Care and Maintenance
To obtain the highest accuracy for measurements it is important to follow these tips:
• Fresh buffer should be used for each calibration. Calibration is only as good as the
buffer being used. The pH buffer values change over time once the sachets are opened.
• The probe should be rinsed with purified water each time before placing in buffer or
sample to be tested.
• When the meter is not in use it is important to add several drops of storage solution to
the protective cap to keep the probe hydrated. If storage solution is not available, pH
4.01 or pH 7.01 buffer can be used.
• For improved accuracy it is recommended to calibrate in two buffers.
• It is important to calibrate and measure samples at the same temperature as there
is no temperature compensation. A dramatic change in temperature between buffer
solutions and samples to be tested will give inaccurate readings.
Refilling the Electrode
• If fouled, remove the sleeve and rinse sleeve and sensor tip with purified water.
Turn over the probe, remove the sleeve by carefully rotating it and pull straight
along the axis of the electrode. Use
care as the pH stem is made of glass.
Rinse off any traces of electrolyte gel.
• Soak the sensor tip in HI70061G,
HI700661P, HI700663P or HI700664P
cleaning solution for 20 minutes then
rinse with purified water.
• Refill the reference well with HI9071
Gelled Bridge Electrolyte.
• Replace the reference sleeve: insert and
push the sleeve onto the electrode. Make
sure the black O-ring is fixed inside the
electrode body. Any excess of gel will be
expelled from the end of the electrode through the open junction.
• Rinse any excess gel off with purified water and gently dry off
body with a soft cloth or tissue.
• Soak assembled probe in Electrode storage solution for a
minimum of 30 minutes.
• Rinse probe with purified water.
• Shake the electrode down as you would do with a clinical
thermometer to eliminate any air bubbles inside the glass bulb.
• Calibrate in fresh buffers before using for measurements.
Warranty
The meter is warranted for a period of one year against defects in workmanship and materials when
used for their intended purpose and maintained according to instructions. This warranty is limited to
repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed
maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under
warranty, report the model number, date of purchase, serial number and the nature of the problem. If the
repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to
be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the
Technical Service department and then send it with shipping costs prepaid. When shipping any instrument,
make sure it is properly packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of
the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.
Auto-off
From measurement mode, press and hold the
ON/OFF button. The meter will cycle through
«OFF,» «CAL,» then current auto-off setting.
The default setting is 8 minutes («d08»). Press
ON/OFF button to change. «d60» is auto-off after
60 minutes, and «d—» disables the auto-off
feature. Press and hold the button to exit the
menu.
Clear Calibration
Place meter in calibration mode. Press and hold
ON/OFF until «CLr» is displayed. The meter will
now be at default calibration.
«Err» Message
In calibration mode, if the meter displays an
«Err» message when in the correct fresh buffer
solution then the probe should be cleaned.
Place the probe in the cleaning solution for
20 minutes. Rinse with purified water and place
in storage solution for a minimum of 30 minutes
before calibrating.
Battery Indicator
The meter features a low battery indicator. When
the battery is running low, the tag will blink on
screen. When the battery has been depleted,
«Erb» will appear on screen and the meter will
turn off.
Recommendations for Users
Before using this product, make sure it is entirely suitable for your specific application and for the
environment in which it is used. Any variation introduced by the user to the supplied equipment may
degrade the meters’ performance. For yours and the meter’s safety do not use or store the meter in
hazardous environments.
Certification
All Hanna Instruments conform to the CE European Directives.
Disposal of Electrical & Electronic Equipment. The product should not be treated as household
waste. Instead hand it over to the appropriate collection point for the recycling of electrical and electronic
equipment which will conserve natural resources.
Disposal of waste batteries. This product contains batteries, do not dispose of them
with other household waste. Hand them over to the appropriate collection point for recycling.
Ensuring proper product and battery disposal prevents potential negative consequences for the
environment and human health. For more information, contact your city, your local household waste
disposal service, the place of purchase or go to www.hannainst.com.
Battery Replacement
To change the CR2032 Li-ion battery, turn the battery cover located on the back of the meter
counterclockwise to unlock. Remove cover and replace the battery with + side facing up.
Note: Batteries should only be replaced in a safe area using the battery
type specified in this instruction manual. Old batteries should be disposed in
accordance with local regulations.
Accessories
pH Buffer Solution
Code
Description
HI70004G
GroLine pH 4.01 buffer solution, 20 mL sachets (25 pcs.)
HI70007G
GroLine pH 7.01 buffer solution, 20 mL sachets (25 pcs.)
HI77400P
pH 4.01 & 7.01 buffer solution, 20 mL sachets (10 pcs., 5 ea.)
Electrode Cleaning Solutions
Code
Description
HI70061G
GroLine general purpose cleaning solution, 20 mL sachets (25 pcs.)
HI700661P
General purpose cleaning solution for agriculture, 20 mL sachets (25 pcs.)
HI700663P
Cleaning solution for soil deposits, 20 mL sachets (25 pcs.)
HI700664P
Cleaning solution for humus deposits, 20 mL sachets (25 pcs.)
Electrode Storage Solutions
Code
Description
HI70300G
GroLine electrode storage solution, 20 mL sachets (25 pcs.)
HI70300-023
GroLine electrode storage solution, 230 mL bottle
HI9072
Electrode storage solution, 13 mL dropper
Electrode Fill Solution
Code
Description
RoHS
HI9071
Gelled Bridge Electrolyte
compliant
Other Accessories
Code
Description
HI721319
Soil auger
HI7051M
Soil preparation solution, 230 mL bottle
IST981030 05/18-1
INSTRUCTION MANUAL
HI981030
Soil pH Tester
Улучшенные характеристики для высокоточных измерений
Индикатор стабильности показаний на дисплее существенно облегчает работу, указывая на окончание измерения.
Активируемая функция авто-отключения поможет существенно продлить срок службы батареи, если прибор случайно забыли выключить.
Карманный прибор
Компактный размер тестера делает его идеальным для измерений «в поле». Крупные и яркие цифры на LCD дисплее отчетливо видны даже под углом
Энергоэффективность
Батарейки CR2032 Li-ion, входящей в комплект поставки, хватает на 1000 часов непрерывного использования. Прибор имеет индикатор батареи, если индикатор мигает, значит пришло время сменить батарею.
Для обеспечения точности измерений нужно поддерживать электрод в оптимальном состоянии. Электрод HI981030 разработан специально для проведения прямых измерений в почв.
Уникальная открытая диафрагма со съемным рукавом
Если Вы заметили, что прибор стал работать медленнее, просто снимите пластиковый рукав с электрода, очистите его и замените электролит. Чистый датчик со свежим электролитом повысит точность Ваших измерений и продлит срок службы электрода.
Коническая стеклянная мембрана
Конструкция мембраны позволяет внедрять электрод в толщу почвы или жидкой глины, обеспечивая максимальный контакт измерительной части электрода с образцом.
Корпус из ПЭТФ
Полиэтилентерефталат (ПЭТФ) — это высококачественный надежный пластик, стойкий к действию большинства растворителей и химикатов. ПЭТФ выдерживает дезинфекцию хлоркой (отбеливателем)
Анализ показателя рН почвы – важная мера, позволяющая существенно улучшить процесс выращивания цветов и растений, плодовых и декоративных деревьев, овощных культур и других зелёных насаждений, так как данный показатель позволяет судить о состоянии почвы и её пригодности для эффективного роста и развития конкретных видов растений, необходимости добавлять те или иные удобрения и т. д. Вот почему регулярные исследования грунта практикуют как частные садоводы, так и профессионалы сельскохозяйственной сферы и других аналогичных областей. Проявляя заботу об упрощении их ежедневного труда, компания Ханна Инструментс создала специализированный портативный электронный рН метр HI981030 для почвы, жидкой глины и других аналогичных образцов, качество исследования и эксплуатационные характеристики которого позволили прибору войти в российский Госреестр средств измерений, включающий в себя лишь самое достойное оборудование. В свою очередь, компания-изготовитель тестера HI981030 предоставляет на него годовую гарантию, что также является убедительным доказательством надёжности модели.
Особенности измерения рН почвы с помощью тестера HI981030 Hanna
С тестером HI981030 Hanna Instruments процедура анализа рН почвы максимально упрощается, во многом благодаря специальной конструкции измерительного электрода модели, которая позволяет проводить исследования, как говорится, прямо «в поле», без необходимости достаточно трудоёмкой предварительной подготовки образца измерений. Кроме того, мобильность исследований обеспечивается независимостью устройства от электросети, так как для его работы достаточно комплекта батарей CR2032, поддерживающих непрерывную работу тестера в режиме измерений в течение 1000 часов. Если на дисплее будет мигать индикатор батареи, значит, необходимо заменить их на свежие, а дополнительную энергоэффективность модели обеспечивает активируемая функция автоотключения питания.
Измерения рН почвы с помощью миниатюрного тестера HI981030 можно производить в диапазоне значений 0-14 рН. При рабочей температуре внешней среды от 0 до 50 °C и максимальной влажности 95% точность исследований составит ±0,05 pH.
Бренд | HANNA Instruments |
Диапазон измерения | от 0 до 14,0 pH |
Госреестр | Внесен |
Относительная влажность RH % | RH 95% max |
Диапазон рабочих температур, С | от 0 до 50 °C |
Дисплей | LCD |
Материал корпуса | полиэтилентерефталат (ПЭТФ) |
Точность | ±0,05 pH при 25 °C |
Калибровка | автоматическая, по одной или двум точкам |
Термокомпенсация | автоматическая, от 0 до 50 °C |
Разрешение | 0,01 pH |
Автоотключение | 8 минут, 60 минут, или выкл. |
Дополнительный функционал | индикатор состояния батареи, индикатор стабильности показаний |
Питание | батарея CR2032 Li-ion |
Срок службы батареи | приблизительно 1000 ч непрерывного использования |
Комплектация | тестер HI981030, буферы pH 4,01 и 7,01 (по 2 саше), очищающий раствор от почвенных отложений, раствор для очистки от гумуса, раствор для хранения электрода, пипетка, раствор для заполнения электрода, инструкция по эксплуатации, заводской сертификат |
Габариты | 51 x 151 x 21 мм |
Вес | 44 г |
Гарантия | 12 месяцев |
Note: Never immerse the tester over the maximum immersion level. Auto-Off From measurement mode, press and hold the ON/OFF button. The LCD will show "OFF", "CAL" followed by "d08" (default setting, 8 minutes). Release the ON/OFF button. A single press on the ON/OFF button will change the auto-off timer to "d60" (60 minutes). To disable the auto-off feature press the ON/ OFF button again, LCD will show "d--". Press and hold to exit. Clear Calibration To clear the user calibration and restore the tester to factory default. From calibration mode, press and hold the ON/OFF button, the LCD will show "CLr". “Err” Message In calibration mode, if the probe is in the correct buffer solution and the "Err" message is displayed, the probe should be cleaned. Soak the probe in cleaning solution for 20 minutes. Rinse with water and hydrate electrode in storage solution for a minimum of 30 minutes before calibrating. Battery Indicator When the battery level is low, the tag on the LCD will blink. When the battery is depleted the "Erb" message is displayed and the tester is powered off. Care & Maintenance General Information Please read the information below, to ensure the highest possible accuracy: • Fresh buffer should be used for each calibration, once the sachets are opened the buffer value can change over time. • For improved accuracy a two-point calibration is recommended. • If the electrode is slow or sluggish, soak it in cleaning solution for 20 minutes. Rinse with water and hydrate the electrode in storage solution for a minimum of 30 minutes before calibrating. • If measurements are taken successively, rinse the probe thoroughly in distilled or deionized water to eliminate cross-contamination. • When not in use, add a few drops of storage solution to the protective cap to keep the glass tip and the junction hydrated. If storage solution is not available, pH 4.01 or pH 7.01 buffer can be used. Never store the probe in distilled or deionized water. • The outer junction sleeve can be removed and cleaned with purified water. Once cleaned a small amount of supplied gel electrolyte should be added to refresh the junction and improve the pH measurement. Refilling the Electrode • To remove the electrode sleeve, rotate it carefully and slide it off the probe body, keeping it parallel to the pH electrode. Note: Handle the probe with care, the pH electrode stem is made of glass. • Rinse off any traces of electrolyte gel from the sleeve. Soak the tip of the pH electrode in HI70061 General purpose cleaning solution (or HI700661, HI700663, HI700664) for 20 minutes. Rinse with distilled or deionized water. • Refill the reference well with HI9071 Gelled bridge electrolyte. • Replace the sleeve making sure that the black O ring is fixed inside the electrode. • Rinse off excess gel with distilled or deionized water. • Shake the probe down as you would do with a clinical thermometer to eliminate any trapped air bubbles. • Soak the probe in HI70300 Electrode storage solution for a minimum of 30 minutes before calibrating. Accessories pH Buffer Solution HI70004G HI70007G HI77400P GroLine pH 4.01 buffer solution, 20 mL sachet (25 pcs.) GroLine pH 7.01 buffer solution, 20 mL sachet (25 pcs.) pH 4.01 & 7.01 buffer solution, 20 mL sachet (10 pcs., 5 each) Electrode Cleaning Solution HI70061G HI700661P HI700663P HI700664P GroLine general purpose cleaning solution, 20 mL sachet (25 pcs.) General purpose cleaning solution for agriculture, 20 mL sachet (25 pcs.) Cleaning solution for soil deposits, 20 mL sachet (25 pcs.) Cleaning solution for humus deposits, 20 mL sachet (25 pcs.) Electrode Storage Solution HI70300G GroLine electrode storage solution, 20 mL sachet (25 pcs.) HI70300-023 GroLine electrode storage solution, 230 mL HI9072 Electrode storage solution, 13 mL dropper Electrode Fill Solution HI9071 Gelled bridge electrolyte Other Accessories HI721319 HI7051M Soil auger Soil preparation solution, 230 mL Battery Replacement INSTRUCTION MANUAL To change the CR2032 lithium-ion battery, turn the battery cover located on the back of the tester counterclockwise to unlock. Remove cover and replace the battery with positive (+) side facing out. Note: Only use the battery type specified in the manual. Old batteries should be disposed in accordance with local regulations. Warranty HI981030 is warranted for a period of one year against defects in workmanship and materials when used for its intended purpose and maintained according to instructions. This warranty is limited to repair or replacement free of charge. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered. If service is required, contact your local Hanna Instruments Office. If under warranty, report the model number, date of purchase, serial number and the nature of the problem. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization (RGA) number from the Technical Service department and then send it with shipping costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. Recommendations for Users Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any variation introduced by the user to the supplied equipment may degrade the tester’s performance. For yours and the tester’s safety do not use or store the tester in hazardous environments. Certification All Hanna Instruments conform to the CE European Directives. Disposal of Electrical & Electronic Equipment. The product should not be treated as household waste. Instead hand it over to the RoHS appropriate collection point for the recycling of electrical and compliant electronic equipment which will conserve natural resources. Disposal of waste batteries. This product contains batteries, do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling. Ensuring proper product and battery disposal prevents potential negative consequences for the environment and human health. For more information, contact your city, your local household waste disposal service, the place of purchase or go to www.hannainst.com. Hanna Instruments reserves the right to modify the design, construction, or appearance of its products without advance notice. All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA. IST981030 10/19 HI981030 Soil pH Tester Preliminary Examination Remove the tester and accessories from the packing material and examine it carefully. If you require any further information, please contact Hanna Instruments technical support team at [email protected] Each HI981030 is delivered in a cardboard box and is supplied with: • HI70004 pH 4.01 buffer solution, 20 mL sachet (2 pcs.) • HI70007 pH 7.01 buffer solution, 20 mL sachet (2 pcs.) • HI700663 Cleaning solution for soil deposits • HI700664 Cleaning solution for humus deposits • HI9072 Electrode storage solution, 13 mL dropper • HI9071 Gelled bridge electrolyte • CR2032 3V Lithium-ion battery • Instrument quality certificate • Instruction manual Note: Save all packing material until you are sure that the tester works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories. General Description & Intended Use The HI981030 Soil pH Tester is designed to measure the pH or acidity of the soil as a key factor in ensuring strong and healthy plant growth. It features a single button operation system and is easy to use. It has a compact and waterproof casing and automatic pH calibration at one or two points. All readings are automatically compensated for temperature variations with a built-in temperature sensor. Direct Soil Measurement Direct soil pH testing offers the benefit of not needing to take soil samples, as the pH is tested right in the ground. 1. Inspect the pH probe to ensure the bridge electrolyte gel has not dried out. Refill if necessary. 2. Calibrate the pH probe prior to measurement. 3. Using an auger (HI721319) or ruler, make a hole down into the soil (app. 20 cm). The hole needs to be the same depth each time to avoid pH discrepancies. 4. Add some distilled or deionized water to the hole; the soil should be damp but not saturated. 5. Wash the electrode with tap water and insert into the hole. 6. Allow for the reading to develop or stabilize. 7. After measurement, gently wash off leftover soil (avoid using a cloth or a wipe). Note: Direct soil pH testing is not recommended if the ground is stony or hardened as the electrode can be damaged. To prepare the soil prior to measurement, take a representative soil sample and add HI7051 Soil preparation solution. Conical Tip: The conical tip allows for easy penetration into damp soil. If the soil is hardened (or stony) it is best to use an auger (HI721319 ) to make a hole for the pH electrode to be inserted into. Open Junction Reference: The open junction design consists of a solid gel interface between the soil sample and the internal Ag/AgCl reference. This interface prevents silver from entering the sample and makes it impermeable to clogging, resulting in fast response and stable reading. Polyvinylidene Fluoride (PVDF) Body: PVDF is a food-grade plastic that is resistant to most chemicals and solvents, including sodium hypochlorite, it is also resistant to fungal growth. It has high abrasion resistance, mechanical strength, and resistance to ultraviolet. Removable Polyvinylidene Fluoride (PVDF) Reference Sleeve: The PVDF outer junction sleeve can be removed and cleaned. Once cleaned, a small amount of supplied gel electrolyte is added and the junction is refreshed, improving the measurement and extending the life of the tester. Specifications Range Resolution Accuracy Calibration Temperature compensation Electrode Battery type Battery life Auto-off Environment Dimensions Weight 0.00 to 12.00 pH 0.01 pH ±0.05 pH Automatic, one or two-point Automatic, 0 to 50 °C Built-in probe for specific application CR2032 Lithium-ion (included) Approximately 800 hours of continuous use 8 minutes, 60 minutes or disabled 0 to 50 °C (32 to 122 °F); RH 95% max 51 x 151 x 21 mm (2 x 5.9 x 0.9“) 46 g (1.62 oz.) Functional Description & LCD Display Liquid Crystal Display (LCD) ON/OFF button Maximum immersion level Calibration tags pH probe Recommended immersion level Preparation • Remove the protective cap. Do not be alarmed if salt deposits are present. Rinse the probe with water and blot dry. • If the glass and / or junction are dry soak the electrode (bottom 4 cm / 1.5 “) in storage solution for a minimum of 30 minutes.Rinse with water and blot dry. • Calibrate the electrode before using. For best results is recommended to recalibrate periodically. Calibration From measurement mode, press and hold the ON/OFF button until “CAL” is displayed. Storage • To ensure a quick response, the glass tip and junction should be kept moist. • Replace the protective cap with a few drops of storage solution when not in use. Do not store the electrode in distilled or deionized water. Operation Press the ON/OFF button to turn the meter on. The tester displays all LCD segments for a few seconds. The tester will enter measurement mode, the current reading and calibrated buffers will be shown. A When "7.01" is displayed, place the tip of the electrode in pH 7.01 buffer. When the reading is stable, the stability icon will disappear. When "4.01" is displayed, follow the procedure to the right for a one or two-point calibration. B Measurement unit Calibration tags Stability indicator A For one or two-point calibration using pH 7.01 buffer solution, follow procedure A. B For one-point calibration using pH 4.01 buffer solution, follow procedure B. Note: It is recommended to calibrate the electrode with buffers at the temperature it will be used at. Low battery indicator One or Two-Point Calibration with pH 7.01 One-Point Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the tester. For more information about Hanna Instruments and our products, visit www.hannainst.com or e-mail us at [email protected] For technical support, contact your local Hanna Instruments Office or e-mail us at [email protected] Probe Features Two-Point Dear Customer, Press the ON/OFF button to save the one-point calibration. "Sto" will be displayed when the calibration is saved. The tester will return to measurement mode and the calibration tag will be displayed. Place the tip of the electrode in pH 4.01 buffer. The buffer is automatically recognized, the stability indicator will blink. Wait until the measurement is stable and the stability indicator disappears. "Sto" will be displayed when the calibration is saved. The tester will return to measurement mode and the calibration tags will be displayed. When the reading is stable, the stability indicator will disappear.“Sto” will be displayed when the calibration is saved. The tester will return to measurement mode and the calibration tag will be displayed. One-Point Calibration with pH 4.01 When "7.01" is displayed, place the tip of the electrode in pH 4.01 buffer. The buffer value will be recognized automatically and "4.01" will be displayed.
Описание pH-тестера для для почвы HI981030:
HI981030 сконструирован для для упрощения измерений рН в почве путем прямых измерений.
pH-тестер HI981030 способен производить измерения в почвах различного типа, навозе, субстрате, глине и т.п. Благодаря съемной конструкции датчика с коническим наконечником использование измерителя pH почвы делает его идеальным в полевых условиях.
Отличительные особенности:
- Точность +/-0,2 pH
- Поставляется в комплекте с батарейками и всеми необходимыми растворами, достаточными на первое время.
- Специализированный электрод для точечного измерения рН в почве и жидких глинах.
Технические характеристики:
Параметр | Значение |
---|---|
Диапазоны | рН от 0.00 до 14.00 pH |
Разрешение рН | 0.1 pH |
Точность | +/- 0,2 pH при 25 ° C (77 ° F) |
Калибровка pH | Автоматически, в одной или двух точках |
pH температурная компенсация | НЕТ |
Автоматическое отключение | 8 минут, 60 минут (возможно отключение функции) |
Тип батареи/долговечность | CR2032 Li-ion / Приблизительно 1000 часов непрерывного использования |
Условия использования | От 0 до 50 ° C (от 32 до 122 ° F); Относительная влажность 95%, макс. |
Размеры | 51 х 151 х 21 мм (2 х 5,9 х 0,9 «) |
Вес | 44 г |