BeneHeart D3

Defibrillator/Monitor

Operator’s Manual

loading

Related Manuals for Mindray BeneHeart D3

Summary of Contents for Mindray BeneHeart D3

  • Page 1
    BeneHeart D3 Defibrillator/Monitor Operator’s Manual…
  • Page 3
    © 2010 — 2013 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. For this Operator’s Manual, the issue date is 2013-09.
  • Page 4
    Mindray is strictly forbidden. are the trademarks, registered or otherwise, of Mindray in China and other countries. All other trademarks that appear in this manual are used only for informational or editorial purposes. They are the property of their respective owners.
  • Page 5
    Contents of this manual are subject to change without prior notice. All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this manual.
  • Page 6
    Mindray’s obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel.
  • Page 7
    Preface Manual Purpose This manual contains the instructions necessary to operate the product safely and in accordance with its function and intended use. Observance of this manual is a prerequisite for proper product performance and correct operation and ensures patient and operator safety. This manual is based on the maximum configuration and therefore some contents may not apply to your product.
  • Page 8
    FOR YOUR NOTES…
  • Page 9: Table Of Contents

    Contents 1 Safety……………………………… 1-1 1.1 Safety Information………………………………..1-1 1.1.1 Dangers ………………………………….1-2 1.1.2 Warnings …………………………………..1-2 1.1.3 Cautions ………………………………….1-3 1.1.4 Notes ………………………………….1-3 1.2 Equipment Symbols………………………………..1-4 2 The Basics …………………………….2-1 2.1 Overview……………………………………2-1 2.2 Intended Use ………………………………….2-1 2.2.1 AED……………………………………2-2 2.2.2 Manual Defibrillation…………………………….2-2 2.2.3 Noninvasive Pacing ……………………………….2-2 2.2.4 ECG……………………………………2-2 2.2.5 Resp ……………………………………2-2 2.2.6 SpO…

  • Page 10
    4 Managing Patients…………………………. 4-1 4.1 Overview ……………………………………4-1 4.2 Editing Patient Information …………………………….. 4-1 5 Alarms…………………………….. 5-1 5.1 Alarm Categories ………………………………… 5-1 5.2 Alarm Levels ………………………………….5-2 5.3 Alarm Indicators………………………………….. 5-2 5.3.1 Alarm Lamps ………………………………..5-2 5.3.2 Audible Alarms ………………………………. 5-3 5.3.3 Alarm Message ……………………………….
  • Page 11
    6.7.4 Changing Arrhythmia Threshold Settings ……………………..6-10 6.7.5 Initiating Arrhythmia Relearning Manually……………………..6-11 6.7.6 Automatic Arrhythmia Relearn…………………………6-11 6.8 Calibrating ECG…………………………………. 6-11 7 AED ………………………………7-1 7.1 Overview……………………………………7-1 7.2 Safety……………………………………..7-1 7.3 AED View……………………………………7-2 7.4 AED Procedure………………………………….7-3 7.5 Shock Advised………………………………….7-4 7.6 No Shock Advised (NSA)………………………………7-4 7.7 CPR ……………………………………..7-5 7.7.1 CPR Metronome ………………………………7-5 7.8 AED Sound Recording………………………………..7-6…
  • Page 12
    10.4.2 Changing Resp Wave Settings…………………………10-3 11 Monitoring PR …………………………..11-1 11.1 Overview ……………………………………11-1 11.2 Adjusting Pulse Tone Volume ……………………………..11-1 12 Monitoring SpO …………………………12-1 12.1 Introduction………………………………….12-1 12.2 Safety……………………………………12-2 12.3 Identifying SpO Modules…………………………….12-2 12.4 SpO Monitoring Procedure…………………………….12-2 12.5 Changing SpO Settings ……………………………….12-3 12.5.1 Setting SpO Sensitivity …………………………..12-3 12.5.2 Monitoring SpO…
  • Page 13
    16 Review …………………………….16-1 16.1 Reviewing Events ………………………………..16-1 16.2 Reviewing Tabular Trends…………………………….16-2 17 Data Management…………………………17-1 17.1 Introduction ………………………………….17-1 17.2 Reviewing Patient Events …………………………….17-2 17.3 Exporting Data ………………………………..17-2 18 Recording …………………………… 18-1 18.1 Using a Recorder………………………………..18-1 18.2 Recording Types………………………………..
  • Page 14
    20.2 Installing the Batteries ………………………………20-2 20.3 Battery Alarms………………………………….20-2 20.3.1 No Battery Alarm……………………………….20-2 20.3.2 Low Battery Alarm …………………………….20-2 20.3.3 Battery Aged Alarm…………………………….20-3 20.3.4 Battery Error Alarm…………………………….20-3 20.4 Checking the Batteries………………………………20-3 20.5 Charging batteries………………………………..20-3 20.6 Storing Batteries………………………………..20-4 20.7 Recycling the Batteries………………………………20-4 21 Care and Cleaning…………………………21-1 21.1 General Points ………………………………….21-1 21.2 Cleaning …………………………………….21-2 21.3 Disinfecting ………………………………….21-2…
  • Page 15
    A.7 Alarm Specifications………………………………..A-10 A.8 Data Management Specifications …………………………..A-11 A.9 Environmental Specifications …………………………….A-11 B EMC………………………………B-1 C BeneHeart Defibrillator Shift Checklist ……………………C-1 D Alarm Messages …………………………..D-1 D.1 Physiological Alarm Messages …………………………….D-1 D.2 Technical Alarm Messages ………………………………D-2 E Electrical Safety Inspection……………………….E-1 E.1 Power Cord Plug…………………………………..E-1 E.2 Device Enclosure and Accessories…………………………..E-1 E.3 Device Labelling…………………………………..E-2…
  • Page 16
    FOR YOUR NOTES…
  • Page 17: Safety

    Safety 1.1 Safety Information DANGER Indicates an imminent hazard that, if not avoided, will result in death or serious injury. WARNING Indicates a potential hazard or unsafe practice that, if not avoided, could result in death or serious injury. CAUTION Indicates a potential hazard or unsafe practice that, if not avoided, could result in minor personal injury or product/property damage.

  • Page 18: Dangers

    1.1.1 Dangers DANGER The equipment delivers up to 360 J of electrical energy. Unless properly used as described in these Operating Instructions, this electrical energy may cause serious injury or death. Do not attempt to operate this defibrillator unless thoroughly familiar with these operating instructions and the function of all controls, indicators, connectors, and accessories.

  • Page 19: Cautions

    For the treatment of patients with implantable pacemakers, place therapy pads or paddles away from internal pacemaker generator if possible to help prevent damage to the pacemaker. To avoid inadvertent disconnection, route all cables in a way to prevent a stumbling hazard. Wrap and secure excess cabling to reduce risk of entanglement or strangulation by patients or personnel.

  • Page 20: Equipment Symbols

    1.2 Equipment Symbols Caution (Attention, consult Status indicator accompanying documents) Alternating current Battery indicator Audio paused Alarm off Audio off Alarm paused Marker Lead select Gain select Event summary NIBP start/stop key Graphical recorder Menu Unlocking Network connector Shock button USB connector Input/Output Equipotentiality…

  • Page 21
    Fragile Right side up Keep dry Maximum stacks Manufacturer Date of manufacture General symbol for recovery/recyclable Electrostatic sensitive devices Mark of conformity to European Medical Device Directive 93/42/EEC Authorised representative in the European community DEFIBRILLATION-PROOF TYPE CF APPLIED PART DEFIBRILLATION-PROOF TYPE BF APPLIED PART Dispose of in accordance to your country’s requirements…
  • Page 22
    FOR YOUR NOTES…
  • Page 23: The Basics

    The Basics 2.1 Overview The BeneHeart (hereinafter called the equipment) is a lightweight and portable defibrillator/monitor. It provides four operating modes: Monitor, Manual Defib, AED and Pacer. In Monitor Mode, the equipment is intended for monitoring, displaying, reviewing, storing and printing multiple physiological parameters and waveforms including ECG, pulse oximetry (SpO ), respiration (Resp), and non-invasive blood pressure (NIBP).

  • Page 24: Aed

    2.2.1 AED The AED mode is to be used only on cardio arrest patients who are at least 8 years. The patients must be: Unresponsive Not breathing or not breathing normally 2.2.2 Manual Defibrillation Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive.

  • Page 25: Main Unit

    2.3 Main Unit 2.3.1 Front View External paddle Handle Alarm lamp Area 1 Area 3 Microphone Area 2 Speaker…

  • Page 26
    Area 1 Display screen AC power indicator Illuminated: when AC mains is connected. Off: when AC mains is not connected. Battery indicator Yellow: when the battery is being charged. Green: when the battery is fully charged or the equipment is run on battery. Off: when no battery is installed or battery fails.
  • Page 27
    Area 2 Lead Select button Press this button to select the lead of the first ECG waveform. Gain select button Press this button to select the size of the first ECG waveform. NIBP button (for equipments configured with NIBP function) Press this button to start or stop NIBP measurements.
  • Page 28
    Area 3 Mode Select knob Rotate this knob to select the operating mode or turn the equipment off. Energy Select button In Manual Defib mode, press this button to select energy level. Charge button Press this button to charge the defibrillator. Shock button Press this button to deliver a shock to the patient.
  • Page 29
    Recorder Start/Stop Key Press this key to start a recording or stop the current recording. Indicator Illuminated: when the recorder works correctly. Flashes: when an error occurred to the recorder, or the recorder runs out of paper. Paper outlet Recorder door Latch…
  • Page 30: Side View

    2.3.2 Side View Therapy port Therapy port is used to connect paddles cable or pads cable. Recorder ECG: ECG cable connector : SpO sensor connector NIBP: NIBP cuff connector…

  • Page 31: Rear View

    2.3.3 Rear View Hook Battery Equipotential grounding terminal When the defibrillator/monitor and other devices are to be used together, their equipotential grounding terminals should be connected together to eliminate the potential difference between them. External power input It connects an AC power cord or a DC/AC adapter to run the equipment respectively on the external AC mains or DC power supply.

  • Page 32: External Paddles

    2.3.4 External Paddles Sternum paddle Apex paddle Shock button Energy Select button Shock indicator Charge button Shock button 2-10…

  • Page 33: Display Views

    2.4 Display Views A typical screen in Manual Defib Mode is shown below. Patient Information area This area shows patient name, patient category, paced status, and current date and time. : indicates that the patient has an implanted pacemaker. Alarm status symbols indicates alarms are paused;…

  • Page 34
    Waveform area This area shows measurement waveforms. The waveform label is displayed at the upper left corner of the waveform. Parameter area This area shows measurement parameters. Each measurement module has a parameter block and the parameter name is displayed at the upper left corner. Manual Defib information area This area shows the selected defibrillation energy, shock counter as well as prompt related to manual defibrillation.
  • Page 35: Basic Operations And Settings

    Basic Operations and Settings 3.1 Installation WARNING The equipment shall be installed by personnel authorized by the manufacturer. The software copyright of the equipment is solely owned by the the manufacturer. No organization or individual shall resort to juggling, copying, or exchanging it or to any other infringement on it in any form or by any means without due permission.

  • Page 36: Environmental Requirements

    3.1.2 Environmental Requirements The operating environment of the equipment must meet the requirements specified in this manual. The environment where the equipment is used shall be reasonably free from noises, vibration, dust, corrosive, flammable and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind shall be left for convenient operation, maintenance and repair.

  • Page 37: Disconnecting From Power

    3.2.3 Disconnecting from Power To disconnect the equipment from the AC power source, follow this procedure: Confirm that the patient monitoring or therapy is completed. Disconnect the patient cables and sensors from the patient. Make sure to save or clear the patient data as required. Turn the Mode Select Knob to Off.

  • Page 38: Adjusting The Screen Brightness

    You can also set system time by selecting [Configuration >>]→[View Config]→[General Setup >>]. However, you cannot select date format and time format in this case. After the completion of setting system time, exit the configuration mode, and then the system will restart. 3.4.2 Adjusting the Screen Brightness Press the Main Menu button on the front panel, and then select [Others >>].

  • Page 39: Managing Patients

    Managing Patients 4.1 Overview Patient information management function enables you to edit and manage information of the current patient. 4.2 Editing Patient Information You can edit patient information in Monitor, Manual Defib and Pacer mode. To edit patient information, Press the Main Menu button on the front panel, and then select [Patient Demographics >>] and then make changes as desired.

  • Page 40
    FOR YOUR NOTES…
  • Page 41: Alarms

    Alarms Alarms, triggered by a vital sign that appears abnormal or by technical problems of the equipment, are indicated to the user by visual and audible alarm indications. WARNING A potential hazard exists if different alarm presets are used for the same or similar device in any single area, e.g.

  • Page 42: Alarm Levels

    5.2 Alarm Levels By severity, alarms can be classified into three categories: high level alarms, medium level alarms and low level alarms. Physiological alarms Technical alarms High level Indicate that your patient is in a life Indicate a severe device malfunction or an improper operation, threatening situation, such as Asystole, which could make it possible that the equipment cannot detect Vfib/Vtac and so forth, and an…

  • Page 43: Audible Alarms

    5.3.2 Audible Alarms The equipment uses different alarm tone patterns to match the alarm level: High level alarms triple+double+triple+double beeps. Medium level alarms triple beeps. Low level alarms single beep. 5.3.3 Alarm Message When an alarm occurs, the alarm message will appear in the technical or physiological alarm area. For physiological alarms, the asterisk symbols (*) before the alarm message match the alarm level as follows: High level alarms Medium level alarms…

  • Page 44: Alarm Tone Configuration

    5.4 Alarm Tone Configuration 5.4.1 Changing the Alarm Volume → Alm Volume >>]. Press the Main Menu button on the front panel, and then select [Alarm Setup >>] [ Set [Alm Volume] to an appropriate level: If [Audio Off] is enabled, alarm volume can be set to a value between 0 and 10, in which 0 means audio off and 10 the maximum volume level.

  • Page 45: Setting Alarm Properties For All Parameters

    5.5.1 Setting Alarm Properties for All Parameters → Para. Alarm >>] to enter the Para. Alarm setup menu, where you can In the main menu, select [Alarm Setup >>] [ review and set alarm limits, alarm switches, alarm level and alarm recordings for all parameters. When a parameter alarm is switched on, the equipment gives alarm indications in accordance with the preset alarm level and stores related waveforms and parameter values.

  • Page 46: Adjusting Alarm Limits Automatically

    5.5.2 Adjusting Alarm Limits Automatically The defibrillator/monitor can automatically adjust the patient’s alarm limits according to the measured vital signs. When [Auto Limits] is selected, the equipment automatically calculates alarm limits based on the latest measured parameter values. To enable auto alarm limits, press the Main Menu button on the equipment’s front panel, and then select [Alarm Setup →…

  • Page 47: Pausing Alarms

    5.6 Pausing Alarms You can temporarily disable alarm indicators by pressing the hardkey on the equipment’s front. When alarms are paused: For physiological alarms, no alarm indication is shown. New physiological alarm will not be presented. The remaining alarm pause time is displayed in the physiological alarm area. For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain presented.

  • Page 48: Pausing Alarm Sounds

    5.8 Pausing Alarm Sounds You can press the [Audio Pause] softkey to pause alarm tones. In this case, the symbol will be displayed in the sound symbol area indicating all system sounds are silenced temporarily. In the audio paused status, all alarm indicators except audible alarm tones works properly.

  • Page 49: Latching Alarms

    5.11 Latching Alarms The latching setting for physiological alarms defines how alarm indicators behave when you do not acknowledge them. If an alarm is latched, alarm indications remain presented even though alarm conditions end, except that: The parameter reading and violated alarm limit stop flashing. The time when the alarm is last triggered is displayed behind the alarm message.

  • Page 50: When An Alarm Occurs

    5.13 When an Alarm Occurs When an alarm occurs, observe the following steps and take proper actions: Check the patient’s condition. Confirm the alarming parameter or alarm category. Identify the alarm source. Take proper action to eliminate the alarm condition. Make sure the alarm condition is corrected.

  • Page 51: Monitoring Ecg

    Monitoring ECG 6.1 Overview The electrocardiogram (ECG) measures the electrical activity of the heart and displays it as waveforms and numerics. The equipment enables ECG monitoring through 3-, 5- lead ECG sets, external paddles and multifunction electrode pads. If both ECG sets and paddles/pads are connected, the configured ECG waveforms are displayed in the waveform area. 6.2 Safety WARNING Periodically inspect the electrode application site to ensure skin quality.

  • Page 52: Monitoring View

    6.3 Monitoring View You can access Monitor mode by switching the Mode Select knob to the Monitor position. When operating in Monitor mode, the equipment displays up to two ECG waveforms, the heart rate reading, other available parameter values and active alarm settings.

  • Page 53
    5-Lead Placement The following is a typical AHA electrode placement for a 5-lead ECG set: RA placement: directly below the clavicle and near the right shoulder. LA placement: directly below the clavicle and near the left shoulder. RL placement: on the right lower abdomen. LL placement: on the left lower abdomen.
  • Page 54: Ecg Monitoring With Paddles/Pads

    6.4.2 ECG Monitoring with Paddles/Pads Prepare the patient’s skin. Apply the paddles/pads to the patient. If multifunction electrode pads are used, apply pads according to the instructions for use indicated on pads package. Use anterior-lateral placement. If external paddles are used, remove the paddle set from the paddle tray by grasping the handles and pulling them straight up.

  • Page 55: Checking Paced Status

    6.4.3 Checking Paced Status It is important to set the paced status correctly when you start monitoring ECG. The paced symbol is displayed in the patient information area when the [Paced] is set to [Yes]. The pace pulse markers “︱” are shown on the ECG wave when the patient has a paced signal.

  • Page 56: Changing Ecg Settings

    6.6 Changing ECG Settings 6.6.1 Changing Lead Settings 6.6.1.1 Selecting Lead Type Select the ECG parameter area to enter the [ECG Setup] menu. Select [Lead Set] and toggle between [3-lead] and [5-lead]. You can also set lead type in the configuration mode: →…

  • Page 57: Changing Ecg Wave Settings

    6.6.2 Changing ECG Wave Settings You can select the ECG parameter area to enter the [ECG Setup] menu to set ECG cascade and wave speed. You can also select the hot keys above the ECG waveform to change ECG lead, size and filter. You can press the Lead Select button on the equipment’s front panel or use the Navigation knob to select the lead hot key above the first ECG waveform to select a lead.

  • Page 58: Adjusting Heartbeat Volume

    6.6.4 Adjusting Heartbeat Volume In the case that ECG alarm is switched on, or both ECG alarm and PR alarm are switched off, heartbeat tone is issued. To adjust the heartbeat volume, → QRS Volume], or Select the ECG parameter window to enter the [ECG Setup] menu, and then select [Others >>] [ →…

  • Page 59: Switching Arrhythmia Analysis On And Off

    Arrhythmia event Description Category Bigeminy A dominant rhythm of N, V,N, V, N, V. Trigeminy A dominant rhythm of N, N, V,N, N, V, N, N, V. R ON T R on T detected in normal heartbeats. Missed Beats* No beat detected for 1.75x average R-R interval for HR <120, or No beat for 1 second with HR >120 (for non-paced patients only), or No beat detected for more than the set pause threshold.

  • Page 60: Changing Arrhythmia Alarm Settings

    6.7.3 Changing Arrhythmia Alarm Settings To change arrhythmia alarm settings, select the ECG parameter area to enter the [ECG Setup] menu, and then select [Arrhythmia >>]→[Arrh. Alarm] menu, where you can set alarm switch, alarm level and alarm record switch for all the arrhythmia events..

  • Page 61: Initiating Arrhythmia Relearning Manually

    6.7.5 Initiating Arrhythmia Relearning Manually Normally arrhythmia relearning allows the equipment to learn new ECG patterns to correct arrhythmia alarms and heart rate value. We suggest you to manually initiate arrhythmia relearning when you suspect the result of arrhythmia analysis. To initiate relearning manually, select the ECG parameter window to enter the [ECG Setup] menu, select [Arrhythmia →…

  • Page 62
    FOR YOUR NOTES 6-12…
  • Page 63: Aed

    7.1 Overview This chapter describes how to operate the equipment in AED Mode. While operating in AED Mode, the equipment analyses the patient’s ECG waveforms and guides you through the defibrillation process. The equipment starts analyzing the patient’s heart rhythm immediately after entering AED mode. When a shockable rhythm is detected, the equipment gives a prompt and automatically starts charging.

  • Page 64: Aed View

    WARNING During defibrillation, air pockets between the skin and multifunction electrode pads can cause patient skin burns. To help prevent air pockets, make sure defibrillation pads are completely adhered to the skin. Do not use dried-out pads. In AED mode, this equipment is not designed to administer energy at pediatric joule settings. The American Heart Association recommends AEDs be used only on patients who are more than eight years old.

  • Page 65: Aed Procedure

    7.4 AED Procedure Confirm that the patient is unresponsive, not breathing or not breathing normally. Then: Remove clothing from the patient’s chest. Wipe moisture from the patient’s chest and, if necessary, clip or shave excessive chest hair. Apply multifunction electrode pads to the patient as directed on the pads package. Use anterior-lateral placement. Connect the pads with pads cable, and then plug the pads cable in the equipment’s therapy port.

  • Page 66: Shock Advised

    overcome to deliver an effective discharge of energy. The degree of impedance differs from patient to patient and is affected by several factors including the presence of chest hair, moisture, and lotions or powders on the skin. If the “Impedance too high. Shock not delivered” message appears, make sure that the patient’s skin has been washed and dried and that any chest hair has been clipped.

  • Page 67: Cpr

    7.7 CPR If [Initial CPR Time] is not configured as Off, the system enters initial CPR if AED mode is entered. You can set [Initial CPR Time] to an appropriate time or switch it off through configuration management. After the shock series, ECG analysis pauses and the equipment enters the CPR status. Analysis resumes at the completion of the pause period or when you press the [Resume Analyzing] soft key in CPR status.

  • Page 68: Aed Sound Recording

    Warning The CPR metronome sounds do not indicate information regarding the patient’s condition. Because patient status can change in a short time, the patient should be assessed at all times. Do not perform CPR on a patient who is responsive or is breathing normally. NOTE CPR metronome and its volume is affected by the settings of [Voice Prompt] and [Voice Volume] in the AED Setup menu.

  • Page 69: Manual Defibrillation

    Manual Defibrillation 8.1 Overview This chapter explains how to prepare for and perform asynchronous defibrillation and synchronous cardioversion using multifunction electrode pads and external paddles. In Manual Defib Mode, you must assess the ECG waveforms, decide if defibrillation or cardioversion is indicated, select appropriate energy setting, charge the equipment, and deliver the shock.

  • Page 70
    WARNING During synchronous cardioversion, if monitoring patient’s ECG through external paddles, artifact introduced by paddle movement may resemble an R-wave and trigger a defibrillation shock. Do not use conductive liquid. Use only conductive gel specified by the equipment manufacturer. If external paddles are used for defibrillation, apply the paddles tightly and evenly to the patient’s chest to ensure good skin contact.
  • Page 71: Manual Defibrillation View

    8.3 Manual Defibrillation View A typical screen in Manual Defib Mode is shown below. In the enlarged ECG area, an ECG waveform and related parameters are displayed. In the middle of the screen, defibrillation mode, synchronous icon, prompt message, selected energy, contact impedance indicator, and a shock counter are displayed.

  • Page 72
    WARNING Hold only the insulating parts of the paddle handles to avoid shock hazard during charging or shock delivery. Turn the Mode Select knob to Manual Defib. You can access manual therapy directly, by confirmation or by password, which can be defined through configuration management.
  • Page 73: Using Pediatric Paddles

    NOTE Defibrillation is always performed through paddles or pads. However, during defibrillation you may choose to monitor the ECG using an alternate ECG source (3- or 5-lead monitoring electrodes). If an alternate ECG source is connected, any available lead may be displayed. When external paddles are used, the Shock button on the equipment’s front panel is disabled.

  • Page 74: Synchronized Cardioversion

    NOTE When internal paddles are used for defibrillation, the energy selection is automatically limited to 50 joules because of possible cardiac damage from higher energies. Sterilize the internal paddles before each use. Otherwise, severe infection may result. Clean the internal paddles after each use. 8.5 Synchronized Cardioversion Synchronized Cardioversion allows you to synchronize delivery of the defibrillator shock with the R-wave of the ECG.

  • Page 75: Performing Synchronized Cardioversion

    8.5.1 Performing Synchronized Cardioversion Connect the therapy cable and apply the multifunction electrode pads or external paddles to the patient. If ECG set is used for ECG monitoring, connect the ECG trunk cable and apply the ECG electrodes to the patient, referring to 6 Monitoring ECG.

  • Page 76: Remote Synchronized Cardioversion

    8.6 Remote Synchronized Cardioversion The equipment can be configured to receive an ECG source from a remote patient monitor (such as a bedside patient monitor) to perform synchronized cardioversion. To do so, the remote patient monitor shall have a sync out connector and shall be connected to the defibrillator/monitor’s multifunctional connector with a synchronous cable.

  • Page 77: Contact Impedance Indicator

    NOTE During remote synchronous defibrillation, the local defibrillator/monitor does not display the ECG waveform. To view the patient’s ECG, check the remote monitor. When you use an remote monitor as the ECG source, a biomedical technician must verify that the remote monitor and the defibrillator/monitor combination will deliver a synchronized shock within 60 ms of the peak of the R-wave.

  • Page 78
    FOR YOUR NOTES 8-10…
  • Page 79: Noninvasive Pacing

    Noninvasive Pacing 9.1 Overview In pacer mode, the patient’s ECG is monitored through ECG lead set and pace pulses are delivered through multifunction electrode pads. The pads cannot be used to monitoring ECG rhythm and deliver pacing current at the same time. A white pacing marker is shown on the ECG waveform each time a pacer pulse is delivered to the patient.

  • Page 80: Pacing View

    CAUTION Use of Pacer mode may be password protected. Make sure the operator knows and remembers the password as defined in Configuration. Failure to enter correct password will prevent the delivery of pacing therapy. For treatment of patients with implanted devices such as permanent pacemakers or cardioverter-defibrillators, consult a physician and the instructions for use provided by the device’s manufacturer Prolonged noninvasive pacing may cause patient skin irritation and burns.

  • Page 81: Demand Mode Versus Fixed Mode

    9.4 Demand Mode versus Fixed Mode The equipment can deliver paced pulses in either demand or fixed mode. In demand mode, the pacer only delivers paced pulses when the patient’s heart rate is lower than the selected pacing rate. In fixed mode, the pacer delivers paced pulses at the selected rate. During pacing, you can change pacer mode.

  • Page 82: Demand Mode Pacing

    9.5.1 Demand Mode Pacing To pace in demand mode: Turn the Mode Select knob to the Pacer position. Thus the pacing function is enabled in demand mode automatically. ECG waveform of Lead II is displayed in the waveform area by default. You can access manual therapy directly, by confirmation or by password, which can be defined through configuration management.

  • Page 83: Fixed Mode Pacing

    Verify the presence of a peripheral pulse. You can temporarily withhold pacing pulse and observe the patient’s underlying rhythm by pressing and holding the [4:1] soft key. This causes pacing pulse to be delivered at 1/4 of the defined pacer rate. To resume pacing at set rate, release this key.

  • Page 84
    WARNING Use care when handling the multifunction electrode pads on the patient to avoid shock hazard during pacing. If you are using the pacing function with battery power and the Low Battery alarm is presented, connect the equipment to external power or install a fully charged battery. CAUTION The monitoring or pacing function may be unstable in the presence of ESU or other electronic devices.
  • Page 85: Monitoring Resp

    Monitoring Resp 10.1 Overview Impedance respiration is measured across the thorax. When the patient is breathing or ventilated, the volume of air changes in the lungs, resulting in impedance changes between the electrodes. Respiration rate (RR) is calculated from these impedance changes, and a respiration waveform appears on the equipment screen. 10.2 Safety WARNING When monitoring the patient’s respiration, do not use ESU-proof ECG cables.

  • Page 86: Placing Resp Electrodes

    10.4 Placing Resp Electrodes As the skin is a poor conductor of electricity, preparing the skin is necessary for a good Respiration signal. You can refer to the ECG section for how to prepare the skin. As the Respiration measurement adopts the standard ECG electrode placement, you can use different ECG cables (3-lead or 5-lead).

  • Page 87: Optimizing Lead Placement For Resp

    10.4.1 Optimizing Lead Placement for Resp If you want to measure Resp and you are already measuring ECG, you may need to optimize the placement of the two electrodes between which Resp will be measured. Repositioning ECG electrodes from standard positions results in changes in the ECG waveform and may influence ST and arrhythmia interpretation.

  • Page 88
    FOR YOUR NOTES 10-4…
  • Page 89: Monitoring Pr

    Monitoring PR 11.1 Overview The pulse numeric counts the arterial pulsations that result from the mechanical activity of the heart. You can display a pulse from SpO . The displayed pulse numeric is color-coded to match SpO parameter. PR unit PR alarm high limit PR value PR alarm low limit…

  • Page 90
    FOR YOUR NOTES 11-2…
  • Page 91: Monitoring Spo

    Monitoring SpO 12.1 Introduction monitoring is a non-invasive technique used to measure the amount of oxygenated haemoglobin and pulse rate by measuring the absorption of selected wavelengths of light. The light generated in the probe passes through the tissue and is converted into electrical signals by the photodetector in the probe. The SpO module processes the electrical signal and displays a waveform and digital values for SpO and pulse rate.

  • Page 92: Safety

    Change the application site every four hours. For neonates or patients with poor peripheral blood circulation or sensitive skin, inspect the sensor site more frequently. 12.3 Identifying SpO Modules The equipment can be configured with any of the following SpO modules. Mindray SpO module; Masimo SpO module; Nellcor SpO module.

  • Page 93: Changing Spo Settings

    12.5.1 Setting SpO Sensitivity For Mindray SpO module, you can set [Sensitivity] to [High], [Med] or [Low] from the [SpO2 Setup] menu. For Masimo module, you can set [Sensitivity] to [Normal] or [High], in which [Normal] is equivalent to [Med].

  • Page 94: Sat-Seconds Alarm Management

    12.5.4 Sat-Seconds Alarm Management With traditional alarm management, high and low alarm limits are set for monitoring oxygen saturation. During monitoring, as soon as an alarm limit is violated, an audible alarm immediately sounds. When the patient’s SpO value fluctuates near an alarm limit, the alarm sounds each time the limit is violated. Such frequent alarm can be distracting. The Sat-Seconds feature is available with the Nellcor SpO module to decrease the likelihood of false alarms caused by motion artifacts.

  • Page 95: Changing The Speed Of The Pleth Wave

    12.5.5 Changing the Speed of the Pleth Wave In the [SpO2 Setup] menu, select [Sweep] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is. 12.6 SpO Desat Alarm The defibrillator/monitor provides an SpO Desat alarm.

  • Page 96: Masimo Information

    Inappropriate positioning of the SpO2 sensor, or use of incorrect SpO2 Drop of arterial blood flow to immeaurable level caused by shock, anemia, low temperature or vasoconstrictor. 12.9 Masimo Information Masimo Patents This device may be covered by one or more of the following US patents and foreign equivalents: 5,758,644, 6,011,986, 6,699,194, 7,215,986, 7,254,433, 7,530,955.

  • Page 97: Nibp

    NIBP 13.1 Introduction Automatic non-invasive blood pressure monitoring uses the oscillometric method of measurement. It is intended for adult, pediatric and neonatal patients. To understand how this method works, we’ll compare it to the auscultative method. With auscultation, the clinician listens to the blood pressure and determines the systolic and diastolic pressures. The mean pressure can then be calculated with reference to these pressures as long as the arterial pressure curve is normal.

  • Page 98: Measurement Limitations

    Do not use the NIBP cuff on a limb with an intravenous infusion or arterial catheter in place. This could cause tissue damage around the catheter when the infusion is slowed or blocked during cuff inflation. If you doubt the NIBP readings, determines the patient’s vital signs by alternative means and then verify that the equipment is working correctly.

  • Page 99: Starting And Stopping Nibp Measurements

    Warning Continuous CUFF pressure due to connection tubing kinking may cause blood flow interference and resulting harmful injury to the patient. 13.5.2 Starting and Stopping NIBP Measurements You can start and stop NIBP measurements by using the hardkey on the equipment’s front panel. 13.5.3 Correcting the Measurement The cuffed limb should be at the same level as the patient’s heart.

  • Page 100: Understanding The Nibp Numerics

    13.6 Understanding the NIBP Numerics The NIBP display shows numerics only as below. Your display may be configured to look slightly different. Measurement mode Pressure unit: mmHg or kPa NIBP alarm high limit NIBP alarm low limit Time of last measurement Systolic pressure Diastolic pressure Mean pressure at the completion of measurement, or cuff pressure during the measurement…

  • Page 101: Marking Events

    Marking Events During patient monitoring or therapy, some events may exert effects on the patient and as a result change related waveforms and parameter values. To help analysing the waveforms or numerics at that time, you can mark these events. Before marking an event, you can define events A to H, for example, define event D as injecting Atropine.

  • Page 102
    FOR YOUR NOTES 14-2…
  • Page 103: Freezing Waveforms

    Freezing Waveforms During patient monitoring, the freeze feature allows you to freeze the currently displayed waveforms on the screen so that you can have a close examination of the patient’s status. Besides, you can select any frozen waveform for recording. Waveforms can be frozen only in the Monitor Mode.

  • Page 104: Unfreezing Waveforms

    15.3 Unfreezing Waveforms To unfreeze the frozen waveforms, you can either: Press the [Unfreeze] soft key, or Select [Exit] to return to the Main screen, or Perform any other action that causes the screen to be readjusted or opens a menu, such as plugging in or out a module, pressing the [Lead Select] or [Main Menu] button, etc.

  • Page 105: Review

    Review 16.1 Reviewing Events The equipment can automatically record and save patient events. You can review patient events following this procedure: To review events, you can: In the Monitor, Manual Defib or Pacer mode, press the Main Menu button on the front panel, and then select [Review>>]→[Review Events >>] to enter the [Review Events] menu, or In the Monitor mode, repeatedly press the [Trends] soft key to enter the [Review Events] menu.

  • Page 106: Reviewing Tabular Trends

    16.2 Reviewing Tabular Trends In the Monitor, Manual Defib o Pacer mode, press the Main Menu button on the front panel; select [Review>>]→[Trends>>] or, if you are operating in the Monitor mode, select the [Trends] soft key to enter the tabular trends menu, as shown below: 16-2…

  • Page 107: Data Management

    Data Management 17.1 Introduction The data management function enables you to: Edit patient information; Review patient events; and Export patient data to USB memory. To access data management, press the Main Menu button on the front panel to enter the Main Menu, and then select →…

  • Page 108: Reviewing Patient Events

    17.2 Reviewing Patient Events To view patient events, select a patient in the Archives Main screen, and then press the navigation knob to confirm the selection. In this case, you can select the [Return] soft key to return to the Archives Main screen. To edit patient information, select the [Patient Info] button and change the patient information as desired.

  • Page 109: Recording

    Recording 18.1 Using a Recorder The thermal recorder records patient information, measurement numerics and up to three waveforms. 18.2 Recording Types By the way recordings are triggered, they can be classified into the following categories: Manually-triggered realtime waveform recordings. Event-triggered recordings. Alarm recordings triggered by an alarm limit violation or an arrhythmia event.

  • Page 110: Setting The Recorder

    Automatic recordings will be triggered in the following conditions: If both [Alarm] and [Alm Rec] for a measurement are switched on, an alarm recording will be triggered automatically as an alarm occurs. When related event is triggered. To manually stop a recording, you can press the hardkey again.

  • Page 111: Changing The Recording Speed

    18.4.4 Changing the Recording Speed Enter the [Record Setup] menu. Select [Paper Speed] and toggle among [6.25 mm/s], [12.5 mm/s], [25 mm/s] and [50 mm/s]. This setting is for all recordings containing waveforms. 18.4.5 Switching Gridlines On or Off Enter the [Record Setup] menu. Select [Gridlines] and toggle between [On] and [Off].

  • Page 112: Removing Paper Jam

    18.6 Removing Paper Jam If the recorder works incorrectly or produces unusual sounds, check if there is a paper jam first. If a paper jam is detected, follow this procedure to remove it: Open the recorder door. Take out the paper and tear off the draped part. Reload the paper and close the recorder door.

  • Page 113: Configuration Management

    Configuration Management 19.1 Introduction Configurations management enables you to customize you equipment to best meet your needs. With this function, you can: Change system configuration; Record system configuration; Restore the factory default configuration. After the system configurations have been changed, the equipment restarts and new configuration settings take effect immediately.

  • Page 114: General Setup Menu

    Enter the required password and then select [OK] to enter the Configuration Main menu as shown below: Selecting [Factory Config] and confirming the selection restores all the current settings to factory default settings: Selecting [Record] records the settings of all system configurations. Selecting [Exit] pops up a dialog box as shown below: WARNING Never connect the equipment with the patient while performing configuration management.

  • Page 115: Manual Defib Setup Menu

    Menu Item Options/Range Default Remark Year 2007 to 2099 2007 Month 01 to 12 01 to 31 System 24 h: 00 to 23 24 h: 00 Time Hour 12 h: 12AM to 11PM 12 h: 12AM Minute 00 to 59 Second 00 to 59 19.3.2 Manual Defib Setup Menu…

  • Page 116: Pacer Setup Menu

    Menu Item Options/Range Default Remark Voice Prompts On, Off Voice Volume High, Med, Low High Voice Prompt Interval Off, 30s, 60s, 90s, 120s, 150s, 180s Voice Recording On, Off 19.3.4 Pacer Setup Menu Menu Item Options/Range Default Pacer Rate 40 to 170 ppm 70 ppm Pacer Output 0 to 200 mA…

  • Page 117
    Menu Item Options/Range Default Remark Arrhythmia On, Off ARR Alarm On, Off ARR Alm Lev PVCs/min High, Med, Low R ON T VT>2 Couplet Vent. Rhythm Bigeminy Trigeminy Tachy Brady Missed Beat Multif. PVCs Nonsus. Vtac Pause Irr. Rhythim Asystole Delay 3 to 10 V-Tach Rate 100 to 200…
  • Page 118: Resp Setup Menu

    Desat Limit to (High-1) Desat 50 to (High-1) Averaging Masimo SpO 2-4s, 4-6s, 8s, 10s, 12s, 14s, 16s For Masimo SpO module only. Mindray SpO High, Med, Low Different options are available to Sensitivity match the SpO module used. Masimo SpO Normal, Maximum…

  • Page 119: Nibp Setup Menu

    Menu Item Options/Range Default Remark PR Low 25 to (High-2) This setting is linked with the [QRS QRS Volume 0 to 10 Volume] setting in the [ECG Setup] menu 19.3.9 NIBP Setup Menu Menu Item Options/Range Default Remark Manual, 1 min, 2 min, 2.5 min, 3 min, 5 Interval min, 10 min, 15 min, 20 min, 30 min, 1 h, Manual…

  • Page 120: Alarm Setup Menu

    19.3.10 Alarm Setup Menu Menu Item Options/Range Default Alarm Pause Time 1, 2, 3, 5, 10, 15 min, Permanent 2 min Audio Off Enabled, Disabled Disabled 0 to 10 (If Audio Off is enabled), Alarm Volume 1 to 10 (If Audio Off is disabled) Reminder Tone On, Off Reminder Volume…

  • Page 121: Mark Event Setup Menu

    19.3.12 Mark Event Setup Menu Menu Item Options/Range Default Remark Event A Generic Generic Unchangeable Event B Adrenalin Adrenalin, Lidocaine, Atropine, Nitroglycerin, Event names that have been Event C Lidocaine Morphine, Intubation, IV Access, Adenosine, selected by previous events will Event D Atropine Amiodarone, Vasopressin, Isoprenaline,…

  • Page 122: User Test Setup Menu

    19.3.15 User Test Setup Menu Menu Item Options/Range Default User Test Prompt On, Off 12h time format: 12:00 AM ~ 05:00 AM 03:00 AM Auto Test Time 24h time format: 00:00 ~ 05:00 03:00 19.3.16 Others Menu Menu Item Options/Range Default Brightness 1 to 10…

  • Page 123: Battery

    Battery 20.1 Introduction The equipment is designed to operate on battery power when external power supply is not available. The battery is charged whenever the equipment is connected to AC mains or the DC power supply through an external DC/AC adapter, regardless of whether or not the equipment is currently turned on.

  • Page 124: Installing The Batteries

    20.2 Installing the Batteries To install the batteries: Align a battery with the battery compartment. Insert the battery, and press until you hear it click into the place. To replace a battery, press the latch on the battery and push the battery to the right until you remove it. Insert a new battery into the battery compartment.

  • Page 125: Battery Aged Alarm

    20.3.3 Battery Aged Alarm If the battery runtime is significantly shorter than the specification, a low level technological alarm “Battery Aged” will be presented. We recommend you to contact our company and replace it with a new one. 20.3.4 Battery Error Alarm In the situation that the battery has a failure, a high level technological alarm “Battery Err”…

  • Page 126: Storing Batteries

    20.6 Storing Batteries When storing batteries, make sure that the battery terminals do not come into contact with metallic objects. If batteries are stored for an extended period of time, they should be placed in a cool place with a partial charge of 40% to 60% capacity (3 LEDs illuminated).

  • Page 127: Care And Cleaning

    Care and Cleaning Use only the substances approved by the equipment manufacturer and methods listed in this chapter to clean or disinfect your equipment. Warranty does not cover damages caused by unapproved cleaning and disinfection substances or methods. We make no claims regarding the efficacy of the listed chemicals or methods as a means for controlling infection. For the method to control infection, consult your hospital’s infection control officer or epidemiologist.

  • Page 128: Cleaning

    21.2 Cleaning Your equipment should be cleaned on a regular basis. If there is heavy pollution or lots of dust and sand in your place, the equipment should be cleaned more frequently. Before cleaning the equipment, consult your hospital’s regulations for cleaning the equipment.

  • Page 129: Maintenance And Testing

    Maintenance and Testing WARNING Failure for the responsible individual, hospital or institution employing this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards. The safety checks or maintenance involving any disassembly of the equipment should be performed by professional servicing personnel.

  • Page 130: Maintenance And Testing Schedule

    22.2 Maintenance and Testing Schedule The following tests, except recorder check and user test, shall be carried out by the service personnel only. Contact your service personnel if any maintenance is required. Make sure to clean and disinfect the equipment before any test and maintenance.

  • Page 131: Shift Check

    Check the display of technical alarm area, prompt area and battery status indicator on the upper right corner of the main screen to judge whether the equipment runs normally. 22.3.2 Shift Check In order to ensure your defibrillator/monitor is ready when needed, we recommended you to inspect your equipment and complete a check list at the change of every shift.

  • Page 132: User Test

    22.3.4 User Test WARNING Do not perform user test when a patient is connected to the equipment. User test covers the following items: Routine Test, Energy delivery test, and Controls test. NOTE Before user test or after each use, thoroughly clean the paddles and properly place them in the paddle tray. User test passes only when paddles properly contact the metal parts of the paddle tray.

  • Page 133
    22.3.4.2 Routine Test Routine Test covers the following items: Batteries, Mainboard, Defib/Pacer function, and Monitor function If any of above items fails, the red status indicator will flash. If mainboard, Defib/Pacer function, or monitor function fails, a low level technical alarm “Last User Test Failed” will be displayed in the technical alarm area when the equipment is restarted.
  • Page 134: Recorder Inspection

    22.3.4.5 Test Summaries The results of User Test are automatically saved as summaries. You can select the [History] button from the User Test Main menu to review the test summaries. The equipment can store up to 300 historical test summaries which are listed in the sequence of time, with the latest on the top.

  • Page 135: Ecg Cable Test

    22.3.6 ECG Cable Test It is recommended to perform ECG cable test once a year. Test tool: ECG simulator Follow this procedure to perform ECG cable test: Turn the Mode Select knob to “Monitor”. Connect the ECG cable to the defibrillator and the electrodes to the simulator. Turn on the simulator and select a normal ECG rhythm.

  • Page 136
    Press the “Disarm” soft key to discharge the energy internally. Verify that a prompt “Charge Removed” appears on the screen and the charge done tone stops. Verify that the value measured by the analyzer is 0J or blank. 10. Enter the Configuration Main menu, select [Manual Therapy Setup] and set [Time to Auto Disarm] to [60s]. 11.
  • Page 137: Pacing Test

    22.3.8 Pacing Test Test tools: defibrillator/pacer analyzer Run the equipment on fully charged battery. Move the Mode Select knob to Pacer. Select Fixed mode.. Connect the pads cable to the equipment and properly place the pads on the defibrillator/pacer analyzer. Set the analyzer to Pacing Measurement mode.

  • Page 138
    22.3.9.3 NIBP Accuracy Test The NIBP accuracy test is required at least once every two years or whenever you doubt the NIBP reading. Tools required: T-shape connector Tubing Balloon pump Metal Vessel, volume 500±25 ml Calibrated manometer for reference, accuracy superior to 1 mmHg Follow this procedure to perform the accuracy test: Connect the equipment as shown below.
  • Page 139
    22.3.9.4 NIBP Leakage Test The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once every two years or whenever you doubt the NIBP reading. Tools required: An adult cuff An air tubing A correct sized cylinder Follow this procedure to perform the leakage test: Set the patient category to [Adu].
  • Page 140: Nibp Overpressure Protection Test

    22.3.9.6 Formatting Storage Card You can format the storage card if data in the card is useless, or if the card has a failure. To format the storage card, select → Format] through the Installation Mode Main menu. Then a dialog box pops up as shown below: [Format Data Card] If storage card is formatted successfully, a prompt “Formatting is completed!”…

  • Page 141: Accessories

    Accessories WARNING Use accessories specified in this chapter. Using other accessories may cause damage to the equipment or not meet the claimed specifications. Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the measurement accuracy.

  • Page 142
    Lead Sets 3-Electrode Lead Sets Type Compatible with Model Applicable patient Remark EL6302A 0010-30-42725 Adult, pediatric EL6304A 0010-30-42732 Long EL6306A Neonate 0010-30-42897 EL6308A Pediatric 0010-30-42899 Clip EL6301A 0010-30-42726 Adult, pediatric EL6303A 0010-30-42731 Long EL6305A Neonate 0010-30-42896 EL6307A Pediatric 0010-30-42898 EL6302B Adult, pediatric 0010-30-42733 EL6308B…
  • Page 143: Spo 2 Accessories

    23.2 SpO Accessories Extension Cables Module type Applicable patient Remark Mindray SpO module 0010-20-42710 040-000332-00 8 pins, purple connector Adult, pediatric, neonate Masimo SpO module 0010-30-42738 7 pins, white connector Nellcor SpO module 0010-20-42712 Sensors Mindray SpO module Type Model…

  • Page 144
    9000-10-07308 OXI-A/N Adult, neonate 9000-10-07336 Wavelength of Mindray 518B, 512E, 512F, 512G and 512H SpO sensors: red light 660 nm, infrared light 905 nm. Wavelength of Masimo SpO sensors: red light: 660 nm, infrared light: 940 nm. Wavelength of Nellcor SpO sensors: red light: 660 nm, infrared light: 890 nm.
  • Page 145: Nibp Accessories

    23.3 NIBP Accessories Tubing Type Applicable patient Adult, pediatric 6200-30-09688 Reusable Neonate 6200-30-11560 Cuff Applicable Limb Circumference Bladder Type Model Applied site patient (cm) Width (cm) CM1201 Infant 10 to 19 0010-30-12157 CM1202 Pediatric 18 to 26 12.2 0010-30-12158 Upper arm CM1203 Adult 24 to 35…

  • Page 146: Miscellaneous

    23.5 Miscellaneous Description Model Rechargeable lithium ion battery LI24I001A 022-000047-00 Test load MR6901 0651-20-77032 Test load MR6905 040-000413-00 Synchronous defibrillation input cable 009-001404-00 Grounding cable UL1015/14AWG 1000-21-00122 DC/AC adapter 0010-30-12471 Patient data management software kit 0651-30-77145 Carrying case and shield cover 115-018610-00 D3 back pouch 115-008708-00…

  • Page 147: A Specifications

    Specifications A.1 General Specifications Class I, equipment energized from an external and internal electrical power source. Type of protection against If you suspect the integrity of the external protective earthing or the protective earthing electrical shock wire, you should run the equipment on internal electrical power supply (battery). Type BF defibrillation proof for external defibrillation.

  • Page 148: Defibrillator Specifications

    Audio Indicator Gives alarm tones (45 to 85 dB), key tones, QRS tones; Speaker Supports PITCH TONE and multi-level tone modulation; Alarm tones comply with IEC60601-1-8. Multifunctional connector Standard Meets the requirements of EN60601-1 for short-circuit protection and leakage current Output impedance Typically 50Ω…

  • Page 149
    360 J defibrillation waveform into impedance of 25Ω, 50Ω, 75Ω, 100Ω, 125Ω, 150Ω, 175Ω Time (MS) Selected energy accuracy Impedance 25Ω 50Ω 75Ω 100Ω 125Ω 150Ω 175Ω Accuracy Energy ±2J ±2J ±2J ±2J ±2J ±2J ±2J ±2J ±2J 10 J ±2J 15 J ±15%…
  • Page 150
    Charge time (Note: at 20 °C of ambient temperature) Manual Defib From initiation of From initial power From initial power on to Charge time rhythm analysis to on to charge done charge done charge done 200J 360J 200J 360J 200J 360J 200J 360J…
  • Page 151: Pacer Specifications

    A.3 Pacer Specifications Pacing mode Demand, fixed Monophasic square wave pulse Output waveform pulse width 20 ms Accuracy: ±5% 40ppm to 170ppm Pacing rate Accuracy: ±1.5% Resolution: 5 ppm 0mA to 200mA, Pacing output Accuracy: ±5% or ±5mA, whichever is greater Resolution: 5mA 200 to 300 ms (depending on pacing rate) Refractory period…

  • Page 152
    Neonate 15 to 350 bpm HR measurement range Pediatric 15 to 350 bpm Adult 15 to 300 bpm HR accuracy ±1% or ±1bpm, which ever is greater HR resolution 1 bpm Measuring electrode: <0.1 μA Lead-off detection current Drive electrode: <1 μA Baseline recovery time <5 s (after defibrillation, in monitor mode and therapy mode) When the test is performed based on part 4.1.2.1 c) of ANSI/AAMI EC 13-2002,…
  • Page 153
    Difference input impedance >2.5 MΩ Apnea alarm time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s Mindray SpO Module *Measurement accuracy verification: The SpO accuracy has been verified in human experiments by comparing with arterial blood sample reference measured with a CO-oximeter.
  • Page 154
    Measurement range 20 to 254 bpm ±3 bpm (measured without motion) Accuracy ±5 bpm (measured with motion) Masimo SpO Module Measurement range 1 to 100% Resolution 70 to 100%: ±2% (measured without motion in adult/pediatric mode) 70 to 100%: ±3% (measured without motion in neonate mode) Accuracy 70 to 100%: ±3% (measured with motion)
  • Page 155: Power Supply Specifications

    NIBP Standards Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1, EN1060-3, EN1060-4 and SP10 Technique Oscillometry Mode of operation Manual, Auto and STAT Static pressure measurement 0kPa to 40.0kPa (0mmHg to 300mmHg) range Static pressure measurement ±0.4kPa (±3mmHg) accuracy 120s for adult and pediatric patients Maximum measurement time 90s for neonatal patients Adult…

  • Page 156: Recorder Specifications

    Battery 14.8V/3Ah, smart lithium ion battery, rechargeable and free of maintenance, one battery Battery type can be configured Less than 2 hours to 80% and less than 3 hours to 100% with equipment power off; Charge time Less than 3.5 hours to 80% and less than 4.5 hours to 100% with equipment power on. New fully charged battery Testing condition Without recording, typical ECG…

  • Page 157: Data Management Specifications

    A.8 Data Management Specifications Data Storage Internal CF card, 1G Bytes Marking Events 16 types of events, user customized Event recording Up to 1000 events for each patient. Waveform storage Up to 24 hours of consecutive ECG waveform Voice recording Max.

  • Page 158
    Vibration Complies with requirements of 21.102, ISO9919. Bump Complies with the requirements of 6.3.4.2, EN1789. Peak acceleration: 15g Duration: 6ms Number of impacts: 1000 Impact direction: vertical impacts are applied when the equipment under test is placed at normal operating position. Free fall Complies with the requirements of 6.3.4.3, EN1789.
  • Page 159
    The equipment meets the requirements of IEC 60601-1-2. NOTE Use of accessories, transducers, and cables other than those specified may result in increased emission and/or decreased electromagnetic immunity of the defibrillator/monitor. The equipment or its components should not be used adjacent to or stacked with other devices. If adjacent or stacked use is necessary, the equipment should be observed to verify normal operation in the configuration in which it will be used.
  • Page 160
    Guidance and Declaration — Electromagnetic Immunity The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment — guidance…
  • Page 161
    Guidance and Declaration — Electromagnetic Immunity The equipment is suitable for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment. IEC 60601 Immunity test Test level Compliance level Electromagnetic environment — guidance…
  • Page 162
    a The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. b Compliance level in the ISM frequency bands between 150 kHz to 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease the likelihood that portable/ mobile communication equipment could cause interference if it is inadvertently brought into patient areas.
  • Page 163: C Beneheart Defibrillator Shift Checklist

    BeneHeart Defibrillator Shift Checklist Inspect the defibrillator/monitor at the change of every shift. Place a “√” in the “Pass/Fail” box as you check the item , or place a “-” if not applicable. Describe the problem if there is any abnormity. Equipment Name: Serial Number: Department:…

  • Page 164
    Perform this test only when automatic selftest is not performed using pads cable or when selftest fails. Perform this test only when automatic selftest is not performed using paddles or when selftest fails. NOTE Remember to disconnect the test load when the test is finished. Otherwise, delay could happen in patient treatment.
  • Page 165: D Alarm Messages

    Alarm Messages This chapter lists only the most important physiological and technical alarm messages. Some messages appearing on your equipment may not be included. In this chapter: The “I” column indicates how indications of technological alarms are cleared after the hardkey or [Audio Pause] softkey is pressed: “A”…

  • Page 166: Technical Alarm Messages

    Measurement Alarm Message Cause and solution Nonsus. Vtac Pause Irr. Rhythm The pacer appears abnormal. Check the pacer. Resp Resp Apnea The respiration signal was so weak that the equipment cannot perform respiration analysis. Check the patient’s condition and the Resp connections. SpO2 SpO2 Desat The SpO…

  • Page 167
    Measurement Alarm Message Cause and solution SpO2 Too Much Light There is too much light on the SpO sensor. Move the sensor to a place with lower level of ambient light or cover the sensor to minimize the ambient light. SpO2 Low Signal The SpO signal is too low or too weak.
  • Page 168
    Measurement Alarm Message Cause and solution Keyboard Comm Err An error occurred to the keypad board, or there is a problem with the communications between the keypad board and the host. Restart the equipment. Therapy Module Comm Err An error occurred to the therapy module, or there is a problem with the communications between the therapy module and the host.
  • Page 169
    Measurement Alarm Message Cause and solution Monitoring Monitor Module Selftest Err An error occurred during MPM module power-on self test. module Contact your service personnel. Monitor Module Reset Err MPM module reset abnormally. In this case, the MPM module restores to default configuration. You can ignore this problem.
  • Page 170
    FOR YOUR NOTES…
  • Page 171: E Electrical Safety Inspection

    Electrical Safety Inspection The following electrical safety tests are recommended as part of a comprehensive preventive maintenance program. They are a proven means of detecting abnormalities that, if undetected, could prove dangerous to either the patient or the operator. Additional tests may be required according to local regulations. All tests can be performed by using commercially available safety analyzer test equipment.

  • Page 172: Device Labelling

    E.3 Device Labelling Check the labels provided by the manufacturer or the healthcare facilities are present and legible. Main unit label Integrated warning labels E.4 Protective Earth Resistance Plug the probes of the analyzer into the device’s protective earth terminal and protective earth terminal of the AC power cord.

  • Page 173: Patient Leakage Current

    E.6 Patient Leakage Current Patient leakage currents are measured between a selected applied part and mains earth. All measurements have a true RMS only The following outlet conditions apply when performing the Patient Leakage Current test. normal polarity (Normal Condition) reverse polarity (Normal Condition) normal polarity with open neutral (Single Fault Condition) reverse polarity with open neutral (Single Fault Condition)

  • Page 174: Patient Auxiliary Current

    E.8 Patient Auxiliary Current Patient Auxiliary currents are measured between any selected Applied Part connector and the remaining Applied Part connector s. All measurements may have a true RMS only response. The following outlet conditions apply when performing the Patient Auxiliary Current test. normal polarity (Normal Condition) reverse polarity (Normal Condition) normal polarity with open neutral (Single Fault Condition)

  • Page 175: F Symbols And Abbreviations

    Symbols and Abbreviations F.1 Units μA microampere μV microvolt ampere ampere hour beat per minute bit per second ºC centigrade cubic centimeter centimeter decibel dyne second ºF fahrenheit gram gigahertz gutta hour hertz inch Joule kilo kilogram kilopascal litre pound meter Milliampere hour mega byte…

  • Page 176: Symbols

    mmHg millimeters of mercury millisecond millivolt milliwatt MΩ megaohm nanometer breaths per minute second volt volt ampere Ω watt F.2 Symbols negative, minus – percent per; divide; or plus equal to < less than > greater than ≤ less than or equal to ≥…

  • Page 177
    arterial left foot augmented lead left arm augmented lead right arm augmented lead awRR airway respiratory rate brachial aterial pressure bispectral index blood pressure BPSK binary phase shift keying body surface area blood temperature BTPS body temperature and pressure, saturated C.I.
  • Page 178
    ethylene oxide EtO2 end-tidal oxygen femoral arterial pressure Federal Communication Commission Food and Drug Administration FEV1.0% first second forced expiratory volume ratio fraction of inspired FiCO2 fraction of inspired carbon oxygen FiN2O fraction of inspired nitrous oxide FiO2 fraction of inspired oxygen FPGA field programmable gate array flow-volume…
  • Page 179
    MetHb methemoglobin magnetic resonance imaging expiratory minute volume inspiratory minute volume not applied neonate NIBP noninvasive blood pressure negative inspiratory pressure oxygen O2CI oxygen consumption index oxygen extraction ratio operating room oxyCRG oxygen cardio-respirogram pulmonary artery airway pressure PAWP pulmonary artery wedge pressure photodetector pediatric PEEP…
  • Page 180
    sevoflurane self-maintenance stroke index satellite module rack SpO2 arterial oxygen saturation from pulse oximetry signal quality index suppression ratio systolic time ratio systemic vascular resistance SVRI systemic vascular resistance index Sync synchronization systolic pressure Taxil axillary temperature temperature difference Temp temperature thoracic fluid content thoracic fluid index…
  • Page 181: G Device Tracking

    In order to provide high quality product and perform better service, we are going to track our product. Please contact us with the device tracking information when you have received your defibrillator/monitor: Please fill the information in the next page, cut the table and fax it to +86 755 26582934. You can also email your information to service@mindray.com.

  • Page 182
    FOR YOUR NOTES…
  • Page 186
    P/N: 046-001653-00 (6.0)

  • Author

  • View
    6.318

  • Download
    2.348

Embed Size (px)

DESCRIPTION

Service manual

Text of BeneHeart D3 Service Manual v2.0 English (1)

BeneHeart D3 Defibrillator/Monitor Service Manual I Intellectual
Property Statement SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO.,
LTD. (hereinafter called Mindray) owns the intellectual property
rights to this product and this manual. This manual may refer to
information protected by copyrights or patents and does not convey
any license under the patent rights of Mindray, nor the rights of
others. Mindray does not assume any liability arising out of any
infringements of patents or other rights of third parties. ,,andare
the registered trademarks or trademarks owned by Mindray in China
and other countries. Revision History This manual has a revision
number. This revision number changes whenever the manual is updated
due to software or technical specification change. Contents of this
manual are subject to change without prior notice. Version
number2.0 Release time:March 2011 2010 — 2011 Shenzhen Mindray
Bio-Medical Electronics Co., Ltd. All rights reserved. Company
Contact Manufacturer:Shenzhen Mindray Bio-Medical Electronics Co.,
Ltd. E-mail Address:[email protected] Tel:+86 755 26582479, +86
755 26582888 Fax:+86 755 26582934, +86 755 26582500 II Preface
Manual Purpose This manual provides detailed information about the
assembling, dissembling, testing and troubleshooting of the
equipment to support effective troubleshooting and repair. It is
not intended to be a comprehensive, in-depth explanation of the
product architecture or technical implementation. Observance of the
manual is a prerequisite for proper equipment maintenance and
prevents equipment damage and personnel injury. Intended Audience
This manual is for biomedical engineers, authorized technicians or
service representatives responsible for troubleshooting, repairing
and maintaining the defibrillator/ monitors Passwords Passwords may
be required to access different modes. The passwords are listed
below: Installation mode:888888 Service mode: 332888 Configuration
mode:315666 1 Contents 1 Safety
…………………………………………………………………………………………………………………
1-11.1 Safety Information
…………………………………………………………………………………………….
1-11.1.1 Dangers
………………………………………………………………………………………………..
1-21.1.2 Warnings
………………………………………………………………………………………………
1-21.1.3 Cautions
……………………………………………………………………………………………….
1-21.1.4 Notes
……………………………………………………………………………………………………
1-31.2 Equipment Symbols
…………………………………………………………………………………………..
1-32 Theory of Operation
…………………………………………………………………………………………..
2-12.1 The Basics
………………………………………………………………………………………………………..
2-12.1.1 Overview
………………………………………………………………………………………………
2-12.1.2 Main Functions
……………………………………………………………………………………..
2-12.2 Components
……………………………………………………………………………………………………..
2-22.3 Main Unit
…………………………………………………………………………………………………………
2-22.4 Front Housing Assembly
…………………………………………………………………………………….
2-42.5 Paddle Tray
………………………………………………………………………………………………………
2-52.6 Rear Housing Assembly
……………………………………………………………………………………..
2-52.6.1 Power System
………………………………………………………………………………………..
2-52.6.2 Main Control System
……………………………………………………………………………..
2-62.6.3 Therapy System
……………………………………………………………………………………..
2-62.6.4 Parameter Measurement System
………………………………………………………………
2-72.7 External Device Connectors
………………………………………………………………………………..
2-83 Unpacking and Installation
…………………………………………………………………………………
3-13.1 Unpacking the Equipment
…………………………………………………………………………………..
3-13.2 Preparation for Installation
………………………………………………………………………………….
3-23.2.1 Preparation for Installation
Site………………………………………………………………..
3-23.2.2 Electrical Requirements
………………………………………………………………………….
3-33.3 Preparation for Power On
……………………………………………………………………………………
3-33.4 User Test
………………………………………………………………………………………………………….
3-44 Testing and Maintenance
…………………………………………………………………………………….
4-14.1 Introduction
………………………………………………………………………………………………………
4-14.1.1 Test Report
……………………………………………………………………………………………
4-24.1.2 Recommended Frequency
……………………………………………………………………….
4-24.2 Preventive maintenance
……………………………………………………………………………………..
4-34.2.1 Visual Test
…………………………………………………………………………………………….
4-34.3 Power On Test
…………………………………………………………………………………………………..
4-34.4 User Test
………………………………………………………………………………………………………….
4-3 2 4.5 Password for Installation Mode
…………………………………………………………………………..
4-44.6 Module Performance Tests
………………………………………………………………………………….
4-54.6.1 Manual Defibrillation Test
………………………………………………………………………
4-54.6.2 Pacing Test
……………………………………………………………………………………………
4-74.6.3 ECG Test
………………………………………………………………………………………………
4-84.6.4 Resp Test
………………………………………………………………………………………………
4-94.6.5 SpO2 Test
………………………………………………………………………………………………
4-94.7 Electrical Safety Tests
………………………………………………………………………………………..
4-94.8 Recorder Check
……………………………………………………………………………………………….
4-104.9 Factory Service
……………………………………………………………………………………………….
4-104.9.1 Password for Service Mode
…………………………………………………………………..
4-104.9.2 Accessing Service Mode Menu
………………………………………………………………
4-104.9.3 Calibrating/Zeroing Impedance
……………………………………………………………….
4-114.9.4 Device Information
………………………………………………………………………………
4-124.9.5 Checking Failure Code
………………………………………………………………………….
4-124.9.6 Inputting Serial Number
………………………………………………………………………..
4-134.9.7 Paddle Open Circuit Display
………………………………………………………………….
4-135 Hardware and Software Upgrade
………………………………………………………………………..
5-15.1 Hardware Upgrade
…………………………………………………………………………………………….
5-15.1.1 Upgrade MPM module from ECG only to ECG plus SPO2
………………………… 5-15.1.2 Upgrade the Therapy Module
………………………………………………………………….
5-25.2 Software Upgrade through a PC
………………………………………………………………………….
5-35.2.1 Installing Mindray Patient Monitor Software Upgrade Tool
………………………… 5-45.2.2 Software Upgrade Procedure
…………………………………………………………………..
5-65.3 Software Upgrade through a USB Memory
…………………………………………………………..
5-75.3.1 Precautions
……………………………………………………………………………………………
5-75.3.2 Software Upgrade Procedure
…………………………………………………………………..
5-76 Troubleshooting
………………………………………………………………………………………………….
6-16.1 Overview
………………………………………………………………………………………………………….
6-16.2 Part Replacement
………………………………………………………………………………………………
6-16.3 Checking Defibrillator/Monitor Status
………………………………………………………………….
6-16.4 Checking Device Information
……………………………………………………………………………..
6-26.5 Checking Technical Alarm
………………………………………………………………………………….
6-26.6 Troubleshooting Guide
……………………………………………………………………………………….
6-36.6.1 Defibrillation Problems
…………………………………………………………………………..
6-36.6.2 Pacing Problems
…………………………………………………………………………………….
6-56.6.3 Power On/Off Problems
………………………………………………………………………….
6-56.6.4 Display Problems
…………………………………………………………………………………..
6-66.6.5 Alarm Problems
……………………………………………………………………………………..
6-76.6.6 Button and Knob Problems
……………………………………………………………………..
6-86.6.7 Recorder Problems
…………………………………………………………………………………
6-96.6.8 Output Interface Problems
………………………………………………………………………
6-9 3 6.6.9 CF Card Problems
………………………………………………………………………………..
6-106.6.10 Power Supply Problems
………………………………………………………………………
6-106.6.11 Software Upgrade Problems
………………………………………………………………….
6-116.7 Technical Alarm Messages
………………………………………………………………………………..
6-126.8 Error Codes
…………………………………………………………………………………………………….
6-156.8.1 Therapy Module Error Codes
…………………………………………………………………
6-156.8.2 Power Module Error Codes
……………………………………………………………………
6-176.8.3 Main Control Error Codes
……………………………………………………………………..
6-176.8.4 MPM Error Codes
………………………………………………………………………………..
6-187 Disassembly and Repair
……………………………………………………………………………………..
7-17.1 Tools Required
………………………………………………………………………………………………….
7-17.2 Preparations for Disassembly
………………………………………………………………………………
7-27.3 Disassembling the Main Unit
………………………………………………………………………………
7-37.3.1 Removing Hook Mount (if configured)
…………………………………………………….
7-37.3.2 Removing Paddle Tray
……………………………………………………………………………
7-47.3.3 Separating the Housing
…………………………………………………………………………..
7-57.3.4 Discharging the Capacitor
……………………………………………………………………….
7-67.3.5 Disassembling the MPM Module Assembly
……………………………………………… 7-77.3.6
Removing the Parameter Panel
Assembly………………………………………………….
7-87.3.7 Removing the Therapy Module
………………………………………………………………..
7-97.3.8 Disassembling the Power Base Assembly
……………………………………………….. 7-107.3.9
Disassembling the AC/DC Power Supply Board
………………………………………. 7-117.3.10
Disconnecting the Therapy Port Cable
…………………………………………………..
7-127.3.11 Removing the Main Board
Assembly…………………………………………………….
7-137.3.12 Checking Waterproof Strips before Reassembling
………………………………….. 7-147.4 Disassembling the
Front Housing Assembly
……………………………………………………….
7-157.4.1 Removing the Keypad Board
…………………………………………………………………
7-167.4.2 Removing Display Assembly
…………………………………………………………………
7-177.4.3 Removing the Alarm Lamp Board and Assistant Keypad Board
………………… 7-187.4.4 Removing the Speaker
………………………………………………………………………….
7-187.4.5 Removing the Mode Select Knob
…………………………………………………………..
7-197.4.6 Removing the Encoder
………………………………………………………………………….
7-197.4.7 Checking Waterproof Material on the Front Housing
……………………………….. 7-207.5 Removing the
Recorder
……………………………………………………………………………………
7-207.6 Disassembling the Recorder
………………………………………………………………………………
7-218 Parts
………………………………………………………………………………………………………………….
8-18.1 Introduction
………………………………………………………………………………………………………
8-18.2 Main Unit
…………………………………………………………………………………………………………
8-28.2.1 Exploded View
………………………………………………………………………………………
8-28.2.2 Parts List
………………………………………………………………………………………………
8-28.3 Front Housing Assembly (115-007249-00)
……………………………………………………………
8-58.3.1 Exploded View
………………………………………………………………………………………
8-5 4 8.3.2 Parts List
………………………………………………………………………………………………
8-58.4 Rear Cover Assembly (115-007255-00)
………………………………………………………………..
8-88.4.1 Exploded View
………………………………………………………………………………………
8-88.4.2 Parts List
………………………………………………………………………………………………
8-88.5 Power Base Assembly115-007253-00
……………………………………………………………
8-98.5.1 Exploded View
………………………………………………………………………………………
8-98.5.2 Parts List
………………………………………………………………………………………………
8-98.6 Main Board Assembly (115-007254-00)
……………………………………………………………..
8-108.6.1 Exploded View
…………………………………………………………………………………….
8-108.6.2 Parts List
…………………………………………………………………………………………….
8-108.7 Treatment Board Subassembly (115-007913-00)
…………………………………………………..
8-118.7.1 Exploded View
……………………………………………………………………………………..
8-118.7.2 Parts List
……………………………………………………………………………………………..
8-118.8 Parameter Panel Assembly (115-007258-00)
……………………………………………………….
8-128.8.1 Exploded View
…………………………………………………………………………………….
8-128.8.2 Parts List
…………………………………………………………………………………………….
8-128.9 MPM Assembly (ECG + masimo SpO2) (115-007264-00)
…………………………………………….. 8-138.9.1
Exploded View
…………………………………………………………………………………….
8-138.9.2 Parts List
…………………………………………………………………………………………….
8-138.10 Capacitor Assembly (801-0651-00040-00)
……………………………………………………..
8-148.10.1 Exploded View
…………………………………………………………………………………..
8-148.10.2 Parts List
…………………………………………………………………………………………..
8-148.11 Paddle Tray Assembly
…………………………………………………………………………………….
8-158.11.1 Exploded View
…………………………………………………………………………………..
8-158.11.2 Parts List
……………………………………………………………………………………………
8-158.12 External Paddles Assembly (0651-30-76937)
…………………………………………………….
8-168.12.1 Exploded View
…………………………………………………………………………………..
8-168.12.2 Parts List
…………………………………………………………………………………………..
8-168.13 Pediatric Sternum Paddle Kit (0651-30-76920)
………………………………………………….
8-178.13.1 Exploded View
…………………………………………………………………………………..
8-178.13.2 Parts List
…………………………………………………………………………………………..
8-178.14 Adult Sternum Paddle Kit (0651-30-76922)
………………………………………………………
8-188.14.1 Exploded View
…………………………………………………………………………………..
8-188.14.2 Parts List
…………………………………………………………………………………………..
8-188.15 Pediatric Apex Paddle Kit (0651-30-76921)
………………………………………………………
8-198.15.1 Exploded View
…………………………………………………………………………………..
8-198.15.2 Parts List
…………………………………………………………………………………………..
8-198.16 Adult Apex Paddle Kit (0651-30-76923)
…………………………………………………………..
8-208.16.1 Exploded View
…………………………………………………………………………………..
8-208.16.2 Parts List
…………………………………………………………………………………………..
8-208.17 External Paddle Cable (0651-20-76827)
……………………………………………………………
8-218.17.1 Exploded View
…………………………………………………………………………………..
8-218.17.2 Parts List
…………………………………………………………………………………………..
8-218.18 Pothook Assembly (0651-30-76864)
………………………………………………………………..
8-22 5 8.18.1 Exploded View
…………………………………………………………………………………..
8-228.18.2 Parts List
…………………………………………………………………………………………..
8-228.19 Replacement Parts
………………………………………………………………………………………….
8-228.19.1 Main Unit
………………………………………………………………………………………….
8-238.19.2 Connecting Cables
……………………………………………………………………………..
8-24A Electrical Safety Inspection
……………………………………………………………………………………
1 6 FOR YOUR NOTES 1-1 1 Safety 1.1 Safety Information DANGER
Indicates an imminent hazard that, if not avoided, will result in
death, serious personal injury or property damage. WARNING
Indicates a potential hazard or unsafe maintenance practice that,
if not avoided, could result in death, serious personal injury,
product / property damage. CAUTION Indicates a potential hazard or
unsafe maintenance practice that, if not avoided, could result in
minor personal injury or product/property damage NOTE Provides
application tips or other useful information to ensure that you can
better service your product. 1-2 1.1.1 Dangers WARNING Do not open
the equipment cases to avoid shock hazard. All servicing and future
upgrades must be carried out by the personnel trained and
authorized by our company only. 1.1.2 Warnings WARNING To avoid
high voltage shock, disconnect the defibrillator/monitor from AC
adapter and remove the batteries before disassembly. The equipment
must be connected to a properly installed power socket with
protective earth contacts only. If the installation does not
provide a protective earth conductor, do not use this socket and
operate the equipment on rechargeable batteries. When disposing of
the packaging material, be sure to observe the applicable waste
control regulations and keep it out of childrens reach. 1.1.3
Cautions CAUTION Make sure that no electromagnetic radiation
interferes with the performance of the equipment when preparing to
carry out performance tests. Mobile phone, X-ray equipment or MRI
devices are a possible source of interference as they may emit
higher levels of electromagnetic radiation. Before connecting the
equipment to the power line, check that the voltage and frequency
ratings of the power line are the same as those indicated on the
equipments label or in this manual. Protect the equipment from
damage caused by drop, impact, strong vibration or other mechanical
force during servicing. 1-3 1.1.4 Notes NOTE Refer to Operation
Manual for detailed operation and other information. 1.2 Equipment
Symbols Attention: Please read this manual carefully before
servicing. Equipotential terminal Danger: High-voltage Status
indicator Alternating current(AC) Network connector Battery USB
connector ESD warning symbol for Electrostatic sensitive devices.
Type CF applied part. Defibrillator-proof protection against
electric shock. Type BF applied part. Defibrillator-proof
protection against electric shock. 1-4 FOR YOUR NOTES 2-1 2 Theory
of Operation 2.1 The Basics 2.1.1 Overview The BeneHeart D3
defibrillator/monitor (hereinafter called the equipment) provides
four operating modes: Manual Defib, AED, Pacer, and Monitor. The
equipment is for use in hospital and pre-hospital settings. It
adopts the most advanced biphasic defibrillation technology and can
deliver up to 360J of defibrillation energy. The equipment has an
7.0 inch color TFT LCD display with LED Backlight. 2.1.2 Main
Functions The equipment has the following main functions: Manual
Defib Mode In Manual Defib Mode, the operator analyzes the patients
ECG, and, if appropriate, follows this procedure: 1Select the
Manual Defib mode, adjust the energy level if necessary 2Charge;
and 3Deliver the shock. Defibrillation may be performed through
external paddles or multifunction electrode pads. In Manual Defib
Mode, you can also perform synchronized cardioversion. AED Mode In
AED mode, the equipment automatically analyzes the patients ECG
rhythm and indicates whether or not a shockable rhythm is detected.
Voice prompts provide easy-to-follow instructions and patient
information to guide you through the defibrillation process.
Messages and flashing buttons are also presented to reinforce the
voice prompts. Pacer Mode The Pacer Mode offers non-invasive
transcutaneous pacing therapy. Pace pulses are delivered through
multifunction electrode pads using a monophasic square waveform.
2-2 Monitor Mode In Monitor Mode, the equipment is intended for
monitoring, displaying, reviewing, storing and printing multiple
physiological parameters and waveforms including ECG and pulse
oximetry (SpO2), 2.2 Components The equipment consists of a main
unit, accessories and PC software. The main unit is the core of the
equipment. It provides: Overall system control; System power
supply; Display; Defibrillation and pacing; AED ; Man-mahcine
interface; Audible and visible alarms; Multiple parameter
measurements; External connectors and communication; and Recording,
printing and data storage. 2.3 Main Unit The main unit is composed
of the front housing assembly, rear housing assembly and the paddle
tray assembly. External paddles are rested in the paddle tray. The
front housing assembly mainly consists of LCD, keypad board,
speaker, microphone, Mode Select knob, navigation knob, alarm lamp
board, front housing and front housing sheet metal, etc. The rear
housing assembly consists of CPU board, therapy module, high
voltage capacitors, MPM module, power management board, fan,
measurement module panel, therapy port, recorder and rear housing,
etc. The paddle tray is for holding the external paddles. 2-3 The
main unit consists of the following subsystem: Input subsystem: Its
input includes keypad board, microphone, Mode Select knob,
navigation knob, and paddle handle controls. Output subsystem:
includes display screen, alarm lamp board, recorder, and speaker
Processing and communication subsystem: includes CPU board, therapy
module, MPM module and power manager board. Power management
subsystem: includes batteries, AC/DC board and power management
board. External device connection subsystem: includes USB
connector, network connector, and multifunction connector for
synchronous input. System Structure B12 mode select knobB13
navigation knobB14 speaker Kaypad boardLCD & inverterStatus
indicatorMain control boardTherapy modulePaddle Tray assemblyPaddle
on-position detectionMPMPower management boardPaddle trayTherapy
port50discharge resistanceECG connectorSPO2 connectorExternal
paddleFanW1Front housing assemblyExternal componentsW13W14RJ45
USBI/O, extendedW2 W4W5W7 B15 microphoneW3AC/DC
module18V/GNDW8C1C2C3C4RecorderBattery W11W6W9W10Rear housing
assemblyConnecting cableSocket connectorHigh-voltage
capacitorW12W15 2-4 System Signal Flow 2.4 Front Housing Assembly
The front housing assembly consists of display assembly, a keypad
board, a speaker, a microphone, a Mode Select knob, a navigation
knob, an alarm lamp board, a front housing and front housing sheet
metal, etc. Navigation Knob You can rotate the knob clockwise or
counterclockwise and then press it to confirm a selection. The knob
is connected to the keypad board. Mode Select Knob A 8-position
encoder is used to select the operating mode (Monitor, Manual
Defib, AED and Pacer) and power-off. The unused positions are
mechanically disabled. Speaker The speaker emits alarm tones,
key-stroke tone, heart beats and PR sound. It supports the
functions of PITCH TONE and the multi-level volume. The speaker is
connected to the keypad board. Microphone It provides the function
of voice recording. 2-5 Alarm Lamp Board The keypad board
interfaces with the alarm lamp board. The alarm lamp transmits
signals to drive the green and yellow alarm lamp. 2.5 Paddle Tray
The paddle tray is used to hold paddles. It has a 50 ohm test load
and position detective switch inside. When the equipment runs self
tests, test current will pass through the test load. 2.6 Rear
Housing Assembly Rear housing assembly consists of the CPU board,
the therapy module, high voltage capacitors, a MPM module, a power
management board, a fan, a rear housing, a measurement module
panel, and a therapy port, etc. 2.6.1 Power System 1.AC/DC board It
has AC mains as an input and outputs 18VDC. 2.Battery Its rated
voltage is 14.8V, 3000mAh. 3.Power Management Board It is
responsible for power transform and battery charge control. The
system has four power supplies: 18V (when AC mains is used) or
14.8V (when batteries are used), 12V, 5V, and 3.3V. The priority of
system power supply is AC mains, Battery. That is to say, when AC
is not available, Battery is used. 2-6 2.6.2 Main Control System
The CPU board is connected with the power management board with
stacking connectors, as shown below. The main control module mainly
consists of the CPU and FPGA. CPU is used to provide least required
internal storage, program memory, large capacity non-volatile
storage, and the watch dog. It connects EEPROM and other peripheral
ICs such as Ethernet PHY chip. FPGA performs the main functions of
display and audio. Besides, it has the function of adapting
interfaces from CPU to MPM module, the keypad board and the
recorder. CPU controls FPGA via Flexbus. 2.6.3 Therapy System The
therapy system provides the functions of defibrillation, pacing and
AED analysis. Therapy module is undetachable. The therapy module
adopts DSP+MCU framework. MCU is responsible for therapy control
while DSP for ECG and impedance detection, AED algorithm,
monitoring algorithm, pacing algorithm, auxiliary therapy control,
etc. 2-7 Recorder The recorder receives data from the CPU board and
then sends the data to a thermal head for printing. The recorder
front panel has a key for starting/ stopping the recorder and a
green indicator which is lit when working normally. The recorder is
connected to the keypad board which board provides connection for
the TR6F recorder. The block diagram and functional modules of the
recorder are shown as below. ModuleDescription Power
InterfaceIntroduces DC power supply from the CPU board. Recorder
Power ModuleAdjusts input voltage to run each module. Recorder CPU
Coordinates module communication, controls and processes module
status. Keypad board Interface Serves as the data communication
channel between the keypad board and the recorder CPU. Motor Drive
Circuit Receives control signals sent by the recorder CPU to drive
the step motor. Keypad and Indicator Interface Sends keypad
commands to CPU and receives CPU commands to control the indicator.
FPC Interface Sends print head information to CPU and receives CPU
commands to control the print head. 2.6.4 Parameter Measurement
System MPM module is used to provide parameter monitoring. However,
ECG monitoring can also be implemented by the therapy module. . 2-8
2.7 External Device Connectors 1.Paddle Tray 2.Hook mount 3.Battery
4.External power input: It connects an AC power cord or a DC/AC
adapter to run the equipment respectively on the external AC mains
or DC power supply. 5.Equipotential grounding terminal: When the
defibrillator/monitor and other devices are to be used together,
their equipotential grounding terminals should be connected
together to eliminate the potential difference between them.
6.Handle 7.Multifunctional connector: defibrillation
synchronization input. 8.USB connector: It connects the USB memory
for data export. Data stored in the internal CF card can be
transferred to the USB memory and then export to a PC via the data
management software. 9.Network connector: It is a standard RJ45
connector, through which software can be upgraded. 124 36 7 8 9 5
3-1 3 Unpacking and Installation This chapter provides information
you need to install a defibrillator/monitor ready for use. 3.1
Unpacking the Equipment Open the package and take out the packing
list. Check that all the articles included in the packing list are
available and the quantity and specification are correct. All the
optional parts purchased by the customer shall also be checked.
Notify the supplier if provided components are not correct as
compared to the packing list. In case of damage during
transportation, keep the packing material and notify the supplier
immediately. Keep the packing material till new equipment is
accepted. The following pictures show the defibrillator/monitor and
accessory packing. Main unit packing Accessory packing 3-2 3.2
Preparation for Installation 3.2.1 Preparation for Installation
Site 1.Ensure that the site meets all safety, environmental and
power requirements 2.Check that required power sockets are
available. 3.Check that a network connector is available if the
defibrillator/monitor needs to be connected to network. WARNING
Only power cables provided with the system may be used. For reasons
of safety, power (mains) extension cables or adapters shall not be
used. Environmental Requirements WARNING To avoid explosion hazard,
do not use the equipment in the presence of flammable anaesthetics,
vapours or liquids. CAUTION The environment where the
defibrillator/monitor will be used should be reasonably free from
vibration, dust and corrosive substances. If these conditions are
not met, the accuracy of the system may be affected and damage may
occur. The environmental specification is as follows: Operating
Temperature0 to 45 Operating humidity15% to 95%, (non-condensing)
Operating altitude-381m to +4575 m (-1250 ft to 15000 ft, or
106.2kPa to 57kPa) Storage temperature-30 to 70 Storage humidity10%
to 95%, (non-condensing) Storage altitude-381m to +4575 m (-1250 ft
to 15000 ft, or 106.2kPa to 57kPa) 3-3 3.2.2 Electrical
Requirements Check cables and power cords. Make sure that: 1.All
system cables, power cords and power plugs are not damaged, and
pins are not loose. Otherwise, remove it from use. 2.The insulation
of patient cables and leadwires is not damaged, and connectors are
not loose. WARNING Only power sockets with protective grounding can
be used. The electrical specification is as follows: Line
voltage:100 to 240VAC Current:1.8 to 0.8 A Frequency:50/60Hz 3.3
Preparation for Power On Before connecting the power cord to the
defibrillator/monitor’s power input, check that The mains voltage
meets the requirement. 3-wire power cord is used. The power socket
should be 3-wire also. This ensures that the defibrillator/monitor
is properly grounded. Do not use 2-wire power cord or socket. The
equipotential grounding terminals should be connected together when
the defibrillator/monitor and other devices are to be used
together. The defibrillator/monitor is not placed under the
infusion bag or placed where their might be liquid spillage. This
protects the defibrillator/monitor from liquid ingress. 3-4 3.4
User Test A user test shall be performed after the
defibrillator/monitor is installed. Follow this procedure:
1.Connect AC mains or install the battery. 2.Connect the external
paddles. If pads are used, connect the test load. 3.Select the Main
Menu button on the equipments front panel and select [User Test
>>]. Select all test items and press [Start] to perform user
test. NOTE Install the battery and properly place the external
paddles in the paddle tray or connect the pads cable and 50 test
load. Otherwise the User Test will fail. Refer to BeneHeart D3
Operating Manual for the detailed information on user test. 4-1 4
Testing and Maintenance 4.1 Introduction To ensure the equipment
always functions normally, qualified service personnel should
perform regular inspection, maintenance and test. This chapter
provides a checklist of the testing procedures for the equipment
with recommended test equipment and frequency. The service
personnel should perform the testing and maintenance procedures as
required and use appropriate test equipment. The testing procedures
provided in this chapter are intended to verify that the equipment
meets the performance specifications. If the equipment or a module
fails to perform as specified in any test, repairs or replacement
must be done to correct the problem. If the problem persists,
contact our Customer Service Department. CAUTION All tests should
be performed by qualified service personnel only. Care should be
taken to change the settings in [Installation Mode] and [Service
Mode] menus to avoid loss of data. Before testing, service
personnel should acquaint themselves with the test tools and make
sure that test tools and cables are applicable. When testing
monitoring parameters, move the Mode Select knob to Monitor to
access the Monitor Mode. When performing therapy function tests,
move the Mode Select knob to corresponding mode. 4-2 4.1.1 Test
Report After completing the tests, service personnel are required
to record test results in this table and report them to Mindray
Customer Service Department. See the Test Report at the end of this
chapter. 4.1.2 Recommended Frequency Test item After repairFunction
suspected6 months 12 months 24 months Visual inspection Power-on
Test User test Recorder check Manual defibrillation tests Charge/
discharge Energy disarming Synchronous defibrillation Pacing test
ECG Performance test Module calibration RespPerformance test
SpO2Performance test Electrical safety tests as per IEC60601-1
Earth leakage current Patient leakage current Patient auxiliary
current 4-3 4.2 Preventive maintenance 4.2.1 Visual Test Inspect
the equipment for obvious signs of damage. The test is passed if
the equipment has no obvious signs of damage. Follow these
guidelines when inspecting the equipment: Carefully inspect the
housing, the display screen and the buttons for physical damage.
Inspect accessories for signs of damage. Inspect all external
connections for loose connectors, bent pins or frayed cables.
Inspect all connectors on the equipment for loose connectors or
bent pins. Make sure that safety labels and data plates on the
equipment are clearly legible. 4.3 Power On Test This test is to
verify that the defibrillator/ monitor can power on normally. The
test is passed if the defibrillator/ monitor starts up by following
this procedure: 1.Place the external paddles on paddle tray, insert
the battery in the battery compartment, and then connect the
equipment with AC mains. In this case, both the AC indicator and
battery indicator shall light. 2.Turn the Mode Select knob to
Monitor. Check that the equipment passes the self test and is
turned on properly. 3.Check the display of technical alarm area,
prompt area and battery status indicator on the upper right corner
of the main screen to judge whether the equipment runs normally.
4.4 User Test Follow this procedure to perform user test: 1.If you
use external paddles, place them on the paddle tray; if you use a
pads cable, connect it to the test load. 2.Insert the battery into
the equipment. Connect the AC mains if no battery is available.
3.Select the Main Menu button on the equipments front panel. In the
Main Menu, select [User Test>>]. Then a prompt Enter user
test? pops up. Select Yes to enter the User Test screen. 4.Check
the test items you want to perform and select [Start] to start user
test The test results indicate the condition of the system. If any
item fails, the Red Cross status indicator flashes. 4-4 If you
cannot pass User Test or the message Connect paddles cable, and
place paddles in paddle tray is shown when paddle cable is
connected and paddles are placed in paddle tray, check paddles
status. Select the Monitor mode. Press and hold the [Event]
hardkey, and then press the [Lead Select] hardkey on the front
panel, the following screen appears. Observe the reading of Lead
Stat: 0 x 382: Paddles are properly placed in paddle tray. 0 x 182:
The travel switch indicating paddle status may fail, but impedance
is correct. 0 x 102 :Paddles are not properly placed in paddle tray
and the impedance value is not correct. 4.5 Password for
Installation Mode Accessing installation mode is password
protected. The required password is set to 888888 before the
equipment leaves the factory. 4-5 4.6 Module Performance Tests
4.6.1 Manual Defibrillation Test Test tools: Defibrillator/pacer
analyzer Charge/Discharge 1.Remove the batteries and connect the
equipment with AC mains. Turn the Mode Select knob to Manual Defib.
2.Connect the external paddles to the equipment and place the
paddles on the defibrillator/pacer analyzer. 3.Enter the
Configuration-Main screen. From the Record Setup menu set [Shock
Event] to [On] so that shock events can be recorded automatically
if happened. 4.Set the analyzer to Energy Measurement mode. In this
case, the energy value should be displayed as 0 or blank. 5.Select
the energy level to 1J. 6.Charge/discharge the equipment to verify
the energies measured by the analyzer meet the following accuracy:
Selected Energy (J)Measured Value (J) 10 to 3 10085 to 115 360306
to 414 7Set the energy to 100J and 360J respectively. Repeat step
6. 8.Disconnect the equipment from the AC mains. Run the equipment
on fully charged battery. Move the Mode Select knob to Manual
Defib. Repeat steps 5 to 7. 9.Use multifunctional electrode pads.
Repeat steps 5 to 7. 10.Verify that the equipment records the shock
events automatically and correctly. 4-6 Energy Disarming 1.Run the
equipment on fully charged battery. Move the Mode Select knob to
Manual Defib. 2.Connect the external paddles to the equipment and
place the paddles on the defibrillator/pacer analyzer. 3.Set the
analyzer to Energy Measurement mode. In this case, the energy value
should be displayed as 0 or blank. 4.Select the energy level to
360J. 5.Charge the equipment. 6.Verify that the charge tone is
issued during charging. 7.Press the Disarm soft key to discharge
the energy internally. 8.Verify that a prompt Charge Removed
appears and the charge done tone stops. 9.Verify that the value
measured by the analyzer is 0J or blank. 10.Enter the
Configuration-Main menu, select [Manual Therapy Setup] and set
[Time to Auto Disarm] to [60s]. 11.Exit Configuration Management.
The equipment restarts automatically. 12.Set the analyzer to Energy
Measurement mode. In this case, the energy value should be
displayed as 0 or blank. 13.Select the energy level to 360J.
14.Charge the equipment. Count time after charging is completed..
Verify that the prompt Shock Removed appears on the equipment and
the energy measured by the analyzer is 0J or blank after 60
seconds. 15.Use multifunctional electrode pads. Repeat steps 3 to
14. Synchronous Defibrillation 1.Connect the external paddles and
ECG cable to the equipment. Place the paddles ECG electrodes on the
defibrillator/pacer analyzer. 2.Set the analyzer to Measurement
Mode and output normal sinus rhythms, e.g. amplitude value 1mV and
HR 60bpm. 3.Enter Configuration Management. In the [Manual Therapy
Setup] menu, set [Sync After Shock] to [On]. 4.Adjust the energy
setting of the equipment to be 10J. 5.Press the [Sync On] soft key
to start synchronous defibrillation. If Remote Sync is switched on,
press the [Sync On] soft key and select [Local] to start
synchronous defibrillation 6.Select [Pads] or [Paddles] as the ECG
source and begin charging. 4-7 7.When charging finishes, press and
hold the Shock button to deliver a shock. 8.Verify that synchronous
discharge succeeds and the delivery energy measured by the analyzer
is 10J2J. 9.Verify that the delay time of synchronous
defibrillation measured by the analyzer is less than 60ms.
10.Verify that the synchronous discharge mark appears on the R
wave. 11.Verify that the prompt messages are correct during
testing. 12.Select lead II as ECG source and perform charging.
Repeat steps 7 to 11. 13.Use multifunctional electrode pads. Repeat
steps 2 to 12. 4.6.2 Pacing Test Test tools: Defibrillator/pacer
analyzer 1.Run the equipment on fully charged battery. Move the
Mode Select knob to Pacer. Set [Pacer Mode] to [Fixed]. 2.Connect
the pads cable to the equipment and properly place the pads on the
defibrillator/pacer analyzer. 3.Set the analyzer to Pacing
Measurement mode. Use test load of 50. 4.On the equipment, set
[Pacer rate] to [70ppm] and [Pacer Output] to [30mA]. 5.Press the
[Start Pacing] soft key. Verify that the pacer rate measured by the
analyzer is 70 ppm1ppm and the pacer output measured is 30 mA5mA.
6.Press the Stop Pacing soft key, and then set [Pacer rate] to
[170ppm] and [Pacer Output] to [200mA]. 7.Press the [Start Pacing]
soft key. Verify that the pacer rate measured by the analyzer is
170 ppm2ppm, and the measured current is 200 mA10mA. 4-8 4.6.3 ECG
Test Performance Test Test tools ECG simulator 1.Connect the
simulator to the equipments ECG connector with ECG leadwires. 2.Set
the patient simulator as follows: ECG sinus rhythm, HR=80 bpm with
the amplitude as 1mV. 3.Check the ECG waves are displayed correctly
without noise and the displayed HR value is within 80 1 bpm.
4.Disconnect the simulator from the equipments ECG connector.
Verify that ECG Lead Off alarm behaves correctly. 5.On the
equipment, set [Paced] to [Yes], the simulator is configured as
pace signals. Verify that pace signals are detected and pace pulse
marks are displayed. 6.Connect the simulator to the equipments
therapy module with pads. 7.Set the patient simulator as follows:
ECG sinus rhythm, HR=80 bpm with the amplitude as 1mV. 8.Check the
ECG waves are displayed correctly without noise and the displayed
HR value is within 80 1 bpm. 9.Disconnect the simulator from the
equipments therapy module. Verify that ECG Lead Off alarm behaves
correctly. 10.On the equipment, set [Paced] to [Yes], the simulator
is configured as pace signals. Verify that pace signals are
detected and pace pulse marks are displayed. ECG Calibration Tool
required: Vernier caliper 1.Connect the simulator to the equipments
ECG connector with ECG leadwires. 2.Select the ECG parameter area
to enter the [ECG Setup] menu. 3.Select [Others>>]
[Calibrate]. A waveform signals appear on the screen and the
message [ECG Calibrating] is displayed in the prompt information
area in the lower left corner of the screen. 4.Compare the
amplitude of the waveform with the wave scale. The difference
should be within 5%. If needed, you can also print out the waveform
and the wave scale. 4-9 5.After ECG calibration is completed,
select [Stop Calibrating]. 6.Connect the simulator to the
equipments therapy module with pads. 7.Repeat steps 35. 4.6.4 Resp
Test Test tools Resp Patient simulator 1.Connect the patient
simulator to the ECG connector on the module. 2.On the
defibrillator/monitor, select the Resp widow to enter the Resp
Setup menu. Set [Lead] to [II]. 3.Configure the simulator as
follows: set Lead to II, base impedance line to 1500 ; delta
impedance to 0.5 , and respiration rate to 40 rpm. 4.Check that
respiration waveform is not distorted and the displayed Resp value
does not exceed 402 rpm. 4.6.5 SpO2 Test Test tool Patient
simulator. 1.Connect the patient simulator to the equipments SpO2
connector. 2.Select the model and manufacturer of the SpO2 module
under test. Configure the parameter as SpO2 96% and PR 80 bmp.
3.The displayed SpO2 and PR values should be within the ranges
listed below SpO2 (%)PR (bmp) Mindray96% 2%803 Masimo96% 2%803
Nellcor MAX-A, MAX-N, MAX-P, MAX-I96% 2% 803 DS-100A, OXI-A/N,
OXI-P/I96% 3% 4.7 Electrical Safety Tests See A Electrical Safety
Inspection.. 4-10 4.8 Recorder Check Tools required: None. 1.Print
ECG waveforms. The recorder should print correctly and the printout
should be clear. 2.Simulate some recorder problems, such as out of
paper, paper jam, etc. the defibrillator/ monitor should give
corresponding prompt messages. After the problem is removed, the
recorder should be able to work correctly. 3.Switch automatic alarm
recording for each parameter ON and then set each parameters limit
outside set alarm limits. Corresponding alarm recordings should be
triggered when parameter alarms occur. 4.9 Factory Service 4.9.1
Password for Service Mode Accessing service mode is password
protected. The required password is set to 332888 before the
equipment leaves the factory. 4.9.2 Accessing Service Mode Menu To
access the factory service menu, Press the Main menu button on the
equipments front panel. Select [Others>>]
[Maintenance>>] [Service Mode>>] enter the required
passwords. The Service Mode-Main menu is shown below. 4-11 4.9.3
Calibrating/Zeroing Impedance Normally impedance calibration and
zeroing is unnecessary. However, you can perform impedance checking
after replacing the therapy module. 1.If not pre-connected, connect
the pads cable to the equipment. 2.Connect a test load of 300 ohms
to the pads cable. 3.Start the equipment and select the Monitor
mode. Press and hold the [Event] hardkey, and then press the [Lead
Select] hardkey on the front panel, the following screen appears.
4.Verify that the reading of RT Imped is between 3000450. NOTE If
300 ohms test load is not available, you can use a 50 ohms test
load to perform impedance checking. In this case, Verify that the
reading of RT Imped is between 50075. If the reading of RT Imped is
not correct, perform impedance calibration/zeroing. 1. Press the
Main menu button on the equipments front panel. Select
[Others>>] [Maintenance>>] [Service Mode>>] enter
the required passwords[Calibrate/Zero Impedance] to enter the
Calibrate/Zero Impedance screen. 2.Connect a test load of 0 ohm to
the pads cable; then select Zero. A message Zero Completed shall be
shown. If the message Zero Failed is displayed, check the
connection of pads cable. 3.Connect a test load of 100 ohms to the
pads cable; then select Calibrate. A message Calibration Completed
shall be shown. If the message Calibration Failed is displayed,
check the connection of pads cable. Replace the therapy module if
impedance calibration/zeroing fails. 4-12 4.9.4 Device Information
Press the Main menu button on the equipments front panel. Select
[Others>>] [Maintenance>>] [Service Mode>>] enter
the required passwords [Device Information]. In the Device
Information list, you can view the device information such as
software version, system status, and etc, as shown below. In the
Device Information screen, you can select [Export] to export error
codes and shock delivery data to a USB flash memory. 4.9.5 Checking
Failure Code Press the Main Menu button on the equipments front
panel. Select [Others>>] [Maintenance>>] [Service
Mode>>] enter the required passwords [Failure Code] to check
error codes. This helps the service personnel to identify failures.
Refer to 6.8 Error Codes for the description of each error code.
4-13 4.9.6 Inputting Serial Number Press the Main Menu button on
the equipments front panel. Select [Others>>]
[Maintenance>>] [Service Mode>>] enter the required
passwords [Input Serial Number] to input the equipments serial
number. After inputting the serial number, you can view it by
accessing Installation Mode and select [Version]. 4.9.7 Paddle Open
Circuit Display This [Paddle Open Circuit Display] switch is for
testing only. In normal operation, it should be set to [Off]]. 4-14
Test Report Customer name Customer address Servicing person
Servicing company Equipment under test (EUT) Model of EUT SN of EUT
Hardware version Software version Test equipmentModel/No.Effective
date of calibration 4-15 Test itemsTest records Test results
(Pass/Fail) Visual inspection The case, display screen, buttons,
knob, modules, power cord, and accessories have no obvious signs of
damage. The external connecting cables are not frayed and the
connector pins are not loose or bent. The external connectors are
not loose or their pins are not bent. The safety labels and data
plate are clearly legible. Power-on test The power-on test is
passed. The power indicator and alarm system work correctly and the
equipment start up properly. Performance test Manual Defibrillation
Test When the equipment runs on AC mains and external paddles are
used, the equipment can be properly charged and discharged; the
energy delivered meets the requirement for accuracy, and the shock
information is correctly recorded. When the equipment runs on fully
charged battery and external paddles are used, the equipment can be
properly charged and discharged; the energy delivered meets the
requirement for accuracy, and the shock information is correctly
recorded. When the equipment runs on AC mains and multifunctional
electrode pads are used, the equipment can be properly charged and
discharged; the energy delivered meets the requirement for
accuracy, and the shock information is correctly recorded. When the
equipment runs on fully charged battery and multifunctional
electrode pads are used, the equipment can be properly charged and
discharged; the energy delivered meets the requirement for
accuracy, and the shock information is correctly recorded. 4-16
Performance test Manual Defibrillation Test When external paddles
are used, the charge tone is correctly issued when the equipment is
being charged. The prompt «Charged Removed» is shown on the screen
and the charge done tone stops when the Disarm hotkey is pressed.
The equipment does not discharge externally. When [Time to Auto
Disarm] is set to [60s], the prompt «Charged Removed» is shown on
the screen and the charge done tone stops after 60 seconds at the
completion of charging. The equipment does not discharge
externally. When pads are used, the charge tone is correctly issued
when the equipment is being charged. The prompt «Charged Removed»
is shown on the screen and the charge done tone stops when the
Disarm hotkey is pressed. The equipment does not discharge
externally. When [Time to Auto Disarm] is set to [60s], the prompt
«Charged Removed» is shown on the screen and the charge done tone
stops after 60 seconds at the completion of charging. The equipment
does not discharge externally. When external paddles are used for
synchronous defibrillation and ECG source is paddles and lead II
respectively, the prompt is correct and a Sync mark appears above
each R wave. The delivered energy measured is 10J2J and the
synchronous shock is delivered within 60 ms of the peak of the
R-wave. When pads are used for synchronous defibrillation and ECG
source is paddles and lead II respectively, the prompt is correct
and a Sync mark appears above each R wave. The delivered energy
measured is 10J2J and the synchronous shock is delivered within 60
ms of the peak of the R-wave. Pacing Test When set [Pacer rate] to
[70ppm] and [Pacer Output] to [30mA], the pacer rate measured by
the analyzer is 70 ppm1ppm and the pacer output measured is 30
mA5mA. When set [Pacer rate] to [170ppm] and [Pacer Output] to
[200mA], the pacer rate measured by the analyzer is 170 ppm2ppm,
and the measured current is 200 mA10mA. 4-17 Performance test ECG
performance test ECG waves are displayed correctly without noise
and the HR value is within 801 bpm. ECG Lead Off alarm behaves
correctly. Paced signals are detected and pace pulse marks are
displayed when [Paced] is set to [Yes] The difference between the
amplitude of the ECG calibration square wave and that of the wave
scale is not greater than 5%. Resp test The Resp wave is not
distorted and the Resp value is within 402 rpm. SpO2 test The
displayed SpO2 and PR values should be within the specified ranges.
Electrical safety tests Refer to Appendix A Electrical Safety
Inspection. Recorder check The recorder can print ECG waves
correctly and the printout is clear. Set the recorder to some
problems such as out of paper, paper jam, etc. the equipment gives
corresponding prompt messages. After the problem is removed, the
recorder is able to work correctly. Automatic alarm recording for
each parameter functions correctly when parameter alarms occur.
Tested by: _________________________Date: ________________________
4-18 FOR YOUR NOTES 5-1 5 Hardware and Software Upgrade 5.1
Hardware Upgrade 5.1.1 Upgrade MPM module from ECG only to ECG plus
SPO2 You can upgrade MPM module from ECG only to any of the
following configuration: ECG plus Mindray SpO2: 801-0652-00036-00
MPM upgrade kit ECG plus Masimo SpO2: 801-0652-00037-00 MPM upgrade
kit ECG plus Nellcor SpO2: 801-0652-00038-00 MPM upgrade kit The
upgrade procedure is as follows: 1.Remove the MPM module assembly
and parameter panel assembly as described in. 7.3.5 Disassembling
the MPM Module Assembly and 7.3.6 Removing the Parameter Panel
Assembly. 2.Replace the old measurement module panel and MPM module
with those in the MPM upgrade kit. 3.Reassemble the equipment. If
the upgrade MPM module is equipped with Masimo SpO2, you need to
stick a Masimo label at the lower left corner of the front housing
and a No Implied License label below the measurement module panel,
as indicated in the following pictures. Masimo label No Implied
License label 5-2 If the upgraded MPM module is equipped with
Nellcor SpO2, you need to stick a Nellcor label at the lower left
corner of the front housing, as indicated in the following
pictures. After upgrade the MPM module, perform the tests described
in 4.6.3 ECG Test, 4.6.4 Resp Test, 4.6.5 SpO2 Test, 4.7Electrical
Safety Tests, and 4.8 Recorder Check. 5.1.2 Upgrade the Therapy
Module You can use 801-0652-00039-00 pacer function upgrade kit to
upgrade the therapy module so that equipment has pacing function.
After upgrading the therapy module, choose Mode label (with pacing
function) with the language you need. Follow this procedure to
upgrade the therapy module: 1.Remove therapy module as described in
7.3.7 Removing the Therapy Module. Be noted that you need not to
remove the parameter panel assembly and MPM module assembly. 2.Take
off the Mode Select knob. Peel off the Mode label, see the picture
below. 3.Replace the old therapy module using the one with pacing
function in the upgrade kit. 4.Replace the old Mode Select knob and
Mode label using new ones in the upgrade kit. 5.Reassemble the
equipment. Nellcor label 5-3 After upgrade the therapy module,
perform the tests described in 4.6.1 Manual Defibrillation Test,
4.6.2 Pacing Test, and 4.7 Electrical Safety Tests. 5.2 Software
Upgrade through a PC You can upgrade system software and module
software using Mindray Patient Monitor Software Upgrade Tool. You
can also view software upgrade log. Mindray Patient Monitor
Software Upgrade Tool can directly run on a PC. By connecting the
defibrillator/monitor to a PC via a crossover network cable, you
can upgrade the following software: Bootstrap System software
Language library BMP files (including screen icons, start-up
screens, standby screens) General configurations (including
passwords, company logo) System functional configuration FPGA
program Power module software MPM module software and therapy
module software Mode Select knob Mode label 5-4 5.2.1 Installing
Mindray Patient Monitor Software Upgrade Tool 1.Find the
installation programand double click it to start installation.
2.Select installation language. 3.Click [Ok] and the following
screen appears. Click [Next] to go to the next step. 4.Enter User
Name, Company Name and Serial Number. 5-5 5. Specify the
destination folder for installing this program. Then select [Next].
7.Select Program Folder. Then select [Next]. 8.Click [Finish] to
complete installation. 5-6 5.2.2 Software Upgrade Procedure
1.Connect the defibrillator/monitor to be upgraded with a PC. 2.Set
IP address to 77.77.XX.XX and subnet mask to 255.255.255.0. 3.Run
Mindray Patient Monitor Software Upgrade Tool on the PC and set
Machine to BneneHeart. 4.On the Mindray Patient Monitor Software
Upgrade Tool screen, select [Select Package] and select packages
you want to upgrade. Then select [Start]. 5.Simultaneously hold the
[Record] key and [Menu] key on the defibrillator/monitors front
panel, and then turn on the equipment. After software upgrade is
finished, turn off the equipment, and then disconnect AC mains and
remove the battery. If you do not disconnect the power supply after
software upgrade, the status indicator will flash and the beeper
will sound. For the details of software upgrade, refer to help and
instructions for use of Mindray Patient Monitor Software Upgrade
Tool. CAUTION Disconnect the equipment from the patient and make
sure the important data are saved before upgrade. Do not shut down
or power off the equipment when upgrading the boot program.
Otherwise, it may cause the equipment to break down. Program
upgrade should be performed by qualified service personnel only.
Crossover network cable shall be used if a PC is connected for
equipment upgrade. NOTE After upgrading the boot program,
re-upgrade the system program and other programs to ensure
compatibility. Make sure the version of the upgrade package is the
one that you desired. If you want to obtain the latest upgrade
package, contact Mindray Customer Service Department. 5-7 5.3
Software Upgrade through a USB Memory You can upgrade the
defibrillator/monitors system software and module software using a
USB memory. All software that can be upgraded through a PC can be
upgraded through a USB memory. 5.3.1 Precautions 1.Software upgrade
through USB memory is only applied to defibrillator/monitors with
BIOS version 2.0 or later. 2.If you are going to upgrade the
software through a PC, do not plug a USB memory to the
defibrillator/monitors USB connector. 3.Up to 15 upgrade files can
be placed to the UPGRADE_0651 folder. If more files are placed,
only the first 15 can be displayed and chosen. 4.Only one upgrade
file can be chosen at a time. Repeat steps 3 to 5 as described in
5.3.2 Software Upgrade Procedure to upgrade files one by one. 5.The
name of files in directory UPGRADE_0651 should be in English.
Otherwise the file name will be unreadable. 6.Both .pkg and .mpkg
files can be upgraded. Bootstrap cannot be included in the .mpkg
files. 7.The bootstrap of main control should be upgrade
separately. That is to say, it cannot be included in the upgrade
package. 5.3.2 Software Upgrade Procedure 1.Create a folder named
UPGRADE_0651 in the root directory of the USB memory. 2.Copy the
file updatecontrol.bin and upgrade files to directory UPGRADE_0651.
3.Turn off the defibrillator/monitor. Plug the USB memory to the
equipments USB connector. 4.Simultaneously hold the [Record] key
and [Menu] key on the defibrillator/monitors front panel, and then
turn on the equipment. 5.Follow the on-screen instructions to
upgrade the software. Choose a file to be upgraded by rotating the
navigation knob. Press down the navigation knob to confirm the
selection. You have to choose the upgrade file within 120 s. If
not, software upgrade fails. The system will upgrade automatically
when the upgrade file is selected and confirmed. 5-8 FOR YOUR NOTES
6-1 6 Troubleshooting 6.1 Overview In this chapter, the
defibrillator/monitor problems are listed along with possible
causes and recommended corrective actions. Refer to the tables to
check the defibrillator/monitor, identify and eliminate the
problems. The problems we list here are frequently arisen
difficulties and the actions we recommend can correct most
problems, but not all of them. For more information on
troubleshooting, contact our Customer Service Department. 6.2 Part
Replacement Printed circuit boards (PCBs), major parts and
components in the defibrillator/monitor are replaceable. Once you
isolate a defective PCB, follow the instructions in 0 Disassembly
and Repair to replace the PCB with a known good one and check that
the trouble disappears or the defibrillator/monitor passes all
performance tests. If the trouble remains, replace the PCB with the
original suspicious PCB and continue troubleshooting as directed in
this chapter. To obtain information on replacement parts or order
them, refer to 8 Parts. 6.3 Checking Defibrillator/Monitor Status
Some troubleshooting tasks may require you to identify the hardware
version and status of your defibrillator/ monitor. To check
equipment status, 1.Select [Main Menu] [Review >>][Event
Review >>]. Then you can view the information on system start
time, self check, etc. 2.You can also view the information on the
defibrillator/monitors current status by pressing the Menu key on
the equipments front panel, and then selecting
[Others>>][Maintenance >>][Service Mode>>] enter
the required password [Device Information>>]. 6-2 6.4
Checking Device Information Some troubleshooting may involve
software compatibility. Thus it requires you to know your
defibrillator/monitor configuration and software version. For
detailed information on version compatibility, please contact our
Customer Service Department. To identify your software version,
press the Menu key on the equipments front panel, and then select
[Others>>][Maintenance>>][Installation
Mode>>]enter the required password [Version]. In the Version
screen, you can view system software version and module software
version. You can also press the Menu key on the equipments front
panel, and then select
[Others>>][Maintenance>>][Service Mode>>] enter
the required password [Device Information>>] to check system
software version, module software version, and device status . 6.5
Checking Technical Alarm Before troubleshooting the
defibrillator/monitor, check for technical alarm message. If an
alarm message is presented, eliminate the technical alarm first.
For detailed information on technical alarm message, possible cause
and corrective action, refer to the defibrillator/monitors
Operation Manual. 6-3 6.6 Troubleshooting Guide 6.6.1
Defibrillation Problems SymptomPossible CauseCorrective Action The
equipment does not charge by pressing the Charge button on the
front panel. Keypad board failure.1. Connect external paddles.
Press the Charge button on the Apex paddle to start charging. If
the Charge button on the paddles works, it indicates the keypad
board fails. 2. Replace the keypad board. The Charge button fails
to be pressed down effectively due to the damaged or dislocated
silica gel keypad. Disassemble the keypad board to replace or
reshuffle the keypad. The equipment cannot be charged by pressing
the Charge button on the external paddles. Paddles not connected
properly.Reconnect the paddles. Paddles failure.1. Isolate the
problem by connecting pads to perform charging/discharging. 2. If
the paddles are failed, replace them. Failure of therapy
module.Replace the therapy module. Failure of connection wire to
the therapy module. Replace the connection wire. The equipment is
charged too slowly Battery failure.Replace the battery or connect
the equipment with external power supply. Power management board
failure.Replace the power management board. Therapy module
failed.Replace the therapy module. A shock cannot be delivered by
pressing the Shock button on the equipments front panel in Manual
Defib Mode or AED Mode. Keypad board failure.1. Locate the problem
by connecting paddles to perform charging / discharging. 2. If the
keypad board is defective, replace it. The Charge Button fails to
be pressed down effectively due to the failure or dislocated silica
gel keypad. Disassemble the keypad board to replace or reshuffle
the keypad. A shock cannot bePaddles not connected
properly.Reconnect the paddles. 6-4 SymptomPossible CauseCorrective
Action delivered by pressing the Shock button on the paddles.
Paddles failure.1. Locate the problem by connecting pads to perform
charging/discharging. If normal discharge can be performed, it
indicates the paddles are defective. 2. Replace the paddles.
Failure of therapy module.Replace the therapy module. Connection
wire to the therapy module broken. Replace the connection wire. The
message. Disarming Failed is displayed. Failure of therapy
module.Replace the therapy module. The equipment can be properly
charged, but the energy is disarmed automatically at the completion
of charging or when the equipment is being discharged. Too high or
too low patient impedance detected. 1. Pads/paddles are detached
from the patient. 2. Pads/paddles failure. 3. Pads cable failure.
Ensure good connection between the patient and pads/paddles. If the
problem persists, replace Pads, paddles or Pads cable. Failure of
therapy module.Replace the therapy module. Defibrillation
malfunction. Defibrillation hardware circuit defective. Replace the
therapy module. Energy Select buttons on the equipment front panel
do not work. Keypad board failure.1. Locate the problem by
connecting paddles to perform energy setting. 2. If the keypad
board is defective, replace it. The Energy Select Buttons fail to
be pressed down due to the damaged or dislocated silica gel keypad.
Disassemble the keypad board to replace or reshuffle the keypad.
Energy Select buttons on the paddles do not work Paddles not
connected properly.Reconnect the paddles. Paddles failure.Replace
the paddles. Failure of therapy module.Replace the therapy module.
Connection wires to the therapy module broken. Replace the
connection wires. 6-5 6.6.2 Pacing Problems SymptomPossible
CauseCorrective Action Does not deliver correct pacing current.
Failure of the therapy moduleReplace the therapy module. Does not
deliver correct pacing rate. Failure of the therapy module.Replace
the therapy module. Pacer Equip Malfunction Pacer hardware
failure.Replace the therapy module. 6.6.3 Power On/Off Problems
SymptomPossible CauseCorrective Action The defibrillator/monitor
fails to start. AC LED or battery LED does not light AC mains not
connected or battery too low. Check that AC mains is properly
connected or battery capacity is sufficient. Power supply
protection.Refer to 6.6.10 Power Supply Problems. Cables defective
or poorly connected. 1. Check that the cables between the power
switch and the keypad board, the keypad board and the power
management board and between the power module and the power
management board are correctly connected. 2. Check that wires and
connectors are not defective. Power switch or keypad board failure
Replace the power switch or keypad board. AC/DC board
defectiveReplace the AC/DC board. Power management board failure
Replace the power management board. 6-6 6.6.4 Display Problems
SymptomPossible CauseCorrective Action The LCD screen is blank, but
the defibrillator/monitor works properly. Connection cable
defective or poorly connected. 1. Check that wires between the
display and the keypad board, and between the keypad board and the
power management board are correctly connected. 2. Check that the
cables and connectors are not defective. LCD Display failureReplace
the display. Keypad board failureReplace the keypad board. CPU
board failure.Replace the CPU board. Power management board
failure. Replace the power management board. Secondary display does
not function. Cable failure1. Check that the wires between the
display and the defibrillator/monitor are correctly connected. 2.
Check that the cables and connectors are not defective. CPU board
failure.Replace the CPU board. Power management board failure.
Replace the power management board. Secondary display displays
snows or flashing specks Connection cable defective or poorly
connected. 1. Check that the wires between the display and the
defibrillator/ monitor are correctly connected. 2. Check that the
cables and connectors are not defective. CPU board failure.Replace
the CPU board. Power management board failure. Replace the power
management board. Images overlapped or distorted FPGA failureUpdate
or upgrade FPGA. Connection cable defective or poorly connected. 1.
Check that the wire between the display and keypad board is
correctly connected. 2. Check that the cables and connectors are
not defective. 6-7 SymptomPossible CauseCorrective Action The
colour of images deviates from the standard configuration.
Connection cable defective or poorly connected. 1. Check that wires
between the display and the keypad board and between the keypad
board and the power management board are correctly connected. 2.
Check that the cables and connectors are not defective. Display
failureReplace the display. Keypad board failureReplace the keypad
board. CPU board failure.Replace the CPU board. Power management
board failure. Replace the power management board. 6.6.5 Alarm
Problems SymptomPossible CauseCorrective Action The alarm lamp is
not light or extinguished but alarm sound is issued Connection
cable defective or poorly connected. 1. Check that wire between
alarm LED board and keypad board are properly connected. 2. Check
that connection wires and connectors are not defective. Alarm LED
board failureReplace the alarm LED board. Keypad board
failureReplace the keypad board. Power management board failure.
Replace the power management board. No alarm sound is issued but
alarm lamp is lit Audio alarm disabledCheck that alarm tone volume
is set to a value other than zero by pressing the Menu key on the
equipments front panel, and then selecting [Alarm Setup >>].
Connection cable defective or poorly connected. 1. Check that the
wire between the speaker and keypad board is properly connected. 2.
Check that connection wires and connectors are not defective. FPGA
audio logic ERRORUpgrade the audio logic part of the FPGA program.
Speaker failureReplace the speaker. Power management board failure.
Replace the power management board. 6-8 6.6.6 Button and Knob
Problems SymptomPossible CauseCorrective Action Buttons do not
respond. Connection cable defective or poorly connected. 1. .Check
that the wire between the keypad and the keypad board is properly
connected. 2. Check that that the wire between the keypad board and
the power management board is properly connected. 3. Check if the
connection wires and connectors are defective. Keypad board failure
Replace the keypad board. CPU board failure.Replace the CPU board.
Power management board failure. Replace the power management board.
Mode Select knob does not respond. Connection cable defective or
poorly connected. 1. Check that wires between the knob to keypad
board, and between the keypad board and the power management board
are properly connected. 2. Check that connecting wires and
connectors are not defective. Knob failureReplace Mode Select knob.
Keypad board failure Replace the keypad board. CPU board
failure.Replace the CPU board. Power management board failure.
Replace the power management board. Navigation knob does not
respond. Connection cable defective or poorly connected. 1. Check
that wires between the knob to keypad board, and between the keypad
board and the power management board are properly connected. 2.
Check that connection wires and connectors are not defective. Knob
failure.Replace the navigation knob. Keypad board failure Replace
the keypad board. CPU board failure.Replace the CPU board. Power
management board failure. Replace the power management board. 6-9
6.6.7 Recorder Problems SymptomPossible CauseCorrective Action No
printoutConnection cable defective or poorly connected. 1. Check
the wire between the recorder and the keypad board is connected
properly. 2. Check that connection wires and connectors are not
defective. Recorder power supply failure Check if the power module
outputs 5 V DC and 12V DC correctly. Recorder failureReplace the
recorder. Poor print quality or paper not feeding properly Paper
roll not properly installed Stop the recorder and re-install the
paper roll. Print head dirty1. Check the thermal print head and the
paper roller for foreign matter. 2. lean the thermal print head
with an appropriate clean solution. Recorder failureReplace
recorder. Blank printoutPaper-roll installed reversely. Reload
paper-roll. Recorder failureReplace the recorder. 6.6.8 Output
Interface Problems SymptomPossible CauseCorrective Action Sync
input failureCPU board failure. Power management board failure.
Replace the CPU board. Replace the power management board. USB
Device does not function (provided that the peripheral devices are
good) CPU board failure.Replace the CPU board. Power management
board failure. Replace the power management board. 6-10 6.6.9 CF
Card Problems SymptomPossible CauseCorrective Action CF card
malfunctionsWrong CF card or limited memory space Use only
INNODISK-manufactured CF storage cards. Those with minimum 1GB
memory space are recommended. CF card defectiveInsert a known good
CF card into the defibrillator/monitor. If it works normally, the
original CF card fails. CF card slot failureReplace the CPU board.
CPU board failure.Replace the CPU board. CF card failureFormat the
CF card. 6.6.10 Power Supply Problems SymptomPossible
CauseCorrective Action Battery failureBattery damaged.Replace
battery. Battery interface failure. 1. Check batteries are
installed properly. 2. Check if the battery interface is defective.
3. If the battery interface is defective, replace the power
management board. Power management board failure. Replace the power
management board. Batteries can not be fully charged. Battery
damaged.Replace batteries. Battery interface failure. 1. Check
batteries are installed properly. 2. Check if the battery interface
is defective. 3. If the battery interface is defective, replace the
power management board. Power management board failure. Replace the
power management board. Battery cannot be charged Battery
failure.Replace battery and recharge the replacement battery. If
the replacement battery can be recharged, the original one fails.
Cable defective or poorly connected 1. Check batteries are
installed properly. 2. Check if the battery interface is defective.
3. If the battery interface is defective, replace the power
management board. Power management board failure. Replace the power
management board. 6-11 SymptomPossible CauseCorrective Action No
+3.3 V A output1. Power supply failure 2. Power management board
failure. 1. Turn off the defibrillator/monitor then restart it. 2.
If the problem remains, disconnect the AC mains for 5 s and
reconnect it, then restart the defibrillator/ monitor. 3. If the
problem still remains, replace the power management board. No +3.3
V B output Not +5.0 V output No +12 V output NOTE When the power
module has a failure, it may cause problems to other components,
e.g. the defibrillator/monitor suddenly breaks down during
start-up, as the power module may have a power supply protection.
In this case, troubleshoot the power module per the procedure
described in the table above. Components of the main unit, SMR and
parameter modules are powered by the power module. In the event
that a component malfunctions, check if the operating voltage is
correct. Refer to chapter 2 Theory of Operation for the operating
voltage and measurement points of each component. 6.6.11 Software
Upgrade Problems SymptomPossible CauseCorrective Action Boot file
upgrade fails Power failure or unintended power off during boot
file upgrade Return the CPU board to factory for repair. Program
upgrade failsIncorrect network connection 1. Check that the network
cable is properly connected and is not too long (shorter than 50m).
2. Make sure that the network cable is of the right type. Network
cable with crossed wires inside is used for LAN upgrade and those
with parallel wires inside for WAN. Wrong upgrade packageUpgrade
package shall be .pkg files. Select package according to system
requirement. Incorrect IP address configuration Configure a fixed
IP address in range C as specified for the defibrillator/monitor.
We recommend not to upgrade a program when the
defibrillator/monitor is connected to a network with multiple PCs.
6-12 6.7 Technical Alarm Messages MeasurementAlarm MessageCause and
solution XXXX SelfTest ErrAn error occurred to the XX module, or
there is a problem with the communications between the module and
the host. Restart the equipment. XX Init Err XX Comm Err XX Comm
Stop XX OverrangeThe measured XX value is not within the specified
range for XX measurement. Check input signal and equipment setting.
ECGECG Lead OffThe ECG electrode has become detached from the
patient or the lead wire has become disconnected from the trunk
cable. Check the connection of the electrodes and leadwires. ECG YY
Lead Off (YY represents the leadwires V, LL, LA, and RA, as per AHA
standard, or C, F, L and R as per IEC standard.) Pads/Paddles
offThe pads/paddles have been detached from the patient or the
therapy cable is loose. Check that the pads/paddles and therapy
cable are properly connected. ECG NoiseThe ECG signal is noisy.
Check for any possible sources of signal noise form the area around
the cable and electrode, and check the patient for excessive
motion. ECG Signal Invalid ECG amplitude is so low that ECG signal
is undetectable. Check for any possible source of interference from
the area around the cable and electrode; check the patients
condition. SpO2SpO2 Sensor OffThe SpO2 sensor has become detached
from the patient or the module, or there is a fault with the SpO2
sensor, or an unspecified SpO2 sensor has been used. Check the
sensor application site and the sensor type, and make sure the
sensor is not damaged. Reconnect the sensor or use a new sensor.
SpO2 Sensor Fault SpO2 No Sensor SpO2 Unknow Sensor SpO2 Sensor
Incompatible SpO2 Too Much Light There is too much light on the
SpO2 sensor. Move the sensor to a place with lower level of ambient
light or cover the sensor to minimize the ambient light. 6-13
MeasurementAlarm MessageCause and solution SpO2 Low SignalThe SpO2
signal is too low or too weak. Check the patients condition and
change the sensor application site. If the error persists, replace
the sensor. SpO2 Weak Signal SpO2 Weak Pulse SpO2 Low Perf SpO2
InterferenceThe SpO2 signal has been interfered. Check for any
possible sources of signal noise form the area around the sensor,
and check the patient for excessive motion. SpO2 Non-PulsatileSpO2
Board FaultThere is a problem with the SpO2 measurement board.
Replace the SpO2 module. Main control system No FanMake sure that
the fan is connected. No SpeakerMake sure that the speaker is
connected. No Data CardMake sure that the CF card is in place or
format the CF card. Power Board Comm Err An error occurred to the
power board, or there is a problem with the communications between
the power board and the host. Restart the equipment. Keyboard Comm
Err An error occurred to the keypad board, or there is a problem
with the communications between the keypad board and the host.
Restart the equipment. Therapy Module Comm Err An error occurred to
the therapy module, or there is a problem with the communications
between the therapy module and the host. Restart the equipment. If
the problem persists, replace the therapy module. Main Control
Selftest Err The main control voltage is abnormal. Replace the main
control board. RT Clock Need Reset Reset system time. RT Clock
ErrAn error occurred to the RTC chip, or the button cell is
depleted. Replace corresponding part. Data Card ErrThere is a
problem with the data card. Check or replace the data card if
necessary. Power boardPower System Selftest Err An error occurred
to the system power supply. Restart the equipment. Power Board Volt
Err Low BatteryChange battery or connect the equipment to the AC
power source to charge the batteries. 6-14 MeasurementAlarm
MessageCause and solution Battery ErrThere is a problem with the
batteries. Check the batteries for damage; verify that correct
batteries are used. Replace the batteries if necessary. Battery
Depleted! System will shut shown imminently. Connect to AC Mains or
Replace Battery. Connect the equipment to AC mains. Battery
AgedReplace the battery. Battery failed charging Battery failure or
power board hardware failure. Replace the battery. If the problem
persists, replace the power board. Therapy module Therapy Equip
selftest Err An error occurred during therapy module self test.
Restart the equipment or replace the therapy module. Defib
MalfunctionThe defibrillation function fails or both the
defibrillation and pacing functions fail. Restart the equipment and
test defibrillation function. If the problem persists, replace the
therapy module. Pacer Malfunction!The pacing function fails.
Restart the equipment and test pacer function. If the problem
persists, replace the therapy module. Disarming FailedThere is a
problem with the therapy module disarming circuit. Replace the
therapy module. Last User Test Failed Run a succesful user test.
Last Auto Test Failed Run a succesful user test again. Monitoring
module Mornitor Module Selftest Err An error occurred during MPM
module power-on self test. Replace the MPM module. Mornitor Module
Reset Err MPM module reset abnormally. In this case, the MPM module
restores to default configuration. You can ignore this problem.
Mornitor Module Voltage Err The voltage of MPM module is abnormal.
Replace the MPM module. RecorderRecorder Init ErrRestart the
equipment. Recordhead Overheated The recorder has been working for
a prolonged time. Clear the recording tasks and resume the
recording till the recorders print head cools down. 6-15
MeasurementAlarm MessageCause and solution Recorder Overcurrent
Re-load the recorder paper. OthersConfig ErrCheck if the
configuration is correct, or restore the factory configuration.
PacerPads cable OffCheck that pads cable is properly connected.
Pads OffCheck that pads are properly connected. ECG Lead OffCheck
that ECG leadwires are properly connected. Pacer Stopped Abnormally
Check paddles. Check that pads well contact with patients skin.
Make sure pads are properly applied, and then start pacing again.
6.8 Error Codes 6.8.1 Therapy Module Error Codes Error codesError
description 00No error 01MCU register selftest error 02MCU RAM
selftest error 03MCU FLASH selftest error 04MCU watchdog selftest
error 05MCU ADC selftest error 06DSP register selftest error 07DSP
RAM selftest error 08DSP FLASH selftest error 09DSP watchdog
selftest error 10ECG hardware selftest error 11Power selftest error
12MCU abnormal reset 13DSP abnormal reset 14Computation selftest
error 15 to 20Reserved 21Impedance circuit error 22Impedance
zeroing failed 23Impedance calibration failed 6-16 Error codesError
description 24Impedance measurement is interfered 25Great
measurement difference between two impedance measuring methods 26
to 30Reserved 31Charging is not complete after 25 s 32Charging is
not complete after 30 s 33Over-charging 34Charging trend error 35
to 40Reserved 41Self-discharging failure 42IGBT failure
43Defibrillation relay failure 44Abnormal discharging current 45 to
70Reserved 71Abnormal pacer power 72Pacer relay failure 73Pacer
frequency error 74Abnormal pacer current 75Pacer DA error 76Pacer
voltage error 77Pacer overcurrent protection point high 78Pacer
overcurrent protection failure 79 to 90Reserved 91Comm is
interfered (data error) 92 to 99Reserved 6-17 6.8.2 Power Module
Error Codes Error codesError contents 100/register valueBattery
permanent failure 101/register valueBattery recoverable failure
102Battery comm error 103Abnormal battery voltage (not enough
voltage) 104Battery charging failure 105Power-on selftest error
106Main control failure (comm handshake overtime) 107 to
199Reserved 6.8.3 Main Control Error Codes Error codesError
contents 200Fan not connected 201Speaker not connected 20

Дефибриллятор Mindray BeneHeart D3

Чтобы узнать актуальные цены, необходимо зарегистрироваться на сайте

Регистрация

Указанная стоимость является информационной, не является публичной офертой, а также не может служить в качестве предложения для обоснования НМЦК для процедур закупки в рамках 44 и 224ФЗ

BeneHeart D3

Компактный и легкий дефибриллятор / монитор со встроенными функциями ручной дефибрилляции, автоматической наружной дефибрилляции, кардиостимуляции и мониторинга основных показателей жизнедеятельности. Быстрая зарядка и подача разряда на заданном уровне энергии помогут в проведении эффективной реанимации.

BeneHeart D3 предлагает медицинскому персоналу функцию мониторинга основных показателей жизнедеятельности пациента, включая ЭКГ, дыхание,НИАД и дополнительные данные о концентрации SpO2, позволяя осуществлять непрерывный мониторинг состояния пациента.

Благодаря степени защиты от губительного проникновения пыли и воды IP44, дефибриллятор / монитор D3 обладает надежными рабочими характеристиками и может использоваться в самых разных условиях. В режиме АНД BeneHeart D3 автоматически анализирует частоту сердечных сокращений и определяет целесообразность подачи электрического импульса. Голосовые и текстовые подсказки помогают пользователю в эксплуатации оборудования. Для анализа работы персонала и проверки имеется функция записи речи (180 минут).

Шнур питания

Шнур питания — EU (Европейский)


Состав комплекта

Шнур питания — EU (Европейский)

999

Дефибриллятор-монитор BeneHeart D3 с дисплеем 7 дюймов, о встроенным аккумулятором, термопринтером, функцией мониторинга ЭКГ 3/5 отведений

999

Дефибриллятор-монитор BeneHeart D3 с дисплеем 7 дюймов, со встроенным аккумулятором, термопринтером, функцией мониторинга ЭКГ 3/5 отведений, функцией кардиостимуляции

999

Дефибриллятор-монитор BeneHeart D3 с дисплеем 7 дюймов, со встроенным аккумулятором, термопринтером, функцией мониторинга ЭКГ 3/5 отведений, SpO2, функцией кардиостимуляции

999

Дефибриллятор-монитор BeneHeart D3 с дисплеем 7 дюймов, со встроенным аккумулятором, термопринтером, функцией мониторинга ЭКГ 3/5 отведений, SpO2, НИАД, функцией кардиостимуляции

999

Дефибриллятор-монитор BeneHeart D3 со встроенным аккумулятором, термопринтером, Wi-Fi картой, функцией мониторинга ЭКГ 3/5 отведений, SpO2, НИАД, функцией кардиостимуляции

999

Внешние электроды для дефибрилляции

Набор внешних электродов для дефибрилляции, взр/дет

999

Набор внешних многофункцональных электродов, одноразовый, взр/дет (для пациентов более 25 кг)

999

Набор внешних многофункцональных электродов, одноразовый, дет (для пациентов менее 25 кг)

999

Набор для ЭКГ в составе:
1. Магистральный кабель ЭКГ: 3/5 отведений (взр/дет), с защитой от дефибрилляции
2. Провод ЭКГ на 5 отведений, взрослый, тип защелка 3. ЭКГ электроды: взр, 5 шт, INTCO

999

Набор для ЭКГ в составе:
1. Магистральный кабель ЭКГ: 3/5 отведений (взр/дет), с защитой от дефибрилляции
2. Провод ЭКГ на 3 отведения, взрослый, тип защелка 3. ЭКГ электроды: взр, 5 шт, INTCO

999

Набор для ЭКГ в составе:
1. Магистральный кабель ЭКГ: 3/5 отведений (взр/дет), с защитой от дефибрилляции
2. Провод ЭКГ на 3 отведения, детский, тип защелка
3. ЭКГ электроды:дет, 10 шт, Kendall

999

Набор для ЭКГ в составе:
1. Магистральный кабель ЭКГ: 3/5 отведений (взр/дет)
2. Провод ЭКГ на 5 отведений, взрослый, тип зажим
3. ЭКГ электроды: взр, 5 шт, INTCO

999

Набор для ЭКГ в составе:
1. Магистральный кабель ЭКГ: 3/5 отведений (взр/дет), с защитой от дефибрилляции
2. Провод ЭКГ на 3 отведения, взрослый, тип зажим
3. ЭКГ электроды: взр, 5 шт, INTCO

999

Набор для ЭКГ в составе:
1. Магистральный кабель ЭКГ: 3/5 отведений (взр/дет), с защитой от дефибрилляции
2. Провод ЭКГ на 3 отведения, детский, тип зажим
3. ЭКГ электроды: дет, 10 шт, Kendall

999

Набор для ЭКГ в составе:
1. Магистральный кабель ЭКГ: 3/5 отведений (дет/нео), с защитой от дефибрилляции
2. Провод ЭКГ на 3 отведения, неонатальный, тип зажим 3. ЭКГ электроды: нео, 10 шт, Kendall

999

Набор для Mindray SpO2, в составе: 1. датчик пульсоксиметрический, взрослый, напалечный, 512F (для пациентов более 30 кг) 2. кабель SpO2 (2,5м)

999

Набор для Mindray SpO2, в составе: 1. датчик пульсоксиметрический, детский, напалечный, 512F (для пациентов 10-30 кг) 2. кабель SpO2 (2,5м)

999

Набор для Mindray SpO2, в составе: 1. датчик пульсоксиметрический, для новорожденных (для пациентов <5 кг) 2. кабель SpO2 (2,5м)

999

Набор для Mindray SpO2, в составе: 1. датчик пульсоксиметрический, для взрослых, напалечный DS100A (для пациентов >40 кг) 2. кабель SpO2 (2,5м)

999

Набор для Mindray SpO2, в составе: 1. датчик пульсоксиметрический, для взрослых, напалечный (указательный/большой палец) DS100A (для пациентов 3-40 кг) 2. кабель SpO2 (2,5м)

999

Набор для Mindray SpO2, в составе: 1. датчик пульсоксиметрический, неонатальный OXI-A/N (для пациентов <3кг) 2. кабель SpO2 (2,5м)

999

Набор для НИАД В соответствии с регистрационным удостоверением количество на 1 основную единицу не указано

Набор для НИАД в составе: 1. Трубка для неинвазивного измерения давления 2. Манжета CM1203 , взр (25-35 см)

999

Набор для НИАД в составе: 1. Трубка для неинвазивного измерения давления 2. Манжета CM1202, дет (18-26 см)

999

Набор для НИАД в составе: 1. Трубка для неинвазивного измерения давления 2. Манжета CM1500C, нео, одноразовая (5.8-10.9 см) — 1 шт.

999

Дополнительное отделение сумки для транспортировки

Заднее дополнительное отделение сумки для транспортировки.
Примечание: Крючок на поручень кровати и заднее отделение не могут быть установлены одновременно

999

Сумка для транспортировки

Сумка для транспортировки c защитным покрытием

999

Программное обеспечение на CD

999

Литий-ионный аккумулятор, 15.1 В, 5600 мАч

999

Внешнее зарядное устройство + шнур питания

Внешнее зарядное устройство + шнур питания

999

Кабель п для внешнего зарядного устройства

Кабель постоянного тока для внешнего зарядного устройства, разъем для прикуривателя

999

Крючок на поручень кровати

Крючок на поручень кровати

999

Крепление настольное/транспортное

Крепление настольное/транспортное

999

Панель для крепления внешнего зарядного устройства

Панель для крепления внешнего зарядного устройства

999

Верхнее дополнительное отделение

999

Заднее дополнительное отделение

999

Сумка для D3 c защитным экраном

999

Мультифункциональные электроды

Мультифункциональные электроды для дефибрилляции, взр, одноразовые (1 пара)

999

Мультифункциональные электроды для дефибрилляции, взр, одноразовые (5 пар 115-040517-00 /уп)

999

Мультифункциональные электроды для дефибрилляции, дет, одноразовые (5 пар 115-040518-00/уп)

999

Набор внешних электродов для дефибрилляции, взр/дет

999

Набор внешних электродов для кардиостимуляции, АНД, одноразовый, взр

999

Набор внешних электродов для кардиостимуляции, АНД, одноразовый, дет

999

Датчик SpO2, многоразовый, взр, напалечный, 512F

999

Датчик SpO2, многоразовый, дет, напалечный, мягкий, 512G

999

Датчик SpO2, многоразовый, дет, напалечный, 512H

999

Датчик SpO2, многоразовый, нео, на ногу (Взр/дет, напалечный), 518B

999

Датчик SpO2, для одного пациента, дет, 10 — 50 кг, 520P, 20 шт/уп

999

Датчик SpO2, для одного пациента, дет. мал., 3 — 20 кг, 520I, 20 шт/уп

999

Датчик SpO2, для одного пациента, нео, <3 кг, 520N, 20 шт/уп

999

Кабель для мультифункциональных эелектродов с тестовой нагрузкой (50 Ом)

999

Кабель удлинительный, Mindray SpO2, 7 Pin, 2.5м

999

Кабель Mindray SpO2, 7 Pin, 1.2 м

999

Кабель удлинительный, Nellcor SpO2, 8 Pin

999

Кабель вывода аналогового сигнала

999

Кабель синхронизации с дефибриллятором

999

Тестовая нагрузка (используется с 040-000545-00)

999

Электроды для дефибрилляции на открытом сердце

Электроды для дефибрилляции на открытом сердце, многоразовые, 1 дюйм

999

Электроды для дефибрилляции на открытом сердце, многоразовые, 2 дюйма

999

Электроды для дефибрилляции на открытом сердце, многоразовые, 3 дюйма

999

Кабель ЭКГ с электродами: взр, 5 отведений, Snap, IEC, защита от дефибрилляции

999

Кабель ЭКГ с электродами: взр, 5 отведений, Clip, IEC, защита от дефибрилляции

999

Кабель ЭКГ с электродами: взр/дет, 3 отведения, Snap, IEC, защита от дефибрилляции

999

Кабель ЭКГ с электродами: взр, 3 отведения, Clip, IEC, защита от дефибрилляции

999

Кабель ЭКГ с электродами: дет, 3 отведения, Snap, IEC, защита от дефибрилляции

999

Кабель ЭКГ с электродами: дет, 3 отведения, Clip, IEC, защита от дефибрилляции

999

Кабель ЭКГ с электродами: нео, 3 отведения, Clip, IEC, защита от дефибрилляции

999

Датчик SpO2, многоразовый, дет/мал. дет, OXI-P/I

999

Датчик SpO2, многоразовый, взр/нео, с 50 фиксаторами

999

Датчик SpO2, взр/дет/нео/мал. дет., Dura-Y Multi-Site

999

Датчик SpO2, одноразовый, взр, >30kg, MAX-A, 24 шт/уп

999

Датчик SpO2, одноразовый, дет, 10 — 50 кг, MAX-P , 24 шт/уп

999

Датчик SpO2, одноразовый, дет. мал., 3 — 20 кг, MAX-I , 24 шт/уп

999

Датчик SpO2, одноразовый, взр/нео, >40 кг или <3 кг, MAX-N , 24 шт/уп

999

ЭКГ электроды: взр, 10 шт/уп, Kendall

999

ЭКГ электроды: нео, 50 шт/уп, Coviden

999

Магистральный кабель ЭКГ: 3 отведения, нео, 12 Pin, TPU, AHA/IEC, DIN

999

Магистральный кабель ЭКГ: 3/5 отведений, взр/дет, 12 Pin, защита от дефибрилляции, AHA/IEC

999

Магистральный кабель ЭКГ: 3 отведения, дет/нео, 12 Pin, защита от дефибрилляции, AHA/IEC

999

3-отведения ЭКГ, кнопка, взр/дет, 1 м

999

3-отведения ЭКГ, зажим, взр/дет, 1 м

999

3-отведения ЭКГ, зажим, нео, 1 м

999

5-отведений ЭКГ, кнопка, взр/дет, 1 м/1.4 м

999

5-отведений ЭКГ, зажим, взр/дет (длинный, 1 м/1.4 м)

999

5-отведений ЭКГ, зажим, взр/дет, 0,6 м/1 м

999

Датчик SpO2, многоразовый, взр, напалечный, мягкий, 512E

999

Датчик SpO2, для одного пациента, взр, >30кг, 520A, 20 шт/уп

999

Датчик SpO2, многоразовый, взр, DS-100A

999

Фиксаторы для датчика SpO2, 518C (12шт/уп)

999

Трубка + манжета, взр (25-35 см)

999

Трубка + манжета, взр (25-35 см)

999

Трубка + манжета, дет (18-26 см)

999

Трубка + манжета, дет (18-26 см)

999

Трубка + манжета, нео, одноразовая (5.8-10.9 см)

999

Трубка для НИАД, взр/дет/мал.дет., 3 м

999

Трубка для НИАД, нео, 3 м

999

Манжета многоразовая, мал. дет., с коннектором, 7-13 см

999

Манжета многоразовая, мал. дет., с коннектором, 10-19 см, CM1201

999

Манжета многоразовая, дет, с коннектором, 18-26 см, CM1202

999

Манжета многоразовая, взр, с коннектором, 25-35 см, CM1203

999

Манжета многоразовая, бол. взр, с коннектором, 33-47 см, CM1204

999

Манжета многоразовая, набедренная, с коннектором, 46-66 см, CM1205

999

Манжета многоразовая, дет, бескамерная, с коннектором, 18-26 см, CM1302

999

Манжета многоразовая, взр, бескамерная, с коннектором, 25-35 см, CM1303

999

Манжета многоразовая, бол.взр, бескамерная, с коннектором, 33-47 см, CM1304

999

Манжета многоразовая, набедренная, бескамерная, с коннектором, 46-66 см, CM1305

999

Манжета многоразовая, взр, длинная, бескамерная, с коннектором, 25-35 см, CM1306

999

Манжета многоразовая, бол. взр, длинная, бескамерная, с коннектором, 33-47 см, CM1307

999

Манжета одноразовая, дет, 18-26 см, CM1502, 10 шт/уп

999

Манжета одноразовая, взр, 25-35 см, CM1503, 10 шт/уп

999

Манжета одноразовая, набедренная, 46-66 см, CM1505, 10 шт/уп

999

Манжета одноразовая, длинная взр, бескамерная, 25-35 см, CM1506, 10 шт/уп

999

Манжета одноразовая, длинная взр, бескамерная, 33-47 см, CM1507, 10 шт/уп

999

Манжета одноразовая, нео, 3.1-5.7 см, CM1500A, 20 шт/уп

999

Манжета одноразовая, нео, 4.3-8.0 см, CM1500B, 20 шт/уп

999

Манжета одноразовая, нео, 5.8-10.9 см, CM1500C, 20 шт/уп

999

Манжета одноразовая, нео, 7.1-13.1 см, CM1500D, 20 шт/уп

999

Манжета одноразовая, нео, 8-15 см, CM1500Е, 20 шт/уп

999

Термобумага (50 мм х 20 м)

999

Адаптер для постоянного/переменного тока

Адаптер постоянного/переменного тока (DC/AC)

999

Оставить отзыв

У данного товара нет отзывов. Станьте первым, кто оставил отзыв об этом товаре!

p7-s1-web

BeneHeart D3 — это компактный, надежный и легкий дефибриллятор со встроенными функциями мониторинга, ручной дефибрилляции, автоматической наружной дефибрилляции и кардиостимуляции. Этот профессиональный бифазный дефибриллятор-монитор подходит для использования в медицинских учреждениях во всем мире.

Характеристики

Быстрое проведение реанимационных мероприятий

· Включение питания за 2 секунды

· Зарядка до 200 Дж за 3 секунды

· Восстановление ЭКГ всего за 2,5 секунды

Технология 360 Дж BTe

В дефибрилляторах BeneHeart D3 используется двухфазная технология 360 Дж с автоматической компенсацией импеданса, что повышает шансы на спасение пациентов со сложными случаями дефибрилляции.

Простота в использовании

Следуя интуитивно понятным последовательным шагам (1) выбор уровня мощности; 2) заряд; 3) разряд), врач может проводить простую ручную дефибрилляцию, не отвлекаясь от пациента и не теряя драгоценное время.

Постоянная готовность в режиме ожидания

Благодаря ежедневной и еженедельной самодиагностике с автоматической загрузкой отчетов о результатах на центральную станцию дефибриллятор BeneHeart D3 может применяться для спасения жизни в любой момент. Надежный аккумулятор, выдерживающий 200 разрядов при 360 Дж и 6 часов непрерывного мониторинга, также гарантирует работоспособность устройства.

Эффективное ИТ-решение

Наше эффективное ИТ-решение позволяет управлять всей информацией с BeneHeart D3 и передавать ее в мощную систему централизованного мониторинга через WiFi 5G/2.4G. Совместимо с международным стандартным протоколом IHE HL7 и DHCP.

*  Mindray предоставляет 2 варианта измерения SpO2: с помощью аппаратов Mindray,    и Nellcor. Для получения информации о наличии измерителей SpO2  Masimo  и Nellcor свяжитесь с местными торговыми представителями.

Комплектация

Профили конфигурации можно легко настроить и загрузить для различных клинических сценариев или эксплуатационных требований, включая значения по умолчанию, структуру экрана и конфигурацию системы.

p-10

  • Способ управления: ручной, автоматический
  • Импульс дефибрилляции: усеченный экспоненциальный бифазный
  • Режим работы: асинхронная внешняя дефибрилляция
  • Число разрядов: 100 (360 Дж)
  • Каналы для измерения: ЭКГ
  • Для новорожденных, детей и взрослых
  • Регистрационное удостоверение

Арт. 6591

Дефибриллятор-монитор Mindray BeneHeart D3

  • Описание
  • Характеристики
  • Комплектация
  • Данные для транспортировки
  • Документация
  • Дефибриллятор-монитор Mindray BeneHeart D3 компактная и легкая конструкция «четыре в одном» объединяет функции мониторинга, ручной дефибрилляции, автоматической наружной дефибрилляции и кардиостимуляции
  • Подходит для применения в медицинских учреждениях, а также для условий вне стационара
  • Большой и яркий дисплей обеспечивает отображение 3 кривых и позволяет просматривать кривые ЭКГ
  • Дефибрилляция, синхронизированная кардиоверсия и автоматическая наружная дефибрилляция используют бифазную технологию
  • Возможность подачи разряда энергией до 360 Дж обеспечивает максимальную вероятность успешной дефибрилляции
  • Мощная аккумуляторная батарея поддерживает длительный непрерывный мониторинг и подачу разрядов во время транспортировки пациента без использования внешних источников питания

Перед применением необходимо ознакомиться с инструкцией по применению (паспортом изделия) или получить консультацию специалиста

Способ управления

ручной, автоматический

Импульс дефибрилляции

усеченный экспоненциальный бифазный

Режим работы

асинхронная внешняя дефибрилляция

Уровни энергии, Дж

1–10, 15, 20, 30, 50, 70, 100, 150, 170, 200, 300, 360 — ручной режим, 100–360 — режим АНД

Число разрядов

100 (360 Дж) с интервалом в 1 мин

Разрешение, пиксели

800х480

Время от начала анализа до завершения набора энергии, не более, с

20

Электроды

взрослые/детские, многоразовые

Время заряда батареи, не более, ч

3 — при выключенном устройстве, 4,5 — при включенном

Работа от встроенной батареи, ч

2,5 — мониторинг, 100 разрядов по 360 Дж с интервалом в 1 мин — дефибрилляция

Электропитание от сети переменного тока, В/Гц

220/50

Электропитание от встроенной батареи, В

14,8

Электробезопасность

класс I, тип BF

Гарантия производителя, мес.

12

Дефибриллятор-монитор Mindray BeneHeart D3

Перед применением необходимо ознакомиться с инструкцией по применению (паспортом изделия) или получить консультацию специалиста

  • Дефибриллятор-монитор — 1 шт.
  • Кабель ЭКГ с электродами — 1 шт.
  • Комплект электродов внешних (взрослые/детские) — 1 шт.
  • Аккумулятор многоразовый — 1 шт.
  • Гель 250 гр. — 1 шт.
  • Рулонная бумага для термопринтера — 3 шт.
  • Руководство по эксплуатации

Габариты в упаковке, м

0,550х0,410х0,375

Количество изделий в упаковке

1

Упаковка

картонная коробка

Дополнительная упаковка

требуется

Понравилась статья? Поделить с друзьями:
  • Дефибриллятор lifepak 1000 инструкция по применению
  • Дефибриллятор heartstart frx инструкция по эксплуатации
  • Дефероксамин инструкция по применению цена отзывы
  • Деферазирокс инструкция по применению цена отзывы аналоги
  • Дефенсор 3 инструкция по применению цена