Measurement and Calibration
For better accuracy, frequent calibration of the meter is recommended. In
addition, the instrument must be recalibrated:
• When high accuracy is required.
• At least once a month.
Measurement
Press the ON/OFF button to turn the meter ON. Place the probe to the
desired solution. The conductivity value, automatically compensated for
temperature, will be shown on the LCD.
The measured temperature will be shown on the secondary LCD.
Note: Before taking any measurement make sure the meter has been
calibrated.
Calibration Procedure
• Press CAL button to enter calibration mode from measurement mode.
• The meter will enter the calibration mode, displaying «mS/cm 12.88
USE» with CAL tag blinking.
1. Pour 2″ (5 cm) of standard solution into a container.
2. Place the probe in calibration solution. The probe tip should be centered
in the solution and submersed 1.18″ (3 cm). The meter will automatically
recognize the solution.
3. If the solution is not recognized or is out of the accepted range «—- Err»
is displayed.
4. If the calibration solution is recognized «REC» is displayed until the
reading is stable and the calibration is accepted.
• After acceptance, the «Stor» message is displayed and meter returns to
measurement mode.
Clear Calibration
Place meter in calibration mode. Press ON/OFF and «Clr» is displayed. The
meter will now be at default calibration.
Error messages
During user calibration, if the reading is out of the accepted range, the meter
will display «—- Err».i.e. 20.00 mS/cm blinking means the limit of the
measured value is out of range.
If the measured temperature is higher than 50.0 ºC or lower than 0.0 ºC, the
50.0 ºC or 0.0 ºC temperature value will blink on the LCD in measure mode.
Care and Maintenance
To obtain the highest accuracy for measurement it is important to follow
these tips:
• Calibration is only as good as the solution being used. The calibration
solution values change over time. Fresh solution should be used for each
calibration.
• The probe should be rinsed with purified water each time before placing
in calibration solution or sample to be tested.
Battery Replacement
The meter features a low battery indicator. When the battery is running low
(under 10%), the battery indicator will blink on the LCD. When the battery
is discharged «dEAd bAtt» will be displayed on the LCD for 2 seconds and
the meter will turn off.
To change the CR2032 Li-ion battery, turn the battery cover located on the
back of the meter counterclockwise to unlock. Remove cover and replace with
new battery + side facing up.
Note: Batteries should only be replaced in a safe area using the
battery type specified in this instruction manual. Old batteries should
be disposed in accordance with local regulations.
Recommendations for Users
Before using Hanna Instruments products, make sure that they are entirely
suitable for your specific application and for the environment in which they
are used. Any variation introduced by the user to the supplied equipment
may degrade the meter’ s performance. Avoid touching the electrode area.
To avoid damages or burns, do not put the meter in microwave oven. For
yours and the meter’s safety do not use or store the meter in hazardous
environments.
Accessories
Solutions
Code
Description
12.88 mS/cm calibration solution,
HI70030P
20 mL sachets (25 pcs.)
12.88 mS/cm calibration solution,
HI7030M
230 mL bottle
12.88 mS/cm calibration solution,
HI7030L
500 mL bottle
Warranty
These meters are warranted for a period of one year against defects in
workmanship and materials when used for their intended purpose and
maintained according to instructions. Probes are warranted for a period
of six months. This warranty is limited to repair or replacement free of
charge. Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered. If service is required, contact your local Hanna
Instruments Office. If under warranty, report the model number, date of
purchase, serial number and the nature of the problem. If the repair is not
covered by the warranty, you will be notified of the charges incurred. If the
instrument is to be returned to Hanna Instruments, first obtain a Returned
Goods Authorization (RGA) number from the Technical Service department
and then send it with shipping costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
IST98304N 06/18
INSTRUCTION MANUAL
DiST 4
HI98304
EC Tester
Тестеры серии DiST представляют собой прочные и надежные карманные тестеры для получения быстрых и точных показаний электрической проводимости. Серия тестеров DiST® широко используется для мониторинга электрической проводимости питьевой воды, в процессах водоподготовки, обратном осмосе, градирнях, сточных водах, лабораториях, сельском хозяйстве, аквакультуре и аквариумах, гидропонике и полиграфической промышленности. Эти тестеры имеют неокисляемый амперометрический графитовый электрод, обеспечивающий лучшую воспроизводимость при измерениях. Амперометрическое измерение электрической проводимости основано на применении закона Ома, I = V / R, где R зависит от расстояния между двумя электродами и величины их поверхности. Окисление электродов изменяют и расстояние, и величину поверхности, что влияет на точность измерений. Неокисляемые графитовые электроды тестеров DiSTs обеспечивают оптимальную поверхность для получения точных и надежных результатов. Калибровка автоматическая.
OPERATION:
• Remove the protective cap.
• Turn the DiST
DiST
DiST
DiST
DiST on.
• Immerse it in the solution to be tested without
exceeding the maximum immersion level.
• Stir gently and wait for the reading
to stabilize.
DiST
DiST
DiST
DiST
DiST automatically compensates for
temperature variations.
• The reading on display is directly expressed in:
ppm (mg/L) unit for DiST1
DiST1
DiST1
DiST1
DiST1 (
HI98300
HI98300
HI98300
HI98300
HI98300 & HI98301
HI98301
HI98301
HI98301
HI98301
),
ppt (g/L) unit for DiST2
DiST2
DiST2
DiST2
DiST2 (HI98302
HI98302
HI98302
HI98302
HI98302),
µS/cm for DiST3
DiST3
DiST3
DiST3
DiST3 (HI98303
HI98303
HI98303
HI98303
HI98303) and
mS/cm for DiST4
DiST4
DiST4
DiST4
DiST4 (HI98304
HI98304
HI98304
HI98304
HI98304).
CALIBRATION:
• Immerse the DiST
DiST
DiST
DiST
DiST in the indicated calibration
solution (see Specifications), without exceeding
the maximum immersion level.
• Allow the reading to stabilize and with a small
screwdriver turn the calibration trimmer until the
display shows the required solution value.
E.g. for calibrating HI98301
HI98301
HI98301
HI98301
HI98301 the correct calibra-
tion solution is HI70032
HI70032
HI70032
HI70032
HI70032, and the trimmer should
be turned until the display shows «
1382
«.
BEPS AND BATTERY REPLACEMENT:
DiST
DiST
DiST
DiST
DiST is supplied with BEPS (Battery Error Preventing
System) which avoids any erroneous reading due to
low batteries level, by switching the meter off.
To replace the batteries, slide off the battery com-
partment cover and replace all four 1.5V batteries
while paying attention to their polarity.
Batteries should only be replaced in a nonhazardous
area using the battery type specified in this instruction
sheet.
DiST
HI98300/1/2/3/4
Conductivity/TDS Meters
with Automatic Temperature
Compensation
Manufacturers since 1978
Visit our Internet Home Page:
http://www.hannainst.com
ACCESSORIES:
HI 70030P
12880 µS/cm calibration kit
(25 x 20 mL sachets)
HI 70031P
1413 µS/cm calibration kit
(25 x 20 mL sachets)
HI 70032P
1382 ppm (mg/L) calibration kit
(25 x 20 mL sachets)
HI 70038P
6.44 ppt (g/L) calibration kit
(25 x 20 mL sachets)
HI 70039P
5000 µS/cm calibration kit
(25 x 20 mL sachets)
HI 70442P
1500 ppm calibration kit
(25 x 20 mL sachets)
HI 73301
Spare electrode for DiST1
HI 73302
Spare electrode for DiST2
HI 73303
Spare electrode for DiST3
HI 73304
Spare electrode for DiST4
HI 731326
Calibration screwdriver (20 pcs)
Thank You
Thank you for choosing a Hanna Instruments product. Please read
this instruction manual carefully before using the instrument.
For more information about Hanna Instruments and our products,
visit www.hannainst.com or e-mail us at sales@hannainst.com.
For technical support, contact your local Hanna Instruments
Office or e-mail us at tech@hannainst.com
Find your local Hanna Instruments Office on www.hannainst.com
Preliminary Examination
Remove the meter from the packing material and examine it carefully to
make sure that no damage has occurred during shipment. If noticeable
damage is evident, contact your local Hanna Instruments Office.
Each meter is supplied with:
• CR2032 battery (1 pc.)
• Storage / Protection sleeve
• 12.88 mS/cm calibration solution sachet (4 pcs.)
• Instruction manual
• Quality Certificate
Note: Save all packing material until you are sure that the instrument
functions correctly. All defective items must be returned in the original
packaging with the supplied accessories.
Preparation
The probe is shipped dry. Before using the electrode, remove the protective
cap. Then follow the calibration procedure.
• Turn the tester on by pressing ON/OFF button.
• Remove the protective cap and immerse the tip of the probe in the
sample to be tested.
• Stir gently and wait for a stability tag to disappear.
• The electrode automatically compensates for temperature variations.
• The reading on display is directly expressed in: mS/cm.
• For best results, recalibrate periodically.
• After use rinse the probe with water.
• Always replace the protective cap after each use.
Operation
mS/cm
Liquid Crystal
Display
CAL
CAL button
ON/OFF button
DiST 4
Recommended
Immersion
Level
Probe
Temperature
Sensor (inside)
All rights are reserved. Reproduction in whole or in part is prohibited
without the written consent of the copyright owner.
Specifications
0.00 to 20.00 mS/cm
Range
0.0 to 50.0 ˚C (32.0 to 122.0 ˚F)
0.01 mS/cm
Resolution
0.1 ˚C (0.1 ˚F)
Accuracy
±2% F.S.
(@25 °C/77 °F)
±0.5 °C (±1.0 °F)
Calibration
HI70030
Solutions
(12.88 mS/cm)
Calibration
automatic, single point (12.88 mS/cm)
Temperature Compensation automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF)
Battery type
CR2032 Li-Ion 3V (1 pc.)
Battery life
approximately 250 hours of continuous use
0 to 50 ºC (32 to 122 ºF);
Environment
RH 100% max
Weight
68 g (2.4 oz.)
(without battery)
Hanna Instruments reserves the right to modify the design,
construction, or appearance of its products without advance notice.
Operational Guide
To turn the meter ON and check the battery status
Press the ON/OFF button to turn the meter on. At start-up, all the LCD
segments are displayed for 1 second, then the percent indication of the
remaining battery life is displayed for another second. The meter then enters
the normal measuring mode.
Note: Keeping the ON/OFF button pressed while turning the meter on
will display all LCD segments as long as the button is pressed.
To enter calibration mode
Press the CAL button. «CAL» message is displayed.
To enter setup mode
Remove the battery cover and press the Setup button located on the side
of the battery.
Meter Setup
While in measurement mode, remove the battery cover. Press
the Setup button located on the side of the battery in the battery
compartment. The meter will enter in setup mode. Press the
ON/OFF button to move through setup parameters. Change option by
pressing CAL button. The default settings are: «Set t» measure unit — °C,
«AOFF» — 8 min.
Setup
Button
To select the temperature unit (°C/°F)
To select the temperature unit when «SET t» is displayed press the CAL
button to change between °C or °F.
To select the Auto-Off time
To select the AUTO OFF when «AOFF» is displayed press the CAL button to
change between 8 min, 60 min or — (disabled).
To return to measurement mode
Press the ON/OFF button.
Specifications:1893/1893382-dist_4.pdf file (04 Feb 2023) |
Accompanying Data:
Hanna Instruments DiST 4 Test Equipment PDF Instruction Manual (Updated: Saturday 4th of February 2023 04:32:46 AM)
Rating: 4.9 (rated by 57 users)
Compatible devices: DiST 1, HI 3812, 3815, DiST 3, HI 96721C, HI 93739, Gro Line HI98318, HI98201.
Recommended Documentation:
Text Version of Instruction Manual
(Ocr-Read Summary of Contents of some pages of the Hanna Instruments DiST 4 Document (Main Content), UPD: 04 February 2023)
-
1, Hanna Instruments DiST 4 INSTRUCTION MANUAL HI98304 EC Tester Warranty These meters are warranted for a period of one year against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. Probes are warranted for a period of six months. This warranty is limited to repair or replace…
-
2, Liquid Crystal Display CAL button Operation Thank You Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the instrument. For more information about Hanna Instruments and our products, visit www.hannainst.com or e-mail us at [email protected]. For t…
-
Hanna Instruments DiST 4 User Manual
-
Hanna Instruments DiST 4 User Guide
-
Hanna Instruments DiST 4 PDF Manual
-
Hanna Instruments DiST 4 Owner’s Manuals
Recommended: WA4003-G2, 2015 RX350, Q-D6 SRS
-
Klein Tools VDV500-063 Toner-Pro
INSTRUCTION MANUALENGLISHVDV500-820Toner-Pro / Probe-Pro KitFRANÇAIS p. 17ESPAÑOL pág. 9• TRACE INDIVIDUAL OR PAIRED WIRES• TEST RJ11, RJ12, AND RJ45 JACKS• DETECTS CONTINUITY AND POLARITY• 5 DISTINCT TONES (3 CONSTANT, 2 ALTERNATING)• EASY-TO-UNDERSTAND STATUS LEDs• REPLACEABLE PROBE TIP …
VDV500-063 Toner-Pro 24
-
TEN N2 PRO-1234
Version 3.4 Sept 2014 N2 PRO / N2 PRO-1234 REFRIGERANT RECOVERY & RECHARGING SYSTEM OPERATORS MANUAL TEN Automotive Equipment Limited www.ten-automotive.co.uk …
N2 PRO-1234 34
-
LEYBOLD 451281
Physics Chemistry ⋅ BiologyTechnics LEYBOLD DIDACTIC GMBH11/97-kem-Instruction sheet 451 281451 292The stroboscope with microprocessor-controlled xenon flashtube can be used• to observe rapid periodic motions, e.g. oscillations of a string,as a stationary image• to measure the frequency of periodic mechanical phe …
451281 2
-
HVTest TAG-8000
1Dear ClientThank you for Purchasing our TAG-8000 Wireless HV PhaseTester. Please read the manual in detail prior to first use, which willhelp you use the equipment skillfully.Our aim is to improve and perfect thecompany’s products continually, so there may beslight differences between your purchase equipmen …
TAG-8000 22
Popular Right Now:
Operating Impressions, Questions and Answers: