  • Page 2: Blood Glucose Monitoring System

    The VivaChek™ Ino Blood Glucose Control Solution is for use with the VivaChek™ Ino Plus Blood Glucose Meter and VivaChek™ Ino Strips as a quality control check to verify that the meter and test strip are working together properly, and that the…

  • Page 3: Table Of Contents

    TABLE OF CONTENTS Understanding Your Testing Tools………………………………….. Your Meter System Overview………………………………………….. Your Meter Display………………………………………………………. Important Safety Information……………………………………………… 4 Limitations…………………………………………………………………… 5 Setting Up Your System……………………………………………….. 6 Set the Time and Date…………………………………………………..6 Pairing your meter with the Smartphone………………………………. 7 Syncing to send results wirelessly to the app…………………………… 7 Taking a Test……………………………………………………………..

  • Page 4: Understanding Your Testing Tools

    CHARPTER 1: UNDERSTANDING YOUR TESTING TOOLS Your Meter System Overview The VivaChek™ Ino Plus Blood Glucose Meter and Strip…

  • Page 5: Your Meter Display

    All of the segments should be clear and exactly like the picture below. If not, contact VivaChek Customer Support at 800-951-8430 (7 days a week. 9 a.m.-5 p.m. Eastern Time) for help. Please contact your health care professional if you need help.

  • Page 6
    Indicates that a low test result may cause hypoglycaemia. Bluetooth Sound on or off Notes: Your VivaChek™ Ino Plus meter is pre-set with beep sound function, the meter will beep when: turn on the meter. • the test strip is inserted and ready to apply blood or control solution.
  • Page 7: Important Safety Information

    Check the expiration dates and discard dates on your test strips vial label • (or on the foil pouch) and control solution bottle label. Use only VivaChek™ Ino test strip with your VivaChek™ Ino Plus meter. • Use only VivaChek™ Ino control solution with your VivaChek™ Ino Plus •…

  • Page 8: Limitations

    Blood samples from patients in shock, or with severe dehydration or from • patients in a hyperosmolar state (with or without ketosis) have not been tested and are not recommended for testing with VivaChek™ Ino Plus Blood Glucose Monitoring System. The VivaChek™ Ino Plus Blood Glucose Monitoring System should not be •…

  • Page 9: Setting Up Your System

    You will hear a long beep. Pairing your meter with the Smartphone Pairing prepares your VivaChek™ Ino Plus Meter and Smartphone to communicate with each other. The devices must be within 5 meters of each other to pair and sync.

  • Page 10: Pairing Your Meter With The Smartphone

    4. Look for a device named VivaChek. This is your meter. 5. Tap on the entry that represents your VivaChek™ Ino Plus meter. 6. Enter your meter passcode using the keypad. The passcode is the last 5 digits from the Serial Number on the back of your meter.

  • Page 11: Taking A Test

    Year-Month format. 2019/01 indicates January, 2019. Your VivaChek™ Ino test strips have 4 months shelf life after you first open the test strip vial. Write the discard date on the vial label when you first open it. Make sure the test strip does not appear damaged.

  • Page 12: Getting A Blood Drop And Testing

    Getting a Blood Drop and Testing Here is an example of how to use a VivaChek™ Safety Lancet for fingertip blood sampling. If you use a different single-use auto-disabling safety lancet or lancing device, please follow the manufacturer’s instructions. Carefully rotate and pull off the protective cap.

  • Page 13: Discard The Used Test Strip

    Immediately touch the tip of the test strip to the drop of blood. The blood pulled into the test strip through the tip. Make sure that the blood sample has been fully filled the check window on the strip tip. Hold the tip of the test strip in the blood drop until the meter beeps.

  • Page 14: Testing With Control Solution

    • Do not change your medication therapy based on VivaChek™ Ino Plus test result before consult your health care provider. If you want to mark your result (only in Advanced Mode (L-2)) you will need •…

  • Page 15
    • Your test result does not match with how you feel • About the Control Solutions Only use VivaChek™ Ino Control Solutions (1, 2 or 3) to practice on the • system. • Your meter automatically recognizes the control solution.
  • Page 16
    Note: Do not apply control solution to the test strip directly from the bottle. Immediately touch the tip of the test strip to the drop of control solution. The control solution is pulled into the test strip through the strip tip. Note: If the control solution sample does not fill the check window, do not add a second drop.
  • Page 17: Using The Meter Memory

    (or the foil pouch). • Make sure the control solution was mixed well. • Confirm that you are using VivaChek™ Ino brand control solution. • Make sure that you followed the test procedure correctly. After checking all of the conditions listed above, repeat the control solution test with a new test strip.

  • Page 18: Maintenance And Troubleshooting

    Use only Clorox Healthcare Bleach Germicidal Wipes, which has been proven to be safe to use with the VivaChek™ Ino Plus Blood Glucose Meter. Cleaning is part of your normal care and maintenance and should be performed…

  • Page 19
    Clorox Healthcare Bleach Germicidal Wipes containing Sodium hypochlorite 0.55%, which has been proven to be safe to use with the VivaChek™ Ino Plus system. Clorox Healthcare Bleach Germicidal Wipes are available through CLOROX PROFESSIONAL PRODUCTS CO.
  • Page 20: Troubleshooting Guide

    Customer Support for help. If you have questions about cleaning or disinfecting, or if you see evidence of physical damage, contact VivaChek Customer Sopport at 800-951-8430 (available 7 days a week, 8 a.m.- 5 p.m. Eastern Time). Please contact your health care professional if you need help.

  • Page 21
    Please contact your health care professional if you need help. Repeat test apply enough Insufficient sample. sample to fill the test strip check window. Replace new batteries. Indicates a low battery. Test result above Wash and dry your hands well and 600 mg/dL.
  • Page 22: Symptoms Of High Or Low Blood Glucose

    Symptoms of High or Low Blood Glucose You can better understand your test results by being aware of the symptoms of high or low blood glucose. According to the American Diabetes Association, some of the most common symptoms are: High blood glucose Low blood glucose (Hyperglycemia): (Hypoglycemia):…

  • Page 23: Technical Information

    CHAPTER 5: TECHNICAL INFORMATION System Specifications: Feature Specification Measurement range 45 to 600 mg/dL Result calibration Plasma-equivalent Sample Fresh capillary whole blood Sample volume: 0.8 µL Test time 5 seconds Power source Two AAA LR03 1.5V batteries Glucose units of measure The meter is pre-set to milligrams per deciliter (mg/dL) Memory Up to 1000 records with date and time…

  • Page 24: Warranty

    Warranty Please complete the warranty card that came with this product and mail it to VivaChek Laboratories, Inc. 913 N market Street Wilmington, De, 19801, USA If the meter fails to work for any reason other than obvious abuse within the first five (5) years from purchase, we will replace it with a new meter free of charge.

Глюкометр VivaChek® Ino

Система VivaChek™ Ino предназначена для количественного измерения концентрации глюкозы в свежей капиллярной цельной крови.Система VivaChek™ Ino полностью соответствует новому международному стандарту EN ISO 15197:2015, предназначена для удобной проверки уровня глюкозы в крови, и помогает держать Ваш уровень глюкозы в крови под контролем.

ü Базовый режим и расширенный режим для выбора

ü Отсутствие технологии кодирования, меньше этапов тестирования, больше доверия к тестированию

ü Съемный эжектор плавно отбрасывает использованную полоску во избежание загрязнения

ü Счетчик памяти: 900 с датой и временем, 7,14 и 30 дней усреднения

ü Технология проприетарной полосы приводит к более точному результату в широком диапазоне (20-70%) уровня HCT

üРезультат передачи порта Micro-USB на смарт-терминал для поддержки распознавания шаблонов и определения того, как лечение и образ жизни влияют на уровень глюкозы

#товара: 12127661577

Состояние Новый

Счет-фактура Я выставляю счет-фактуру НДС

Вес с индивидуальной упаковкой 0.25 кг

Измерение глюкозы

Бренд Вивачек

Модель Ино Смарт

Код производителя VGM04

Объем образца крови 0.5 мкл

Время измерения 5 s

Средняя измерений 90 дней

Диапазон измерений глюкозы 10-600 мг/дл

Число измерений сохраняются Пятьсот

Количество ремней Десять

Количество lancetów Десять

приложение для смартфона

автоматическое кодирование

автоматическое отключение

Bluetooth

датчик минимального объема пробы крови

ремни самовсасывающие

дисплей с подсветкой

выбор глубины прокола

маркировка измерений

батарейки

дневник

чехол

ланцеты

прокол

ремни

жидкость контрольная

Питание батарейки

Количество
168 штук

  • Количество

  • Проблемы? Сомнения? Вопросы? Задайте вопрос!

    i

    Blood Glucose Monitoring System

    Dear VivaChek™ Ino Sound System User,

    Thank you for choosing the VivaChek™ Ino Sound Blood Glucose Monitoring

    System! VivaChek™ Ino Sound Blood Glucose Monitoring System is designed for

    easy test of blood glucose and helps you keep blood glucose under control.

    Read this User’s Manual carefully before you use your meter system. This manual

    will help you to get comfortable using the VivaChek™ Ino Sound Blood Glucose

    Monitoring System and get reliable test results. Please keep your User’s Manual in

    a safe place; you may want to refer it in the future.

    Thank you again for choosing the VivaChek™ Ino Sound System.

    Principle and Intended Use

    VivaChek™ Ino Sound Blood Glucose Monitoring System is comprised of the

    VivaChek™ Ino Sound Blood Glucose Meter and the VivaChek™ Ino Blood

    Glucose Test Strips. The VivaChek™ Ino Sound Blood Glucose Monitoring System

    (Meter Model: VGM09) is designed to quantitatively measure the glucose

    concentration in fresh capillary whole blood. It is used at home as a way to monitor

    the effectiveness of diabetes control programs. This system is intended to be used

    by a single patient and should not be shared.

    The VivaChek™ Ino Sound Blood Glucose Monitoring System is not used for the

    diagnosis of or screening of diabetes or for neonatal use.

    The VivaChek™ Ino Blood Glucose Control Solution is for use with the VivaChek

    Ino Sound Blood Glucose Meter and VivaChek™ Ino Strips as a quality control

    check to verify that the meter and test strip are working together properly, and that

    the test is performing correctly.

    ii

    TABLE OF CONTENTS

    1. Understanding Your Testing Tools………………………………….. 1

    Your Meter System Overview………………………………………….. 1

    Your Meter Display………………………………………………………. 2

    Important Safety Information…………………………………………

    4

    Limitations…………………………………………………………………

    5

    2. Setting Up Your System……………………………………………….. 6

    Set the Time and Date……………………………………………………

    Set sound feature………….……………………………………………….

    Pairing your meter with the Smartphone……………………………….

    Using your meter without the app

    3. Taking a Test…………………………………………………………….. 9

    Preparing the Test Strip…………………………………………………. 9

    Getting a Blood Drop and Testing………………………………………

    10

    Discard the Used Test Strip..…………………………………………… 11

    Testing with Control Solution…………………………….…………………

    15

    Using the Meter Memory………………………………………………….. 17

    4. Maintenance and Troubleshooting………………………………… 18

    Replace Batteries………………………………………………………….

    Caring for Your Glucose Monitoring System…………………………….

    Cleaning and Disinfection………………………………………………….

    18

    Troubleshooting Guide……………………………………………………..

    Symptoms of High or Low Blood Glucose……………………………….

    5. Technical Information……………………………………………………. 23

    System Specifications……………………………………………………

    23

    Warranty……………………………………………………………………..

    24

    1

    CHARPTER 1: UNDERSTANDING YOUR TESTING TOOLS

    Your Meter System Overview

    The VivaChek™ Ino Sound Blood Glucose Meter and Strip

    2

    Your Meter Display

    The picture below shows all the symbols that appear on your meter display.

    Please make sure the display is working properly before testing. When the meter is

    off, press and hold for 3 seconds to see the complete display. All display

    segments will appear for 10 seconds. If you need more than 10 seconds to check

    the display, repeat the above operation.

    All of the segments should be clear and exactly like the picture below. If not,

    contact VivaChek Customer Support at 8009518430 (7 days a week. 9 a.m.5

    p.m. Eastern Time) for help. Please contact your health care professional if you

    need help.

    Icon What it Means

    The top left area on the screen indicates the date.

    The top right area on the screen indicates the time.

    Date, month.

    Indicates average memory

    Indicates the alarms have been set.

    Indicates a low battery or it needs to be replaced.

    3

    Center area on the display that shows test results or error

    codes.

    Indicates the system is ready to test.

    Control test result.

    Pre-meal marker.

    Post-meal marker.

    Test results are displayed as mg/dL or mmol/L according to

    local government regulation.

    Indicates the temperature is not suitable for testing.

    Ketone warning.

    Indicates that a low test result may cause hypoglycaemia.

    Bluetooth

    Sound on or off

    Notes:

    Your VivaChek™ Ino Sound meter is preset with beep sound function, the meter

    will beep when:

    turn on the meter.

    the test strip is inserted and ready to apply blood or control solution.

    sufficient blood or control solution is pulled into the test strip.

    the test is complete.

    it is time to perform a test if you set the test alarms.

    if any error occurs during operation.

    Meter Use and Precautions

    The meter is preset to display blood glucose concentration in either millimoles per

    liter (mmol/L) or milligrams per deciliter (mg/dL). This depends on your country’s

    standard unit of measurement. This unit of measure cannot be adjusted. The

    meter will be set to mg/dL by default when sold in the United States.

    Meter will shut off by itself after 2 minutes of inactivity.

    4

    Do not get water or other liquids inside the meter.

    Keep the strip port area clean.

    Keep your meter dry. Avoid exposing it to extreme temperatures or humidity. Do

    not leave it in your car.

    Do not drop the meter or get it wet. If you do, check the meter by running a

    quality control test. Refer to Quality Control Test for instructions.

    Do not take the meter apart. This will void the warranty.

    Refer to the Caring for Your Meter section for details on cleaning the meter.

    Keep the meter and all associated parts out of reach of children.

    Note: Follow proper precautions and all local regulations when disposing of the

    meter and used batteries.

    Important Safety Information

    The meter is single patient use. Do not share them with anyone including

    other family members! Do not use on multiple patients!

    Always keep the test strips in the original vial. Tightly close the vial

    immediately after you have removed the test strip.

    Do not use the meter if it is dropping into water or splashing water on to it.

    Wash and dry your hands well before and after testing.

    Test strips and Safety lancets are for single use only.

    Do not drop blood directly on the flat surface of the test strip.

    Check the expiration dates and discard dates on your test strips vial label

    (or on the foil pouch) and control solution bottle label.

    Use only VivaChek™ Ino test strip with your VivaChek™ Ino Sound meter.

    Use only VivaChek™ Ino control solution with your VivaChek™ Ino Sound

    meter and VivaChek™ Ino strip.

    For selftesters, consult your physician or diabetes health care professional

    before making any adjustments to your medication, diet or activity routines.

    If the system is used in a manner not specified by the manufacturer,

    the protection provided by the system can be impaired.

    Potential Biohazard

    All parts of the kit are considered biohazardous and can potentially transmit

    infectious diseases, even after you have performed cleaning and disinfection.

    Note:

    1. The meter and safety lancet are for single patient use. Do not share them with

    anyone!

    2. All parts of the kit are considered biohazardous. They can potentially transmit

    infectious diseases from blood borne pathogens, even after you have

    performed cleaning and disinfection. Please follow proper precautions when

    handling your meter and lancing device.

    3. Crossuse of meter and lancing device may cause Bloodborne Pathogens

    Transmission. For more information, please refer to the FDA Public Health

    Notification: Use of Fingerstick Devices on More than One Person Poses Risk

    for Transmitting Bloodborne Pathogens: Initial Communication” (2010) at

    5

    http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm.

    You may also refer to the CDC Clinical Reminder: Use of Fingerstick Devices

    on More than One Person Poses Risk for Transmitting Bloodborne Pathogens”

    (2010) at http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html.

    Limitations

    Very high (above 70%) and very low (below 20%) hematocrit levels can

    cause false results. Talk to your health care professional to find out your

    hematocrit level.

    If you are taking vitamin C (ascorbic acid > 3 mg/dL) then your glucose

    results using this meter may not be reliable.

    The system is tested to accurately read the measurement of glucose in

    whole blood within the range of 45 to 600 mg/dL.

    The VivaChek™ Ino Sound Blood Glucose Monitoring System has been

    tested and shown to work properly up to 9777ft (2980 meters).

    Critically ill patients should not be tested with this system.

    Patient who is taking oxygen therapy is not recommended for testing with

    VivaChek™ Ino Sound Blood Glucose Monitoring System.

    Blood samples from patients in shock, or with severe dehydration or from

    patients in a hyperosmolar state (with or without ketosis) have not been

    tested and are not recommended for testing with VivaChek™ Ino Sound

    Blood Glucose Monitoring System.

    The VivaChek™ Ino Sound Blood Glucose Monitoring System should not

    be used following xylose absorption procedures.

    Not for neonatal use.

    Not for screening or diagnosis of diabetes mellitus.

    Note: This device complies with part 15 of the FCC Rules. Operation is subject to the following

    two conditions:

    (1) This device may not cause harmful interference, and

    (2) This device must accept any interference received, including interference that may cause

    undesired operation.

    Please note that changes or modifications not expressly approved by the party responsible for

    compliance could void the user’s authority to operate the equipment.

    Note: This equipment has been tested and found to comply with the limits for a Class B digital

    device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable

    protection against harmful interference in a residential installation. This equipment generates,

    uses and can radiate radio frequency energy and, if not installed and used in accordance with

    the instructions, may cause harmful interference to radio communications. However, there is no

    guarantee that interference will not occur in a particular installation. If this equipment does cause

    harmful interference to radio or television reception, which can be determined by turning the

    equipment off and on, the user is encouraged to try to correct the interference by one or more of

    the following measures:

    Reorient or relocate the receiving antenna.

    Increase the separation between the equipment and receiver.

    Connect the equipment into an outlet on a circuit different from that to which the receiver

    is connected.

    Consult the dealer or an experienced radio/TV technician for help.

    6

    CHAPTER 2: SETTING UP YOUR SYSTEM

    Before you first time using your meter or if you change your meter batteries, you

    should check and update your meter settings.

    Set the Time and Date

    1. Enter the setting mode and set the date

    The year will now flash. Press side button to adjust it then press and hold power

    button until the meter beeps to set. Then it will shift to next digit for setting. Repeat

    the above action until the year setting is completed.

    The month will now flash. Press side button to adjust the month, press and hold

    power button until the meter beeps to set.

    The date will now flash. Press side button to adjust the date, press and hold

    power button until the meter beeps to set. Then it will shift to next digit for setting.

    Repeat the above action until the year setting is completed.

    Note:

    Before you first time use your meter system for testing, please adjust the meter

    settings to set the date and time, ensuring that results stored in the memory are

    shown with the correct date and time.

    2. Set the Time

    The hour will now flashing. Press side button to adjust the current hour, press and

    hold power button until the meter beeps to set. Then it will shift to next digit for

    setting. Repeat the above action until the hour setting is completed.

    The minute will now flash. Press side button to adjust the minute, press and hold

    power button to set. You will hear a long beep.

    Set the sound feature

    After set the time, the and Onwill now flash. Press side button to selectOn

    or OFF”. Press and hold power button to set.

    Now you have completed your meter set up. A symbol of a test strip appears letting

    you know the meter is ready to test.

    7

    Pairing your meter with the Smartphone

    Pairing prepares your VivaChek™ Ino Sound Meter and Smartphone to

    communicate with each other. The devices must be within 5 meters of each other

    to pair and sync. Download the GlucoWell™ app before pairing your meter and

    Smartphone.

    Note: it is compatible with iPhone mobile phones with iOS 8.0 or above operating

    system.

    WARNING

    DO NOT pair another person’s meter with your Smartphone.

    To pair the Smartphone with your meter, turn your meter on and follow these steps:

    1. When you turn on your meter, the ( ) symbol will appear to indicate

    Bluetooth® is on.

    2. To turn on Bluetooth® on your Smartphone tap the Settings icon on the home

    screen.

    3. Next select Bluetooth and set Bluetooth to ON. Your devices are now

    searching for each other.

    4. Look for a device named VivaChek. This is your meter.

    5. Tap on the entry that represents your VivaChek™ Ino Sound meter.

    6. Enter your meter passcode using the keypad. The passcode is the last 5 digits

    from the Serial Number on the back of your meter.

    7. Tap Pair

    8. Wait for the Smartphone status to say “Connected”. Congratulations! Your

    meter and Smartphone are now paired. Before testing, you must now initially

    sync the meter with the app.

    NOTE: The button will continue to blink with blue light after successful pairing.

    Using your meter without the app

    The meter can be used without a Smartphone or the app. You can still test your

    blood glucose and review your results on the meter screen.

    Your meter automatically stores up to 1000 results with the time and date. Test

    results are stored from the newest to the oldest. If there are already 1000 records

    in memory, the oldest record will be erased to make room for a new one.

    8

    CHAPTER 3: TAKING A TEST

    Set up your meter correctly and have all the materials you will need ready before

    you begin testing. This including your VivaChek™ Ino Sound meter, the

    VivaChek™ Ino test strips and VivaChek™ Safety Lancets.

    Preparing the Test Strip

    1. Wash and dry your hands well before testing.

    2. Remove a test strip from the test strip vial (or the foil pouch). Tightly close the

    vial cap immediately after you have removed the test strip.

    3. Insert the test strip into the meter in the direction of the arrows. Meter turns

    on after a beep.

    4. A symbol with a test strip with a flashing blood () will appear

    letting you know the meter is ready to test.

    Note:

    Check the expiration and discard dates on the test strip vial. All expiration dates

    are printed in YearMonth format. 2019/01 indicates January, 2019. Your

    VivaChek™ Ino test strips have 4 months shelf life after you first open the test strip

    vial. Write the discard date on the vial label when you first open it. Make sure the

    test strip does not appear damaged. Prior to testing, wipe the test site with an

    alcohol swab or soapy water. Use warm water wash hands to increase blood flow if

    necessary. Then dry your hands and the test site thoroughly. Make sure there is no

    cream or lotion on the test site.

    9

    Getting a Blood Drop and Testing

    Here is an example of how to use a VivaChek™ Safety Lancet for fingertip

    blood sampling. If you use a different singleuse autodisabling safety lancet

    or lancing device, please follow the manufacturer’s instructions.

    1. Carefully rotate and pull off the protective cap.

    Avoid placing your thumb or fingers over the

    uncapped end of the safety lancet.

    2. Hold the safety lancet against the puncture site.

    Note: Before using a singleuse autodisabling

    safety lancet for blood sampling, wash both hands

    with soap and warm water and disinfect the puncture

    site with a topical skin antiseptic such as an alcohol

    swab.

    3. Gently press the safety lancet against the puncture

    site to lance the skin. Listen for an audible click. This

    indicates that the safety feature of the device has

    been activated. Dispose of the lancet in an

    appropriate biohazard sharps container. Please see

    the Caution Statement.

    4. Gently massage from the surrounding area toward

    the puncture site to collect the required blood

    volume.

    For the greatest reduction in pain, lance on the sides

    of the fingertips. Rotation of sites is recommended. Repeated punctures in

    the same spot can make fingers sore and callused.

    Caution Statement

    Do not use the singleuse autodisabling safety lancet if the cap is

    missing or loose.

    Always use a new, sterile single-use autodisabling safety lancet. Do

    not reuse the singleuse autodisabling safety lancet.

    Avoid getting the singleuse autodisabling safety lancet dirty with hand

    lotion, oil, dirt or debris.

    Singleuse autodisabling safety lancets appropriate biohazard sharps

    container. If a biohazard sharps container is not available use a heavy

    duty plastic container such as a laundry detergent bottle. Ensure the

    container is leakresistant and has a hard puncture proof lid. Do not

    place loose sharps into the trash and do not flush down the toilet. Do

    not place sharps into the recycle bin as they are not recyclable.

    10

    5. Immediately touch the tip of the test strip to the drop of blood. The blood

    pulled into the test strip through the tip. Make sure that the blood sample

    has been fully filled the check window on the strip tip. Hold the tip of the test

    strip in the blood drop until the meter beeps.

    Note:

    If the blood sample does not fill the check window, do not add a second drop.

    Discard the test strip and start over with a new test strip.

    6. The meter counts down 5 seconds and your result appears on the display

    after a beep. The test result will automatically be stored in the meter

    memory. Please do not touch the test strip during the countdown as this

    may result in an error.

    Discard the Used Test Strip

    You can discard the used test strip by slide the strip ejector to the direction of arrow.

    Meter turns off automatically after a beep.

    Potential Biohazard

    Dispose of the used test strips as medical waste.

    Alternative Site Testing

    The forearm and palm areas have less nerve endings than the fingertip. Because

    of this, you may find that obtaining blood from these sites is less painful than from

    the fingertip. The procedure for forearm and palm sampling is different. You need

    the lancing device with clear cap to draw blood from these sites. The clear cap is

    not adjustable for puncture depth. And our starter kit is not included the lancing

    device, and please contact Customer Support at 8009518430 for information

    about purchasing the lancing device. Please contact your health care professional

    if you need help.

    Blood samples for glucose testing may be taken from sites other than your

    fingertips. Alternative site testing using blood from the forearm or palm may give

    glucose results that significantly differ from fingertip blood. Differences occur when

    11

    blood glucose levels are changing rapidly, such as after a meal, after insulin, and

    during or after exercise.

    The lancing device with clear cap is not included in the kit. Please purchase it

    separately if you need. Please refer to the lancing device insert which you

    purchase for AST testing.

    1. Unscrew the lancing device cover from the body of the lancing device. Insert a

    sterile lancet into the lancing device and push it until the lancet is fully inserted.

    2. Hold the lancet firmly in the lancing device and twist the safety tab of the lancet

    until it is loose, then pull the safety tab off the lancet. Save the safety tab for

    disposing of the used lancet.

    3. Screw the clear cap onto the lancing device.

    4. Choose a puncture site on the forearm or palm. Select a soft and fleshy area of

    the forearm or palm that is clean and dry, away from bone, and free of visible

    veins and hair. To bring fresh blood to the surface of the puncture site, massage

    the puncture site vigorously for a few seconds until you feel it getting warm.

    12

    5. Place the lancing device against the puncture site. Press and hold the clear

    cap against the puncture site for a few seconds. Press the release button of

    the lancing device, but do not immediately lift the lancing device from the

    puncture site. Continue to hold the lancing device against the puncture site

    until you can confirm a sufficient blood sample has formed.

    Notes:

    Consult your doctor or health care professional to determine if alternative site

    testing is right for you.

    Alternative site testing is not recommended if you have hypoglycemic

    unawareness (you do not recognize the symptoms of or cannot tell when you

    have low blood glucose). Please consult with your doctor or health care

    professional if you have a low blood glucose level.

    Select a soft, fleshy area of skin that is free from hair, moles and visible veins

    for alternative site testing. Wash the site with soap and warm water, then rinse

    and dry thoroughly.

    Use alternative site testing for blood glucose tests only when it is more than 2

    hours after: ·A meal ·Taking medication · Exercise

    IMPORTANT: There are important differences among forearm, palm and fingertip

    samples that you should know. Important Information about forearm and palm

    glucose testing:

    You should consult your doctor before choosing to perform forearm or palm

    testing.

    When blood levels are changing rapidly, such as after a meal, insulin dose or

    exercise, blood from the fingertips may show these changes more rapidly than

    blood from other areas.

    You should use fingertips if testing is within 2 hours of a meal, insulin dose or

    exercise. You should also use fingertip testing any time you feel glucose levels

    are changing rapidly.

    You should test with the fingertips anytime there is a concern for hypoglycemia

    or you suffer from hypoglycemia unawareness.

    13

    Expected Diabetes Control Goal:

    Blood glucose values will vary depending on food intake, medication dosages,

    health, stress, or exercise. Ideally, to control the glucose level as close to a normal

    (nondiabetic) blood glucose level as you safely can.

    Expected blood glucose levels for people without diabetes:1

    Time Range, mg/dL Range, mmol/L

    Fasting and Before Meals 70100 3.9 5.6

    2 hours after meals Less than 140 Less than 7.8

    Reference:

    1. ADA Clinical Practice Recommendations, 2015.

    Warning:

    If your blood glucose reading is under 50 mg/dL or you see LO (less than

    20 mg/dL) on the meter display, contact your health care professional as

    soon as possible.

    If you test result is above 250 mg/dL or you see HI (greater than 600 mg/dL)

    on the meter display, contact your health care professional as soon as

    possible.

    Do not change your medication therapy based on VivaChek™ Ino Sound

    test result before consult your health care provider.

    If you want to mark your result (only in Advanced Mode (L-2)) you will need

    to do it before removing the test strip.

    Questionable or Inconsistent Results:

    If your blood glucose result does not match how you feel, please:

    Check the expiration date and the discard date of the test strip. Make sure

    that the test strip vial has not been opened for more than 4 months.

    Confirm the temperature in which you are testing is between 41113°F.

    Make sure that the test strip vial has been tightly capped.

    Make sure the test strip has been stored at 3686°F, 1090% humidity..

    Make sure the test strip was used immediately after removing from the test

    strip vial (or the foil pouch).

    Make sure that you followed the test procedure correctly.

    Perform a control solution test (See Performing a Control Test for

    instructions).

    After checking all of the conditions listed above, repeat the test with a new test strip.

    If you are still unsure of the problem, please contact VivaChek Customer Support

    at 8009518430 (7 days a week. 9 a.m.5 p.m. Eastern Time). Please contact your

    health care professional if you need help.

    14

    Testing with Control Solution

    Why Perform Control Tests

    Performing a control test lets you know that your meter and test strips are working

    properly to give reliable test results. You should perform a control test when:

    At least once a week

    You open a new box of test strips

    You want to check the meter and test strips

    Your test strips were stored in extreme temperature or humidity

    After cleaning your meter

    You dropped the meter

    Your test result does not match with how you feel

    About the Control Solutions

    Only use VivaChek™ Ino Control Solutions (1, 2 or 3) to practice on the

    system.

    Your meter automatically recognizes the control solution.

    The control solution results are not including in the average value

    calculation.

    Store the control solution at 3686°F, 1090% humidity.

    All expiration dates are printed in YearMonth format. 2019/01 indicates

    January, 2019.

    Do not use control solution that is past the expiration date or discard date

    (the control solution will expire 4 months after the bottle is opened for the

    first time).

    Shake the bottle well before use.

    Close the bottle tightly after use.

    Performing a Control Test

    1. Remove test strip from the test strip vial (or the foil pouch). Tightly close the

    vial cap immediately after you have removed the test strip.

    Note: Check the expiration and discard dates of the test strips. Do not use the

    expired test strip.

    2. Insert a test strip into the meter in the direction of the arrows.

    3. The meter turns on after a beep. An image of a test strip with a flashing blood

    drop () will appear letting you know the meter is ready to

    test.

    15

    4. Shake the control solution bottle thoroughly. Squeeze the control solution

    bottle gently and discard the first drop. Squeeze out a second small drop on

    a clean nonabsorbent surface.

    Note: Do not apply control solution to the test strip directly from the bottle.

    5. Immediately touch the tip of the test strip to the drop of control solution. The

    control solution is pulled into the test strip through the strip tip.

    Note:

    If the control solution sample does not fill the check window, do not add a second

    drop. Discard the test strip and start over with a new test strip.

    6. Hold it in the drop until the meter beeps, and then you see the meter count

    down on the screen and followed with your control test result after a beep.

    Note:

    The meter will automatically recognize and mark the control result for you. Control

    results are not included in the 7, 14 and 30 day average calculation.

    Understand Your Control Test Result

    Compare your control test result with the ranges printed on the test strip vial label.

    16

    Notes:

    If your control test result is out of range:

    Check the expiration dates and discard dates of the test strip and control solution.

    Make sure that the test strip vial and the control solution bottle have not been

    opened for more than 4 months. Discard any expired test strips or control solution.

    Confirm the temperature in which you are testing is between 59104°F.

    Make sure that you stored strip and control solution at 3686°F, 1090%

    humidity.

    Make sure that the test strip vial and the control solution bottle have been tightly

    capped.

    Make sure the test strip was used immediately after removing from the test strip

    vial (or the foil pouch).

    Make sure the control solution was mixed well.

    Confirm that you are using VivaChek™ Ino brand control solution.

    Make sure that you followed the test procedure correctly.

    After checking all of the conditions listed above, repeat the control solution test with

    a new test strip. If your results still fall out of the range indicated on the test strip

    vial label, your meter or test strips may not be working properly. DO NOT use the

    system to test blood. Contact VivaChek Customer Support at 8009518430 (7

    days a week. 9 a.m.5 p.m. Eastern Time) for help. Please contact your health care

    professional if you need help.

    To turn your meter off, just remove the test strip. Dispose of the used test strips as

    medical waste. The result will be automatically marked and stored in the meter

    memory. Control results will be not included in your blood glucose averages.

    Using the Meter Memory

    Your meter automatically stores up to 1000 results with the time and date. Test

    results are stored from the newest to the oldest.

    Notes:

    If there are already 1000 records in memory, the oldest record will be erased

    to make room for a new one.

    It is very important to set the correct time and date in the meter, please

    make sure the time and date are correct after you change your batteries.

    Viewing Your Test Results

    When your meter is off, press to turn the meter on. After a beep, a symbol of a

    test strip flashes on the display. Continue to press to review previous results in

    order. Results will be shown starting with the most recent. Each result will show the

    date and time the test was taken.

    When END appears on the display, you have viewed all of the results in the

    memory.

    17

    CHARTER 4MAINTENANCE AND TROUBLESHOOTING

    Proper maintenance is recommended for best results.

    Replace the Battery

    When the meter needs to be replace, the Battery symbol ( ) will appear.

    1 Turn off your meter before changing the batteries.

    2 Press firmly on the battery cover and slide.

    Note:

    After you change the batteries, your meter prompts you to confirm the meter’s time

    and date settings. All the test results are saved in the memory.

    3 Lift out the old batteries.

    4 Place the new batteries under the prongs and into the battery compartment.

    5 Slide the battery cover back into place, lining up with the open slots, and close

    firmly.

    Warning:

    Keep batteries away from children. The batteries are poisonous. If swallowed,

    immediately contact your doctor or poison control centre. Discard batteries

    according to your local environmental regulations.

    Caring for Your Glucose Monitoring System

    Store meter in the carrying case provided whenever possible.

    Wash and dry hands well before handling to keep the meter and test strips

    free of water and other contaminants.

    VivaChek™ Ino Sound Blood Glucose Meter is a precision electronic

    instrument. Please handle it with care.

    Avoid exposing meter and test strips to excessive humidity, heat, cold, dust,

    or dirt.

    Cleaning and Disinfection

    Your VivaChek™ Ino Sound Blood Glucose Meter should be cleaned and

    disinfected a minimum of once per week. Use only CloroxTM Healthcare Bleach

    Germicidal Wipes, which has been proven to be safe to use with the VivaChek™

    Ino Sound Blood Glucose Meter.

    18

    Cleaning is part of your normal care and maintenance and should be performed

    prior to disinfection, but cleaning does not kill germs. After use and exposure to

    blood, all parts of this kit can potentially transmit infectious diseases. Disinfecting

    reduces the risk of transmitting infectious diseases.

    Cleaning your meter

    The meter should be cleaned whenever they are visibly dirty or a minimum of once

    per week. For cleaning, first use one piece of Clorox™ Healthcare Bleach

    Germicidal Wipes (EPA Registration No. 6761912) to clean the entire meter

    surface for 1 minute. This precleaning is to prepare the meter surface for a

    disinfection process.

    Disinfecting your meter

    The meter should be disinfected a minimum of once per week. The meter

    disinfection process has been validated through repeated disinfection cycles,

    which is equivalent to 5 years of meter usage based on once daily cleaning and

    disinfection processes. This is to ensure that your meter will operate properly, even

    with the disinfection process applied to it for 5 years.

    After cleaning your meter, use one piece of Clorox™ Healthcare Bleach Germicidal

    Wipes (EPA Registration No. 6761912) to wipe the meter. The wipe will be wiped

    using 2 pass, where one pass equals a back and forth motion for a total of 4

    motions. And then, please expose the meter to air dry for at least 2 minute as

    specified on the disinfectant wipe manufacturers label.

    Disinfection is performed using one wipe to ensure that the treated the surface of

    the meter remained visibly wet during the contact time.

    CloroxTM Healthcare Bleach Germicidal Wipes containing Sodium hypochlorite

    0.55%, which has been proven to be safe to use with the VivaChek™ Ino Sound

    system. CloroxTM Healthcare Bleach Germicidal Wipes are available through

    CLOROX PROFESSIONAL PRODUCTS CO. You can purchase them by visiting

    their website at www.cloroxprofessional.com. CloroxTM Healthcare Bleach

    Germicidal Wipes are also available at internet retailers such as www.amazon.com,

    or for more information call VivaChek Customer Support at 8009518430

    (available 7 days a week, 8 a.m.5 p.m. Eastern Time). Please contact your health

    care professional if you need help.

    Wash hands thoroughly with soap and water after disinfecting procedure.

    Notes:

    Do not use alcohol or any other solvent.

    Do not allow liquid, dirt, dust, blood, or control solution to enter the test strip

    port or the data port.

    Do not squeeze the wipe or gauze into test strip port.

    19

    Do not spray cleaning solution on the meter.

    Do not immerse the meter in any liquid.

    Notes:

    Although it has not been observed, some alterations may appear on your meter

    due to the cleaning and disinfection procedure. Such as: cloudy display window,

    plastic housing cracking, meter buttons do not function, partial display on full

    screen, unable to execute the meter’s initial set up, etc. Stop using the meter and

    please contact Customer Support for help.

    If you have questions about cleaning or disinfecting, or if you see evidence of

    physical damage, contact VivaChek Customer Support at 8009518430 (available

    7 days a week, 8 a.m. 5 p.m. Eastern Time). Please contact your health care

    professional if you need help.

    Troubleshooting Guide

    E1

    Blood or control solution

    was applied to the test strip

    before the flashing drop

    appeared on the display

    Discard the test strip and repeat

    the test with a new test strip. Wait

    until you see the flashing blood

    drop on the display before testing.

    E2

    used or contaminated test

    strip.

    Discard the test strip and repeat

    the test with a new test strip. Wait

    until you see the flashing blood

    drop on the display before testing.

    E3

    Discard the test strip and repeat

    the test with a new test strip. Make

    sure that you are using a

    VivaChek™ Ino test strip from

    VivaChek Laboratories, Inc..

    E4

    Incorrect sample. Discard the test strip and repeat

    the test with a new test strip. Make

    sure that only human capillary

    blood and VivaChek™ Ino control

    solution can be used for the test.

    E5

    Temperature out of range.

    Move to an area that is within the

    operating range for the meter. Let

    temperature for 20 minutes before

    20

    E6,E7

    hardware issue. Take out batteries and restart the

    meter. If the problem continues,

    contact VivaChek Customer Support

    at 8009518430 (available 7 days a

    week, 8 a.m.5 p.m. Eastern Time).

    Please contact your health care

    professional if you need help.

    E10 Insufficient sample.

    Repeat test and apply enough

    sample to fill the test strip check

    window.

    Indicates a low battery. Replace new batteries.

    Hi

    600 mg/dL.

    Wash and dry your hands well and

    the test site. Repeat the test using

    a new test strip. If your result still

    flashes HI, contact your health

    care professional as soon as

    possible.

    Caution:

    Glucose levels above 250 mg/dL(13.9 mmol/L) may indicate a potential serious

    Lo

    45 mg/dL.

    Repeat the test using a new test

    strip. If your result still flashes LO,

    professional as soon as possible.

    Caution:

    Glucose levels below 50 mg/dL (2.8 mmol/L) may indicate a potential serious

    21

    Symptoms of High or Low Blood Glucose

    You can better understand your test results by being aware of the symptoms of

    high or low blood glucose. According to the American Diabetes Association, some

    of the most common symptoms are:

    Warning:

    If you are experiencing any of these symptoms, test your blood glucose. If your test

    result is under 50 mg/dL or above 250 mg/dL, contact your health care professional

    immediately.

    Low blood glucose

    (Hypoglycemia):

    shakiness

    sweating

    fast heartbeat

    blurred vision

    confusion

    passing out

    irritability

    seizure

    extreme hunger

    dizziness

    High blood glucose

    (Hyperglycemia):

    frequent urination

    excessive thirst

    blurred vision

    increased fatigue

    Ketones (ketoacidosis):

    shortness of breath

    nausea or vomiting

    very dry mouth

    22

    CHAPTER 5: TECHNICAL INFORMATION

    System Specifications

    Feature Specification

    Measurement range 45 to 600 mg/dL

    Result calibration Plasma-equivalent

    Sample Fresh capillary whole blood

    Sample volume: 0.8 µL

    Test time 5 seconds

    Power source Two AAA LR03 1.5V batteries

    Glucose units of measure The meter is pre-set to milligrams per deciliter (mg/dL)

    Memory Up to 1000 records with date and time

    Automatic shutoff 2 minutes after last action

    Dimensions 89 mm x 52 mm x 25 mm

    Display size 32 mm x 32 mm

    Weight Approximately 63g

    Operating temperature 41113°F

    Operating relative humidity 10-90% (noncondensing)

    Hematocrit range 2070%

    Data port Micro USB

    Bluetooth Version 4.1

    23

    Warranty

    Please complete the warranty card that came with this product and mail it to

    VivaChek Laboratories, Inc.

    913 N market Street

    Wilmington, De, 19801, USA

    If the meter fails to work for any reason other than obvious abuse within the first

    five (5) years from purchase, we will replace it with a new meter free of charge. For

    your records, also write the purchase date of your product here.

    Date of purchase: ________________________________________________

    Note:

    This warranty applies only to the meter in the original purchase, and does not apply

    to the batteries supplied with the meter.

    Описание

    Цена по запросу

    Передовые технологии

    Расширенные возможности

     8-электродная тест полоска

     возможность использования у новорожденных, беременных, пациентов с анемией и других

     5 сек для получения результата

    достаточно 0,5 мкл образца крови

    Особенности:

    — Хранение 900 результатов с отображением времени и даты
    — 5 видов напоминаний от системы мониторирования
    — Сигналы о кетоацидозе и гипогликемии снизят риски крайних значений глюкозы крови
    — Micro USB  порт позволяет перенести данные в персональный компьютер
    — Доступны индивидуальные упаковки из фольги, которые обеспечивают гигиену и больший срок службы тест полосок (до 24 мес)
    — Автовыключение после 2-х минут бездействия
    — Производит поиск вероятных причин поврежедния тест полоски
    — 8-электродная тест-полоска устраняет влияние температуры и влажности воздуха окружающей среды на результат
    — Возможность использования при очень низких значениях гематокрита
    — Автокодировка

    Производитель: Китай

    Характеристики

    Диапазон измерений 0,6-33,3 ммоль/л
    Тип образца свежая цельная капиллярная кровь
    Фермент оксидаза глюкозы
    Калибровка результата по плазме
    Образец приблизительно 0,5 мкл
    Время получения результата около 5 сек
    Гематокрит 20-70%
    Температурный режим от 5°C и до 45°C
    Допустимая влажность воздуха от 10% до 90% (без конденсации)
    Батарея 2 шт
    Длительность использования батареи свыше 1000  измерений
    Память 900 результатов с отображением времени и даты
    Порт для переноса данных Micro USB

    Понравилась статья? Поделить с друзьями:

    Новое и полезное:

  • Vitumnus магний в6 раствор инструкция
  • Vitotronic 333 инструкция по эксплуатации
  • Vitotronic 200 typ kw5 инструкция
  • Vitotronic 200 kw5 инструкция по эксплуатации
  • Vitotronic 200 kw4 инструкция по эксплуатации для специалистов

  • 0 0 голоса
    Рейтинг статьи
    Подписаться
    Уведомить о
    guest

    0 комментариев
    Старые
    Новые Популярные
    Межтекстовые Отзывы
    Посмотреть все комментарии