags.
Before loading the chamber again, dry it carefully and wait 10 minutes to allow the water to evaporate and be drained completely.
DOMINA PLUS B 15
DentalX
4.4 TOPPING UP AND DRAINING THE TANKS
The sterilizer is fitted with two 4-liters tanks: main tank for the clean demineralized water and recovery tank for the used water.
The hydraulic system does not reuse the condensate obtained during the sterilization process; this condensate is collected in the recovery tank which must be periodically drained. This mode of operation involves the progressive main tank emptying and recovery tank filling.
4.4.1 Topping up the main tank
The average water consumption per sterilization cycle is 520 cc, hence 7 cycles can be performed with a full tank .
The of the Led H2O MIN signals a water level into the main tank insufficient to perform a new cycle.
Provides for the topping up of the main tank, taking care to not exceed the grid on the entrance. The lighting of the Led H2O MAX and a acoustic warning signals that the tank is full.
MAX
MIN
4.4.2 Draining the used water tank
The lighting of the Led H2O MAX relative to the used water tank warns that the maximum level has been reached. In this case :
MAX
Get a bucket or a tank of at least 4 l capacity,
Fix the supply drain tube into the left fast fitting (gray),
Wait for a complete draining,
Unfit the tube pushing the ring nut against the machine and drawing the tube.
CAUTION! The water contained in the used water tank is to be regarded as biologically contaminating, therefore when this tank is emptied, suitable precautions should be taken.
The disposal of recovery water needs to be done in accordance with the national or local regulation.
16
DOMINA PLUS B
DentalX
5. PROGRAMMING
5.1 DATE AND TIME
Push the key
Set
and use the keys
1
and
2
to adjust the value. Whenever key
Set
is pushed, a different time parameter can be controlled:
PRESS IN A
SEQUENCE
SET
SET
SET
SET
SET
SET
DISPLAY
TIME
SET YEAR
PARAMETER
TO BE SET
YEAR
SET MONTH MONTH
SET DAY
SET HOUR
DAY
HOUR
SET MIN MINUTES
Exit the programming mode
USE KEY
1
to increase the value, or
2
to decrease the value
Example: for hour adjustment, push key
Set
four times and set the time by the keys
1
and
2
5.2 MEASUREMENT UNIT AND OPTIONS
Push in a sequence the keys
Set
access the programming mode
Press key and
Set
F
to
Display PRESS is showing:
SET UNIT °C or
SET UNIT F
Display TIME is showing:
L1
L2
L3
L4
L5
Display PRESS is showing:
SET UNIT BAR or
SET UNIT PSI
Push the key
1
to set the desired temperature measurement unit
Push more times the key
2
:
L1 = Italian
L2 = English
L3 = German
L4 = French
L5 = Spanish
Push the key to set the language of the print report
1
to set the desired pressure measurement unit
Push again the key
Set
to exit the programming mode
The sterilizers are generally factory preset on measurement units and options normally used in the destination countries (i.e. measurement units: °C, bar / printout language: L2).
DOMINA PLUS B 17
DentalX
5.3 SPECIAL CYCLE S5
The operator can set a customized sterilization cycle as follows:
Push the keys
Set
and
S
in a sequence
Display PRESS is showing:
SET TEMP
Set the process temperature value between 105 and 135 °C by the keys
1
and
2
Push again the key
Set
Display PRESS is showing:
Set the process time between 3 and 90 minutes by the keys
1
and
2
Set the number of vacuum pulses (1, 3 or 4) by key
1
; the value is shown close to the message VAC.
Push again the key
Set
Display PRESS is showing:
VAC or
DRY
Set the duration of the drying phase (vacuum + ventilation = 3+2, 6+4 or 8+6 minutes) by the key
2
DRY.
; the value is shown close to the message
Push again the key
Set
to exit the programming mode.
The parameter values for the special cycle S5 are automatically stored and maintained until new values are set through the same procedure.
The combination of time/temperature parameters set by the operator can lead to cycles that do no assure the sterilization. It is necessary to check the effectiveness of the customized cycle by means of suitable tests.
At the end of the special cycle S5, only the Led READY will switch on — not STERILIZE — to signal that the efficiency of the cycle selected by the operator has not been validated by the manufacturer.
18
DOMINA PLUS B
DentalX
6. MAINTENANCE
6.1 AUTOMATIC PERIODIC CLEANING CYCLE
For a proper operation of your sterilizer it is indispensable that a correct and regular maintenance is carried out. To this end, it is important to perform a cleaning procedure, as described on the following, at least once every 15 days or, in case of intense usage, every two water full of the main tank.
For further safety, after 60 cycles carried out without an intermediate cleaning cycle, the equipment will display the message <NEED CLEANING
>
.
IMPORTANT Take the basket and the trays out of the chamber and wash them with an ordinary dish washing powder, rinse with water and wipe.
WARNING
NOTE
DO NOT USE ABRASIVE PRODUCTS.
DO NOT PERFORM THE CLEANING CYCLE WITH
TRAYS IN THE CHAMBER.
Carry out the chamber surface cleaning when the equipment is cold.
To open the door and carry out these preliminary operations, it is necessary to switch on the equipment; ended the maintenance, switch off the unit to avoid excessive heating. To close the door switch on the unit again.
1. Put a cleaning tab in the chamber and close the door
THE CHAMBER MUST BE EMPTY !
2. Set the equipment in stand-by mode by pushing the key
Power
(
OFF on the display).
3. Hold down the key
Start
and push the key
Power
to start the automatic cleaning cycle; this cycle takes approx. 15 minutes.
4. At the end of the cycle, the Led READY will turn on; open the door and wipe the chamber with a clean cloth slightly soaked with demineralized water and pure alcohol.
Do not use spong-
es, brushes, abrasive steel wool or paper.
Thank to the electronic control system, the number of maintenance cycles actually performed is continuously recorded and updated.
The missing of appropriate and regular maintenance according to the above guidelines could cause an early and more frequent service activities and the
lapse of the warran-
ty.
6.2 CLEANING THE INSTRUMENTS BEFORE THE STERILIZATION
In order to extend the sterilizer life, we recommend to carry out an accurate cleaning of instruments; one of the main causes of an early equipment wear is the settlement and accumulation of debris and fragments for inadequately instrument cleaning, and consequent stains, fouling and progressive clogging of filters, electro-valves and tubing.
DOMINA PLUS B 19
DentalX
6.3 CLEANING/REPLACING THE WATER INLET FILTER
To carry out the clean or the replacement of the filter on the left of the front panel, proceed as follows:
1. Drain completely the main tank by plugging the supply tube (unthreaded end) into the right fast fitting (white) of the front panel.
2. Use a coin to unscrew the cap (1) on the filter seat; pay attention for possible water outflow from the internal tubes. Unscrew the filter holder (2) using the supply spanner.
3. Clean the filter (3) by compressed air (or ultrasonic cleaner) or replace it if damaged.
4. Remount filter, filter holder and cap; screw down by hand without tightening excessively.
5. Fill the main tank with demineralized water as for the standard operation.
6. Switch on and put the equipment in stand-by mode ( OFF on the display), hold down the key
2
and push the key
Power
.
The equipment starts an automatic initialization procedure, with the exhausting of the residual air out the filter. The end of the procedure is signaled by the lighting of the Led READY.
6.4 REGULAR STERILITY TESTS
1
2
3
During the factory tests and according to the standing regulations, the sterilizers are deeply tested and the calibrations checked; these tests warrant the sterilizer performances except for unauthorized interventions, tampering or an improper usage. Although the unit is equipped with an advanced diagnosis and process evaluation system, is user responsibility of checking periodically the performance in the course of the equipment life. The frequency of these checks is regulated at local country level; check the regulations in force.
For any explanations or information, contact your authorized reseller or directly Dental X.
6.4.1 Bowie & Dick test
The test can be performed at any time with the sterilizer turned on and operating.
— Load the sterilization chamber with a B&D test (i.e. 3M
TM standards for the test procedure.
COMPLY
TM
cod. 1300) according to the
— Hold down the key
Test
and push the key
Start/Stop
.
The test cycle starts with cycle phases characterized by 3 pre-vacuum pulses, process temperature of
134°C and process time of 3,5 min.
Simulators of the B&D test are available on the shelf, developed for big sterilizers and equivalent to a texture load of 7Kg; it is clear that the chamber dimensions do not allow such a load, but a positive result of this test is indication of efficient sterilizer features.
6.4.2 Vacuum Test
We suggest to carry out the test at the beginning of the working day in order to check the chamber
and system tightness.
The test should be started with equipment in stand-by mode ( OFF on the display) and chamber temperature lower than 35°C (normal conditions at the beginning of the working day).
—
—
Push the key
Test
The vacuum test starts automatically and takes about 15 minutes.
In case of negative result, the display Alarm/Time will show TEST FAIL to warn for insufficient tightness of the chamber (see Chapter 8 — Alarms).
20
DOMINA PLUS B
Domina
Plus
B
DentalX
7. TROUBLESHOOTING
7.1 MANUAL DIAGNOSIS
The operator or technician can perform at whenever time a test to check the correct operation of the equipment; proceed as follows:
Step 1
Action
Push the keys
Set
and
Test
in a sequence
Push the key
1
Push the key
2
Push the key
F
Push the key
S
Push the key
Test
Push the key
Power
Push the key
Set
Message/Signaling on the display
The displays show respectively (from the one at the top): message “ TEST”, chamber temperature value and pressure value
.
Temperature value of the chamber’s top wall
Temperature value of the chamber‟s bottom wall
Message “
CICL” and number of the performed cycles
Message “ ABOR” and number of the aborted cycles
Number of the automatic cleaning cycles actually performed
Message “ ALARM” and codes of the last three alarms occurred
Exit the diagnosis and return to the normal mode
During the manual diagnosis, it is impossible to set the unit in stand-by mode with the key POWER.
First you need exit the diagnosis mode by pushing the key
Set
The use of these procedures is reserved for qualified personnel
.
.
Step 2
Action
Push the keys
Set
and
Power
in a sequence
Push the key
1
Push the key
2
Push the key
F
Push the key
S
Push the key
Power
Push the key
Test
Push the key
Set
Message on the display / Result
Message “ TEST OUT”
Electro-valve 1 energized (open)
Electro-valve 2 energized (closed)
Electro-valve 3 energized (open), electro-valve 5 energized (closed), vacuum and water pumps energized
Electro-valve 4 energized (open)
Electro-valve 5 energized (closed)
Condenser‟s fan powered
Exit the diagnosis and return to the normal mode
DOMINA PLUS B 21
DentalX
7.2 INITIAL AUTO-TEST
Each time the equipment is switched on, starts an automatic test (duration approx. 15 s) checking sequentially any main component.
Three beeps signal the end of the auto-test and, if positively passed, the message Card Good is displayed .
Whatever fault detected, it will be shown on the display and stored with the alarm code of the Table C (see chapter ALARM).
To skip the auto-test, turn on the unit holding down any key.
7.2.1 Water quality check
To prevent the accidental use of poor quality demineralized water, the sterilizer is equipped with a device that monitors the water conductivity; this check starts at the switching-on (initial auto-test), provided that the unit is at environment temperature and with the main tank full.
At the end of the auto-test, the display will show the message « H2O good » or « H2O hard » if the measured conductivity value is respectively lower or higher than 15 µS.
ATTENTION
The negative result of the water quality check does not stop the operation of the sterilizer; anyway, it is strictly recommended to replace the water with a type of better quality.
The following table shows the minimum values suggested for the feeding water.
Polluting agents
evaporation residue silicon oxide iron cadmium lead heavy metal residues, included iron, cadmium and lead chlorine phosphate
Conductivity (at 20°C)
Ph appearance hardness
Feeding water
< 10mg/l
< 1 mg/l
< 0,2 mg/l
< 0,005 mg/l
< 0,05 mg/l
< 0,1 mg/l
< 2 mg/l
< 0,5 mg/l
< 15 uS from 5 to 7,5 colorless, clean, without sediment
< 0,02 mmol/l
Condensate
< 1,0 mg/l
< 0,1 mg/l
< 0,1 mg/l
< 0,005 mg/l
< 0,05 mg/l
< 0,1 mg/l
< 0,1 mg/l
< 0,1 mg/l
< 3 uS from 5 to 7 colorless, clean, without sediment
< 0,02 mmol/l
22
DOMINA PLUS B
DentalX
8. ALARMS
8.1 GENERAL
At equipment turned on and during every cycle, the internal supervising system constantly monitors the parameters characterizing the sterilization phases, the proper operation and the status of the main components. Any detected anomaly or fault is promptly signaled through specified messages, coded alarms and acoustic signaling.
To easy the interpretation and identification, the alarms have been divided into four classes, as shown in tables A, B, C and D.
8.2 WARNING MESSAGES
Table A reports the warning messages displayed.
TABLE A
Message
OPEN DOOR
FAIL
DRY FAIL
ADD H2O
FULL H2O
MANU STOP
Cause
Door not opened at the end of the cycle.
Start command entered with the door open.
Failed cycle
Drying phase not completed due to manual interference (the load has been removed before the drying cycle completion). However the sterilization process has been achieved.
Insufficient water in the main tank (the message appears before starting a cycle)
The used water tank is full (the message appears before starting a cycle)
Cycle manually interrupted. Sterilization process not completed
Open the door.
Close the door
See table C
Solution
Push the key
STOP.
Top up the main tank.
BLAC OUT
NEED CLEANING
NEED SERVICE
NEED INST
NEED TEST
TEST FAIL
Black-out occurred during the cycle
60 cycles carried out from the last automatic cleaning cycle
One year from the installation or over 1500 cycles performed without any service check-up
Need for the installation procedure.
Detected a pre-warning alarm
Negative result of the Vacuum Test
Drain the used water tank
Wipe the chamber ( if wet) and start again the cycle
Check the AC plug and socket.
Wipe the chamber and repeat the cycle.
Perform the automatic cleaning cycle (see
Chapter 6.1)
The warning message disappears as soon as a cycle is selected, but will appear again at the next switching on.
Call for a complete check-up by a qualified technical service; the message will be reset after the servicing.
Perform the installation procedure (see
Chapter 3.2)
See table B
Clean the door gasket and repeat the test.
Call for a technical service
DOMINA PLUS B 23
DentalX
8.3 PRE-WARNING ALARMS
The alarms shown on table B do not stop the operation of the sterilizer, but warn that a problem might interfere with the correct working of the sterilizer.
Check the trouble and promptly carry out the recommended action.
In case of fault, the message Need Test will appear along with the code number of the detected alarm.
Example: Need Test cd 1.
TABLE B
Alarm code cd 1
Drain filter dirty
Cause
Solution
Clean or replace the filter
cd 2 cd 3 cd 4 cd 5 cd 6 cd 7
Slow heating of the top wall of the chamber Perform a cycle with reduced load. In case, call for a technical service.
Check the mains voltage.
Slow heating of the bottom wall of the chamber
Water dose distributor clogged
Water inlet filter dirty
Water entry valve dirty
Perform a cycle with reduced load. In case, call for a technical service.
Check the mains voltage.
Impurities in the main tank .
Carry out the filter maintenance (see § 6.3).
Carry out the automatic cleaning cycle (see
6.1)
If the problem occurs more than 3 times, call for a technical service
Bacterial filter clogged
Vacuum phase too slow
Replace the filter
Wipe the chamber and perform the automatic cleaning cycle (see § 6.1).
8.4 ABORTED CYCLE ALARMS
The alarms shown on table C refer to faults keep the sterilization process from being completed.
Identify the fault on the table and carry out the recommended action.
The alarm condition is signaled by the Led ALARM, and the Alarm/Time display will show the message
FAIL (intermittent) along with the code number of the detected alarm. Example: FAIL AL 6.
TABLE C
Alarm code
AL 1
AL 2
AL 3
AL 4
AL 5
AL 6
AL 7
AL 8
AL 9
AL 10
AL 11
Cause
Electro-valve 1 faulty
Electro-valve 2 faulty
Electro-valve 3 faulty
Electro-valve 4 faulty
The pressure has not reached the set-point value within the preset time
Too long time of the initial vacuum phase
Door not properly locked
Air into the sterilization chamber
Stop of the countdown for over 60 sec. during the sterilization phase
Pressure too high
Pressure too low
Solution
Call for a technical service
Call for a technical service
Call for a technical service
Call for a technical service
Overload or pressure weak.
Perform the automatic cleaning cycle (§ 6.1)
Perform the automatic cleaning cycle ( § 6.1)
Make sure that the door is correctly closed.
Check the door tightness. Clean the gasket.
Check the door tightness. Perform, if needed, the automatic cleaning cycle (see § 6.1), and the Vacuum Test.
Call for a technical service.
Check the door tightness. Perform, if needed,
24
DOMINA PLUS B
DentalX
Alarm code
AL 12
Cause
Temperature out the normal range
Solution
the automatic cleaning cycle (see § 6.1).
Perform the vacuum test
Perform the automatic cleaning cycle (§ 6.1).
Call for a technical service
AL 13
AL 14
AL 15
Chamber‟s temperature sensor faulty
Chamber‟s top wall temperature sensor faulty Call for a technical service
Chamber‟s bottom wall temperature sensor faulty
Call for a technical service
Pressure sensor faulty
Call for a technical service
AL 16
8.5 CLASS B ADDITIONAL ALARMS
Alarm code
18
31
Phase involved
Drying phase
Drying phase
TABLE D
Cause
Drying interrupted Dry the load
Vacuum not sufficient Excess of load
Solution
NOTE: The Class B alarms may occur only for the programs 1, 2, S2, S3 and S4.
DOMINA PLUS B 25
DentalX
9. CONNECTIONS
9.1 CONNECTION TO AN EXTERNAL PRINTER
The equipment not supplied with an internal printer, is provided with serial interface for the connection to an external printer the process data are sent to, in order to document and certify the sterilization process performed. The use of a printer, mandatory in some countries, will become more and more frequent with the growing need to document the effective sterilization of the instruments, also for its forensic and medical aspects.
The printer cable cannot exceed 3 meters.
The serial port can be connected to any DOS compatible, standard CEN-
TRONIX printer. Most printers support these standards. Please contact
Dental X for further information.
1. Switch-on the printer,
2. Switch-on the sterilizer.
The report is automatically printed on the course of the cycle and with the following data content: date and time of the process – progressive cycle number — selected program and parameters — cycle type: sterilization or disinfecting – start/end time of the sterilization phase — end time of the drying phase
In case of trouble or cycle interruption, the printout will report the message
ABORTED CYCLE — NOT
STERILE along with the indication of the detected alarm.
Remember to switch off the printer at the end of the working day.
To set the language for the report printout see the chapter 5.
The PRINTER port interfaces only a printer directly. Through a dedicated interface, this port can be also con-
nected to a PC for storing the file sterilization cycles. Contact the reseller or directly Dental X for more infor-
mation.
9.2 INTEGRATED PRINTER (OPTION)
The model with a thermal printer does not require an additional installation for the report printing. The printer‟s Led POWER indicates the correct power supply, while the Led
ERROR indicates a lack of paper.
Replacing the paper roll
Open the door by leveraging on the tab at the top.
Remove the empty roll and plug in the new, taking care to put upwards the sensitive side of the paper (see figure).
Close the door, leaving about 5 cm of the outside
Use original chemical paper, 57 mm wide and 40 mm diameter.
The report is printed automatically on the course of the cycle as for the external printer. If report printout is not required, leave open the printer door before starting the cycle.
NOTES:
1. In the model with integrated printer, the serial interface connector is not installed.
2. If one or more parameters exceed the set-point values, the sterilizer enters automatically in alarm status and the relative indication will be reported on the printout.
3.
If the cycle appears as correctly completed, there is no need to
check on the report that any parameter value is within the limits.
_____CLASS B____
DOMINA PLUS
Serial No 001091
Date 14.02.11
Prog. 1
134 C 5‟
Cycle 001343
START
Time 13:45:34
ABORTED
Time 13:45:35
ALARM N. 7
26
DOMINA PLUS B
DentalX
9.3 CONNECTION TO A COMPUTER (ONLY FOR SERVICE)
The serial port can be also used for the connection to a PC. This function is available for Service technician only, and allows to perform more accurate tests as well as a new approach to the service and certification aspect.
By a special SW tool, the Service technician is able to detect any significant data of the machine, and to perform more quickly and cheaply the troubleshooting and the repairing.
Moreover, the system allows the connection (via external modem) to an authorized Service Center in order to perform a remote check-up and a periodic certification of the effective operation of the sterilizer.
Do not connect devices not supported by the manufacturer.
THE RJ45 CONNECTOR IS NOT DESIGNED FOR A LAN CONNECTION, THIS OPERATION
WILL DESTROY THE MOST OF THE ELECTRONIC PART OF THE AUTOCLAVE.
DOMINA PLUS B 27
DentalX
10. OPERATOR’S NOTES
28
DOMINA PLUS B
APPENDIX
SERVICE BOOK
Although the sterilizer‟s average life is about 8-12 years, it is compulsory to carry out periodic controls in order to check the correct calibration and for possible wear of components. The control interval, as reported on the
Service Book, is 1 year or 1500 cycles for the ordinary maintenance, and 4 years or 10.000 cycles for the special maintenance. The sterilizer is configured to signal on the display, as the first term comes to expire, the message “ NEED SERVICE”. This feature is in compliance with the norm EN13060 and answers to specific safety requirements. The maintenance activities must be performed by authorized technicians (provided with card or certificate released by Dental X) and reported on this Service Book.
Model ………………………………………………………..
Installation Date ……………………………………………
Serial Number ………………………..
Altitude …………………………………
Installer ……………………………………………………………………………………………………………
Reseller …………………………………………………………………………………………………………….
Sterilization Manager: ………………………………………………………………………………………… .
Standard periodic checks
Pos. Check/Activity
1 Door adjustment
— Gasket replacement
— Disk-door clearance check
— Closing force check
— Lubrication
— Component wear check
— Closing pin replacement
— Screw hinge replacement
— Screw tightening
2 Calibration / Validation
— Altitude setting check
3 Filter cleaning / replacement
— Bacterial filter replacement
4 Pump feature check
— Pump replacement
5 Tank cleaning
6 Condenser cleaning
7 Safety valve replacement
Ordinary maintenance
1 Year / 1.500 cycles
1 Year / 1.500 cycles
1 Year / 1.500 cycles
1 Year / 1.500 cycles
1 Year / 1.500 cycles
1 Year / 1.500 cycles
1 Year
1 Year / 1.500 cycles
1 Year / 1.500 cycles
6 Months / 500 cycles
1 Year / 1.500 cycles
1 Year / 1.500 cycles
1 Year / 1.500 cycles
Special maintenance *
4 Years / 10.000 cycles
4 Years / 10.000 cycles
4 Years / 10.000 cycles
10.000 cycles
4 Years / 10.000 cycles
*) To be carried out in DentalX service department
In case of direct ship or consign for shipping the sterilizer for service center or factory repairing, include a copy of the filled Service Book pages .
DOMINA PLUS B
APPENDIX
SERVICE BOOK
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles …………………………
Notes
Aborted cycles ………. Alarm code ………………………………
Technician name: …………………………………………………….
Door adjustment
Calibration
Pump
□
□
□
1
Filter
Cond. cleaning
Tank cleaning
□
□
□
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
Sterilizer OK
6
YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles …………………………
Aborted cycles ………. Alarm code ………………………………
Technician name: …………………………………………………….
Door adjustment
Calibration
Pump
□
□
□
Filter
Cond. cleaning
Tank cleaning
□
□
□
Notes
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
6
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles ………………………..
Notes
Aborted cycles ………. Alarm code ……………………………..
Technician name: ……………………………………………………
Door adjustment
Calibration
Pump
□
□
□
2
Filter
Cond. cleaning
Tank cleaning
□
□
□
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
Sterilizer OK
6
YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles ………………………..
Aborted cycles ………. Alarm code ……………………………..
Technician name: ……………………………………………………
Door adjustment
Calibration
Pump
□
□
□
Filter
Cond. cleaning
Tank cleaning
□
□
□
Notes
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
6
DOMINA PLUS B
APPENDIX
SERVICE BOOK
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
Notes
No. of cycle …………… Cleaning cycles ………………………..
Aborted cycles ………. Alarm code ……………………………..
Technician name: ……………………………………………………
Door adjustment
Calibration
Pump
□
□
□
5
Filter
Cond. cleaning
Tank cleaning
□
□
□
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
Sterilizer OK
……………………………………………………………………………..
6
YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles ………………………..
Aborted cycles ………. Alarm code ……………………………..
Technician name: ……………………………………………………
Door adjustment
Calibration
Pump
□
□
□
Filter
Cond. cleaning
Tank cleaning
□
□
□
Notes
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
DOMINA PLUS B
6
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
Notes
No. of cycle …………… Cleaning cycles …………………………
Aborted cycles……….. Alarm code ………………………………
Technician name: …………………………………………………….
Door adjustment
Calibration
Pump
□
□
□
6
Filter
Cond. cleaning
Tank cleaning
□
□
□
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
Sterilizer OK
………………………………………………………………………………
6
YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles …………………………
Aborted cycles……….. Alarm code ………………………………
Technician name: …………………………………………………….
Door adjustment
Calibration
Pump
□
□
□
Filter
Cond. cleaning
Tank cleaning
□
□
□
Notes
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
6
APPENDIX
SERVICE BOOK
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles …………………………
Notes
Aborted cycles ………. Alarm code ………………………………
Technician name: …………………………………………………….
Door adjustment
Calibration
Pump
□
□
□
9
Filter
Cond. cleaning
Tank cleaning
□
□
□
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles …………………………
Aborted cycles ………. Alarm code ………………………………
Technician name: …………………………………………………….
Door adjustment
Calibration
Pump
□
□
□
Filter
Cond. cleaning
Tank cleaning
□
□
□
Notes
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
………………………………………………………………………………
60
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles ………………………..
Notes
Aborted cycles ………. Alarm code ……………………………..
Technician name: ……………………………………………………
Door adjustment
Calibration
Pump
10
□
□
□
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
Sterilizer OK YES
□
NO
□
Date …………………………………………………………………..
No. of cycle …………… Cleaning cycles ………………………..
Aborted cycles ………. Alarm code ……………………………..
Technician name: ……………………………………………………
Door adjustment
Calibration
Pump
□
□
□
Filter
Cond. cleaning
Tank cleaning
Notes
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
……………………………………………………………………………..
60
DOMINA PLUS B
□
□
□
DENTAL X S.p.A.
Via Marzotto 11
36031 Dueville (VI) Italy
Tel. +39 0444 367400
Fax +39 0444 367436 e-mail: [email protected]
http://www.dentalx.it
SERVICE
Italy — Tel. +39 0444 367415
Foreign — Tel. +39 02 76275234
Copyright Dental X 2011
Änderungen im Zuge technischer Weiterentwicklung vorbehalten.
We reserve the right to make any alterations which may be due to technical improvements.
Ci riserviamo il diritto di apportare modifiche a seguito di migliorie tecniche.
Printed in Italy
Sous réserve de modifications dues au progrès technique.
Reservados los derechos de modificatiòn en virtud del progreso tècnico.
-
- Производитель: Dental X
- Состояние на складе: Доступен для заказа
- Код товара: 2255-AMF
- Артикул 401004254010
Доставка Отправляем заказы от 1 дня при наличии товара на складе. |
Гарантия качества Все продаваемые товары имеют соответствующие документы. |
Онлайн 24/7 Заказы на сайте принимаются 24/7. |
В связи с изменением курса ЕВРО и ДОЛЛАРА |
Характеристики
Технические характеристики
- Быстрый цикл20 минут
- Вид сушкиВакуумная
- Класс автоклаваB
- Наличие принтераОпционально
- Объем камеры (л)18
- Страна производителяИталия
Описание
Модель соединила в себе передовые комплектующие, компактные формы, эргономичный дизайн и замечательные показатели работы. Автоклав укомплектован мощной вакуумной помпой, благодаря которой в нем можно обрабатывать материалы и инструменты, в частности наконечники, любого вида. К тому же помпа создает вентиляцию и необходимую сушку загруженных в камеру аппарата инструментов. Процесс стерильной обработки инструментов полностью автоматизирован, благодаря современной компьютерной техники.
Все этапы цикла стерилизации контролируются мощным микропроцессором, по окончании работы система самостоятельно отключается. Увеличить время службы аппарата и, как следствие, уменьшить расходы на техническое обслуживание позволяет система самотестирования, выявления и исправления неполадок. Удобство работы с автоклавом обеспечивается и возможностью индивидуального программирования параметров процесса стерилизации. Мощная вентиляция, система нагрева обеспечат не только качественную сушку, но и предупредят повреждения инструментов.
Ускоренный цикл Flash позволит простерилизовать инструменты и материалы под большим давлением меньше, чем за 20 минут. Надежная работа автоклава обеспечивается и возможностью подсоединения его к ПК. Автоклав можно доукомплектовать принтером, который ставиться спереди, в панель управления. С помощью принтера можно автоматически в печатном виде получать информацию о проводимом стерилизационном цикле. Однако к автоклаву можно подключить и стационарный принтер. Для архивирования протоколов процесса стерилизации в автоклав встроена опция USB-комплект, позволяющая сохранять информацию на любом внешнем носителе до 2 Гб.
Технические характеристики:
- Объем камеры 18 л
- 4 лотка
- Фракционный вакуум
- 4 цикла предвакуума
- Вакуумная сушка
- 7 стандартных программ + программа, задаваемая пользователем
- Тестовые программы: Bowie Dick-тест, Vacuum-тест
- Температура стерилизации 105-135оС
- Подключение к внешнему принтеру
- Встроенный принтер (опция)
- USB-комплект (опция)
- Габариты 443x590x428 мм
- Размеры камеры oslash,240х384 мм
- Максимальная потребляемая мощность 1900 Вт
- Вес нетто 55 кг
10 причин купить автоклав Domina Plus B:
- Ускоренный рабочий цикл — 18 мин (цикл Flash только для неупакованных инструментов и предметов), 35 мин для цикла B экономит Ваше время и позволяет принять больше пациентов.
- Повышенная вместимость! Более чем на 50% больше чем у аналогов позволяет Вам стерилизовать больше инструментов за один цикл.
- Энергосбережение (800Вт) позволяет Вам сократить расходы на электроэнергию.
- Стерилизация с бережным режимом предотвращает термический удар и не допускает повреждения инструментов, что позволяет продлить их срок службы.
- Своевременное и быстрое сервисное обслуживание по всей России
- Высокая функциональность обеспечивается благодаря интерактивной системе- инструменты в камере автоклава размещаются, эффективно используя весь полезный объем.
- Прямое подключение к водопроводной сети (опция)
- Возможность контроля без лишних хлопот. Участие человека сведено к минимуму. Система автоматически отключает автоклав по окончании цикла, освобождая Вас от необходимости контролировать ее работу.
- Простота в использовании- выберите программу и нажмите старт
- Вы платите один раз, а остаетесь довольны на много лет вперед
Быстрота работы:
Высокий темп работы зачастую требует выполнения быстрых циклов стерилизации. Эта задача должна быть выполнена быстро и безопасно. Однако такая практика приводит к быстрому износу инструмента.
В автоклаве Domina Plus B предусмотрено два быстрых режима:
- 18 минутный Flash (для неупакованных инструментов и материалов)
- 35 минутный Универсальный B
Системы Warm Up и Adaptive heat а так же датчики, контролирующие температуру по всей рабочей камере автоклава, сделали возможным бережную обработку без теплового удара инструментов. Domina Plus B экономит ваше время и деньги
Для выполнения быстрых циклов стерилизации рекомендуется:
- Предварительно прогреть камеру.
- Провести надлежащее техническое обслуживание.
- Убедиться, что автоклав имеет достаточную вентиляцию.
- Не перегружайте камеру: расположите инструменты так, чтобы пар мог свободно циркулировать.
- Проверьте напряжение питания.
- Domina Plus B экономит Ваше время и сохраняет инструменты в целости, позволяя избежать лишних расходов.
Надежность:
Основными компонентами, обеспечивающими надежность автоклава являются:
- Система нагрева Adaptive heat(не нуждающаяся в активации парогенератора)
- Система защиты для электромагнитных клапанов и вакуумного насоса
- Камера из меди.
Все эти факторы увеличивают долговечность аппарата.
С покупкой автоклава Domina Plus B Вы забываете о таком явлении как внеплановый ремонт и расходы на постгарантийное техобслуживание. Ведь ресурс работы Domina Plus B подразумевает многолетнюю бесперебойную эксплуатацию, по сравнению с бюджетными автоклавами азиатских производителей. Вы платите один раз и получаете результат надолго.
Вместимость:
Полезное рабочее пространство автоклава Domina Plus B увеличено на 50% за счет того, что применен новый подход интерактивная система загрузки инструментов. Грамотное расположение лотков позволяет использовать полезный объем камеры 18 литров на 50 % эффективнее, чем в автоклавах с объемом камеры 24 л.
Комфорт работы синоним Domina Plus B
Презентация DOMINA PLUS B (скачать)
Инструкция DOMINA PLUS B (скачать)
Рекомендуемые
Быстрый поиск: Автоклавы Италии
Технические характеристики | |
Быстрый цикл | 20 минут |
Вид сушки | Вакуумная |
Класс автоклава | B |
Наличие принтера | Опционально |
Объем камеры (л) | 18 |
Страна производителя | Италия |
- Самовывоз в Санкт-Петербурге
- Курьерская доставка
- Транспортными компаниями по Москве и России
- Самовывоз или доставка заказов СДЭК
Подробная информация о доставке в Информация о доставке
- Оплата наличными или картой при самовывозе в Санкт-Петербурге
- Оплата наличными при курьерской доставке в Санкт-Петербурге
- Безналичным переводом для юр. лиц
- Оплата в отделении банка по счету
Подробная информация в Способы оплаты