iMark
™
Microplate Absorbance Reader
Instruction Manual
iMark™
Microplate Absorbance Reader
Instruction Manual
Warranty Statement
This warranty may vary outside of the continental United States. Contact your local Bio-Rad office for the exact terms of your warranty.
Bio-Rad Laboratories warrants to the customer that the iMark Microplate Absorbance Reader will be free from defects in material and workmanship, and will meet all performance specifications for a period of one year from the date of shipment. This warranty covers all parts and labor.
In the event that the instrument must be returned to the factory for repair under warranty, the instrument must be packed for return in original packaging. Please contact your local Bio-Rad representative for the necessary return documents. The required documentation includes a decontamination certificate.
Bio-Rad shall not be liable for any incidental, special, or consequential loss, damage, or expense directly or indirectly arising from the use of the iMark Microplate Absorbance Reader. Bio-Rad makes no warranty whatsoever in regard to products or parts furnished by third parties, such being subject to the warranty of their respective manufacturers. Service under this warranty shall be requested by contacting your nearest Bio-Rad office.
The following items are considered customer-installable consumables: thermal printer paper and light bulbs. These parts are not covered by this warranty. All customer-installed parts are warranted only to be free from defects in workmanship. This warranty does not extend to any instrument or parts thereof that have been subject to misuse, neglect, or accident, or that have been modified by anyone other than Bio-Rad or that have been used in violation of Bio-Rad instructions.
The foregoing obligations are in lieu of all other obligations and liabilities including negligence and all warranties of merchantability, fitness for a particular purpose, or otherwise expressed or implied in fact or by law, and state Bio-Rad’s entire and exclusive liability and buyer’s exclusive remedy for any claims or damages in connection with the furnishing of goods or parts, their design, suitability for use, installation or operation. Bio-Rad will in no event be liable for any special, incidental or consequential damages whatsoever, and Bio-Rad’s liability under no circumstances will exceed the contract price for the goods for which liability is claimed.
i
Regulatory Notices
Electro Magnetic Compatibility:
This instrument is designed to meet the Class-A emissions and immunity levels of the EN 61326 product family standard for the CE mark. This rating requires that it only be used in a laboratory environment.
NOTE: These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his/her own expense.
Safety
This instrument is designed to meet EN 61010-1 requirements for the CE mark, which are internationally accepted safety standards for electrical equipment for measurement, control and laboratory use. EN 61010-1 applies to equipment which is used under the following conditions:
•Indoor use
•Altitude up to 2,000 m
•MAINS supply voltage fluctuations up to ±10%
This instrument has been tested, and its compliance has been confirmed under the following environmental conditions:
•Overvoltage category II
•Pollution degree 2
If the instrument is used in a manner not specified in this manual, the protection provided by the instrument may be impaired. This instrument should not be modified or altered in any way. Alteration of this instrument will result in the following:
•Void the manufacture’s warranty.
•Void regulatory compliance.
•Create a potential safety hazard.
Protective Ground
This instrument has a detachable, three-wire power cord for connection to both the power source and protective ground. The protective ground contact is connected to the accessible metal parts of the instrument. To prevent electrical shock, always use a power source outlet that has a properly grounded protective-ground contact.
ii
Table of Contents |
||
Introduction …………………………………………………………………………………….. |
1 |
|
Description …………………………………………………………………………………………… |
1 |
|
USB Interface ……………………………………………………………………………………….. |
1 |
|
Built-in Printer ………………………………………………………………………………………. |
1 |
|
External Features……………………………………………………………………………………. |
3 |
|
Keypad Guide ……………………………………………………………………………………….. |
4 |
|
Instrument Setup ……………………………………………………………………………….. |
5 |
|
Unpacking the Instrument…………………………………………………………………………. |
5 |
|
Initial Instrument Startup………………………………………………………………………….. |
5 |
|
Instrument Shutdown ………………………………………………………………………………. |
6 |
|
Software Overview…………………………………………………………………………….. |
7 |
|
Capabilities…………………………………………………………………………………………… |
7 |
|
Quick Guide to Reading a Plate …………………………………………………………………. |
7 |
|
Instrument Functions ………………………………………………………………………….. |
8 |
|
Calculations ………………………………………………………………………………………….. |
8 |
|
Security password…………………………………………………………………………………… |
8 |
|
Memory Back-Up…………………………………………………………………………………… |
8 |
|
Limits ………………………………………………………………………………………………… |
9 |
|
Report Types ………………………………………………………………………………….. |
10 |
|
End-point Protocol Reports……………………………………………………………………… |
10 |
|
1. |
Raw data report……………………………………………………………….. |
10 |
2. |
Absorbance report ……………………………………………………………. |
10 |
3. |
Limit report ……………………………………………………………………. |
10 |
4. |
Matrix report ………………………………………………………………….. |
10 |
5. |
Cutoff report…………………………………………………………………… |
11 |
6. |
Curve Fit Report ……………………………………………………………… |
12 |
7. |
Concentration Report………………………………………………………… |
12 |
8. |
Difference Report…………………………………………………………….. |
12 |
9. |
TeSeE Screening Report ……………………………………………………. |
13 |
Kinetic Protocol Reports ………………………………………………………………………… |
15 |
|
1. |
Absorbance Report …………………………………………………………… |
15 |
2. |
Kinetic Plots Report …………………………………………………………. |
15 |
3. |
Linear Regression Report…………………………………………………… |
15 |
4. |
GALT Report …………………………………………………………………. |
15 |
iii
Screen by Screen Procedures………………………………………………………………. |
17 |
Initial Startup ………………………………………………………………………………………. |
17 |
Configuring the System………………………………………………………………………….. |
18 |
Printing Reports and Protocol information…………………………………………………… |
20 |
Memory Recall Operation ………………………………………………………………………. |
21 |
Editing the Kinetic Protocol…………………………………………………………………….. |
22 |
Checkmark Value Setting (Administrator only)…………………………………………….. |
24 |
Editing End Point Protocol ……………………………………………………………………… |
25 |
Cutoff Setting Procedure ………………………………………………………………………… |
26 |
Standard Setting Procedure……………………………………………………………………… |
29 |
Mode Setting Procedure …………………………………………………………………………. |
32 |
Plate Map Setting Procedure……………………………………………………………………. |
33 |
Instrument Maintenance…………………………………………………………………….. |
36 |
Installing Interference Filters …………………………………………………………………… |
36 |
Changing the Lamp ………………………………………………………………………………. |
36 |
Troubleshooting and Error Messages…………………………………………………….. |
37 |
Instrument Specifications…………………………………………………………………… |
39 |
iv
iMark Microplate Absorbance Reader Instruction Manual
Introduction
Description
The iMark™ Microplate Absorbance Reader (catalog #168-1130) is an eight-channel, vertical pathlength photometer that measures the absorbance of the contents in the wells of a 96-well microtitration plates. It can perform single or dual wavelength measurements and can report absorbance values to three decimal places.
The iMark reader can be programmed by entering commands through the membrane keypad to define plate reading and data analysis parameters as testing protocols and select report types. Hard copy reports can be produced by an optional on-board thermal printer or external printer.
The iMark reader can also be controlled by the Windows-based Microplate Manager® software (catalog # 168-9520) program, through its built-in USB 2.0 interface device. Microplate Manager software provides a complete, user-friendly solution for protocol programming as well as data analysis, for all microplate absorbance readers from Bio-Rad Laboratories.
USB Interface
The iMark Microplate Reader has a built-in USB interface. This allows external computers to control the instrument. When an external computer is in control of the microplate reader, the printer and the membrane keys, except for the Start/Stop and Open/Close keys, are automatically deactivated, and the LCD display appears as below.
Remote mode
The reader will remain in remote control mode until it is released or until the Start/Stop key is pressed on the reader.
Built-in Printer
The iMark Microplate Reader has a built-in printer. It can generate all types of reports, including graphs and kinetic plots.
1
iMark Microplate Absorbance Reader Instruction Manual
Catalog # |
Description |
168-1130 |
iMark Microplate Absorbance Reader with built-in printer |
168-1135 |
iMark Microplate Absorbance Reader with built-in printer, |
including Microplate Manager 6 software |
|
168-6940 |
Checkmark Absorbance Reader Performance Validation Kit |
Accessories |
|
Catalog # |
Description |
168-1011 |
405 nm filter |
168-1013 |
415 nm filter |
168-1020 |
450 nm filter |
168-1028 |
490 nm filter |
168-1038 |
540 nm filter |
168-1040 |
550 nm filter |
168-1044 |
570 nm filter |
168-1049 |
595 nm filter |
168-1054 |
620 nm filter |
168-1056 |
630 nm filter |
168-1061 |
655 nm filter |
168-1080 |
750 nm filter |
168-1006 |
Replacement lamp for iMark Microplate Absorbance Reader |
168-2007 |
Replacement printer paper, package contains 3 rolls |
Custom filters between 400 and 750 nm may be ordered. Corresponding catalog numbers can be found on www.bio-rad.com, or you can order by specifying the wavelength and the model number of the reader.
Checkmark Absorbance Reader Performance Validation Kit (catalog item 168-6940) consists of a plate with absorbance standards and software to calculate the accuracy and precision of the iMark Microplate Absorbance Reader.
2
iMark Microplate Absorbance Reader Instruction Manual |
|||
External Features |
Release latch for rear cover |
Rear cover |
|
Power button |
Printer slot |
Liquid crystal display (LCD)
Keypad
Reading chamber door
Cooling vents and fan
USB 2.0 interface |
Serial number label |
Power cord receptacle |
3
iMark Microplate Absorbance Reader Instruction Manual
Keypad Guide
Main |
Returns to MAIN SCREEN. |
Start/Stop |
Initiates plate reading using current active protocol. |
Stops plate reading and printing. |
|
Paper Feed |
Advances paper strip in internal printer. |
|
Prints out the plate data and protocol information. |
Edit |
Enters the Edit Menu, and sets up the instrument. |
Memory Recall |
Reads out protocol and plate data. |
Open/Close |
Opens or closes the reading chamber door. |
Open/Close |
Memory |
Edit |
|
Recall |
|||
START |
|||
/ |
|||
STOP |
|||
Main |
Paper Feed |
|
Up Arrow |
Moves the cursor upward. |
Selects the alphabet or the symbol. |
|
Left Arrow |
Returns to the previous screen. |
Moves the cursor to the left. |
|
Down Arrow |
Moves the cursor downward. |
Selects the alphabet or the symbol. |
|
Right Arrow |
Moves the cursor to the right. |
Changes or selects the value and type. |
|
Enter |
Completes or seals a field entry. |
A => Z
+−••
ENTER
Back Change
Z => A
Dot/Function |
Inputs the dot. |
||||||||||||||||||
Changes the input mode. |
|||||||||||||||||||
Ten Keys |
Inputs numbers or well type in plate mapping. |
||||||||||||||||||
CP |
CN |
CW |
|||||||||||||||||
0 |
/ EMP |
: Empty |
|||||||||||||||||
7 |
8 |
9 |
|||||||||||||||||
5 |
/ QC |
: QC control |
|||||||||||||||||
1 |
/ SMP |
: Sample |
CO |
QC |
CAL |
||||||||||||||
6 |
/ CAL |
: Calibrator |
4 |
5 |
6 |
||||||||||||||
2 |
/ BLK |
: Blank |
|||||||||||||||||
SMP |
BLK |
STD |
|||||||||||||||||
7 |
/ CP |
: Positive Control |
|||||||||||||||||
1 |
2 |
3 |
|||||||||||||||||
3 |
/ STD |
: Standard |
|||||||||||||||||
8 |
/ CN |
: Negative Control |
|||||||||||||||||
EMP |
FUNC |
||||||||||||||||||
4 |
/ CO |
: Cutoff control |
0 |
0 |
|||||||||||||||
9 |
/ CW |
: Weak Positive Control |
4
iMark Microplate Absorbance Reader Instruction Manual
Instrument Setup
Unpacking the Instrument
The shipping carton contains the following items:
•iMark™ Microplate Reader, with 415 nm, 450 nm, 490 nm, 595 nm, 655 nm, and 750 nm interference filters installed on the filter wheel
•Power cord
•USB cable
•Dust cover
•Instruction manual
•Warranty card
•One roll of thermal printer paper
Inspect the exterior of the instrument for any signs of shipping damage. Contact your local Bio-Rad representative if any of these items are damaged or missing.
Initial Instrument Startup
1.Place the instrument on a clean, sturdy table or bench. It is important to keep the instrument in a clean, relatively dust free environment to ensure optimal performance.
2.Connect the power cord to the back of the instrument. Before connecting the instrument to the main electrical supply, check that the AC voltage is appropriate for the instrument.
3.To turn the power on press the green power button on the top of the instrument. The LCD will display the version number of the on-board firmware. After about three seconds, the instrument will perform an initial self-diagnosis that requires about fifteen seconds. The login screen displays for you to choose your user level (administrator or common user), and enter the password (initial password: 00000). Please allow three minutes for the instrument to warm up (reach thermal equilibrium) before reading plates.
System Login
User: Administrator
Password:*****
Press ENTER
Change password
Select “Change Password” in the Security Menu.
Enter the current Password then move the cursor to the 3rd line and enter a new password. Press Enter to complete the field entry. The system will return to the previous Security Menu.
Change user mode
Select “Change user” in the Security Menu.
Select a desired user mode by pressing the Right Arrow key. Enter the appropriate password for the selected user mode. Press Enter to complete the field entry. The system will return to the previous Security Menu.
5
iMark Microplate Absorbance Reader Instruction Manual
4.Install the printer paper. The thermal printer paper is printable on one side only and must be properly installed for the printer to function. The outside of the roll is the printable side.
a.Open the rear cover of the instrument.
b.Tear off a small piece at the beginning of a new roll at an angle to form a point.
c.Place the roll of paper in the small pan-like holder positioned below the printer. The roll should be positioned so that it will feed from the bottom.
d.While pressing the PAPER FEED key, feed the pointed end of the paper into the paper feed slot in the underside of the printer until the printer grabs the paper and feeds it through the slot in the top of the instrument.
e.Securely close the rear compartment.
Multi-language support
The iMark LCD can produce text in three languages; English, Russian and Chinese. To switch to English from Russian or Chinese, press the Main button, then the Edit button to get to the configuration screen. Press the downward-pointing arrow on the key pad six times to navigate to the language options line. Press Enter twice to select English, and then press Enter a third time to shut down the reader. When restarted, the iMark will display in English.
Instrument Shutdown
To shut down the instrument, press the green power button on the top of the instrument for a few seconds. Then the LCD will display a Power off, Yes or No screen. Select “Yes” to shut down the instrument.
Power off
>>Yes No
6
iMark Microplate Absorbance Reader Instruction Manual
Software Overview
Capabilities
The iMark Microplate Reader has on-board software that allows the user to set the plate reading and data analysis conditions, to save these settings as a test protocol, to read a microplate under this protocol, and to print reports from the built-in printer. The software communicates through the 4-line, 20-character LCD and is controlled through the instrument’s membrane keypad.
The software has three different protocol types: End-point analysis, Kinetic analysis, and Checkmark validation. The Checkmark validation function is used only by the administrator to verify quality control.
The display shows the current mode. The user can choose between these protocol types, or can press the Memory Recall key to use the current protocol type.
Quick Guide to Reading a Plate
1.Turn on the instrument. Allow about 30 seconds for self-diagnosis. The instrument needs to warm up for about three minutes prior to reading a plate.
2.Upon power up, the login screen will appear. Enter the login password (initial password: 00000), press Enter.
3.Program a protocol for your assay. See page 22 for information on Editing a Kinetic protocol, or page 25 for Editing an End Point Protocol. Make sure filters installed correlate with information entered in reader firmware.
4.Select the protocol of choice, if programmed. See the Memory Recall section on page 21 for detailed operation.
5.Review or print parameters set in the protocol. See the Editing End Point Protocol section on page 25 for details.
6.Verify settings on Cutoff, Report, Limit, STDs, Mode, Mapping, and kit name parameters are as desired.
7.Protocol setting can be printed; see the Printing Reports and Protocols section on page 20 for information.
8.All reports other than the raw data report require plate mapping. Set the plate map with instructions from the Plate Map Setting Procedure section on page 33.
9.The Matrix and Limit reports require assignment of upper and lower limits. Set the upper and lower limits in the Editing End Point Protocol Limit setting window.
10.The Cutoff report requires definition of the cutoff setting. Refer to the Cutoff Setting Procedure section on page 26.
11.The Curve fit and Concentration reports require that standard concentration and locations must be defined. Refer to page 29 for details on how to define the standards and the curve fit parameters.
12.Set the general reading mode (single or dual wavelength reading, plate shaking, read mode) following the instructions in the Mode Setting Procedure section on page 32.
After warm-up is complete:
13.Press the “Main” key. The main screen appears on the LCD.
14.Carefully place the microplate in the reading chamber. Press the “Start/Stop” key to read.
7
iMark Microplate Absorbance Reader Instruction Manual
Instrument Functions
Calculations
The Microplate Reader uses Beer’s Law to calculate the absorbance value of each well. Beer’s Law states that absorbance is equal to the log10 of the ratio of the baseline measurement intensity (Io) to the sample measurement intensity (I).
Beer’s Law: Absorbance = Log10 (Io/I)
Before measuring the plate, the reader takes a reading for all eight photodiode channels. These values are recorded as the baseline measurement (Io) values for each channel, respectively. The reader then records the sample measurement (I) value for each well, and calculates the absorbance using these values. Channel-to-channel error is significantly reduced because the Io value for a given channel is used only in determining the absorbance of the wells of that channel.
Security password
The iMark Microplate Reader requires a user login with password. This security feature helps to prevent any unwarranted modifications to the analysis conditions/protocols and data stored in the instrument, and identifies the operator on reports. The reader has two user settings, the Administrator and the Common user. Each user level has its own security password.
Memory Back-Up
The battery back-up provides memory even after the reader is turned off.
The following information will be saved in the memory until new reading parameters are set by the user or a new plate is read by the instrument. Note that if a run is aborted before it is finished, all the previous plate reading data remain in memory and none of the data from the aborted reading is stored.
•The ten latest sets of End-point protocol plate reading results
•The two latest sets of Kinetic plate reading results (max. 30 series of readings for each Kinetic Protocol)
•64 End-point Protocol settings and 2 Kinetic Protocol settings
•A set of Checkmark reference data and the one latest set of Checkmark reading result
•A set of standard curve data for the storage of standard curves and graphs
•Eight wavelength values for filters
•Laboratory name
•Two security passwords (for the Administrator and the Common user)
•Instrument serial number
In addition to the battery back-up memory, the instrument has a small amount of non-volatile memory to keep the system information after the battery is dead. The serial number is stored in this memory. The average battery life is 5 years.
8
iMark Microplate Absorbance Reader Instruction Manual
The first time the instrument is turned on, or after a battery failure, the following default information is held in the memory.
1.Both Administrator and Common user security passwords are set to «00000».
2.Laboratory name is «Bio-Rad Laboratories».
3.All wavelength values of the filters are set to «—» nm.
4.The date and time of the calendar chip are set to 1 day, 1 month, 00 year, 0 hour and 0 minute.
5.No plate reading data is available.
6.End-point protocol is chosen and the protocol number is set to #1.
7.No standard curve data for the storage of standard curves and graphs is available.
Limits
The Microplate Reader displays absorbance readings with absolute values as 3.500. Out-of-range absorbance values, (those with absolute values greater than 3.500) are displayed as either «*.***» or «-*.***». For example, if the absorbance is 4.500, then the display will read «*.***», and if the absorbance is -4.500, the display will read «-*.***».
9
Loading…
Bio-Rad Laboratories (США) – глобальная компания, производитель оборудования, питательных сред, антител, расходных материалов, а также программного обеспечения для исследований в области геномики и протеомики и безопасности пищевых продуктов. За более полувековую историю компании удалось стать одним из лидеров рынка в производстве современного оборудования для молекулярно-биологических исследований, с каталогом, насчитывающим более 8 000 наименований, штатом сотрудников в 6 тыс. человек, работающих в том числе в России, Казахстане и Украине. Большое внимание компания уделяет поддержке пользователей, организовывая регулярные практические семинары по протеомному анализу и применению ПЦР «в реальном времени», проводя сервисное обслуживание на территории России и стран СНГ и многое другое.
-
Life Science Group
Model 680 and 680 XR Microplate Reader
Service Manual
Part Number 100-2414 Revision E
-
100-2414E 02/06 i
PROPRIETARY NOTICE
Information contained in this manual is copyrighted by Bio-Rad
Laboratories, Inc. None of the contents may bereproduced in full or
in part by any person without the prior written approval of Bio-Rad
Laboratories, Inc. Thismanual is intended to provide information to
assist the qualified field or instrument service technician to
repair,maintain, and order replace parts for the instrument
supplied. Every reasonable effort has been made to keepthe
information in this manual current and accurate as of the date of
publication or revision. However, no guar-antee is given or implied
that the manual is error-free or accurate about any
specification.Copyright 2002-2006 by Bio-Rad Laboratories, Incorporated. All
rights reserved.Bio-Rad Laboratories Inc, Life Science Group
2000 Alfred Nobel Drive
Hercules, California 94547
Telephone: 1-800-424-6723
-
ii 100-2414E 02/06
About this manual This document is part of Life Science Groups
commitment to provideyou, the service professional, with adequate
information to effectivelysupport Bio-Rad instruments. Additional
copies of this service documentmay be ordered using part number
100-2414.We welcome and appreciate any suggestions you may have to
improveour service documentation. Please address your comments or
sugges-tions to:Technical Editor, Global Service Support Dept.
Bio-Rad Laboratories, Life Science Group
2000 Alfred Nobel Drive
Hercules, CA 94547
FAX Number: 510 741-5802
-
100-2414E 02/06 iii
Revision Control List
Document: Model 680 and 680 XR Microplate Reader Service
ManualPart Number: 100-2414Date of Latest Revision: February
2006Each page of this manual has the catalog number, revision level
and revision date for that page. Subsequentrevisions to a page will
be indicated by changes to the revision level and revision date for
that page. A sum-mary of changes for all revisions is listed in the
following table.RevisionPage
NumberDescription
ChangedBy
Date
A — Initial Release A. Gandy Feb. 2002
B 4-11 to 4-12 Updated Checkmark Plate procedure A. Gandy June
20025-1 to 5-4 Added optics cleaning procedure
6-1 to 6-25 Added part numbers to sections; Updated Figure
6-96-16 Added check of plate position
7-1 to 7-2 Added Checkmark Plate; Updated Figure 7-3
7-4 Added Section AdjustingPlate Carrier Belt Tension
7-6 Added Section Alignment of Plate Carrier
8-5 Added Table 8-4 Checkmark Plate Failures
C Sec 1 thru 10 Manual Updated to Add Model 680XR A. Gandy Mar
20045-1 Added Note
6-3 Added 6.3.3 Model 680 XR Lamp
6-12 Added Table 6-1
6-15 to 6-16 Added 6.3.11 Model 680 XR Lamp Control PCB
6-22 Added 6.3.17 Model 680 XR Light Balancing Filter
7-8 Added Table 2 Firmware History
8-1 Updated Table 8-1
9-3 to 9-11 Updated parts information
App D Added Appendix D
D 6-12 Added Backlit LCD to Table 6-1 A. Gandy Sept 2005
7-6 Updated to reference Section 7.7
7-9 Updated Firmware Table 7-2 and added 7.6.3
7-10 to 7-11 Added Section 7.7 QC Software
9-5 to 9-11 Updated parts
E 7-8 to 7-9 Updated Firmware version 1.71; Table 7-2 update A.
Gandy Feb 20067-11 to 7-15 Updated 7.6.4 Firmware Downloading
7-17 Added Section 7-8 External Printer Setup
8-1 Updated troubleshooting printer
9-11 Added Shield Cover
-
Table of Contents
Section 1 Introduction 1.1 General Inforation . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . 1-11.2 Service Philosophy . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-11.3 Safety and Precautions . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-11.4 Conventions . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 1-1Section 2 System Overview 2.1 Introduction . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . 2-12.2 Product Description . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-12.3 Product Configurations . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-22.4 Regulatory Notices . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-22.5 Product Overview . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-32.5.1 Calculations . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32.5.2 Limits . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-32.5.3 Memory Backup . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 2-32.5.4 Reports . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-42.5.5 RS-232 Interface . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-42.6 Instrument Specifications . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-52.6.1 Features . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-52.6.2 Optical . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-52.6.3 Performance . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-52.6.4 Thermal . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-62.6.5 Electrical . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-62.6.6 Physical . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-6Section 3 Installation 3.1 Introduction . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . 3-13.2 Visual check . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 3-13.3 Installation Procedure . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-23.4 Install Paper Roll (Optional Internal Printer) . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23.5 Computer Interface (Optional). . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3Section 4 System Checkout 4.1 Introduction . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . 4-14.2 Tools and Materials . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-14.3 Summary of Tests and Specifications . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24.4 System Checkout Procedure . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34.4.1 General. . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-34.4.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-34.4.3 Preparing Yellow Dye Test Plate . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . 4-3 -
Table of Contents
4.4.4 Startup . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-34.4.5 Touch Pad/Mixing Function . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 4-34.4.6 Verify Printer . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44.4.7 Start Microplate Manager. . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 4-44.4.8 Air Read . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-54.4.9 Accuracy . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-74.4.10 Channel-to-Channel Variation . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 4-84.4.11 Stability and Drift . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-94.4.12 Verify Incubator Temperature (Optional) . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 4-104.5 Checkmark Validation Program (Optional Method) . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 4-114.5.1 General . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-114.5.2 Procedure. . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-11Section 5 Preventative Maintenance5.1 Introduction . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . 5-15.2 Tools and materials
5.3 Preventative Maintenance. . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-15.3.1 Firmware . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15.3.2 Inspection of Instrument . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 5-15.3.3 Perform System Checkout . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 5-1Section 6 Removal and Replacement 6.1 Introduction . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . 6-16.2 Tools and Materials . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6-16.3 Removal and Replacement Procedures . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . 6-16.3.1 Interference Filters . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16.3.2 Model 680 Lamp . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16.3.3 Model 680 XR Lamp . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 6-36.3.4 Top Cover . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-46.3.5 Printer Assembly . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-56.3.6 Keypad Membrane. . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 6-76.3.7 Inlet Assembly . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-86.3.8 Plate Carrier Assembly and Analog Board (#282) . . . . . .
. . . . . . . . . . . . . . . . . . 6-96.3.9 CPU Board (#281) . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-126.3.10 Incubator Control Board . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-136.3.11 Model 680 XR Lamp Control PCB . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . 6-156.3.12 Power Supply, 12 Volt . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-176.3.13 Power Supply, 24 Volt . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-186.3.14 Photo Sensor (Analog Board) . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . 6-19 -
Table of Contents
6.3.15 Photo Sensor (Filter Wheel) . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 6-206.3.16 Model 680 Light Balancing Filter . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 6-216.3.17 Model 680 XR Light Balancing Filter . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . 6-226.3.18 Plate Carrier Motor . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-236.3.19 Optical Fiber Assembly . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-256.3.20 Filter Wheel/Filter Wheel Motor . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 6-276.3.21 Fan . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6-286.3.22 Lithium Battery . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6-29Section 7 Adjustment and Calibration 7.1 Introduction . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . 7-17.2 Tools and Materials . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7-17.3 Electronics Adjustment . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7-17.3.1 Setting of Jumper on the CPU Board . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . 7-17.3.2 Setting of Jumpers on the Printer Control Board . . . . .
. . . . . . . . . . . . . . . . . . . . 7-27.3.3 Incubator Electronics Adjustment . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . 7-37.4 Adjusting Plate Carrier Belt Tension . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-47.5 Plate Carrier Alignment . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7-67.5.1 Checking of Plate Alignment . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 7-67.5.2 Adjusting Plate Alignment — Y Direction . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . 7-77.5.3 Adjusting Plate Alignment — X Direction . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . 7-77.6 Firmware . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 7-87.6.1 General . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7-87.6.2 Firmware History . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-97.6.3 Integration of English and Chinese Firmware Version . . .
. . . . . . . . . . . . . . . . . . 7-107.6.4 Downloading Firmware. . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 7-117.7 QC Software . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 7-167.7.1 Entering QC Software . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 7-167.7.2 Monitor Mode . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-167.7.3 Plate Positioning. . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-177.8 External Printer Setup . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7-17Section 8 Troubleshooting 8.1 About this Section . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . 8-18.2 Troubleshooting . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
8-1Table 8-1 General Problems . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1Table 8-2 Error Codes. . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2Table 8-3 Firmware Downloading . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 8-4 -
Table of Contents
Section 9 Parts Information9.1 About this Section . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . 9-19.2 Parts Orders . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 9-19.2.1 US Domestic Operations . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 9-19.2.2 International Subsidiaries . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 9-19.3 Returning Defective Parts . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9-19.3.1 US Domestic Operations . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . 9-19.3.2 International Subsidiaries . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . 9-29.4 Service Parts . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 9-3Table 9-1 Replaceable Parts for the Model 680 Microplate Reader
. . . . . . . . . . . . . . . . 9-3Table 9-2 Consumables . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-119.5 Filters . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . 9-12Table 9-3 Filters . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9-129.6 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . 9-13Table 9-4 Special Service Tools . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . 9-139.7 Manuals . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . 9-14Table 9-5 Manuals . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9-14Section 10 Drawings 10.1 About this Section . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . 10-110.2 Drawings . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. 10-1Block Diagram Wiring Diagram
AppendixAppendix A Model 680 System Checkout Report . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . A-1Appendix B Model 680 XR System Checkout Report . . . . . . . . .
. . . . . . . . . . . . . . . . . . . B-1Appendix C Checkmark Check Plate . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . C-1Appendix D Error Flow Diagrams . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . D-1 -
Introduction
100-2414C 03/04 1-1
Section 1Introduction
1.1 General Information
This manual is the reference to be used for troubleshooting,
repair, and maintenance of the Model 680 andModel 680 XR Microplate
Readers. It is intended to be used by trained technicians and
engineers familiar withsafety precautions relative to working with
instrumentation. The procedures in this manual assume that
thereader is familiar with the operation of the Model 680 and 680
XR Microplate Readers. Complete operatinginstructions are contained
in the Model 680 and 680 XR Microplate Reader instruction manual.
Informationpresented in the instruction manual is not repeated in
this Service Manual unless there is a compelling reasonto do
so.1.2 Service Philosophy
This manual is designed to support mail in products. Regional
service centers are the first level of service,where faulty
assemblies can be most effectively repaired at a central repair
facility. The second level of serviceis performed at our factory
instrument service facility in Richmond, California, where
additional service proce-dures requiring resources not available at
regional service centers are performed.1.3 Safety and Precautions
The Model 680 and 680 XR Microplate Reader contains the
following Hazards:Hazardous voltages
Thermally hot components
The following should be observed when servicing this unit:
When power is required, the unit should be plugged into an
outlet with a properly grounded receptacle.The use of 2-prong plug
adapters is not recommended and will void the safety certification
of thisproduct.The use of extension cords may compromise the safety of the
operator and is not recommended.Ensure that fuses installed in the unit are of the correct
rating.Ensure that none of the ventilation openings in the unit are
blocked. Excessive heat buildup in the unitmay cause failures.Bio-Rad is not responsible for any injury or damage caused by
the use of this instrument for purposes otherthan for which it is
intended or by modifications of the instrument not performed by
Bio-Rad or an authorizedagent.1.4 Conventions
The following conventions are used throughout this manual.:
Left and right sides of the unit are as viewed from the front,
operators position, unless otherwise stated.Dimensions are in millimeters, unless otherwise stated.
-
System Overview
100-2414C 03/04 2-1
Section 2System Overview
2.1 Introduction
The Model 680 and Model 680 XR Microplate Readers are an
eight-channel, vertical pathlength photometerthat measures
absorbance of the contents of 96 well mictotitration plates. They
can perform single or dualwavelength measurements, and can report
absorbance values to three decimal places.The Model 680 and 680 XR Microplate Readers can be programed by
entering commands through the keypadto define plate reading
conditions and data analysis conditions as a protocol data, and
select report types. Aprinted report is produced by an optional
on-board thermal printer or external printer. An optional incubator
isalso available.The Model 680 and 680 XR Microplate Readers feature a built-in
RS-232C serial interface device for conve-nient computer
interfacing. The Microplate Manager software for personal computers
offers a complete analy-sis program. Microplate Manager supports
Windows 95, Windows NT and Windows 2000.Microplate Manager Version 5.2.1 added support for the Model
680XR reader.2.2 Product Description
Figure 1: Model 680/680 XR Microplate Reader
The following items are included with the Model 680/680 XR
Microplate Reader system:Microplate Reader with 415 nm, 450 nm, 490 nm and 655 nm
interference filters installedPower Cord
Spare Fuse
Dust Cover
Instruction Manual
Warranty Card
-
System Overview
2-2 100-2414C 03/04
2.3 Product Configurations
The Model 680 Microplate Reader is available in four
configurations.2.4 Regulatory Notices
This instrument is designed to meet Class A emissions and
immunity levels of EN61326 product family stan-dard for CE mark.
This rating requires that it only be used in a laboratory
environment.This instrument is designed and certified to meet EN61010
requirements which are internationally acceptedelectrical safety
standards. Certified products are safe to use when operated in
accordance with the InstructionManual. This instrument should not
be modified or altered in any way. Alteration of this instrument
will result invoiding of the manufacturers warranty, void
regulatory certification and create a potential safety hazard.This equipment has been tested and found to comply with the
limits for Class A digital device, pursuant to Part15 of FCC rules.
These limits are designed to provide a reasonable protection
against harmful interferencewhen the equipment is operated in a
commercial environment. This equipment generates, and can
radiateradio frequency energy and if not installed and used in
accordance with the Instruction Manual, may causeharmful
interference to radio communications. Operation of equipment in a
residential area is likely to causeharmful interference in which
case the user will be required to correct the interference at their
expense.Table 1: Product Configurations
Catalog Number Description
168-1000 Model 680 Microplate Reader No Temperature Control and
No Internal Printer168-1001 Model 680 Microplate Reader With Temperature Control
and No Internal Printer168-1002 Model 680 Microplate Reader No Temperature Control and
With Internal Printer168-1003 Model 680 Microplate Reader With Temperature Control
and With Internal Printer168-1100 Model 680 XR Microplate Reader No Temperature Control
and No Internal Printer168-1101 Model 680 XR Microplate Reader With Temperature Control
and With Internal Printer -
System Overview
100-2414C 03/04 2-3
2.5 Product Overview
2.5.1 Calculations
The Model 680 and 680 XR Microplate Readers uses Beers Law to
calculate the absorbance value of each well. Beers Law states that
absorbance is equal to the Log10 of the ratio on the baseline
measurement intensity (Io) to the Sample measurement intensity
(I).Beers Law: Absorbance = Log10 (Io/I)
Before measuring the plate, the reader takes a reading for all
eight photodiode channels. These values are recorded as the
baseline measurement (Io) values for each channel, respectively.
The reader then records the sample measurement (I) value for each
well, and calculates the absorbance using these values.
Chan-nel-to-channel error is significantly reduced because the Io
value for a given channel is used only in deter-mining the
absorbance of the wells of that channel.2.5.2 Limits
The Model 680 and 680 XR Microplate Readers display absorbance
readings with absolute values up to 3.500. Absolute values greater
than 3.500 are out of range and are displayed as either *.*** or
-*.***.2.5.3 Memory Backup
The battery backup provides memory even after the reader is
turned off. The following information will be saved in memory until
new reading parameters are set by the user or a new plate is read
by the instrument. Note: If reading a new plate is aborted before
it is finished, all the previous plate reading date will remain in
memory and none of the new data from the new plate is retained.1. 10 latest plate reading data for End point protocol
2. 2 latest sets of Kinetic plate reading data (maximum 30
series of readings for each Kinetic)3. End Point and kinetic protocol data
4. A set of Checkmark protocol data and the latest set of
Checkmark read data5. 8 wavelength values for filters
6. Laboratory name
7. 2 security passwords (Administrator and Common user)
The first time the instrument is turned on, or after a battery
replacement, the following default information will hold in
memory:1. Both Administrator and Common user security password are set
to 000002. Laboratory name is Bio-Rad Laboratories
3. All wavelength values of the filters are set to — nm
4. The date and time of the calendar chip are set to 1 day, 1
month, 00 year, 0 hour and 0 minutes5. None of the plate reading data is available
6. End Point protocol is chosen and the protocol number is set
to #1 -
System Overview
2-4 100-2414C 03/04
2.5.4 Reports
Eight types of reports for End Point protocol can be generated
by the Model 680 Microplate Reader:1. Raw Data Report2. Absorbance report3. Limit Report4. Matrix
Report5. Cut Off report6. Curve Fit Report7. Concentration Report8.
Difference report2.5.5 RS-232 Interface
The Model 680 Microplate Reader has a built-in RS232C interface
(9 pin D-sub connector). This allows external computers to control
the instrument. When an external computer is in control of the
microplate reader, the printer and the keypad keys (except
Start/Stop) are automatically deactivated and the display will
indicate Remote Mode.The reader will remain in the remote control mode until it
receives RL command from the host computer or until the Start/Stop
key is pressed.RS-232 communication parameters are as follows:
Signals: TDX: Transmit DataRXD: Receive DataDTR: Data Terminal
ReadyDSR: Data Set ReadyCTS: Clear to SendRTS: Ready to SendTransmission: Baud Rate 9600, Data Width 8, Stop Width 1, Parity
NoneData: ASCII (no distinction between upper and lower case)
Signal Assignment:
Host Computer9 Pin D-sub
Model 680 Reader9 Pin D-sub
Receive Data 2 ———————— 3 Transmit Data
Transmit Data 3 ———————— 2 Receive Data
Data Terminal Ready 4 ———————— 6 Data Set
ReadySignal Ground 5 ———————— 5 Signal Ground
Data Set Ready 6 ———————— 4 Data Terminal
ReadyRequest to Send 7 ———————— 8 Clear to Send
Clear to Send 8 ———————— 7 Request to Send
-
System Overview
100-2414C 03/04 2-5
2.6 Instrument Specifications
The Model 680 and 680 XR Microplate Reader specifications are as
follows:2.6.1 Features
Operating Panel: Membrane pad with 4 arrow keys, 7 function keys
and 11 numeral keysDisplay: LCD, 4 Lines x 20 Alphabet Characters
Computer Interface: Bi-directional RS-232C serial communication
port, 9 pin D-sub receptacleExternal Printer Port (only when internal printer option is not
installed): Centronics interface with ASCIIcharacter code; 25 pin
D-sub receptacle; 80 characters output per lineReport Types: Raw, absorbance, matrix, limit, cutoff,
concentration, curve fit, difference, kineticApplicable Plates: Standard 96 well microtitre plate of
polystyrene — flat, U, V rigid bottomMixing: 3 speeds (low, medium, high); Programmable time period
0-999 secondsWarm-up Time: 3 minutes
2.6.2 Optical
Photometric Methods: Single and dual wavelength
Photo detectors: 8 silicon diodes for measurement and 1 silicon
photodiode for referenceLight Source: Tungsten halogen lamp, 20 Watts with 3000 hours
average lifetimeModel 680 Spectral Range: 400 nm to 750 nm
Model 680 XR Spectral Range: 340 nm to 800 nm
Interference Filters: Wavelength Tolerance +/- 3 nm; Band Width:
10 nm typicalFilter Capacity: 8 filters
Model 680 Standard Filters: 415, 450, 490, 655 nm
Model 680 XR Standard Filters: 340, 405, 415, 655 nm
Optional Filters: 340 through 800 nm (in 5 nm increments)
Fast Reading Speed: 6 seconds with single wavelength reading; 10
seconds with dual wavelengthreadingStandard Reading Speed: 15 seconds with single wavelength
reading; 30 seconds with dual wave-length readingIndication Range: 0.000 to 3.5 OD
2.6.3 Performance
Resolution: 0.001 OD
Accuracy: +/- 1.0% (or 0.010 from 0 to 3.0 OD at 490 nm)
Linearity: +/- 1.0% (from 0 to 2.0 OD); +/- 2.0% (from 2.0 to
3.0 OD)Reproducibility: 1.0% (or 0.005 from 0 to 2.0 OD); 1.5% (from
2.0 to 3.0 OD)Inter-Channel Variation: 1.5% (or 0.005 from 0 to 3.0 OD)
Stability and Drift: 0.010 OD at OD=1 with single wavelength
reading -
System Overview
2-6 100-2414C 03/04
2.6.4 Thermal
Incubator (Optional): Temperature Set Point: 25oC to 45oC (at
least 5oC above ambient)Set Point Accuracy: +/- 0.5oC (set point = 37oC and room
temperature = 25oC)Well-to-Well Uniformity: 0.5oC (set point = 37oC and room
temperature = 25oC)Warm-up Time: 5 minutes (set point = 37oC and room temperature =
25oC)2.6.5 Electrical
Line Voltage: Universal Input: 100-120 VAC, 200-240 VAC, 50/60
HzFuse: T3.15 Amp, 250 Volts, 20mm x 5mm
Power Consumption: 100 VA maximum
Battery: Lithium, CR2032, 3 Volts; average lifetime 5 years
2.6.6 Physical
Dimensions: 340 x 330 x 152 mm
Weight: 5.5 kg
Operating Environment: 5oC to 35oC, 0 to 95% RH, non
condensingStorage Environment: -20oC to 50oC, 0 to 95% RH, non
condensing -
Installation
100-2414C 03/04 3-1
Section 3Installation
3.1 Introduction
The Model 680 and 680 XR Microplate Readers are easy to install.
Prior to installation, verify that all systemcomponents have been
received.3.2 Visual Check
1. Check this instrument for any signs of damage which may have
occurred during shipping.2. Check the filters: Open the rear cover by pushing the release
latch on the top. Check that the filters are installed into the
proper positions per Table 1:Table 1 — Filter Installation
3. Check to insure filters are properly seated (Figure 1).
Figure 1: Filter Check
Filter Position Model 680 Model 680XR
1 415 nm 340 nm
2 450 nm 405 nm
3 490 nm 415 nm
4 655 nm 655 nm
5 User User
6 User User
7 User User
8 User User
-
Installation
3-2 100-2414C 03/04
3.3 Installation Procedure
1. Place the instrument on a clean, sturdy table or bench. It is
important to keep the instrument in a clean, relatively dust free
environment to ensure maximum performance.2. Connect the power cord to the connection on the back if the
instrument.3. Turn ON the power switch on the rear panel. The LCD will
display the version number of the on board firmware. After about
three seconds, the instrument will perform a self test diagnosis
that takes about fifteen seconds.4. When prompted by the Log In screen, enter the password
(password = 00000). Press Enter.5. Allow three minutes for the instrument to warm up prior to
reading plates.3.4 Install Paper Roll (Optional Internal Printer)
1. Open the rear cover by pushing the release latch on the
top.2. Tear off a small piece of paper to form a diagonal point on
the end of the paper roll.3. Place the paper roll in the holder positioned below the
printer and orient the roll so that it will feed from the bottom
(Figure 2).Figure 2: Orientation of Paper Roll
4. While pressing the Paper Feed key, direct the pointed end of
the paper into the paper feed slot in the underside of the printer
until the printer grabs the paper and feeds it through the slot in
the top of the instrument. -
Installation
100-2414C 03/04 3-3
3.5 Computer Interface (Optional)
1. Some customers may utilize the system with a computer and
microplate software such as Microplate Manager (or Checkmark).2. Set up computer as required and install software.
3. Install serial cable from computer to Model 680 oe 680
XR.4. Turn on instrument, enter password.
5. Verify communication with instrument using Microplate
Manager.Note: Microplate Manager version 5.2.1 is required for Model
680XR support. -
System Checkout
100-2414C 03/04 4-1
Section 4System Checkout
4.1 Introduction
This section provides the system checkout procedure for the
Model 680 and 680 XR Microplate Reader. Suc-cessful completion of
the tests in this section represent the Operational Qualification
of the instrument. A Sys-tem Checkout Report is available in
Appendix A and B.The system checkout procedures for the Model 680 and 680 XR
Microplate Readers require the use of aremote computer using
Microplate Manager software.The Model 680 and 680 XR Microplate Reader contains an on board
checkout procedure, Checkmark Valida-tion Program, which requires
the use of a special Checkmark test plate.4.2 Tools and Materials
Microplate Manager Software, 170-9520
RS-232 Cable, 931-0184
Printer
Digital Temperature Multi-meter with 0.1 mm diameter bead type K
thermoprobe (minimum +/- 0.5 C temperature meter accuracy)96 Well Microtitre Plate, Flat Bottom, 224-0096
Micro Pipette, 50L
Pipette Tips
50 ml beaker
Acid Yellow 99, Fisher AC19178-100
Distilled Water
Tween 20, 170-6531
Checkmark Reader Performance Check Plate, 170-6940
(Optional) -
System Checkout
4-2 100-2414C 03/04
4.3 Summary of Tests and Specifications
The following is a summary of the tests and specifications for
the Model 680 and 680 XR Microplate Readers.Table 1: Summary of Tests and Specifications
Test Parameter Specification
Firmware Firmware Version Record Version
Instrument Function Keypad and Display Proper Function
Internal Printer (optional feature) Proper Function
Startup and Self Diagnosis Proper Function
RS-232 Communication Proper Communication
Mixing Feature Proper Function
Incubator Set Point 37 C (optional feature) 37 C +/- 0.5 C
System Checkout Air Read < 0.005 OD
Accuracy CV < 5%
Channel-to-Channel CV < 5%
Stability and Drift < 0.010 OD
-
System Checkout
100-2414C 03/04 4-3
4.4 System Checkout Procedure
4.4.1 General
1. All system checkout procedures in this section will be
performed with Microplate Manager via PC computer.2. The Checkmark Validation Program, if available, may be used
for system checkout as an alternate method (paragraph 4.5).4.4.2 Setup
1. Setup computer.
2. Install Microplate Manager.
3. Install HPK protection key.
4. Connect RS-232 cable from computer to reader.
4.4.3 Preparing Yellow Dye Test Plate
1. System checkout requires the preparation of a yellow dye
pipetted into a microtitre plate. Pipetting must be of extreme
accuracy as the Model 680 reader is sensitive enough to detect
pipetting errors.2. Prepare the yellow dye as follows:
Mix: 2 mg Acid Yellow 9950 ml Distilled Water0.01 ml Tween
203. Pipette 200 microliters yellow dye into each well of the
microtitre plate (96 total).4. The value of the prepared test plate is 0.424 — 0.444 OD.
4.4.4 Startup
1. Turn ON the reader power switch. The LCD will display the
version number of the on board firmware. Record Firmware
Version.2. After about three seconds, the instrument will perform a self
test diagnosis that takes about ten sec-onds.3. When prompted by the Log In screen (Administrator), enter the
password 00000. Press Enter.4.4.5 Touch Pad/Mixing Function
1. Open chamber door (leave open for this test).
2. Press Memory Recall button.
3. Select Protocol and press Enter.
4. Select End Point and press Enter.
5. Select End Point 01 and press Enter twice.
6. Press the Start button.
7. Observe that plate carrier vibrates for 3 seconds prior to
read. -
System Checkout
4-4 100-2414C 03/04
4.4.6 Verify Printer (Optional Feature)
1. After completion of the above End Point 01 test, press the
Print button.2. Select Reports and press Enter.
3. Verify the printer is functioning properly.
4.4.7 Start Microplate Manager
1. Start Microplate Manager.
2. Select File…Select Reader. Select Model 680 or Model 680 XR
and click OK.3. Select New Endpoint Protocol. The Microplate Manager window
will appear (Figure 1).Figure 1: Microplate Manager — Endpoint Protocol
4. The instrument will be placed in remote mode. Remote Mode
will appear on display and touch keys will become
nonfunctional. -
System Checkout
100-2414C 03/04 4-5
4.4.8 Air Read
1. This test is performed without the use of a microtiter
plate.2. Press Show Template button.
3. Click the Turn Off Autofill Icon.
4. Click the Format as Unknowns Icon.
5. Using the curser, click and drag highlighting all rows (A1 to
G12).6. All wells on the template should be squares filled with 1s
(see Figure 2).Figure 2: Plate Template
7. Minimize the plate template.
8. Click on Reports.
9. Check the Raw Data Report. Click OK. (See Figure 3).
Figure 3: Raw Data Report Setup
-
System Checkout
4-6 100-2414C 03/04
10. Check the Reading Parameters Single box.
11. Select Measurement filter: 490.
12. Click the Run button.
13. Plate will take a few moments to read and display the Raw
Data Report.14. Ensure that all values are < 0.005. Reference Figure
4.Figure 4: Raw Data Report
-
System Checkout
100-2414C 03/04 4-7
4.4.9 Accuracy
1. This test is performed with the use of the yellow dye test
plate.2. Use the formatted plate template from the Air Read above.
3. Click on Reports.
4. Check the Raw Data Report and Unknown Concentrations. Click
OK. See Figure 5.Figure 5: Report Selection
5. Check the Reading Parameters Single box.
6. Select Measurement filter: 490.
7. For Mix Time (sec) enter 3.
8. Click the Run button.
9. Ensure the CV is < 5% on the Unknown Concentrations Report
(Figure 6) and target values are within expected ranges (Mean
OD).Figure 6: Unknown Concentrations Results
-
System Checkout
4-8 100-2414C 03/04
4.4.10 Channel-to-Channel Variation
1. This test is performed with the use of the yellow dye test
plate.2. Press Show Template button.
3. Click the Turn Off Autofill Icon.
4. Click the Format as Unknowns Icon.
5. Using the curser, click and drag highlighting rows A1 to A12.
The number 1 should appear in each box. Repeat highlighting rows B1
to B12. The number 2 should appear in each box. Repeat for the
remaining rows C1 to C12, D1 to D 12, etc. All wells on the
template should appear as shown in Fig-ure 7.Figure 7: Channel to Channel Setup
6. Minimize the plate template.
7. Use the same format for reports (Raw Data Report and Unknown
Concentrations), filter (490) and mix time (3).8. Click the Run button.
9. Plate will take a few moments to read and display the Raw
Data Report and Unknown Concentra-tions Report.10. Ensure the CVs are < 5% on the Unknown Concentrations
Report (Figure and target values are within expected ranges
(Mean OD). -
System Checkout
100-2414C 03/04 4-9
Figure 8: Unknown Concentrations Report
4.4.11 Stability and Drift
1. This test uses the same setup as specified in 4.4.11
Channel-to-Channel Variation above.2. Rerun the Channel-to-Channel variation test 3 times allowing
10 minutes between tests.3. Compare the data from the three tests. The average mean from
each row shall be within 0.010 OD of each other. -
System Checkout
4-10 100-2414C 03/04
4.4.12 Verify Incubator Temperature (Optional Feature)
1. Place checkmark in Incubator On box. Insure set point is 37
C.2. Place a standard 96 well microtitre plate filled with 200
microliters water in wells A1, A6, A12, D6, D9, H1, H6, and H12 in
the plate carrier. See Figure 9.Figure 9: Designated Temperature Test Points
3. Allow microtiter plate to warm for 30 minutes.
4. After 30 minutes, immerse the thermoprobe in the designated
well filled with the 200 microliters of water without touching any
portion of the inner well wall. The immersed probe must be in the
middle of the well and immersed in water. See Figure 10.Figure 10: Well Temperature Measurement
5. Verify that the actual well temperature is 37 C +/- 0.5 C for
each well (A1, A6, A12, D6, D9, H1, H6, and H12). -
System Checkout
100-2414C 03/04 4-11
4.5 Checkmark Validation Program (Optional Method)
4.5.1 General
Figure 11: Checkmark Check Plate
1. The Checkmark program is used with the Checkmark Reader
Performance Check Plate for validat-ing the performance of the
Model 680 Microplate Reader. It will test the following
parameters:Default Measurements
Accuracy
Precision
Linearity
Optional Measurements
Plate Positioning
Cross-talk
Filter Check
0 OD and Overflow
IR Spectral Blocking
2. The Checkmark program resides in the Model 680 Microplate
Reader programs.4.5.2 Procedure
Entering Checkmark Plate Reference Data
1. Turn on Model 680 Reader and enter password.
2. Press Memory Recall button.
3. Select Protocol and press Enter.
4. Select Checkmark and press Enter. Press any key if
instructed.5. Press the Edit button.
6. Select Protocol and press Enter.
-
System Checkout
4-12 100-2414C 03/04
7. Select Serial/Report and press Enter
8. Enter serial number of Checkmark plate. Select report type
Simple or Full. Press Enter.9. Select Linearity and press Enter.
10. Enter appropriate data from Checkmark reference plate data.
Press enter when complete.11. Select Max CV and press Enter.
12. Enter appropriate data from Checkmark reference plate data.
Press enter when complete.13. Select Filter Check Reference and press Enter.
14. Enter appropriate data from Checkmark reference plate data.
Press enter when complete.Running the Test
1. Open chamber door, insert Checkmark Reader Performance Check
Plate, and close door.2. Press Memory Recall button.
3. Select Protocol and press Enter.
4. Select Checkmark and press Enter. Press any key if
instructed.5. Press the Start button.
6. Test will take about 3 minutes to read plate.
-
Preventative Maintenance
100-2414C 03/04 5-1
Section 5Preventative Maintenance
5.1 Introduction
This section provides the preventative maintenance procedure for
the Model 680 and 680 XR MicroplateReader. Preventative maintenance
is recommended at least once a year.5.2 Tools and Materials
Small Hand Tools
Cleaning Materials (Lint Free Cloth, Lens Tissue, Isopropyl
Alcohol)5.3 Preventative Maintenance
5.3.1 Firmware
1. Connect the power cord to the connection on the back if the
instrument.2. Turn ON the power switch on the rear panel. The LCD will
display the version number of the firm-ware.3. Record firmware version.
4. Turn OFF instrument.
5. Update to latest firmware if applicable (see Section
7.6).5.3.2 Inspection of Instrument
1. Remove top cover.
2. Visually inspect for cleanliness of plate carrier. Clean
plate carrier and rails with lint free cloth and isopropyl
alcohol.Note: Do not use grease on the plate carrier or rails.
3. Visually inspect for cleanliness of lenses. Clean as required
with lens tissue and isopropyl alcohol (see Section 5.3.3).4. Reassemble instrument.
-
Preventative Maintenance
5-2 100-2414C 03/04
5.3.3 Optics Cleaning
1. Remove top cover.
2. To access top lenses, remove the two screws securing the
fiber optics to the cover (Figure 5-1).Figure 5-1: Remove Screws Securing Fiber Optics
3. Remove two hex standoffs securing top lens block.
4. To remove lenses, remove ten securing screws (Figure
5-2).Figure 5-2: Top Lens Block
5. Clean lenses with lint free cloth and isopropyl alcohol.
6. Reassemble lenses into holder and remount.
Note: Check for proper orientation when remounting.
-
Preventative Maintenance
100-2414C 03/04 5-3
7. To access bottom lenses, remove cover (Figure 5-3).
Figure 5-3: Bottom Lenses
8. Clean lenses as required. If removal of bottom lenses is
required for cleaning, remove plate carrier and analog board (see
Section 6.3.12).9. Remove bottom lens block (Figure 5-4).
Figure 5-4: Bottom Lens Block Screws
-
Preventative Maintenance
5-4 100-2414C 03/04
10. Remove lenses from holder to clean (Figure 5-5).
Figure 5-5: Bottom Lens Holder
11. Clean with lint free cloth and isopropyl alcohol.
12. Reassemble instrument. Note proper orientation of bottom
lens holder upon resaaembly.5.3.4 Perform System Checkout
1. Perform system checkout per Section 4.0 of this Service
Manual. -
Removal and Replacement
100-2414D 09/05 6-1
Section 6Removal and Replacement
6.1 Introduction
This section provides the parts removal and replacement
procedures for the Model 680 and 680 XR MicroplateReader.6.2 Tools and Materials
Standard Hand Tools
Voltmeter
6.3 Removal and Replacement Procedures
6.3.1 Interference Filters
1. Open the rear cover by pushing the release latch on the
top.2. Rotate the filter wheel until the desired filter is in top
position. Grasp top of filter frame to and pull up to remove
filter.3. To install filter, insert into desired slot until filter
locks into place.6.3.2 Model 680 Lamp (1681006)
1. Open the rear cover by pushing the release latch on the
top.Caution: Surface of Cover may be Hot!
2. Remove the two screws that hold the black plastic cover in
place (Figure 6-1) and remove cover.Figure 6-1: Screws Securing Cover
-
Removal and Replacement
6-2 100-2414D 09/05
3. Remove the screw which holds the metal cover in place (Figure
6-2) and remove cover.Figure 6-2: Screw Securing Metal Cover (Caution: Hot)
4. Carefully pull the lamp free and remove (Figure 6-3).
Figure 6-3: Lamp Removal
5. Install new lamp in reverse order of the above steps.
-
Removal and Replacement
100-2414D 09/05 6-3
6.3.3 Model 680 XR Lamp (1681104)
1. Open the rear cover by pushing the release latch on the
top.Note: Surface of Cover may be Hot! Allow to cool before
replacing lamp.2. Remove the two screws (arrows) that hold the black plastic
cover in place (Figure 6-4) and remove cover assembly.Figure 6-4: Screws Securing Cover
3. Remove the lamp (Figure 6-5).
Figure 6-5: Lamp Removal
4. Install new lamp.
Note: Do not tough new lamp with fingers. Oil on lamp may reduce
lamp life.5. Reassemble in reverse order of the above steps.
-
Removal and Replacement
6-4 100-2414D 09/05
6.3.4 Top Cover (9310161)
1. Remove the six screws from the underside of the instrument
(Figure 6-6).Figure 6-6: Remove Six Screws from Underside of Instrument
2. Remove the top cover to left side of the instrument (Figure
6-7).Figure 6-7: Remove Top Cover to Left Side of Instrument
3. Install cover in reverse order of the above steps.
-
Removal and Replacement
100-2414D 09/05 6-5
6.3.5 Printer Assembly (9310160)
1. Remove top cover.
2. Cut the cable tie securing the cable and remove cable from
the circuit board (Figure 6-8).Figure 6-8: Cut Cable Tie and Remove Cable
3. Remove three screws securing printer assembly and remove
assembly (Figure 6-9).Figure 6-9: Three Screws Securing Printer Assembly
4. Prior to installing printer assembly, set the jumpers on the
printer control board as follows: -
Removal and Replacement
6-6 100-2414D 09/05
5. Check printer head type on flexible cable (Figure 6-10).
Figure 6-10: Location of Head Type
6. Configure the JP4 jumpers in accordance with the Figure
6-11.Figure 6-11: JP4 Jumper Settings
7. Install new printer in reverse order of the above steps.
-
Removal and Replacement
100-2414D 09/05 6-7
6.3.6 Keypad Membrane (9310177, 9310277, 9310229)
1. Remove top cover.
2. Cut cable tie and disconnect three cables from keypad board
(Figure 6-12).Figure 6-12: Keypad Board
3. Remove three screws and lift out keypad board (Figure
6-13).Figure 6-13: Screws Securing Keypad Board
4. Peel off keypad membrane (Figure 6-14).
Figure 6-14: Peel Off Keypad Membrane
5. Install new keypad membrane in reverse order of the above
steps. -
Removal and Replacement
6-8 100-2414D 09/05
6.3.7 Inlet Assembly (9310167)
1. Remove top cover.
2. Disconnect the terminals from board (Figure 6-15).
Figure 6-15: Disconnect Terminals
3. Remove nuts from interface connectors and remove (Figure
6-16).Figure 6-16: Remove Nuts from Connectors
4. Remove four screws securing the fan (Figure 6-17).
Figure 6-17: Fan Mounting Screws
5. Remove inlet assembly.
6. Install new inlet assembly in reverse order of the above
steps. -
Removal and Replacement
100-2414D 09/05 6-9
6.3.8 Plate Carrier Assembly and Analog Board
Note: Check the model and serial number for the correct
replacement board.Table 6-1: Analog Board Part Numbers
1. Remove top cover.
2. Loosen the two screws which secure the fiber and pull out
fiber (Figure 6-18).Figure 6-18: Fiber Mounting
3. Remove the nine mounting screws (Figure 6-19).
Figure 6-19: Plate Carrier Mounting Screws
Model 680 Model 680 XR
Analog Board 9310182 9310289
-
Removal and Replacement
6-10 100-2414D 09/05
4. Lift the plate carrier assembly, disconnect the cables and
remove assembly (Figure 6-20).Figure 6-20: Disconnect Cables
5. Remove two screws securing the cover over the analog board
(Figure 6-21).Figure 6-21: Analog Board Mounted to Plate Carrier
-
Removal and Replacement
100-2414D 09/05 6-11
6. Carefully remove fiber from mounting clips (Figure 6-22).
Figure 6-22: Clips Securing Fiber
7. Remove four screws securing analog board and remove (Figure
6-23).Figure 6-23: Analog Board Mounting Screws
8. Install new plate carrier assembly or analog board in reverse
order of the above steps. -
Removal and Replacement
6-12 100-2414D 09/05
6.3.9 CPU Board
Note: Check the model and serial number for the correct
replacement board.Table 6-1: CPU Board Part Numbers
1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Remove all cables to CPU board.
4. Remove four mounting screws and remove board (Figure
6-24).Figure 6-24: CPU Board Mounting
5. Install new CPU board in reverse order of the above
steps.6. Set the jumper on the CPU Board (#281) per the following
table:Model 680Old LCD
Model 680New LCD
Model 680Backlit LCD
Model 680 XR
Serial Number 10001 to 10735 10736 to 13850 13851 and higher
AllCPU Board 9310181 9310225 9310353 9310278
Table 6-2: JP2 Jumper Settings for CPU Board
Instrument Configuration
Jumper Jumper Placement
Without Incubator JP2 Jump
With Incubator JP2 No Jumper (Open)
-
Removal and Replacement
100-2414D 09/05 6-13
6.3.10 Incubator Control Board (9310180)
1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Remove all cables to incubator control board.
4. Remove four mounting screws and remove board (Figure
6-25).Figure 6-25: Incubator Control Board Mounting
5. Install new incubator control board in reverse order of the
above steps.6. Perform incubator electronics adjustment as follows.
Disconnect cable from plate carrier to CN52.7. Connect cable into Incubation Adjustment Tool, 9310183,
107.79 ohm connector on tool.8. Turn on instrument.
9. Measure voltage between TP5 and TP0 (see Figure 6-26).
-
Removal and Replacement
6-14 100-2414D 09/05
Figure 6-26: Incubator Board Adjustments
10. Adjust trim pot VR1 until the voltage measures 0.000V +/-
0.001V.11. Connect cable into Incubation Adjustment Tool, 116.47 ohm
connector on tool.12. Measure voltage between TP5 and TP0.
13. Adjust trim pot VR1 until the voltage measures 5.528V +/-
0.006V.14. If voltage is out of range, repeat steps 3 through 9.
-
Removal and Replacement
100-2414D 09/05 6-15
6.3.11 Model 680 XR Lamp Control PCB (9310276)
1. Open top cover.
2. Remove ground screw connection and four mounting screws and
lift up the filter wheel/lamp assem-bly (Figure 6-27).Figure 6-27: Lamp Control PCB is Located Under Lamp Assy
3. For ease of disassembly, turn over assembly. To access lamp
control board, remove the five screws securing the lamp PCB cover
(Figure 6-28).Figure 6-28: Bottom View — Lamp PCB Cover Screws
4. Remove cover.
Note: Heat sink will be removed as a result of removing the two
side screws.Ground Screw
Two Screws Backside (Not Shown)
-
Removal and Replacement
6-16 100-2414D 09/05
5. Remove the two remaining screws securing the PCB (Figure
6-29).Figure 6-29: Lamp Control PCB
6. Remove three cables to lamp control PCB and lift out.
7. Install new lamp control PCB by securing with two mounting
screws and attaching the three cables.8. Install the cover and heat sink. For ease of assembly,
position the heat sink in place prior to placing over lamp control
PCB (Figure 6-30).Figure 6-30: Position Heat Sink
9. Install the two screws heat sink side first, then the two
remaining screws securing the cover.10. Reassemble in reverse order of the above steps.
-
Removal and Replacement
100-2414D 09/05 6-17
6.3.12 Power Supply, 12 Volt (9310178)
1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Remove four screws securing lamp housing and remove (Figure
6-31).Figure 6-31: Lamp Housing Mounting
4. Remove all cables to 12 volt power supply.
5. Remove four mounting screws and remove power supply (Figure
6-32).Figure 6-32: 12 Volt Power Supply Mounting
6. Install new 12 volt power supply in reverse order of the
above steps. -
Removal and Replacement
6-18 100-2414D 09/05
6.3.13 Power Supply, 24 Volt (9310179)
1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Remove four screws securing lamp housing and remove (Figure
6-33).4. Remove all cables to 24 volt power supply.
5. Remove four mounting screws and remove power supply (Figure
6-33).Figure 6-33: 24 Volt Power Supply Mounting
6. Install new 24 volt power supply in reverse order of the
above steps. -
Removal and Replacement
100-2414D 09/05 6-19
6.3.14 Photo Sensor — Plate Carrier/Analog Board (9310174)
1. Remove top cover.
2. Open the clip securing the optical fiber assembly. Remove the
three screws securing the cover and remove (Figure 6-34).Figure 6-34: Remove Clip and Mounting Screws
3. Remove the screw securing the photo sensor and remove (Figure
6-35).Figure 6-35: Removing the Photo Sensor
4. Install new photo sensor in reverse order of the above
steps.5. Check plate positioning per Section 7.5.
-
Removal and Replacement
6-20 100-2414D 09/05
6.3.15 Photo Sensor — Filter Wheel (9310174)
1. Remove top cover.
2. Remove cable from photo sensor. Remove screw securing photo
sensor and remove (Figure 6-36).Figure 6-36: Photo Sensor Removal
3. Install new photo sensor in reverse order of the above
steps. -
Removal and Replacement
100-2414D 09/05 6-21
6.3.16 Model 680 Light Balancing Filter (9310173)
1. Remove top cover.
2. Remove plastic cover and metal cover from atop lamp.
3. Remove two screws securing lamp housing (Figure 6-37).
Figure 6-37: Lamp Housing
4. Remove two screws securing the light balancing filter (Figure
6-38).Figure 6-38: Light Balancing Filter
5. Install new light balancing filter in reverse order of the
above steps. -
Removal and Replacement
6-22 100-2414D 09/05
6.3.17 Model 680 XR Light Balancing Filter (9310275)
1. Remove top cover.
2. Remove two screws (arrows) securing lamp housing assembly
(Figure 6-39).Figure 6-39: Lamp Housing
3. Remove two recessed screws securing the light balancing
filter (Figure 6-40).Figure 6-40: Light Balancing Filter
4. Install new light balancing filter in reverse order of the
above steps. -
Removal and Replacement
100-2414D 09/05 6-23
6.3.18 Plate Carrier Motor (9310172)
1. Remove top cover.
2. Remove the three cover screws and lift away (Figure
6-41).Figure 6-41: Remove Inner Cover
3. Remove the two screws from the rail cover and remove cover
(Figure 6-42).Figure 6-42: Cover Screws
-
Removal and Replacement
6-24 100-2414D 09/05
4. Loosen the screw to the belt and remove belt Figure
6-43).Figure 6-43: Loosen Belt Adjustment Screw
5. Remove two screws securing motor (Figure 6-44). Remove and
unplug motor.Figure 6-44: Motor Mounting Screws
6. Install new motor in reverse order of the above steps.
-
Removal and Replacement
100-2414D 09/05 6-25
6.3.19 Optical Fiber Assembly (9310169)
1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Open the clip securing the optical fiber assembly. Remove the
three screws securing the cover and remove (Figure 6-45).Figure 6-45: Removing Optical Fiber Assembly
4. Remove two screws securing the cover over the analog
board5. Carefully remove fiber from mounting clips (Figure 6-46).
Figure 6-46: Remove Cover Screws
6. Install new optical fiber assembly in reverse order of the
above steps. -
Removal and Replacement
6-26 100-2414D 09/05
7. When re-installing fiber into filter wheel housing, ensure
that end is protruding 2 mm from plate (Fig-ure 6-47).Figure 6-47: Fiber Installation
-
Removal and Replacement
100-2414D 09/05 6-27
6.3.20 Filter Wheel (9310170) & Filter Wheel Motor
(9310171)1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Cut cable tie and remove two screws securing the filter wheel
assembly (Figure 6-48).Figure 6-48: Filter Wheel Assembly Mounting
4. Unplug cable from motor to analog board.
5. Remove two screws securing filter wheel to motor (Figure
6-49).Figure 6-49: Filter Wheel Mounting
6. Remove two screws securing motor to mounting bracket.
7. Install new filter wheel or motor in reverse order of the
above steps. -
Removal and Replacement
6-28 100-2414D 09/05
6.3.21 Fan (9310168)
1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Remove 4 screws and move lamp assembly in order to facilitate
unpluging of fan cable (Figure 6-50).Figure 6-50: Lamp Assembly Mounting
4. Remove four mounting screws and remove fan (Figure 6-51).
Figure 6-51: Fan Mounting
5. Install new fan in reverse order of the above steps.
-
Removal and Replacement
100-2414D 09/05 6-29
6.3.22 Lithium Battery (9006020)
1. Open top cover.
2. Remove plate carrier assembly (see 6.3.8).
3. Replace the lithium battery located on the CPU board (Figure
6-52).Figure 6-52: Lithium Battery
4. Install new lithium battery in reverse order of the above
steps. -
Adjustments, Calibration and Checks
100-2414E 02/06 7-1
Section 7Adjustments, Calibration and Checks
7.1 Introduction
This section provides the adjustments, calibration and checks
applicable to the Model 680 and Model 680 XRMicroplate Readers.7.2 Tools and Materials
Standard Hand Tools
Voltmeter
9310183 Incubator Adjustment Tool
9310184 RS-232C Cable
1002446 Firmware CD
7.3 Electronics Adjustment
7.3.1 Setting of Jumper on the CPU Board (#281)
1. Open the top cover and remove plate carrier assembly to
access CPU Board (#281).2. Locate JP2 (Figure 7-1).
Figure 7-1: Jumper JP2
3. Set the jumper on the CPU Board (#281) per the following
table:Table 7-1: JP2 Jumper Settings for CPU Board
Instrument Configuration
Jumper Jumper Placement
Without Incubator JP2 Jump
With Incubator JP2 No Jumper (Open)
-
Adjustments, Calibration and Checks
7-2 100-2414E 02/06
7.3.2 Setting of the Jumpers on the Printer Control Board
(#284)1. Open the top cover to access Printer Control Board
(#284).2. Check printer head type on flexible cable (Figure 7-2).
Figure 7-2: Location of Head Type
3. Configure the jumpers to JP4 in accordance with the Figure
7-3.Figure 7-3: JP4 Jumper Settings
-
Adjustments, Calibration and Checks
100-2414E 02/06 7-3
7.3.3 Incubator Electronics Adjustment (Board #284)
Note: This procedure is used to set the proper temperature set
points for the incubator.1. Open the top cover and remove plate carrier assembly to
access Incubator board (#284).2. Disconnect cable from plate carrier to CN52.
3. Connect cable into Incubation Adjustment Tool, 9310183,
107.79 ohm connector on tool.4. Turn on instrument.
5. Measure voltage between TP5 and TP0 (Figure 7-4).
Figure 7-4: Incubator Board Adjustments
6. Adjust trim pot VR1 until the voltage measures 0.000V +/-
0.001V.7. Connect cable into Incubation Adjustment Tool, 116.47 ohm
connector on tool.8. Measure voltage between TP5 and TP0.
9. Adjust trim pot VR1 until the voltage measures 5.528V +/-
0.006V.10. If voltage is out of range, repeat steps 3 through 9.
-
Adjustments, Calibration and Checks
7-4 100-2414E 02/06
7.4 Adjusting Plate Carrier Belt Tension
1. This procedure is used to adjust the belt tension of the
plate carrier.2. Open the top cover.
3. Remove fiber optics cover (Figure 7-5)
Figure 7-5: Remove Fiber Optics Cover
4. Remove two screws securing the belt cover as shown in Figure
7-6.Figure 7-6: Remove Belt Cover
-
Adjustments, Calibration and Checks
100-2414E 02/06 7-5
5. Loosen the screw and adjust the belt tension. Tension should
be firm but not overly tight.Figure 7-7: Belt Tension Adjustment
-
Adjustments, Calibration and Checks
7-6 100-2414E 02/06
7.5 Alignment of Plate Carrier
7.5.1 Checking Plate Alignment
1. This procedure is used to check positioning and alignment of
the plate carrier.2. Open the top cover.
3. Install Checkmark Plate into reader. Make sure plate is
properly seated.4. Apply clear tape over small holes in columns 1 and 12 (Figure
7-8).Figure 7-8: Placement of Tape
5. Use QC software to move plate beneath optics path per Section
7.7.3.6. Check that the light is centered about the holes in the
Checkmark Plate in columns 1 and 12 (Figure 7-9).Figure 7-9: Checking Plate Position at Columns 1 & 12
-
Adjustments, Calibration and Checks
100-2414E 02/06 7-7
7.5.2 Adjusting Plate Alignment — Y Direction
1. This procedure is used to adjust the Y direction (top/bottom)
position of the plate carrier by reposi-tioning of the rails.2. Open the top cover.
3. Loosen the 12 screws shown in Figure 7-10 and adjust the
rails as required.Figure 7-10: Y Direction Adjustment — Reposition Rails
7.5.3 Adjusting Plate Alignment — X Direction
1. This procedure is used to adjust the X direction (left/right)
position of the plate carrier by repositioning the photo
sensor.2. Open the top cover.
3. Loosen the screws shown in Figure 7-11 and reposition the
photo sensor as required.Figure 7-11: X Direction Adjustment — Reposition Sensor
-
Adjustments, Calibration and Checks
7-8 100-2414E 02/06
7.6 Firmware
7.6.1 General
1. Downloading firmware is achieved through a PC (Windows 95,
NT, 2000, XP) and an RS-232 cable.2. The communication requirements are as follows:
RS232CPort: COM1Baud Rate: 9600, 57600Data: 8 bit 1 stop
3. The downloading program and additional files are required for
downloading:m680_up.exe mscomctl.ocxcomdlg32.ocx msvbvm60.dllmscomm32.ocx
asycfilt.dllcomcat.dll mci32.ocxoleaut32.dll
olepro32.dllvb6stkit.dll4. The current firmware files are:
M680_V171OL.txt for Original Lens models (SN 10001 through
10735)M680_V171NL.txt for New LCD models (SN 10736 through
13850)M680_V171.txt for Backlit LCD models (SN 13851 and higher)
M680_V171.txt for Chinese version instrumentsM680XR_V163-XR.txt
(for Model 680XR readers)5. When updating firmware versions 1.60 and earlier with
m680_up.exe, plate reading results, security password, Checkmark
reference values, Checkmark reading results and laboratory name
will be lost. Protocol settings (except for #5-8) and filter
wavelengths will be retrieved and restored in the update process.
Date and time will need to be reset.6. When updating firmware versions 1.61 and later with
m680_up.exe, all data and settings except for security password,
Checkmark reading results, Protocol settings (except for #5-8) and
Plate Reading results (#5-8) be retrieved and restored in the
update process. Date and time will need to be reset. -
Adjustments, Calibration and Checks
100-2414E 02/06 7-9
7.6.2 Firmware HistoryTable 7-2: Firmware History
Instrument Style
Firmware Version
History
Compatibility
Original LCD
New LCD
Backlit LCD
Original LCDSerial Numbers 10001 — 10735
1.59 Initial release (compatible with serial num-bers 10001
through 10735)Yes No No
1.60 Minor corrections (compatible with serial numbers 10001
through 10735)Yes No No
1.62 Added TSE protocols (compatible with serial numbers 10001
through 10735)Yes No No
1.63 Changed TSE validation criteria (compatible with serial
numbers 10001 through 10735)Yes No No
1.71OL Added functionality for ESC/P printer. Added protocols
for TSE Sheep/Goat and Rabies (compatible with serial numbers 10001
through 10735)Yes No No
New LCDSerial Numbers 10736 — 13850
1.60NL Released for new LCD display (SN 10736 through 13850).
Minor corrections.No Yes No
1.60cNL Released for Chinese version readers (SN 11201 through
13850).No Yes No
1.62NL Added TSE protocols (compatible with serial numbers 10736
through 13850)No Yes No
1.63NL Changed TSE validation criteria (compatible with serial
numbers 10736 through 13850)No Yes No
1.71NL Added functionality for ESC/P printer. Added protocols
for TSE Sheep/Goat and Rabies (compatible with serial numbers 10736
through 13850)No Yes No
Model 680XR 1.63XR For Model XR functionality only No Yes No
Backlit LCDSerial Numbers
13851 and higher
1.70 Released for backlit LCD and integration of English and
Chinese version firmware. Added functionality for ESC/P printer
(effec-tivity serial number 13851 and higher).No No Yes
1.71 Added protocols for TSE Sheep/Goat and Rabies (effectivity
serial number 13851 and higher).No No Yes
-
Adjustments, Calibration and Checks
7-10 100-2414E 02/06
7.6.3 Integration of English and Chinese Firmware Versions
Note: With firmware version 1.70, the English and Chinese
versions have been integrated. A language choice has been added
with this version.1. To enter the Select Language screen, press the 0 key and 4
key and turn ON instrument power. Hold for approximately 5 seconds
and the Select Language screen will appear.2. Use Up and Down arrow keys to select language choice. Press
Enter.3. Reader resets automatically and restarts with the selected
language.Note: If the language selected is different, the reader will
clear all memory and restore default values. -
Adjustments, Calibration and Checks
100-2414E 02/06 7-11
7.6.4 Downloading Firmware
1. Load the m680_up.exe program. The screen shown in Figure 7-12
will appear.Figure 7-12: Microplate Reader Model 680 Update Screen
2. Click on Model 680 Update and the following screen will
appear:Figure 7-13: Click Yes after Log-in
3. Connect a RS-232C cable between Model 680 and PC COM1
port.4. Turn ON Model 680 reader and input password for log in
(default 00000). Press enter.5. When complete, press Yes button on screen.
-
Adjustments, Calibration and Checks
7-12 100-2414E 02/06
6. The following screen will appear:
Figure 7-14: Confirmation Screen
7. Press Yes to update to new version. If the instrument is at
the latest version, the screen will indicate no update is
required.8. The screen shown in Figure 7-15 will indicate the status as
existing data and settings are retrieved from the Model 680
reader.Figure 7-15: Retrieving Data Screen
-
Adjustments, Calibration and Checks
100-2414E 02/06 7-13
9. When data retrieving is complete, the following screen will
appear:Figure 7-16: Firmware Update Screen
10. The following steps are also indicated on the screen shown
in Figure 7-1611. Turn OFF the Model 680 Reader.
12. While pushing both Main and 0 key on the keypad, turn on
reader power switch. After about 10 seconds, the following screen
on LCD appears:Firmware updateAre you sure?Yes: Start KeyNo: 7 Key
13. Push Start/Stop key switch on the Model 680 reader to begin
reader firmware erasing.14. After the erasing finishes, the LCD screen RS Receive will
appear.15. Click the Start button on PC screen. The firmware will begin
uploading and bar graph at the bottom of the screen will display
status. This will take approximately 7 minutes.16. When the firmware update operation is complete, the LCD on
the reader will displayed Finished and the restore previous data
screen will appear (Figure 7-17). -
Adjustments, Calibration and Checks
7-14 100-2414E 02/06
Figure 7-17: Restore Previous Data Screen
17. Turn OFF reader power switch then turn ON. A Battery Low
error message will occur. This is nor-mal; press Enter.18. At the Calendar settings screen, enter Date and Time and
press Enter.19. Enter password and press Enter.
20. After completion of log-in press the Yes button on the PC
screen.21. The restoring data screen will appear (Figure 7-18).
Figure 7-18: Storing Data Screen
-
Adjustments, Calibration and Checks
100-2414E 02/06 7-15
22. When the update is complete, the firmware complete message
will appear (Figure 7-19).Figure 7-19: Firmware Update Complete
23. Click OK to continue.
24. A password change reminder screen will appear (Figure
7-20).Figure 7-20: Change Password Reminder Screen
25. Follow on screen instructions to change password or select
Exit to quit. -
Adjustments, Calibration and Checks
7-16 100-2414E 02/06
7.7 QC Software
Note: There are 4 modes in this program; Measure Mode, Monitor
Mode, Extension Mode and Incubator. The instructions in this
section apply to the Monitor Mode which allows step-by-step
positioning of the plate carrier.7.7.1 Entering QC Software
1. While pressing the 0 and Upper key buttons, turn ON power
switch. Continue to hold down but-tons until QC Software screen
appears.2. Select YES by pressing the 1 key.
3. Input the password 00000 and press Enter.
4. Press the Dot key . to enter the QC Mode menu (Figure
7-13).Figure 7-21: QC Mode
7.7.2 Monitor Mode
1. Select Monitor Mode from the menu by pressing 2 (Figure
7-14).Figure 7-22: Monitor Mode
Definitions:MON = AD Value (voltage conversion)DEC = AD Value
(decimal)CH = Row PositionPLT = Plate Column PositionFL = Filter
NumberGain = Gain A, BCommands:Main = Back to QC Mode Screen
Change of Gain A
Change of Gain B
Change of Plate Position
Change of Filter Position
. (Dot) = Change of Channel
QC 1: Measure Mode2: Monitor Mode3: Extension Mode4:
IncubatorMON: VDEC:CH: A PLT: 12FL: 1 Gain=1.1.1.0
-
Adjustments, Calibration and Checks
100-2414E 02/06 7-17
7.7.3 Plate Positioning
1. To use the QC software to check for plate positioning, enter
the QC Mode and press 2 for Monitor Mode.2. The reader will initialize the plate carrier and place column
12 in the optics path.3. Press the down arrow key to change filter position for
desired filter position.4. Press the up arrow key to step the plate through column
positions 11 to 1.7.8 External Printer Setup
Note: The Model 680 Reader with firmware version 1.70 added
support for ESC/P external printer mode. Previous versions of
firmware support only ASCII print mode.1. To change the external printer print mode, perform the
following steps:a. Select Edit from the top screen menub. Select Printerc.
Select desired print mode: ASCII or ESC/Pd. Press Enter2. Connect printer and check print application by performing any
run. -
Troubleshooting
100-2414E 02/06 8-1
Section 8Troubleshooting
8.1 About this Section
This section of the manual provides information on
troubleshooting the Model 680 and 680 XR MicroplateReader. The
information in this section will help diagnose the common operating
difficulties and componentfailures. If additional assistance in
resolving a problem is required, please contact the Technical
ServiceDepartment at 1-800-424-6723.8.2 Troubleshooting
The following troubleshooting tables are intended as a guide to
assist in diagnosing some common problems.Table 8-1: General Problems
Problem Possible Cause Solution
Will not communicate with com-puter
Improper COM Port selected Check that COM1 port is used.
Improper serial cable Check for proper serial cable .
CPU board problem Replace CPU board.
Blank LCD upon Power Up Fuse Blown Check Fuse. If bad,
replaceCable connection Check display cable and reseat.
Bad LCD Replace LCD.
Memory Error Bad CPU Board Replace CPU Board.
Calendar Error Data error Reset time and date. Reboot.
Bad CPU Board Replace CPU Board.
Printer Error Printer malfunction Replace printer.
Will not print to external printer Check printer setup Section
7.8 (ESC/P requires firmware ver-sion 1.70 or later)Battery Error Battery dead Replace battery.
Old version firmware Download new firmware.
LCD Hardware Error LCD malfunction Replace LCD.
A/D Conversion Error Loose cable connection Check cables to CPU
board and Analog BoardBad A/D convertor Replace Analog Board.
Grinding Sound or Plate Error Carrier binding Check for source
of binding. Clean and lubricate plate carrier rods as
necessary. -
Troubleshooting
8-2 100-2414E 02/06
Table 8-2: Error Codes
Problem Possible Cause Solution
Code 8071- Invalid Command Received data error Resend
commandCode 8072 — Parameter Out of range
Received command parameter is out of range
Resend command and parame-ter
Code 8073 — Not in Remote Mode
Instrument is not in Remote Mode
Send the AQ command to acquire Remote Control Mode
Code 8074 — Device Busy Instrument is busy Wait until ready
Code 8075 — Filter Disk Jammed CPU board bad Replace CPU
boardMotor bad Replace motor
Photo interrupter for Filter disk damaged
Replace photo interrupter
Filter disk damage Replace the filter disk
Code 8076 — Carrier Jammed After Reading
Plate is not properly seated Reseat plate
Interference or blockage Check for Interference or block-age
Code 8077 — Light Bulb Error Lamp defective Replace lamp
Lamp out of alignment Check that lamp is properly seated
Optical fiber damaged Replace optical fiber
Bad analog board Replace Analog Board
Code 8082 — Filter Wheel Filter wheel jammed Check filter
wheelCode 8083 — Incubator Error Thermo sensor damaged Replace plate
carrierIncubation control board bad Replace Incubation control
boardCode 8084 — Carrier Home Sen-sor Closed Status Error
Photo interrupter for plate carrier damaged
Replace photo interrupter
CPU board bad Replace CPU board
Code 8085 — Carrier Home Sen-sor Open Status Error
Plate is not properly seated Reseat plate
Interference or blockage Check for Interference or block-age
Photo interrupter for plate carrier damaged
Replace photo interrupter
CPU board bad Replace CPU board
Motor bad Replace motor
-
Troubleshooting
100-2414E 02/06 8-3
Code 8086 — Carrier Jammed Before Reading
Plate is not properly seated Reseat plate
CPU board bad Replace CPU board
Motor bad Replace motor
Code 8087 — Plate Carrier Posi-tion
Plate carrier jammed after read-ing
Check plate carrier for obstruc-tion.
Code 8088 — Light Level Error Between Channel
Dirty Lenses Clean lenses
Optical fiber damaged Replace optical fiber
Bad analog board Replace Analog Board
Code 8089 — Filter Missing Filter missing from filter wheel
allowing light to exceed accept-able rangeCheck filter wheel and install filter
Code 8090 — Security Code in Error
Security code did not match Remove security lock
Code 8091 — RS-232C Error Data communication problem Check
RS-232 cableCode 8092 — A/D Conversion Error
Bad A/D convertor Replace Analog Board
Code 8093 — Stop Key Pressed Stop key pressed during
mea-surementUser error. Reset command
Table 8-2: Error Codes
Problem Possible Cause Solution
-
Troubleshooting
8-4 100-2414E 02/06
Table 8-3: Firmware Downloading
Problem Possible Cause Solution
Checksum Error Corrupted Firmware File Repeat using uncorrupted
fileCommunication Time Out RS-232 not connected Ensure RS-232 is
properly con-nected.Wrong button clicked Repeat download procedure
RS-232C Error Parity, Overrun or Framing errors occurred
Ensure RS-232 is properly con-nected. Repeat download
proce-dureVerify Error Communicated data and written data is different
Repeat download procedure (make sure instrument power is
cycled)Flash Write Error Flash memory could not be writ-ten
Repeat download procedure (make sure instrument power is
cycled)Flash Erase Error Flash memory could not be erased
Repeat download procedure (cycle power; first screen to appear
is downloading screen)Erase Check Error No erasing memory exists Repeat download
procedure (make sure instrument power is cycled)Battery Error Occurs after firmware download Normal. Press
Enter. Reset date, time and filter data. -
Troubleshooting
100-2414E 02/06 8-5
Table 8-4: Checkmark Plate Failures
Problem Possible Cause Solution
Fails Accuracy, Linearity or Precision
Erratic or non-uniform well-to-well OD values
Check for cleanliness of optics and lenses. Clean as
required.Check lamp. Reseat or replace as necessary.
Check filter. Reseat or replace as necessary.
Replace fiber optics assembly
Replace analog board
Replace CPU board
Fails Plate Positioning Plate binding Check for proper movement
of plate.Belt tension too tight or too loose Check belt tension and
tighten/loosen as necessary per Section 7.4Plate carrier out of alignment Check plate alignment and
cor-rect as necessary per Section 7.5Failed Cross Talk Wells A2, C2, E2, G2, B3, D3, F3, and H3 are
exceeding 0.003 OD due to some light path blockage.Check for cleanliness of optics and lenses. Clean as
required.Check lamp. Reseat or replace as necessary.
Check filter. Reseat or replace as necessary.
Check for proper movement of plate.
Check belt tension and tighten/loosen if necessary per Section
7.4Check plate alignment and cor-rect as necessary per Section
7.5Replace fiber optic assembly
Fails Filter Check Filter deterioration Replace filter
Fails 0 OD/Overflow Not detecting infinite absorbance Check
plate alignment and cor-rect as necessary per Section 7.5Replace analog board
Fails IR Blocking Detected IR light passing through plate; not
blocked by filtersCheck filter for damage or deteri-ation. Replace as
necessary. -
Parts Information
100-2414E 02/06 9-1
Section 9Parts Information
9.1 About this Section
This section of the manual provides information on the field
replaceable parts and service tools for the Model680 and 680 XR
Microplate Readers. Information on parts and tools ordering is also
included in this section.9.2 Parts Orders
9.2.1 US Domestic Operations
Parts may be ordered through our Service Parts department. You
can reach the Service Parts department during Pacific time regular
office hours by calling 1-510-741-5038 or 5058. You can help to
make processing your order more efficient by having the following
information before you call:The part number(s) or catalog number(s) of the parts you want to
orderShipment information
Your account or inventory location number
9.2.2 International Subsidiaries
The procedures for ordering parts vary from region to region
outside the United States. Please contact your Bio-Rad Export Sales
Representative through your regular sales and service channels.9.3 Returning Defective Parts
9.3.1 US Domestic Operations
Defective Assemblies returned for evaluation and repair are an
important part of the logistics and inventory control process.
Returned parts and assemblies are refurbished when possible, tested
and returned to ser-vice stock. Maintaining the effectiveness and
efficiency of this process enables us to deliver the high standard
of service that our customers expect, at a reasonable cost that
allows us to remain competitive.You are an integral part of our inventory process. When you
return parts or assemblies to Bio-Rad to be eval-uated and
repaired, please attach a tag to each assembly. The tag should
contain the following information:Model and serial number of the unit which the part was
removedThe part number or catalog number of the part being returned
The date of removal
A description of the problem or symptoms exhibited
Your name and other identification information
This information will help our service centers to make repairs
efficiently. When you receive replacement parts,sa
Назначение
Описание
Программное обеспечение
Технические характеристики
Знак утверждения типа
Комплектность
Поверка
Сведения о методах измерений
Нормативные документы
Назначение
Фотометры для микропланшетов Mark исполнение iMark (далее по тексту — фотометры) предназначены для измерений оптической плотности в лунках стандартного 96-луночного планшета при проведении различного рода исследований (иммуноферментный анализ, биохимические исследования).
Описание
Принцип действия фотометра основан на измерении оптической плотности анализируемых проб в 96-луночных планшетах. Для коррекции нестабильности источника света используется опорный световой канал. Выделение необходимой длины волны в световом потоке производится при помощи узкополосных интерференционных фильтров.
Световой поток падает через верхний блок линз сверху, на лунку с измеряемым образцом. Ослабленный пробой световой поток регистрируется фотодетектором, расположенным под образцом. В момент измерения каретка с микропланшетом перемещается относительно светового потока.
Фотометр имеет встроенный интерфейс USB, что позволяет работать как автономно, так и под управлением внешнего компьютера.
Защита от несанкционированной модификации обеспечивается конструкцией фотометра.
Общий внешний вид фотометра и схема пломбировки показаны на рисунке 1.
Программное обеспечение
предназначено для управления фотометром, обеспечения функционирования интерфейса, обработки информации, полученной в процессе проведения измерений, хранения результатов измерений.
Идентификационные данные (признаки) метрологически значимой части программного обеспечения фотометров указаны в таблице 1.
Идентификационные данные (признаки) |
Значение |
1 |
2 |
Идентификационное наименование ПО |
Microplate Manager |
Номер версии (идентификационный номер) ПО |
6.x* и выше |
Цифровой идентификатор ПО |
Данные являются собственностью производителя и являются защищёнными для доступа дилера и пользователей |
*где 6 — версия метрологически значимой части Г |
О; х — версия сборки ПО |
Защита программного обеспечения от непреднамеренных и преднамеренных изменений соответствует уровню «средний» в соответствии с Р 50.2.077 — 2014.
Технические характеристики
приведены в таблице 2.
Таблица 2
Диапазон измерений оптической плотности, Б |
от 0,030 до 3,000 |
Пределы абсолютной погрешности измерений оптической плотности, Б — в диапазоне измерений от 0,030 до 2,000 Б — в диапазоне измерений от 2,001 до 3,000 Б |
±0,06 ±0,1 |
Напряжение питания, В При частоте, Гц |
от 100 до 240 50/ 60 |
Потребляемая мощность, В-А, не более |
100 |
Габаритные размеры (ШхДхВ), мм, не более |
346x377x164 |
Масса, кг, не более |
5,5 |
Условия эксплуатации: — температура окружающей среды, °С — относительная влажность воздуха, %, не более |
от +15 до +40 90 |
Знак утверждения типа
наносится на титульный лист руководства по эксплуатации типографским способом и на заднюю панель фотометра методом наклеивания.
Комплектность
указана в таблице 3.
Таблица 3
Наименование |
Количество |
Фотометр с установленными фильтрами 415, 450, 490, 595, 655, 750 нм |
1 шт. |
Сетевой кабель |
1 шт. |
Кабель USB |
1 шт. |
Рулон термобумаги |
1 шт. |
Г арантийный талон |
1 экз. |
Руководство по эксплуатации (CD) |
1 шт. |
Поверка
осуществляется по документу Р 50.2.066-2009 «ГСИ. Анализаторы иммуноферментные. Методика поверки».
Основное средство поверки — комплект светофильтров поверочный КСП-02 (регистрационный номер 38817-08).
Допускается применение аналогичных средств поверки, обеспечивающих определение метрологических характеристик поверяемых СИ с требуемой точностью.
Знак поверки наносится на свидетельство о поверке.
Сведения о методах измерений
приведены в эксплуатационном документе.
Нормативные документы
Техническая документация фирмы «Bio-Rad Laboratories, Inc.», США.