Medrad Spectris Solaris EP Operation Manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

loading

Related Manuals for Medrad Spectris Solaris EP

Summary of Contents for Medrad Spectris Solaris EP

  • Page 3
    The Netherlands Tel: +31 (0) 43-3585600 Fax:+31 (0) 43-3656598 © 2009-2012 MEDRAD, INC. All rights reserved. Reproduction of this manual is strictly prohibited without express written consent of MEDRAD, INC. For more information about MEDRAD products and services, please visit www.medrad.com…
  • Page 5: Table Of Contents

    TABLE OF CONTENTS 1 — Introduction………………….1-3 Important Safety Notice …………………. 1-3 Certifications ……………………1-3 Indications for Use ………………….1-3 Contraindications …………………… 1-3 Restricted Sale……………………1-3 Required Training ………………….. 1-3 Trademarks……………………. 1-3 Disclaimers ……………………. 1-3 The Equipotential Connector (EPC)………………1-4 Understanding Symbols…………………. 1-4 Warnings……………………..

  • Page 6
    MEDRAD Spectris Solaris EP MR Injection System Storing a Protocol ………………….3-33 Recalling a Stored Protocol ………………..3-34 4 — Arming and Injecting ………………4-35 Arming……………………..4-35 Single and Multi-Arm………………….4-35 Insufficient Volume………………….4-36 Injecting……………………..4-36 Disarming…………………….. 4-37 Injection History …………………… 4-39 Clean Up ………………………
  • Page 7: Introduction

    IEC 60601-1-1. To obtain on-site consulting or consulting references, contact MEDRAD Service. The MEDRAD Spectris Solaris EP MR Injection System is not intended for portable use. 1 — 3…

  • Page 8: The Equipotential Connector (Epc)

    The EPC’s function is to minimize any voltage potentials differences between all connected equipment. The EPC is not designed to be an electrical safety ground. Understanding The following symbols are used on the MEDRAD Spectris Solaris EP Mobile Symbols MR Injection System and components.: Attention, consult accompanying instructions.

  • Page 9
    1 — Introduction Indicates the current charge level of the system battery. Identifies Integrated Continuous Battery Charger system activity on Graphical User Interface. When illuminated yellow this indicates that the Continuous Battery Charger system is present and functioning. Indicates the AIR EXPELLED button on the injector head. When illuminated yellow on the touch screen, also indicates that the operator has acknowledged inspecting the fluid path for air.
  • Page 10: Warnings

    This manual contains important information about use of the MEDRAD Spectris Solaris EP MR Injection System. MEDRAD urges you to read this manual carefully, become familiar with the procedures and system functions that it describes, and follow its recommendations to assure proper use of the system.

  • Page 11
    Contact MEDRAD for service or replacement. The MEDRAD Spectris Solaris EP MR Injection System is a dual syringe system. Always ensure that the proper syringes are loaded with contrast media and flush solution prior to the injection. Failure to properly load and install the syringes may require the procedure to be repeated.
  • Page 12: Cautions

    Third Edition Standards. Special precautions regarding ElectroMagnetic Compatibility (EMC), are required for installation and use of this injector system. Detailed EMC information can be found in the MEDRAD Injector Service Manual — Addendum, (label number: 202559). Damage can occur as a result of incorrect voltage. Before plugging in the system, check the following: •…

  • Page 13: System Basics

    2 — System Basics 2 — System Basics About the Injection The MEDRAD Spectris Solaris EP MR Injection System is a programmable, System dual syringe system, designed to accurately administer controlled doses of intra-venous MR contrast agents and common flushing solutions to patients undergoing a contrast enhanced MR scan.

  • Page 14: Pressure Safety Limit

    MEDRAD Spectris Solaris EP MR Injection System Pressure Safety The MEDRAD Spectris Solaris EP MR Injection System is designed to allow Limit varied flow rates for contrast injections. By automatically reducing the flow rate, the system can limit the pressure produced during an injection to prevent damage or failure of any connecting devices or tubing.

  • Page 15: Control Room Unit

    2 — System Basics Control Room Unit 1. Handswitch 2. System Power Switch 3. Touch Screen At rear of Touch Screen Assembly — Display Contrast Controls 2 — 11…

  • Page 16: Scan Room Unit

    MEDRAD Spectris Solaris EP MR Injection System Scan Room Unit 1. Injector Head 2. Handswitch 3. Lower Console 4. System Battery Pack 5. Middle Pivot Clamp Not shown — Contrast Holder (optional) 2 — 12…

  • Page 17: Injector Head

    2 — System Basics Injector Head 1. Manual piston movement knobs 2. Armed indicator lights 3. ENABLE button — Used to activate the forward/reverse controls — the appropriate direction must be selected within 5 seconds. 4. Syringe A forward/reverse controls 5.

  • Page 18: Battery Charger

    MEDRAD Spectris Solaris EP MR Injection System Battery Charger 1. Battery Pack 2. Battery Charging Unit 3. Charging Indicator — Amber 4. Power Indicator — Green 5. Battery Charger Head 2 — 14…

  • Page 19: Optional Control Room Unit Accessories

    2 — System Basics Optional Control Room Unit Accessories Adjustable Height Pedestal WARNING: Injury or equipment damage may result if the adjustable height pedestal is taken into the scanner room. Do not take the adjustable height pedestal in the scanner room. It contains ferrous material that could be attracted toward the magnet.

  • Page 20: Touch Screen Calibration

    MEDRAD Spectris Solaris EP MR Injection System Touch Screen To enter Touch Screen Calibration mode, simultaneously press both the Calibration Contrast UP and DOWN keys on the rear of the touch screen housing. A series of screens with instructions to press the appropriate calibration circles will appear.

  • Page 21: Setup Mode

    2 — System Basics Setup Mode The Setup screen can be accessed by pressing the SETUP button at the lower right corner of the Main screen. The Setup screen allows user configurable options and preferences to be selected, along with setting of date and time parameters.

  • Page 22
    MEDRAD Spectris Solaris EP MR Injection System 2 — 18…
  • Page 23: Preparing To Inject

    3 — Preparing to Inject 3 — Preparing to Inject Applying Power Place the power switch located on the right side of the Control Room Unit in the ON position. The System Logo screen will appear while the system performs a series of self diagnostic tests. Note: Do not touch the screen or activate any controls while self diagnostics are in progress.

  • Page 24: Main Screen

    MEDRAD Spectris Solaris EP MR Injection System Main Screen The Main screen is entered from the Safety screen after power-up. The Main screen is used during programming, arming, and injecting, with applicable screen controls made visible based on the task currently being performed.

  • Page 25: Battery Maintenance

    WARNING: Explosion Hazard. Serious injury or death may result from improper use of the battery charger. The battery charger, MEDRAD Catalog Number 3012424, is intended for use in a well ventilated area, with the injection system battery, MEDRAD Catalog Number 3012070, only. Do not use the charger with non-rechargeable batteries.

  • Page 26: Syringe And Disposable Accessory Installation

    MEDRAD Spectris Solaris EP MR Injection System Syringe and Disposable Accessory Installation Retracting the Fully retract each piston by using the reverse switches on the injector head. Pistons Note: When using the reverse switches, first press the Enable switch; then within 5 seconds, press the reverse switch(es).

  • Page 27: Installing A Syringe

    3 — Preparing to Inject Installing a Syringe WARNINGS: Patient or operator injury may result if damaged components are used. Do not use damaged components. Visually inspect all components before use. Patient infection may result from the use of non-sterile components.

  • Page 28
    To help avoid air injection, MEDRAD syringes are equipped with FluiDot indi- cators. These FluiDot indicators should be observed as part of an arming pro- cedure.
  • Page 29: Loading A Syringe

    ENABLE. 3. Attach a sterile filling device (spike or Female-to- Female Adaptor — MEDRAD Catalog Number FFA 50) onto the tip of the syringe. If loading contrast media and/or saline from a bag or bottle, use a spike.

  • Page 30
    MEDRAD Spectris Solaris EP MR Injection System 6. Attach the long end of the T-connector to syringe B. 7. While the injector head is still in a vertical position, attach the short section of the T- connector to syringe A.
  • Page 31: Reinstalling A Syringe

    3 — Preparing to Inject Reinstalling a WARNING: Patient injury could result if the syringe is not Syringe properly engaged. Ensure the alignment marks on the syringe and injector head are properly aligned, and the piston and plunger are interlocked. Improper engagement may cause syringe damage or under-volume delivery.

  • Page 32: Programming

    MEDRAD Spectris Solaris EP MR Injection System Programming If a program has not been previously entered or stored on the Main screen when the unit is powered up, the Main screen will display default settings; 1.0 ml/s flow rate and 1.0 ml volume, KVO off and No Delay.

  • Page 33: Multiple Phases

    3 — Preparing to Inject Multiple Phases If appropriate, select a second phase for the injection protocol by pressing the triangle block below the first phase of the injection. The Phase Type selector will appear in order to select the function of the new phase. Hold and Pause A Hold or Pause phase can be programmed into a multi-phase injection.

  • Page 34: Programmed Delay

    MEDRAD Spectris Solaris EP MR Injection System Programmed Delay After entering Flow Rate and Volume parameters, press SET in the Delay Timer field to select the delay type (Scan Delay, Inject Delay, Stopwatch, or No Delay.) Note: There is no direct interface between the scanner and the injector.

  • Page 35: Stopwatch

    3 — Preparing to Inject Stopwatch The Stopwatch function initiates an incremental count of elapsed time from initial fluid injection. After selecting the delay type, enter the delay duration on the numeric keypad. To lock in values, press ENTER. Press CANCEL to eliminate a selection if an error is made.

  • Page 36
    MEDRAD Spectris Solaris EP MR Injection System The KVO field displays the time available to support KVO based on the configured interval and the volume remaining in syringe B less any volume programmed from syringe B in the protocol. Starting KVO: On the Main Screen, press START in the KVO field to initiate KVO.
  • Page 37: Storing A Protocol

    3 — Preparing to Inject Storing a Protocol To store a protocol for future use, press the STORE button on the upper right corner of the Main screen. An alpha-numeric keypad will appear with a flashing cursor in the title block. Type in a title of up to 20 characters, including spaces.

  • Page 38: Recalling A Stored Protocol

    MEDRAD Spectris Solaris EP MR Injection System Recalling a Stored To access program memory, press RECALL on the Main screen. Protocol Select a previously stored injection protocol by pressing one of the names on either side of the screen. Key parameters of the selected injection will be displayed in the center of the screen.

  • Page 39: Arming And Injecting

    4 — Arming and Injecting 4 — Arming and Injecting Before beginning the arming process, ensure that the casters on the Scan Room Unit are locked, verify that all air has been expelled from the fluid path, and that the programmed parameters are correct. Carefully inspect all tubing and syringe(s), then acknowledge that the inspection has occurred by pressing the AIR EXPELLED button/indicator on the injector head.

  • Page 40: Insufficient Volume

    MEDRAD Spectris Solaris EP MR Injection System Insufficient Volume If an insufficient volume condition occurs during a multi arm sequence, the system will remain armed to permit the injection of the remaining volume. However, the screen will update to display only the phases that are achievable with the volume that remains.

  • Page 41: Disarming

    4 — Arming and Injecting On the Injecting Screen: • As each phase is activated, the phase parameters will be highlighted to display injection progress. • The Duration window will also increment to display elapsed time. • The Delivered window will increment as the injection proceeds to display volume delivery (including KVO volume, if selected in Setup mode).

  • Page 42
    MEDRAD Spectris Solaris EP MR Injection System Note: A MEDRAD SSIT 96VLD low pressure connector tube (LPCT) holds approximately 7 ml of fluid. If syringe B is used to flush, use at least 8 ml of flush to deliver this volume to the patient.
  • Page 43: Injection History

    4 — Arming and Injecting Injection History To review injection parameters used in a procedure, along with actual achieved values for the injection, press the HISTORY button on the Main screen. The Injection History screen displays an injection summary block containing the following data: •…

  • Page 44: Clean Up

    System malfunction may be caused by failure to perform regular maintenance. Regular preventive maintenance is recommended to ensure that the system stays calibrated and functions properly. Refer to Appendix B of this manual or contact MEDRAD for additional information. Do not expose system components to excessive amounts of water or cleaning solutions.

  • Page 45
    4 — Arming and Injecting Scan Room Unit Using a soft non-abrasive cloth, warm water, and a mild disinfectant, carefully clean the assembly, paying particular attention to the following: • Injector Head • Syringe Piston Plunger • Syringe Interface • SRU Lower Console covers To clean the injector head, piston, and syringe interface: 1.
  • Page 46
    MEDRAD Spectris Solaris EP MR Injection System 4 — 42…
  • Page 47: Appendix A: System Messages

    If a fault message is displayed that cannot be corrected, and/or the system is not operating correctly, do not use the injection system. Call MEDRAD for assistance. Type 1 Messages Type 1 messages are messages which provide information regarding the current status of the system, and will clear automatically from the screen.

  • Page 48: Type 2 Messages

    MEDRAD Service for assistance. System Error Detected Immediately disconnect the patient from the injector. Record the code below and contact MEDRAD Service. Symcode = HSW Errnum = -2600 Refer to the operation manual or www.Medrad.com for contact information. A — 44…

  • Page 49: Appendix B: Maintenance And Checkout

    Ensure continued performance of the injection system • Reduce the possibility of equipment malfunction Recommended Your MEDRAD Spectris Solaris EP MR Injection System must be properly Maintenance maintained to ensure that it is in peak operating condition. Your individual Schedule maintenance system and schedule depends upon how your injection system is used, the type of procedures performed, and frequency of use.

  • Page 50: Inspection Procedures

    Inspection Procedures The following procedures are recommended for daily inspection of all components in the MEDRAD Spectris Solaris EP MR Injection System. If any defects are detected, either repair the system, or call MEDRAD for service. Do not use the system until the problem is corrected.

  • Page 51
    Appendix B: Maintenance and Checkout Wall Mount Bracket 1. Inspect all parts of the bracket for cracks and other defects that would weaken the assembly. 2. Ensure that the bracket is securely attached to the wall. 3. Ensure that all cables are secured to the display control unit and do not interfere with the movement of the mounting bracket.
  • Page 52: Cleaning Guidelines

    Operational Checkout A basic functional checkout of the MEDRAD Spectris Solaris EP MR Injection System should be included as part of regular maintenance. Verifying proper operation of the injection system will help in detection of any problems that may not be noticed in day to day operation.

  • Page 53
    Appendix B: Maintenance and Checkout System Labels Ensure that all system safety and warning labels are in place and legible. Power Up Apply power to the system. Verify that the Safety screen is displayed after system diagnostics occur. Press OK to acknowledge the messages on the Safety screen.
  • Page 54
    MEDRAD Spectris Solaris EP MR Injection System Verify that when the injection completes, KVO resumes. Remove and discard the syringe. Remove power from the system. B — 50…
  • Page 55: Scan Room Unit

    Appendix C: Specifications Appendix C: Specifications Scan Room Unit Weight: 60 lbs. (27.3 kg.) 18.5″ 46.99 cm 12.75″ / 32.28 cm 51.75″ 131.45 cm 21.75″ 19.0″ 55.25 cm 48.26 cm C — 51…

  • Page 56: Control Room Unit

    MEDRAD Spectris Solaris EP MR Injection System Control Room Unit Weight: 15 lbs. (6.8 kg.) 11.995″ 30.46 cm 11.80″ 29.97 cm 2.375″ 6.03 cm 10.922″ 27.91 cm 9.25″ 23.5 cm 2.50″ 6.35 cm Battery Dimensions Weight: 7.7 lbs. (3.5 kg.) 2.1″…

  • Page 57: Battery Charger

    Appendix C: Specifications Battery Charger Weight: 2 lbs. (0.9 kg.) 2.765″ 3.05″ 7.02 cm 7.74 cm 4.664″ 5.5 ft 5.08″ 11.84 cm 1.68 m 12.90 cm 1.114″ 1.59″ 2.90 cm 4.03 cm Power Cords American 12 ft. (3.6 m), Continental 9.8 ft. (3 m) C — 53…

  • Page 58: System Capabilities

    Executable Flow These Flow Rates, with pressure safety limit set to 325psi/2240kPa, are Rates achievable with the MEDRAD Spectris Solaris EP MR Injection System, using the Becton Dickinson catheters listed below and the SSQK 65/115VS MEDRAD Syringe/Disposables Kit. 18 g IV Catheter 20 g IV Catheter 22 g IV Catheter 24 g IV Catheter…

  • Page 59: System Performance

    Appendix C: Specifications Multihance, Gadovist, Magnevist, Optimark, Prohance, and Omniscan are trademarks of their respective companies. System Performance Volume Accuracy: Syringe A: +/- (1% + 0.1 ml) Syringe B: +/- (5% + 0.1 ml) Flow Rate Accuracy +/- (10% + 0.005 ml/s) when rate is 0.01 to 0.99 ml/s +/- (10% + 0.02 ml/s) when rate is 1 to 10 ml/s Programmed Delay/ +/- (5% + 0.2 second)

  • Page 60: Emi/Rfi

    MEDRAD Spectris Solaris EP MR Injection System EMI/RFI The MEDRAD Spectris Solaris EP MR Injection System is designed to be in compliance with IEC/EN 60601-1 Second/Third Edition, EN 60601-1-2 Second Edition and IEC 60601-1-2 Second/Third Edition. Electrical 100-240 VAC Requirements…

  • Page 61: Classifications

    Appendix C: Specifications Classifications Protection Against Electrical Shock: Per IEC/EN 60601-1, the MEDRAD Spectris Solaris EP MR Injection System is designed as Class 1 equipment with Type BF applied parts. Type BF corresponds to the degree of protection against electrical shock via the applied parts.

  • Page 62
    MEDRAD Spectris Solaris EP MR Injection System C — 58…
  • Page 63: Appendix D: Options And Accessories

    Appendix D: Options and Accessories Appendix D: Options and Accessories Catalog Number Part Number Power Cord American SPC 300A 535-0243-012 Continental SPC 300C 535-0127-012 Integrated Continuous Battery Charger 3012080 3012080 System Battery Charger Kit 3012424 3012424 Enhanced Battery Pack 3012070 3012070 Handswitch SSMR START…

  • Page 64
    MEDRAD Spectris Solaris EP MR Injection System D — 60…
  • Page 65: Appendix E: System Installation

    Plug the system directly into a properly grounded AC outlet or contact MEDRAD for installation assistance. Injury or equipment damage may result from improper placement of the Battery Charger. Do not install the Battery Charger in the Scan Room.

  • Page 66: Unpacking The Injection System

    MEDRAD Spectris Solaris EP MR Injection System Unpacking the The entire standard configuration of the MEDRAD Spectris Solaris EP MR Injection System Injection System is shipped in one shipping carton. The optional Control Room Unit mounting accessories, the Adjustable Height Stand and the Wall Mounting Bracket, along with the optional IV Pole, are packaged individually and shipped in separate cartons.

  • Page 67: Installation Considerations

    Appendix E: System Installation Installation Considerations WARNING: Injury or equipment damage may result from use of tools containing ferrous materials. Use only non-magnetic tools to install any scanner/magnet room components. NOTE: System Installation requires that the suite have a 1.5 inch (3.81 cm) minimum tuned port (either separate or within the penetration panel) available for connections between the Scan and Control rooms.

  • Page 68
    MEDRAD Spectris Solaris EP MR Injection System Review the following general connections prior to installing the MEDRAD Spectris Solaris EP MR Injection System. Be sure to consider all specifica- tions and requirements outlined in Appendix C of this manual, and follow all applicable regulations of your locality.
  • Page 69: Fiber Optic Cable Installation

    Appendix E: System Installation Fiber Optic Cable Connection at the Scan Room Unit: Installation Special care should be taken with routing the fiber optic cable to ensure: • there is no bend radius less than 1 in. (2.54 cm) • the connector dust caps are not removed until connections are made •…

  • Page 70: Control Room Unit Setup

    MEDRAD Spectris Solaris EP MR Injection System Strain Relief The strain relief should be placed 17″ (43 cm) inches form the end of the Fiber Location optic connectors. This ensures there is sufficient length to maneuver the cable and minimize severe cable bends, which may damage the Fiber Optic cable.

  • Page 71: Handswitch Mounting

    Appendix E: System Installation the equal potential stud, and the equal potential bus. 6. Plug the AC power cord into an appropriate AC power outlet. 7. Insert a fully charged system battery in the Scan Room Unit battery receptacle 8. Apply system power at the Control Room Unit power switch, then perform an system operation checkout as outlined in Appendix B of this manual.

  • Page 72
    MEDRAD Spectris Solaris EP MR Injection System E — 68…

  • Contents

  • Table of Contents

  • Bookmarks

  • ENGLISH, page 3

  • РУССКИЙ, страница 312

  • FRANÇAIS, page 74

  • ESPAÑOL, página 120

  • DEUTSCH, seite 51

  • ITALIANO, pagina 97

  • DUTCH, pagina 168

  • PORTUGUÊS, página 144

  • SVENSKA, sida 194

  • POLSKI, strona 504

  • DANSK, side 216

  • SUOMI, sivu 264

  • ČEŠTINA, strana 408

  • NORSK, side 242

  • ΕΛΛΗΝΙΚΆ, σελίδα 288

  • MAGYAR, oldal 480

  • SLOVENČINA, strana 338

  • TÜRKÇE, sayfa 384

  • ROMÂNĂ, pagina 530

  • SLOVENŠČINA, stran 552

  • 汉语, 第 434 页

  • БЪЛГАРСКИ, страница 360

  • 日本語, 32ページ

  • HRVATSKI, stranica 456

Quick Links

Operation Manual

Integrated Continuous Battery Charging System

Language Key

English

1–24

25–48

Deutsch

49–71

Français

72–94

Italiano

95–118

Español

119–142

Português

143–166

Nederlands

167–190

Svenska

191–214

Language Key

Dansk

215–238

Norsk

239–262

Suomi

263–286

Ελληνικά

287–310

Русский

311–334

Slovenský

335–358

Български

359–382

Türkçe

383–405

Česky

407–430

Language Key

431–454

Hrvatski

455–478

Magyar

479–502

Polski

503–526

Română

527–550

Slovenščina

551–574

Chapters

loading

Summary of Contents for Medrad Spectris Solaris EP

Table of Contents for Medrad Spectris Solaris EP:

  • Sistema integrado de carga continua de batería 122

  • Integrerede Kontinuerlige Batteriopladningssystem 234 Reservedelsliste for ekstraudstyr iCBC Efter udskiftning af en samling eller et kabel anbefaler Bayer, at følgende grundlæggende trin udføres for at bekræfte korrekt funktion. 1. Med alle kabler fastgjort skal det sikres, at det Integrerede Kontinuerlige Batteriopladningssystems statusindikator på kontrolrummenhedens display viser, at batteriopladeren er driftsklar. Kabel Strømforsyning Netledning Katalognummer Netledning Kina 3013369 Brasilien 3029224 Ledninger 6,1 m. M/ferrit 3009961 Filte

  • Entegre Sürekli Batarya Şarj Edici Sistemi 398 Güç Kablosu Özellikleri MEDRAD Spectris Solaris’in Entegre Sürekli Batarya Şarj Edici Sistem ile kullanma konusunda güç kablosuna (fiş, muhafaza ve kablo) ilişkin gereklilikler şunlardır: •Çalıştırma Sıcaklığı: minimum 60°C • Muhafaza Tipi: IEC-60320 C13 • Normal Kablo Voltajı: minimum 300 VAC •Tel Çapı: minimum 1.00 mm 2 • Kablo Tipi: IEC 60245-1, Ek A, Tahsis 53 veya IEC 60227-1, Ek A, Tahsis 53 Sertifikalı • Kablo

  • 1 Innledning 243 1 Innledning Denne håndboken omhandler innebygde, kontinuerlige batteriladersystem til bruk med MEDRAD ® Spectris Solaris og MEDRAD ® Spectris Solaris EP MR injeksjonssystemer. Les all informasjonen i dette kapittelet. God forståelse av informasjonen vil hjelpe deg å betjene apparatet på en trygg måte. MEDRAD Spectris Solaris MR injeksjonssystem må ha MEDRAD Spectris Solaris EP strømoppgradering for å kunne bruke innebygde, kontinuerlige batterilader

  • 4Instalace 429 6. Zapojte zdroj napájení do sítě. 7. Vložte baterii do jednotky pro vyšetřovnu. 8. Zkontrolujte, zda je jednotka pro vyšetřovnu zapnutá a ověřte, zda stavový indikátor integrovaného systému nepřetržitého napájení baterie na displeji jednotky pro řídicí místnost ukazuje, že je nabíječka pro nepřetržité nabíjení baterie v provozu. Kruhový spojovací otvor (použijte spojovací desku) Pokud na spojovacím panelu nejsou volné spojovací otvory pro připojení konektoru, použijte kruhový spojovací otvor. 1. Um

  • Kiinteä jatkuvatoiminen akun latausjärjestelmä 276

  • 1 Wstęp 509 Zakres temperatur. Zakres wilgotności. Zakres ciśnień atmosferycznych. Góra. Delikatne. Przechowywać w suchym miejscu. Medyczne – ogólne urządzenia medyczne Dotyczy jedynie ryzyka porażenia prądem elektrycznym, zagrożenia pożarem lub ryzyka mechanicznego Zgodnie z ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012) CAN/CSA-C22.2 nr 60601-1 (2014). Maksymalna waga systemu wstrzykiwania i akcesoriów w

  • 4 Instalación 139 1. Con todos los cables conectados, compruebe que el indicador de estado del Sistema integrado de carga continua de batería situado en la pantalla de la unidad de la sala de control indica que el sistema de carga continua de batería está en funcionamiento. 2. Con el juego de baterías enchufado en la unidad de la sala de exploración, la batería aparece en la pantalla de la unidad de la sala de control como “Completamente cargada”. 3. Antes de enviar los componentes de repuesto, se comprueban pos

  • 1 Introduzione 101 Range temperatura Range umidita Range pressione atmosferica Alto Fragile Tenere all’ asciutto Apparecchiature medicali — mediche generali Solo per scosse elettriche, incendi e rischi meccanici In conformità alla norma ANSI/AAMI ES60601-1 (2005) + AMD 1 (2012) CAN/CSA-C22.2 n. 60601-1 (2014) Peso massimo del sistema di iniezione e degli accessori durante il normale utilizzo: SSEP, Peso totale: 35 kg Piedistallo CRU SSEP

  • 479 Magyar Gyártó Bayer Medical Care Inc. 1 Bayer Drive Indianola, PA 15051-0780 Amerikai Egyesült Államok Telefonszám: +1.412.767.2400 +1.800.633.7231 Fax: +1.412.767.4120 Hivatalos európai képviselő Bayer Medical Care B.V. Horsterweg 24 6199 AC Maastricht Airport Hollandia Telefonszám: +31 (0) 43-3585601 Fax: +31 (0) 43-3656598 © 2009, 2012, 2014, 2016-2017 Bayer. A dokumentumban szereplő információk nem sokszorosíthatók, jeleníthetők meg, módosíthatók vagy adhatók át egyéb feleknek a Bayer vállalat előzetes é

  • Zintegrowany system ciągłego ładowania baterii 508 Znaczenie symboli Na zintegrowanym systemie ciągłego ładowania baterii oraz jego elementach zastosowano następujące symbole. Uwaga: Należy zapoznać się z ostrzeżeniami i przestrogami zawartymi w instrukcji obsługi znajdującej się w każdym kartonie. Oznacza zgodność urządzenia z wymaganiami europejskiej Dyrektywy dotyczącej wyrobów medycznych 93/42/EWG. Wskazuje napięcia niebezpieczne. Wskazuje p

Questions, Opinions and Exploitation Impressions:

You can ask a question, express your opinion or share our experience of Medrad Spectris Solaris EP device using right now.

Medrad Spectris Solaris EP Operation Manual | Manualzz

North America Headquarters
MEDRAD Service Department
One Medrad Drive
Indianola, PA 15051-0780
U.S.A.
Tel: 412-767-2400
Fax: 412-767-4120
Nihon MEDRAD KK
2-4-9, Umeda, Kita-ku,
Osaka, 530-0001
Japan
Tel: +81(0)66-133-6250
Fax:+81(0)66-344-2395
European Headquarters
MEDRAD Europe B.V.
P.O. Box 205
6190 AE Beek
The Netherlands
Tel: +31 (0) 43-3585600
Fax:+31 (0) 43-3656598
© 2009-2012 MEDRAD, INC. All rights reserved.
Reproduction of this manual is strictly prohibited without express written
consent of MEDRAD, INC.
For more information about MEDRAD products and services, please visit
www.medrad.com
TABLE OF CONTENTS
1 - Introduction......................................................................................................1-3
Important Safety Notice ...................................................................................................... 1-3
Certifications ....................................................................................................................... 1-3
Indications for Use .............................................................................................................. 1-3
Contraindications ................................................................................................................ 1-3
Restricted Sale.................................................................................................................... 1-3
Required Training ............................................................................................................... 1-3
Trademarks......................................................................................................................... 1-3
Disclaimers ......................................................................................................................... 1-3
The Equipotential Connector (EPC).................................................................................... 1-4
Understanding Symbols...................................................................................................... 1-4
Warnings............................................................................................................................. 1-6
Cautions.............................................................................................................................. 1-8
2 - System Basics .................................................................................................2-9
About the Injection System ................................................................................................. 2-9
Pressure Safety Limit........................................................................................................ 2-10
Response to Occlusions ................................................................................................... 2-10
Volume and Rate Protection ............................................................................................. 2-10
Control Room Unit ............................................................................................................ 2-11
Scan Room Unit................................................................................................................ 2-12
Injector Head..................................................................................................................... 2-13
Battery Charger................................................................................................................. 2-14
Optional Control Room Unit Accessories.......................................................................... 2-15
Touch Screen Calibration ................................................................................................. 2-16
Help Mode......................................................................................................................... 2-16
Setup Mode....................................................................................................................... 2-17
3 - Preparing to Inject .........................................................................................3-19
Applying Power ................................................................................................................. 3-19
Main Screen...................................................................................................................... 3-20
Battery Maintenance ......................................................................................................... 3-21
Syringe and Disposable Accessory Installation ................................................................ 3-22
Retracting the Pistons................................................................................................ 3-22
Installing a Syringe .................................................................................................... 3-23
Loading a Syringe...................................................................................................... 3-25
Reinstalling a Syringe ................................................................................................ 3-27
Programming .................................................................................................................... 3-28
Flow Rate and Volume .............................................................................................. 3-28
Pressure Limit ............................................................................................................ 3-28
Multiple Phases ......................................................................................................... 3-29
Hold and Pause Phases ............................................................................................ 3-29
Programmed Delay .................................................................................................... 3-30
Scan Delay ......................................................................................................... 3-30
Inject Delay ......................................................................................................... 3-30
Stopwatch ........................................................................................................... 3-31
KVO (Keep Vein Open) ............................................................................................. 3-31
i
MEDRAD Spectris Solaris EP MR Injection System
Storing a Protocol ............................................................................................................. 3-33
Recalling a Stored Protocol .............................................................................................. 3-34
4 - Arming and Injecting .....................................................................................4-35
Arming............................................................................................................................... 4-35
Single and Multi-Arm......................................................................................................... 4-35
Insufficient Volume............................................................................................................ 4-36
Injecting............................................................................................................................. 4-36
Disarming.......................................................................................................................... 4-37
Injection History ................................................................................................................ 4-39
Clean Up ........................................................................................................................... 4-40
Appendix A: System Messages........................................................................ A-43
Type 1 Messages..............................................................................................................A-43
Type 2 Messages..............................................................................................................A-44
Type 3 Messages..............................................................................................................A-44
Appendix B: Maintenance and Checkout ....................................................... B-45
Recommended Maintenance Schedule ............................................................................B-45
Inspection Procedures ......................................................................................................B-46
Cleaning Guidelines..........................................................................................................B-48
Operational Checkout .......................................................................................................B-48
Appendix C: Specifications .............................................................................. C-51
Scan Room Unit............................................................................................................... C-51
Control Room Unit ........................................................................................................... C-52
Battery Dimensions.......................................................................................................... C-52
Battery Charger................................................................................................................ C-53
Power Cords .................................................................................................................... C-53
System Capabilities ......................................................................................................... C-54
Executable Flow Rates .................................................................................................... C-54
System Performance ....................................................................................................... C-55
Forward and Reverse Controls ........................................................................................ C-55
EMI/RFI............................................................................................................................ C-56
Electrical Requirements ................................................................................................... C-56
Power Supply DC Output Voltage.................................................................................... C-56
Electrical Leakage............................................................................................................ C-56
Ground Continuity ............................................................................................................ C-56
Environmental Specifications........................................................................................... C-56
Classifications .................................................................................................................. C-57
Appendix D: Options and Accessories............................................................ D-59
Appendix E: System Installation ...................................................................... E-61
Unpacking the Injection System........................................................................................E-62
Installation Considerations................................................................................................E-63
Fiber Optic Cable Installation............................................................................................E-65
Control Room Unit Setup ..................................................................................................E-66
Handswitch Mounting........................................................................................................E-67
ii
1 - Introduction
1 - Introduction
This manual applies to the MEDRAD Spectris Solaris® EP MR Injection
System, Catalog Number 3012011. Read all of the information contained in
this section. Understanding the information will assist you in operating the
device in a safe manner.
Important Safety
Notice
This device is intended to be used by medical professionals with adequate
training and experience in magnetic resonance imaging (MRI) studies.
Certifications
This device is equipped to operate at 100-240 VAC, 50/60 Hz, 180 VA
(Single), and is designed to comply with EN 60601-1/IEC 60601-1 Second/
Third Edition, and EN 60601-1-2 Second Edition and IEC 60601-1-2 Second/
Third Edition Standards.
Indications for Use
This system is intended for the purposes of injecting intravenous MR contrast
media and common flushing solutions into the human vascular system for
diagnostic studies in magnetic resonance imaging (MRI) procedures.
Contraindications
This device is not to be used in the arterial side of the vascular system, for
drug infusion, chemotherapy, or any other use for which the device is not
indicated. The system should not be used with a magnetic resonance imaging
scanner having a magnetic field strength greater than 3.0 Tesla.
Restricted Sale
Federal (USA) law restricts this device to sale by or on the order of a
physician.
Required Training
This device is intended to be used by individuals with adequate training and
experience in diagnostic image studies.
Trademarks
MEDRAD, FluiDot, Qwik-Fit, Spectris Solaris, MEDRAD Radiology,
Performance for Life are federally registered trademarks of MEDRAD, INC.
The trademarks Becton Dickinson, Daiichi, NSKK, Multihance, Gadovist,
Magnevist, Optimark, Prohance, and Omniscan appear in this manual, and
are the property of their respective companies.
Disclaimers
External wiring and modifications disclaimers: MEDRAD disclaims liability for
any modifications or interfaces with other equipment which are not in
conformity with the specifications and information contained within this
manual.
Accessory equipment connected to the device must be certified according to
IEC 60601-1 Second/Third Edition standard. Furthermore, all configurations
shall comply with system standard EN 60601-1/IEC 60601-1-1. Anyone who
connects additional equipment to the signal input or output part configures a
medical system and is therefore responsible that the system complies with the
requirements of the standard IEC 60601-1-1. To obtain on-site consulting or
consulting references, contact MEDRAD Service.
The MEDRAD Spectris Solaris EP MR Injection System is not intended for
portable use.
1-3
MEDRAD Spectris Solaris EP MR Injection System
The Equipotential
Connector (EPC)
The Equipotential Connector (EPC) is an electrically bonded terminal on the
injector that is used as a connection point between other medical electrical
equipment. The EPC’s function is to minimize any voltage potentials
differences between all connected equipment. The EPC is not designed to be
an electrical safety ground.
Understanding
Symbols
The following symbols are used on the MEDRAD Spectris Solaris EP Mobile
MR Injection System and components.:
Attention, consult accompanying instructions.
Indicates that this device conforms to the requirements of the European
Medical Device Directive 93/42/EEC.
Indicates on/off switch for the Control Room Unit.
Indicates hazardous voltages.
Indicates alternating current.
Identifies a type BF applied part complying with EN 60601-1 standards.
CLASS 1
IPX1
Indicates the injection system is Class 1 medical equipment as defined by
EN 60601-1 standards.
Identifies the degree of protection against fluid as drip proof for the
Spectris Solaris EP Injector system.
Identifies connection of the handswitch.
Identifies injector head forward and reverse piston control keys.
Identifies the direction of manual knob rotation relative to plunger
movement.
Identifies the ENABLE key.
Identifies polarity of the battery pack terminals.
Indicates DC power supply.
1-4
1 - Introduction
Indicates the current charge level of the system battery.
Identifies Integrated Continuous Battery Charger system activity on
Graphical User Interface. When illuminated yellow this indicates that the
Continuous Battery Charger system is present and functioning.
Indicates the AIR EXPELLED button on the injector head. When
illuminated yellow on the touch screen, also indicates that the operator has
acknowledged inspecting the fluid path for air.
Identifies the Equipotential connection.
Identifies the Earth Ground point.
IOIO
Identifies the Service Connection Port.
Identifies the Locking Bracket. Indicates which direction to turn Locking
Bracket knob to “lock” and “unlock” the bracket.
TX
Identifies the Communication Cable Transmit connection.
RX
Identifies the Communication Cable Receive connection.
Indicates design for indoor use only.
Identifies the Integrated Continuous Battery Charger System power supply
connection.
Indicates the presence of no serviceable parts.
I
Indicates the presence of AC power at the battery charger.
Identifies the Control Room Unit brightness controls.
P109
Reserved for future use.
Indicates the status of the battery charger. When a battery is properly
inserted, the LED will illuminate while charging, and extinguish when the
battery is fully charged.
Pushing Prohibited. Do not push at or above this point on the Injector.
1-5
MEDRAD Spectris Solaris EP MR Injection System
This manual contains important information about use of the MEDRAD
Spectris Solaris EP MR Injection System.
MEDRAD urges you to read this manual carefully, become familiar with the
procedures and system functions that it describes, and follow its
recommendations to assure proper use of the system.
Labels on the system or statements in this manual preceeded by any of the
following words and/or symbols are of special significance, intended to help
you to operate the system in a safe and successful manner:
WARNING: Indicates that the information is a warning. Warnings
advise you of circumstances that could result in injury or death to the
patient or operator. Read and understand the warnings before
operating the injection system.
CAUTION: Indicates that the information is a caution. Cautions advise
you of circumstances that could result in damage to the device. Read
and understand the cautions before operating the injection system.
Note:
Warnings
Indicates that the information that follows is additional
important information or a tip that will help you recover
from an error or point you to related information within the
manual.
Patient injury may result from a system malfunction. If a system
malfunction occurs, immediately remove unit power (by pulling the battery
from the Scan Room Unit), and disconnect the unit from the patient. If a fault
message is displayed that cannot be corrected, and/or the system is not
operating correctly, do not use the injection system. Call MEDRAD for
assistance.
Patient injury could result from leaks or ruptures during an injection. To
prevent leaks or ruptures in the event of a blockage, use only catheters and
connectors with pressure ratings compatible with this system.
Explosion hazard. The MEDRAD Spectris Solaris EP MR Injection System is
not suitable for use in the presence of a flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
Fire hazard. To avoid an electrical fire, assure the correct type of fuse is used
for replacement. The fuse must be replaced with Type F, 250 V, 2.5 A fuse by
qualified personnel only.
Electrical shock hazard. Hazardous voltages exist within system
components. Do not remove or open any enclosure.
Electrical shock hazard. Avoid fluid entry into system components. Do not
immerse any components in water or cleaning solutions. Use a damp cloth
when cleaning on or around the battery and the Integrated Continuous Battery
Charger system power supply.
Electrical shock hazard. Serious injury or death may result from exposure to
hazardous voltages existing within the system. Disconnect the Battery
Charging System from line power and remove the battery from the Scan
Room Unit before cleaning.
1-6
1 - Introduction
Electrical shock hazard. Equipment must only be connected to a supply
mains with protective earth.
Ventilation hazard. To avoid a build up of hydrogen gas from the battery,
assure the room is well ventilated while battery is charging.
Improper disposal of the battery pack may result in explosion, leakage,
or personal injury. Do not open, or dispose of in a fire! Follow all local
regulations concerning the disposal of spent lead-acid based batteries, or
contact MEDRAD for assistance.
System electronic assemblies contain potentially hazardous materials.
Dispose of system components or accessories properly. Follow local
regulations for proper disposal or contact MEDRAD Service for assistance.
Unsafe operation may result from using improper accessories. Use only
accessories and options provided by MEDRAD designed for this system.
Chemical burn hazard. Always carry the battery pack firmly by the battery
pack hand grips. Damage to the housing may result in a chemical burn
hazard. Do not use if the housing is severely cracked or damaged.
Voltage hazard from worn cabling or unit disassembly. To avoid exposure
to potentially hazardous voltages, do not disassemble the injection system in
any way. Worn cabling also creates voltage hazards. If any worn or damaged
cables are detected, do not use the injection system. Contact MEDRAD for
service or replacement.
The MEDRAD Spectris Solaris EP MR Injection System is a dual syringe
system. Always ensure that the proper syringes are loaded with contrast
media and flush solution prior to the injection. Failure to properly load and
install the syringes may require the procedure to be repeated. Syringe A is
designated for contrast agent use only. Syringe B is designated for flush
solutions only.
Injury or equipment damage may result from use of tools containing
ferrous materials. Use only non-magnetic tools to install any scanner/
magnet room components.
Patient injury and/or catheter damage may result from using connector
tubing (LPCT) that is too short. Operator must consider tubing length and
stretch limitations when moving the injector or the patient.
Serious injury or death may result from syringe failure. Do not retract
pistons with connector tubing installed. Retracting the pistons with the
connector tubing installed on syringes will create a vacuum in the syringe due
to the check valve in the connector tubing. This vacuum may accelerate the
plunger rapidly toward the tip of the syringe when it is removed from the
injector causing the syringe to break.
1-7
MEDRAD Spectris Solaris EP MR Injection System
Cautions
Condensation may cause electrical damage to the injection system. Do
not use the system immediately after it has been brought indoors from
extreme outside temperatures. Allow the system to stabilize at room
temperature before use.
Injector may disarm or fail to operate upon exposure to high
electromagnetic fields that may be generated by radio transmitters or cellular
phones, or upon exposure to high levels of electrostatic discharge.
This injector system is in compliance to IEC-60601-1-2 / Second and
Third Edition Standards. Special precautions regarding ElectroMagnetic
Compatibility (EMC), are required for installation and use of this injector
system. Detailed EMC information can be found in the MEDRAD Injector
Service Manual - Addendum, (label number: 202559).
Damage can occur as a result of incorrect voltage. Before plugging in the
system, check the following:
•
Verify that the voltage and frequency marked on the serial tag on
the back of the unit matches the voltage and frequency of the
electrical outlet.
•
Verify that the Control Room Unit and the Battery Charger power
supply have the appropriate power cord plugs for the power
outlet.
Additional warnings, cautions, and notes are located throughout this manual,
where applicable.
1-8
2 - System Basics
2 - System Basics
About the Injection
System
The MEDRAD Spectris Solaris EP MR Injection System is a programmable,
dual syringe system, designed to accurately administer controlled doses of
intra-venous MR contrast agents and common flushing solutions to patients
undergoing a contrast enhanced MR scan.
The system consists of two basic components that communicate by a direct
connection of fiber optic lines.
•
The Control Room Unit houses the Touch Screen and electronic
components used to program the injection system.
•
The Scan Room Unit, positioned near the magnet bore, contains
the Injector Head, system battery pack, and the mechanical
assemblies required for fluid delivery.
A battery charger is also supplied with the system, used to charge the Scan
Room Unit battery pack. For convenience, the charger can be used in the
control room, but should never be installed or operated in the scan room.
Note:
Follow all institutional, local, or national safety regulations
related to routing cabling on the floor.
2-9
MEDRAD Spectris Solaris EP MR Injection System
Pressure Safety
Limit
The MEDRAD Spectris Solaris EP MR Injection System is designed to allow
varied flow rates for contrast injections. By automatically reducing the flow
rate, the system can limit the pressure produced during an injection to prevent
damage or failure of any connecting devices or tubing. This feature is called
Pressure Safety Limit.
Inability to maintain the desired flow rate while remaining below the Pressure
Safety Limit can be caused by various conditions including contrast viscosity,
catheter sizing, connector tube sizing, and stopcock restrictions. If the system
is unable, for a period of three seconds, to maintain a flow rate of at least 10%
of the programmed rate, the system will disarm due to a stall condition.
If unable to automatically achieve the required level of flow rate reduction,
thus reaching the Pressure Safety Limit, the system will terminate the
injection and move to a disarm state.
Response to
Occlusions
When injecting into an occlusion, a stall condition (flow rate less than 10% of
programmed rate) will result. A stall condition lasting more than 3 seconds (3
minutes for programmed rates less than 0.1 ml/sec) will result in the injection
being automatically terminated.
If an occlusion occurs during KVO (Keep Vein Open) the system will detect
the condition after 4 or less KVO boluses fail to be delivered. This will
correspond to from 1 minute with a KVO interval of 15 seconds configured, to
5 minutes with a KVO interval of 75 seconds. Refer to the Setup screen to
determine the current KVO setting.
If a stall occurs due to an occlusion, and the blockage is subsequently
removed, less than 10 ml will be delivered as the pressure in the
administration set dissipates.
Volume and Rate
Protection
The following means are provided to protect against over and under volume
or rate conditions:
•
Warnings displayed on the Safety screen and during the arming
sequence remind the operator to check the programmed injection
parameters prior to the system being armed.
•
An onscreen indication of insufficient volume is provided whenever the
total volume programmed to be delivered is greater than the amount of
fluid in the syringe.
•
Injection monitoring is performed to detect over rate or over volume
conditions due to system faults. If either of these conditions is detected,
the injection will be stopped before an additional 10 ml of fluid above
programmed volume is delivered.
2 - 10
2 - System Basics
Control Room Unit
1
3
2
1. Handswitch
2. System Power Switch
3. Touch Screen
At rear of Touch Screen Assembly - Display Contrast Controls
2 - 11
MEDRAD Spectris Solaris EP MR Injection System
Scan Room Unit
5
2
1
3
4
1. Injector Head
2. Handswitch
3. Lower Console
4. System Battery Pack
5. Middle Pivot Clamp
Not shown - Contrast Holder (optional)
2 - 12
2 - System Basics
Injector Head
1. Manual piston movement knobs
2. Armed indicator lights
3. ENABLE button - Used to activate the forward/reverse controls - the
appropriate direction must be selected within 5 seconds.
4. Syringe A forward/reverse controls
5. AIR EXPELLED button/indicator
6. Syringe B forward/reverse controls
7. Syringe A: Contrast agent
8. Syringe B: Flush solution
2 - 13
MEDRAD Spectris Solaris EP MR Injection System
Battery Charger
1
2
3
4
5
1. Battery Pack
2. Battery Charging Unit
3. Charging Indicator - Amber
4. Power Indicator - Green
5. Battery Charger Head
2 - 14
2 - System Basics
Optional Control
Room Unit
Accessories
Adjustable Height Pedestal
WARNING: Injury or equipment damage may result if the
adjustable height pedestal is taken into the scanner room. Do not
take the adjustable height pedestal in the scanner room. It
contains ferrous material that could be attracted toward the magnet.
Wall Mounting Bracket
Note:
These accessories contain ferrous material and are
designed to be used in the Control Room only. Do not
install or operate in the Scan Room.
2 - 15
MEDRAD Spectris Solaris EP MR Injection System
Touch Screen
Calibration
To enter Touch Screen Calibration mode, simultaneously press both the
Contrast UP and DOWN keys on the rear of the touch screen housing. A
series of screens with instructions to press the appropriate calibration circles
will appear.
CAUTION: Do not touch the screen with a sharp object in order to
perform the calibration.
Help Mode
The Help screen can be accessed by pressing the HELP button on the lower
right corner of the Main screen. Besides safety information, the Help screen
displays a variety of topics as displayed below.
2 - 16
2 - System Basics
Setup Mode
The Setup screen can be accessed by pressing the SETUP button at the
lower right corner of the Main screen. The Setup screen allows user
configurable options and preferences to be selected, along with setting of
date and time parameters.
Select the appropriate option, then choose from the available selections in the
display window. Select the DEFAULT key to return all options to original
factory settings.
The system provides a calibration and maintenance reminder. This reminder
will be displayed on the System Logo screen at each startup, beginning 30
days before the system is due to be recalibrated. The duration of time from
one calibration to the next is programmed during system installation or by
selecting the Calibration Reminder key and entering the correct due date.
2 - 17
MEDRAD Spectris Solaris EP MR Injection System
2 - 18
3 - Preparing to Inject
3 - Preparing to Inject
Applying Power
Place the power switch located on the right side of the Control Room Unit in
the ON position. The System Logo screen will appear while the system
performs a series of self diagnostic tests.
Note:
Do not touch the screen or activate any controls while
self diagnostics are in progress. If this occurs, diagnostic
tests will interpret this activity as a hardware failure and
halt the system. The system must then be powered
down/up to reset the error.
After diagnostics have successfully completed, the System Logo screen will
be replaced by the Safety screen
After reading the Safety screen, press CONTINUE to view the Main screen.
Apply power to the Scan Room Unit by inserting the battery into the
receptacle on the bottom of the Scan Room Unit. Upon power up of the CRU
and SRU, verify that the indicators, lamps, and speaker are operational.
Note:
The Control Room Unit can be programmed for an
injection without power applied at the Scan Room Unit.
3 - 19
MEDRAD Spectris Solaris EP MR Injection System
Main Screen
The Main screen is entered from the Safety screen after power-up. The Main
screen is used during programming, arming, and injecting, with applicable
screen controls made visible based on the task currently being performed.
3 - 20
3 - Preparing to Inject
Battery Maintenance
WARNING: Explosion Hazard. Serious injury or death may result
from improper use of the battery charger. The battery charger,
MEDRAD Catalog Number 3012424, is intended for use in a well
ventilated area, with the injection system battery, MEDRAD Catalog
Number 3012070, only. Do not use the charger with non-rechargeable
batteries.
When the Main screen appears, check the status of the system battery in the
lower left corner of the screen. The battery icon will contain three horizontal
bars when indicating full charge, two when indicating medium charge, one for
low charge, and no indicators in the icon when the battery is fully depleted,
not connected, or communication is not detected between the Control Room
and Scan Room Units.
If the depleted battery is not replaced when only one bar is displayed, the
system will complete any injection that is in progress. However, the system
may not initiate a single or multi arm injection, the forward and reverse
controls on the injector head may not function and system communications
may be lost.
Each battery pack should last for 4 to 6 typical injections using a 20 minute
KVO, or approximately 5 hours in an idle state before requiring a recharge.
Monitor battery status per injection and on a daily basis. Each battery is
capable of being recharged approximately 300 times. When the life of a
battery pack becomes shortened, noticeably sustaining fewer injections per
charge, this signals that battery life is expiring and the battery pack should be
replaced. Call MEDRAD Service for battery pack replacement.
To charge the battery, place the 3-pronged, charging head into the battery,
then connect the charger to AC power. A green LED on the charger indicates
that AC power is applied. An amber LED indicates that the battery is charging.
The amber LED will turn off when full charge is reached. Battery charge time
is approximately 5 hours.
3 - 21
MEDRAD Spectris Solaris EP MR Injection System
Syringe and Disposable Accessory Installation
Retracting the
Pistons
Fully retract each piston by using the reverse switches on the injector head.
Note:
When using the reverse switches, first press the Enable
switch; then within 5 seconds, press the reverse
switch(es). Both pistons may be reversed simultaneously.
The forward and reverse switches have dual speed capabilities:
When the switches are partially depressed, the pistons will move slowly.
When fully depressed, the pistons will move quickly. Forward and reverse
piston speeds are fully configurable (1 to 10 ml/sec) in the Setup mode.
The manual knobs may also be used to move the pistons in the forward or
reverse direction. Turn the knobs clockwise to advance the piston,
counterclockwise to retract.
3 - 22
3 - Preparing to Inject
Installing a Syringe
WARNINGS:
Patient or operator injury may result if damaged components are
used. Do not use damaged components. Visually inspect all
components before use.
Patient infection may result from the use of non-sterile
components. Maintain sterility of all disposable components. Do not
store pre-loaded syringes.
The use of single-use disposable devices on more than one
patient is a biological hazard. Do not reuse single-use disposable
components.
Patient injury could result if the syringe is not properly engaged.
Ensure the alignment marks on the syringe and injector head are
properly aligned, and the piston and plunger are interlocked. Improper
engagement may cause syringe damage or under-volume delivery.
CAUTION: Improperly engaged syringes may leak or be
damaged. Ensure proper engagement of syringe and injector. Syringe
and injector engagement points must align.
Note:
Syringe A is intended for contrast media, and Syringe B
is intended for flushing solution only.
1. Align the flanges on the syringe with the notches in the injector head.
(The syringe is keyed to properly fit in only one way.)
2. Insert the syringe.
3. Twist 1/4 turn clockwise until the syringe snaps into place. (Graduations
will be facing the front of the injector head.)
3 - 23
MEDRAD Spectris Solaris EP MR Injection System
WARNING: Air embolism can cause patient injury or death. Expel
all trapped air from the syringe(s), connectors, tubing, and catheterover-needle before injecting.
Note:
Do not hit or tap the syringe to remove air bubbles.
To reduce the volume and size of air bubbles drawn into the syringe during
loading, a MEDRAD Fluid Dispensing Device (FDD or “spike”) is
recommended. Air removal from the syringe(s) will be much more difficult if a
small diameter tube, such as a catheter-over-needle, needle, or a tube longer
than 10 inches (25 cm), is used for loading.
Operator vigilance and care, coupled with a set procedure, is essential to
minimizing the possibility of an air embolism. The injector head should be
pointed upward during loading, enabling any air to accumulate at the syringe
tip and to be expelled. The injector head should be pointed downward during
an injection, enabling any small air bubbles which could still be in the fluid to
float to the rear of the syringe(s).
To help avoid air injection, MEDRAD syringes are equipped with FluiDot indicators. These FluiDot indicators should be observed as part of an arming procedure. When the FluiDot is viewed through an empty syringe, the dots
appear as small narrow ellipses as illustrated below in figure 1. However,
when viewed through a full syringe, the dots become larger, almost round (or
wider than round) as illustrated below in figure 2.
Fig. 2 Filled
Fig. 1 Empty
FluiDot indicators must be viewed in a properly illuminated environment, with
a light source behind the operator providing enough light to permit easy
viewing.
To minimize air embolization risks, ensure that one operator is designated the
responsibility of filling the syringe(s). Do not change operators during the
procedure. If an operator change must occur, ensure that the new operator
verifies that the fluid path is purged of air.
WARNING: If a blockage occurs, disposable components with a
lower pressure rating may leak or rupture. Use only catheter and
connectors with ratings that are compatible with the MEDRAD
Spectris Solaris EP MR Injection System.
During installation of the low pressure connector tubing (LPCT) with Tconnector to the syringe(s), and before arming, manually advance the syringe
plunger(s) to provide a very slow flow of fluid at the connection. An absence of
flow is an obvious indication of air or a blockage in the fluid path.
3 - 24
3 - Preparing to Inject
Loading a Syringe
WARNINGS:
Remove all trapped air from the syringe, connector tubing, and
catheter-over-needle before connecting the patient to the injector.
Syringe sterility will be compromised, and patient infection may
result, if the plunger is removed from the syringe. Do not remove the
plunger to fill the syringe.
Bacterial contamination can occur if syringes are used to store
contrast media. Use loaded syringes immediately. Do not store loaded
syringes for later use. Discard any unused syringes.
Note:
The presence of rounded FluiDot indicators does not
indicate the total absence of air bubbles in the syringe tip.
Note:
FluiDot indicators must be viewed in a properly
illuminated environment, with a light source behind the
operator providing enough light to permit easy viewing.
1. Position the injector head so that the syringes are
pointing upward.
2. Fully advance each piston plunger. The plungers
may be advanced simultaneously after pressing
ENABLE.
3. Attach a sterile filling device (spike or Female-toFemale Adaptor - MEDRAD Catalog Number FFA
50) onto the tip of the syringe. If loading contrast
media and/or saline from a bag or bottle, use a
spike. If loading contrast media from a pre-filled
syringe, use an FFA 50.
A. If using a spike, open the bottle(s) of contrast
and/or bag(s) of flushing solution, then draw the
contents (contrast for syringe A; flush for syringe B)
into the syringe(s) by depressing ENABLE, and
then the reverse load button for each syringe.
B. If using the FFA 50, attach it to the tip of syringe
A, then attach the pre-filled syringe to the FFA
50. Draw the contents from the prefilled syringe
into syringe A by pressing ENABLE, then the
reverse load button.
4. With the filling device still attached, advance the
plunger to expel any air that may remain in the top
of the syringe; then, if necessary, draw more fluid
into the syringe to replace fluid loss.
5. Remove the filling device and expel any air bubbles
from the syringes.
3 - 25
MEDRAD Spectris Solaris EP MR Injection System
6. Attach the long end of the T-connector to
syringe B.
7. While the injector head is still in a vertical
position, attach the short section of the Tconnector to syringe A.
8. Starting with syringe A, then syringe B,
prime the T-connector, then fill the
connector tubing with the appropriate
fluid. Ensure that all air is expelled from
the entire length of the tubing.
Saline Flush
Contrast
Note:
A MEDRAD SSIT 96VLD low pressure connector tube
(LPCT) holds approximately 7 ml of fluid. If syringe B is
used to flush, use at least 8 ml of flush to deliver this
volume to the patient.
Note:
If the connector tube is filled with saline, contrast will be
delivered to the patient with a delay dependent on the
flow rate selected for syringe A.
Note:
When the connector tube is filled with contrast the
volume remaining displayed on the protocol screen is
approximately 7 ml less than what was loaded into the
syringe.
9. Tilt the injector head downward before attaching to the vascular entry
device in the patient. After attaching the connector tube to the vascular
entry device verify that the connector luer fittings are secured. The
injector head must be maintained in this position during the injection.
WARNINGS:
Patient injury could result from movement of the Scan Room Unit
after the patient is connected to the fluid path. Lock the casters at
the base of the unit and the Middle Pivot Clamp to prevent unintended
movement.
Patient injury and/or catheter damage may result from using
connector tubing (LPCT) that is too short. Operator must consider
tubing length and stretch limitations when moving the injector or the
patient.
10. Secure the Scan Room Unit by locking the casters and the Middle Pivot
Clamp, then verify that all air has been expelled from the fluid path by
carefully inspecting all tubing and syringe(s). Acknowledge that the
inspection has occurred by pressing the AIR REMOVED confirmation
button/indicator on the injector head. The Air Removed Indicator will then
illuminate yellow on the touch screen.
Note:
Reverse movement of the pistons after the AIR
EXPELLED button has been pressed will cancel the Air
Expelled status. Re-check the fluid path for air, then
press the AIR EXPELLED button again to continue.
3 - 26
3 - Preparing to Inject
Reinstalling a
Syringe
WARNING: Patient injury could result if the syringe is not
properly engaged. Ensure the alignment marks on the syringe and
injector head are properly aligned, and the piston and plunger are
interlocked. Improper engagement may cause syringe damage or
under-volume delivery.
If you remove a syringe from the injector, and then wish to reinstall it, perform
the following steps:
1. Insert the end of the syringe in
the horizontal cutouts in the
injector head.
2. Advance the piston until it is
past the plunger feet and the
piston/plunger interlock.
3. Rotate the syringe 1/4 turn
clockwise until the syringe
locks and alignment marks are
positioned.
4. Proceed as normal by
aspirating and dislodging any
air bubbles.
Note:
If bubbles appear in the syringe DO NOT hit the syringe
to remove them. Reverse the plunger 3 to 5 ml, then rock
the head on the pivot to gather and accumulate the small
bubbles. Expel the remaining air.
3 - 27
MEDRAD Spectris Solaris EP MR Injection System
Programming
If a program has not been previously entered or stored on the Main screen
when the unit is powered up, the Main screen will display default settings; 1.0
ml/s flow rate and 1.0 ml volume, KVO off and No Delay.
Flow Rate and
Volume
Begin programming by selecting any programmable block, such as FLOW
RATE or VOLUME. When a programmable block on the screen is touched, a
keypad will be displayed to permit the selection of numeric values. The
numeric keypad is displayed when a Flow Rate, Volume or Delay value is
selected. The keypad window will also display the appropriate programmable
range for the parameter selected. To lock in values, press ENTER. Press <<
to edit a selection, or CANCEL to eliminate a selection if an error is made.
Pressure Limit
The pressure limit can be programmed by choosing a value between 100-325
PSI up to the greater of the syringes’ maximum pressure.
3 - 28
3 - Preparing to Inject
Multiple Phases
If appropriate, select a second phase for the injection protocol by pressing the
triangle block below the first phase of the injection. The Phase Type selector
will appear in order to select the function of the new phase.
Hold and Pause
Phases
A Hold or Pause phase can be programmed into a multi-phase injection. A
Pause phase will stop the total injection process for a preprogrammed length
of time, while the Hold phase will stop the injection until input from the
operator resumes the injection. A Hold phase can be maintained for up to 20
minutes, at which time the system will disarm.
After selecting the phase type, continue programming by entering Flow Rate
and Volume values for the new phase.
3 - 29
MEDRAD Spectris Solaris EP MR Injection System
Programmed Delay
After entering Flow Rate and Volume parameters, press SET in the Delay
Timer field to select the delay type (Scan Delay, Inject Delay, Stopwatch, or
No Delay.)
Note:
There is no direct interface between the scanner and the
injector. The scanner cannot trigger the injector, nor can
the injector trigger the scanner.
Scan Delay
Scan Delay time will elapse in the timer block on the screen. The time
remaining before the scan should be activated will decrement in one second
intervals. (The scan delay countdown will continue through multiple arm
injections.) When countdown is complete, the system will emit 5 beeps.
Inject Delay
Inject Delay will also countdown in one second intervals, commencing when
the handswitch is pressed. The clock will display, in one second decrements,
the time remaining before the injection will begin. When inject delay is chosen
and the handswitch is pressed, the injection will automatically occur unless
the injector is disarmed. When countdown is complete, the system will emit 5
beeps and the injection will automatically begin.
If Hold is activated during an inject delay or a scan delay, the timer will stop
during the hold interval and resume when the handswitch is pressed.
For scan or inject delays that are longer than 3 minutes, the unit will beep 30
seconds before the delay is to terminate, then will beep every second from 5
seconds through 1 second before the delay is due to terminate.
3 - 30
3 - Preparing to Inject
Stopwatch
The Stopwatch function initiates an incremental count of elapsed time from
initial fluid injection.
After selecting the delay type, enter the delay duration on the numeric keypad.
To lock in values, press ENTER. Press CANCEL to eliminate a selection if an
error is made.
KVO (Keep Vein
Open)
The KVO function delivers small boluses of fluid from syringe B at
configurable intervals. KVO can run during:
•
Programming
•
Pre and post injection
•
Between multiple injections
•
During Pause and Hold
The delivery interval can be selected in the Setup mode, which is accessed
using the Setup button on the Main Screen.
After an initial KVO pulse of 2 ml, KVO delivery intervals include 0.25 ml
pulsed every:
15 seconds
20 seconds
30 seconds (default)
45 seconds
60 seconds
75 seconds
3 - 31
MEDRAD Spectris Solaris EP MR Injection System
The KVO field displays the time available to support
KVO based on the configured interval and the volume
remaining in syringe B less any volume programmed
from syringe B in the protocol.
Starting KVO:
On the Main Screen, press START in the KVO field to initiate KVO. When
KVO is running, “KVO” will appear, and the KVO Injecting arrows will flash in
the Syringe B touch screen indicator.
KVO will function during Pause, Hold and/or inject delay periods. KVO will
resume post-injection until no fluid remains in syringe B, or until STOP KVO is
pressed in the Injection Complete window.
Note:
Volume displayed in the Volume Delivered window can
be configured in the Setup mode to include the total KVO
volume delivered in addition to volume delivered by the
programmed injection.
KVO may be stopped at any time by pressing STOP in the KVO field, or by
pressing any injector head control button (this will also disarm the system and
terminate any injection in progress). Other actions that disarm the injector,
such as syringe removal, disarm button press and injection stall, will also stop
KVO.
KVO and Occlusions:
If an occlusion occurs during KVO the system will detect the condition after 4
or less KVO boluses fail to be delivered. This will correspond to from 1 minute
with a KVO interval of 15 seconds configured, to 5 minutes with a KVO interval of 75 seconds. Refer to the Setup screen to determine the current KVO
setting.
3 - 32
3 - Preparing to Inject
Storing a Protocol
To store a protocol for future use, press the STORE button on the upper right
corner of the Main screen.
An alpha-numeric keypad will appear with a flashing cursor in the title block.
Type in a title of up to 20 characters, including spaces. Use the arrow key to
backspace, individually erasing characters, and the CLEAR key to clear a
string of text. When title entry is completed, press ENTER.
To exit the Store screen without making changes, press the CANCEL button
in the upper right corner.
3 - 33
MEDRAD Spectris Solaris EP MR Injection System
Recalling a Stored
Protocol
To access program memory, press RECALL on the Main screen.
Select a previously stored injection protocol by pressing one of the names on
either side of the screen. Key parameters of the selected injection will be
displayed in the center of the screen. Once selected, the protocol can be
deleted by selecting DELETE at the upper right corner of the screen, or
brought to the Main screen by selecting OK.
3 - 34
4 - Arming and Injecting
4 - Arming and Injecting
Before beginning the arming process, ensure that the casters on the Scan
Room Unit are locked, verify that all air has been expelled from the fluid path,
and that the programmed parameters are correct. Carefully inspect all tubing
and syringe(s), then acknowledge that the inspection has occurred by
pressing the AIR EXPELLED button/indicator on the injector head. A yellow
illuminated Air Expelled Indicator on the touch screen confirms that the button
has been pressed.
WARNINGS:
Air embolization can cause death or serious injury to the patient.
Do not connect a patient to the injector until all trapped air has been
cleared from the syringe and fluid path.
Patient injury could result from high flow rate venous injections.
Use extreme care when selecting flow rate and duration. Before
arming the injector, verify that high flow rate injection parameters have
not been unintentionally programmed.
Patient injury could result from inadvertent aspiration. To
minimize the possibility of inadvertent aspiriation and injection, ensure
the patient is disconnected from the injector when utilizing the forward/
reverse plunger control(s).
Extravasation can cause injury to the patient. Follow commonly
accepted good clinical procedures to minimize the possibility of
extravasation.
Arming
To begin the arming and injecting process, press ARM on the Main screen. If
necessary, changes can be made to programmed injection parameters after
the arming sequence is complete. Select the required parameter, then enter
the correct value with the on-screen keypad. Pressure safety limit programmed by the user is indicated to the user and cannot be changed when
the injector is armed.
Note:
Single and MultiArm
If the AIR EXPELLED button on the injector head has not
been pressed, the system will request user confirmation
that air has been expelled before proceeding.
Select either a Single or Multiple arming sequence by pressing either SINGLE
or MULTI. (The default is Single arm.)
A single arm injection will perform the protocol once, then disarm.
A multi-arm injection allows the protocol to be repeated, creating a series of
injections. After the protocol is completed, the system will automatically rearm in preparation for the protocol to be repeated. Each injection in the series
must be started with the handswitch.
4 - 35
MEDRAD Spectris Solaris EP MR Injection System
Insufficient Volume
If an insufficient volume condition occurs during a multi arm sequence, the
system will remain armed to permit the injection of the remaining volume.
However, the screen will update to display only the phases that are
achievable with the volume that remains. In a single arm sequence, the
screen will update when arming occurs to display only the phases that are
achievable.
While the system is armed, pressing DISARM or activating any injector head
controls will return the system to the idle state.
Injecting
After the system has been armed, press the handswitch to begin the injection.
Additional presses of the handswitch will alternately “hold” and resume the
injection. The maximum duration for Hold is 20 minutes. If the maximum hold
time is exceeded the injection will abort automatically.
If an inject delay has been programmed, pressing the handswitch will
activate the countdown timer. The programmed injection will automatically
begin when the timer counts down to zero. If the handswitch is pressed during
an inject delay, the countdown timer will stop counting until the switch is
pressed again, or the Hold time is exceeded.
If a scan delay is programmed, the scan delay countdown and the injection
will start simultaneously. During the injection, additional presses of the hand
switch will alternately “hold” and resume the injection and the scan delay
timer.
If KVO is running: KVO will function during Pause, Hold and/or inject delay
periods as long as sufficient fluid remains in syringe B to complete the
programmed injection. KVO will run post-injection until no fluid remains in
syringe B, or until STOP KVO is pressed in the Injection Complete window. To
stop KVO the operator can also press any injector head control buttons.
If a Hold phase is entered, parameters for the remaining portion of the
injection protocol can be altered.
4 - 36
4 - Arming and Injecting
On the Injecting Screen:
•
As each phase is activated, the phase parameters will be highlighted to
display injection progress.
•
The Duration window will also increment to display elapsed time.
•
The Delivered window will increment as the injection proceeds to display
volume delivery (including KVO volume, if selected in Setup mode).
•
The Volume Remaining display will decrement.
•
The Programmed pressure limit and the current pressure will be indicated
on the display. If a pressure limit condition occurs, it will indicate on the
display.
•
If KVO is selected, the time available for KVO will decrement in the KVO
Time Remaining window while KVO is active. (During an injection, KVO
will stop and the time display will not count-down.)
On the Injector Head:
Disarming
•
While injecting, indicator lamps on the back of the injector head will be
illuminated (white for syringe A, blue for syringe B). The appropriate
lamps will be lit solid while injecting, and flash while either Armed or on
Hold.
•
If multi-arm is selected, the indicator lamps will flash when the system
rearms.
•
During KVO the blue indicator lamp for syringe B is illuminated.
•
AIR EXPELLED Indicator is illuminated.
Pressing DISARM, activating any injector head controls, or touching any
portion of the touch screen while the system is injecting, will cause the system
to disarm.
The Hold mode can be entered at any time during an injection by pressing the
handswitch. The system will remain in this state until the handswitch is
pressed a second time, or the maximum hold time of 20 minutes is exceeded.
4 - 37
MEDRAD Spectris Solaris EP MR Injection System
Note:
A MEDRAD SSIT 96VLD low pressure connector tube
(LPCT) holds approximately 7 ml of fluid. If syringe B is
used to flush, use at least 8 ml of flush to deliver this
volume to the patient.
Note:
If the connector tube is filled with saline, contrast will be
delivered to the patient with a delay dependent on the
flow rate selected for syringe A.
Note:
When the connector tube is filled with contrast, the
volume remaining displayed on the screen is
approximately 7 ml less than what is present in the
system.
When an injection (single or all multi-arm sequences) is completed, the
following window, with a brief summary of the injection parameters, will be
displayed.
4 - 38
4 - Arming and Injecting
Injection History
To review injection parameters used in a procedure, along with actual
achieved values for the injection, press the HISTORY button on the Main
screen.
The Injection History screen displays an injection summary block containing
the following data:
•
Time and Date Started
•
Programmed Flow Rate
•
Programmed Volume
•
Programmed Protocol
•
Total Fluid (plus KVO)
•
Delay Type
•
Delay Duration
•
Pressure Limit Programmed
•
Peak Pressure
•
Pressure Limit Status (YES/NO)
•
Premature Termination Status (YES/NO)
The system maintains status information of the 20 most recent injections,
sorted by date and time.
To delete any injection protocol history from the system, press DELETE while
the protocol is selected. To scroll to the next page of protocols, press the
ARROW key. To exit the Injection History screen, press CANCEL.
4 - 39
MEDRAD Spectris Solaris EP MR Injection System
Clean Up
Note:
Do not resterilize or reuse any disposable items.
When cleaning the injector, remove and discard all used disposable items.
(Syringes should be removed without retracting the pistons.) It is not
necessary to remove the connector tubing when removing and discarding
syringes.
WARNING:
Serious injury may result from syringe failure. Do not retract
pistons with connector tubing installed.
•
Disconnect the Control Room Unit from line power and remove
the battery from the Scan Room Unit before cleaning.
•
Avoid fluid entry into system components. Do not immerse any
components in water or cleaning solution.
•
Do not remove any covers or disassemble the injector.
Periodically inspect for loose or frayed cables, loose covers,
cracks, dents, or loose hardware. Contact MEDRAD Service for
repairs.
•
Retracting the pistons with the connector tubing installed on
syringes will create a vacuum in the syringe due to the check
valve in the connector tubing. This vacuum may accelerate the
plunger rapidly toward the tip of the syringe when it is removed
from the injector causing the syringe to break.
CAUTIONS:
System malfunction may be caused by failure to perform regular
maintenance. Regular preventive maintenance is recommended to
ensure that the system stays calibrated and functions properly. Refer
to Appendix B of this manual or contact MEDRAD for additional
information.
Do not expose system components to excessive amounts of
water or cleaning solutions. Wipe components with a soft cloth or
paper towel dampened with cleaning solution.
Do not use strong cleaning agents and solvents. Warm water and
a mild disinfectant are all that is required. Do not use strong industrial
cleaning solvents such as acetone.
Note:
For all body fluid spills, follow institutional
decontamination procedures.
Note:
If contrast medium has leaked inside any component of
the system, the affected subassembly should be
disassembled and cleaned by MEDRAD Service
personnel or returned to MEDRAD Factory Service.
4 - 40
4 - Arming and Injecting
Scan Room Unit
Using a soft non-abrasive cloth, warm water, and a mild disinfectant, carefully
clean the assembly, paying particular attention to the following:
•
Injector Head
•
Syringe Piston Plunger
•
Syringe Interface
•
SRU Lower Console covers
To clean the injector head, piston, and syringe interface:
1. Fully advance the piston.
2. Remove the battery from the Scan Room Unit.
3. Place the injector head in a vertical position.
4. Clean the piston with a soft cloth or paper towel dampened with cleaning
solution.
5. Thoroughly dry the piston with a paper towel.
6. Re-install the system battery, then fully retract the piston.
7. Remove the battery from the Scan Room Unit again.
8. Clean the inner area of the syringe interface with a soft cloth or paper
towel dampened with cleaning solution.
9. Wipe the injector head case and control panel with a soft cloth or paper
towel dampened with cleaning solution.
10. Thoroughly dry the injector head case and control panel with a paper
towel.
Control Room Unit
CAUTION: Do not spray cleaning solutions directly onto the
touch screen. To prevent damage, wipe the touch screen with a soft
non-abrasive cloth or paper towel dampened with cleaning solution.
4 - 41
MEDRAD Spectris Solaris EP MR Injection System
4 - 42
Appendix A: System Messages
Appendix A: System Messages
The system will display messages on the screen as conditions or events
occur. There are three basic types of messages:
WARNING: Patient injury may result from a system malfunction.
If a system malfunction occurs, immediately remove Scan Room Unit
power (by pulling the battery from the head stand), and disconnect the
system from the patient. If a fault message is displayed that cannot be
corrected, and/or the system is not operating correctly, do not use the
injection system. Call MEDRAD for assistance.
Type 1 Messages
Type 1 messages are messages which provide information regarding the
current status of the system, and will clear automatically from the screen.
These messages are typically displayed in the lower right corner of the
screen.
A - 43
MEDRAD Spectris Solaris EP MR Injection System
Type 2 Messages
Type 2 messages are messages that convey information that must be
explicitly acknowledged before proceeding. The message is displayed within
a yellow dialog box - a button (or buttons) must be pressed to acknowledge
and remove the message from the screen.
Type 3 Messages
Type 3 messages are system malfunction messages which require power to
be removed from the system. Some Type 3 messages provide suggestions to
prevent the condition from recurring. If the condition cannot be corrected,
record the code and number from the lower left corner of the dialog box, then
call MEDRAD Service for assistance.
System Error Detected
Immediately disconnect the patient from the injector.
Record the code below and contact MEDRAD Service.
A
Symcode = HSW
Errnum = -2600
Refer to the operation manual or www.Medrad.com for contact information.
A - 44
Appendix B: Maintenance and Checkout
Appendix B: Maintenance and
Checkout
This section contains recommended procedures for maintenance, and an
operational checkout of the MEDRAD Spectris Solaris EP MR Injection
System. Routine maintenance and inspection will:
Recommended
Maintenance
Schedule
•
Ensure continued performance of the injection system
•
Reduce the possibility of equipment malfunction
Your MEDRAD Spectris Solaris EP MR Injection System must be properly
maintained to ensure that it is in peak operating condition. Your individual
maintenance system and schedule depends upon how your injection system
is used, the type of procedures performed, and frequency of use. The
following maintenance schedule is recommended for the system:
Daily:
The piston rod should be thoroughly cleaned after each use. Before use each
day, the system should be cleaned and inspected, using the procedures
outlined in this section. Ensure that all system safety and warning labels are in
place and are legible.
Monthly:
Once a month, the entire system should be thoroughly inspected and
cleaned, and an Operational Checkout should be performed.
Annually:
As part of an annual maintenance program performed by a qualified
MEDRAD Service Representative or authorized dealer, both Electrical
Leakage and Ground Continuity checks should be performed.
NOTE:
Local regulations or hospital protocol may require
electrical leakage checks at more frequent intervals. If
this applies, local regulations for leakage must be
followed.
MEDRAD also recommends that a complete system calibration and
performance checkout be performed annually. Contact MEDRAD Factory
Service, or your local MEDRAD office for complete details.
In the United States, Canada, and Europe, the MEDRAD Service Department
offers Preventive Maintenance Programs. These annual programs greatly
assist in maintaining accuracy and reliability, and can also extend the life of
the system. Contact MEDRAD for details. In Europe, contact your local
MEDRAD office or your local authorized dealer for further information. Refer
to the back of the title page of this manual for address, telephone and FAX
information.
B - 45
MEDRAD Spectris Solaris EP MR Injection System
NOTE:
MEDRAD Service
Failures which occur due to lack of proper maintenance
will not be covered under warranty.
MEDRAD Service will make available upon request:
•
Circuit diagrams, component parts lists, or other information that will
assist qualified technicians to repair components classified as repairable.
•
On-site consulting or consulting references upon request.
Inspection Procedures
The following procedures are recommended for daily inspection of all
components in the MEDRAD Spectris Solaris EP MR Injection System. If any
defects are detected, either repair the system, or call MEDRAD for service.
Do not use the system until the problem is corrected.
Scan Room Unit
1. Inspect the housing for any damage or cracks that could allow fluid to leak
inside, or weaken the structural integrity of the unit.
2. Inspect all cables connected to the unit: Look for cuts, cracks, worn spots
or other obvious damage to the cables. Ensure that all connectors are
properly seated.
3. Inspect for contrast media build-up in the syringe interface area. Follow
the cleaning guidelines outlined in this section.
4. Inspect the stand, base, and support arm for cracks and other defects that
could weaken the structure.
5. Ensure that all mounting bolts and screws are secure.
6. Ensure that all locking mechanisms on the casters are functional.
7. Inspect the pivot points. The head and support arm must pivot freely. The
injector head should rotate on the support arm no more than 330o. The
support arm should not rotate on the center post more than 350o.
NOTE:
Control Room Unit
All relevant guidelines for institutional, local, or national
safety recommendations related to cable routing and
installation should be followed.
1. Inspect all cables connected to the unit: Look for cuts, cracks, or worn
spots, or other obvious damage. Ensure that all connectors are properly
seated.
2. Inspect the housing for any damage or cracks that could allow fluid to leak
inside, or weaken the structural integrity of the unit.
B - 46
Appendix B: Maintenance and Checkout
Wall Mount Bracket
1. Inspect all parts of the bracket for cracks and other defects that would
weaken the assembly.
2. Ensure that the bracket is securely attached to the wall.
3. Ensure that all cables are secured to the display control unit and do not
interfere with the movement of the mounting bracket.
Height Adjustable
Pedestal
1. Inspect the stand, base and support arm for cracks and other defects that
could weaken the structure.
2. Ensure all mounting bolts and screws are secure.
3. Ensure that the casters roll smoothly with no binding or scraping.
4. Ensure all locking mechanisms on the casters are functional.
5. Verify that the vertical height adjustment of the column shaft moves freely
without binding or scraping.
Battery Charger
1. Inspect all cables connected to the unit: Look for cuts, cracks, or worn
spots, or other obvious damage. Ensure that all connectors are properly
seated.
2. Inspect the housing for any damage or cracks that could allow fluid to leak
inside, or weaken the structural integrity of the unit.
3. Inspect all parts of the wall mounting bracket for cracks or other defects
that would weaken the assembly. If applicable, ensure that the bracket
remains firmly attached to the wall.
Communication Link
1. Inspect the cables for cuts, cracks or worn spots. Ensure that the
connectors are properly seated.
B - 47
MEDRAD Spectris Solaris EP MR Injection System
Cleaning Guidelines
Deposits of contrast media can interfere with proper operation of the
MEDRAD Spectris Solaris EP MR Injection System. The following guidelines
should be followed when removing deposits, or cleaning any portion of the
system.
WARNING: Serious injury or death may result from exposure to
hazardous voltages existing within the system. Disconnect the
system from line power before cleaning or attempting to perform any
maintenance. Ensure that the system is completely dry before
connecting to the power source and applying power.
CAUTION: Improper or careless cleaning methods may result in
equipment damage. Do not soak or immerse any part of the injection
system in water. While cleaning any outside portion of the system,
avoid allowing any water to leak inside system components.
•
If contrast medium has leaked inside any component of the system, the
affected subassembly should be disassembled and cleaned. This
cleaning procedure can be done in the field by trained MEDRAD Service
personnel, or returned to MEDRAD Service. If the cleaning will be
performed in the field, do not disturb any internal wiring or components.
•
Care must be taken not to get water or cleaning solutions inside any
system components. Do not use strong industrial cleaning agents or
solvents such as acetone. Warm water and a mild disinfectant such as
antibacterial hand soap are all that is required.
•
To clean the syringe interface area of the injector head, fully retract the
piston. Using a paper towel moistened with warm water or a mild
disinfectant, gently wipe the inner syringe installation area. Do not insert
any sharp instruments into this area during the cleaning process.
•
Check all System Safety and Warning Labels for legibility. Ensure that the
labels are not damaged or missing.
Operational Checkout
A basic functional checkout of the MEDRAD Spectris Solaris EP MR Injection
System should be included as part of regular maintenance. Verifying proper
operation of the injection system will help in detection of any problems that
may not be noticed in day to day operation. The following procedure
represents a suggested series of activities which encompass typical operation
of the system. Read the following procedure carefully before beginning the
checkout. If problems are detected, contact MEDRAD Service.
NOTE:
Any problems detected during this or any other
procedure should be corrected before using the injection
system in patient procedures.
B - 48
Appendix B: Maintenance and Checkout
System Labels
Ensure that all system safety and warning labels are in place and legible.
Power Up
Apply power to the system. Verify that the Safety screen is displayed after
system diagnostics occur. Press OK to acknowledge the messages on the
Safety screen. Upon power up of the CRU and SRU, verify that the indicators,
lamps, and speaker are operational.
Programming
After the Main screen is displayed, verify that the following controls are
functioning properly.
At the rear of the Control Room Unit, Press the Lighten Display Contrast
key until the screen is lightened to its fullest extent. Press the Darken Display
Contrast Key until the screen is darkened to its fullest extent. Adjust the
screen appearance to return to a desirable contrast level.
Fully advance and reverse the pistons by using the ENABLE key and the
forward/reverse controls. Verify that both pistons respond to the forward and
reverse controls.
Enter the following protocol:
Phase 1:
Phase 2:
Phase 3:
Phase 4:
Phase 5:
Flow Rate
10 ml/s
2.5
PAUSE
5.0
0.1
No Delay
KVO Off
Syringe A:
Syringe B:
Syringe A:
Syringe B:
Volume
20 ml
10
5 seconds
10
1
Arm in single injection mode and inject. In one of the phases, activate the
HOLD feature for at least 10 seconds.
Verify that the injection completes normally.
When the injection is complete, access the Injection History screen and verify
volume accuracy; actual volume and programmed volume should be the
same (41 ml).
Add a 15 second inject delay to the program and activate KVO.
Install syringes and fill them with water.
Arm in single injection mode and inject.
Verify that:
A. When the handswitch is pressed, the inject delay begins counting
down.
B. The inject delay beeps 5 times when the delay timer elapses and that
the injection begins automatically at that time.
B - 49
MEDRAD Spectris Solaris EP MR Injection System
Verify that when the injection completes, KVO resumes.
Remove and discard the syringe.
Remove power from the system.
B - 50
Appendix C: Specifications
Appendix C: Specifications
Scan Room Unit
Weight: 60 lbs. (27.3 kg.)
18.5"
46.99 cm
12.75" / 32.28 cm
51.75"
131.45 cm
21.75"
55.25 cm
19.0"
48.26 cm
C - 51
MEDRAD Spectris Solaris EP MR Injection System
Control Room Unit
Weight: 15 lbs. (6.8 kg.)
11.995"
30.46 cm
11.80"
29.97 cm
2.375"
6.03 cm
10.922"
27.91 cm
9.25"
23.5 cm
2.50"
6.35 cm
Battery Dimensions
Weight: 7.7 lbs. (3.5 kg.)
2.1"
5.3 cm
4.0"
10.2 cm
5.0"
12.7 cm
C - 52
8.2"
20.8 cm
Appendix C: Specifications
Battery Charger
Weight: 2 lbs. (0.9 kg.)
2.765"
7.02 cm
3.05"
7.74 cm
4.664"
11.84 cm
5.5 ft
1.68 m
1.114"
2.90 cm
Power Cords
5.08"
12.90 cm
1.59"
4.03 cm
American 12 ft. (3.6 m), Continental 9.8 ft. (3 m)
C - 53
MEDRAD Spectris Solaris EP MR Injection System
System Capabilities
SYRINGE A:
Disposable 63 ml
SYRINGE B:
Disposable 115 ml
VOLUME:
Syringe A:
0.5 ml to max. syringe volume in:
0.1 ml increments between 0.5 and 31 ml
1 ml increments for 31 ml and above
Syringe B:
FLOW RATE
(Programmable):
1 ml to max. syringe volume in 1 ml increments
0.01 to 10 ml/s in:
0.01 ml/s increments between .01 and 3.1 ml/s
0.1 ml/s increments between 3.1 and 10 ml/s
KVO (Programmable):
15 seconds
20 seconds
0.25 ml pulsed every: 30 seconds (default)
45 seconds
60 seconds
75 seconds
PRESSURE SAFETY LIMIT:
Factory set below 325 psi (2240 kPa)
PROGRAMMABLE
PRESSURE LIMIT (PSI/kPa):
100/690
150/1035
200/1380
250/1725
300/2070
325/2240 (default)
DELAY:
1 to 300 seconds in 1 second increments
PAUSE PHASE:
1 to 900 seconds in 1 second increments
INJECTION CAPABILITIES
6 phases per protocol
STORAGE CAPACITY
32 Protocols of up to 6 phases each
Protocol and User Configuration memory is maintained when system power is
off.
Executable Flow
Rates
These Flow Rates, with pressure safety limit set to 325psi/2240kPa, are
achievable with the MEDRAD Spectris Solaris EP MR Injection System, using
the Becton Dickinson* catheters listed below and the SSQK 65/115VS
MEDRAD Syringe/Disposables Kit.
18 g IV Catheter 20 g IV Catheter 22 g IV Catheter 24 g IV Catheter
BD pn 381144
BD pn 381134
BD pn 381123
BD pn 381112
Multihance*
7.0
5.8
4.0
2.6
Gadovist*
7.3
6.0
4.2
2.7
Magnevist*
9.0
7.2
5.2
3.2
Optimark*
9.7
7.8
5.5
3.7
Prohance*,
Omniscan*
10.0
8.3
5.8
4.0
Saline
10.0
9.1
6.1
4.3
* Becton Dickinson (BD) is a trademark of Becton Dickinson, Inc.
C - 54
Appendix C: Specifications
Multihance, Gadovist, Magnevist, Optimark, Prohance, and Omniscan are
trademarks of their respective companies.
System
Performance
Volume Accuracy:
Syringe A:
+/- (1% + 0.1 ml)
Syringe B:
+/- (5% + 0.1 ml)
Flow Rate Accuracy
+/- (10% + 0.005 ml/s) when rate is 0.01 to 0.99 ml/s
+/- (10% + 0.02 ml/s) when rate is 1 to 10 ml/s
Programmed Delay/
Pause Accuracy
+/- (5% + 0.2 second)
Displayed Pressure
Accuracy
+/- 10 psi
KVO Volume Accuracy
+/- 0.05 ml, averaged over 10 consecutive boluses
KVO Flow Rate Accuracy
1 ml/s +/- 0.2 ml/s
Forward and
Reverse Controls
Low Speed:
2.5 ml/s (default)
High Speed:
10 ml/s (default)
Low speed is selectable from 1.0 to 10.0 ml/s in 0.5 ml/s increments
High speed is selectable from 1.0 to 10.0 ml/s in 0.5 ml/s increments
C - 55
MEDRAD Spectris Solaris EP MR Injection System
EMI/RFI
The MEDRAD Spectris Solaris EP MR Injection System is designed to be in
compliance with IEC/EN 60601-1 Second/Third Edition, EN 60601-1-2
Second Edition and IEC 60601-1-2 Second/Third Edition.
Electrical
Requirements
100-240 VAC
50/60 Hz
Scanner Room Unit (Powered by Integrated Continuous Battery Charger): 100VA
Control Room Unit: 50VA
Power Supply DC
Output Voltage
Nominal 15.5 VDC
Electrical Leakage
Unit < 100 microamperes
Patient < 10 microamperes
Ground Continuity
The resistance from the earth ground connector at the plug-end of the AC
mains power cord to any exposed metal on the Control Room Unit shall be
less than 0.2 ohms.
Environmental
Specifications
Non-Operating: (Transportation and Storage)
Temperature:
-25° C to 70° C (-13° F to +158° F)
Humidity:
5% to 100% R.H.
Air Pressure:
48kPa to 110 kPa (6.96 psi-16 psi)
Operating: (The system may not meet all performance specifications if
operated outside of the following conditions.)
Temperature:
+10° C to + 40° C (+50° F to +104° F)
Humidity:
20% to 90% R.H., non-condensing
Air Pressure:
69 kPa to 110 kPa
C - 56
Appendix C: Specifications
Classifications
Protection Against Electrical Shock: Per IEC/EN 60601-1, the MEDRAD
Spectris Solaris EP MR Injection System is designed as Class 1 equipment
with Type BF applied parts.
Type BF corresponds to the degree of protection against electrical shock via
the applied parts. Class 1 applies to equipment which includes a means for
the connection of the equipment to protective earth in such a way that
accessible metal parts cannot become live in the event of failure of basic
insulation.
Flammable Anesthetics: The MEDRAD Spectris Solaris EP MR Injection
System is not suitable for use in the presence of a flammable anesthetic
mixture with air, oxygen, or nitrous oxide.
Protection Against Ingress of Fluids: Per IEC/EN 60601-1, the Scan Room
and Control Room Units have been classified as drip proof equipment. The
components of the MEDRAD Spectris Solaris EP MR Injection System Scan
Room and Control Room Units are provided with an enclosure that prevents
the entry of such an amount of falling liquid as might interfere with the safe
operation of the injector, indicated by the IPX1 designation. The battery
charger is not classified for protection against the ingress of fluids.
Mode of Operation: Per IEC/EN 60601-1 the mode of operation for the
Control Room Unit is continuous operation. It will operate under normal load
for an unlimited period, without excessive temperature being developed.
The Integrated Continuous Battery Charger power supply will operate under
normal load for an unlimited period, without excessive temperature being
developed.
The mode of operation for the Scan Room Unit is continuous operation with
intermittent loading. Although power is applied to the Scan Room Unit
continuously, intermittent use for loading and injecting will result in an internal
temperature less than continuous load operating temperatures, but greater
than no-load operating temperatures. Under normal operating conditions with
a minimum of 10 minutes between injections, the internal temperature of
the Scan Room Unit will not rise enough to degrade system performance or
reliability.
C - 57
MEDRAD Spectris Solaris EP MR Injection System
C - 58
Appendix D: Options and Accessories
Appendix D: Options and
Accessories
Catalog Number
Part Number
SPC 300A
SPC 300C
535-0243-012
535-0127-012
Integrated Continuous Battery Charger
System
3012080
3012080
Battery Charger Kit
3012424
3012424
Enhanced Battery Pack
3012070
3012070
Handswitch
SSMR START
3006265
Contrast Holder
Tray (optional)
CHD 100 MR
CHD 400 MR
404002295
404003227
Control Room Unit Mounting System
Height Adjustable Pedestal
Wall Mounting Bracket
SDP 300
SDW 300
401001277
401000775
Service Manual
SSMR-SERV
200880
Power Cord
American
Continental
D - 59
MEDRAD Spectris Solaris EP MR Injection System
D - 60
Appendix E: System Installation
Appendix E: System Installation
WARNINGS:
Serious injury or death may result from exposure to hazardous
voltages existing within the system. Use of an unapproved
extension cord, adapter, inverter, or multi-outlet strip may compromise
electrical safety. Plug the system directly into a properly grounded AC
outlet or contact MEDRAD for installation assistance.
Injury or equipment damage may result from improper placement
of the Battery Charger. Do not install the Battery Charger in the Scan
Room. Install the Battery Charger in the Control Room, or any
convenient location other than the Scan Room.
Serious injury or death can result from placing the Adjustable
Height Pedestal in the Scan Room. Do not install or operate the
Adjustable Height Pedestal in the Scan Room.
Injury or equipment damage may result from use of tools
containing ferrous materials. Use only non-magnetic tools to install
any scanner/magnet room components
CAUTIONS:
Condensation may cause electrical damage to the injection
system. Do not use the system immediately after it has been brought
indoors from extreme outside temperatures. Allow the system to
stabilize at room temperature before use.
Damage can occur as a result of incorrect voltage. Before
plugging in the system, check the following:
•
Verify that the voltage and frequency marked on the serial tag on
the back of the power supply matches the voltage and frequency
of the electrical outlet.
•
Verify that the injector has the appropriate power cord plug for the
power outlet.
Damage can occur to fiber optic cabling due to mishandling
during installation. Install following proper handling procedures or
contact MEDRAD Service for further assistance.
E - 61
MEDRAD Spectris Solaris EP MR Injection System
Unpacking the
Injection System
The entire standard configuration of the MEDRAD Spectris Solaris EP MR
Injection System is shipped in one shipping carton. The optional Control
Room Unit mounting accessories, the Adjustable Height Stand and the Wall
Mounting Bracket, along with the optional IV Pole, are packaged individually
and shipped in separate cartons. Prior to beginning installation, verify that the
following items are present:
Standard:
•
Scan Room Unit
•
Control Room Unit and power cord (110V or 220V)
•
Fiber Optic Interconnection Cable - 200 ft. (60.96 m)
•
Battery Charging Unit and power cord (110V or 220V) - with
mounting bracket and all hardware
•
System Batteries (2)
•
Hand Switch - with mounting bracket and all hardware
•
Syringe Kit (w/ 65 and 115 ml syringes)
•
Operation Manual
Optional*:
•
Additional Battery Charging Unit and power cord (110V or 220V) with mounting bracket and all hardware
•
Additional System Battery
•
Additional Hand Switch - with mounting bracket and all hardware
•
Service Manual
Optional (packaged separately)*:
•
Adjustable Height Pedestal for Control Room Unit
•
Wall Mounting Bracket for Control Room Unit
•
IV Pole for Scan Room Unit Mounting
•
Integrated Continuous Battery Charger System
*Refer to ”Instructions For Use” sheets for accessory and optional equipment
installation.
E - 62
Appendix E: System Installation
Installation
Considerations
WARNING: Injury or equipment damage may result from use of
tools containing ferrous materials. Use only non-magnetic tools to
install any scanner/magnet room components.
NOTE:
System Installation requires that the suite have a 1.5 inch
(3.81 cm) minimum tuned port (either separate or within
the penetration panel) available for connections between
the Scan and Control rooms.
NOTE:
Follow all institutional, local, or national safety regulations
related to routing cabling on the floor.
Tools Required: Non-magnetic #2 Phillips Head Screwdriver
E - 63
MEDRAD Spectris Solaris EP MR Injection System
Review the following general connections prior to installing the MEDRAD
Spectris Solaris EP MR Injection System. Be sure to consider all specifications and requirements outlined in Appendix C of this manual, and follow all
applicable regulations of your locality.
NOTE:
When using the Spectris Solaris EP MR Injector with a
Siemens MAGNETOM Avanto 1.5T or Siemens
MAGNETOM Espree 1.5T, it is recommended that the
Spectris Solaris EP be placed a minimum of 18 inches
from the facade of the scanner.
Scan Room
Control Room
Rear Cover View
Opening for Fiber
Optic Cable
Detail of Handswitch Connection
Detail of battery Charger Assembly
E - 64
Appendix E: System Installation
Fiber Optic Cable
Installation
Connection at the Scan Room Unit:
Special care should be taken with routing the fiber optic cable to ensure:
•
there is no bend radius less than 1 in. (2.54 cm)
•
the connector dust caps are not removed until connections are
made
•
the cable does not run across any sharp edges
•
the cable is run across a low traffic area of the floor
•
established standards for fiber optic cabling installation should be
followed if routing through conduit.
1. Remove the four screws that secure the covers to the lower console of
the Scan Room Unit.
2. Install the strain relief 17 inch (43 cm) from the tip of the fiber optic cable.
3. Route the fiber optic cable (the end with the strain relief) through the hole
in the lower corner of the rear console cover and snap the strain relief into
the hole.
4. Remove the caps from the fiber optic cable connectors, then establish the
connections at the lower console - TX to TX, and RX to RX.
5. Carefully re-position the covers on the lower console and secure with the
four screws previously removed. Ensure that there are no tight bends in
the fiber optic cable.
6. Gather and loop any extra fiber optic cable, then hang the loop from the
cable hanger on the Scan Room Unit column.
E - 65
MEDRAD Spectris Solaris EP MR Injection System
Strain Relief
Location
The strain relief should be placed 17" (43 cm) inches form the end of the Fiber
optic connectors. This ensures there is sufficient length to maneuver the
cable and minimize severe cable bends, which may damage the Fiber Optic
cable.
Fib
Strain
Fiber Optic
Relief
Rx/Tx
17 inches
(43 cm)
Recommended
Routing
Cable
Sheathing
The recommended Routing of the FO cable after it enters the SRU rear cover
is as follows:
1. Route the cable to the right side of the Electric box
2. Next, route the cable up about 4-6 inches (10-15 cm).
3. Make a gradual bend and connect the FO cable to the Electronic box.
(left side of drawing shows a gradual bend, right side of drawing shows a
severe bend)
Severe Bend
Gradual Bend
Cable Routing:
Route the fiber optic cable from the Scan Room Unit through the tuned port in
the wall between the Scan and Control Rooms. (The tuned port may be part
of the penetration panel connecting the two rooms).
NOTE:
Control Room Unit
Setup
Follow all institutional, local, or national safety regulations
related to routing cabling on the floor.
1. Position the Control Room Unit near an appropriate AC power outlet.
2. Remove the caps from the fiber optic cable connectors, then establish the
connections at the rear of the Control Room Unit - TX to TX, and RX to
RX.
3. Connect the Handswitch at either the Scan Room Unit, or Control Room
Unit Handswitch connection port.
NOTE:
Refer to the following procedure for Handswitch
mounting hardware installation instructions.
4. Attach the power cord to the Control Room Unit power inlet.
5. If required by local codes, connect the optional equal potential cable to
E - 66
Appendix E: System Installation
the equal potential stud, and the equal potential bus.
6. Plug the AC power cord into an appropriate AC power outlet.
7. Insert a fully charged system battery in the Scan Room Unit battery
receptacle
8. Apply system power at the Control Room Unit power switch, then perform
an system operation checkout as outlined in Appendix B of this manual.
Handswitch
Mounting
To allow convenience in use, the handswitch mount can be mounted in one of
the following ways:
On the wall:
Attach the mount to the bracket using the two screws supplied. Apply the
supplied double sided adhesive tape to the back of the solid side (without
holes) of the mounting bracket. Affix the mount/bracket assembly to a
properly primed wall to ensure maximum adhesion.
On the Scan Room Unit:
Route the supplied strap through the recesses in the back of the mount.
Secure the mount and strap to the upright column of the Scan Room Unit.
On either side of the Control Room Unit:
Determine which side of the unit the mount and bracket will be attached to.
Secure the mount to the bracket using the two screws supplied. Apply the
supplied double sided adhesive tape to the front of the solid side (without
holes) of the mounting bracket. Affix the mount/bracket assembly to the
backof the Control Room Unit display.
E - 67
MEDRAD Spectris Solaris EP MR Injection System
E - 68

Download Operation manual of Medrad Spectris Solaris EP Medical Equipment for Free or View it Online on All-Guides.com.

Medrad Spectris Solaris EP Operation manual - Page 1

1

Medrad Spectris Solaris EP Operation manual - Page 2

2

Medrad Spectris Solaris EP Operation manual - Page 3

3

Medrad Spectris Solaris EP Operation manual - Page 4

4

Medrad Spectris Solaris EP Operation manual - Page 5

5

Medrad Spectris Solaris EP Operation manual - Page 6

6

Medrad Spectris Solaris EP Operation manual - Page 7

7

Medrad Spectris Solaris EP Operation manual - Page 8

8

Medrad Spectris Solaris EP Operation manual - Page 9

9

Medrad Spectris Solaris EP Operation manual - Page 10

10

Medrad Spectris Solaris EP Operation manual - Page 11

11

Medrad Spectris Solaris EP Operation manual - Page 12

12

Medrad Spectris Solaris EP Operation manual - Page 13

13

Medrad Spectris Solaris EP Operation manual - Page 14

14

Medrad Spectris Solaris EP Operation manual - Page 15

15

Medrad Spectris Solaris EP Operation manual - Page 16

16

Medrad Spectris Solaris EP Operation manual - Page 17

17

Medrad Spectris Solaris EP Operation manual - Page 18

18

Medrad Spectris Solaris EP Operation manual - Page 19

19

Medrad Spectris Solaris EP Operation manual - Page 20

20

Medrad Spectris Solaris EP Operation manual - Page 21

21

Medrad Spectris Solaris EP Operation manual - Page 22

22

Medrad Spectris Solaris EP Operation manual - Page 23

23

Medrad Spectris Solaris EP Operation manual - Page 24

24

Medrad Spectris Solaris EP Operation manual - Page 25

25

Medrad Spectris Solaris EP Operation manual - Page 26

26

Medrad Spectris Solaris EP Operation manual - Page 27

27

Medrad Spectris Solaris EP Operation manual - Page 28

28

Medrad Spectris Solaris EP Operation manual - Page 29

29

Medrad Spectris Solaris EP Operation manual - Page 30

30

Medrad Spectris Solaris EP Operation manual - Page 31

31

Medrad Spectris Solaris EP Operation manual - Page 32

32

Medrad Spectris Solaris EP Operation manual - Page 33

33

Medrad Spectris Solaris EP Operation manual - Page 34

34

Medrad Spectris Solaris EP Operation manual - Page 35

35

Medrad Spectris Solaris EP Operation manual - Page 36

36

Medrad Spectris Solaris EP Operation manual - Page 37

37

Medrad Spectris Solaris EP Operation manual - Page 38

38

Medrad Spectris Solaris EP Operation manual - Page 39

39

Medrad Spectris Solaris EP Operation manual - Page 40

40

Medrad Spectris Solaris EP Operation manual - Page 41

41

Medrad Spectris Solaris EP Operation manual - Page 42

42

Medrad Spectris Solaris EP Operation manual - Page 43

43

Medrad Spectris Solaris EP Operation manual - Page 44

44

Medrad Spectris Solaris EP Operation manual - Page 45

45

Medrad Spectris Solaris EP Operation manual - Page 46

46

Medrad Spectris Solaris EP Operation manual - Page 47

47

Medrad Spectris Solaris EP Operation manual - Page 48

48

Medrad Spectris Solaris EP Operation manual - Page 49

49

Medrad Spectris Solaris EP Operation manual - Page 50

50

Medrad Spectris Solaris EP Operation manual - Page 51

51

Medrad Spectris Solaris EP Operation manual - Page 52

52

Medrad Spectris Solaris EP Operation manual - Page 53

53

Medrad Spectris Solaris EP Operation manual - Page 54

54

Medrad Spectris Solaris EP Operation manual - Page 55

55

Medrad Spectris Solaris EP Operation manual - Page 56

56

Medrad Spectris Solaris EP Operation manual - Page 57

57

Medrad Spectris Solaris EP Operation manual - Page 58

58

Medrad Spectris Solaris EP Operation manual - Page 59

59

Medrad Spectris Solaris EP Operation manual - Page 60

60

Medrad Spectris Solaris EP Operation manual - Page 61

61

Medrad Spectris Solaris EP Operation manual - Page 62

62

Medrad Spectris Solaris EP Operation manual - Page 63

63

Medrad Spectris Solaris EP Operation manual - Page 64

64

Medrad Spectris Solaris EP Operation manual - Page 65

65

Medrad Spectris Solaris EP Operation manual - Page 66

66

Medrad Spectris Solaris EP Operation manual - Page 67

67

Medrad Spectris Solaris EP Operation manual - Page 68

68

Medrad Spectris Solaris EP Operation manual - Page 69

69

Medrad Spectris Solaris EP Operation manual - Page 70

70

Medrad Spectris Solaris EP Operation manual - Page 71

71

Medrad Spectris Solaris EP Operation manual - Page 72

72

Medrad Spectris Solaris EP Operation manual - Page 73

73

Medrad Spectris Solaris EP Operation manual - Page 74

74

Medrad Spectris Solaris EP Operation manual - Page 75

75

Medrad Spectris Solaris EP Operation manual - Page 76

76

Medrad Spectris Solaris EP Operation manual - Page 77

77

Medrad Spectris Solaris EP Operation manual - Page 78

78

Medrad Spectris Solaris EP Operation manual - Page 79

79

Medrad Spectris Solaris EP Operation manual - Page 80

80

Medrad Spectris Solaris EP Operation manual - Page 81

81

Medrad Spectris Solaris EP Operation manual - Page 82

82

Medrad Spectris Solaris EP Operation manual - Page 83

83

Medrad Spectris Solaris EP Operation manual - Page 84

84

Medrad Spectris Solaris EP Operation manual - Page 85

85

Medrad Spectris Solaris EP Operation manual - Page 86

86

Medrad Spectris Solaris EP Operation manual - Page 87

87

Medrad Spectris Solaris EP Operation manual - Page 88

88

Medrad Spectris Solaris EP Operation manual - Page 89

89

Medrad Spectris Solaris EP Operation manual - Page 90

90

Medrad Spectris Solaris EP Operation manual - Page 91

91

Medrad Spectris Solaris EP Operation manual - Page 92

92

Medrad Spectris Solaris EP Operation manual - Page 93

93

Medrad Spectris Solaris EP Operation manual - Page 94

94

Medrad Spectris Solaris EP Operation manual - Page 95

95

Medrad Spectris Solaris EP Operation manual - Page 96

96

Medrad Spectris Solaris EP Operation manual - Page 97

97

Medrad Spectris Solaris EP Operation manual - Page 98

98

Medrad Spectris Solaris EP Operation manual - Page 99

99

Medrad Spectris Solaris EP Operation manual - Page 100

100

Medrad Spectris Solaris EP Operation manual - Page 101

101

Medrad Spectris Solaris EP Operation manual - Page 102

102

Medrad Spectris Solaris EP Operation manual - Page 103

103

Medrad Spectris Solaris EP Operation manual - Page 104

104

Medrad Spectris Solaris EP Operation manual - Page 105

105

Medrad Spectris Solaris EP Operation manual - Page 106

106

Medrad Spectris Solaris EP Operation manual - Page 107

107

Medrad Spectris Solaris EP Operation manual - Page 108

108

Medrad Spectris Solaris EP Operation manual - Page 109

109

Medrad Spectris Solaris EP Operation manual - Page 110

110

Medrad Spectris Solaris EP Operation manual - Page 111

111

Medrad Spectris Solaris EP Operation manual - Page 112

112

Medrad Spectris Solaris EP Operation manual - Page 113

113

Medrad Spectris Solaris EP Operation manual - Page 114

114

Medrad Spectris Solaris EP Operation manual - Page 115

115

Medrad Spectris Solaris EP Operation manual - Page 116

116

Medrad Spectris Solaris EP Operation manual - Page 117

117

Medrad Spectris Solaris EP Operation manual - Page 118

118

Medrad Spectris Solaris EP Operation manual - Page 119

119

Medrad Spectris Solaris EP Operation manual - Page 120

120

Medrad Spectris Solaris EP Operation manual - Page 121

121

Medrad Spectris Solaris EP Operation manual - Page 122

122

Medrad Spectris Solaris EP Operation manual - Page 123

123

Medrad Spectris Solaris EP Operation manual - Page 124

124

Medrad Spectris Solaris EP Operation manual - Page 125

125

Medrad Spectris Solaris EP Operation manual - Page 126

126

Medrad Spectris Solaris EP Operation manual - Page 127

127

Medrad Spectris Solaris EP Operation manual - Page 128

128

Medrad Spectris Solaris EP Operation manual - Page 129

129

Medrad Spectris Solaris EP Operation manual - Page 130

130

Medrad Spectris Solaris EP Operation manual - Page 131

131

Medrad Spectris Solaris EP Operation manual - Page 132

132

Medrad Spectris Solaris EP Operation manual - Page 133

133

Medrad Spectris Solaris EP Operation manual - Page 134

134

Medrad Spectris Solaris EP Operation manual - Page 135

135

Medrad Spectris Solaris EP Operation manual - Page 136

136

Medrad Spectris Solaris EP Operation manual - Page 137

137

Medrad Spectris Solaris EP Operation manual - Page 138

138

Medrad Spectris Solaris EP Operation manual - Page 139

139

Medrad Spectris Solaris EP Operation manual - Page 140

140

Medrad Spectris Solaris EP Operation manual - Page 141

141

Medrad Spectris Solaris EP Operation manual - Page 142

142

Medrad Spectris Solaris EP Operation manual - Page 143

143

Medrad Spectris Solaris EP Operation manual - Page 144

144

Medrad Spectris Solaris EP Operation manual - Page 145

145

Medrad Spectris Solaris EP Operation manual - Page 146

146

Medrad Spectris Solaris EP Operation manual - Page 147

147

Medrad Spectris Solaris EP Operation manual - Page 148

148

Medrad Spectris Solaris EP Operation manual - Page 149

149

Medrad Spectris Solaris EP Operation manual - Page 150

150

Medrad Spectris Solaris EP Operation manual - Page 151

151

Medrad Spectris Solaris EP Operation manual - Page 152

152

Medrad Spectris Solaris EP Operation manual - Page 153

153

Medrad Spectris Solaris EP Operation manual - Page 154

154

Medrad Spectris Solaris EP Operation manual - Page 155

155

Medrad Spectris Solaris EP Operation manual - Page 156

156

Medrad Spectris Solaris EP Operation manual - Page 157

157

Medrad Spectris Solaris EP Operation manual - Page 158

158

Medrad Spectris Solaris EP Operation manual - Page 159

159

Medrad Spectris Solaris EP Operation manual - Page 160

160

Medrad Spectris Solaris EP Operation manual - Page 161

161

Medrad Spectris Solaris EP Operation manual - Page 162

162

Medrad Spectris Solaris EP Operation manual - Page 163

163

Medrad Spectris Solaris EP Operation manual - Page 164

164

Medrad Spectris Solaris EP Operation manual - Page 165

165

Medrad Spectris Solaris EP Operation manual - Page 166

166

Medrad Spectris Solaris EP Operation manual - Page 167

167

Medrad Spectris Solaris EP Operation manual - Page 168

168

Medrad Spectris Solaris EP Operation manual - Page 169

169

Medrad Spectris Solaris EP Operation manual - Page 170

170

Medrad Spectris Solaris EP Operation manual - Page 171

171

Medrad Spectris Solaris EP Operation manual - Page 172

172

Medrad Spectris Solaris EP Operation manual - Page 173

173

Medrad Spectris Solaris EP Operation manual - Page 174

174

Medrad Spectris Solaris EP Operation manual - Page 175

175

Medrad Spectris Solaris EP Operation manual - Page 176

176

Medrad Spectris Solaris EP Operation manual - Page 177

177

Medrad Spectris Solaris EP Operation manual - Page 178

178

Medrad Spectris Solaris EP Operation manual - Page 179

179

Medrad Spectris Solaris EP Operation manual - Page 180

180

Medrad Spectris Solaris EP Operation manual - Page 181

181

Medrad Spectris Solaris EP Operation manual - Page 182

182

Medrad Spectris Solaris EP Operation manual - Page 183

183

Medrad Spectris Solaris EP Operation manual - Page 184

184

Medrad Spectris Solaris EP Operation manual - Page 185

185

Medrad Spectris Solaris EP Operation manual - Page 186

186

Medrad Spectris Solaris EP Operation manual - Page 187

187

Medrad Spectris Solaris EP Operation manual - Page 188

188

Medrad Spectris Solaris EP Operation manual - Page 189

189

Medrad Spectris Solaris EP Operation manual - Page 190

190

Medrad Spectris Solaris EP Operation manual - Page 191

191

Medrad Spectris Solaris EP Operation manual - Page 192

192

Medrad Spectris Solaris EP Operation manual - Page 193

193

Medrad Spectris Solaris EP Operation manual - Page 194

194

Medrad Spectris Solaris EP Operation manual - Page 195

195

Medrad Spectris Solaris EP Operation manual - Page 196

196

Medrad Spectris Solaris EP Operation manual - Page 197

197

Medrad Spectris Solaris EP Operation manual - Page 198

198

Medrad Spectris Solaris EP Operation manual - Page 199

199

Medrad Spectris Solaris EP Operation manual - Page 200

200

Medrad Spectris Solaris EP Operation manual - Page 201

201

Medrad Spectris Solaris EP Operation manual - Page 202

202

Medrad Spectris Solaris EP Operation manual - Page 203

203

Medrad Spectris Solaris EP Operation manual - Page 204

204

Medrad Spectris Solaris EP Operation manual - Page 205

205

Medrad Spectris Solaris EP Operation manual - Page 206

206

Medrad Spectris Solaris EP Operation manual - Page 207

207

Medrad Spectris Solaris EP Operation manual - Page 208

208

Medrad Spectris Solaris EP Operation manual - Page 209

209

Medrad Spectris Solaris EP Operation manual - Page 210

210

Medrad Spectris Solaris EP Operation manual - Page 211

211

Medrad Spectris Solaris EP Operation manual - Page 212

212

Medrad Spectris Solaris EP Operation manual - Page 213

213

Medrad Spectris Solaris EP Operation manual - Page 214

214

Medrad Spectris Solaris EP Operation manual - Page 215

215

Medrad Spectris Solaris EP Operation manual - Page 216

216

Medrad Spectris Solaris EP Operation manual - Page 217

217

Medrad Spectris Solaris EP Operation manual - Page 218

218

Medrad Spectris Solaris EP Operation manual - Page 219

219

Medrad Spectris Solaris EP Operation manual - Page 220

220

Medrad Spectris Solaris EP Operation manual - Page 221

221

Medrad Spectris Solaris EP Operation manual - Page 222

222

Medrad Spectris Solaris EP Operation manual - Page 223

223

Medrad Spectris Solaris EP Operation manual - Page 224

224

Medrad Spectris Solaris EP Operation manual - Page 225

225

Medrad Spectris Solaris EP Operation manual - Page 226

226

Medrad Spectris Solaris EP Operation manual - Page 227

227

Medrad Spectris Solaris EP Operation manual - Page 228

228

Medrad Spectris Solaris EP Operation manual - Page 229

229

Medrad Spectris Solaris EP Operation manual - Page 230

230

Medrad Spectris Solaris EP Operation manual - Page 231

231

Medrad Spectris Solaris EP Operation manual - Page 232

232

Medrad Spectris Solaris EP Operation manual - Page 233

233

Medrad Spectris Solaris EP Operation manual - Page 234

234

Medrad Spectris Solaris EP Operation manual - Page 235

235

Medrad Spectris Solaris EP Operation manual - Page 236

236

Medrad Spectris Solaris EP Operation manual - Page 237

237

Medrad Spectris Solaris EP Operation manual - Page 238

238

Medrad Spectris Solaris EP Operation manual - Page 239

239

Medrad Spectris Solaris EP Operation manual - Page 240

240

Medrad Spectris Solaris EP Operation manual - Page 241

241

Medrad Spectris Solaris EP Operation manual - Page 242

242

Medrad Spectris Solaris EP Operation manual - Page 243

243

Medrad Spectris Solaris EP Operation manual - Page 244

244

Medrad Spectris Solaris EP Operation manual - Page 245

245

Medrad Spectris Solaris EP Operation manual - Page 246

246

Medrad Spectris Solaris EP Operation manual - Page 247

247

Medrad Spectris Solaris EP Operation manual - Page 248

248

Medrad Spectris Solaris EP Operation manual - Page 249

249

Medrad Spectris Solaris EP Operation manual - Page 250

250

Medrad Spectris Solaris EP Operation manual - Page 251

251

Medrad Spectris Solaris EP Operation manual - Page 252

252

Medrad Spectris Solaris EP Operation manual - Page 253

253

Medrad Spectris Solaris EP Operation manual - Page 254

254

Medrad Spectris Solaris EP Operation manual - Page 255

255

Medrad Spectris Solaris EP Operation manual - Page 256

256

Medrad Spectris Solaris EP Operation manual - Page 257

257

Medrad Spectris Solaris EP Operation manual - Page 258

258

Medrad Spectris Solaris EP Operation manual - Page 259

259

Medrad Spectris Solaris EP Operation manual - Page 260

260

Medrad Spectris Solaris EP Operation manual - Page 261

261

Medrad Spectris Solaris EP Operation manual - Page 262

262

Medrad Spectris Solaris EP Operation manual - Page 263

263

Medrad Spectris Solaris EP Operation manual - Page 264

264

Medrad Spectris Solaris EP Operation manual - Page 265

265

Medrad Spectris Solaris EP Operation manual - Page 266

266

Medrad Spectris Solaris EP Operation manual - Page 267

267

Medrad Spectris Solaris EP Operation manual - Page 268

268

Medrad Spectris Solaris EP Operation manual - Page 269

269

Medrad Spectris Solaris EP Operation manual - Page 270

270

Medrad Spectris Solaris EP Operation manual - Page 271

271

Medrad Spectris Solaris EP Operation manual - Page 272

272

Medrad Spectris Solaris EP Operation manual - Page 273

273

Medrad Spectris Solaris EP Operation manual - Page 274

274

Medrad Spectris Solaris EP Operation manual - Page 275

275

Medrad Spectris Solaris EP Operation manual - Page 276

276

Medrad Spectris Solaris EP Operation manual - Page 277

277

Medrad Spectris Solaris EP Operation manual - Page 278

278

Medrad Spectris Solaris EP Operation manual - Page 279

279

Medrad Spectris Solaris EP Operation manual - Page 280

280

Medrad Spectris Solaris EP Operation manual - Page 281

281

Medrad Spectris Solaris EP Operation manual - Page 282

282

Medrad Spectris Solaris EP Operation manual - Page 283

283

Medrad Spectris Solaris EP Operation manual - Page 284

284

Medrad Spectris Solaris EP Operation manual - Page 285

285

Medrad Spectris Solaris EP Operation manual - Page 286

286

Medrad Spectris Solaris EP Operation manual - Page 287

287

Medrad Spectris Solaris EP Operation manual - Page 288

288

Medrad Spectris Solaris EP Operation manual - Page 289

289

Medrad Spectris Solaris EP Operation manual - Page 290

290

Medrad Spectris Solaris EP Operation manual - Page 291

291

Medrad Spectris Solaris EP Operation manual - Page 292

292

Medrad Spectris Solaris EP Operation manual - Page 293

293

Medrad Spectris Solaris EP Operation manual - Page 294

294

Medrad Spectris Solaris EP Operation manual - Page 295

295

Medrad Spectris Solaris EP Operation manual - Page 296

296

Medrad Spectris Solaris EP Operation manual - Page 297

297

Medrad Spectris Solaris EP Operation manual - Page 298

298

Medrad Spectris Solaris EP Operation manual - Page 299

299

Medrad Spectris Solaris EP Operation manual - Page 300

300

Medrad Spectris Solaris EP Operation manual - Page 301

301

Medrad Spectris Solaris EP Operation manual - Page 302

302

Medrad Spectris Solaris EP Operation manual - Page 303

303

Medrad Spectris Solaris EP Operation manual - Page 304

304

Medrad Spectris Solaris EP Operation manual - Page 305

305

Medrad Spectris Solaris EP Operation manual - Page 306

306

Medrad Spectris Solaris EP Operation manual - Page 307

307

Medrad Spectris Solaris EP Operation manual - Page 308

308

Medrad Spectris Solaris EP Operation manual - Page 309

309

Medrad Spectris Solaris EP Operation manual - Page 310

310

Medrad Spectris Solaris EP Operation manual - Page 311

311

Medrad Spectris Solaris EP Operation manual - Page 312

312

Medrad Spectris Solaris EP Operation manual - Page 313

313

Medrad Spectris Solaris EP Operation manual - Page 314

314

Medrad Spectris Solaris EP Operation manual - Page 315

315

Medrad Spectris Solaris EP Operation manual - Page 316

316

Medrad Spectris Solaris EP Operation manual - Page 317

317

Medrad Spectris Solaris EP Operation manual - Page 318

318

Medrad Spectris Solaris EP Operation manual - Page 319

319

Medrad Spectris Solaris EP Operation manual - Page 320

320

Medrad Spectris Solaris EP Operation manual - Page 321

321

Medrad Spectris Solaris EP Operation manual - Page 322

322

Medrad Spectris Solaris EP Operation manual - Page 323

323

Medrad Spectris Solaris EP Operation manual - Page 324

324

Medrad Spectris Solaris EP Operation manual - Page 325

325

Medrad Spectris Solaris EP Operation manual - Page 326

326

Medrad Spectris Solaris EP Operation manual - Page 327

327

Medrad Spectris Solaris EP Operation manual - Page 328

328

Medrad Spectris Solaris EP Operation manual - Page 329

329

Medrad Spectris Solaris EP Operation manual - Page 330

330

Medrad Spectris Solaris EP Operation manual - Page 331

331

Medrad Spectris Solaris EP Operation manual - Page 332

332

Medrad Spectris Solaris EP Operation manual - Page 333

333

Medrad Spectris Solaris EP Operation manual - Page 334

334

Medrad Spectris Solaris EP Operation manual - Page 335

335

Medrad Spectris Solaris EP Operation manual - Page 336

336

Medrad Spectris Solaris EP Operation manual - Page 337

337

Medrad Spectris Solaris EP Operation manual - Page 338

338

Medrad Spectris Solaris EP Operation manual - Page 339

339

Medrad Spectris Solaris EP Operation manual - Page 340

340

Medrad Spectris Solaris EP Operation manual - Page 341

341

Medrad Spectris Solaris EP Operation manual - Page 342

342

Medrad Spectris Solaris EP Operation manual - Page 343

343

Medrad Spectris Solaris EP Operation manual - Page 344

344

Medrad Spectris Solaris EP Operation manual - Page 345

345

Medrad Spectris Solaris EP Operation manual - Page 346

346

Medrad Spectris Solaris EP Operation manual - Page 347

347

Medrad Spectris Solaris EP Operation manual - Page 348

348

Medrad Spectris Solaris EP Operation manual - Page 349

349

Medrad Spectris Solaris EP Operation manual - Page 350

350

Medrad Spectris Solaris EP Operation manual - Page 351

351

Medrad Spectris Solaris EP Operation manual - Page 352

352

Medrad Spectris Solaris EP Operation manual - Page 353

353

Medrad Spectris Solaris EP Operation manual - Page 354

354

Medrad Spectris Solaris EP Operation manual - Page 355

355

Medrad Spectris Solaris EP Operation manual - Page 356

356

Medrad Spectris Solaris EP Operation manual - Page 357

357

Medrad Spectris Solaris EP Operation manual - Page 358

358

Medrad Spectris Solaris EP Operation manual - Page 359

359

Medrad Spectris Solaris EP Operation manual - Page 360

360

Medrad Spectris Solaris EP Operation manual - Page 361

361

Medrad Spectris Solaris EP Operation manual - Page 362

362

Medrad Spectris Solaris EP Operation manual - Page 363

363

Medrad Spectris Solaris EP Operation manual - Page 364

364

Medrad Spectris Solaris EP Operation manual - Page 365

365

Medrad Spectris Solaris EP Operation manual - Page 366

366

Medrad Spectris Solaris EP Operation manual - Page 367

367

Medrad Spectris Solaris EP Operation manual - Page 368

368

Medrad Spectris Solaris EP Operation manual - Page 369

369

Medrad Spectris Solaris EP Operation manual - Page 370

370

Medrad Spectris Solaris EP Operation manual - Page 371

371

Medrad Spectris Solaris EP Operation manual - Page 372

372

Medrad Spectris Solaris EP Operation manual - Page 373

373

Medrad Spectris Solaris EP Operation manual - Page 374

374

Medrad Spectris Solaris EP Operation manual - Page 375

375

Medrad Spectris Solaris EP Operation manual - Page 376

376

Medrad Spectris Solaris EP Operation manual - Page 377

377

Medrad Spectris Solaris EP Operation manual - Page 378

378

Medrad Spectris Solaris EP Operation manual - Page 379

379

Medrad Spectris Solaris EP Operation manual - Page 380

380

Medrad Spectris Solaris EP Operation manual - Page 381

381

Medrad Spectris Solaris EP Operation manual - Page 382

382

Medrad Spectris Solaris EP Operation manual - Page 383

383

Medrad Spectris Solaris EP Operation manual - Page 384

384

Medrad Spectris Solaris EP Operation manual - Page 385

385

Medrad Spectris Solaris EP Operation manual - Page 386

386

Medrad Spectris Solaris EP Operation manual - Page 387

387

Medrad Spectris Solaris EP Operation manual - Page 388

388

Medrad Spectris Solaris EP Operation manual - Page 389

389

Medrad Spectris Solaris EP Operation manual - Page 390

390

Medrad Spectris Solaris EP Operation manual - Page 391

391

Medrad Spectris Solaris EP Operation manual - Page 392

392

Medrad Spectris Solaris EP Operation manual - Page 393

393

Medrad Spectris Solaris EP Operation manual - Page 394

394

Medrad Spectris Solaris EP Operation manual - Page 395

395

Medrad Spectris Solaris EP Operation manual - Page 396

396

Medrad Spectris Solaris EP Operation manual - Page 397

397

Medrad Spectris Solaris EP Operation manual - Page 398

398

Medrad Spectris Solaris EP Operation manual - Page 399

399

Medrad Spectris Solaris EP Operation manual - Page 400

400

Medrad Spectris Solaris EP Operation manual - Page 401

401

Medrad Spectris Solaris EP Operation manual - Page 402

402

Medrad Spectris Solaris EP Operation manual - Page 403

403

Medrad Spectris Solaris EP Operation manual - Page 404

404

Medrad Spectris Solaris EP Operation manual - Page 405

405

Medrad Spectris Solaris EP Operation manual - Page 406

406

Medrad Spectris Solaris EP Operation manual - Page 407

407

Medrad Spectris Solaris EP Operation manual - Page 408

408

Medrad Spectris Solaris EP Operation manual - Page 409

409

Medrad Spectris Solaris EP Operation manual - Page 410

410

Medrad Spectris Solaris EP Operation manual - Page 411

411

Medrad Spectris Solaris EP Operation manual - Page 412

412

Medrad Spectris Solaris EP Operation manual - Page 413

413

Medrad Spectris Solaris EP Operation manual - Page 414

414

Medrad Spectris Solaris EP Operation manual - Page 415

415

Medrad Spectris Solaris EP Operation manual - Page 416

416

Medrad Spectris Solaris EP Operation manual - Page 417

417

Medrad Spectris Solaris EP Operation manual - Page 418

418

Medrad Spectris Solaris EP Operation manual - Page 419

419

Medrad Spectris Solaris EP Operation manual - Page 420

420

Medrad Spectris Solaris EP Operation manual - Page 421

421

Medrad Spectris Solaris EP Operation manual - Page 422

422

Medrad Spectris Solaris EP Operation manual - Page 423

423

Medrad Spectris Solaris EP Operation manual - Page 424

424

Medrad Spectris Solaris EP Operation manual - Page 425

425

Medrad Spectris Solaris EP Operation manual - Page 426

426

Medrad Spectris Solaris EP Operation manual - Page 427

427

Medrad Spectris Solaris EP Operation manual - Page 428

428

Medrad Spectris Solaris EP Operation manual - Page 429

429

Medrad Spectris Solaris EP Operation manual - Page 430

430

Medrad Spectris Solaris EP Operation manual - Page 431

431

Medrad Spectris Solaris EP Operation manual - Page 432

432

Medrad Spectris Solaris EP Operation manual - Page 433

433

Medrad Spectris Solaris EP Operation manual - Page 434

434

Medrad Spectris Solaris EP Operation manual - Page 435

435

Medrad Spectris Solaris EP Operation manual - Page 436

436

Medrad Spectris Solaris EP Operation manual - Page 437

437

Medrad Spectris Solaris EP Operation manual - Page 438

438

Medrad Spectris Solaris EP Operation manual - Page 439

439

Medrad Spectris Solaris EP Operation manual - Page 440

440

Medrad Spectris Solaris EP Operation manual - Page 441

441

Medrad Spectris Solaris EP Operation manual - Page 442

442

Medrad Spectris Solaris EP Operation manual - Page 443

443

Medrad Spectris Solaris EP Operation manual - Page 444

444

Medrad Spectris Solaris EP Operation manual - Page 445

445

Medrad Spectris Solaris EP Operation manual - Page 446

446

Medrad Spectris Solaris EP Operation manual - Page 447

447

Medrad Spectris Solaris EP Operation manual - Page 448

448

Medrad Spectris Solaris EP Operation manual - Page 449

449

Medrad Spectris Solaris EP Operation manual - Page 450

450

Medrad Spectris Solaris EP Operation manual - Page 451

451

Medrad Spectris Solaris EP Operation manual - Page 452

452

Medrad Spectris Solaris EP Operation manual - Page 453

453

Medrad Spectris Solaris EP Operation manual - Page 454

454

Medrad Spectris Solaris EP Operation manual - Page 455

455

Medrad Spectris Solaris EP Operation manual - Page 456

456

Medrad Spectris Solaris EP Operation manual - Page 457

457

Medrad Spectris Solaris EP Operation manual - Page 458

458

Medrad Spectris Solaris EP Operation manual - Page 459

459

Medrad Spectris Solaris EP Operation manual - Page 460

460

Medrad Spectris Solaris EP Operation manual - Page 461

461

Medrad Spectris Solaris EP Operation manual - Page 462

462

Medrad Spectris Solaris EP Operation manual - Page 463

463

Medrad Spectris Solaris EP Operation manual - Page 464

464

Medrad Spectris Solaris EP Operation manual - Page 465

465

Medrad Spectris Solaris EP Operation manual - Page 466

466

Medrad Spectris Solaris EP Operation manual - Page 467

467

Medrad Spectris Solaris EP Operation manual - Page 468

468

Medrad Spectris Solaris EP Operation manual - Page 469

469

Medrad Spectris Solaris EP Operation manual - Page 470

470

Medrad Spectris Solaris EP Operation manual - Page 471

471

Medrad Spectris Solaris EP Operation manual - Page 472

472

Medrad Spectris Solaris EP Operation manual - Page 473

473

Medrad Spectris Solaris EP Operation manual - Page 474

474

Medrad Spectris Solaris EP Operation manual - Page 475

475

Medrad Spectris Solaris EP Operation manual - Page 476

476

Medrad Spectris Solaris EP Operation manual - Page 477

477

Medrad Spectris Solaris EP Operation manual - Page 478

478

Medrad Spectris Solaris EP Operation manual - Page 479

479

Medrad Spectris Solaris EP Operation manual - Page 480

480

Medrad Spectris Solaris EP Operation manual - Page 481

481

Medrad Spectris Solaris EP Operation manual - Page 482

482

Medrad Spectris Solaris EP Operation manual - Page 483

483

Medrad Spectris Solaris EP Operation manual - Page 484

484

Medrad Spectris Solaris EP Operation manual - Page 485

485

Medrad Spectris Solaris EP Operation manual - Page 486

486

Medrad Spectris Solaris EP Operation manual - Page 487

487

Medrad Spectris Solaris EP Operation manual - Page 488

488

Medrad Spectris Solaris EP Operation manual - Page 489

489

Medrad Spectris Solaris EP Operation manual - Page 490

490

Medrad Spectris Solaris EP Operation manual - Page 491

491

Medrad Spectris Solaris EP Operation manual - Page 492

492

Medrad Spectris Solaris EP Operation manual - Page 493

493

Medrad Spectris Solaris EP Operation manual - Page 494

494

Medrad Spectris Solaris EP Operation manual - Page 495

495

Medrad Spectris Solaris EP Operation manual - Page 496

496

Medrad Spectris Solaris EP Operation manual - Page 497

497

Medrad Spectris Solaris EP Operation manual - Page 498

498

Medrad Spectris Solaris EP Operation manual - Page 499

499

Medrad Spectris Solaris EP Operation manual - Page 500

500

Medrad Spectris Solaris EP Operation manual - Page 501

501

Medrad Spectris Solaris EP Operation manual - Page 502

502

Medrad Spectris Solaris EP Operation manual - Page 503

503

Medrad Spectris Solaris EP Operation manual - Page 504

504

Medrad Spectris Solaris EP Operation manual - Page 505

505

Medrad Spectris Solaris EP Operation manual - Page 506

506

Medrad Spectris Solaris EP Operation manual - Page 507

507

Medrad Spectris Solaris EP Operation manual - Page 508

508

Medrad Spectris Solaris EP Operation manual - Page 509

509

Medrad Spectris Solaris EP Operation manual - Page 510

510

Medrad Spectris Solaris EP Operation manual - Page 511

511

Medrad Spectris Solaris EP Operation manual - Page 512

512

Medrad Spectris Solaris EP Operation manual - Page 513

513

Medrad Spectris Solaris EP Operation manual - Page 514

514

Medrad Spectris Solaris EP Operation manual - Page 515

515

Medrad Spectris Solaris EP Operation manual - Page 516

516

Medrad Spectris Solaris EP Operation manual - Page 517

517

Medrad Spectris Solaris EP Operation manual - Page 518

518

Medrad Spectris Solaris EP Operation manual - Page 519

519

Medrad Spectris Solaris EP Operation manual - Page 520

520

Medrad Spectris Solaris EP Operation manual - Page 521

521

Medrad Spectris Solaris EP Operation manual - Page 522

522

Medrad Spectris Solaris EP Operation manual - Page 523

523

Medrad Spectris Solaris EP Operation manual - Page 524

524

Medrad Spectris Solaris EP Operation manual - Page 525

525

Medrad Spectris Solaris EP Operation manual - Page 526

526

Medrad Spectris Solaris EP Operation manual - Page 527

527

Medrad Spectris Solaris EP Operation manual - Page 528

528

Medrad Spectris Solaris EP Operation manual - Page 529

529

Medrad Spectris Solaris EP Operation manual - Page 530

530

Medrad Spectris Solaris EP Operation manual - Page 531

531

Medrad Spectris Solaris EP Operation manual - Page 532

532

Medrad Spectris Solaris EP Operation manual - Page 533

533

Medrad Spectris Solaris EP Operation manual - Page 534

534

Medrad Spectris Solaris EP Operation manual - Page 535

535

Medrad Spectris Solaris EP Operation manual - Page 536

536

Medrad Spectris Solaris EP Operation manual - Page 537

537

Medrad Spectris Solaris EP Operation manual - Page 538

538

Medrad Spectris Solaris EP Operation manual - Page 539

539

Medrad Spectris Solaris EP Operation manual - Page 540

540

Medrad Spectris Solaris EP Operation manual - Page 541

541

Medrad Spectris Solaris EP Operation manual - Page 542

542

Medrad Spectris Solaris EP Operation manual - Page 543

543

Medrad Spectris Solaris EP Operation manual - Page 544

544

Medrad Spectris Solaris EP Operation manual - Page 545

545

Medrad Spectris Solaris EP Operation manual - Page 546

546

Medrad Spectris Solaris EP Operation manual - Page 547

547

Medrad Spectris Solaris EP Operation manual - Page 548

548

Medrad Spectris Solaris EP Operation manual - Page 549

549

Medrad Spectris Solaris EP Operation manual - Page 550

550

Medrad Spectris Solaris EP Operation manual - Page 551

551

Medrad Spectris Solaris EP Operation manual - Page 552

552

Medrad Spectris Solaris EP Operation manual - Page 553

553

Medrad Spectris Solaris EP Operation manual - Page 554

554

Medrad Spectris Solaris EP Operation manual - Page 555

555

Medrad Spectris Solaris EP Operation manual - Page 556

556

Medrad Spectris Solaris EP Operation manual - Page 557

557

Medrad Spectris Solaris EP Operation manual - Page 558

558

Medrad Spectris Solaris EP Operation manual - Page 559

559

Medrad Spectris Solaris EP Operation manual - Page 560

560

Medrad Spectris Solaris EP Operation manual - Page 561

561

Medrad Spectris Solaris EP Operation manual - Page 562

562

Medrad Spectris Solaris EP Operation manual - Page 563

563

Medrad Spectris Solaris EP Operation manual - Page 564

564

Medrad Spectris Solaris EP Operation manual - Page 565

565

Medrad Spectris Solaris EP Operation manual - Page 566

566

Medrad Spectris Solaris EP Operation manual - Page 567

567

Medrad Spectris Solaris EP Operation manual - Page 568

568

Medrad Spectris Solaris EP Operation manual - Page 569

569

Medrad Spectris Solaris EP Operation manual - Page 570

570

Medrad Spectris Solaris EP Operation manual - Page 571

571

Medrad Spectris Solaris EP Operation manual - Page 572

572

Medrad Spectris Solaris EP Operation manual - Page 573

573

Medrad Spectris Solaris EP Operation manual - Page 574

574

Medrad Spectris Solaris EP Operation manual - Page 575

575

Medrad Spectris Solaris EP Operation manual - Page 576

576

Medrad Spectris Solaris EP Operation manual - Page 577

577

Operation Manual

Language Key

English 1–24

᪥ᮏㄒ 25–48

Deutsch 49–71

Français 72–94

Italiano 95–118

Español 119–142

Português 143–166

Nederlands 167–190

Svenska 191–214

Language Key

Dansk 215–238

Norsk 239–262

Suomi 263–286

Ελληνικά 287–310

Русский 311–334

Slovenský 335–358

Български 359–382

Türkçe 383–405

Česky 407–430

Language Key

ㆰփѝ᮷ 431–454

Hrvatski 455–478

Magyar 479–502

Polski 503–526

Română 527–550

Slovenščina 551–574

Integrated Continuous Battery Charging System

Понравилась статья? Поделить с друзьями:
  • Medohar vati инструкция по применению на русском языке отзывы
  • Media converter 10 100base tx to 100base fx инструкция
  • Medebiotin fuerte инструкция на русском
  • Medi sas 15 инструкция по применению
  • Medan shiffa home инструкция по применению