User Manual
Blood Glucose/ Cholesterol/ Uric Acid Monitoring System
Before you Begin
Before using the products, read all of the instructions carefully.
It includes all the information you need to know to get accurate
readings. For EU customer service, dial +49-6894-581020.
Warnings
• Do not change your medical plan without a doctor’s approval.
The EasyTouch
®
GCU system should not be used for diagnosis.
®
• If using the EasyTouch
GCU system to monitor an existing
disease, only adapt treatment with proper training.
• Do not use the EasyTouch
®
GCU system to test newborns.
• Be sure to follow your local regulations for proper disposal of
used test strips, lancets, and batteries.
• The monitoring system will not work properly at altitudes greater
than 8,000 feet (2,400 meters) above sea level.
• This system is designed for use at temperatures between 14°C
and 40°C (57.2°F and 104°F) and less than 85% relative humidity.
If you use the system outside of the proposed conditions, it can
give false results.
Precautions
• Do not use the meter in a very dry environment, especially if
synthetic materials are present.
• Synthetic clothes, carpets, etc., may cause damaging static
discharges in a dry environment.
• Do not use this meter near cellular or cordless telephones, walkie
talkies, garage door openers, radio transmitters, or other electrical
or electronic equipment that are sources of electromagnetic
radiation, as these may interfere with the proper operation of the
meter.
• Store the device and the components in a clean, safe location.
• Store the test strips between 4 and 30 °C (39 and 86°F).
• Remove the batteries before putting the meter into storage
Limitations
The following compounds (left column) in the blood at the given
concentrations can cause inaccurate results on the test.
Low glucose concentration range: 50mg/dL ~ 100 mg/dL
High glucose concentration range: 250 mg/dL ~ 350 mg/dL
Compound
Glucose Conc.
Calculated Value
Acetaminophen
Low
7.5 mg/dL
Ascorbic acid
Low
3.2 mg/dL
65.1 mg/dL
Low
Cholesterol
High
105.0 mg/dL
Low
0.2 mg/dL
Gentisic acid
High
0.2 mg/dL
39.3 mg/dL
Low
Glutathione
113.2 mg/dL
High
Low
-5.8 mg/dL
Hemoglobin
High
-3.9 mg/dL
Low
2.5 mg/dL
L-Dopa
5.0 mg/dL
High
Low
0.3 mg/dL
Methyl-Dopa
High
0.6 mg/dL
Low
23.3 mg/dL
PAM (pralidoxime Iodide)
High
44.7 mg/dL
7.4 mg/dL
Low
Tolazamide
High
24.8 mg/dL
Low
6.2 mg/dL
Uric acid
High
7.3 mg/dL
Compound
Uric Acid
Compound
Cholesterol
Acetaminophen
Acetaminophen
2 mg/dL
20 mg/dL
Amiloride
20 mg/dL
Ascorbic acid
10 mg/dL
Ascorbic acid
5 mg/dL
Bilirubin
21.2 mg/dL
Cholesterol
358 mg/dL
Dopamine
10 mg/dL
Creatinine
30 mg/dL
L-Dopa
5 mg/dL
Diclofenac
75 mg/dL
Uric acid
19.5 mg/dL
Dopamine
2 mg/dL
Ketoprofen
500 mg/dL
L-Dopa
5 mg/dL
Methyl-Dopa
1.0 mg/dL
Cholesterol up to 500 mg/dL or triglycerides up to 3,000 mg/dL do
not significantly affect results. Grossly lipaemic patient samples
have not been tested and are not recommended for testing with the
®
EasyTouch
GCU System.
Intended Use
The EasyTouch
®
GCU Monitoring System is designed for in
vitro diagnostic use only (external use only), and is suitable
for self-testing. The system is for healthcare professionals and
persons with diabetes, hypercholesterolemia, or hyperuricemia to
measure glucose, cholesterol, and uric acid. Simply add a drop
of blood to the test strip, and the result is displayed on the screen
in 10 seconds for glucose, 150 seconds for cholesterol, and 20
seconds for uric acid.
Test Principle
Blood glucose, cholesterol and uric acid measurements are
based on electrical current changes caused by the reaction
of glucose, cholesterol, or uric acid with the reagent on the
electrode of the corresponding strip.
Items in the Package
1. EasyTouch Meter
5. Lancets
2. EasyTouch Kit and
6. Check Strip
puncturer User Manual
7. Pouch
3. Puncturer — lancing device
8. 3V CR2032 battery
4. Simple User Guide
9. EasyTouch Test Strips/
Code
1
2
3
GCU
Blood Glucose / Cholesterol / Uric Acid Monitoring System
7
8
9
4
5
6
Getting to Know the System
Test Strip Slot
Code Key Slot
Unit-Changing Switch
(To Switch between mg/dL
and mmol/L)
Screen(Display Messages
and Test Result)
Memory Button
(Recall Stored Data)
Battery Cover
Setting Date and Time Button
Battery Compartment
Test Strips
Control Solution
3ml for glucose and uric acid; 1ml for
Blood Glucose/Cholesterol/
cholesterol
Uric Acid Test Strips
*Control Solution not included in the
package
Sample target area
Strip handle
Contact bars
Structure of Puncturer and Lancet
Lancet
Puncturer
Checking the Unit of Measure
®
The EasyTouch
GCU Meter can measure your blood glucose,
cholesterol, and uric acid in mg/dL, mmol/L (mM), or μmol/L (μM).
Just adjust the switch in the battery compartment:
Left Position
Right Position
Unit
Decimal point Unit
Decimal point
Glucose
mg/dL
No
mmol/L (mM) Yes
Cholesterol
mg/dL
No
mmol/L (mM) Yes
Uric acid
mg/dL
Yes
μmol/L (μM) No
Inserting the Batteries
If display shows «conn», please press any botton or re-install the
batteries again.
1. Remove the battery cover
from the back of the meter.
2. Insert two AAA batteries
into the battery
compartment.
3. Replace the battery cover.
Setting the Time and Date
When you insert new batteries, the meter automatically enters date
and time setting. Set the correct date before using the meter.
1. Press «S» to change to the current year.
2. Press «M» to shift to the month setting.
3. Adjust the month by pressing «S» until the current month appears.
4. Repeat this process to set the date, hour and minute.
5. Simply wait for the meter to turn off automatically.
Using the Check Strip
Insert the check strip into the test strip slot on the
meter.The screen should display «OK». If «X» appears
on the screen, remove the check strip and reinsert it.
If «X» is displayed again, call our customer support.
Do not test until the meter is working properly.
Use the check strip to check the meter when:
— You use the EasyTouch
®
GCU Monitoring System for the first time.
— You drop the meter.
— You think there is something wrong with the results you got and you
want to check whether the meter and test strips are working correctly.
Control Testing —
What control test is
By using control solution that has a measured amount of glucose,
cholesterol, or uric acid, you can check if your meter and strips are
accurate. If your control results fall within the range on the test strip
vial, you can be sure that the system is working properly.
When to perform a control test
Perform a control test when you think there is something wrong with
your measurement and you want to check if the meter and test strips
are working correctly.
Items you need
®
GCU Series Meter
EasyTouch
EasyTouch
®
Glucose Control Solution (not included)
EasyTouch
®
Cholesterol Control Solution (not included)
EasyTouch
®
Uric Acid Control Solution (not included)
EasyTouch
®
Glucose/Cholesterol/Uric Acid Test Strips
®
EasyTouch
Glucose/Cholesterol/Uric Acid Code Key
Always write down the opening date of your glucose, cholesterol or uric
acid control solution bottle and test strip vial.
Glucose strips and control solution expire three months after first
opening or until the expiration date, whichever comes first. Cholesterol
strips and control solution expire two months after first opening or until
the expiration date, whichever comes first. Uric acid strips and control
solution expire three months after first opening or until the expiration
date, whichever comes first. The meter stores control test results in its
memory, where they appear just like test results.
Control Testing Steps
1. Insert the glucose, cholesterol or uric acid code key
into the code key slot on the back of the meter.
2. Take one glucose, cholesterol, or uric acid strip from
the vial. Close the vial quickly.
3. Insert the test strip into the test strip slot on the meter.
The meter first displays the code number, and then
the blood symbol «S».
4. Hold the glucose, cholesterol, or uric acid control
solution bottle upside down and slowly squeeze the
bottle to form a small drop. Discard the first drop of
the solution. Squeeze out a second drop and let it
touch the edge of the sample target area on the test
strip. The test strip draws the control solution into the
targeting zone.
5. Immediately after the meter sounds a beep, close the
cap of the control solution. The meter then counts
Glu
down from 10 for glucose, 150 for cholesterol, or 20
for uric acid, and displays your result on the screen.
6. Check if the displayed value falls within the
pm
acceptable range shown on the test strip vial.
Glu
7. Remove the test strip from the meter and discard it.
pm
Understanding Control Test Results
You will find the acceptable range of glucose, cholesterol, and uric acid
levels for normal and high controls on the test strip vial label. If your
control results fall within the acceptable range, you can begin to test your
blood glucose, cholesterol, and uric acid levels. If your control results are
not within the acceptable range, consider the following items:
— Are your test strips or control solutions expired?
— Have you ever forgotten to close your test strip vials or control
solution bottles?
— Does the code in the meter match the code on the test strip vial?
— Did you follow all the operating instructions correctly?
Repeat the control test with a new test strip, following the steps exactly.
If you continue to receive control values that are outside of the acceptable
range, discontinue your blood testing and call customer service or your
healthcare provider immediately.
Blood Testing
Items You Need
1. EasyTouch
®
GCU Meter
®
2. EasyTouch
Glucose/Cholesterol/Uric Acid Test Strips
3. Puncturer (Lancing Device) and lancets
4. Log sheet
5. Alcohol swab (not included)
Preparing the Puncturer
Blood glucose, cholesterol, and uric acid monitoring devices and
lancing devices can transmit bloodborne pathogens.
To avoid infection, you should:
— Never reuse the lancets.
— Never share your puncturer with others.
— Always use a new, sterile lancet each time a test is performed.
— Always use an alcohol pad to wipe blood off of the puncturer’s
tip after use.
For safety and to prevent cross-contamination, always place
the protective cover back on before discarding the used lancets.
To avoid injury, do not leave a used lancet in the puncturer.
Always remove used lancets immediately after a test.
Please read puncturer’s instruction for use manual before use
Blood Glucose Testing
®
The first time you use the EasyTouch
GCU Meter or
open a new test strip vial, insert the code key from
the test strip vial. Each test strip vial contains one
code key. Make sure the number on the code key
matches the code number on the vial of test strips
you use.
1. Take one strip from the vial of glucose test strips.
Close the vial quickly.
2. Insert the test strip into the test strip slot on the
meter. The meter first displays the code number,
and then the blood symbol «S».
3. Clean your finger with an alcohol swab. Let it dry
completely.
To increase blood flow to the lancing site, use warm water to wash
your hands, and let your arm hang down at your side.
4. Place the puncturer on your finger.
5. Press the trigger on the puncturer.
6. Withdraw the puncturer.
It is very important that you put enough blood on your test strip so that
the entire reaction zone is covered. This ensures that your meter can
give accurate and reliable results.
7. Wipe away the first drop of blood and allow a
second drop to form.
8. Touch the drop of blood to the top of the test strip
target area. The target area turns red as the strip
absorbs the blood. The testing reaction starts
when the meter sounds a beep.
9. The meter then counts down from 10, shows
your result on the screen, and stores your result
in its memory automatically.
10. Record your glucose value on the provided log
sheet.
Fasting Blood Glucose
Normal
70 ~ 100 mg/dL
Early Diabetes
101 ~ 125 mg/dL
Diabetic
> 126 mg/dL
The following values are expected glucose levels for non-diabetics
Upon waking — fasting
70 ~ 99 mg/dL
After meals
70 ~ 140 mg/dL
The following values are targets for diabetics
Before meals
70 ~ 130 mg/dL
1 ~ 2 hours after starting a meal
< 180 mg/dL
Source: American Diabetes Association
Consult your healthcare provider for the appropriate range for you.
Always consider your symptoms when interpreting your blood glucose
results. If your results conflict with how you feel, perform a control test.
The following are just some of the symptoms for low and high blood
sugar.
Hypoglycemia
Hyperglycemia
•
Dizziness/Headache
•
Frequent urination
•
Sweating
•
Thirst
•
Hunger
•
Headache
•
Pounding heartbeat
•
Fatigue
•
Shakiness
•
Blurred vision
Blood Cholesterol Testing
The first time you use the EasyTouch
®
GCU
Meter or open a new test strip vial, insert the
code key from the test strip vial. Each test
strip vial contains one code key. Make sure
the number on the code key matches the code
number on the vial of test strips you use.
1. Take one strip from the vial of cholesterol
test strips. Close the vial quickly.
2. Insert the test strip into the test strip slot on
the meter. The meter first displays the code
number, and then the blood symbol «S».
3. Clean your finger with an alcohol swab. Let
it dry completely.
4. Place the puncturer on your finger.
5. Press the trigger on the puncturer.
6. Withdraw the puncturer.
7. Wipe away the first drop of blood and allow
a second one to form.
The blood volume needed in cholesterol test is
significantly more than that in glucose test. You
need to apply a 15 μl blood sample in cholesterol
test and a 4 μl blood sample in glucose test.
8. Touch the drop of blood to the side of the
test strip target area. The target area turns
red as the strip absorbs the blood. Apply
blood until the meter sounds a beep. The
meter then counts down from 150 seconds,
shows your result on the screen, and stores
your result in its memory automatically.
9. Record your cholesterol value on the
provided log sheet.
The following are ranges for blood cholesterol
levels:
Desirable
160 — 200 mg/dL
Borderline High
200 — 239 mg/dL
High
> 240 mg/dL
Source: American Heart Association
The above range is just for reference, and it may not apply to
everyone. If your blood cholesterol is borderline high or high,
discuss these results with your healthcare provider.
Consult your healthcare provider for the appropriate range for you.
Тест-полоски на мочевую кислоту (10 шт.).
Особенности
В основу работы аппарата положена система Easy Touch. Она не имеет аналогов на современном российском рынке. Преимущество системы в удобном использовании, поскольку она позволяет сделать сразу три анализа. Все пациенты смогут ежедневно контролировать интересующие показатели. Для этого достаточно капельки крови из кончика пальца.
Применение Easy Touch основывается на:
Мониторинге биохимического метода измерения. Это позволяет использовать минимум крови при анализе.
Быстром получении результатов. Данные о глюкозе и мочевой кислоте отображаются на экране через 6 сек., холестерина — через 150 сек.
Функции сохранении информации. Это помогает сравнивать уровень показателей самостоятельно и отслеживать ход любых изменений.
Показания
Прибор показан к использованию тем, кому требуется постоянно контролировать количество гемоглобина, глюкозы и холестерина. Прежде всего, это пациенты с анемией, диабетом, гиперхолестеринемией.
Способ применения
К аппарату прилагается инструкция по применению. Для проведения анализа достаточно нанести капельку крови на полоску. Обратите внимание, тест включает разное количество полосок, предназначенных для измерения. Прибор работает от батареек стандарта ААА. Прилагается проверочная полоса. Для удобства в применении предназначен дневник самоконтроля и памятка.
Использование основывается на определенных диапазонах измерения:
На глюкозу: в рамках 20-600 мг/дл.
На холестерин в рамках 100-400 мг/дл.
На мочевую кислоту: в рамках 179-1190 мкмоль/л.
Условия хранения
Хранить и транспортировать следует в чехле, поставляемом в комплекте, при температуре от -15 до 25° С.
Условия отпуска из аптеки
Без рецепта.
Биохимический анализатор крови EasyTouch GCu
Биохимический анализатор EasyTouch GCu — многофункциональная система мониторинга уровня глюкозы / холестерина / мочевой кислоты в крови.
Применяемость
Для самостоятельного контроля содержания глюкозы/холестерина/мочевой кислоты в крови,
Многофункциональная система мониторинга предназначена только для диагностики in vitro (только для наружного применения).
Система предназначена для работников здравоохранения и людей, страдающих диабетом, гиперхолестеринемией или гиперурикемией, для количественного измерения содержания глюкозы, холестерина и мочевой кислоты в свежей капиллярной цельной крови из кончика пальца.
Частый контроль содержания в крови глюкозы, холестерина, мочевой кислоты – дополнительная забота о людях, страдающих диабетом, гиперхолестеринемией и гиперурикемией.
Просто нанесите каплю крови на тестполоску, и результат содержания глюкозы будет отображен на экране спустя 6 секунд, холестерина спустя 150 секунд и мочевой кислоты спустя 6 секунд.
● Многофункциональная система EasyTouch® GCU подходит для самоконтроля при диабете, гиперхолестеринемии или гиперурикемии в домашних условиях или для профессионального использования.
● Многофункциональная система EasyTouch® GCU может быть использована только с тестполосками EasyTouch® на глюкозу, тест-полосками EasyTouch® на холестерин и тестполосками EasyTouch® на мочевую кислоту.
● Перед тем, как использовать прибор для измерения уровня содержания глюкозы, холестерина и мочевой кислоты в Вашей крови, прочтите внимательно все инструкции. Они содержат всю информацию, необходимую Вам для получения точных результатов содержания глюкозы, холестерина и мочевой кислоты в крови. может привести к неверным результатам.
● Не изменяйте план лечения без разрешения врача. Система EasyTouch® GCU не может быть использована для постановки диагнозов диабета, гиперхолестеринемии и гиперурикемии, а также не предназначена для тестирования новорожденных.
Описание EasyTouch GCu:
— большой экран;
— тип тест-полосок прибор определяет самостоятельно;
— результат через 6 секунд (глюкоза и мочевой кислоты);
— результат через 150 секунд (холестерин);
— память (200 тестов глюкоза / 50 тестов мочевой кислоты / 50 тестов холестерин).
Комплектация EasyTouch GCu:
— биохимический анализатор крови EasyTouch GCu;
— проверочная тест-полоска;
— сумочка;
— батарейки (ААА — 2 шт.);
— автопрокалыватель;
— ланцеты (25 шт.);
— тест-полоски на глюкозу (10 шт.);
— тест-полоски на холестерин (2 шт.);
— тест-полоски на мочевую кислоту (10 шт.);
— дневник самоконтроля;
— руководство пользователя;
— гарантийный талон;
— товарный и кассовый чеки.
Биоптик ( Bioptik Technology, Inc., Тайвань ) Гарантия: 2 года
Многофункциональная система мониторинга ИзиТач / EasyTouch GCU для контроля уровня глюкозы, холестерина, мочевой кислоты
Система мониторинга EasyTouch GCU уникальна и не имеет аналогов на Российском рынке. В отличие от других известных приборов для самоконтроля, эта система позволяет провести три вида анализа при помощи только одного прибора, что отличает EasyTouch невероятным удобством. Пациенты, использующие систему EasyTouch GCU, ежедневно могут контролировать свои результаты.
Прибор создан для количественного измерения содержания глюкозы, холестерина и мочевой кислоты в свежей капиллярной цельной крови из кончика пальца. Работа системы мониторинга построена на электрохимическом методе измерения. Это позволяет обходиться минимальным количеством крови при анализе.
Результаты измерения:
глюкоза — 6 секунд, холестерин — 150 секунд, мочевая кислота — 6 секунд.
Прибор обладает функцией сохранения данных, что поможет Вам наблюдать за ходом изменений уровня глюкозы, холестерина и мочевой кислоты в крови.
г.Красноармейск, Новая Жизнь, д. 1
Чкалова ул, 7, Красноармейск, МО
Сергиев Посад-6, 1А, Сергиево-Посадский р-н, МО (на въезде в Военный городок, на территории рынка)
г.Сергиев Посад, площадь Красногорская, д. 1
Мичуринский пр, 9 к 1, Москва
г.Жуковский, Чкалова, д. 41
Энтузиастов ш, 3 к 1, Москва
Семеновская пл, 7 к 17А, Москва
Юшуньская малая. ул, 3, Москва
Нахимовский пр, 22, Москва
Братиславская ул, 18к1, Москва
Новоугличское ш, 46, Сергиево-Посадский р-н, Сергиев Посад, МО
Новоостаповская ул, 4к1, Москва
Петра Романова ул, 6, Москва
г.Москва, Михайлова, д. 8, корп. 1
г.Внииссок, Березовая, д. 1
Рижское ш, 41А, Волоколамск, МО
Ласточкин пр-д, 16, Балашиха, МО
Маршала Жукова пр, 49, Москва
Рязанский пр, 71к1, Москва